Dynasonics TFXL Specifications

Transit Time Flow Meters
TFXL Ultrasonic
DESCRIPTION
The TFXL ultrasonic transit time flow meter measures most clean liquids and liquids with small amounts of suspended solids or aeration, such as surface water or sewage.
Bi-directional flow measurement system
Totalizer options include forward, reverse and net total
UltraLink® software utility enables in-field flow meter configuration, calibration and troubleshooting, via laptop PC
Compact enclosure uses large, easy-to-read digital display
BENEFITS
Reduced material costs: Clamp-on sensor eliminates the need for in-line flanges, pipe fittings, strainers and filters.
Reduced installation time: May be installed and fully operational within minutes. No need to break into pipelines.
Reduced maintenance costs: With no moving parts, there is nothing on the TFXL to wear down—no repair kits or replacement parts are needed.
Easy retrofit: With three standard outputs (4…20 mA, TTL pulse and simulated turbine frequency), the TFXL drops easily into existing DCS and flow monitoring systems.
Reduced downtime: Installation may be performed on full pipes. No need to shut the process down for installation or maintenance.
APPLICATION
The TFXL ultrasonic flow meter clamps onto the outside of a pipe and does not contact the internal liquid. This advanced product provides instantaneous rate and accumulated flows along with 4…20 mA and pulse outputs. Compact integral mount systems can accommodate pipes/tubing two inches and smaller. Remote mount systems are also available for pipe/tubing sizes 1/2 in. (DN 15) and higher.
OPERATION
Transit time flow meters use two transducers that function as both ultrasonic transmitters and receivers. The flow meters operate by alternately transmitting and receiving a frequency-modulated burst of sound energy between the two transducers. The burst is first transmitted in the direction of fluid flow and then against fluid flow. Since sound energy in a moving liquid is carried faster when it travels in the direction of fluid flow (downstream) than it does when it travels against fluid flow (upstream), a differential in the times of flight will occur. The sound’s time-of-flight is accurately measured in both directions and the difference in time-of-flight calculated.
TTM-DS-00960-EN-07 (April 2016)
Product Data Sheet
Specications
SPECIFICATIONS
System
Liquid Types Most clean liquids or liquids containing small amounts of suspended solids or gas bubbles
Velocity Range 0.1…40 FPS (0.03…012 MPS)
Flow Accuracy
Ambient Temperature
Repeatability ±0.5% of reading
Transducer Type Clamp-on, uses time-of-flight ultrasonics
Protection Reverse polarity, surge suppression
Certifications
DTTR/DTTN/DTTH DTTS/DTTC
General purpose Hazardous locations integral mount Hazardous locations DTTN
Remote mount transmitter and integral mount transmitter with transducers
Hazardous location transducers (DTTN with I.S. option)
±1% of reading or ±0.01 FPS (0.003 MPS), whichever is greater 1 in. (25 mm) and larger = ±1 % above 1 FPS (0.3 MPS) and ±0.01 FPS below 1 FPS 3/4 in. (19 mm) and smaller = ±1% of full scale
–40…185° F (–40…85° C) 0…105° F (–17…40° C) –40…185° F (–40…85° C)
General purpose standards: UL 61010-1 and CSA C22.2 No. 61010-1
Hazardous location designation and standards: Class I, Division 2, Groups C and D UL1604, CSA C22.2 No. 213
Hazardous location designation and standards: Class I, Division 1, Groups C and D, T5 UL913:2002, UL916 CAN/CSA C22.2 No. 0-10, C22.2 No. 142-M1987, C22.2 No. 157-92 Install with I.S. barrier D070-1010-002
Transmitter
Power Requirements
Display
Enclosure Rating
Units of Measure
Outputs
12…28V DC @ 0.25A
Type Rate Total
NEMA Type 3 (Type 3) ABS, PVC and Ultem (integral system), brass or SS hardware
Engineering units
Rate
Analog and TTL Frequency (Output option 1)
Totalizer pulse (Output option 3)
2 line x 8 character LCD 8 maximum digits with lead zero blanking 8 maximum digits with exponential multipliers from –1…6
Feet, US gallons, cubic feet, million gallons, barrels (liquid and oil), acre-feet, lbs, meters, cubic meters, liters, million liters, kilograms Second, minute, hour, day
4…20 mA: 900 ohms max, internally powered, 12-bit resolution
Selectable turbine meter simulation or square wave 0…1000 Hz, duty cycle 50% +/- 10% Square wave: 5V DC Turbine meter simulation: 500 mVpp minimum
Source or sink, 5V DC, 2 mA maximum, pulse duration 30 ms, external resistor Normal state high with pulses low
Page 2 April 2016TTM-DS-00960-EN-07
Transducers
?
