Dynasonics D601 User Manual

Series D601
Series D601
Series D601Series D601
Doppler Ultrasonic Flow Meter
Operations & Maintenance
REV 4/00
PART 1 - TABLE OF CONTENTS
Description Pages
Quick Start Installation 1.2-1.3 Operating Theory 1.4 Product Limitations 1.5 Model Number Matrix 1.6 Specifications 1.7
Transducer Installation 2.1-2.5
Pre-Installation Functional Test 3.1-3.2 Transmitter Installation 3.3-3.4 Power Source Jumper Settings 3.5 Electrical Connections 3.5-3.6
Power Up and Configuration 3.7 Keypad Operations 3.8-3.19 Trouble Shooting 3.20-3.21
Appendix Fluid Sound Speed Conversions Ductile Iron Pipe Data Cast Iron Pipe Data Steel, SS, PVC Pipe Data FPS to GPM Conversion Chart Intrinsic Safety Installation Statement of Warranty Service Information
Rev. 4/00 -1.1- D601
PART 1 - QUICK START
General
Transducer Cable
This manual contains detailed operating instructions for all aspects of the D601 instrument. The following condensed instructions are provided to assist the operator in getting the instrument started up and running as quickly as possible. This pertains to basic operation only. If specific instrument features are to be used or if the installer is unfamiliar with this type of instrument, refer to the appropriate section in the manual for complete details.
1. TRANSDUCER LOCATION
A. In general, select a mounting location on the piping
system with a minimum of 10 pipe diameters (10 X the pipe inside diameter) of straight pipe upstream and 5 straight diameters downsteam. See Table 2.1 for additional configurations.
B. On horizontal pipe, select a position that is between 2
and 4 o’clock on the pipe, with 12 o’clock representing the top. Installations on vertical pipe should be made in an area where the flow moves from bottom to top— ensuring a full pipe of liquid.
2. PIPE PREPARATION AND TRANSDUCER MOUNTING
A. The piping surface, where the transducers are to be
mounted, needs to be clean and dry. Remove loose scale, rust and paint to ensure satisfactory acoustical bonds.
B. Loosely wrap the appropriate length of strap around
FLOW
Figure 1.1
Top View of Pipe
the pipe at the location determined in Step 1. Refer to Figure 1.1 for proper orientation of the transducer. For greatest accuracy, point the cable of the transducer in the primary flow direction.
C. Apply a liberal amount of couplant onto the transducer
face. Place the transducer onto the pipe ensuring square and true placement. If an RTV type of couplant (requiring curing time) was utilized, allow sufficient time for curing before applying power to the instrument or moving the cable.
Rev. 4/00 -1.2- D601
PART 1 - QUICK START
3. TRANSDUCER/POWER CONNECTIONS
Figure 1.2
Transducer (top picture)
and Power (bottom picture)
Connections
A. Do not
transducers.
B. Refer to the DIAGRAMS in Figure 1.2 for proper power
and transducer connections. Verify proper jumper selections are in place for the power source. See Figure 3.4.
4. INITIAL SETTINGS AND POWER UP
attempt to add additional cable to the
IMPORTANT!
In order to successfully complete the configuration of the D601 flow meter, the transducer must be mounted on a pipe which is full of a flowing liquid. It is normal to have a zero reading and no signal strength indication with empty pipes or zero flow rate.
A. Adjust the GAIN control [R13] to 1/4 turn from full
counter-clockwise rotation.
B. Apply power to the instrument. C. If the pipe is full of a flowing liquid, the flow meter
signal strength will increase from a zero reading.(press the
2nd FUNCT
Signal Strength does not increase to a minimum of 000125 counts, gradually turn the GAIN contr ol [R13] clockwise until the indication is between 000125 and
000200. (Do not over adjust this setting as ambient noise can influence readings.)
key, then press
SIGNAL STR)
. If the
D. If possible, turn off the flow in the pipe. Verify that
SIGNAL STR. is lower than 000100. If SIGNAL STR. is greater than 000100, verify that the sensor/ transmitter are not located near electrically noisy components. (VFDs, inverters, motors, power relays, etc) Verify that DT6 transducer connections are proper and secure. If the SIGNAL STR. remains greater than 000100, consult the Dynasonics Factory for assistance. It is possible that the GAIN control [R13] is set too far clockwise and ambient noise is influencing the readings. Turn the control counter-clock wise until the signal strength decreases to below 000070 counts.
