Dynamica Velocity 15HR Instruction Manual

Velocity 15HR
Micro Centrifuge
Instruction manual
Model Velocity 15HR High Speed Micro Centrifuge
Version: 1.0 Revision: D01-201802
Dynamica Velocity 15HR
Refrigerated
Microcentrifuge
Copyright © 2018 Dynamica Scientific Ltd. All rights reserved. No part of this document may be reproduced or transmitted in any form or any means without permission from Dynamica Scientific Ltd.
The names of actual companies and products mentioned herein may be the trademarks of their respective owners.
Safety Reminder
Common safety precautions
The centrifuge is designed to separate liquid-suspended materials having different
densities and particle size.
Carefully read the following safety precautions for a thorough understanding.
Follow the instructions and procedures described in this manual to operate this
centrifuge safely.
Carefully read all safety messages in this manual and safety instructions on your
equipment.
Safety messages are labeled as indicated below. They are in combination with
signal words of “WARNING” and “CAUTION” with the safety alert symbol to call your
attention to items or operations that could be dangerous to you or other persons
using this equipment. The definitions of signal words are as follows:
WARNINGWarning notes indicate any condition or practice, which if not
strictly observed, could result in serious injury or possible death.
CAUTIONCaution notes indicate any condition or practice, which if not
strictly observed or remedied, could result in personal injury, damage or destruction of the equipment.
NOTE:Notes indicate an area or subject of special merit, emphasizing
either the product’s capabilities or common errors in operation or maintenance.
Do not operate this centrifuge in any manner not described in this Instruction
Manual. Should you have any troubles with this centrifuge, call a Dynamica
authorized sales/service representative.
The precautions described in this Instruction Manual are carefully developed in an
attempt to cover all the possible risks. However, it is also important that you are
alert for unexpected incidents. Be careful when operating this centrifuge.
i
Safety Reminder
WARNING
1. This centrifuge is not explosion-proof. Never use explosive or flammable samples or materials that chemically react vigorously.
2. Do not install the centrifuge in or near places where inflammable gases are generated or chemicals are stored.
3. Take all necessary safety measures before using samples that are toxic or radioactive, or blood samples that are pathogenic or infectious. You use such samples at your own responsibility.
1) If the centrifuge, rotor, or an accessory is contaminated by samples that are toxic or radioactive, or blood samples that are pathogenic or infectious, be sure to decontaminate the item according to good laboratory procedures and methods.
2) If there is a possibility that the centrifuge, rotor, or an accessory is contaminated by samples that might impair human health (for example, samples that are toxic or radioactive, or blood samples that are pathogenic or infectious), it is your responsibility to sterilize or decontaminate the centrifuge, rotor or the accessory properly before requesting repairs from a Dynamica authorized sales/service representative.
3) It is your responsibility to sterilize and/or decontaminate the centrifuge, rotor, or parts properly before returning them to a Dynamica authorized sales/service representative.
4. Your centrifuge must be grounded properly.
5. To avoid electrical shocks, do not handle the power cord or turn on or off the POWER switch with wet hands.
6. For safety purposes, do not go within 30 cm of this centrifuge while it is in operation.
7. While the rotor is rotating, never forcedly release the door lock.
8. Repairs, disassembly, and other modifications to the centrifuge are strictly prohibited unless performed by a Dynamica authorized sales/service representative.
ii
Safety Reminder
If you observe some abnormality, contact a
Dynamica
authorized sales/service
CAUTION
1. Be careful not to get your fingers or hands caught between the door hook and the table when closing the door. Do not insert you fingers or an object into the door hook hole.
2. Do not move or relocate this centrifuge while the rotor is rotating.
3. Do not pour any solution such as water, detergent, or disinfectant directly into the rotor chamber. If you do so, the bearings of the drive unit might corrode or deteriorate.
4. Before running this centrifuge, remove any dropped objects and tube fragments from inside the rotor chamber.
5. Prevent dew drops in the centrifuge. When moving the centrifuge from the cold room to the ordinary room, perform a) or b).
a) Warm up the centrifuge by performing the continuous operation for more than 30 minutes. b) Place the centrifuge for more than three hours in a dried room without connecting
the power cord into the inlet of the centrifuge.
6. Cautions on rotors
1) Carefully read the rotor instruction manual before use.
2) Always check for corrosion and damage on the rotor, adapters, and tubes before using them. Do not use the rotor, adapters, or tubes if you find such a problem.
3) Place the rotor in position until it contacts the drive shaft. You should feel a click when the rotor is properly placed on the drive shaft. If you do not feel anything, there might be foreign matter stuck between the rotor and the drive shaft, and the rotor might be tiltedCheck and remove foreign matter if found.
4) Before running this centrifuge, check whether the nut for fixing the rotor to the drive shaft and the rotor cover have been tightened firmly. If the nut has not been tightened sufficiently, the rotor might be improperly installed and come off during operation.
5) Do not run this centrifuge over the allowable maximum speed of the rotor, adapters, and tubes. If their maximum speeds vary, run the centrifuge at the lowest maximum speed among them.
6) Do not exceed the allowable imbalance. For the details, refer to section 3.1 or
the rotor instruction manual.
7) Use the rotor tubes within their actual capacities.
8) Be sure to cap the microtubes if caps are provided.
9) When using filters attached to microtubes, use microtubes specifically designed for use with filters. If you use an ordinary microtube with a filter and the microtube is not capped, the rotor might abnormally vibrate or come off.
10) When storing the rotor on a shelf, make sure that the shelf is secured (for
example, to avoid the rotor from dropping during an earthquake).
11) Check the chemical resistance chart attached to the rotor, and do not use any sample
inapplicable to the rotor, adapters, tubes, tube caps, etc. Using such a sample could corrode or deteriorate such parts.
7. If you observe some abnormality in this product, stop using it immediately and contact a Dynamica authorized sales/service representative. Notify the service representative of the alarm code if displayed.
8. If the centrifuge will not be used for a long time, remove the power cord from the receptacle.
9. Depending on the magnitude, an earthquake might damage the centrifuge.
iii
Contents
Page
Safety Reminder i
1. Structure 1-1
1.1
Location of the component 1-1
1.2
Operation panel 1-2 Standard accessories 1-4
1.3
2. Installation 2-1
3. Preparation 3-1
3.1
List of Applicable Rotors 3-1
3.2
Installing the rotor 3-2
3.3
Preparing samples 3-3
4. Operation 4-1
4.1
Setting the operating conditions 4-1
4.1.1
Setting the speed 4-2
4.1.2
Setting the RCF 4-3
4.1.3
Setting the running time 4-4
4.1.4
Setting the rotor temperature (Velocity 15HR) 4-4
4.1.5
Setting the acceleration and deceleration rate 4-5
4.2
Normal Operation 4-6
4.3
Programmed Operation 4-8
4.4
Spin-down Operation 4-9
4.5
Calculating Relative Centrifugal Force (RCF) 4-10
5. Other Functions 5-1
5.1
Selecting Rotor Stop Signal 5-1
5.2
Indicating the total run hours 5-2
5.3
Actual run timer 5-3
5.4
Rotor condition indicator 5-4
6. Troubleshooting 6-1
6.1
Corrective actions when alarm code is displayed 6-1
6.2
Troubleshooting when no alarm code is displayed 6-3
6.3
Opening the door during power outage 6-4
7. Maintenance 7-1
7.1
Maintenance after operation 7-1
7.2
DefrostingVelocity 15HR 7-2
7.3
Tubes and bottles 7-3
7.4
Service decontamination policy 7-5
8. Specifications 8-1
APPENDIX
Decontamination Sheet
WEEE Compliance Marking for the restriction of the use of hazardous substances in electrical and electronic product (THE PEOPLE’S REPUBLIC OF CHINA)
Rotor Guide Rotor Instruction Manual
Warranty back cover
After-sales Service back cover
Drive Shaft
1. Structure
Door
Operation Panel
Window of the door
reflection type rotation meter.
1.1 Location of the component
Velocity 15HR
Door Hook
Rotor Chamber
Left side
Inlet
POWER switch
Cap
Drain hose
You can measure the rotor speed through the window of the door by a optical
1-1
1. Structure
Deceleration rate
1.2 Operation panel
Velocity 15HR
SPEED/RCF setting button
indicator
Acceleration rate
indicator
ACCEL button
DECEL button
PROG button
SPEED/RCF display
TEMP display
TIME display
DISP/CE button
START button
TIME setting button
PULSE button
OPEN button
⑧ ♪ button
TEMP setting button
STOP button
1-2
1. Structure
Indicates the rate of acceleration from the stop to 1,000 rpm by two
Indicates the rate of deceleration from 1,000 rpm and coasting
F is less
RCF is
Displays an alarm code if any abnormality in the centrifuge is
button,
it
is
button, the value
running and the setting status and fixes the
when the
button, the value
No.
Acceleration rate
indicator
Deceleration rate
indicator
SPEED/RCF display
SPEED/RCF
setting button
TIME display
TIME setting button
Name Function
LEDs.
deceleration state by three LEDs.
(1) Displays a speed or RCF setting or status value.
1)SPEED (speed indicator) The speed is displayed in increments of 100-rpm.
2)RCF(RCF indicator)
The RCF is displayed in increments of 10- xg when RC
than 1,000xg.
The RCF is displayed in increments of 100- xg when
1,000xg or more.
(2)
observed.
Sets the speed or the RCF. When you keep pressing this the value of the speed or the RCF changes rapidly.
(1) Displays the setting or status value of the running time.
( Not sets the time)
The running time is displayed in increments of 1-second when
is below 1 minute.
When the running time is between 1 minute and 99minutes, it
displayed in increments of 1- minute .
“Hd” is displayed when the centrifuge is run continuously.
Sets a running time. When you keep pressing this of the running time changes rapidly.
PROG button
♪ button
⑧ ⑨
ACCEL button Sets an acceleration rate.
DECEL button Sets a deceleration rate.
DISP/CE button
START button Rotor rotation start button
STOP button Rotor rotation stop button
OPEN button
PULSE button
TEMP display
TEMP setting button
Stores and recalls running conditions.
Changes rotor stop signal.
(1) Shifts between the
setting value.
(2) Clears an alarm code.
Door lock release button ( only when the rotor is stopping)
Accelerates the rotor while this button is pressed.
Displays the setting or status value of the temperature. The temperature is displayed in increments of 1 temperature is between -19 and 50.
(When the temperature is –20 or less, it displays “Lo”.)
Sets a temperature. When you keep pressing this of the temperature changes rapidly. Temperature setting range : from “Lo”(–20) to 40.
