Dwyer Instruments HM28D3H30000, HM28D3H10000, HM28D3F30000, HM28D3F10000, HM28D3C30000 User manual

...
Operating instructions Digital handheld pressure gauge
HM28
www.GlobalTestSupply.com
Contents
2 Safety details ...................................4
3 Operating .........................................4
3.1 Switching on and off ........................4
3.2 Pneumatic connection .....................4
3.3 Operating modes .............................4
3.4 Configuration ...................................6
3.5 RS 232 mode ................................... 7
3.6 Battery change ..............................10
3.7 Recalibration.................................. 10
4 Specifications ................................ 10
4.1 Technical data ............................... 10
4.2 Measuring ranges ..........................12
4.3 Mains supply unit connection ........12
5 Maintenance ..................................12
6 Warning messages and faults ....... 13
7 Accessories ...................................13
8 Calibration instructions ..................14
2
www.GlobalTestSupply.com
Operating instructions
ON = On OFF = Off MODE = Select operating mode
- Pressure measurement
- Min./Max.
- Leak rate
- Datalogging CLEAR/ZERO = Zero setting PRINT/HOLD = Print and freeze display START/STOP = Data recording / printing SET = Select config. modes SELECT = Select config. data
c Matrix display d Connection for pneumatic hoses e Connection for plug-in power supply unit f Connection for RS 232 interface
d
c
e
f
Important!
Please note warning symbol in the operating instructions.
Copyright (c) by LOVE CONTROLS. The contents of this publication may not be copied without the express permission of LOVE CONTROLS. Subject to dimension and design changes.
www.GlobalTestSupply.com
3
1 Description
The „LOVE CONTROLS“ handheld pressure gauge HM28 is a digital pressure measuring instrument with integrated pressure sensor for the measurement of differential, relative or absolute pressures and vacuum. Its versatile range of functions and high precision makes it suitable for a wide range of applications.
Configuration possibilities:
Autom. switching off, selectable time
Selectable resolution and damping
Selectable units of time
Selectable measurement units
General reset to standard setting
Selectable data transfer rate
Configuration printout
Selectable measuring intervals
2 Safety details
Switching off Press the OFF key, or auto-
matic switching off 1,10 or 60
min after the last keystroke. In case of a temperature change the instrument requires at least 30 minutes to adjust to the new ambient temperature.
3.2 Pneumatic connection
For measuring ranges up to 7000 mbar, 4/6 mm or NPT 1/8" connectors can be connected depending on model. Higher ranges are equipped with NPT 1/8".
Ensure correct connection of the pneumatic hoses!
+P high pressure S- low pressure (not available
with the absolute and relative pressure version)
The pressure values stated on the rating plate and mentioned in these operating instructions must not be exceeded otherwise the pressure sensor can be destroyed.
Do not use the instrument in danger zones (explosives zones)! Wear eye protection for pres­sures > 1bar!
3 Operating
3.1 Switching on and off
Switching on Press the ON key (the maxi-
mum permissible measuring range and the class appears on the display). For precision measurements the instrument must be switch­ed on for at least 1 minute (warm-up phase).
www.GlobalTestSupply.com
Measuring negative differential or relative pressures:
connect “S-” or
change ports “S-“ and “+P”.
The HM28 will display a positive value.
3.3 Operating modes
The modes in the matrix below can be selected in succession by pressing the MODE key.
After switching ON, the instrument is always in the mode „normal“.
4
Î negative pressure,
When screwing on a coupling it is important to hold with a span­ner in the opposite direction. Never hold at case only!
Switching on
Normal (differential) and analogue/hold
Normal (absolute) and differential
Zoom
Min/Max
Leak rate
Tendency (absolute)
Record 1)
Print Record
print hold
print
print
print
print
print
start/stop
print
quick keystroke: max. output speed
Keystroke approx.1s: single or fast output start/stop printing
zero
display
zero
diff
zero
display
zero
(reset)
zero
display
clear
memory
Display example
Max. Range:
200 mbar
Full scale error
0.05 %
123.45 mbar ====== or HLD 123.4 mbar
1013.2 mbar
DIF 1013.2 mbar
123.45
MAX 150.0 mbar MIN 100.0 mbar
1234.5 mbar
LEK 2.1 /min
1013.2 mbar
TND - 15.7 /min REC 432.1 mbar
STOP 30 s MEM 1 REC 432.1 mbar
RUN 30 s MEM 321
432.1 mbar STOP 30 s MEM 901
PRINT RECORD PUSH PRINT
PRT 432.1 mbar 321 00:13:30
PRT Record stopped
Notes
Measuring range displayed in desired unit
Accuracy class as a % of full scale reading
Only with differential or relative pressure models
Hold, store indicated value
Only with absolute pressure mode Zero
sets Diff to 0 Measurement in large
figures With absolute pressure,
reset to actual value
Only with Diff./Rel.
