Dwyer Instruments SERIES IEF Catalog Page

®
E
7-3/32 [180.2]
SERIES IEF
INSERTION ELECTROMAGNETIC FLOW TRANSMITTER
FLOW
Field Congurable, High Accuracy, BACnet or Modbus® Protocol
5-53/64
[148.0]
®
YEAR LIMITED
WARRANTY
Shown with A-IEF-VLV-BR** accessory valve kit
E
O
D
I
V
O
N
L
I
N
The Series IEF Insertion Electromagnetic Flow Transmitter is an adjustable
insertion owmeter featuring electromagnetic technology that accurately and reliably measures uid velocity in addition to providing several continuous signal outputs. This series is specically designed to offer superior performance paired with simple installation and use. One unit is adjustable to t pipe sizes from 4 to 36˝ (102 to 914 mm), and offers several output options including selectable BACnet MS/TP or
Modbus® RTU communications protocol over 2-wire RS-485 in addition to the standard
analog, frequency and alarm outputs.
FEATURES/BENEFITS
• Field congurable setup displays (-LCD integral option or remote accessory A-IEF-DSP) allow for ultimate exibility by accommodating a variety of application congurations with one model through multiple display congurations i.e. pipe size, pipe material, liquid type, analog output, pulse/frequency output, alarm outputs, communication outputs, damping, and calibration factor.
• High performance accuracy is maintained through changes in temperature, density or viscosity.
• Setup Wizard and installation tool are simple to use allowing for quick and precise
installation.
• Accessory setup kit A-IEF-KIT ensures exact installation application depth with included thickness gage and measuring tape.
• Long Life Cycle and minimal maintenance requirements with no moving parts to wear or break and electrodes that discourage fouling.
• Isolation valve accessory options allow for installation in operational systems via hot-tap kit or easy removal without system downtime.
• NIST traceable pass/fail verication certicate included standard for Carbon Steel Schedule 40 pipes sized 4˝ (102 mm), 6˝ (150 mm), 8˝ (200 mm), and 10˝ (250 mm) with high accuracy option; 10˝ (250 mm) with standard option.
APPLICATIONS
• Boiler feed water
Adjustable Insertion
• Chilled water
• Open and closed loop condenser water
• Irrigation system
Flow Transmitters, Electromagnetic,
• Municipal water distribution
• Process and coolant ow
• Ground water remediation
• Chemical processing
• Pump protection
• Wastewater
• Mining
MODEL CHART Example IEF -H N -CND -LCD IEF-HN-CND-LCD Series IEF Insertion electromagnetic ow transmitter Accuracy L
Process Connection
Housing Electrical Connection
G S F I E T H
N
B
CND
PG 10
Options LCD
Note: For CC option, must provide completed conguration paperwork. Note: For maximum performance select -LCD option or setup display accessory.
DWYER INSTRUMENTS, INC. 292
Standard accuracy <10˝ (250 mm) pipe; 1% FS Standard accuracy >10˝ (250 mm) pipe; 1% FS Standard accuracy 4 to 36˝ (100 to 900 mm) pipe; 1% FS High accuracy 4˝ (100 mm) pipe; 1% of reading High accuracy 6˝ (150 mm) pipe; 1% of reading High accuracy 8˝ (200 mm) pipe; 1% of reading High accuracy 10˝ (250 mm) pipe; 1% of reading High accuracy 4 to 10˝ (100 to 250 mm) pipe; 1% of reading
1˝ male NPT 1˝ male BSPT
1/2˝ female NPT conduit connection without cable PG gland without cable PG gland with 10´ (3 m) cable
Integral LCD display
COM
BACnet or Modbus
NIST
Six point NIST traceable calibration certicate
FC
Factory calibration certicate for 0.5% of reading at single point
CC
Custom congured for specic installation
NW
NSF certied
®
communication protocol (display selectable)
-LCD option shown
A-IEF-KIT
SPECIFICATIONS
Service: Compatible clean or dirty non
coating, conductive liquids.
