Dynasonics D601 User Manual

Series D601
Series D601
Series D601Series D601
Doppler Ultrasonic Flow Meter
Operations & Maintenance
REV 4/00
PART 1 - TABLE OF CONTENTS
Description Pages
Quick Start Installation 1.2-1.3 Operating Theory 1.4 Product Limitations 1.5 Model Number Matrix 1.6 Specifications 1.7
Transducer Installation 2.1-2.5
Pre-Installation Functional Test 3.1-3.2 Transmitter Installation 3.3-3.4 Power Source Jumper Settings 3.5 Electrical Connections 3.5-3.6
Power Up and Configuration 3.7 Keypad Operations 3.8-3.19 Trouble Shooting 3.20-3.21
Appendix Fluid Sound Speed Conversions Ductile Iron Pipe Data Cast Iron Pipe Data Steel, SS, PVC Pipe Data FPS to GPM Conversion Chart Intrinsic Safety Installation Statement of Warranty Service Information
Rev. 4/00 -1.1- D601
PART 1 - QUICK START
General
Transducer Cable
This manual contains detailed operating instructions for all aspects of the D601 instrument. The following condensed instructions are provided to assist the operator in getting the instrument started up and running as quickly as possible. This pertains to basic operation only. If specific instrument features are to be used or if the installer is unfamiliar with this type of instrument, refer to the appropriate section in the manual for complete details.
1. TRANSDUCER LOCATION
A. In general, select a mounting location on the piping
system with a minimum of 10 pipe diameters (10 X the pipe inside diameter) of straight pipe upstream and 5 straight diameters downsteam. See Table 2.1 for additional configurations.
B. On horizontal pipe, select a position that is between 2
and 4 o’clock on the pipe, with 12 o’clock representing the top. Installations on vertical pipe should be made in an area where the flow moves from bottom to top— ensuring a full pipe of liquid.
2. PIPE PREPARATION AND TRANSDUCER MOUNTING
A. The piping surface, where the transducers are to be
mounted, needs to be clean and dry. Remove loose scale, rust and paint to ensure satisfactory acoustical bonds.
B. Loosely wrap the appropriate length of strap around
FLOW
Figure 1.1
Top View of Pipe
the pipe at the location determined in Step 1. Refer to Figure 1.1 for proper orientation of the transducer. For greatest accuracy, point the cable of the transducer in the primary flow direction.
C. Apply a liberal amount of couplant onto the transducer
face. Place the transducer onto the pipe ensuring square and true placement. If an RTV type of couplant (requiring curing time) was utilized, allow sufficient time for curing before applying power to the instrument or moving the cable.
Rev. 4/00 -1.2- D601
PART 1 - QUICK START
3. TRANSDUCER/POWER CONNECTIONS
Figure 1.2
Transducer (top picture)
and Power (bottom picture)
Connections
A. Do not
transducers.
B. Refer to the DIAGRAMS in Figure 1.2 for proper power
and transducer connections. Verify proper jumper selections are in place for the power source. See Figure 3.4.
4. INITIAL SETTINGS AND POWER UP
attempt to add additional cable to the
IMPORTANT!
In order to successfully complete the configuration of the D601 flow meter, the transducer must be mounted on a pipe which is full of a flowing liquid. It is normal to have a zero reading and no signal strength indication with empty pipes or zero flow rate.
A. Adjust the GAIN control [R13] to 1/4 turn from full
counter-clockwise rotation.
B. Apply power to the instrument. C. If the pipe is full of a flowing liquid, the flow meter
signal strength will increase from a zero reading.(press the
2nd FUNCT
Signal Strength does not increase to a minimum of 000125 counts, gradually turn the GAIN contr ol [R13] clockwise until the indication is between 000125 and
000200. (Do not over adjust this setting as ambient noise can influence readings.)
key, then press
SIGNAL STR)
. If the
D. If possible, turn off the flow in the pipe. Verify that
SIGNAL STR. is lower than 000100. If SIGNAL STR. is greater than 000100, verify that the sensor/ transmitter are not located near electrically noisy components. (VFDs, inverters, motors, power relays, etc) Verify that DT6 transducer connections are proper and secure. If the SIGNAL STR. remains greater than 000100, consult the Dynasonics Factory for assistance. It is possible that the GAIN control [R13] is set too far clockwise and ambient noise is influencing the readings. Turn the control counter-clock wise until the signal strength decreases to below 000070 counts.
E. If the instrument passes steps 4C and 4D, the basic
setup of the instrument is complete.
Rev. 4/00 -1.3- D601
PART 1 - GENERAL
General
Operating Theory
The Dynasonics D601 flowmeter is designed to measure the flow of liquids and slurries in full-pipe closed systems. The transmitter is field configured to measure flow on a variety of pipes and liquids. The standard product is typically used on pipe sizes ranging from 1 - 120 inch [ 25 - 1524 mm ] pipe I.D. (With the small pipe transducer option, the pipe size range is 0.25 - 1 inch [ 6 - 25 mm]). A variety of liquid applications can be accommodated: sewage, sludges, concrete, mining slurries, dredging, etc. Because the transducers are non-contacting and have no moving parts, the flow meter is not affected by system pressure, fouling or wear. Standard transducers are rated to 250°F [121°C]. Optional high temperature transducers are rated to operate to 400°F [204°C].
The basic principle of operation is the measurement of the frequency shift “Doppler” of a reflected ultrasonic signal from discontinuity in the flowing liquid. In theory, these discontinuities can be virtually any amount of suspended bubbles, solids, or interfaces caused by turbulent flow. In practice the degree to which this can be reliably accomplished is a function of the sensitivity and frequency of the transducer and associated transmitter. The D601 design requires greater than 100 PPM of suspended solids or bubbles over 100 microns in size. Should your application be cleaner than this, please check with the factory for information on transit time ultrasonic flowmeters. The transducer which generates and receives the ultrasonic signal supplies the data to the transmitter. The transmitter processes the signal and provides an analog and pulse output for velocity indicating and volumetric totalizing. In addition, the transmitter contains a signal strength indicator which determines satisfactory operation.
Rev. 4/00 -1.4- D601
PART 1 - GENERAL
Measuring Limits
The flowmeter is typically used as a unidirectional meter and is most accurate when the transducer is mounted in the orientation detailed in this manual. But, the meter will measure flow in both directions — although flow direction will not be indicated or totalized properly. The flowmeter will operate from signals returned from turbulence alone (such as installation directly at pump discharges or downstream from elbows and valves); however, it should be noted that turbulence may vary with flow rates and result in non-linear results. The repeatability of the device is not dependent on most process liquids.
The flowmeter is designed to measure the flow of liquids and slurries, as long as a small, homogeneous quantity of entrained air or suspended solids are present. Without the presence of continuing supply of air or solids, the transmitted pulses are not reflected back to the transducer and the indicator will indicate zero flow.
The signal strength value will indicate a value grea ter than 100 counts when a minimum size and concentration of suspended particles are available for a reliable flow reading (100 micron and 100 PPM minimum) and the liquid is moving at least 0.1 FPS [0.03 MPS]. Most water-based liquids can be measured from a factory calibrated flowmeter. However, liquids with a heavy solids level (i.e. over 2% by volume), liquids with sound speeds that vary from water (see Appendix - Liquid Sound Speed list) or pipes with liners may have to be field calibrated. This is done by adjusting the Calibration value on the keypad to make the indicator agree with a known flow velocity or a mathematically corrected fluid velocity. All standard flow meters are calibrated to measure the flow of a water-based liquid at 25 degrees C. If the scale range or units need to changed, the process to do so is covered in detail later in this manual.
Rev. 4/00 -1.5- D601
Serial Number
PART 1 - GENERAL
The D601 employs modular construction and provides electrical safety for the operator. The display face contains voltages no greater than 9 Vdc and any exposed metal work is electrically connected to Earth Ground. The display face swings open to allow access to user connections.
The serial number and complete model number of your D601 is located on the inside of the transmitter front cover. Should technical assistance be required, please provide the Dynasoncs’ Customer Service Department with this information. Email: dynasonics_technical@racinefed.com.
A part number breakdown of the transducer and transmitter portions of the flowmeter are described below.