1-3
1. Structure
1.3 Standard accessories
The standard accessories of the Velocity 15HR centrifuge are listed in Table 1.3. Check
that all the standard accessories are provided before installation.
Table 1.3 Standard accessories for Velocity 15HR
Item name Part No. Q’ty Item drawing
Instruction manual D3-S999931
1
S204746
Power cord*
(Velocity 15HR)
1
(120V)
S204816
(Velocity 15HR)
1
(220-230V)
Tool for setting a rotor S413606
* The provided power cord is different depending on the country.
1
1-4
2. Installation
From 15
℃ to 25℃
15℃
CAUTION
Before moving the centrifuge, the rotor must be removed.
This section describes instructions that you need to follow when installing the centrifuge. The instructions ensure your safety and maximum performance.
(1) Location
Position this centrifuge on a firm and level floor. Avoid installing it on a slippery
floor, or on a floor that conveys vibrations (for example, from people’s footsteps).
This centrifuge can be used at the ambient temperature range of 2°C to 32°C.
We recommend that this centrifuge will be used at the ambient temperature range of 15°C to 25°C to ensure maximum performance. Do not position this centrifuge in direct sunlight.
This centrifuge has an air-cooling system. To secure efficient cooling, keep clearances
of at least 10 cm on both sides of the centrifuge, and at least 30 cm behind it.
Do not install near a heat-generating device or water source from where water
might drip or splash. Such a location might increase the temperature of samples or cause the centrifuge to malfunction.
Do not dew the centrifuge. When moving the centrifuge from the cold room to the
ordinary room, perform a) or b).
a) Warm up the centrifuge by performing the continuous operation for more than 30 minutes. b) Place the centrifuge for more than three hours in a dried room without
connecting the power cord into the inlet of the centrifuge.
On a firm and level floor. A room with a good ventilation
A room that is not exposed to the direct sunlight
Room temperature:
25
Humidity
From 20 to 90
90%
20%
A room with no chemicals and no inflammable gas
A place where water do not drip or splash
Noncondensing
(Be careful when moving the centrifuge
from a cold room to a ordinary room.)
2-1
2. Installation
This section describes instructions you need to follow when installing the centrifuge. This instructions ensure your safety and the maximum performance.
(2) Power source
Velocity 15HR:
120 V instrument: 108-126 V AC, 15 A, 60 Hz
220-230 V instrument: 198-253 V AC, 10 A, 50 Hz ; 198-242 V AC, 10 A, 60 Hz
Install an emergency switch (circuit breaker) that turns off the main power supply
in the event of a malfunction. Optimally, install the emergency switch outside the room or near the exit.
(3) Connection of the power cord
Connect the attached power cord to the inlet of the centrifuge.
Check that the POWER switch is turned off. Then plug in to the outlet.
WARNING
To avoid electrical shocks, do not touch the power cord with wet hands.
Do not hold the cord when disconnecting the power cord from the outlet.
Hold the plug when disconnecting it.
This centrifuge must be grounded properly.
The power cord plug has a three-prong plug. One of the three prongs is provided for grounding. This centrifuge can be grounded just by putting the three- prong plug into a three-prong outlet. Do not connect the three-prong plug to a two-prong outlet using an adapter.
Velocity 15HR
Inlet
Power cord
CAUTION
For connection to a different outlet, the power cord may be needed to be replaced.
Follow local electrical codes.
(4) Removing the packing for transportation from the rotor chamber
Unlock the door by pressing the OPEN button after turning on the POWER
switch. Remove the packing for transportation from the rotor chamber after opening the door.
2-2
3. Preparation
Do not run the centrifuge over the allowable maximum speeds of the rotor,
3.1 List of Applicable Rotors
CAUTION
Only rotors listed in Table 3.1 can be used in the Velocity 15HR centrifuge.
To use the rotor properly, read its instruction manual carefully.
adapters, tubes. Some adapters and tubes cannot withstand the speed of the rotor in use.
Never use any rotor and adapter that is not designated for the centrifuge by
Dynamica.
Check the chemical resistance chart attached to the rotor, and do not use any sample
inapplicable to the rotor, the adapters, tubes, tube caps, etc. Using such a sample could corrode or deteriorate such parts.
If it is necessary to use commercial tubes or adapters other than those
recommended by Dynamica, contact a Dynamica authorized sales/service representative.
Table 3.1 List of applicable rotors of the Velocity 15HR centrifuge
Rotor type
(Shape)
FAS15E
(T15A61)
FAS15F
(T15A62)
FAS15G
(T15A63)
Maximum speed (Maximum RCF)
15,000rpm
(21,500×g)
15,000rpm
(21,500×g)*
15,000rpm
(21,500×g)**
Tube (mL)
1.5mL / 2.0mL X 24tubes 0.2 g/tube
1.5mL / 2.0mL X 24tubes or
0.5mL X 24tubes
0.2mL X 32tubes (8-strip tubes) 0.1 g/tube
Allowable imbalance
Mass difference
g
0.2 g/tube***
*The FAS15E rotor used with 0.5mL tube has a maximum RCF of 18,100xg.
**The FAS15G rotor has a maximum RCF of 21,500xg when a tube is loaded in the
corner tube cavity of the rotor.
***Allowable imbalance is 0.2g / tube when the same kinds of tubes are loaded in the
rotor. Allowable imbalance is 0.1g / tube when both the 1.5/2 mL micro tubes and the
0.5ml micro tubes are loaded in the rotor at the same time.
3-1
3. Preparation
3.2 Installing the rotor
(1) Check the rotor for corrosion or crack before use. The rotor is corroded if the
surface is discolored, dented or cracked.
CAUTION
If any abnormality such as corrosion or crack is found, stop using the rotor and
contact a Dynamica authorized sales/service representative.
(2) Remove the rotor cover from the rotor and check that there is no foreign matter in tube
cavities.
(3) Place the rotor body on the drive shaft without the rotor cover. Then hold the rotor body
to avoid moving the rotor body and tighten the nut with the tool for setting rotor.
There is the mark in the side of the nut. If you can see the mark after you finish tightening the nut, the nut has not been tightened sufficiently. Tighten the nut sufficiently until you can not see the mark on the side of the nut.
Proper setting:
If you can not see the
mark after you finish tightening the nut, the nut has been tightened sufficiently.
Nut
Tool for setting the rotor
Rotor
Improper setting:
If you can see the mark after you finish tightening the nut, the nut has not been tightened sufficiently.
CAUTION
Place the rotor in position until it contacts the drive shaft. You should feel a click when
the rotor is properly placed on the drive shaft. If you do not feel anything, there might be foreign matter stuck between the rotor and the drive shaft, and the rotor might be tiltedCheck and remove foreign matter if found.
When the rotor is installed improperly, the rotor may abnormally vibrate or come off .
If the nut is tightened insufficiently, the rotor may stick in the drive shaft. As the result, the rotor may not be able to be removed from the drive shaft. Alarm E20 and Alarm E21 may be displayed when the rotor is installed improperly. These alarm codes are displayed when the drive shaft is rotated at high speed without the rotor. When these alarm codes are displayed, turn off the POWER switch to clear the alarm code.
Be careful not to injure your fingers or hands with the threaded portion of the rotor
when holding the rotor.
Mark
Drive shaft
3-2
3. Preparation
3.3 Preparing samples
(1) Prepare the sample to be centrifuged.
WARNING
This centrifuge and the rotor are not explosion-proof. Never use explosive or
flammable samples.
There are restrictions on the usage of biological samples and radioactive substances
that require biological isolation such as pathogens and recombinant DNA for safety purposes. User must prepare necessary safety measures at user’s responsibility before treating samples containing such substances.
CAUTION
Be careful that imbalance operation may occur in the following cases. Fill the
same sample in the tubes/bottles and load them in the buckets that are placed symmetrically with respect to the drive shaft in the rotor.
If samples that are equal in volume but different in composition are used, the
precipitation levels may be different by centrifugation and such operation may increase the level of imbalance.
If samples that are equal in weight but different in volume (density) are used or if
the tubes/bottles are different in inside diameter or shape, there may be variations in position of center of gravity and such operation may cause imbalance.
(2) Inject the sample into the tubes.
Balance the tubes which are loaded into the rotor symmetrically with respect to the
center of rotation of the rotor.
CAUTION
Use the rotor tubes within their actual capacities. Depending on the kind of tubes
used, a sample might leak from the gap between a tube and its cap if the tube is full.
Do not exceed the allowable imbalance. For the details, refer to section 3.1 or the
rotor instruction manual.
Never wittingly run under unbalanced condition even though the allowable
imbalance is not exceeded.
3-3
3. Preparation
(3) Load the tubes into the tube cavities of the rotor.
Load the tubes which are balanced symmetrically in the step (2) into the tube
cavities.
Example of correct tube settings and incorrect tube settings
Correct settings:
Incorrect settings:
CAUTION
Be sure to cap the micro tubes if caps are provided, otherwise it may cause the
contamination of the sample with dust or water.
When using filters attached to micro tubes, use micro tubes specifically designed
for use with filters. If you use an ordinary micro tubes with a filter and the micro tube is not capped, the rotor might abnormally vibrate.
rotor
Tube
3-4
4. Operation
4.1 Setting the operating conditions
WARNING
Never open the door while the rotor is rotating or touch the rotating rotor.
For your purposes, do not go within 30 cm of this centrifuge while it is in operation.
CAUTION
Do not place any object or your hands on the centrifuge.
Do not push or lean against the centrifuge while it is running.
Do not place containers holding liquid on or near the centrifuge. If they spill,
liquid may get into the instrument and damage electrical or mechanical components.
Do not run the centrifuge with fragments of tubes or dew drops left in the rotor
chamber. The foreign matter might get mixed with samples or might become the cause of rise in the rotor retention temperature. Keep the interior of the rotor chamber always clean.
When operating the rotor at high speed after turning on the POWER switch in the
Velocity 15HR centrifuge, the temperature of the rotor chamber might increase temporarily. However something is not wrong with the Velocity 15HR. To cool the rotor chamber completely, wait for at least three minutes after turning on the POWER switch.
If this centrifuge makes peculiar sound during its operation, stop it immediately
and contact a Dynamica authorized sales/service representative.
NOTE
The setting value can be indicated by pressing DISP/CE button. By pressing it
again, the status value can be indicated.
When changing the setting value, press the button or button (These
buttons are placed at the lower of the display) to blink the indication. If three seconds elapses without operation, the indication does not blink and the centrifuge does not accept the intended value. Press the button or button again to blink the indication.