(unit/time unit)
Only with absolute pressure (unit/time unit)
If measurement memory cleared
Recording runs up to 964 values
Recording stopped
Appears only if values are stored
Printing / display with time
Recording stopped
1) at 70 bar the measurements are stored in [bar]
5
www.GlobalTestSupply.com
3.4 Configuration
Select mode: Press >2 seconds
Store and exit: Press >2 seconds
Î on the display appears
CONFIGURATION
Î on the display appears
SAVE CONFIGURATION
set unit mbar, kPa, bar, 2)
mmH2O, mH2O, mmHg, psi, inH2O, inHg, hPa, Pa, MPa
set high resolution1) low
set damping off
on
set baudrate 9600, 4800, 2400,
1200
set auto-off 1, 10, 60 min continuous continue
set time unit minutes hours hours
set 1, 5, 10, 20, 30, 60 s Rec. interval 2, 3, 5, 10, 30, 60 min
1) manual, off
set 2.5 Hz (400 ms) display rate 5 Hz (200 ms)
Display example
UNIT mbar
RESOLUTION Display resolution, high influences measuring rate
in RS 232 mode
DAMPING 90% recovery after 4 off measurements (when ON)
BAUDRATE 9600
AUTO OFF Automatic switching off or 10 min continuous mode
TIME UNIT For leak rate, tendency minutes
REC INTERVAL Off = record mode is 5 s (1.3h) deactivated
(max. record interval)
DISPLAY RATE Normal mode indication/zoom
2.5 Hz (400 ms) (influences DAMPING)
Notes
general reset? set on default values zero
print print now configuration?
1) If changed, the measurement memory is cleared 2) See in 4.1 "Technical data"
GENERAL RESET? Reset all settings and clear PUSH ZERO measurement memory
GENERAL RESET? After actuating ZERO RESET OK
PRINT CONFIG? Unit, ... PUSH PRINT
PRINT CONFIG? After actuating PRINT PRINTING
6
www.GlobalTestSupply.com
3.5 RS 232 mode
The HM28 can be remotely operated from a personal computer via an RS 232 interface. The automatic switching off (chapter 3.1) is inactive. Connection by means of a RS232­cable.
Plug pin assignment RS 232, DB 9 (female)
The answer is terminated with a CR
- The checksum is formed from the least significant byte of the command string
Code-decoding of control command
«readconfig»
Pin Assignment Pin Assignment 1 DCD 6 DSR 2 TXD 7 RTS 3 RXD 8 CTS 4 DTR 9 SHIELD 5 GND
Protocol
ASCII-commands 9600/4800/2400/1200 baud, 8 bit, no parity, 1 stop bit, XON/XOFF - protocol (17 dec/19 dec)
Notes
- The HM28 is switched on when the supply voltage is applied
- The HM28 sends "XON" every 3 seconds To determine the baud rate, read-in "XON" in each transfer rate until "XON" is correctly identified
- To go to remote control mode, interrogate "XON", immediately send the "remote" command and then read-in the acknowledgement "ok"
- Only lower case letters are accepted
- Observe > 0,1 s delay before the next command after "setbaud" command
- To acknowledge the answer of the HM28 without reading it back, just send a "XON"
- The symbol "*" followed by the checksum must be suffixed to each string. The string must be terminated with a CR
- A "TAB" is prefixed to each answer of the HM28, a "*" suffixed and the checksum sent.
www.GlobalTestSupply.com
Code Configuration
Pressure unitPressure unit
Pressure unit
Pressure unitPressure unit
5 MPa(7-70bar) 6 Pa(25mbar-7bar) 7 kPa 8 bar 9 mH2O(with 70bar,
otherwise mmH2O)
10 mmHg(not with
70bar) 11 psi 12 inH2O 13 inHg (not with
70bar) 14 hPa 15 mbar
ResolutionResolution
Resolution
ResolutionResolution
0 low 1 high
DampingDamping
Damping
DampingDamping
0on 1 off
BaudrateBaudrate
Baudrate
BaudrateBaudrate
0 1200 Baud 1 2400 Baud 2 4800 Baud 3 9600 Baud
Code Configuration
Switch. off timeSwitch. off time
Switch. off time
Switch. off timeSwitch. off time
0 60 min 1 continous 2 1 min 3 10 min
Time unitTime unit
Time unit
Time unitTime unit
0 /hours 1 /minutes
Storage intervalStorage interval
Storage interval
Storage intervalStorage interval
2 10 s 3 20 s 4 30 s 5 60 s 6 2 min 7 3 min 8 5 min 9 10 min 10 30 min 11 60 min 12 manual 13 off 14 1 s 15 5 s
Display rateDisplay rate
Display rate
Display rateDisplay rate
0 5 Hz 1 2.5 Hz
Reply to control command «readconfig»
A whole number is returned as a reply
msb lsb