Transmitter
Transducer
Rev. 4/00 -1.6- D601
PART 1 - GENERAL
Rev. 4/00 -1.7- D601
PART 2 - TRANSDUCER INSTALLATION
Transducer Mounting Locations
The following list outlines how to install the DT6 transducer for optimal performance, h ighest reliability and greatest accuracy:
1. Select a transducer site at least 10 pipe diameters downstream from bends, or fittings and 5 pipe diameters upstream. A symmetrical flow pattern is necessary for accuracy and repeatability over the
Table 2.1
1
1
The D601 system will provide repeatable measurements on piping systems that do not meet these
requirements, but the accuracy
Rev. 4/00 -2.1- D601
of these readings may be influenced to various degrees.
PART 2 - TRANSDUCER INSTALLATION
Transducer Cable
FLOW
Figure 2.2
Top View of Pipe
operating range of the meter. Down stream from pump or orifices, etc., locate at least 20 diameters. See
Table 2.1
2. On horizontal pipe, select a position that is between 2 and 4 o’clock on the pipe, with 12 o’clock representing the top. If the transducer is to be mounted on a vertical pipe, select a section of pipe where the flow is moving from bottom to top (flow moving vertically down a pipe tends to cavitate and provide unreliable operation.)
3. Mount the transducer in the orientation shown in Figure 2.2. The flow meter will read flow in both directions, but will be most accurate if the cable is mounted in the orientation shown—pointing in the primary flow direction.
4. If totalization of the measured fluid is required, the pipe must remain full. The meter will read when the liquid level is greater than the placement of the transducer, but the volumetric measurement will be based on a full pipe, so totalization will be higher than actual.
5. The flowmeter will achieve proper Doppler signals off of turbulence; however, it should be noted that turbulence may not be linear with pump speed changes, nor is the reading necessarily accurate due to the non-uniformity of turbulence.
6. When a liquid has less than 100 PPM of 100 micron or larger particles, try mounting the transducer within 12 inches of a pump discharge or other source of flow turbulence or cavitation. A reading obtained under these circumstances will be repeatable, but not necessarily accurate or linear.
7. It is a good practice to test the flow meter on the piping system before permanently mounting the transducer using RTV. Function can be verified by applying a water soluble lubricant, such as KY­Jelly, and holding the transducer by hand on the pipe in the location where the transducer will be
Rev. 4/00 -2.2- D601
PART 2 - TRANSDUCER INSTALLATION
permanently mounted. Under flowing liquid conditions, adequate signal is indicated when the Signal Strength indicates between 000125 and 000200 counts.
Acoustic Couplant Types
Small Pipe Transducers
For proper operation, there cannot be air voids between the traducer face and pipe. The space must be filled with a material which is a good transmitter of sound energy such as:
SILICONE GREASE: Dow Corning 111 R or comparable (-100 to +450 F.) The material must be suitable not to flow at temperature of pipe. Used for temporary survey installations and portable flow meters.
SILICONE RUBBER: Dow Corning 732-RTV R. Excellent for permanent bonding. This adhesive is a recommended bonding agent and easily removable.
INSTALLATION AND PIPE PREPARATION
The cable from the DT6 transducer is provided with either dual-coaxial cables, flexible nylon conduit or PVC coated steel conduit with a 1/2” NPT fitting. The coaxial cable was ordered from the factory at a specific length
should the coaxial cable be lengthened as this may de-tune the circuitry and influence performance
Installation of the DT61 and DT63 small-pipe transducers follow the same procedures as the DT60 and DT62 standard pipe type. The only difference is that the small pipe transducers u tilize an integral pipe clamping mechanism with two opposing sensing heads and the standard pipe units use a stainless steel strap.
UNDER NO CIRCUMSTANCES
.
Rev. 4/00 -2.3- D601
PART 2 - TRANSDUCER INSTALLATION
Intrinsic Safety Installations
Installations requiring intrinsic safety should refer to the Appendix drawings covering these applications.
1. Pipe Preparation:
For permanent silicone adhesive mounting, after determining the transducer location, some attention must be given to the pipe condition. Before the transducer head is bonded to the pipe surface, an area slightly larger than the flat surface to the transducer head (black rectangle) must be cleaned to bare metal. This means the re moval of all paint rust, and scale. Some minor pipe pitting will not cause problems, as the acoustic couplant will take up the voids. In the case where plastic pipe is used, remove all paint and grease so that a smooth, dry surface is exposed.
2. Transducer Mounting:
The transducer center line is designed to mount parallel to the pipe center line. The groove in the transducer body will allow the 1/2” stainless steel strap that was enclosed with the meter to align the transducer properly on the pipe. transducer on bends, elbows or fittings. Every effort should be made to mount the transducer parallel to the axis of the pipe as well as flat on the pipe. The transducer cable should run in the “down-stream” direction of liquid flow. See
In horizontal pipe runs, mount the transducer between 2 and 4 o’clock from the top—12 o’clock position; prepare the pipe surface as described. Finish the surface with some emery paper and then wipe the surface with trichlorenthylene to thoroughly degrease the contact surface in a area slightly larger than the flat surface of the transducer.
For permanent mounting, use a good silicone based
Figure 2.2
DO NOT
mount the
.
Rev. 4/00 -2.4- D601
PART 2 - TRANSDUCER INSTALLATION
adhesive (Dow-732). Spread a bead of the adhesive on the flat surface of the transducer face, covering well. Now spread a bead to the prepared pipe surface and press the head lightly to the pipe. Let the adhesive flow enough to fill in all the area beneath the head. At the same time, clamp (clamp supplied) into place until the silicone has set up. Ta ping along the edges of the head will hold the adhesive in place. A pad of adhesive must be formed between the transducer face and the pipe. Ensure that no relative movement between the transducer and the pipe takes place during the setup time (about 24 hours). Clamp transducer only tight enough to hold it in place while the adhesive is curing. Tighten for mechanical strength only after 24 hours. Secure the conduit as well.
3. Temporary Mounting and Spot Checks:
For temporary mounting, clean pipe as described and use silicone grease as the acoustical coupling material, holding by hand for spot readings or with a strap clamp for indefinite periods.
Rev. 4/00 -2.5- D601
PART 3 - PRE-INSTALLATION CHECKOUT
Unpacking
Functional Test
Figure 3.1
After unpacking, it is recommended to save the shipping carton and packing materials in case the instrument is stored or re-shipped. Inspect the equipment and carton for damage. If there is evidence of shipping damage, notify the carrier immediately.
The D601 flow meter can be checked for basic functionality using the following Bench Test procedure. It is recommended that this operation be performed before permanently installing the system.
Procedure:
1. Open the D601 transmitter cover.
2. Connect the transducer cable connector to the terminal locations on the lower left corner of the D601 main circuit card. See Figure 3.1.
3. Connect supply power to the appropriate terminal locations on the upper right corner of the main circuit card. See Figure 3.2. Verify that the power supply selection jumpers are configured properly—See Fig- ure 3.4.
4. Apply power.
5. Verify that the display indicates 0.00 FPS (or 0.0 flow rate of any other unit). If the display does not register
0.0, then press the 2nd FUNCT key, then press SIGNAL STR. Verify that SIGNAL STR. is lower than
000100. If SIGNAL STR. is greater than 000100, verify that the sensor/transmitter are not located near electrically noisy components. (VFDs, inverters, motors, power relays, etc) Verify that DT6 transducer connections are proper and secure. If the SIGNAL STR. remains greater than 000100, consult the Dynasonics Factory for assistance.
6. Press the 2nd FUNCT key to enter SERVICE MODE.
Figure 3.2
Rev. 4/00 -3.1- D601
Press the SIGNAL STR key to display SIGNAL STR. XXXXXX.
7. Rub the face of the transducer lengthwise back and
PART 3 - PRE-INSTALLATION CHECKOUT
forth with your thumb using moderate pressure. The cycle time should be 1-2 times per second.
8. Verify that signal strength increase s with frequenc y of the rubbing. Typical increases will range from 20-30 counts.
9. Verify that signal strength decreases when rubbing ceases.
Bench Test is Complete
Rev. 4/00 -3.2- D601
PART 3 - TRANSMITTER INSTALLATION
PART 2 - ELECTRIC AL CONNECTIONS
Transmitter Installation
1. Place the D601 transmitter in a location that is:
Where little vibration exist. Protected from falling corrosive fluids. Within ambient temperature limits - 22 to 122°F [-30 to
50°C]
Out of direct sunlight. Direct sunlight may increase
temperatures within the transmitter to above maximum limit.
2. Mounting: Refer to Figure 3.3 for enclosure and mounting dimension details. Ensure that enough room is available to allow for door swing, maintenance and conduit entrances. Secure the enclosure to a flat surface with four appropriate fasteners.
3. Conduit holes. Conduit hubs should be used where cables enter the enclosure. Holes not used for cable entry should be sealed with plugs.