4-1
4. Operation
100rpm
100rpm
4.1.1 Setting the speed
Check whether LED (color:green) at the right of the SPEED/RCF display has been
extinguished. When the LED has been extinguished, perform the step .
When the LED described in step lights, shift the mode from the RCF indication
to the speed indication by pressing the button and button (of SPEED/RCF setting) for 0.5 seconds or longer simultaneously (see the below). Then the LED is extinguished and you can hear the short electronic sound.
Press the ▲ button or ▼ button once at the lower of the SPEED/RCF display.
SPEED/RCF display blinks which indicates that the centrifuge is in the speed
input waiting mode.
Press the button or ▲ button to set the intended speed.
When pressing the button continuously, the value of the speed changes rapidly.)
×
×100·g
extinctionthe mode of the speed indication lightingthe mode of the RCF indication
The mode is shifted by pressing two buttons simultaneously for 0.5 seconds or longer.
×
×100·g
■ When pressing the button, the value changes as follows. ▼ button:150 → 149 →‥‥ → 4 → 3 → 150 → 149 → ‥‥ ▲ button:3 → 4 → 5 →‥‥→ 149 → 150 → 3 → 4 → ‥‥
The speed is set in 100-rpm intervals.
Confirm the entered setting by pressing DISP/CE button.
Even if neglecting to press DISP/CE button, the entered setting is confirmed after
about three seconds.
4-2
4. Operation
100rpm
100rpm
4.1.2 Setting the RCF
Check whether LED (color:green) at the right of the SPEED/RCF display has been
lit. When the LED has been lit, perform the step .
When the LED described in step is extinguished, shift the mode from the speed
indication to the RCF indication by pressing the button and button (of SPEED/RCF setting) for 0.5 seconds or longer simultaneously (see the below). Then the LED is extinguished and you can hear the short electronic sound.
Press the button or button once at the lower of the SPEED/RCF display.
SPEED/RCF display blinks which indicates that the centrifuge is in the RCF input
waiting mode.
Press the button or ▲ button to set the intended RCF.
When pressing the button continuously, the value of the RCF changes rapidly.)
×
×100·g
extinctionthe mode of the speed indication lightingthe mode of the RCF indication
The mode is shifted by pressing two buttons simultaneously for 0.5 seconds or longer.
×
×100·g
■ When pressing the button, the value changes as follows. ▼ button:215 → 214 →‥‥ → 0.2 → 0.1 → 215 → 214 → ‥‥
▲ button:0.1 → 0.2 → ‥‥→ 214 → 215 → 0.1 → 0.2 → ‥‥
The RCF which is less than 1,000xg is set in 10xg intervals and the RCF which is
1,000xg or more is set in 100xg intervals. “0.1” is displayed when the value of RCF is 10xg, because the indicated unit in this centrifuge is x100g.
Confirm the entered setting by pressing DISP/CE button.
Even if neglecting to press DISP/CE button, the entered setting is confirmed after
about three seconds.
4-3
4. Operation
manufacturer. You can choose one between two (see the section 5.3).
min
T EM P
T IME
4.1.3 Setting the running time
NOTE
This centrifuge has two kinds of timers. One is “Normal timer mode”, and the other is
“Actual run timer mode”. When you choose “Normal timer mode”, the run time is counted as soon as the rotor is accelerated. When you choose “Actual run timer mode”, the run time is counted as soon as the speed of the rotor reaches the setting speed. “Normal timer mode” has been chosen in your centrifuge by the
Press the button or button once at the lower of the TIME display.
TIME display blinks which indicates that the
centrifuge is in the time input waiting mode.
Press the button or button to
set the intended running time.
When pressing the button continuously,
the value of the running time changes rapidly.
■ When pressing the button, the value changes as follows. ▼ button:Hd →99 → 98 →‥‥ → 2 → 1 → Hd → 99 → 98 →‥ ▲ button:1 → 2 → 3 → ‥‥ → 98 → 99 → Hd → 1 → 2 →‥
The running time you can set is from 1 to 99 minutes. The running time is
set in 1-minute intervals. (The continuous running is indicated as “Hd”.)
Confirm the entered setting by pressing DISP/CE button.
Even if neglecting to press DISP/CE button, the entered setting is
confirmed after about three seconds.
min
NOTE
The following describes the condition that TIME display of a status value
indication indicates the running time by the second.
1) When the elapsed time is less than 1 minute at the continuous running
2)When the remaining time is less than 59 seconds
A dot stays lit at the lower right of number when TIME display indicates the running time by the second.
4.1.4 Setting the rotor temperature
Press the button or button once at the lower of the TEMP display.
TEMP display blinks which indicates that the
centrifuge is in the temperature input waiting mode.
Press the button or button to
set the intended temperature time.
When pressing the button continuously,
the value of the temperature changes rapidly.
■ When pressing the button, the value changes as follows. ▼ button:40 → 39 → 38 → ‥‥ → -18 → -19 → Lo → 40 →‥ ▲ button:Lo → -19 → -18 → ‥‥ → 39 → 40 → Lo → -19 →‥
The temperature you can set is from Lo(-20) to 40℃. The temperature is
set in 1- intervals. (Depending on the room temperature, the rotor temperature might not reach the set temperature.)
Confirm the entered setting by pressing DISP/CE button.
Even if neglecting to press DISP/CE button, the entered setting is
confirmed after about three seconds.
4-4
4. Operation
SLOW
1,000
FAST
FAST
SLOW
4.1.5 Setting the acceleration and deceleration rate
You can select acceleration and deceleration curves tailored to your jobs from two acceleration stages (FAST, SLOW), three braked deceleration stages (FAST, SLOW, FREE (coasting deceleration)). Acceleration and deceleration control has the time from 0 to 1,000 rpm variable, excluding the coasting deceleration. The chart below depicts the relationships between the acceleration and deceleration times against the speed.
Speed (rpm)
Figure 4.1.5 Relationships between acceleration and deceleration rates and curves
To collect precipitate that is difficult to whirl up, set acceleration and deceleration
rate “FAST” to efficiently separate it.
To avoid a disturbance during acceleration or deceleration, choose “SLOW” or
“FREE”.
[Procedure for setting the acceleration and deceleration rate]
FREE(Coasting deceleration)
Time
Press the ACCEL button once to blink the LED of the
acceleration rate indicator.
Press the ACCEL button again to blink the intended LED
of the acceleration rate indicator.
The blinking LED switches from “FAST” to “SLOW” alternately
by pressing the ACCEL button repeatedly.
Press the DECEL button once to blink the LED of the
deceleration rate indicator.
Press the DECEL button again to blink the intended LED
of the deceleration rate indicator.
By pressing the DECEL button repeatedly,
the blinking LED changes as follows.
FASTSLOWFREEFAST
Confirm the entered setting by pressing DISP/CE
button.
Even if neglecting to press DISP/CE button, the entered
setting is confirmed after about three seconds.
4-5
4. Operation
the table when closing the door.
4.2 Normal Operation
(1) Turn on the POWER switch.
All indicators on the operation panel stay lit for about 3 seconds for self-checking.
NOTE
As soon as the power switch is turned on, the function of self-checking works.
Any button does not work during self-checking.
(2) Open the door and then set tubes into the rotor (see Chapter 3).
Door lock is released by pressing the OPEN button while the door is locked.
(Even if pressing the OPEN button, the door lock can not be released when the
rotor is rotating or the door is open.)
(3) Close the door to start running.
CAUTION
Do not insert your fingers or a object into the door hook hole. If the door lock
functions by inserting your fingers into the door hook hole, you might injure your fingers with the door lock of this centrifuge. Malfunction of this centrifuge might be caused by inserting a object into the door hook hole.
Be careful not to get your fingers or hands caught between the door hook and
Velocity 15HR (the right side of the centrifuge)
When closing the door of the centrifuge like pressing the door lightly, the door is
locked automatically. If you need to open the door again, press the OPEN button
to unlock the door.
4-6
4. Operation
S TART
S TO P
S TART
S TO P
(4) Set the operating condition (see section 4.1).
CAUTION
Some tubes that are sold on the market have lower allowable speeds (allowable
RCF) than the rotor in use. Use them at the lowest allowable speed or less.
(5) Start the operation.
Press the START button.
The START LED starts blinking.
When it reaches the set speed,
START LED stops blinking and stays lit.
(6) End the operation.
Wait until the TIME indicator shows “0” or press
the STOP button.
As the rotor starts decelerating,
the STOP LED starts blinking.
When the rotor stops, the STOP LED stops blinking and stays lit. Stop signal
starts sounding. If you select “no tune”, all indicators blink upon halt of the rotor.
In the Velocity 15HR centrifuge, the door lock is not released automatically to
prevent dew drops. When using Velocity 15HR, press the OPEN button to unlock the door.
(7) Open the door gently and take out the sample.
Do not open the door roughly, otherwise the separated sample might be disturbed.
(8) Turn off the POWER switch.
In the Velocity 15HR centrifuge, turn off the POWER switch with its door open to dry
the rotor chamber at the end of daily operation. If the Velocity 15HR centrifuge will not be used for a long time, turn off the POWER switch with its door open as well as the above.
4-7
4. Operation
4.3 Programmed Operation
This centrifuge is featured with programmed operation that is able to store three running conditions in its memory and recall the condition by simply pressing a button. By using
this feature, you are able to save a step for setting conditions.
(1) Programming run conditions
Turn on the POWER switch and set a desired condition to be programmed according
to section 4.1.
Choose one of three PROG buttons and press it for 3 seconds or longer.
NOTE
When programming a new condition, do not release the PROG button until you
can hear the electronic sound. Otherwise, the newly programmed condition is canceled and the preprogrammed condition is recalled.
You can hear the short electronic sound when
pressing the PROG button. Keep pressing it for about 3 seconds or longer to complete the programming the process.
(2) How to perform a programmed operation
Preparing the operation (see Chapter 3). Press the desired PROG button (one of the three buttons) and release it
immediately.
NOTE
Do not keep pressing the PROG button for a long time when recalling the
programmed condition. Otherwise programmed condition is canceled.
You can hear the short electronic sound when
pressing the PROG button, and the programmed condition is displayed on the corresponding display.
The selected PROG LED stays lit during
a programmed operation.
Press the START button to start the operation.
4-8
4. Operation
4.4 Spin-down Operation
This feature is useful to remove the adhered samples on the interior walls of the tubes.
NOTE
In the spin-down operation, the speed is at the maximum for acceleration and
deceleration, regardless of the settings.