ULTRALINK Software Utility
PBT glass filled, Ultem, Nylon cord grip PVC cable jacket; –40…250° F (–40…121° C)
CPVC, Ultem, Nylon cord grip Polyethylene cable jacket; –40…185° F (–40…85° C)
CPVC, Ultem, Nylon cord grip Polyethylene cable jacket; –40…185° F (–40…85° C)
PTFE, Vespel, Nickel-plated brass cord grip PFA cable jacket; –40…350° F (–40…176° C)
PVC, Ultem, Nylon cord grip PVC cable jacket; –40…140° F (–40…60° C)
Transducer Construction
DTTR NEMA 6*/IP67
DTTC NEMA 6*/IP67
DTTN I.S. NEMA 6P*/IP68
DTTH NEMA 6*/IP67
DTTS NEMA 6*/IP67
*NEMA 6 units: to a depth of 3 ft (1 m) for 30 days max. NEMA 6P units: to a depth of 100 ft (30 m) seawater equivalent density indefinitely.
Cable Length 990 ft (300 meter) max. in 10 ft (3 m) increments; Submersible Conduit limited to 100 ft (30 m)
Pipe/Tubing Sizes 1/2 in. (12 mm) and larger
Pipe/Tubing Materials Carbon steel, stainless steel, copper and plastic
Software Utilities
ULTRALINK
Used to configure, calibrate and troubleshoot flow meter. Software is compatible with Windows 2000, Windows XP, Windows Vista and Windows 7
ULTRALINK SOFTWARE UTILITY
The flow meter must be programmed with the UltraLink software utility. The software is used to configure, calibrate and communicate with TFXL flow meters. Additionally, it has numerous troubleshooting tools to make diagnosing and correcting installation problems easier.
UltraLINK Device Addr 127
U
Conguration CalibrationStrategy
Device Addr 127
U
135 Gal/Min
Flow:
237 Gal
Totalizer Net:
Pos:
237 Gal
Neg:
0 Gal
Sig. Strength:
15.6%
Margin:
100%
Delta T:
2.50 ns
Last Update:
12:17:20
Signal Strength too Low!
Reset Totalizers
Data Display Diagnostics
HelpWindowCommunicationsViewEditFile
!
Print About
Errors
60 Min
2000
1600
1200
800
400
0
Flow Rate
-400
-800
-1200
-1600
-2000
-1.00:00
?
