Dwyer Instruments Mark II 41-2, Mark II 41-600 Pa, Mark II 41-2-AV, Mark II 40-1-AV, Mark II 40-250-AV Installation And Operating Instructions Manual

...
10-5/8 [269.88]
5-5/16
[134.95]
MAX
4-5/8
[117.48]
2-31/32 [75.41]
25/32
[19.84]
Ø15/64 [5.95] MOUNTING HOLE
1-1/4
[31.75]
2-1/4 [57.15]
MAX
4 [101.60]
MARK II MODEL 40
Bulletin D-59
Mark II - Model 40 Molded Plastic Manometers
Installation and Operating Instructions
Dwyer Mark II - Model 40 Molded Plastic Manometers are ideally suited
for air velocity and air filter gage applica­tions at minimal cost. These inclined manometers provide linear calibration and excellent resolution throughout their ranges. They are capable of pressure
line spaced 2-31/32˝ apart. Install the gage with the self-tapping screws provided, turning the screws down snug but not tight. Adjust the position of the gage until the bub­ble is centered in the spirit level. Tighten the mounting screws, checking to be sure the instrument remained level, and re-level as necessary.
measurements above and below atmospheric, as well as differential pres­sure measurements.
INSTALLATION
Mount the
Stationary Applications:
Mark II manometer on a convenient vertical surface. The installation should not be exposed to strong chlorine atmospheres or solvents such as benzene, acetone, carbon tetrachloride, etc. The instrument is suitable for total internal pressures of up to 15 psi (100 kPa) and ambient temperatures of 150°F (65°C). DO NOT EXCEED THESE LIMITS. At the mounting location selected, drill two 9/64˝ diameter holes on a vertical
DWYER INSTRUMENTS, INC.
P.O. BOX 373 Fax: 219/872-9057 e-mail: info@dwyer-inst.com MICHIGAN CITY, INDIANA 46361,U.S.A
SPECIFICA T IONS Accuracy
: ±3% of full scale.
Maximum Internal Working Pressure:
15 psi (100 kPa). Maximum Working Temperature: 150°F (65°C).
Scale Length: Approx. 8-1/4˝ (21 cm). Weight: 10 oz (283.49 g). Accessories: 3/4 oz. gage oil, mounting
screws, 1/8˝ NPT tubing adapters, wash­ers, nuts, red and green pointer flags, (2)
4-1/2 ft.
Phone: 219/879-8000 www.dwyer-inst.com
lengths 3/16˝ I. D. vinyl tubing.
Portable Applications: For portable
use, a molded plastic swing-away foot and leveling screw are provided. Referring to the picture on the first page of this bulletin, secure the swing-away foot to the bottom of the gage, using the plastic screw in the tapped hole on the right-hand side. Install the plastic leveling screw in the tapped hole on the bottom left side of the manometer. In use, the swing-away foot is positioned per­pendicular to the gage. With the manometer in operating position, the leveling screw is adjusted until the bubble is centered in the spirit level. For storage, the swing-away foot can be rotated to a position in line with the gage body.
FILLING
T urn the zero adjust knob fully counter clock­wise until it stops; then turn it clockwise approximately four full turns so as to center the adjustment to allow room for adjusting either side of zero. Unscrew the entire low pressure (right) shut-off connection fitting. A 3/4˝ wrench is required. Also, remove the cork, disc, and “O” ring. Vent the left con­nector by turning the elbow 1 - 2 turns counterclockwise. Slowly fill gage with fluid provided until fluid rises in the indicating tube to the vicinity of zero on the scale. Nearly all the oil provided will be required to fill the gage. Replace the cork, disc, “O” ring and the low-pressure shut-off connection fitting.
OPERATION
The Model 40 manometer is provided with convenient shut-off tubing connections to prevent loss of oil when transporting the gage. To use the gage, turn each barbed tubing connection counterclockwise 1-2 turns from the closed (fully clockwise) posi­tion. Rotate the zero adjustment for an exact zero reading and the gage is ready for use. In stationary installations, these steps need only be performed upon initial installation. In portable use, each time the gage is used the connections must be opened, the gage lev­eled and zeroed; prior to storage, the gage connections must be closed.
Pressure Connection: Two 4-1/2˝
lengths of plastic tubing are included with the gage along with two adapters for con­nection of the tubing to 1/8˝ NPT fittings or sheet metal ducts. The tubing from the high pressure pickup should be connected to the high pressure (left) barbed connection at the top of the gage. The tubing from the low pressure pickup should be connected to the low pressure (right) barbed fitting. Stick-on red and green arrows are provided to indi­cate the appropriate range of readings for the particular application. The 1/8˝ NPT tubing adapters can be used as static pres­sure sensors in sheet metal ducts by drilling a 7/16˝ diameter hole in the duct and mounting the adapter in the hole, placing one of the washers on each side of the duct wall and securing the 1/8˝ NPT nut.
CAUTION
USE .826 SPECIFIC GRAVITY RED GAGE OIL FOR MODEL #’S 40-1, 40-1­AV, 40-250 Pa, 40-250 Pa-AV, and 40­25 mm. Use 1.910 SPECIFIC GRAVITY BLUE GAGE OIL FOR MODEL #’S 41­2, 41-2-AV, 41-600 PA, 41-60 mm.
TYPICAL APPLICATIONS Draft Gage: Run iron pipe, 1/8˝ or larg-
er in size, from the source of the draft to a point within 4-1/2˝ of the manometer. Provide a means for periodic clean-out to remove any soot accumulation. Using one of the lengths of tubing provided, make the connection from the pipe to the low pres­sure (right) fitting on the gage.
Loading...