Dwyer Instruments MC Specifications

TEST EQUIPMENT
1/2
[12.70]
7-7/8
[200.0]
3-1/4
[82.55]
1-3/4
[44.45]
3-1/4
[82.55]
MODULES
4-1/4
[108.0]
2-7/8
[73.03]
2-3/64
[51.99]
2-11/64
[55.17]
2-17/64
[57.55]
ON OFF
ZERO
BAT CK
SET UP
PORT SEL
ENG UNIT
DAMP 2
%
_
mA
V .
FLOW 7
HOLD 4
LEAK 8
DATA LOG 9
TRIP DET 6
TARE
ENT
CE
MIN
MAX
DATA STORE
0
1
5
3
MULTI-CAL
Series
MC
Multi-Cal Pressure Calibrator
Interchangeable Ranges, ±0.05% Accuracy, Datalogging Capability, NIST Traceable
Pressure Calibrators
APPLICATIONS
Pressure measurement for clean room control, filter performance monitoring, HVAC testing and setup, draft measurement, airflow measurement control, differential pressure measurement in laminar flowhoods, paint booths, industrial ovens and fume hoods. Use as a secondary standard for calibrating pressure equipment.
Multi-Cal Pressure Modules are interchangeable and available in a wide selection of pressure ranges and accuracies. Handheld calibrator accepts up to two pressure modules. Modules include NIST calibration certification.
Model MC1000 MC1001 MC1004 MC1006 MC2010 MC2012 MC2016
OPTIONS
Consult factory for other pressure ranges and accuracies. FM approved models are also available.
458
Range
O Differential Pressure, ±0.07
0.25˝ H
2
O Differential Pressure, ±0.07
0.50˝ H
2
O Differential Pressure, ±0.06
5.00˝ H
2
O Differential Pressure, ±0.06
25˝ H
2
5.0 psig Gauge Pressure, ±0.05
15.0 psig Gauge Pressure, ±0.05
100.0 psig Gage Pressure, ±0.05
DWYER INSTRUMENTS, INC. | www.dwyer-inst.com
SPECIFICATIONS Service: Clean, dry, nonconductive, noncorrosive gases. Accuracy: Differential pressure modules: ±0.06% FS; Gauge pressure modules:
±0.05% FS; Voltage input: ±0.025% FS @ 0/10 VDC, ±0.10% FS @ 0/30 VDC; Current input: ±0.03% FS @ 0/20 mA, ±0.05% FS @ 0/50 mA.
Sensitivity: ±0.002% of span with dampening 1 part in 50,000 (max). Repeatability: Ranges ≤ 0/2 psi: ±0.05% of span; Ranges ≥ 0/5 psi: ±0.02% of
span.
Output: RS232 serial interface, 9-pin. Alarm Output: SPST form C 110 VDC, 120 VDC (max), 1 A (max), 30 W, 62.5 VA
(resistive). Display: Alphanumeric LCD, 0.37˝ (9.5 mm) height per line, 2 lines, 16 characters/line.
Display Update: 100 ms. Ambient Operating Temperature: 32 to 120°F (0 to 49°C). Storage Temperature: -4 to 158°F (-20 to 70°C). Process Connection: 1/8˝ female NPT. Electrical Connections: Miniature recessed banana jacks. Power Requirements: Internal: (2) 9 V alkaline batteries, included, user
replaceable and VL1220 or BR1225 lithium metal battery, installed functional, user replaceable; External: AC adapter 9 VDC, 500 mA.
Battery Life: 30 hours (approximate). Engineering Units: inH
defined.
O, psi, inHg, kPa, mbar, cmH2O, mmHg, and user-
2
Overpressure: Differential pressure modules: 50 psi positive direction, 15 psi negative direction; Gage pressure modules: 2x range (0/5 psi to 0/1000 psi).
Temperature Compensation: 20 to 120°F (-7 to 49°C). Temperature Error: Maximum of ±0.004% of span per °F over compensated
range for zero and span. Temperature Effect Electrical Measurement: ±0.001% of span per °F over compensated range. Dampening: (Measurement averaging) programmable from 0 to 16 consecutive readings.
Baud Rate: 300, 1200, 2400, or 9600, selectable. Housing Material: ABS plastic. Weight: Calibrator: 2.2 lb (1.3 kg); Pressure module: 0.5 lb (0.3 kg).
Model MC2K, Handheld Calibrator
Loading...