Dwyer Instruments SERIES UFB, SERIES UFC Catalog Page

9-5/16 [236.49]
9-5/16 [236.49]
®
E
SERIES UFB & UFC
ULTRASONIC FLOWMETER SETS
Non-Invasive Pipe Flow Measurement, Easy Operation and Data Logging Option
UFB
UFB UFC
E
O
D
I
V
O
N
L
I
N
UFC
4-11/16 [119.02]
7-7/32 [183.36]
4-1/2
[114.30]
FLOW
7-13/16 [198.30]
The Series UFB & UFC Ultrasonic Flowmeter Sets utilize the transit-time difference
for measuring ow rates in pipes. These units are permanent mount, where the converters can be mounted on a surface or pipe with a 4 to 20 mA and pulse output capabilities for pipe sizes from1/2 to 79˝ (13 to 2000 mm). The Series UFC offers the same features plus data logging capability.
FEATURES/BENEFITS
• Non-invasive pipe measurement
• Easy-to-use compact and lightweight design, intended for homogeneous liquids that
contain no air
• Simple installation with all necessary components included such as converter,
sensor, cables and mounting accessories
• Sturdy IP65 rating, protecting it from dust and direct water contact
APPLICATIONS
• Water treatment
• Industrial systems
• Irrigation applications
• Treated water ow
• River water
• Sea water
• Potable water
• Demineralized water
• Glycol/water mix
• Hydraulic system
• Diesel oil
• Water use data logging
KIT INCLUDES
• Converter
• Set of transducers
• Ruled guide rail
• Steel banding
• Banding clips
• Set of transducer cables
• Set of high temperature interface cables
• Ultrasonic coupling grease
MODEL CHART - STANDARD VERSION
Model UFB-122 UFB-123 UFB-222 UFB-223
Pipe Size Range in (mm)
0.5 to 4.5 (13 to 115) 2 to 79 (50 to 2000)
0.5 to 4.5 (13 to 115) 2 to 79 (50 to 2000)
Power Supply
86 to 264 VAC 86 to 264 VAC 24 VAC/VDC 24 VAC/VDC
SPECIFICATIONS Service: Homogeneous liquids that do not contain more than 3% of air bubbles or
particulate and capable of ultrasonic wave propagation.
Inputs: TNC cable from sensors. Range: 0.33 to 33 ft/s (0.1 to 10 m/s). Display: 240 x 64 pixel graphic display, high contrast black on white with backlight;
Languages: English, French, German, Swedish, Italian, Spanish, Portuguese, Russian, Norwegian, and Dutch; 5˝ W x 1.3˝ H (5 x 33.02 mm). Accuracy: ±0.5 to ±2% of ow reading of ow rate > 0.03 ft/s (0.01 m/s) and pipe OD > 3.0 in (75 mm); ±3% of ow reading for ow rate > 0.03 ft/s (0.01 m/s) and pipe OD 0.5 to 3 in (13 to 75 mm); ±6% of ow reading for ow rate < 0.03 ft/s (0.01 m/s).
Power Requirements: 86 to 264 VAC (50 to 60 Hz) or 24 VAC/VDC (1 A max). Power Consumption: 10.5 W. Temperature Limits: Transducer: -4 to 275°F (-20 to 135°C); Controller: -4 to
122°F (-20 to 50°C). Outputs: Analog 1 opto-isolated output: 4 to 20 mA, 0 to 16 mA or 0 to 20 mA (selectable); Error current: 0 to 26 mA (selectable); Load resistance: 620 Ω max; Alarm: 2 opto-isolated MOSFET NO relays, 48 V at 500 mA, maximum 200 Hz; Pulsed: 1 opto-isolated MOSFET relay, 48 V at 500 mA, 1 to 250 pps; Pulse width: 2 to 500 ms (selectable).
Serial Communications: USB (UFC only). Enclosure Rating: IP65 when using TNC connector; Transducers IP54. Materials: Plastic ABS and aluminum. Repeatability: ±0.5 % of measured value or 0.03 ft/s (0.01 m/s). Electrical Connections: Removable screw-in type terminal block. Mounting: Wall mounted using 3 type M4 screws. Turbidity: < 3 % by volume of particulate content. Permissible Air Content: < 3% by volume. Response Time: < 500 ms. Weight: Unit not including accessories: 2.80 lb (1.26 kg); Unit including
accessories: 9.92 lb (4.5 kg).
Agency Approvals: CE.
ADDITIONAL SPECIFICATIONS Applicable Pipe Material: Carbon steel, SS, copper, UPVC/PVDF, concrete, mild
steel, glass, brass.
Applicable Pipe Lining: Rubber, glass, concrete, epoxy, steel, other*. Pipe Wall Thickness: 0.04 to 3˝ (1 to 75 mm). Pipe Lining Thickness: < 1˝ (< 25 mm).
*Selectable option for special material with known propagation rate of lining material.
Flow Transmitters,
Ultrasonic
MODEL CHART - DATA LOGGING VERSION
Model UFC-122 UFC-123 UFC-222 UFC-223
Pipe Size Range in (mm)
0.5 to 4.5 (13 to 115) 2 to 79 (50 to 2000)
0.5 to 4.5 (13 to 115) 2 to 79 (50 to 2000)
Power Supply
86 to 264 VAC 86 to 264 VAC 24 VDC/VAC 24 VDC/VAC
OPTIONS Use order code: Description NISTCAL-FU NIST traceable calibration certicate
USA: California Proposition 65 WARNING: Cancer and Reproductive Harm - www.P65Warnings.ca.gov
DWYER INSTRUMENTS, INC.
321