Stop
Step View
Go
Stop
Print Preview
2000
Scale:Time:
-50:00 -40:00 -30:00 -20:00 -10:00 -0:00
Stop
Historical Data
Time (mm:ss)
ADDITIONAL PARTS REQUIRED FOR CONFIGURATION
Part Number Description
D010-0204-001 Configuration cable kit
D005-2116-004 USB-to-DB9 converter (required if PC does not have a serial port)
13:26:33
Exit
OK
COMM:For Help, press F1
Page 3 April 2016 TTM-DS-00960-EN-07
Additional Parts Required for Conguration
DIMENSIONS
Enclosures
Integral Enclosure Remote Enclosure
A
A
in. (mm)B in. (mm)C in. (mm)
B
E
C
D
Conduit hole
B
H
F
G
D DIA
in. (mm)E in. (mm)F in. (mm)G in. (mm)H in. (mm)I in. (mm)
K
A
J
Two mounting holes
I
C
D
Conduit hole
E
F
J DIA
in. (mm)K in. (mm)
6.72 (170.7) 3.17 (80.5) 2.57 (65.3) 0.87 (22.2) 1.33 (33.8) 0.85 (21.6) 0.77 (19.6) 1.78 (45.2) 3.74 (95) 0.22 (5.6) 7.01 (178)
Transducers
Remote System with Large Pipes
DTTR DTTN/DTTH
(Min Clearance)
C
B
D
TOP VIEW
OF PIPE
A
DTTR DTTN DTTH
A 3.75 in. (95 mm) 2.95 in. (74.9 mm) 2.95 in. (74.9 mm)
B 2.35 in. (60 mm) 2.75 in. (69.8 mm) 2.75 in. (69.8 mm)
C 3.00 in. (76.2 mm) 3.00 in. (76.2 mm)
D 2.19 in. (56 mm) 1.70 in. (43.2 mm) 1.71 in. (43.4 mm)
D
Page 4 April 2016TTM-DS-00960-EN-07
Remote System with Small Pipes
A
C
Additional Parts Required for Conguration
Integral System
DTTS/DTTC
Pipes and Tubing 1/2…2 in. (12…50 mm)
B
D
A
C
DTTS/DTTC DTTS/DTTC U-Bolt Connections
B
C
D
A
DTTS/DTTC U-Bolt Connections
ANSI/DS and Copper 2 in. (50 mm) Models
D
B
C
B
D
A
Pipe
Size
1/2 in.
3/4 in.
1 in.
1-1/4 in.
1-1/2 in.
2 in.
* Varies due to U-bolt conguration
Pipe
Material
ANSI/DN 2.46 in. (62.5 mm) 2.36 in. (59.9 mm) 2.66 in. (67.6 mm) 0.84 in. (21.3 mm) 2.0…38 gpm (8…144 lpm)
Copper 2.46 in. (62.5 mm) 2.36 in. (59.9 mm) 3.33 in. (84.6 mm) 0.63 in. (15.9 mm) 1.8…27 gpm (7…102 lpm)
Tubing 2.46 in. (62.5 mm) 2.28 in. (57.9 mm) 3.72 in. (94.5 mm) 0.50 in. (12.7 mm) 1.5…18 gpm (6…68 lpm)
ANSI/DN 2.46 in. (62.5 mm) 2.57 in. (65.3 mm) 2.66 in. (67.6 mm) 1.05 in. (26.7 mm) 2.75…66 gpm (10…250 lpm)
Copper 2.46 in. (62.5 mm) 2.50 in. (63.5 mm) 3.56 in. (90.4 mm) 0.88 in. (22.2 mm) 2.5…54 gpm (10…204 lpm)
Tubing 2.46 in. (62.5 mm) 2.50 in. (63.5 mm) 3.56 in. (90.4 mm) 0.75 in. (19.0 mm) 2.5…45 gpm (10…170 lpm)
ANSI/DN 2.46 in. (62.5 mm) 2.92 in. (74.2 mm) 2.86 in. (72.6 mm) 1.32 in. (33.4 mm) 3.5…108 gpm (13…409 lpm)
Copper 2.46 in. (62.5 mm) 2.87 in. (72.9 mm) 3.80 in. (96.5 mm) 1.13 in. (28.6 mm) 3.5…95 gpm (13…320 lpm)
Tubing 2.46 in. (62.5 mm) 2.75 in. (69.9 mm) 3.80 in. (96.5 mm) 1.00 in. (25.4 mm) 3.5…85 gpm (13…320 lpm)
ANSI/DN 2.80 in. (71.0 mm) 3.18 in. (80.8 mm) 3.14 in. (79.8 mm) 1.66 in. (42.2 mm) 5.0…186 gpm (19…704 lpm)