The centrifuge accelerates up to the set speed. This is not the maximum speed of
the rotor in use.
If you release the PULSE button during acceleration, the centrifuge discontinues
acceleration and starts deceleration.
The PULSE button works only while the rotor is stopped.
Turn on the POWER switch. Check the speed of the rotor and change the setting as
necessary (see section 4.1).
Open the door and then set tubes into the rotor (see Chapter 3).
Close the door to start running.
Press the PULSE button.
The rotor continues accelerating while pressing this button, and when the
centrifuge reaches the set speed, it continues to operate.
PULSE
Release the PULSE button.
The rotor starts decelerating by releasing the button. Upon halt of the rotor, the
door lock is released automatically in the Velocity 15HR centrifuge, and the door is opened to a few cm automatically.
PULSE
4-9
4. Operation
4.5 Calculating Relative Centrifugal Force (RCF)
An RCF can be determined by calculation or by referring to the graph provided below. The rotating radius of each rotor is given in the rotor instruction manual.
Calculation formula: RCF = 1,118 x Rotating radius (cm) x Rotating speed 2 (rpm) x 10-8
Calculation graph is shown below. The point of intersection between the rotating radius and the speed indicates the corresponding RCF. The example shows an RCF of 1,500xg with a rotating radius of 15 cm and a speed of 3,000 rpm.
Rotating radius
(cm)
RCF (xg)
Example
Speed (rpm)
4-10
5. Other Functions
5.1 Selecting Rotor Stop Signal
The rotor stop signal can be selected from seven kinds of sounds including five tunes, electronic beep, and no sound.
Currently selected sound starts as the button is pressed. Press again before the sound ends to move to the next sound and the last played sound is chosen as the rotor stop signal.
NOTE
The rotor stop signal can be selected only while the rotor is stopped.
“Amaryllis” starts
“My Grandfather’s Clock ” starts.
“Twinkle, Twinkle, Little Star” starts.
“For Elise” starts.
When the rotor is operated under this condition, the indicators blink upon the rotor stop.
When the rotor is operated under this condition, the beeper sounds two beeps upon the rotor stop.
“The Jolly Finale”
starts.
No sound
The beeper sounds a beep only at the setting. No sound is heard at the rotor stop if this is selected.
The beeper sounds two
beeps.
The tune stops by pressing any button on the operation panel while the selected
tune continues.
When no sound is selected, the indicators blink upon the rotor stop instead of a
tune or beep. By pressing any button on the operation panel, the indicators stop blinking and stays lit.
5-1
5. Other Functions
5.2 Indicating the total run hours
The total run hours can be indicated in this centrifuge. The total run hours can be
displayed on the operation panel. Inspect the centrifuge periodically by referring it.
The total run hours of the centrifuge includes the time for which the rotor reaches
the set speed whether you select the actual run timer or not.
(1) Checking the total run hours
Press the ACCEL button while pressing button to display the total run
hours. You can check the total run hours whether the rotor is stopped or not.
The total run hours is displayed on the SPEED/RCF display and the time display.
The integer part (four-digits) of the total run time and the decimal part (one-digit) of it blink alternately. When the decimal part is displayed, the unit (hr) of the total run time is also displayed.
ex.)123.1 hours
(2) How to finish indicating the total run hours
Return the normal display by selecting one from the following.
• Wait for 15 seconds.
• Press one button on the operation panel.
• Turn off the POWER switch and turn on the POWER switch again.
(2) Procedure for clearing (resetting) the total run hours
Press the DISP/CE button for three seconds when the total run hours is
displayed. Then the following are displayed and blinking.
Press the START button within 10 seconds to clear (reset) the total run hours.
The total run hours is displayed again if you do not press any button or you press a button except the START button.
TIME TIME SPEED / RCF SPEED / RCF
blinking alternatively
blinking
SPEED / RCF
5-2
5. Other Functions
TIME
5.3 Actual run timer
This centrifuge has two kinds of timers. One is normal timer and the other is actual
run timer. The time for which the rotor reaches the set speed varies depending on rotor types or your samples. Actual run timer is effective to get stable separation of your samples because as soon as the rotor reaches the set speed the actual run timer works.
After you have chosen between the normal timer and actual run timer, this is
registered in this centrifuge even if the POWER switch is turned off. Follow the below procedures when checking or changing the timer mode.
(1) Checking the timer mode
Press the button and button (under TIME display) for 0.5 seconds or longer
simultaneously (see the below). Then the setting mode is displayed and you can hear the short electronic sound.
The original indication is displayed when you press any button except TIME
setting button.
TIME
The setting mode is displayed by pressing two buttons simultaneously for 0.5 seconds or longer.
TIME
Normal timer Actual run timer
(2) Changing the timer mode
After displaying the setting mode (see the step (1)), press the button or
button (under TIME display) to display the intended timer mode.
The original indication is displayed when you press any button except TIME
setting button.
NOTE
When you choose the actual run timer, the right figure
Normal timer mode is chosen in the factory.
You can not check and change the setting mode during operation.
shows the TIME display at acceleration time.
TIME
5-3
5. Other Functions
TIME TIME
5.4 Rotor condition indicator
When you choose the rotor condition indicator, you can see easily that the rotor is
rotating at the set speed. After the rotor reaches the set speed, the run time and the rotor condition indicator are displayed by turns.
ex.) When run time is 10 minutes
After you have chosen the rotor condition indicator, this is registered in this
(1) Checking the rotor condition indicator
Press the DECEL button while pressing button to display the setting
TIME
centrifuge even if the POWER switch is turned off. Follow the below procedures when checking or changing the indication.
indication on the TIME display.
displaying by turns
TIME
LED for the TIME display follows the arrows.
The original indication is displayed when you perform either of the following.
Normal indication Rotor condition indication
• You press any button except TIME setting button.
• You do not press TIME setting button for three seconds.
(2) Procedure for clearing (resetting) the total run hours
After displaying the setting indication (see the step (1)), press the button or
button (under TIME display) to display the intended indication.
The original indication is displayed when you press any button except TIME
setting button.
NOTE
Normal indication is chosen in the factory.
Rotor condition indication is not displayed in the following cases, because the
unit of the rotor condition indication is “minute”.
• The first one minute when the “HOLD” is selected as the run time setting.
• The last one minute when the “HOLD” is not selected as the run time setting.
5-4
6. Troubleshooting
The centrifuge stops the rotor upon
power supply voltage.
The door is opened while the rotor
might be
Lower the room temperature. If there
is a heat source nearby, move the
r inlet of the centrifuge is
Keep clearances of at least 10 cm
The cap at the tip of the drain
The allowable imbalance is
ng the
The centrifuge is shocked during
An alarm code may be indicated when
the centrifuge is shocked during
Alarm codes to report control
authorized
Remember to provide the alarm code
6.1 Corrective actions when alarm code is displayed
WARNING
Unspecified repairs, remodeling or disassembly of the centrifuge that is not listed
below is strictly prohibited other than a Dynamica authorized service representative.
Close the door immediately if the rotor is still rotating when the door is opened after
releasing the alarm status.
An alarm code is indicated on the SPEED/RCF display of the operation panel with a beep if any abnormality is found in the centrifuge. Take the corrective actions specified in the table below.
Alarm code Probable cause Action
E 1
POWER
E 2
DOOR
E 4
TEMP
E 9
IMBALANCE
E 10
E 99
While the rotor was rotating, a
power failure occurred or power supply voltage was decreased.
The START button is pressed while the door is open.
is rotating. ( E-48 indicated.)
Cooling ability is degraded.
The ai
blocked
hose removed.
exceeded.
operation.
errors.
For the E13 , E15 , E20 , E21 ,
refer to (1).
recovery from the power failure. Retry operation as needed.
Contact a power company if
abnormality is found about the
Press the START button after
closing the door.
WARNING:
Never unlock the door while the rotor is rotating.
centrifuge away from it.
from the air inlet of the centrifuge.
Put the cap on.
Balance the sample keepi
allowable imbalance.
operation.
Contact a Dynamica
sales/service representative.
and the situation.
6-1
6. Troubleshooting
(1) When an alarm code is displayed
Alarms E-1 to 9 are mainly related to handling errors. You can continue using
the centrifuge if the cause is removed. If alarm codes can not be cleared by taking the proper action, contact a Dynamica authorized sales/service representative.
Alarm E-13 indicates a rotation sensor error. Once this alarm code is displayed,
the alarm code cannot be cleared and the door cannot be opened for six minutes – a time that allows all the rotors to come to a halt with the coasting deceleration.
Wait at least six minutes keeping the power on and press the DISP/CE button
to remove the rotor.(When turning off the POWER switch, you can not clear the alarm code for six minutes after turning on the POWER switch.)
Alarm E-15 indicates a motor over-temperature. When the room temperature is
high or the rear of the centrifuge is covered, this alarm code is displayed even if the centrifuge is not out of order. Lower the room temperature or keep clearances of at least 30 cm behind the centrifuge. And then clear the alarm code.
Alarm E20 and E21 are indicated when the drive shaft is rotated at high speed
without the rotor. When these alarm codes are indicated, turn off the POWER switch to clear the alarm code.
(2) How to clear alarm codes
Alarm codes can be cleared by pressing the DISP/CE button after removing
the cause.
When the alarm code is displayed during operation, the door is not unlocked
automatically even though the rotor has stopped. After clearing the alarm code, unlock the door by pressing the OPEN button.
If alarm codes can not be cleared by taking the proper action, contact a
Dynamica authorized sales/service representative.
6-2
6. Troubleshooting
6.2 Troubleshooting when no alarm code is displayed
If the centrifuge cannot operate properly even though no alarm code is displayed, take the corrective actions specified in the table below.
Symptom Probable cause Action
Nothing appears on the display when the POWER is turned on.
Building power circuit breaker trips or the fuse is blown out while it is running.
The centrifuge does not accept the entries on
the operation panel.
Building power circuit breaker trips or the fuse is blown out.
Power is wired from a single outlet to multiple devices.
The centrifuge may be out of order.
The centrifuge may be out of order.
Remove the cause of the trouble and turn on the circuit breaker or replace the fuse. If alarm codes can not be cleared by taking the proper action, contact a Dynamica authorized sales/service representative. The power capacity may be exceeded. Supply power for this centrifuge and other devices from different sources. Set the rotor and start operation to check for successful rotation (look at the SPEED/RCF display). If normal, continue using the centrifuge.
CAUTION :Turn off the power if any
abnormality is found on the
centrifuge and contact a Dynamica authorized sales/service representative.
