SERIES 672
LOW PRESSURE TRANSDUCER
Single Pressure Connection, Ranges down to 10 in w.c.
Single Pressure Transmitters
The Series 672 Low Pressure Transducer is a perfect solution to any The Series 672
Low Pressure Transducer is a perfect solution to any application where a very accurate
low pressure transducer is necessary. Using variable capacitance technology, the
Series 672 is designed to measure pressures as low as 10 in w.c. up to 400 in w.c.,
very low ranges for a single connection pressure transducer. The 672 also features
a 0.25% FS accuracy. Use the Series 672 in liquid level, ood warning, waste water,
clean room, and open channel ow applications.
FEATURES/BENEFITS
• Low range high accuracy provides precise control for process applications
APPLICATIONS
• Liquid level
• Flood warning
• Waste water
MODEL CHART
Model Operating Range Output
672-1-A
0 to 10 in w.c.
672-2-A
0 to 15 in w.c.
672-3-A
0 to 25 in w.c.
672-4-A
0 to 50 in w.c.
672-5-A
0 to 100 in w.c.
672-6-A
0 to 150 in w.c.
672-7-A
0 to 200 in w.c.
672-8-A
0 to 300 in w.c.
672-9-A
0 to 400 in w.c.
Note: For voltage output models change -A to -V.
• Clean room
• Open ow applications
4-20 mA, 2-wire
4-20 mA, 2-wire
4-20 mA, 2-wire
4-20 mA, 2-wire
4-20 mA, 2-wire
4-20 mA, 2-wire
4-20 mA, 2-wire
4-20 mA, 2-wire
4-20 mA, 2-wire
DWYER INSTRUMENTS, INC. 98
ø1-1/2
CONNECTION
SPECIFICATIONS
Service: Compatible liquids and gases.
Wetted Materials: 318 duplex SS, ceramic, uoroelastomer (FKM).
Housing Material: 318 SS.
Accuracy: ±0.25% FS (RSS). Includes non-linearity, hysteresis, and non-
repeatability.
Stability: 0.25% FS/1 year.
Temperature Limits: -40 to 212°F (-40 to 100°C).
Compensated Temperature Limits: -5 to 140°F (-20 to 60°C).
Pressure Limits: 29 psi (2 bar) for up to 85 in w.c. (0.2 bar) ranges; 58 psi (4 bar)
for 85 to 140 in w.c. (0.2 to 0.35 bar); 73 psi (5 bar) for 141 to 400 in w.c. (0.35 to 1
bar).
Thermal Effects: Zero: 1.0%FS/100°F (2.0%FS/100°C); Span: 1.0%FS/100°F
(2.0%FS/100°C).
Power Requirements: 4-20 mA: 9-35 VDC; 0-5 VDC: 7.5-35 VDC.
Output Signal: 4-20 mA (2-wire) or 0-5 VDC (3-wire).
Zero and Span Adjustment: ±10% FS each (by potentiometer).
Response Time: 5 ms.
Max Loop Resistance: 1.325 kΩ.
Electrical Connections: Large DIN 43650 connector with mating plug.
Process Connection: 1/4˝-18 NPT male.
Enclosure Rating: NEMA 4X (IP66).
Weight: 11.6 oz (330 g).
OPTIONS
Use order code: Description
NISTCAL-PT1 NIST traceable calibration certicate