E. If the instrument passes steps 4C and 4D, the basic
setup of the instrument is complete.
Rev. 4/00 -1.3- D601
PART 1 - GENERAL
General
Operating Theory
The Dynasonics D601 flowmeter is designed to measure the flow of liquids and slurries in full-pipe closed systems. The transmitter is field configured to measure flow on a variety of pipes and liquids. The standard product is typically used on pipe sizes ranging from 1 - 120 inch [ 25 - 1524 mm ] pipe I.D. (With the small pipe transducer option, the pipe size range is 0.25 - 1 inch [ 6 - 25 mm]). A variety of liquid applications can be accommodated: sewage, sludges, concrete, mining slurries, dredging, etc. Because the transducers are non-contacting and have no moving parts, the flow meter is not affected by system pressure, fouling or wear. Standard transducers are rated to 250°F [121°C]. Optional high temperature transducers are rated to operate to 400°F [204°C].
The basic principle of operation is the measurement of the frequency shift “Doppler” of a reflected ultrasonic signal from discontinuity in the flowing liquid. In theory, these discontinuities can be virtually any amount of suspended bubbles, solids, or interfaces caused by turbulent flow. In practice the degree to which this can be reliably accomplished is a function of the sensitivity and frequency of the transducer and associated transmitter. The D601 design requires greater than 100 PPM of suspended solids or bubbles over 100 microns in size. Should your application be cleaner than this, please check with the factory for information on transit time ultrasonic flowmeters. The transducer which generates and receives the ultrasonic signal supplies the data to the transmitter. The transmitter processes the signal and provides an analog and pulse output for velocity indicating and volumetric totalizing. In addition, the transmitter contains a signal strength indicator which determines satisfactory operation.
Rev. 4/00 -1.4- D601
PART 1 - GENERAL
Measuring Limits
The flowmeter is typically used as a unidirectional meter and is most accurate when the transducer is mounted in the orientation detailed in this manual. But, the meter will measure flow in both directions — although flow direction will not be indicated or totalized properly. The flowmeter will operate from signals returned from turbulence alone (such as installation directly at pump discharges or downstream from elbows and valves); however, it should be noted that turbulence may vary with flow rates and result in non-linear results. The repeatability of the device is not dependent on most process liquids.
The flowmeter is designed to measure the flow of liquids and slurries, as long as a small, homogeneous quantity of entrained air or suspended solids are present. Without the presence of continuing supply of air or solids, the transmitted pulses are not reflected back to the transducer and the indicator will indicate zero flow.
The signal strength value will indicate a value grea ter than 100 counts when a minimum size and concentration of suspended particles are available for a reliable flow reading (100 micron and 100 PPM minimum) and the liquid is moving at least 0.1 FPS [0.03 MPS]. Most water-based liquids can be measured from a factory calibrated flowmeter. However, liquids with a heavy solids level (i.e. over 2% by volume), liquids with sound speeds that vary from water (see Appendix - Liquid Sound Speed list) or pipes with liners may have to be field calibrated. This is done by adjusting the Calibration value on the keypad to make the indicator agree with a known flow velocity or a mathematically corrected fluid velocity. All standard flow meters are calibrated to measure the flow of a water-based liquid at 25 degrees C. If the scale range or units need to changed, the process to do so is covered in detail later in this manual.
Rev. 4/00 -1.5- D601
Serial Number
PART 1 - GENERAL
The D601 employs modular construction and provides electrical safety for the operator. The display face contains voltages no greater than 9 Vdc and any exposed metal work is electrically connected to Earth Ground. The display face swings open to allow access to user connections.
The serial number and complete model number of your D601 is located on the inside of the transmitter front cover. Should technical assistance be required, please provide the Dynasoncs’ Customer Service Department with this information. Email: dynasonics_technical@racinefed.com.
A part number breakdown of the transducer and transmitter portions of the flowmeter are described below.