NOTE: Use NEMA 4 [ IP65 ] rated fittings and plugs to maintain the water tight integrity of the enclosure. Generally, the right conduit hole (viewed from front) is used for line power; the left conduit hole for transducer connections.
4. If additional holes are required, (analog outputs, etc.) drill the appropriate size hole in the enclosure’s bottom. Use extreme care not to run the drill bit into the wiring or circuits cards.
To access terminal strips for electronic connectors, loosen the two screws in the enclosure door and open the door.
Rev. 4/00 -3.3- D601
PART 3 - TRANSMITTER INSTALLATION
PART 2 - SERVICE AND M AINTENANCE
FIGURE 3.3
Important !
Rev. 4/00 -3.4- D601
NOTE: The transducer cable carries low level signals. Do not attempt to add additional cable to the factory supplied transducer cable.
PART 3 - TRANSMITTER INSTALLATION
PART 2 - SERVICE AND M AINTENANCE
4-20mA OUTPUT
Power Connections
The 4-20mA output is proportional to the flow rate measuring scale and can drive a load of up to 1000 ohms. The output is isolated from earth ground and circuit low. Connect the load to the 4-20 mA connection terminals located on the inside of the D601 enclosure, matching polarity as indicated.
Line power is connected by supplying power to the appropriate terminals located inside of the D601 enclosure. Use wiring practices that conform to local codes (National Electric Code Hand book in the USA). Use only the standard three wire connection. The ground terminal grounds the instrument, which is mandatory for safe operation.
CAUTION: Any other wiring method may be unsafe or cause improper operation of the instrument.
It is recommended not to run line power with other signal wires within the same wiring tray or conduit. Verify that the power supply jumper connections are oriented cor­rectly for the power source being wired. The electronics can be damaged if improper power is connected or if jumpers are not installed correctly. The DC input is not fuse protected. It is recommended that an external fuse be installed if DC power is selected. The fuse shou ld be a 1A delay action type. See Figure 3.4
Power Source Jumpers 115 VAC JP8, JP10, JP11
Figure 3.4
Power Supply Jumper
Selection
Rev. 4/00 -3.5- D601
230 VAC JP9, JP11 100 VAC JP7 12 VDC JP3, JP5 24 VDC JP4, JP6
NOTE: This instrument requires clean electrical line power. Do not connect the meter on a circuit which oper­ates lighting ballasts, motors, solenoids, etc.
PART 3 - TRANSMITTER INSTALLATION
Transducer Connections
CTR Output
1. Guide the transducer terminations through the transmitter conduit hole located on the left side of the enclosure. Secure the transducer cable with the supplied conduit nut.
2. The terminals on the transducer cable are coded with wire markings. Connect the appropriate wires to the corresponding screw terminals in the transmitter.
NOTE: The transducer cable carries low level signals. Do not attempt to add additional cable to the factory supplied transducer cable.
If additional cable is required, contact the Dynasonics factory to arrange for an exchange transducer with the appropriate length of cable. Cables to 300 feet [90 meters] are available.
NOTE: An additional hole in the transmitter enclosure is required for outputs. Drill the hole in the the enclosure bottom taking care not to drive the drill bit into wiring or the circuit boards with the transmitter.
The CTR pulse output is proportional to the flow rate measuring scale. This output may be used one of two ways:
To drive a 12V logic device or electromechanical total-
izers.
To drive a low impedance, 12V device. Minimum
resistance 50 ohms.
The pulse output pulses with totalizer increments. The connections are located on the right side of the signal processing PCB in the back of the enclosure. The pulse width is fixed at 50 milli-seconds. CTR “ - ” represents circuit low. CTR “ + “ represents 12 Vdc pulse output.
Rev. 4/00 -3.6- D601
PART 3 - STARTUP AND CONFIGURATION
Before Starting the Instrument
Instrument Startup
Note: The D601 flow meter system requires a full pipe of
flowing liquid before a successful startup can be completed. Do not attempt to make adjustments or change configurations until a full pipe is verified.
Procedure:
1. Verify that all wiring is properly connected and routed.
2. Apply power.
3. Adjust the GAIN control [R13] to 1/4 turn from full counter-clockwise rotation.
4. Apply power to the instrument.
5. If the pipe is full of a flowing liquid that contains ade­quate concentrations of suspended solids, the flow me­ter signal strength will increase from a zero reading. (press the 2nd FUNCT key, then press SIGNAL STR). If the Signal Strength does not increase to a minimum of 000125 counts, gradually turn the GAIN control [R13] clockwise until the indication is at between 000125 and 000200. (Do not over adjust this setting as ambient noise can influence readings.)
6. If possible, turn off the flow in the pipe. Verify that SIGNAL STR. is lower than 000100 . If SIGNAL STR. is greater than 000100, verify that the sensor/ transmitter are not located near electrically noisy components. (VFDs, inverters, motors, power relays, etc) Verify that DT6 transducer connections are proper and secure. If the SIGNAL STR. remains greater than 000100, consult the Dynasonics Factory for assistance. It is possible that the GAIN control [R13] is set too far clockwise and ambient noise is influencing the readings. Turn the control counter-clockwise until the signal strength decreases to below 000070 counts.
5. If the instrument passes steps 5 and 6, the basic setup of the instrument is complete.
Important!
Rev. 4/00 -3.7- D601
It is normal to have low/zero SIGNAL STRENGTH indication at ZERO flow.
PART 3 - KEYPAD CONFIGURA TION
After a successful flow meter installation and startup (covered in the previous sections of this manual) the D601 can be keypad configured to provide select engineering unit readings of flow and a scaled 4-20mA output. Configuration inputs are made via the keypad and are stored by the microprocessor. The entries are retained by the flow meter’s E2PROM memory in the event of power failure. If fluid velocity readings, FPS or MPS, are the only required measurement keypad configuration is not required.
UP/DOWN Arrows
Allow changing of the D601 configuration constants. Use the UP arrow to increase constant values and the DOWN arrow to decrease values. The arrows can be momentarily pressed to change values incrementally or held to advance continuously. Constants outside of the valid range of the D601 cannot be displayed. The scrolling rate at which the values will change is two tiered. Scrolling will be relatively slow durin g the first five seco nd s of a continuous keypad press; the scrolling rate will increase after that time to allow rapid changes of large values.
2nd FUNCT Controls access to the commands located on the lower
half of the keys. After pressing this key the word SERVICE MODE will appear on the LCD indicator.
ENTER
Records and activates the configuration constant value that is displayed on the LCD indicator. Can also be used to return the meter to its run mode.
Rev. 4/00 -3.8- D601
PART 3 - KEYPAD CONFIGURA TION
F1 and F2
Not utilized.
RESET Caution: Conducts a system reset. All configuration
constants will be lost and the D601 will load defau lt values for all constants.
I.D. Allows entry of a pipes internal diameter. Internal
diameters must be entered if volumetric flow rates are to be displayed.
! If a UNITS code for U.S. measurements was made the
I.D. value will be entered in inches. Valid ranges for this entry are 0.25 to 120.00 inches.
! If a UNITS code for metric measurement was made
the I.D. value will be entered in millimeters. Valid ranges for this entry are 6 to 3050 millimeters.
The appendices in the back of this manual contain tables of common pipe sizes and schedules. If the pipe size does not appear in the table, consult the pipe manufacture or conduct a physical measurement of the pipe internal diameter. Errors in the entry of this value can result in large inaccuracies.
FULL SCALE
Allows entry of the maximum fluid velocity anticipated within the pipe. This value does not have any bearing on displayed flow rates or values, but is used to scale the span value of the 4-20mA analog output. If the analog output is not going to be utilized set this value to 30 .00 if measuring in U.S. units or 10.00 if measuring in metric units.
! If a UNITS code for U.S. measurements was made the
FULL SCALE value will be entered in FPS (feet per
Rev. 4/00 -3.9- D601
PART 3 - KEYPAD CONFIGURA TION
second). Valid ranges for this entry are 0.00 to 30.00 FPS. Two useful equations that relate liquid velocity to volume:
GPM = 2.45 X I.D.2 X FPS
FPS = ( GPM X 0.408) / I.D. I.D. in inches ! If a UNITS code for metric measurement was made
the I.D. value will be entered in MPS (meters per second). Valid ranges for this entry are 0 to 10.00 MPS. Two equations that relate liquid velocity to volume are
LPM = 0.047 X I.D.
