Dwyer Instruments SERIES 608 Catalog Page

1/4
[76.20]
PLACES
SERIES 608
INTRINSICALLY SAFE DIFFERENTIAL PRESSURE TRANSMITTER
Ranges Down to 0.1 in w.c., FM Approved, NEMA 4X
PRESSURE
[6.35]
The Dwyer Series 608 Intrinsically Safe Differential Pressure Transmitter converts positive, negative (vacuum), or differential pressures of clean, dry air or other non­conductive, non-corrosive gases into a standard two wire, 4-20 mA output signal. The use of an ultra-thin silicon diaphragm enables precision measurement of differential pressures as low as 0.1 in w.c. while withstanding high static working pressures up to 100 psig (6.89 bar). The Series 608 transmitters are FM approved intrinsically safe for
barrier. The rugged NEMA 4X, stainless steel housing makes this transmitter ideal for use in industrial and process plant environments.
FEATURES/BENEFITS
• High accuracy at low pressure ranges provides exceptional measurement for
ensuring tight-control and minimizing costly out of specication conditions
• Intrinsically safe for use in the specied hazardous locations meets specications
where pressure transmission and safety cannot be compromised
• NEMA 4X rated enclosure provides protection in harsh environments permitting outdoor monitoring or in areas where dust and particulate matter exists
APPLICATIONS
• Lab fume hood control
• Clean room applications
• Flow measurements and control
• Filter monitoring
Liquid & Gas
• Furnace draft measurement
• Process control
Differential Pressure Transmitters,
MODEL CHART Model Range (in w.c.) 608-02 608-03 608-04 608-05 608-06 608-07 608-01B 608-13B* 608-04B
*Models have a ±0.25% FS accuracy.
0 to 0.5 0 to 1.0 0 to 2.0 0 to 5.0 0 to 10.0 0 to 25.0 0 to ±0.25 0 to ±1.0 0 to ±2.0
2-11/16
[68.26]
3
3-1/4
4-1/2
[107.95]
5/8
[15.88]
SPECIFICATIONS Service: Clean/dry air and compatible, combustible gases. (see Agency Approvals
for FM ratings).
Wetted Materials: Consult factory. Accuracy: ±0.5% or ±0.25% FS. Stability: ±0.5% FS/year. Pressure Limits: 100 psig (6.89 bar); 15 psid (1.03 bar). Temperature Limits: -20 to 185°F (-28 to 85°C). Compensated Temperature Range: 0 to 160°F (-18 to 71°C). Thermal Effect: 0.5% Accuracy: ±0.02% FS/°F; 0.25% Accuracy: ±0.01% FS/°F. Power Requirements: 12-36 VDC (2-wire). Output Signal: 4-20 mA DC. Zero and Span Adjustments: Potentiometers for zero and span. Response Time: 250 ms. Loop Resistance: DC: 0 to 1045 Ω max. Electrical Connections: Screw terminal: Two 1/2˝ female NPT conduit. Process Connections: Two 1/4˝ female NPT. Enclosure Rating: NEMA 4X (IP66). Weight: 2 lb (0.9 kg). Agency Approvals: FM approved intrinsically safe for use in Class I, Div. 1,
Groups A, B, C, D; Class II, Div. 1, Groups E, F, G; Class III, Div. 1 when wired with approved intrinsically safe barrier. Entity parameters: V
I=12 nF; LI=0 mH.
C
OPTIONS Use order code: Description NISTCAL-PT1 NIST traceable calibration certicate
ACCESSORIES Model Description MTL5541 MTL7706
Intrinsically safe galvanic isolator
Intrinsically safe zener barrier
[82.55]
5-1/32
[127.79]
1/2 NPT TYP 2 PLACES
1-1/2
[38.10]
1/4 NPT TYP 2
max= 36 VDC; Imax= 250 mA;
74
DWYER INSTRUMENTS, INC.
Static Fitting Options: See page 452 (Static Pressure Tips)
Process Tubing Options: See page 453 (Gage Tubing Accessories)
Loading...