When the rotor does not stop even if pressing the STOP button, turn off the POWER switch and turn on the switch again. Take out the rotor after the rotor stops. Contact a Dynamica authorized sales/service representative.
6-3
6. Troubleshooting
complete stop.
6.3 Opening the door during power outage
WARNING
Opening the door while the rotor is rotating is very dangerous. Never unlock the
door while the rotor is rotating.
The door cannot be opened except when the centrifuge is powered on (POWER switch on) and the rotor is stopped. If the door cannot be opened due to a power outage, open the door according to the following steps. (Below are the illustrations about the Velocity 15HR centrifuge.)
(1) Check that the rotor is not rotating.
WARNING
Before opening the door, wait more than six minutes until the rotor comes to a
Observe the rotor through the window of the door and check the halt of the rotor.
Listen carefully to make sure that no rotating sound is heard, because you might think that the rotor does not rotate although the rotor rotates at a high speed. As it might be difficult to observe the rotor through the window of the door due to dew- drops and frost, listen carefully to make sure that no rotating sound is heard.
Touch the centrifuge carefully to make sure that vibration is not felt on the
centrifuge.
(2) Remove the plug after turning off the POWER switch.
WARNING
Be sure to turn off the POWER switch before performing this work. Otherwise it may
cause electrical shock hazards and the centrifuge may be damaged.
There is a plug at the right side of the centrifuge.
Window of the door
Hole for unlocking the door
Plug
6-4
6. Troubleshooting
(3) Insert the attached tool for setting the rotor horizontally into the hole for unlocking
the door to fit the tool for setting rotor into the shaft inside the centrifuge.
NOTE
You can not see the shaft for unlocking the door, because it is inside the centrifuge. Insert
the attached tool for setting the rotor horizontally until the tool contacts the shaft. Then fit the tool for setting rotor into the shaft for unlocking the door by turning the tool clockwise or counterclockwise slowly.
Window of the door
(4) By turning the tool for setting the rotor at 180°, the door is unlocked automatically in
the centrifuge and the door is opened to a few cm automatically.
After the door is opened, remove the tool for setting the rotor. Then be sure to
position the plug into the hole for unlocking the door. It is not necessary to place the shaft to the original position.
(5) Open the door and take out the samples.
WARNING
When you first turn on the POWER switch after you unlock the door as described
above, the door lock might work automatically to adjust it’s position. Do not insert your fingers or an object into the door hook hole. Otherwise your fingers might be injured or the centrifuge might be damaged.
Be sure to remove the tool for setting the rotor from the centrifuge if it is not
necessary to inset it to the centrifuge. It is very dangerous because the tool contacting the shaft rotates if the tool is being inserted into the centrifuge when you close the centrifuge door after turning on the POWER switch.
Tool for setting the rotor
6-5
7. Maintenance
corrosion or deterioration of the centrifuge.
7.1 Maintenance after operation
CAUTION
Using cleaning or sterilization methods other than recommended in this instruction
manual may cause in corrosion or deterioration of this centrifuge. Refer to chemical resistance chart attached to the rotor or contact Dynamica.
Unplug the main power cord before cleaning or sterilizing the instrument.
(1) Centrifuge
CAUTION
If the centrifuge is exposed to ultraviolet rays for a long time, the color of the covers
may be changed or the coating may be peeled off. After use, cover the centrifuge with a cloth to protect it from direct exposure.
Do not sprinkle water, neutral detergent or disinfectant solution directly on the
centrifuge. Otherwise fluids might leak into the inside of the centrifuge from the door hook hole and it might cause corrosion or deterioration of the centrifuge.
If the centrifuge is heavily stained, clean it with a cloth or sponge moistened with a
dilute solution of neutral detergent. Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution.
(2) Rotor chamber
CAUTION
Do not expose the rotor chamber to acids, salts, etc. Use of these materials can cause
a chemical reaction that initiates corrosion such as rust. When samples contain these materials, remove the rotor from the centrifuge after the operation and clean the rotor chamber thoroughly.
Do not pour water, neutral detergent or disinfectant solution directly into the rotor
chamber. Otherwise fluids may leak into the inside of the centrifuge and cause
Wipe frost or moisture in the rotor chamber with a soft cloth to prevent getting in the
sample or corrosion. If the rotor chamber is dirty, clean with a cloth or sponge moistened with a neutral detergent solution. Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution.
When the tubes are broken, remove the rotor and then clean the bottom of the rotor
chamber, because there might be the fragments of the tubes or the sample on the bottom of the rotor chamber.
(3) Drive shaft
Wipe the drive shaft with a soft cloth if the surface is dirty.
To prevent scuffing of the rotor, apply a thin coat of lubricant for screws to the drive shaft once a month (purchase the lubricant for screw (Part No. 84810601) separately).
(4) Door
Clean and sterilize the door in the same manner as specified in the step (1).
The door packing may have been deformed if air leaks through the door or the door
rattles. Contact a Dynamica authorized sales/service representative for an inspection.
(5) Rotor
For more details, refer to the rotor instruction manual.
7-1
7. Maintenance
7.2 DefrostingVelocity 15HR
When the rotor chamber is frosted, the cooling ability might be degraded. Defrost the
rotor chamber by performing the following procedures.
(1) Open the door of the centrifuge and turn off the POWER switch.
(2) Hold up the left side of the centrifuge a little. Then remove two clamps on the bottom
of the centrifuge and pull the drain hose.
Cap
Clamp Drain hose
(3) Remove the cap of the tip of the drain hose and put the drain hose in a beaker etc.
for drain.
(4) Leave centrifuge with the door open until you observe that the rotor chamber is
defrosted.
(5) After completing defrosting the rotor chamber, cap the drain hose and place the drain
hose to the original position.
Remove the cap
Pull the drain hose
NOTE
Be sure to cap the drain hose during the operation. The cooling ability degrades if
you operate this centrifuge whose drain hose is not capped.
7-2
7. Maintenance
7.3 Tubes and bottles
(1) Cleaning and sterilizing tubes and bottles
Use the best method for cleaning and sterilizing tubes and bottles referring to the table below.
Cleaning and sterilizing conditions for tubes and bottles: Applicable ×: Inapplicable
Condition
Acidic (pH5 or lower) × × × Acidic (higher than pH5)
Cleaning fluids
Cleaning
Alkaline (higher than pH9) × Alkaline (pH9 or lower) Neutral (PH7) Warm water (up to 70°C)
Ultrasonic cleaning
Neutral detergent (pH7)
Autoclaving 115°C(0.7kg / cm2)30 minutes ○
121°C(1.0kg / cm2)20 minutes × ○ 126°C(1.4kg / cm2)15 minutes × × ×
Sterilization
Boiling 15 to 30 minutes Ultraviolet sterilization
Gas sterilization
200 to 300 nm × × ×
Ethylene oxide × Formaldehyde
PA: Polypropylene copolymer PC: Polycarbonate PP: Polypropylene
(2) Cleaning PC tubes and bottles
PC materials have low in chemical resistance to alkaline solutions. Avoid using neutral detergents higher than pH9. Note that some neutral detergents are still higher than pH9 even if diluted according to the instruction in the maker’s catalog. Use detergents between pH7 and pH9.
(3) Autoclaving PA, PC and PP tubes and bottles
PA begins softening at about 120°C, and PC and PP at about 130°C. Autoclave PA tubes/bottles at 115°C (0.7 kg/cm2) for 30 minutes and PC and PP tubes/bottles at 121°C (1.0 kg/cm2) for 20 minutes. If a certain temperature is exceeded, the tubes/bottles may be deformed. Observe the following instructions when using a sterilizing chamber:
Place bottles in a vertical position, mouths facing up. If bottles are placed sideways, they
may deform into an oval shape due to their own weight.
Remove screw caps and inner covers to prevent deformation or rupture.
Wait until the sterilizing chamber cools down to room temperature before removing the
bottles.
Material
PA PC PP
7-3
7. Maintenance
Cleaning and
(4) Conditions and life expectancy of tubes and bottles
CAUTION
Do not use tubes/bottles that have exceeded their life expectancy. Failure to do so
could result in damage of tubes/bottles and the rotor and the centrifuge. The life expectancy of tubes/bottles depends on factors such as the characteristics of samples, speed of the rotor used, and temperature. Always check for deterioration and damage (cracks, deformation, and so on) on tubes/bottles before using them. Do not use the tubes/bottles if you find such a problem.
The life expectancies of plastic tubes and bottles depend on factors such as the characteristics of samples, speed of the rotor used, and temperature. When plastic tubes/bottles are used for centrifugation of ordinary aqueous samples (between pH5 and pH9), their life expectancies are specified as follows.
When operated for 1 hour at the maximum speed:
Tubes (PA, PC, PP): 5 operations Thick-walled tubes and bottles (PA, PC, PP): 50 operations himac 50 TC tube and tubes on the market: 1 operation
The life expectancies of the PC bottles are specifically specified as follows according to the pretreatment conditions such as cleaning and sterilization.
sterilization
Sample Neutral (PH7) 50 operations 10 operations Alkalescent (PH7 to 9) 30 operations 5 operations
Do not use crazed (cracked) tubes or bottles.
The life expectancy of a plastic tube/bottle as the above is an approximate guide. We do not warrant the life expectancies of tubes/bottles.
Gas sterilization and
cleaning with warm water
Autoclaving at 121°C for 20 min.
7-4
7. Maintenance
Do not operate this centrifuge in any manner not described in this instruction manual.
7.4 Service decontamination policy
WARNING
If the centrifuge, rotor, or an accessory is contaminated by samples that are toxic or
radioactive, or blood samples that are pathogenic or infectious, be sure to decontaminate the item according to good laboratory procedures and methods.
If there is a possibility that the centrifuge, rotor, or an accessory is contaminated by
samples that might impair human health (for example, samples that are toxic or radioactive, or blood samples that are pathogenic or infectious), it is your responsibility to sterilize or decontaminate the centrifuge, rotor or the accessory properly before requesting repairs from a Dynamica authorized sales/service representative. Note that Dynamica cannot repair the centrifuge, rotor or the accessory unless sterilization or decontamination is completed.
It is your responsibility to sterilize and/or decontaminate the centrifuge, rotor, or an
accessory properly before returning to a Dynamica authorized sales/service representative. In such cases, copy the decontamination sheet at the end of this manual and fill out the copied sheet, then attach it to the item to be returned. Dynamica may ask you about the treatment for the centrifuge, rotor or the part if the decontamination is checked and judged as insufficient by Dynamica. It is your responsibility to bear the cost of sterilization or decontamination. If you have any question, please send e-mail to “info@dynamica-eu.com”. Note that Dynamica cannot repair or inspect the centrifuge, the rotor or the accessory unless sterilization or decontamination is completed.