MPS = ( LPM X 21.28) / I.D.2 I.D. in inches
2
X MPS
2
Note: Attempting to set a FULL SCALE value of less than
0.5 FPS [0.15 MPS] may result in an unstable transmitted output. If flows are typically lower than this range, the LOW FLOW FILTERS and higher DAMPing values may be required.
UNITS
Utilized to set engineering units of measure. There are twelve different selections possible. The Table 3.1 lists the entry code number, flow rate unit of measure and totalizer unit of measure.
Note: After changing the UNITS value, it may be necessary to change other configuration values accordingly. For example, FULL SCALE, I.D., HIGH/LOW ALARM are influenced by the UNITS entry.
Note: Flow already accumulated will not be correctly compensated for if the UNITS of measur e cha nges.
Rev. 4/00 -3.10- D601
PART 3 - KEYPAD CONFIGURA TION
Table 3.1
UNITS Code Flow Rate Totalizer
0 1 2 3
4 5 6 7 8
9
FPS (feet per sec) N/A GPM (gallons per min) GALLONS GPH (gallons per hr) GALLONS MGD (millions of gal
pre day) CFM (ft.3 per min) CF
MPS (meters per sec) N/A CMH (m3 per hr) m3 LPM (liters per min) Liters MLD (millions of liters
per day) BPM (barrels per min) BARRELS
GALLONS
Liters
10 11
Rev. 4/00 -3.11- D601
BPH (barrels per hr) BARRELS BPD (barrels per day) BARRELS
PART 3 - KEYPAD CONFIGURA TION
HIGH ALARM (Labeled RELAY-1 on the Main PCB) Controls the set-point of the SPDT relay labeled RELAY-1
on the Main PCB. Enter a liquid velocity at which a relay­contact action is desired. Relay contacts are utilized for signaling flo w rate conditions tha t are higher of lower th an a desired set point. If a relay setting is made very close to a nominal liquid velocity, relay “chatter” ( rapid opening and closing of the relay ) may result.
! If a UNITS code for U.S. measurements was made the
HIGH ALARM value will be entered in FPS. Valid ranges for this entry are 0.00 to 30.00 FPS.
! If a UNITS code for metric measurements was made
the HIGH ALARM value will be entered in MPS. Valid ranges for this entry are 0.00 to 10.00 MPS.
LOW ALARM (Labeled RELAY-2 on the Main PCB) Controls the set-point of the SPDT relay labeled RELAY-2
on the Main PCB. Enter a liquid velocity at which a relay­contact action is desired. Relay contacts are utilized for signaling flow rate conditions that are higher of lower than a desired set point. If a relay setting is made very close to a nominal liquid velocity, relay “chatter” ( rapid opening and closing of the relay ) may result.
! If a UNITS code for U.S. measurements was made the
HIGH ALARM value will be entered in FPS. Valid ranges for this entry are 0.00 to 30.00 FPS.
! If a UNITS code for metric measurements was made
the HIGH ALARM value will be entered in MPS. Valid ranges for this entry are 0.00 to 10.00 MPS.
TOTAL MULT
Utilized for setting the flow totalizer exponent and changing the External Counter output. This feature is useful for accommodating a very large accumulated flow.
n
The exponent is a “ X 10
0
0 (10
, X 1 multiplier) to 4 (104, X 10,000 multiplier).
” multiplier, were “n” ca n be f rom
Rev. 4/00 -3.12- D601
PART 3 - KEYPAD CONFIGURA TION
The External Counter output, available at the two terminals labeled CTR on the Main PCB, is influenced by the TOTAL MULT value. Since the output is designed to operate electromechanical accumulators, large flow rates will require that the TOTAL MULT be set to a value usable by these types of counters (typically speeds no faster than 3 cps). The following chart tabulates suggested settings vs. flow ranges:
Exponent Multiplier Useable CTR
Range: GPM or LPM
0 1 2 3 4
X1 30-600 X10 300-6,000 X100 3,000-60,000 X1,000 30,000-600,000 X10,000 300,000-6,000,000
TOTAL ON/OFF
This key has three functions:
Key press number Operation
First press Stops the internal totalizer/
external CTR and displays the last value
Second press Resets the internal totalizer
to zero
(continued)
Rev. 4/00 -3.13- D601
PART 3 - KEYPAD CONFIGURA TION
(continued)
Third press Restarts the internal
totalizer/external CTR (The internal totalizer starts
from zero.)
If inhibiting (pausing) the totalizer is necessary, there are two methods suggested:
1. Connect and external totalizer to the CTR terminals.
See the section of this manual related to CTR electrical connections for connection parameters.
2. To inhibit the internal totalizer without resetting the
accumulation, press the TEST key to pause the accumulation. Press the ENTER key to resume accumulation.
LOCK ON
To ensure security of the configuration and accumulated flow, the keypad can be locked. To enable the keyboard lock out, press LOCK ON key, the display will show LOCK ON. Press ENTER to return to Run mode. To turn the lock off, press the LOCK ON key. Use the arrow keys to set a value of 125. Press the ENTER key. The display will show LOCK OFF to acknowledge that all keypad entries can now be made.
CAL
A few factors can influence the readings of a D601 flow meter. The CAL entry allows the user to compensate for flow discrepancies without affecting the factory calibration. Examples of situations that can cause reading discrepancies are:
! Operations on liquids with sonic velocity carrying
properties that are different from water. See the table
Rev. 4/00 -3.14- D601
PART 3 - KEYPAD CONFIGURA TION
of correction factors located in the Appendix of this manual for Liquid Sound Speed and their associated
correction factors. ! Transducer mounted in non-recommended locations. By applying a CAL value other than 100%, the factory-
calibrated readings will be altered by the percentage entered. This CAL value will be reflected in the display, 4­20mA and CTR outputs and relay settings.
For example, if a reading of 175 GPM is displayed and the known flow rate is 160 GPM, a CAL value of
160 GPM x 100 = 91.4% 175 GPM
The D601 will not allow decimal values to be entered as a CAL constant, so round to the nearest whole number; in this case 91%.
Acceptable input ranges for the CAL constant are 0-255%.
DAMP
In installations where very turbulent or erratic flow is encountered, increasing the Damping setting can increase display and output stability. The DAMP setting increases and decreases the response time of the flow meter display and outputs. To set the damping time constant, press the DAMP key. Set a value between 1 and 10, 1 having the fastest response and 10 having the slowest response. Press ENTER to complete the configuration.
TEST
The meter contains a test function for verification of the 4­20mA analog and CTR external counter outputs. To activate the test function, press the TEST key. Verify that 20mA is flowing in the 4-20mA output and verify that the CTR output is supplying 50mS pulses. Press ENTER to exit the test function.
Rev. 4/00 -3.15- D601
PART 3 - KEYPAD CONFIGURA TION
2nd FUNCT - SERVICE MODE
nd
{2
FUNCT} SIGNAL STR
Displays the raw Doppler signal strength value. This value will increase as the velocity of the liquid increases. Typically, a liquid flowing at a velocity greater than 0.2 FPS [0.06 MPS], with adequate suspended solids (100 ppm or 100 micron or larger solids) or aeration, will produce SIGNAL STR readings of at least 000125 counts.
NOTE: If the liquid is not flowing a low SIGNAL STR reading is non-conclusive. If a high SIGNAL STR is indicated at zero flow rate, it indicates that a source of interference (another ultrasonic instrument, VFD, or poor electrical ground) may be present. Verify that SIGNAL STR increases when the flow starts. If it does, increase the SS CUTOFF setting (see SS CUTOFF).
If SIGNAL STR is lower than 100 counts in a flowing liquid, one or more of the following steps may need to be invoked:
1. If the liquid velocity is less than 1 FPS (0.3 MPS) turn
SW-1 “LOW FLOW” switch ON. (This dual DIP switch
is located near the center of the Main PCB.)
2. If SW-1 did not cause an increase in SIGNAL STR to a
level above 100, turn ON SW-2.
3. There may not be adequate reflectors for the Doppler
principle to operate. The DT6 transducer can be
relocated to a source of liquid degasification, such as
would be found a 1-3 diameters down stream of a 90-
degree elbow. A surrogate source of aeration can also
be introduced by bleeding a small amount of
compressed air into the line several diameters
upstream of the DT6 transducer.
Rev. 4/00 -3.16- D601
PART 3 - KEYPAD CONFIGURA TION
{2nd FUNCT} 4 mA The 4-20mA output on standard D601 flow meters is
scaled at zero flow equals 4mA and 20 FPS (6.08 MPS) equals 20mA. The 4mA key allows fine adjustments to be made to the “zero” of the 4-20mA output or allows offset to be placed on the 4-20mA output. To adjust the 4mA output, an ammeter or reliable refer ence con n ecti on to the 4-20mA output must be present.
