Dwyer Instruments TH-10 Calibration And Operating Instructions

Model TH-10 Hygro-Thermometer Pen
Specifications– Calibration and Operating Instructions
DESCRIPTION
This portable advanced meter can measure RH from 10 % to 90% and Temperature from 14 to 113˚ (-10 to 45˚C). Advanced features include MIN/MAX Memory, Push-button Calibration, Display Hold, and Automatic Power Off. The Model TH-10K kit includes Calibration bottles and a hard shell carrying case.
METER DESCRIPTION
BULLETIN H-TH
SPECIFICATIONS
RH Sensor: High precision thin-film capacitance type Temperature Range: 14 to 122˚F (-10 to 50˚C) RH Range: 10 to 95% RH Resolution: Temperature: 0.1˚; RH: 0.1% Accuracy: Temperature: ±1.8˚F (±1.0˚C); RH: ±5% (at 25˚C) DC Power: 2032 button battery (150 hour battery life) Response Time: 120 seconds Dimensions: 6.9 x 1.6 x 06˝ (175 x 42 x 16 mm)
MIN/MAX Memory Recall
1. Press and hold MIN/MAX button until “MIN” appears on the display. The reading will then be the lowest humidity and temper­ature measurements made since the meter was turned on or last reset.
2. Press and hold MIN/MAX again until “MAX” appears on the display. The reading will then be the highest humidity and temperature measurements made since the meter was turned on or last reset. Press the MIN/MAX key again to return to normal operation.
1. RH/Temperature Sensor
2. POWER ON/OFF button
3. Data HOLD button
4. MIN/MAX Memory button
5. Calibration and units select button
6. Memory Reset button
7. RH Display
8. Temperature Display
9. Function Displays
OPERATION
Measurements Press the ON/OFF button to turn the meter ON. After a 5 second initialization, RH will be displayed in the upper display and temperature will be displayed in the lower display.
Data Hold Press the HOLD button to freeze the reading in the display. “HLD” and the last reading will appear in the display. Press and hold the RESET button until the “HLD” icon disappears from the display to return the meter to normal operation.
Min/Max Memory Reset Press and hold the RESET button until “88.8” appears in the display to reset (clear) the MIN/MAX memory.
Automatic Power Off (Sleep Mode) The meter will automatically shut off after a period of approxi­mately 20 minutes. To defeat this mode, press and hold the MIN/MAX button while turning power on. Release the ON/OFF button after the display initializes. “n” will appear in the display, indicating that the non-sleep mode has been activated. Sleep mode will be re-activated when the meter is turned off.
Setting Temperature Units (C or F) CAUTION! The meter’s calibration can be corrupted if this proce­dure is not followed exactly as stated.
1. With the meter off, press and Hold the recessed F/C button (use a paper clip or similar object to press this button) and press the ON/OFF button. Release the F/C button when the F and C appear in the display.
2. Press the MIN/MAX key to select C or F on the display.
3. Press the recessed F/C button to store the selection. “S” will appear in the display.
4. Turn the meter off. Note: Accidently pressing the RESET button during this proce­dure will corrupt the meter’s calibration.
Error Messages An error message will appear on the display if the meter fails an internal diagnostic test. E2: Calibration error. Recalibration is necessary. E5: RH sensor failure. Repair/replacement is necessary.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
CALIBRATION
The following verification and calibration procedures require the 33% and 75% RH reference bottles which are supplied with the model TH-10K kit.
Accuracy Verification Checking the 33% or 75% RH Calibration
1. Insert meter’s sensor into the 33% or 75% salt reference bottle
2. Place the meter and bottle inside the meter case
3. Check the reading after 40 minutes
4. Verify that the reading is within the accuracy specification
Relative Humidity Calibration (33% and 75%) 33% Calibration
1. With power OFF, insert the meter sensor into the 33% salt reference bottle. Allow the meter to stabilize for 40 minutes
2. Press and Hold
the recessed F/C CAL button (use a paper clip or similar object to hold this key) and press ON. Release F/C CAL button when the display blanks and then displays only the small “c” and “f” symbols.
3. Press RESET. “% MIN” will appear in the display.
4. Press F/C CAL. “S” will briefly appear and then “%MAX” will appear.
5. Turn power off. The 33% calibration is now complete.
warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of spec­ification, improper maintenance or repair, or unauthorized modifi­cation. Dwyer Instruments, Inc. specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequen­tial damages. Dwyer’s total liability is limited to repair or replace­ment of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied.
75% Calibration
1. With power OFF, insert the meter into the 75% salt reference bottle. Allow the meter to stabilize for 40 minutes.
2. Press and Hold
the recessed F/C CAL button (use a paper clip or similar object to hold this key) and press ON. Release F/C CAL when the display blanks and then displays only the small “c” and “f” symbols.
3. Press RESET twice. “%MIN” and then “%MAX” will appear in the display.
4. Press F/C CAL. “S” will appear briefly and then the meter will return to normal operation.
5. Turn power off. The 75% calibration is now complete.
Note: The bottles contain a saturated salt solution that generates a known and accurate relative humidity (dependent on the type of salt) in the air above it. The calibration bottles will maintain their accuracy as long as the salt is damp. If the salt in the bottom of the bottle appears dry, the bottle should be replaced.
BATTERY REPLACEMENT
Use a coin to remove the battery compartment cover on the rear of the meter. Replace the 2032 lithium battery installing it face (+) up in the battery compartment.
WARRANTY
Dwyer Instruments, Inc. warrants this instrument to be free of defects in parts and workmanship for one year from date of ship­ment (a six month limited warranty applies on sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department at (800) 879-8000 for authoriza­tion. A Returned Goods Authorization (RGA) number must be issued before any product is returned to Dwyer Instruments, Inc.. The sender is responsible for shipping charges, freight, insur­ance, and proper packaging to prevent damage in transit. This
©Copyright 1999 Dwyer instruments, Inc Printed in U.S.A. 7/03
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Loading...