Transmitter
Transducer
Rev. 4/00 -1.6- D601
PART 1 - GENERAL
Rev. 4/00 -1.7- D601
PART 2 - TRANSDUCER INSTALLATION
Transducer Mounting Locations
The following list outlines how to install the DT6 transducer for optimal performance, h ighest reliability and greatest accuracy:
1. Select a transducer site at least 10 pipe diameters downstream from bends, or fittings and 5 pipe diameters upstream. A symmetrical flow pattern is necessary for accuracy and repeatability over the
Table 2.1
1
1
The D601 system will provide repeatable measurements on piping systems that do not meet these
requirements, but the accuracy
Rev. 4/00 -2.1- D601
of these readings may be influenced to various degrees.
PART 2 - TRANSDUCER INSTALLATION
Transducer Cable
FLOW
Figure 2.2
Top View of Pipe
operating range of the meter. Down stream from pump or orifices, etc., locate at least 20 diameters. See
Table 2.1
2. On horizontal pipe, select a position that is between 2 and 4 o’clock on the pipe, with 12 o’clock representing the top. If the transducer is to be mounted on a vertical pipe, select a section of pipe where the flow is moving from bottom to top (flow moving vertically down a pipe tends to cavitate and provide unreliable operation.)
3. Mount the transducer in the orientation shown in Figure 2.2. The flow meter will read flow in both directions, but will be most accurate if the cable is mounted in the orientation shown—pointing in the primary flow direction.
4. If totalization of the measured fluid is required, the pipe must remain full. The meter will read when the liquid level is greater than the placement of the transducer, but the volumetric measurement will be based on a full pipe, so totalization will be higher than actual.
5. The flowmeter will achieve proper Doppler signals off of turbulence; however, it should be noted that turbulence may not be linear with pump speed changes, nor is the reading necessarily accurate due to the non-uniformity of turbulence.
6. When a liquid has less than 100 PPM of 100 micron or larger particles, try mounting the transducer within 12 inches of a pump discharge or other source of flow turbulence or cavitation. A reading obtained under these circumstances will be repeatable, but not necessarily accurate or linear.
7. It is a good practice to test the flow meter on the piping system before permanently mounting the transducer using RTV. Function can be verified by applying a water soluble lubricant, such as KY­Jelly, and holding the transducer by hand on the pipe in the location where the transducer will be
Rev. 4/00 -2.2- D601
PART 2 - TRANSDUCER INSTALLATION
permanently mounted. Under flowing liquid conditions, adequate signal is indicated when the Signal Strength indicates between 000125 and 000200 counts.
Acoustic Couplant Types
Small Pipe Transducers
For proper operation, there cannot be air voids between the traducer face and pipe. The space must be filled with a material which is a good transmitter of sound energy such as:
SILICONE GREASE: Dow Corning 111 R or comparable (-100 to +450 F.) The material must be suitable not to flow at temperature of pipe. Used for temporary survey installations and portable flow meters.
SILICONE RUBBER: Dow Corning 732-RTV R. Excellent for permanent bonding. This adhesive is a recommended bonding agent and easily removable.
INSTALLATION AND PIPE PREPARATION
The cable from the DT6 transducer is provided with either dual-coaxial cables, flexible nylon conduit or PVC coated steel conduit with a 1/2” NPT fitting. The coaxial cable was ordered from the factory at a specific length
should the coaxial cable be lengthened as this may de-tune the circuitry and influence performance
Installation of the DT61 and DT63 small-pipe transducers follow the same procedures as the DT60 and DT62 standard pipe type. The only difference is that the small pipe transducers u tilize an integral pipe clamping mechanism with two opposing sensing heads and the standard pipe units use a stainless steel strap.
UNDER NO CIRCUMSTANCES
.
Rev. 4/00 -2.3- D601
PART 2 - TRANSDUCER INSTALLATION
Intrinsic Safety Installations
Installations requiring intrinsic safety should refer to the Appendix drawings covering these applications.
