Dwyer Instruments IWP Operating Manual

Series IWP Industrial Weatherproof Pressure Transmitter
1
-43/64
[
42.50]
1
-1/4
[
31.65]
2X 1/2˝ FEMALE NPT PLUGS
2-31/64
[63.25]
1/2˝ FEMALE NPT PRESSURE PORT
1
-3/64
[
26.50]
2-1/64
[51.00]
1-11/32 [34.25]
B
ACK LID REMOVED
POWER
4 TO 20 MA
METER
RL
FIELD
TERMINALS
OUT TEST
Specifications - Installation and Operating Instructions
he Series IWP Pressure Transmitter provides an excellent solution to pressure
T
easurement in industrial conditions. The transmitter delivers a 4 to 20 mA signal
n response to the pressure being measured against its stainless steel diaphragm.
i The 4-digit LED display allows for easy field inspection of pressure in psi units. The IP65 rated or industrial conduit entry enclosure ensures the high quality sensor
ircuitry is able to maintain precise operation under dirty and wet conditions, making
c the Series IWP an ideal choice for petroleum, chemical and metallurgical industry applications.
FEATURES
Industrial weatherproof conduit entry enclosure
• 4-digit LED display
PECIFICATIONS
S
ervice: Gases and liquids compatible
S
ith wetted materials.
w Wetted Materials: 304 and 316 stainless steel.
ccuracy: 0.5% FS.
A Stability: <0.2% FS per year. Temperature Limits: -22 to 185°F (-30
to 85°C).
Compensated Temperature Limits:
4 to 158°F (-10 to 70°C).
1
ressure Limits: 1.5 x pressure range.
P
Bulletin P-IWP
emperature Coefficient: 0.3% F.S.
T
er 10°C.
p
ower Requirements: 12 to 36 VDC.
P Output Signal: 4 to 20mA. Loop Resistance: 1200 Ω max.
lectrical Conduit Connection: 1/2˝
E
female NPT. Process Connection: 1/2˝ female NPT.
Enclosure Rating: IP65.
gency Approvals: CE.
A
INSTALLATION Media Compatibility
Series IWP transmitters are designed to be used with any gas or liquid compatible with both 304 and 316L stainless steel.
Environment
The operating temperature limits of the IWP are as follows:
Temperature Limits: -22 to 185°F (-30 to 85°C); Compensated Temperature Limits: 14 to 158°F (-10 to 70°C).
Pressure Fittings
Typically, standard pipe fittings and installation procedures should be used. However, for very high pressure ranges (such as those in excess of 500 psig), we recommend the use of a sealant such as Loctite Sealant.
Moisture Precautions
The Series IWP is provided with two 1/2˝ female NPT conduit entries for electrical termination. In order to prevent ingress of moisture to the IWP, the ports must be sealed in accordance with standard industry practice. It is recommended that a sealer such as PTFE tape be applied around the threads of the conduits entering the IWP.
Note: Buttons in front of Series IWP transmitters are not meant for customer use. Tampering will void warranty.
©Copyright 2015 Dwyer Instruments, Inc. Printed in U.S.A. 4/15 FR# 444288-00
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyermail.com
®
569 Thread Sealant, Hydraulic
ELECTRICAL INSTALLATION
Wiring is to enter the IWP by one of the 1/2˝ female NPT conduit entries. Remove the rear solid metal cover by twisting counterclockwise.
The terminal strip, as pictured above, will be shown. A power supply with an output of 12 to 36 VDC must be connected to the device at the “OUT” terminals (two left terminals). Unscrew the terminal blocks, insert each wire (observing polarity), and retighten the terminal blocks. The device will deliver a 4 to 20 mA rated current to any external load, such as an ammeter, provided its resistance is between 1 and 1200 ohms.
MAINTENANCE/REPAIR
After final installation of the Series IWP Industrial Weatherproof Pressure Transmitter, no routine maintenance is required. A periodic check of system enclosure integrity is recommended to ensure against ingress of moisture. These devices are not field repairable and should be returned to the factory if service is required. After first obtaining a Return Goods Authorization (RGA) number, send the material, freight prepaid, to the following address. Please include a clear description of the problem plus any application information available.
Dwyer Instruments, Inc. Attn: Repair Department 102 Highway 212 Michigan City, IN 46360
Loctite® is a registered trademark of Henkel Corporation
Phone: 219/879-8000 www.dwyer-inst.com