Series
2-1/2
[63.50]
1/2˝ NPT CONDUIT
KNOCKOUT
3X 1/2˝ NPT
1/4
[6.35]
L
4-5/32
[105.41]
5
[127.00]
2-1/2
[63.50]
1/2˝ NPT
CONDUIT
KNOCKOUT
3X
1/2˝ NPT
5-3/4
[145.80]
1/2˝ NPT CONDUIT
KNOCKOUT
2-1/2
[63.50]
3X 1/2˝ NPT
FOR 1/2˝ NPT CONNECTION
1/4
[6.35]
L
4-5/32
[105.41]
WTP
Wireless Temperature Sensors
Transmits up to 100 Feet, NEMA 4X Enclosure
The Series WTP Wireless Temperature Sensors can be quickly installed on chilled
water lines, air ducts or outside a building without having to run additional wiring. By
sampling the temperature measurement every 10 seconds, the 3.6 V lithium battery is able
to power the unit for up to 8 years.
TEMPERATURE
A reliable 418 MHz signal can carry the measurements up to 100 feet without a repeater.
With the optional 900 MHz repeater, measurements can be transmitted up to 1000 feet
away.
Model
WTP-OSA
WTP-D04
WTP-D08
WTP-D12
WTP-D18
WTP-I02
WTP-I04
WTP-I08
Temperature Sensors
Mounting
Configuration
OSA
Duct
Duct
Duct
Duct
Immersion
Immersion
Immersion
Insertion
Length
N/A
4˝
8˝
12˝
18˝
2˝
4˝
8˝
DUCT & IMMERSION MODEL
SPECIFICATIONS
Temperature Range: -40 to 185°F (-40 to 85°C).
Accuracy: ±0.3°C.
Temperature Limits: Ambient: 32 to 140°F (0 to 60°C).
Humidity Limits: 5 to 95% RH non-condensing.
Power Requirements: (2) 3.6 V ER14505 AA lithium metal batteries, included,
user replaceable.
Transmitter Interval: Approximately 10 s.
Housing Material: ABS plastic.
Enclosure Rating: UL 94 V-0, NEMA 4X (IP66).
Antenna: 418 MHz - Built inside enclosure.
Weight: Duct/OSA: 1.25 lb (566 g); Immersion: 0.8 lb (363 g).
FCC Approval: FCC ID# T4F06811RH.
Agency Approvals: RoHS.
ACCESSORIES (see complete list on page 135)
WM-4SS, 418 MHz Receiver
WM-PSS, 418 MHz Repeater with Standard Antenna
WM-CSF-4G, 4 to 20 mA Temperature Output Module Range 32 to 185°F
WM-VSF-1G, 0 to 10 VDC Temperature Output Module Range 32 to 185°F
WM-TS2, 10K Ω Type 2 Thermistor Output Module
WM-TS3, 10K Ω Type 3 Thermistor Output Module
DUCT MODEL
IMMERSION MODEL
OSA MODEL
308