1. Pipe Preparation:
For permanent silicone adhesive mounting, after determining the transducer location, some attention must be given to the pipe condition. Before the transducer head is bonded to the pipe surface, an area slightly larger than the flat surface to the transducer head (black rectangle) must be cleaned to bare metal. This means the re moval of all paint rust, and scale. Some minor pipe pitting will not cause problems, as the acoustic couplant will take up the voids. In the case where plastic pipe is used, remove all paint and grease so that a smooth, dry surface is exposed.
2. Transducer Mounting:
The transducer center line is designed to mount parallel to the pipe center line. The groove in the transducer body will allow the 1/2” stainless steel strap that was enclosed with the meter to align the transducer properly on the pipe. transducer on bends, elbows or fittings. Every effort should be made to mount the transducer parallel to the axis of the pipe as well as flat on the pipe. The transducer cable should run in the “down-stream” direction of liquid flow. See
In horizontal pipe runs, mount the transducer between 2 and 4 o’clock from the top—12 o’clock position; prepare the pipe surface as described. Finish the surface with some emery paper and then wipe the surface with trichlorenthylene to thoroughly degrease the contact surface in a area slightly larger than the flat surface of the transducer.
For permanent mounting, use a good silicone based
Figure 2.2
DO NOT
mount the
.
Rev. 4/00 -2.4- D601
PART 2 - TRANSDUCER INSTALLATION
adhesive (Dow-732). Spread a bead of the adhesive on the flat surface of the transducer face, covering well. Now spread a bead to the prepared pipe surface and press the head lightly to the pipe. Let the adhesive flow enough to fill in all the area beneath the head. At the same time, clamp (clamp supplied) into place until the silicone has set up. Ta ping along the edges of the head will hold the adhesive in place. A pad of adhesive must be formed between the transducer face and the pipe. Ensure that no relative movement between the transducer and the pipe takes place during the setup time (about 24 hours). Clamp transducer only tight enough to hold it in place while the adhesive is curing. Tighten for mechanical strength only after 24 hours. Secure the conduit as well.
3. Temporary Mounting and Spot Checks:
For temporary mounting, clean pipe as described and use silicone grease as the acoustical coupling material, holding by hand for spot readings or with a strap clamp for indefinite periods.
Rev. 4/00 -2.5- D601
PART 3 - PRE-INSTALLATION CHECKOUT
Unpacking
Functional Test
Figure 3.1
After unpacking, it is recommended to save the shipping carton and packing materials in case the instrument is stored or re-shipped. Inspect the equipment and carton for damage. If there is evidence of shipping damage, notify the carrier immediately.
The D601 flow meter can be checked for basic functionality using the following Bench Test procedure. It is recommended that this operation be performed before permanently installing the system.
Procedure:
1. Open the D601 transmitter cover.
2. Connect the transducer cable connector to the terminal locations on the lower left corner of the D601 main circuit card. See Figure 3.1.
3. Connect supply power to the appropriate terminal locations on the upper right corner of the main circuit card. See Figure 3.2. Verify that the power supply selection jumpers are configured properly—See Fig- ure 3.4.
4. Apply power.
5. Verify that the display indicates 0.00 FPS (or 0.0 flow rate of any other unit). If the display does not register
0.0, then press the 2nd FUNCT key, then press SIGNAL STR. Verify that SIGNAL STR. is lower than
000100. If SIGNAL STR. is greater than 000100, verify that the sensor/transmitter are not located near electrically noisy components. (VFDs, inverters, motors, power relays, etc) Verify that DT6 transducer connections are proper and secure. If the SIGNAL STR. remains greater than 000100, consult the Dynasonics Factory for assistance.
6. Press the 2nd FUNCT key to enter SERVICE MODE.
Figure 3.2
Rev. 4/00 -3.1- D601
Press the SIGNAL STR key to display SIGNAL STR. XXXXXX.
7. Rub the face of the transducer lengthwise back and
PART 3 - PRE-INSTALLATION CHECKOUT
forth with your thumb using moderate pressure. The cycle time should be 1-2 times per second.
8. Verify that signal strength increase s with frequenc y of the rubbing. Typical increases will range from 20-30 counts.
9. Verify that signal strength decreases when rubbing ceases.
Bench Test is Complete
Rev. 4/00 -3.2- D601
Loading...
+ 33 hidden pages