Range: 0 to 20 ft/s (0 to 6 m/s).* Wetted Materials: Body shaft/tting: 316
SS; Electrodes: 316 SS; Electrode cap: Polymer/Polystyrene; O-ring: Silicon. Accuracy: High accuracy units: ±0.5% of reading at calibrated velocity; ±1% of reading from 2 to 20 ft/s (0.6 to 6 m/s); ±0.02 ft/s (±0.006 m/s) at < 2 ft/s (0.6 m/s); Standard accuracy units: ±1% FS. Temperature Limits: Ambient: -20 to 160°F (-29 to 71°C); Process: 15 to 250°F (-9 to 121°C); Storage: -40 to 185°F (-40 to 85°C). Process Connection: 1˝ NPT or BSPT with accessory full port ball valve options. Pressure Limits: 400 psi (27.6 bar) @ 100º F (37.8°C). Pressure Drop: < 0.1 psi at 12 ft/s in 4˝ (101.6 mm) and larger pipe.
Outputs:
(1) Analog: 4-20 mA, 0-5 V, 0-10 V or 2-10 V (display selectable); (1) Pulse/Frequency: 0-15 V peak pulse, 0 to 500 Hz or scalable pulse output (display selectable); (2) Alarm: (1) Empty pipe detection or minimum/maximum velocity, (display selectable); (1) Reverse ow output
indication.
Power Requirements: 12-42.4 VDC, .25 A @ 24 VDC; 12-36 VAC.
*For max owrates >10 ft/s (3 m/s) order option -CC.
Brass ttings and pipe are not to be used with NSF Certied models.
5-53/64 [148.0]
4-51/64
[121.8]
®
61/64 [24.2]
22-1/64
[559.2]
Electrical Connection: Removable terminal blocks, model selectable 1/2˝ female NPT conduit connection, PG 16 gland or PG 16 gland with (2) 10 ft (3 m) 9 conductor 22 AWG plenum rated cables, accessory cable lengths up to 200 ft (61 m) optional. Display (-LCD option): 2˝ ( 5.08 cm) x 2˝ (5.08 cm) graphic LCD with backlight.
Conductivity: >20 microsiemens. Enclosure Material: Powder coated die
cast aluminum.
Enclosure Ratings: NEMA 6P (IP68) (Non display models); NEMA 4X (IP66) (-LCD option). Agency Approvals: BTL, CE, NSF/ANSI 61 and 372.
COMMUNICATIONS (-COM OPTION) Type: BACnet MS/TP or Modbus
RTU communication protocol (default disabled, display selectable). Supported Baud Rates: 9600, 19200, 38400, 57600, 76800, or 115200 bps (display selectable).
Device Load: 1/8 unit load.
ADDITIONAL SPECIFICATIONS Applicable Pipe Material: Most popular
plastic and metal pipes; i.e. Carbon steel, SS, copper, UPVC/PVDF, galvanized
steel, mild steel, and brass.
Applicable Pipe Size: 4-36˝ (101 to 914 mm), model dependent. See model chart. Diameter Length Requirements: >10 upstream; >5 downstream.
Glycol: 0 to 100% display selectable.
ACCESSORIES Model Description A-IEF-KIT
A-IEF-DSP A-IEF-CBL-50 A-IEF-VLV-BR A-IEF-VLV-SS A-IEF-PA
**Brass ttings and pipe are not to be used with NSF Certied models. Brass valves are non-RoHS compliant.
Setup kit (includes setup display, thickness gage and measuring tape), and universal power adapter Setup display Plenum rated cable 50 ft (15.2 m) 1-1/4˝ full port isolation valve brass kit** 1-1/4˝ full port isolation valve 316 SS kit AC wall adapter
Modbus® is a registered trademark of Schneider Automation, Inc.
A-IEF Remote Display now available: See page 295