CAUTION
Should you have any troubles with this centrifuge, call a Dynamica authorized sales/service representative.
(1) Storage period of service parts
Service parts are kept in stock for seven years after the discontinuation of production. The term “service parts” means the parts that are necessary to ensure the correct functioning of the centrifuge.
7-5
8. Specifications
2
Velocity 15HR Maximum speed
Maximum RCF Maximum capacity Range of speed control
Temperature
2 mL x 24 tubes + 0.5 mL x 24 tubes (FAS15E rotor)
300 to 15,000 rpm in 100-rpm increments
( “Lo” is displayed as -20 )
Refrigerator
(For further information, see the instrument labels*1.)
Timer Acceleration/deceleration control Driving motor Memory-based programmed operation
Rotor stop signal
Safety devices
1-99 minutesin 1-minute intervalswith HOLD function
2-stage variable acceleration control, 2-stage braked deceleration,
plus coasting deceleration control
Brushless DC motorinverter-controlled
Selectable three programmed operations (SPEED/RCF, TIME, TEMP
(ACCEL, DECEL, and rotor stop signal)
Selectable from seven kinds of sounds including five tunes, electronic
Door interlock, dual-overspeed detector, imbalance detector and
detector for over-temperature of motor
Applicable standard
Heat dissipation
108-126 V AC, 15 A, 60 Hz
Power requirements
198-253 V AC, 10 A, 50Hz ;
Dimensions
290 (W)×538 (D)×307 (H) mm
Weight
Ambient temperature
*1: Because the instrument labels may differ from one country to another, please contact your local
dealer or a Dynamica authorized sales/service representative.
The Velocity 15HR centrifuges satisfy CE marking requirements. The CE marking is an international symbol, which shows that the product conforms to EU directives.
Standards concerning these directives are as follows:
• Product Safety (EN61010-1 and EN61010-2-020) Environment requirements:
• indoor use;
• altitude up to 2000 m;
maximum relative humidity of 80 % for temperature up to 31 decreasing linearly to 50 %
relative humidity at 40 ;
Pollution degree Installation category
• Electromagnetic compatibility (EN61326-1)
Ambient Temperature Range2 - 32 Ambient Temperature Range for performance guarantee:15 - 25
15,000 rpm
21,500×g
-20 to 40
Hermetically sealed
(refrigerant: HFC134a)
beep, and no sound
CE marking
500 W
120 V instrument:
220-230 V instrument:
198-242 V AC, 10 A, 60Hz
40 kg
8-1
Attention: Dynamica Scientific Ltd.
Date
biological or chemical contaminants
(including radioactive isotope) from this product as follows.
Date of
decontamination
Signature
APPENDIX
It is requested that you return the faulty product with this Decontamination Sheet in order to repair it safely in our plant. Be sure to decontaminate the product according to good laboratory procedures and methods, and fill out this Decontamination Sheet and attach it to the product to be returned to Dynamica for repair.
Decontamination Sheet
Name
Name of company(organization) or school
Division or faculty/Subject of study
Telephone number
Address
I performed decontamination to remove
Model of centrifuge Serial number
Model of rotor Serial number
Accessory Serial number
Contaminants used
Decontamination methods(conditions)
WEEE Compliance
The mark is in compliance with the Waste Electrical and Electronic Equipment Directive 2012/19/EU (WEEE). The mark indicates the requirement NOT to dispose the equipment as unsorted municipal waste, but use the return and collection systems available. For further information regarding return, collection, recycle or disposal, please contact your local dealer or a Dynamica authorized sales/service representative.
Marking for the restriction of the use of hazardous substances in electrical and electronic product (THE PEOPLE’S REPUBLIC OF CHINA)
The mark and separated sheet “Names and Contents of Hazardous Substances in Each Component of This Centrifuge” are in compliance with “People’s Republic of China Electronic Industry Standard SJ/T11364-2014” requirements. The mark indicates that this electrical and electronic product contains certain hazardous substances, and can be used safely during its environment-friendly use period. The number in the middle of the mark indicates the environment-friendly use period of the product. The outer circle indicates that the product can be recovered.
Names and Contents of Hazardous Substances in Each Component of This Centrifuge
    
    
    
    
    
    
    
    
all of the homogeneous materials for this
Name of
Hazardous
Substances
(Model of centrifuge: Velocity 15HR
Component
Printed wiring board Ass’y
Cover
Drive unit
Parts for electric machine
Parts for power source
Cables
Screws
Refrigerator
Parts for refrigerator piping
This table is prepared in accordance with the provisions of SJ/T 11364.
:Indicates that this hazardous substance contained in
component is below the limit requirement of GB/T 26572.
:Indicates that this hazardous substance contained in at least one of the homogeneous materials used
for this component is above the limit requirement of GB/T 26572.
Pb Hg Cd Cr(VI) PBB PBDE
产品中有害物质的名称及含量
产品型号Velocity 15HR
有害物质
部件名称
印刷电路板组件 外壳部件 驱动部分 电源部件 电机部件 各种电缆 各种螺 冷冻机 冷冻配管部件
本表格依据SJ/T 11364 的规定编制
○:表示该有害物质在该部件所有均质材料中的含量均在GB/T 26572规定的限量要求以下。
×:表示该有害物质至少在该部件的某一均质材料中的含量超出 GB/T 26572 规定的限量要求。
(Pb)
(Hg)
(Cd)
六价铬
(Cr (VI))
多溴联苯
(PBB)
× ○ ○ × ○ × ○ × × ○ × ○ × ○ ○ × ○ × ○ ×
多溴二苯醚
(PBDE)
S99993802-D
ROTOR GUIDE
Rotor Factory
rotor code
FAS15E T15A61 D3-90591900 Max Speed: 15,000 rpm,
FAS15F T15A62 D3-90592000 Max Speed: 15,000 rpm,
FAS15G T15A63 D3-90592100 Max Speed: 15,000 rpm,
FA15S.AC N/A D3-S308839B Metal cover for FAS15E rotor
Part number Specification
Max RCF: 21,500 x g Tube Capacity: 1.5/2ml x 24 With plastic cover
Max RCF: 21,500 x g Tube Capacity: 1.5/2ml x 24 With metal cover
Max RCF: 21,500 x g Tube Capacity: 1.5/2ml x 24 With metal cover
(Material:
Instruction
manual
Keep this rotor instruction manual with the centrifuge instruction
manual.
Angle rotor for Tabletop High-Speed Micro Centrifuge
Safety Reminder
Centrifuge rotors rotating at high speed have considerable potential for damage to personal
FAS15E/15F/15G
properties if used improperly. For safe and proper use of this rotor, carefully read the centrifuge instruction manual and this rotor instruction manual before use and observe the instructions.
WARNING: and
CAUTION: notes are used to call your attention in this manual to
prevent personal injury or damage to the rotor and the centrifuge. These notes are defined as follows.
WARNING: indicates a potentially hazardous situation which, if not avoided, could result in
personal severe injury or possible death.
CAUTION : indicates a hazardous situation which, if not avoided, could result in personal
injury or severe damage to the instrument.
Specifications
FAS15E
Rotor cover (Material: Polypropylene) Rotor cover (option) (Material: AL)
Rmin.=5.95 cm
Tube cavity
Rmax.=8.56 cm
45
°
Cover-locking Knob
(Material: Stainless
steel)
Nut (Material:
Stainless steel)
O-ring
(Material: NBR)
Drive hole (Crown hole)
Applicable centrifuge: Velocity 15HR Max. speed: 15,000rpm Max. RCF: 21,500 Allowable mean density of sample: 1.2
Nominal rotor capacity: 2.0mL x 24tubes = 48mL Material of rotor: Aluminum alloy Dimensions of rotor:
φ
Weight: 0.9
×
g
/mL
17.5cm x6.1cm
Cross-sectional view of rotor
FAS15F
O-ring (Material:
Fluorine-contained rubber)
Rotor cover (Material: AL)
Tube cavity
Rmin.=5.02 cm Rmin.=5.95 cm
Nut (Material: Stainless steel)
Cover-locking Knob (Material: AL )
45
50
°
Rmax.=7.18 cm Rmax.=8.56 cm
Right: Tube cavity for 1.5/2.0mL tube (Crown hole) Left: Tube cavity for 0.5mL tube
°
Drive hole
Cross-sectional view of rotor
FAS15G
Rotor cover (Material: AL)
O-ring (Material:
Fluorine-contained rubber)
Tube cavity
Upper: Edge Lower: Center
Cover-locking Knob
(Material: AL )
Rmin.=7.44 cm
6.77 cm
Rmax.=8.56 cm
7.98 cm
Nut (Material:
°
45
Stainless steel)
O-ring
(Material:
Drive hole (Crown hole
Cross-sectional view of rotor
O-ring
NBR)
NBR)
Applicable centrifuge: Velocity 15HR Max. speed: 15,000rpm Max. RCF: 1.5/2.0mL :21,500
0.5mL :18,100
Allowable mean density of sample: 1.2 Nominal rotor capacity: 60mL:
/mL
2.0mLX24tubes=48mL
0.5mLX24tubes=12mL
Material of rotor: Aluminum alloy
φ
Dimensions of rotor: Weight: 1.1
Applicable centrifuge: Velocity 15HR
17.5cm x6.1cm
Max. speed: 15,000rpm Max. RCF: Corner tube hole :21,500
Allowable mean density of sample: 1.2
Center tube hole :20,100
/mL
Nominal rotor capacity: 0.2mLX32=6.4mL Material of rotor: Aluminum alloy
φ
Dimensions of rotor: Weight: 0.9
)
17.5cm x5.6cm
× ×
g g
g
×
×
g
Dynamica Scientific Ltd
All rights reserved. No part of this document may be reproduced or transmitted in any form or any means without permission from Dynamica Scientific Limited
The names of actual companies and products mentioned herein may be the trademarks of their respective owners.
Copyright © 2018 Dynamica Scientific Ltd.
2018.05 S99994105D
試料平均度(
/
ml
1.2
This rotor can separate sample whose mean density is 1.2/mL or less.
Nominal
L
When the mean density of the sample to be separated is over 1.2/mL, calculate the allowable speed according to the following equation.
Allowable speed (rpm)15,000(rpm)×
Mean density of sample (g/mL)
(g/mL)
1.2
To find the RCF: Substitute the rotation radius (Rmax., Rmin., etc.) and the speed in the following equation.
