Dwyer Instruments VT-300 Specifications

L
3-1/64
®
MODEL VT-300
MINIATURE VANE THERMO-ANEMOMETER
Telescoping Probe Measures Air Velocity, Air Volume, Temperature, and Humidity
TEST & DATA
3-21/64
[84.53]
1-23/64
[34.73]
The MODEL VT-300 Miniature Vane Thermo-Anemometer measures air velocity, air volume, temperature, and humidity inside air ducts. This meter includes a
telescoping vane probe that is only 0.7˝ (18 mm) in diameter that allows duct traverse measurements up to 20˝ ducts. User-selectable units include ft/min, m/s, knots, mph, and km/hr. The vane probe has a built-in sensor to record temperature in °F or °C, as well as humidity in %RH. There are three modes which include viewing
data in real time, manual recording, and automatic recording. Model VT-300 can store
measurements that can later be transferred to a PC via RS-232 communication. The vane probe is detachable for easy replacement, if necessary. Each unit is supplied with a hard carrying case, batteries, logging software CD, USB to RS-232 cable and
instruction manual.
FEATURES/BENEFITS
Vane Thermo-Anemometers
3-in-1 parameters: Air Velocity/Flow, RH, and Temperature
Telescopic probe for traversing ducts up to 20”
Built-in data logging for paperless reporting
Vane Anemometer offers wider application use vs pitot or hotwire
Hard case, Batteries, and software included
APPLICATIONS
Building commissioning
Building HVAC test and balance
• Critical environment testing
Industrial process verication
• Instrumentation validation
MODEL CHART Model Description VT-300 Miniature vane thermo-anemometer
51/64
[20.13]
PS/2 PLUG
9V DC POWER PORT
1/8˝ RS232 PORT
3/4
[19.05]
SPECIFICATIONS AIR VELOCITY Range: 98.4 to 3937 ft/min (0.5 to 20
m/s).
Accuracy: ±3% of reading + 0.2 m/s. Resolution: 0.1 m/s. Response Time: 1 s.
TEMPERATURE Range: -4 to 140°F (-20 to 60°C). Accuracy: ±1°F (±0.6°C) from -4 to
122°F (-20 to 50°C); ±2.2°F (±1.2°C) from 122 to 140°F (50 to 60°C).
Resolution: 0.1°F (0.1°C). Response Time: 60 s (typ).
RELATIVE HUMIDITY Range: 0.1 to 99.9% RH. Accuracy: ±3% RH at 25°C (10 to 90%
RH); ±5% RH (0.1 to 10% RH, 90 to
99.9% RH).
Resolution: 0.1% RH. Response Time: 60 s (typ).
AIR VOLUME Range: 0 to 99,999 (CFM or m3/s). Resolution: 0.1 (0 to 9999.9) or 1
(10,000 to 99,999).
TELESCOPING
A-VA3000-P
Retracted Extended
WET BULB Range: -7.6 to 158°F (-22 to 70°C). Resolution: 0.1°F (0.1°C).
METER Temperature Limits: Operating: 32 to
122°F (0 to 50°C); Storage: -4 to 122°F (-20 to 50°C). Humidity Limits: Operating: <80% RH; Storage: <90% RH.
Display: 1 x 1.8˝ (26 x 45 mm). Serial Communications: 9600 bps, 8
data bits, no parity.
Power Requirements: (4) AAA 1.5 V alkaline batteries, included, user
replaceable.
Battery Life: 100 hours. Vane Diameter: 0.7˝ (18 mm). Weight: 7.41 oz (210 g).
“L” in [mm]
11-27/32 [301] 18-1/4 [946]
[76.5]
A-551
ACCESSORIES Model Description A-VT300-P A-551
Replacement probe with miniature vane Replacement probe with large vane
DWYER INSTRUMENTS, INC. | www.dwyer-inst.com202
Loading...