Copper 2.46 in. (62.5 mm) 3.00 in. (76.2 mm) 4.04 in. (102.6 mm) 1.38 in. (34.9 mm) 4.5…152 gpm (17…575 lpm)
Tubing 2.46 in. (62.5 mm) 3.00 in. (76.2 mm) 4.04 in. (102.6 mm) 1.25 in. (31.8 mm) 4.0…136 gpm (15…514 lpm)
ANSI/DN 3.02 in. (76.7 mm) 3.40 in. (86.9 mm) 3.33 in. (84.6 mm) 1.90 in. (48.3 mm) 6.0…250 gpm (23…946 lpm)
Copper 2.71 in. (68.8 mm) 2.86 in. (72.6 mm) 4.28 in. (108.7 mm) 1.63 in. (41.3 mm) 5.0…215 gpm (19…814 lpm)
Tubing 2.71 in. (68.8 mm) 3.31 in. (84.1 mm) 4.28 in. (108.7 mm) 1.50 in. (38.1 mm) 5.0 …200 gpm (19…757 lpm)
ANSI/DN 3.70 in. (94.0 mm) 3.42 in. (86.9 mm)* 5.50 in. (139.7 mm) 2.38 in. (60.3 mm)* 8.0…420 gpm (30…1590 lpm)
Copper 3.70 in. (94.0 mm) 3.38 in. (85.9 mm)* 5.50 in. (139.7 mm) 2.13 in. (54.0 mm)* 8.0…375 gpm (30…1419 lpm)
Tubing 3.21 in. (81.5 mm) 3.85 in. (98.0 mm) 4.75 in. (120.7 mm) 2.00 in. (50.8 mm) 8.0…365 gpm (30…1381 lpm)
A B C D Measuring Range
Page 5 April 2016 TTM-DS-00960-EN-07
Part Number Construction
Y
PART NUMBER CONSTRUCTION
Dynasonics Ultrasonic Flow Meters
DTFXL - -
TFXL - Transit Time
Model
Transit Time Ultrasonic Flow Transmitter DTFXL
Display Options
ABS Enclosure - Blind (No Display) C/US 1
ABS Enclosure - With Rate and Total Display C/US 2
Pipe Size/Measurement Range
1/2 Inch ANSI Pipe A
3/4 Inch ANSI Pipe B
1 Inch ANSI Pipe C
1-1/4 Inch ANSI Pipe D
1-1/2 Inch ANSI Pipe E
2 Inch ANSI Pipe F
1/2 Inch Copper Tube G
3/4 Inch Copper Tube H
1 Inch Copper Tube I
1-1/4 Inch Copper Tube J
1-1/2 Inch Copper Tube K
2 Inch Copper Tube L
1/2 Inch O.D. Std Tube M
3/4 Inch O.D. Std Tube N
1 Inch O.D. Std Tube P
1-1/4 Inch O.D. Std Tube Q
1-1/2 Inch O.D. Std Tube R
2 Inch O.D. Std Tube S
Remote Mount | Use wit
Remote Mount | Use with DTTS/C
h DTTR/N/H
X
Connector Options
None - (Two) 1/2 inch Conduit Holes N
(Two) Water Tight Cable Clamps A
(Two) 1/2 Inch Flexible Conduit Connectors D
Output Options
4 … 20 mA and TTL Pulse 1
Totalizer Pulse 3
Reserved
None (Reserved) N
Options
None N
CPVC Transducer Material | Integral Mount Options Only
I.S. DTTN Transducer | Remote Option X Only
C
F
Page 6 April 2016TTM-DS-00960-EN-07
Part Number Construction—Remote Flow Transducers, Small Pipes 1/2…2 in. (15…50 mm)
PART NUMBER CONSTRUCTIONREMOTE FLOW TRANSDUCERS, SMALL PIPES 1/2…2 IN. 15…50 MM
DTT - -
Piping Environment
PVC -40. . .185° F (-40. . .85° c) S
CPVC -40. . . 140° F (-40. . . 60° C) C
Nominal Pipe Size
1/2 in. D
3/4 in. F
1 in. G
1-1/4 in. H
1-1/2 in. J
2 in. L
Pipe Type
ANSI Pipe P
Copper Pipe C
Tubing T
Cable Length
20 ft (6.1 m) 020
50 ft (15 m) 050
100 ft (30 m) 100
Conduit Type
None - (Bare Twinax Cable) N Flexible Armored (LiquidTite) A