RCF (x g) = 1.118 x 10-5x Rotation radius (cm) x (Speed (rpm))2
CAUTION:Use the FAS15E/FAS15F/FAS15G rotor only in the Velocity 15HRE centrifuge.
These rotors can not be used in other centrifuges.
Applicable Tubes and Adapters
CAUTION:When using tubes outside 4-25℃, tubes may be broken or deformed during operation.
Test them before actual operation and perform actual operation after checking that there is no inconvenience such as the above.
Maximum speeds listed are guidelines only. Because of variances in user methodologies, no guarantee of
performance is expressed or implied.
FAS15E
Max. speed (Max. RCF)
Part No.
Adapter Tube
Name
capacity
Part No.
Name
Actual
capacity
Size (mm)
S413789A
S413789A
487293A
487292A
487292A
1GA2 adapter
(2 pcs.)
(Material:1)
1GA2 adapter
(2 pcs.)
(Material:1)
0.75 adapter (4 pcs.)
(Material:2)
0.4 adapter (4 pcs.)
(Material:2)
0.4 adapter (4 pcs.)
(Material:2)
2.0 mL
1.5 mL
1.0 mL
0.7 mL
0.5 mL
0.5 mL
0.4 mL
0.25 mL
(Marketed)
(Marketed)
(Marketed)
489241A
(Marketed)
489240A
(Marketed)
(Marketed)
(Marketed)
2.0 mL micro tube
1.5 mL micro tube
1.0 round-bottom glass tube
0.7 mL micro tube
0.5 glass tube (100 pcs.)
0.5 mL micro tube
0.4 mL micro tube
0.25 mL micro tube
Centrifugal filter unit
(100 pcs.)
1.7 mL
1.4 mL
0.7 mL
(D) 10.8×(L) 39
(D) 10.8×(L) 39
(D) 9.0×(L) 35
0.6 mL (D) 10.8×(L) 25
0.5 mL
0.5 mL
0.4 mL
0.25 mL
(D) 9.0×(L) 35
(D) 7.8×(L) 30
(D) 5.7×(L) 46
(D) 5.7×(L) 30
(D) 12.3×(L) 45
15,000 rpm
(21,500 xg 3)
: When using tubes on the market, perform operation under the allowable speed or the allowable RCF specified by
the manufacturer. Otherwise the tubes may be broken during operation.
1: PP 2: ABS 3: Max. RCF is 21,500xg, when Rmax. is 8.56cm.
D
L
Unit (mm)
FAS15F
Nominal
D
0.5mL tubes can be used without adapters in this rotor.
Adapter Tube
Part No.
Name
capacity
Part No.
Name
Actual
capacity
Size (mm)
Max. speed (Max. RCF)
S413789A
S413789A
487292A
487292A
1GA2 adapter
(2 pcs.)
(Material:1)
1GA2 adapter
(2 pcs.)
(Material:1)
0.4 adapter (4 pcs.)
(Material:2)
0.4 adapter (4 pcs.)
(Material:2)
2.0 mL
1.5 mL
1.0 mL
0.7 mL
0.5 mL
0.4 mL
0.25 mL
(Marketed)
0.5 mL
(Marketed)
(Marketed)
489241A
(Marketed)
489240A
(Marketed)
(Marketed)
(Marketed)
2.0 mL micro tube
1.5 mL micro tube
1.0 round-bottom glass tube
0.7 mL micro tube
0.5 glass tube
0.4 mL micro tube
0.25 mL micro tube
Centrifugal filter unit
0.5 mL micro tube
L
D
(100 pcs.)
(100 pcs.)
1.7 mL
1.4 mL
0.7 mL
0.6 mL
0.5 mL
0.4 mL
0.25 mL
0.5 mL
L
(D) 10.8×(L) 39
(D) 10.8×(L) 39
(D) 9.0×(L) 35
(D) 10.8×(L) 25
(D) 9.0×(L) 35
(D) 5.7×(L) 46
(D) 5.7×(L) 30
(D) 12.3×(L) 45
(D) 7.8×(L) 30
Unit (mm)
15,000 rpm
(21,500 xg 3)
15,000 rpm
(18,100 xg 4)
: When using tubes on the market, perform operation under the allowable speed or the allowable RCF specified by
the manufacturer. Otherwise the tubes may be broken during operation.
1: PP 2: ABS 3: Max. RCF is 21,500xg, when Rmax. is 8.56cm. 4: Max. RCF is 18,100xg, when Rmax. is 7.18cm.
FAS15G
(21,500 xg
2
)
(20,100 xg
3
)
Single or 8-strip tubes and caps can be used in this rotor.
Adapter Tube
Part No.
Name
Nominal capacity
Part No.
Name
Actual
capacity
Size (mm)
Max. speed (Max. RCF)
: When using tubes on the market, perform operation under the allowable speed or the allowable RCF specified by the
0.2 mL
0.2 mL
(Marketed)
(Marketed)
0.2 mL PCR tube (single)
0.2 mL PCR 8-tube strip
Unit (mm)
15,000 rpm
1
1
(D) 6×(L) 24
(D) 6×(L) 24
L
D
manufacturer. Otherwise the tubes may be broken during operation.
1: Actual capacities of PCR tubes on the market may vary depending on the specifications. Refer to the manufacture’s
catalog.
2: Max. RCF is 21,500xg, when Rmax. is 8.56cm (Corner tube hole).
3: Max. RCF is 20,100xg, when Rmax. is 7.98cm (Center tube hole).
Filter
load the tubes to the rotor not to stick the tube
caps out of the rotor if the tubes can not be
Otherwise the rotor may abnormally vibrate
ape during the
When using tubes on the market, perform operation under the allowable speed or the allowable RCF
specified by the manufacturer. Otherwise the tubes may be broken during operation.
CAUTION:When using tubes outside 4-25℃, tubes may be broken or deformed during operation.
Test them before actual operation and perform actual operation after checking that there is no inconveniences such as the above.
Precautions in using centrifugal filter unit when using the FAS15E/FAS15F rotor. 1
Be sure to cap the tubes before the centrifugation when the tubes can be capped.
Load the tubes to the rotor not to stick the tube caps out of the rotor when the tubes can not be capped.
2
Micro tube
Example of correct setting
Be sure to cap the tubes
Example of centrifugal filter unit
before the centrifugation.
Example of correct setting
Load the tubes to the rotor not to stick the tube caps out of the rotor when the tubes can not be capped.
CAUTION:When using the marketed centrifugal filter unit,
Tube cap
capped.
depending on the tube sh centrifugation.
Example of incorrect setting
Spare parts
Lubricant for O-rings or the threaded portion of the rotor is not included in this rotor. Please order applicable ones selecting from the following list (designate the part numbers). O-rings are consumable parts. We recommend that O-rings are replaced with new ones periodically.
For FAS15E/FAS15F/FAS15G rotor
O-ring for the cover-locking knob
Part No. 483719 Silicone grease (vacuum grease) Part No. 84810601 Lubricant for screw
O-ring for the rotor cover
For FAS15E rotor
Part No. 84521015 O-ring for the cover-locking knob (5pcs./set)
For FAS15F rotor Part No. 84521015 O-ring for the cover-locking knob (5pcs./set) Part No. 4348812 O-ring for the rotor cover (1pc./set)
For FAS15G rotor Part No. 84521015 O-ring for the cover-locking knob (5pcs./set) Part No. 84522007 O-ring for the rotor cover (1pc./set)
Operating Procedure
Mark
For 0.5 mL
micro tube
operation.
Mark
1.
Install the rotor according to the procedures specified in the centrifuge instruction manual.
CAUTION:Be sure to tighten the nut with the attached tool for setting the rotor. Be sure to use the
attached tool to tighten the nut. Otherwise it might be difficult to remove the tool from the nut. If you tighten the nut insufficiently, the rotor might abnormally vibrate or come off. There is the mark in the side of the nut. If you can see the mark after you finish tightening the nut, the nut has not been tightened sufficiently. Tighten the nut sufficiently until you can not see the mark on the side of the nut.
Nut
Improper setting:
If you can see the mark after you finish tightening the nut, the nut has not been tightened sufficiently.
2.
Inject the sample into the tubes.
CAUTION:Check the attached chemical resistance chart, and do not use any sample inapplicable to
If you can not see the mark after you finish tightening the nut, the nut has been tightened sufficiently.
Proper setting:
the rotor, adapters, or tubes. Using such a sample could corrode or deteriorate them.
CAUTION:Do not exceed the actual capacity specified in “Applicable Tubes and Adapters”.
Otherwise the sample leakage may occur during operation.
CAUTION:Use of tubes with caps is recommended to prevent contamination of the sample with dust or water.
3.
Balance the tubes that are symmetrically arranged.
CAUTION:Balance between symmetrically arranged tubes filled with the same-density sample.
Variations in sample density cause imbalance operation.
The allowable imbalance of this rotor is specified in the following list.
Allowable imbalance
Difference in weight of contents( the tube and the sample)
in the tube cavities that are arranged symmetrically.
FAS15E/15F FAS15G
Weight of difference Within 0.2 g Within 0.1 g
Level difference Within 5 mm
:Level difference of sample
Level difference is a guide for balancing samples. Note that the
[A
weight difference does not necessarily agree with the level difference.
CAUTION:The FAS15F rotor has two kinds of tube cavities(for
1.5mL/2.0mL micro tube and for 0.5mL micro tube).
For 1.5mL/2.0 mL micro tube
Check the tube size before operation and load
the tubes into the tube cavities corresponding to the tube.
CAUTION:The caps of the 0.5mL micro tube and 1.5mL/2.0mL micro
tube may overlap depending on the kinds of tubes. Do not load 1.5mL/2.0mL micro tube and 0.5mL micro tube together in this rotor when the caps of them overlap. Otherwise the tubes may be damaged during the
4.
Check that there is no crack or deterioration on the rotor. If the tube cavities are deteriorated, the strength of
the rotor may be significantly decreased. Pay special attention when checking the tube cavities. The rotor is deteriorated if the surface is discolored, dented or cracked.
CAUTION:If there is any abnormality such as deterioration or crack, stop use of the rotor and
contact an authorized sales/service representative.
Check that there is no foreign matter in tube cavities.
5.
Load the balanced tubes and adapters in the rotor symmetrically with respect to the center of rotation of the rotor .
CAUTION:Do not load the tubes asymmetrically. Asymmetrical loading may cause imbalance
operation and the centrifuge may be damaged.