Procedure:
1. Either break the present current loop and connect the
ammeter in series (disconnect either wire at the
terminal block labeled 4-20mA on the Main PCB of the
D601) or, if this output is not being utilized, connect the
ammeter + to the + terminal and – to the – terminal of
the 4-20mA output.
2. Press the 4mA key.
3. With no flow moving through the pipe, adjust the
setting count using the arrow keys until 4.00mA is
indicated on the ammeter. The typical count value
range for this setting is between 3350 and 3850.
4. Press ENTER to store the value.
5. Re connect the 4-20mA output circuitry as required.
{2nd FUNCT} VEL ADC Press VEL ADC to display the raw analog to digital
converter counts that are being processed by the microprocessor. This count value will vary linearly with flow rate from 0000 at zero flow rate to 1024 at max i mum full-scale flow rate. No modifications of this count can be made, this display is for diagnostic purposes only.
nd
FUNCT} BLANKING
{2 This key is unused.
Rev. 4/00 -3.17- D601
PART 3 - KEYPAD CONFIGURA TION
{2nd FUNCT} 20mA The 4-20mA output on standard D601 flow meters is
scaled at zero flow equals 4mA and 20 FPS (6.08 MPS) equals 20mA. The 20mA key allows fine adjustments to be made to the “span” of the 4-20mA output. To adjust the 20mA output, an ammeter or reliable reference connection to the 4-20mA output mus t be pres ent.
Procedure:
1. Either break the present current loop and connect the
ammeter in series (disconnect either wire at the
terminal block labeled 4-20mA on the Main PCB of the
D601) or, if this output is not being utilized, connect the
ammeter + to the + terminal and – to the – terminal of
the 4-20mA output.
2. Press the 20mA key.
3. With maximum flow moving through the pipe, adjust
the setting count using the arrow keys until 20.00mA is
indicated on the ammeter. The typical count value
range for this setting is between 1450 and 1950.
4. Press ENTER to store the value.
5. Re connect the 4-20mA output circuitry as required.
{2nd FUNCT} SS CUTOFF This entry represents the signal strength cutoff level (low
velocity cutoff). At flow rates below 0.2 FPS [0.06 MPS] the readings displayed by the D601 are unreliable. By utilizing the default SS CUTOFF of 100 counts, small unstable readings at low flow rates will be avoided.
Note: If SS CUTOFF is set to a level higher than the SIGNAL STR value the meter will not show flow or outp ut any values.
Rev. 4/00 -3.18- D601
PART 3 - KEYPAD CONFIGURA TION
{2nd FUNCT} INPUT F.S. This key is used to select a maximum velocity that the
instrument will operate. Choices are 10, 20 and 30 FPS. It is not recommended to deviate alter this value from factory settings, as certain spans have been set that correlate to the set maximum velocity. Consult the Dynasonics factory for adjustment procedures.
{2nd FUNCT} DAC 3 This key is not used.
{2nd FUNCT} BACK LIGHT Toggles the electro-luminescent LCD back lighting ON
and OFF. This type of back lighting has an illumination half-life of approximately one year. If the instrument is left unattended for long periods of time, turning the back lighting OFF can preserve the electro-luminescent panel and save a small amount of power.
{2nd FUNCT} CONTRAST This setting allows the adjustment of the LCD contrast.
An LCD’s viewing quality is affected by temperature, ambient lighting, back lighting and viewing angle. Adjust the contrast level to provide the best contrast possible. Default count is 50.
Rev. 4/00 -3.19- D601
PART 3 - TROUBLE SHOOTING
CONDITION
Unit does not turn “ON” when power is applied
FAULT CONDITION is indicated on LCD indicator
POSSIBLE CAUSE
Verify that AC power source is live.
Test the fuse
Verify that proper power supply jum pers are in place.
Fault conditions can arise from several causes: electrical surges, short circuits, open circuits, etc. To clear a “Fault Condi­tion”, press
set
. Use the arrow keys to change the Fault code number to “0”. Press the meter resumes measuring flow, no permanent damage was incurred.
Ensure that the transducer is properly mounted to the pipe.
Verify that transducer connections are correct
2nd Function
then press
Enter
Re-
. If
Ensure that the pipe is full of moving liquid.
If SIGNAL STR is less than 000100 counts and flow rate is greater than 1 FPS [0.3 MPS], adjust GAIN control (R13 on the Main PCB) till SIGNAL STR reaches at least 000125 counts.
On cleaner liquids, move the transducers closer to a 90
If GND connection and pipe are at different potentials, ground D601 to pipe potential.
If Variable Frequency Drives are being utilized, verify that the D601 obt ains a flow indication when the pum p turns O FF . If it does, contact the Dynasonics factory.
o
pipe elbow.
Rev. 4/00 -3.20- D601
PART 3 - TROUBLE SHOOTING
Stability of flow readings are unsatisfactory
Erroneous Reading
Increase the DAMP constant from keypad.
Move transducers to a location further
from piping tees, elbows, valves, filters,
Transducer mounted incorrectly or not true to the pipe.
Another local ultrasonic instrument is operating at about the same frequency [ consult the Dynasonics factory].
Presence of large amounts of suspended solids or aeration. Use CAL constant to compensate.
Sources of radiated interference are present. Apply appropriate shielding.
An electrically noisy power supply is powering the D601. Power the meter with a circuit that does not power motors,
The D601 display indicates flow, when true fluid velocity is zero.
Rev. 4/00 -3.21- D601
Verify that residual leakage and flow is not present. [I.e. leaking check valves]
Verify that GAIN control (R13 on the Main PCB) is not adjusted too high. With nominal flow running through the pipe, adjust GAIN control till the display zeros with no flow.
APPENDIX
Appendix
Fluid Sound Speed Conversions Pipe Dimension Chart: Ductile Iron Pipe Dimension Chart: Cast Iron Pipe Dimension Chart: Steel, SS, PVC Velocity to Volumetric Conversion Chart Intrinsic Safety Connections Statement of Warranty Customer Service
Rev. 4/00 -3.22- D601
Fluid Sound Speeds
g
Original Date: 10/19/99 Revision: none Revision Date: none File: I:/dynasonics/dyna_code/tables/doppler ss conversions.xls
120.0176921
Doppler
Fluid Specific Gravity Sound Speed Calibration Entry