Conduit Length
0 ft (0 m) 000
20 ft (6.1 m) 020
50 ft (15 m) 050
100 ft (30 m) 100
Page 7 April 2016 TTM-DS-00960-EN-07
Transit Time Flow Meters, TFXL Ultrasonic
General Purpose
Flexible Armored |Not available with high temperature DTTH transducer A
PART NUMBER CONSTRUCTIONREMOTE FLOW TRANSDUCERS, PIPES LARGER THAN 2 IN. 50 MM
DTT - - - N
Transmitter Type
Standard (1 MHz), 250° F (121° C) Max Temp. R High Temperature (1 MHz), 350° F (176° C) Max Temp. H
Cable Length
20 ft. (6.1 m) 020 50 ft. (15 m) 050 100 ft. (30 m) 100
Conduit Type
None N
Conduit Length
None 000 20 ft. (6.1 m) 020 50 ft. (15 m) 050 100 ft. (30 m) 100
General Purpose, Submersible (IP68)
DTT - - S 000 - N
Transmitter Type
Standard: 1 MHz N
Cable Length
20 ft. (6.1 m) 020 50 ft. (15 m) 050 100 ft. (30 m) 100
Hazardous Location (Class 1, Division 1, Groups C and D)
DTT N - - - F
Cable Length
20 ft. (6.1 m) 020 50 ft. (15 m) 050 100 ft. (30 m) 100
Conduit Type
None N Flexible Armored A
Conduit Length
None 000 20 ft. (6.1 m) 020 50 ft. (15 m) 050 100 ft. (30 m) 100
Control. Manage. Optimize.
DYNASONICS is a registered trademark of Badger Meter, Inc. Other trademarks appearing in this document are the property of their respective entities. Due to continuous research, product improvements and enhancements, Badger Meter reserves the right to change product or system specications without notice, except to the extent an outstanding contractual obligation exists. © 2016 Badger Meter, Inc. All rights reserved.
www.badgermeter.com
The Americas | Badger Meter | 4545 West Brown Deer Rd | PO Box 245036 | Milwaukee, WI 53224-9536 | 800-876-3837 | 414-355-0400 México | Badger Meter de las Americas, S.A. de C.V. | Pedro Luis Ogazón N°32 | Esq. Angelina N°24 | Colonia Guadalupe Inn | CP 01050 | México, DF | México | +52-55-5662-0882 Europe, Middle East and Africa | Badger Meter Europa GmbH | Nurtinger Str 76 | 72639 Neuen | Germany | +49-7025-9208-0 Europe, Middle East Branch Oce | Badger Meter Europe | PO Box 341442 | Dubai Silicon Oasis, Head Quarter Building, Wing C, Oce #C209 | Dubai / UAE | +971-4-371 2503 Czech Republic | Badger Meter Czech Republic s.r.o. | Maříkova 2082/26 | 621 00 Brno, Czech Republic | +420-5-41420411 Slovakia | Badger Meter Slovakia s.r.o. | Racianska 109/B | 831 02 Bratislava, Slovakia | +421-2-44 63 83 01 Asia Pacic | Badger Meter | 80 Marine Parade Rd | 21-06 Parkway Parade | Singapore 449269 | +65-63464836 China | Badger Meter | 7-1202 | 99 Hangzhong Road | Minhang District | Shanghai | China 201101 | +86-21-5763 5412 Legacy Document Numbers: 06-TTM-BR-00404-EN
Loading...