Examples of incorrect tube setting are shown below.
When using the FAS15E/FAS15F rotor
Center of rotation
Tubes are loaded.
Tubes are not loaded.
When using the FAS15G rotor
Center of rotation
6.
Mount the rotor cover to the rotor. This rotor is available without setting the rotor cover to the rotor. It is recommended that the rotor cover is set to the rotor in order to prevent contamination of the sample with dust or water and avoid dispersing the sample to the rotor chamber when the sample leakage occurs. We recommend that you mount the rotor cover to the rotor because it will prevent the temperature of the sample from increasing and it will prevent the noise of the centrifuge from increasing during the operation.
Mounting the rotor cover (Polypropylene) on FAS15E rotor
(1)
Apply silicone grease (vacuum grease) (Part No.483719) to the rotor packing (O-ring) and then fit the rotor packing
Rotor cover
Cover-locking knob
O-ring
in the groove of the cover-locking knob.
Replace the rotor packing with new one if damaged or deteriorated.
(2)
Apply lubricant for screw (Part No. 84810601) to the inner claw of the cover-locking knob.
(3)
Set the rotor cover onto the rotor. Adjust UNLOCK of the cover-locking knob to the match marks of the rotor. Then,
Inner Claw
push down the cover-locking knob and rotate it for adjusting LOCK to the match marks of the rotor.
(4)
Check that the rotor cover cannot be rotated.
Match marks (Rotor)
Rotor cover
Cover-locking knob
Rotor
CAUTION:In using FAS15E rotor whose cover is made of polypropylene, hold the rotor body when
tightening the cover-locking knob. Holding the rotor body avoids moving the rotor body.
CAUTION:Be sure to adjust LOCK (of the cover-locking knob) to the match marks of the rotor.
Otherwise the rotor cover might be removed during operation and result in damage to the centrifuge and the rotor. If you can rotate the rotor cover after adjusting LOCK to the match marks of the rotor, the rotor cover may have deteriorated or been transformed. In that case, replace it with new one.
Although the contact mark may remain on the rotor when you use the rotor cover (Polypropylene) repeatedly, there is no problem.
Contact part with the
cover-locking knob
Mounting the rotor cover (AL) on FAS15E/15F/15G rotor
(1)
Apply silicone grease (vacuum grease) (Part No.483719) to the rotor packing (O-ring)and then fit the rotor packing in the groove of the cover. Replace the rotor packing with new one if damaged or deteriorated.
(2)
Apply lubricant for screw (Part No. 84810601) to the threaded portion of the cove-locking knob. Place the rotor cover on the rotor body.
Cover-locking knob O-ring
Rotor cover
Tighten the cover-locking knob to fix the rotor
O-ring
Threaded portion
cover to the rotor body securely.
CAUTION:Be sure to tighten the cover-locking knob securely before operation when mounting the rotor
cover to the rotor body. Otherwise the rotor cover might be removed during operation and result in damage to the centrifuge and the rotor.
7.
Run the rotor according to the procedures specified in the centrifuge instruction manual.
8.
After the operation, gently remove the rotor cover from the rotor. Take out the tubes from the rotor and then take out the sample in a proper manner.
Precautions in Cleaning
Wipe the inside of the drive hole (crown hole) of the rotor with a soft cloth dampened with water
sufficiently.
Use tap water or neutral detergent (pH5~pH9) for cleaning the rotor and the adapters.
If sample is spilled in the rotor, wash the stains off and dry it well to avoid corrosion.
In the event a tube is broken while the rotor is running and the fragments or foreign matter remains in the
tube cavities, remove the fragments or foreign matter completely.
Precautions in Sterilization
Sterilize the rotor in any of the following methods; autoclaving (121℃, 20 minutes), gas sterilization (ethylene oxide or formaldehyde), chemicals sterilization (ethanol (70%), hydrogen peroxide (3%) or formalin (3%)), or ultraviolet rays sterilization.
Sterilize the adapters in the following method; chemicals sterilization (hydrogen peroxide (3%)).
CAUTION:Do not dip the rotor in the formalin (3%) solution more than 2 hours.
CAUTION:Be sure to sterilize the rotor after removing the rotor cover from the rotor.
CAUTION:Inspect the cover-packing (O-rings) after sterilization. If you observe deterioration of
the cover-packing (O-rings), replace them with new ones.
Sterilizing the rotor cover (material: Polypropylene) of theFAS15E rotor
Sterilize the rotor cover (material: Polypropylene) in the following method; chemicals sterilization
(ethanol (70%) and hydrogen peroxide (3%)).
CAUTION:Never sterilize the rotor cover by autoclaving. Otherwise the rotor cover might be
deteriorated.
Maintenance
Retire After Years
Rotor
7
centrifuge,
repair or inspect
! WARNING
:●
If the rotor or an accessory is contaminated by samples that are toxic or radioactive, or
blood samples that are pathogenic or infectious, be sure to decontaminate the item according to good laboratory procedures and methods.
If there is a possibility that the rotor or an accessory is contaminated by samples that might impair human health (for example, samples that are toxic or radioactive, or blood samples that are pathogenic or infectious), it is your responsibility to sterilize or decontaminate the rotor or the accessory properly before requesting repairs from an authorized sales/service representative. Note that we cannot repair the rotor or the accessory unless sterilization or decontamination is completed.
It is your responsibility to sterilize and/or decontaminate the rotor or parts properly before returning them to an authorized sales/service representative. In such cases, copy the attached decontamination sheet and fill out the copied sheet, then attach it to the item to be returned. We may ask you about the treatment for the rotor or parts if the decontamination is checked and judged as insufficient by us. It is your responsibility to bear the cost of sterilization or decontamination. Note that we cannot the rotor or the accessory unless sterilization or decontamination is completed.
Rotor Life
! WARNING:Do not use rotors that have exceeded their lifetime: to do so might cause the rotor failure
and the centrifuge might be damaged. The life expectancy of rotors depends on factors such as the speed of the rotor and temperature. Always check for deterioration and damage (cracks, deformation, and so on) on rotors before using them. Do not use the rotors if you find such a problem.
After many years of use, there will be inevitably some corrosion or stress corrosion. At some points, the combination of such damage and metal fatigue could make the rotor vulnerable to a failure. Although a rotor may appear to be in a good condition, you should follow the rotor retirement recommendation shown in Table 1 (see below).
Table 1. Retirement Recommendation
Warranty
Dynamica warrants its compact centrifuge rotors under the terms and conditions set forth below.
Scope of Application
1.
This warranty applies to failure or damage to the rotor due to rotor defects, in materials or workmanship, that are attributable to Dynamica.
Note: This warranty covers the rotor only, and Dynamica shall not be liable for damage to samples or
other items that may arise from failure of or damage to the rotor.
Warranty Period
2.
This warranty is valid for seven years from the date of delivery of the rotor by Dynamica or a Dynamica dealer.
Notes
3.
This warranty extends only to the original buyer from Dynamica or the original buyer from a
(1)
Dynamica dealer. Rotors whose ownership passes from the original buyer to a third person due to resale, transfer or the like are not covered by this warranty. Neither does this warranty cover cases where the ownership of a rotor is reacquired subsequently by the original buyer. The warranty for the rotor proper (warranty period and terms) shall continue to apply in cases where
(2)
buckets and rotor parts and the like are additionally purchased. In such cases, the warranty period shall continue to be measured from the date of delivery of the rotor. If such parts are subject to a restricted warranty period, this warranty shall only cover failure or damage involving such parts that occurs both within the rotor warranty period and within the warranty period for the parts concerned.
This warranty does not cover any of the following, even during the warranty period:
(3)
Failure or damage due to abuse or misuse.
Failure or damage due to operation or maintenance in a manner contrary to the instructions in the
operator's manuals for the rotor and centrifuge.
Failure or damage due to disassembly or modification without the permission of Dynamica.
Failure or damage due to use of a bucket, adapter, tube, bottle, or the like, other than those
approved by Dynamica for use with the product.
Failure or damage due to disaster such as fire or earthquake.
Failure or damage due to use with a centrifuge not of Dynamica manufacture without the
permission of Dynamica.
Failure or damage due to use with a centrifuge that has been modified without the permission of
Dynamica.
Failure or damage due to a failed or damaged part (including a tube, bottle, adapter, etc.) not
submitted to Dynamica.
Replacement of consumable items.
Parts subject to a restricted warranty period that has expired.
MEMO
Warranty
Warranty of the centrifuge
This centrifuge is warranted for one (1) year from the date of delivery provided that it has been operated and maintained properly.
Warranty of the rotor
For information on the warranty on rotors, refer to the instruction manuals of each rotor.
Incidental conditions
We do not warrant this centrifuge and the rotor under the following conditions even before the warranty period expires:
Failures caused by incorrect installation
(1)
Failures caused by rough and/or improper handling
(2)
Failures caused by operation or maintenance in any manner not described in the
(3)
rotor instruction manual and the centrifuge instruction manual
Failures caused by conveyance or relocation after installation
(4) (5)
Failures caused by modification or disassembly without Dynamica’s permission
(6)
Failures caused by use of rotors, buckets, adapters, tubes and bottles that are not
designated for the centrifuge by Dynamica
Failures caused by fire, earthquakes, or other natural disaster
(7)
Consumable parts and parts having a limited warranty period
(8)
Failures caused by use of a rotor that is out of warranty
(9)
This warranty does not apply to samples or other damage caused by a failure of this centrifuge or the rotor.
After-sales Service
Periodic inspection of the centrifuge is recommended to assure safe and efficient operation. If the centrifuge fails to function normally, do not attempt to repair it yourself. Contact a Dynamica authorized sales/service representative.
The Velocity Range
Bench Top Centrifuges
Dynamica Scientific Limited
2 Bain Square, Kirkton Campus, Livingston,
EH54 7DQ, United Kingdom
P: +44 1908 211 900
F: +44 1908 211 909
Email: info@dynamica-eu.com
Web: www.dynamica-eu.com
Australasia
Dynamica Pty Ltd
ABN 16 126 043 052
PO Box 3278 Prahran East Victoria 3181, Australia
P: +61 3 8540 5988
F: +61 3 8540 5981
Email: info@dynamica.com,au
Web: www.dynamica.com.au
Asia
Dynamica (Asia) Limited
6/F Mita Centre
552-566 Castle Peak Road, Kwai Chung
Kowloon, Hong Kong
P: +852 3583 1581
F: +852 3583 1580
Email: info@dynamica-asia.com
Web: www.dynamica-asia.com
Loading...