20 de
Acetate, Butyl (n) 1270 4163.9 85 Acetate, Ethyl 0.901 1085 3559.7 72 Acetate, Methyl 0.934 1211 3973.1 81 Acetate, Propyl 1280 4196.7 85 Acetone 0.79 1174 3851.7 78 Alcohol 0.79 1207 3960.0 81 Alcohol, Butyl (n) 0.83 1270 4163.9 85 Alcohol, Ethyl 0.83 1180 3868.9 79 Alcohol, Methyl 0.791 1120 3672.1 75 Alcohol, Propyl (I) 1170 3836.1 78 Alcohol, Propyl (n) 0.78 1222 4009.2 82 Ammonia (35) 0.77 1729 5672.6 115 Anlline (41) 1.02 1639 5377.3 109 Benzene (29,40,41) 0.88 1306 4284.8 87 Benzol, Ethyl 0.867 1338 4389.8 89 Bromine (21) 2.93 889 2916.7 59 n-Butane (2) 0.60 1085 3559.7 72 Butyrate, Ethyl 1170 3836.1 78 Carbon dioxide (26) 1.10 839 2752.6 56 Carbon tetrachloride 1.60 926 3038.1 62 Chloro-benezene 1.11 1273 4176.5 85 Chloroform (47) 1.49 979 3211.9 65 Diethyl ether 0.71 985 3231.6 66 Diethyl Ketone 1310 4295.1 87 Diethylene glycol 1.12 1586 5203.4 106 Ethanol 0.79 1207 3960.0 81 Ethyl alcohol 0.79 1207 3960.0 81 Ether 0.71 985 3231.6 66 Ethyl ether 0.71 985 3231.6 66 Ethylene glycol 1.11 1658 5439.6 111 Freon R12 774.2 2540 52 Gasoline 0.7 1250 4098.4 83 Glycerin 1.26 1904 6246.7 127 Glycol 1.11 1658 5439.6 111 Isobutanol 0.81 1212 3976.4 81 Iso-Butane 1219.8 4002 81 Isopentane (36) 0.62 980 3215.2 65 Isopropanol (46) 0.79 1170 3838.6 78 Isopropyl alcohol (46) 0.79 1170 3838.6 78 Kerosene 0.81 1324 4343.8 88 Linalool 1400 4590.2 93
rees C m/s ft/s relative to 25C water
Linseed Oil .925-.939 1770 5803.3 118
(
)
Methanol (40,41) 0.79 1076 3530.2 72 Methyl alcohol (40,44) 0.79 1076 3530.2 72 Methylene chloride (3) 1.33 1070 3510.5 71 Methylethyl Ketone 1210 3967.2 81 Motor Oil (SAE 20/30) .88-.935 1487 4875.4 99 Octane (23) 0.70 1172 3845.1 78 Oil, Castor 0.97 1477 4845.8 99 Oil, Diesel 0.80 1250 4101 83 Oil (Lubricating X200) 1530 5019.9 102 Oil (Olive) 0.91 1431 4694.9 96 Oil (Peanut) 0.94 1458 4783.5 97 Paraffin Oil 1420 4655.7 95 Pentane 0.626 1020 3346.5 68 Petroleum 0.876 1290 4229.5 86 1-Propanol (46) 0.78 1222 4009.2 82 Refrigerant 11 (3,4) 1.49 828.3 2717.5 55 Refrigerant 12 (3) 1.52 774.1 2539.7 52 Refrigerant 14 (14) 1.75 875.24 2871.5 58 Refrigerant 21 (3) 1.43 891 2923.2 59 Refrigerant 22 (3) 1.49 893.9 2932.7 60 Refrigerant 113 (3) 1.56 783.7 2571.2 52 Refrigerant 114 (3) 1.46 665.3 2182.7 44 Refrigerant 115 (3) 656.4 2153.5 44 Refrigerant C318 (3) 1.62 574 1883.2 38 Silicone (30 cp) 0.99 990 3248 66 Toluene (16,52) 0.87 1328 4357 89 Transformer Oil 1390 4557.4 93 Trichlorethylene 1050 3442.6 70 1,1,1-Trichloro-ethane 1.33 985 3231.6 66 Turpentine 0.88 1255 4117.5 84 Water, distilled Water 0 degrees C 1402 4596.7 94 Water 20 degrees C 1482 4859.0 99 Water 40 degrees C 1529 5013.1 102 Water 60 degrees C 1551 5085.2 103 Water 80 degrees C 1554 5095.1 104 Water 100 degrees C 1543 5059.0 103 Water 120 degrees C 1519 4980.3 101 Water 140 degrees C 1485 4868.9 99 Water 160 degrees C 1440 4721.3 96 Water 180 degrees C 1390 4557.4 93 Water 200 degrees C 1333 4370.5 89 Water, heavy 1 1400 4593 93 Water, sea 1.025 1531 5023 102 Wood Alcohol (40,41) 0.791 1076 3530.2 72 m-Xylene (46) 0.868 1343 4406.2 90 o-Xylene (29,46) 0.897 1331.5 4368.4 89 p-Xylene (46) 1334 4376.8 89
49,50
0.996 1498 4914.7 100
Cement Lining
Ductile Iron Pipe
Standard Classes
Pipe Size
(inches)
3 3.96 3.46 0.25 3.40 0.28 3.34 0.31 3.28 0.34 3.22 0.37 3.14 0.41 4 4.80 4.28 0.26 4.22 0.29 4.16 0.32 4.10 0.35 4.04 0.38 3.93 0.44 6 6.90 6.40 0.25 6.34 0.28 6.28 0.31 6.22 0.34 6.16 0.37 6.10 0.40 6.04 0.43 .123/.250
8 9.05 8.51 0.27 8.45 0.30 8.39 0.33 8.33 0.36 8.27 0.39 8.21 0.42 8.15 0.45 10 11.10 10.32 0.39 10.46 0.32 10.40 0.35 10.34 0.38 10.28 0.41 10.22 0.44 10.16 0.47 12 13.20 12.58 0.31 12.52 0.34 12.46 0.37 12.40 0.40 12.34 0.43 12.28 0.46 12.22 0.49
14 15.30 14.64 0.33 14.58 0.36 14.52 0.39 14.46 0.42 14.40 0.45 14.34 0.48 14.28 0.51 16 17.40 16.72 0.34 16.66 0.37 16.60 0.40 16.54 0.43 16.48 0.46 16.42 0.49 16.36 0.52 18 19.50 18.80 0.35 18.74 0.38 18.68 0.41 18.62 0.44 18.56 0.47 18.50 0.50 18.44 0.53 .1875/.375 20 21.60 20.88 0.36 20.82 0.39 20.76 0.42 20.70 0.45 20.64 0.48 20.58 0.51 20.52 0.54 24 25.80 25.04 0.38 24.98 0.41 24.92 0.44 24.86 0.47 24.80 0.50 24.74 0.53 24.68 0.56
30 32.00 31.22 0.39 31.14 0.43 31.06 0.47 30.98 0.51 30.90 0.55 30.82 0.59 30.74 0.63
Outside
Diameter
(inches)
Class 53
ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall
Class 54 Class 55 Class 56Class 50 Class 51 Class 52
Std./Double
Thickness
36 38.30 37.44 0.43 37.34 0.48 37.06 0.62 37.14 0.58 37.40 0.45 36.94 0.68 36.84 0.73 42 44.50 43.56 0.47 43.44 0.53 43.32 0.59 43.20 0.65 43.08 0.71 42.96 0.77 42.84 0.83 .250/.500 48 50.80 49.78 0.51 49.64 0.58 49.50 0.65 49.36 0.72 49.22 0.79 49.08 0.86 48.94 0.93 54 57.10 55.96 0.57 55.80 0.65 55.64 0.73 55.48 0.81 55.32 0.89 55.16 0.97 55.00 1.05
March, 2000
Cast Iron Pipe
Standard Classes
CLASS B CLASS C CLASS D CLASS E CLASS F CLASS G CLASS HCLASS A
Size
(Inches)
10 11.10 10.10 0.50 11.10 9.96 0.57 11.40 10.16 0.62 11.40 10.04 0.68 11.60 10.12 0.74 11.60 10.00 0.80 11.84 10.12 0.86 11.84 10.00 0.92
12 13.20 12.12 0.54 13.20 11.96 0.62 13.50 12.14 0.68 13.50 12.00 0.75 13.78 12.14 0.82 13.78 12.00 0.89 14.08 12.14 0.97 14.08 12.00 1.04 14 15.30 14.16 0.57 15.30 13.98 0.66 15.65 14.17 0.74 15.65 14.01 0.82 15.98 14.18 0.90 15.98 14.00 0.99 16.32 14.18 1.07 16.32 14.00 1.16 16 17.40 16.20 0.60 17.40 16.00 0.70 17.80 16.20 0.80 17.80 16.02 0.89 18.16 16.20 0.98 18.16 16.00 1.08 18.54 16.18 1.18 18.54 16.00 1.27 18 19.50 18.22 0.64 19.50 18.00 0.75 19.92 18.18 0.87 19.92 18.00 0.96 20.34 18.20 1.07 20.34 18.00 1.17 20.78 18.22 1.28 20.78 18.00 1.39 20 21.60 20.26 0.67 21.60 20.00 0.80 22.06 20.22 0.92 22.06 20.00 1.03 22.54 20.24 1.15 22.54 20.00 1.27 23.02 20.24 1.39 23.02 20.00 1.51
24 25.80 24.28 0.76 25.80 24.02 0.89 26.32 24.22 1.05 26.32 24.00 1.16 26.90 24.28 1.31 26.90 24.00 1.45 27.76 24.26 1.75 27.76 24.00 1.88 30 31.74 29.98 0.88 32.00 29.94 1.03 32.40 30.00 1.20 32.74 30.00 1.37 33.10 30.00 1.55 33.46 30.00 1.73 36 37.96 35.98 0.99 38.30 36.00 1.15 38.70 35.98 1.36 39.16 36.00 1.58 39.60 36.00 1.80 40.04 36.00 2.02 42 44.20 42.00 1.10 44.50 41.94 1.28 45.10 42.02 1.54 45.58 42.02 1.78 48 50.50 47.98 1.26 50.80 47.96 1.42 51.40 47.98 1.71 51.98 48.00 1.99
O.D.
Inch
3 3.80 3.02 0.39 3.96 3.12 0.42 3.96 3.06 0.45 3.96 3.00 0.48 4 4.80 3.96 0.42 5.00 4.10 0.45 5.00 4.04 0.48 5.00 3.96 0.52 6 6.90 6.02 0.44 7.10 6.14 0.48 7.10 6.08 0.51 7.10 6.00 0.55 7.22 6.06 0.58 7.22 6.00 0.61 7.38 6.08 0.65 7.38 6.00 0.69 8 9.05 8.13 0.46 9.05 8.03 0.51 9.30 8.18 0.56 9.30 8.10 0.60 9.42 8.10 0.66 9.42 8.10 0.66 9.60 8.10 0.75 9.60 8.00 0.8
I.D.
Inch
Wall
O.D.
Inch
I.D.
Inch
Wall
O.D.
Inch
I.D.
Inch
Wall
O.D. Inch
I.D.
Inch
Wall
O.D.
Inch
I.D.
Inch
Wall
O.D.
Inch
I.D.
Inch
Wall
O.D.
Inch
I.D.
Inch
Wall
O.D. Inch
I.D.
Inch
Wall
54 56.66 53.96 1.35 57.10 54.00 1.55 57.80 54.00 1.90 58.40 53.94 2.23 60 62.80 60.02 1.39 63.40 60.06 1.67 64.20 60.20 2.00 64.82 60.06 2.38 72 75.34 72.10 1.62 76.00 72.10 1.95 76.88 72.10 2.39 84 87.54 84.10 1.72 88.54 84.10 2.22
March, 2000
Steel, Stainless Steel, P.V.C.
Standard Schedules
Nominal
Pipe Size
Inches
OUTSIDE DIAMETER
1 1.315 1.185 0.065 1.097 0.109 1.049 1.049 0.133 0.957 0.179 0.957 0.179 0.815 0.250
1.25 1.660 1.530 0.065 1.442 0.109 1.380 1.380 0.140 1.278 0.191 1.278 0.191 1.160 0.250
1.5 1.900 1.770 0.065 1.682 0.109 1.610 1.610 0.145 1.500 0.200 1.500 0.200 1.338 0.281 2 2.375 2.245 0.065 2.157 0.109 2.067 2.067 0.154 1.939 0.218 1.939 0.218 1.687 0.344
2.5 2.875 2.709 0.083 2.635 0.120 2.469 2.469 0.203 2.323 0.276 2.323 0.276 2.125 0.375 3 3.500 3.334 0.083 3.260 0.120 3.068 3.068 0.216 2.900 0.300 2.900 0.300 2.624 0.438
3.5 4.000 3.834 0.083 3.760 0.120 3.548 3.548 0.226 3.364 0.318 3.364 0.318 4 4.500 4.334 0.083 4.260 0.120 4.026 0.237 4.026 0.237 3.826 0.337 3.826 0.337 3.624 0.438 3.624 0.438 3.438 0.531 5 5.563 5.345 0.109 5.295 0.134 5.047 0.258 5.047 0.258 4.813 0.375 4.813 0.375 4.563 0.500 4.563 0.500 4.313 0.625 6 6.625 6.407 0.109 6.357 0.134 6.065 0.280 6.065 0.280 5.761 0.432 5.761 0.432 5.501 0.562 5.501 0.562 5.187 0.719 8 8.625 8.407 0.109 8.329 0.148 8.125 0.250 8.071 0.277 7.981 0.322 7.981 0.322 7.813 0.406 7.625 0.500 7.625 0.500 7.437 0.594 7.187 0.719 7.187 0.719 6.183 1.221
10 10.750 10.482 0.134 10.42 0.165 10.25 0.250 10.13 0.310 10.02 0.365 10.020 0.365 9.750 0.500 9.750 0.500 9.562 0.594 9.312 0.719 9.062 0.844 9.062 0.844 8.500 1.125
12 12.750 12.420 0.165 12.39 0.180 12.25 0.250 12.09 0.330 12.00 0.375 11.938 0.406 11.626 0.562 11.750 0.500 11.370 0.690 11.060 0.845 10.750 1.000 10.750 1.000 10.120 1.315 14 14.000 13.50 0.250 13.37 0.315 13.25 0.375 13.25 0.375 13.124 0.438 12.814 0.593 13.000 0.500 12.500 0.750 12.310 0.845 11.810 1.095 11.810 1.095 11.180 1.410 16 16.000 15.50 0.250 15.37 0.315 15.25 0.375 15.25 0.375 15.000 0.500 14.688 0.656 15.000 0.500 14.310 0.845 13.930 1.035 13.560 1.220 13.560 1.220 12.810 1.595 18 18.000 17.50 0.250 17.37 0.315 17.12 0.440 17.25 0.375 16.876 0.562 16.564 0.718 17.000 0.500 16.120 0.940 15.680 1.160 15.250 1.375 15.250 1.375 14.430 1.785 20 20.000 19.50 0.250 19.25 0.375 19.25 0.375 19.25 0.375 18.814 0.593 18.376 0.812 19.000 0.500 17.930 1.035 17.430 1.285 17.000 1.500 17.000 1.500 16.060 1.970 24 24.000 23.50 0.250 23.25 0.375 23.25 0.375 23.25 0.375 22.626 0.687 22.126 0.937 23.000 0.500 21.560 1.220 20.930 1.535 20.930 1.535 20.930 1.535 19.310 2.345
30 30.000 29.37 0.315 29.00 0.500 29.00 0.500 29.25 0.375 29.250 0.375 29.000 0.500 36 36.000 35.37 0.315 35.00 0.500 35.00 0.500 35.25 0.375 35.250 0.375 35.000 0.500 42 42.000 41.25 0.375 41.250 0.375 41.000 0.500 48 48.000 47.25 0.375 47.250 0.375 47.000 0.500
SCH.
5
ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall
SCH. 10
(LTWALL)
SCH. 20 SCH. 30 STD. SCH. 40 SCH. 60 SCH. 140 SCH. 180
X STG. SCH. 80 SCH. 100 SCH. 120
March, 2000
FPS TO GPM CROSS - REFERENCE (Schedule 40)
Nominal
Pipe
(Inches)
1 1.05 2.6989 4.0484 5.3978 6.7473 8.097 9.4462 10.796 12.145 13.490 14.844 16.190 17.540 18.890 20.240 21.590 22.941 24.290
1.25 1.38 4.6620 6.9929 9.3239 11.655 13.99 16.317 18.648 20.979 23.310 25.641 27.970 30.300 32.630 34.960 37.300 39.627 41.958
1.5 1.61 6.3454 9.5182 12.691 15.864 19.04 22.209 25.382 28.555 31.730 34.900 38.070 41.250 44.420 47.590 50.760 53.936 57.109 2 2.07 10.489 15.734 20.979 26.224 31.47 36.713 41.958 47.202 52.450 57.692 62.940 68.180 73.430 78.670 83.920 89.160 94.405
2.5 2.47 14.935 22.402 29.870 37.337 44.80 52.272 59.740 67.207 74.670 82.142 89.610 97.080 104.50 112.00 119.50 126.95 134.41 3 3.07 23.072 34.608 46.144 57.680 69.22 80.752 92.288 103.82 115.40 126.90 138.40 150.00 161.50 173.00 184.60 196.11 207.65
3.5 3.55 30.851 46.276 61.702 77.127 92.55 107.98 123.40 138.83 154.30 169.68 185.10 200.50 216.00 231.40 246.80 262.23 277.66 4 4.03 39.758 59.636 79.515 99.394 119.3 139.15 159.03 178.91 198.80 218.67 238.50 258.40 278.30 298.20 318.10 337.94 357.82 5 5.05 62.430 93.645 124.86 156.07 187.3 218.50 249.72 280.93 312.10 343.36 374.60 405.80 437.00 468.20 499.40 530.65 561.87 6 6.06 89.899 134.85 179.80 224.75 269.7 314.65 359.60 404.55 449.50 494.45 539.40 584.30 629.30 674.20 719.20 764.14 809.09 8 7.98 155.89 233.83 311.78 389.72 467.7 545.61 623.56 701.50 779.40 857.39 935.30 1013.0 1091.0 1169.0 1247.0 1325.1 1403.0
10 10.02 245.78 368.67 491.56 614.45 737.3 860.23 983.12 1106.0 1229.0 1351.8 1475.0 1598.0 1720.0 1843.0 1966.0 2089.1 2212.0
I.D.
INCH
1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9
12 11.94 348.99 523.49 697.99 872.49 1047.0 1221.5 1396.0 1570.5 1745.0 1919.5 2094.0 2268.0 2443.0 2617.0 2792.0 2966.5 3141.0 14 13.13 422.03 633.04 844.05 1055.1 1266.0 1477.1 1688.1 1899.1 2110.0 2321.1 2532.0 2743.0 2954.0 3165.0 3376.0 3587.2 3798.2 16 15.00 550.80 826.20 1101.6 1377.0 1652.0 1927.8 2203.2 2478.6 2754.0 3029.4 3305.0 3580.0 3856.0 4131.0 4406.0 4681.8 4957.2
FPS TO GPM: GPM = (PIPE ID)² X VELOCITY IN FPS X 2.45 FPS X .3048 = MPS GPM TO FPS: FPS =
GPM
(ID)² X 2.45
GPM X .0007 = GPD GPM X 3.7878 = LPM
FPS TO GPM CROSS - REFERENCE (Schedule 40)
Nominal
Pipe
(Inches)
18 16.88 697.52 1046.3 1395.0 1743.8 2093.0 2441.3 2790.1 3138.8 3488.0 3836.3 4185.0 4534.0 4883.0 5231.0 5580.0 5928.9 6277.7 20 18.81 866.14 1299.0 1732.0 2165.3 2598.4 3031.5 3464.6 3897.6 4330.7 4763.8 5196.8 5629.9 6063.0 6496.0 6929.1 7362.2 7795.3 24 22.63 1253.7 1880.0 2507.0 3134.1 3761.0 4387.8 5014.6 5641.5 6268.3 6895.1 7522.0 8148.8 8775.6 9402.4 10029 10656 11283 26 25.25 1560.7 2341.0 3121.0 3901.9 4682.2 5462.6 6243.0 7023.4 7803.7 8584.1 9364.5 10145 10925 11706 12486 13266 14047 28 27.25 1817.8 2727.0 3636.0 4544.5 5453.4 6362.3 7271.2 8180.0 9088.9 9997.8 10907 11816 12725 13633 14542 15451 16360 30 29.25 2094.4 3142.0 4189.0 5236.0 6283.2 7330.4 8377.6 9424.9 10472 11519 12566 13614 14661 15708 16755 17803 18850 32 31.25 2390.6 3586.0 4781.0 5976.5 7171.9 8367.2 9562.5 10758 11953 13148 14344 15539 16734 17930 19125 20320 21516 34 33.25 2706.4 4060.0 5413.0 6766.0 8119.2 9472.4 10826 12179 13532 14885 16238 17592 18945 20298 21651 23004 24358 36 35.25 3041.8 4563.0 6084.0 7604.5 9125.4 10646 12167 13688 15209 16730 18251 19772 21292 22813 24334 25855 27376 42 41.25 4165.4 6248.0 8331.0 10414 12496 14579 16662 18744 20827 22910 24992 27075 29158 31241 33323 35406 37489 48 47.99 5637.8 8457.0 11276 14095 16913 19732 22551 25370 28189 31008 33827 36646 39465 42284 45103 47922 50740 54 53.98 7133.1 10700 14266 17833 21399 24966 28532 32099 35665 39232 42798 46365 49931 53498 57065 60631 64198
I.D.
INCH
1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9
60 60.09 8839.2 13259 17678 22098 26518 30937 35357 39777 44196 48616 53035 57455 61875 66294 70714 75134 79553 72 72.10 12726 19089 25451 31814 38177 44540 50903 57266 63628 69991 76354 82717 89080 95443 101805 108168 114531 84 84.10 17314 25971 34628 43285 51943 60600 69257 77914 86571 95228 103885 112542 121199 129856 138514 147171 155828
FPS TO GPM: GPM = (PIPE ID)² X VELOCITY IN FPS X 2.45 FPS X .3048 = MPS GPM TO FPS: FPS =
GPM
(ID)² X 2.45
GPM X .0007 = GPD GPM X 3.7878 = LPM
GENERAL TERMS AND CONDITIONS OF SALES
1. PAYMENT – Terms of payment are effective from the actual date of invoice. If, in the Seller’s
opinion, the financial condition of the Buyer at any time – or any other circumstances – do not justify the incurrence of production costs of shipment on the terms of payment specified, the Seller may require partial or full payment in advance. Payment terms are net 30 days unless otherwise stated on invoice.
2. F.O.B. – All shipments are from Racine, Wisconsin, USA, unless otherwise other stated, and title
transfers to the buyer upon leaving factory.
3. QUOTATION AND PRICES – Quoted price s are firm for 30 days unless stated in the quotation and
are subject to change without notice after expiration of this period.
4. TAXES – Any applicable sales, use, revenue, excise or other taxes not specifically stated in the
quotation are to be remitted by the Buyer directly to the appropriate regulatory agency.
5. WARRANTY – Seller’s standard published warranty in effect at the time of shipment shall apply.
This warranty is exclusive and is in lieu of all other warranties, express, implied, or statutory, including the warranty of merchantability.
6. DELIVERY – The Seller shall not be liable for loss or damage of any kind resulting from delay or
inability to deliver on account of flood, fire, labor trouble, riots, civil disturbances, accidents, acts or orders or regulations of civil or military authorities, sh ortages of material, or any other causes beyond Seller’s control.
7. PRODUCT CHANGES – In keeping with our continuing policy of product improvement, we reserve
the right to make changes in our products at any time, without incurring an obligation to change, replace or upgrade equipment previously shipped.
8. CANCELLATIONS – An order placed by Buyer and accepted by Seller may be cancelled only with
the Seller’s consent and upon terms that will indemnify the Seller against loss.
9. RESTOCKING CHARGE – On standard equipment, the charge is 25%, provided the equipment is
returned within 30 days in acceptable condition with a RGA number. Restocking charges for special equipment may vary from standard equipment, and will be handled on a case-by-case basis. No returns will be taken after one year.
Limited Warranty and Disclaimer
Dynasonics, div. of Racine Federated Inc. warrants to the end purchaser, for a period of one year from the date of shipment from our factory, that all new transmitters and transducers manufactured by it are free from defects in materials and workmanship. This warranty does not cover products that have been damaged due to normal use, misapplication, abuse, lack of maintenance, or improper installation. Dynasonics’ obligation under this warranty is limited to the repair or replacement of a defective product, at no charge to the end purchaser, if the product is inspected by Dynasonics and found to be defective. Repair or replacement is at Dynasonics’ discretion. An authorization number must be obtained from Dynasonics before any product may be returned for warranty repair or replacement. The product must be thoroughly cleaned and any process chemicals removed before it will be accepted for return.
The purchaser must determine the applicability of the product for its desired use and assumes all risks in connection therewith. Dynasonics assumes no responsibility or liability for any omissions or errors in connection with the use of its products. Dynasonics will under no circumstances be liable for any incidental, consequential, contingent or special damages or loss to any person or property arising out of the failure of any product, component or accessory.
All expressed or implied warranties, including the implied warranty of
merchantability and the implied warranty of fitness for a particular purpose or application are expressly disclaimed and shall not apply to any products sold or
services rendered by Dynasonics.
The above warranty supersedes and is in lieu of all other warranties, either expressed or implied and all other obligations or liabilities. No agent or representative has any authority to alter the terms of this warranty in any way.
RETURN OF EQUIPMENT/SALES INFORMATION
CONTACTS AND PROCEDURES
Customer Service/Application Engineer:
If you have a question regarding order status, placing an order, reviewing applications for future purchases, or wish to purchase a new flowmeter, please contact our new National Sales and Marketing Headquarters:
DYNASONICS
Division of Racine Federated, Inc.
8635 Washington Avenue
Racine, WI 53406
PHONE: (800)535-3569 or
(262)639-6770
FAX: (262)639-2267
Service/Repair Department:
If you already purchased equipment and have an operation problem, require service, or need to schedule field service, please contact our Service Department:
DYNASONICS
Division of Racine Federated, Inc.
8635 Washington Avenue
Racine, WI 53406
PHONE: (800)535-3569 or
(262)639-6770
FAX: (262)639-2267
Return Goods Authorization:
When returning equipment, it is necessary for you to contact our Service Department at (800)535-3569 or (262)639-6770 to obtain an RGA number for the authority and proper tracking of your material and its prompt inspection and return. The RGA number should be noted on the outside of the box. All returns of equipment go to the following address:
DYNASONICS
Division of Racine Federated, Inc.
8635 Washington Avenue
Racine, WI 53406
Attn: RGA #
8635 WASHINGTON AVENUE
RACINE, WI 53406
TOLL-FREE IN NORTH AMERICA.: TEL: (800) 535-3569 FAX: (800) 732-8354 TEL: (262) 639-6770 FAX: (262) 639-2267
URL: www.dynasonics.com
D601 O&M REV 4/00
Loading...