#### #### #### #### |||| |||| |||| ****- Pressure unit |||| |||| |||*------ Resolution |||| |||| ||*------- Damping |||| |||| **-------- Baudrate |||| ||**----------- Switching off time |||| |*------------- Time unit |***-*-------------- Storage interval *------------------- Display rate
7
block keypad
"RESOLUTION high" output10 M/s
"RESOLUTION low" output 20 M/s
following send "XON"
(full=ok, empty=change)
(tab)[record_stopped] * 200 CR Recording is stopped
(tab)[out_of_range] * 164 CR Measurement out of range
(tab)[record_end] * 0 CR Recording stopped (waits at "XON"
and sends back "ok" )
±4°C (±39 °F)
change to keypad
Control commands/syntax Answer from HM28/syntax Description
remote * 182 CR (tab)[ok] * 13 CR Switch to remote control and
local * 53 CR (tab)[ok] * 13 CR Switch to keypad
off * 101 CR (tab)[ok] * 13 CR Switch off instrument
readpress * 243 CR (tab)"Value" _ "Unit" * Checksum CR Interrogate measurement
readpressfast * 161 CR (tab)"Value" * Checksum CR (tab)"Value" * Checksum CR Fast measurement interrogation
$ (tab)[ok] * 13 CR Exit "readpressfast" mode
readrange * 211 CR (tab)"Range" _ "Unit" * Checksum CR Interrogate measuring range
readbat * 253 CR (tab)"full"/"empty" * 230/98 CR Interrogate battery capacity (lobat)
readrecord * 69 CR (tab)"Recinterval" _ (tab)"Value" _ (tab)"Value" _ (tab)"Value" Interrogate stored measured
8
www.GlobalTestSupply.com
readtemp * 124 CR (tab)"Value" _ °C * Checksum CR Interrogate internal temperature,
clearmem * 112 CR (tab)[ok] * 13 CR Clear datalogging memory
readconfig * 60 CR (tab)Code (see decoding in table page 17) * Checksum CR Readout actual configuration
setzero * 54 CR (tab)[ok] * 13 CR Zero indication
setdefault * 91 CR (tab)[ok] * 13 CR Reset all settings and clear memory,
setunit_kpa * 146 CR (tab)[ok] * 13 CR Corresponding units
setunit_mbar * 248 see 4.1 Technical data
setunit_bar * 139
low: 20 M/s
off at RS232 mode
periodically (default value)
Control commands/syntax Answer from HM28/syntax Description
setunit_mmh2o * 57
setunit_mmhg * 255
setunit_psi * 162
setunit_inh2o * 54
setunit_inhg * 252
setunit_hpa * 143
setunit_pa * 39
setunit_mpa * 148
setbaud_9600 * 1 CR (tab)[ok] * 13 CR Select baud rate (answer with
setbaud_4800 * 254 new baud rate, wait >0,1s
setbaud_2400 * 248 before "XON")
setbaud_1200 * 245
resolution_high * 62 CR (tab)[ok] * 13 CR Select measurement resolution
resolution_low * 240 and indication, high:10 M/s,
settempmode_0 * 33 CR (tab)[ok] * 13 CR Temp. measuring will be switched
settempmode_1 * 34 Temp. measuring will be done
setrecint_off * 86 CR (tab)ok * 13 CR Select storage interval (secondes)
setrecint_man * 87 (tab)ok
setrecint_1 * 76 (tab)ok
setrecint_5 * 80 (tab)ok
setrecint_10 * 124 (tab)ok
setrecint_20 * 125 (tab)ok
setrecint_30 * 126 (tab)ok
setrecint_60 * 129 (tab)ok
setrecint_2m * 186 (tab)ok Select storage interval (minutes)
setrecint_3m * 187 (tab)ok
setrecint_5m * 189 (tab)ok
setrecint_10m * 233 (tab)ok
9
www.GlobalTestSupply.com
3.6 Battery change
- Open battery receptacle
- Insert 9 V-alkali battery (IEC 6LR61) or accumulator
Ensure correct polarity!
Correct disposel of the used batteries according to environment regulations!
3.7 Recalibration
Recalibration to be carried out by specially skilled staff only.
Relevant instructions see in section 8. We recommend to recalibrate the instrument at least once a year.
4 Specifications
Control commands/syntax Answer from HM28/syntax Description
setrecint_30m * 235 CR (tab)ok * 13 CR
setrecint_60m * 238 (tab)ok
setautooff_man * 198 CR (tab)ok * 13 CR Select autom. switching off time (minutes)
setautooff_1 * 187 (tab)ok
setautooff_10 * 235 (tab)ok
4.1 Technical data
Measuring media instrument air or
Media-compatible all media compatible types with stainless steel
(tab)[er] * 10 CR Error
setautooff_60 * 240 (tab)ok
setdamp_off * 115 CR (tab)ok * 13 CR Damping off
setdamp_on * 21 (tab)ok Damping on
setdisrate_2.5 * 23 CR (tab)ok * 13 CR Set display rate (Hz)
setdisrate_5 * 183 (tab)ok
setunit_perh * 5 CR (tab)ok * 13 CR Tendency .../h
setunit_permin * 225 (tab)ok Tendency .../min
Note: The character _ indicates a space
10
www.GlobalTestSupply.com
inert gases
18/8 (DIN 1.4305)
Units Measuring ranges
up to 7 bar 10 to 30 bar 70 bar
mbar x x x bar x x x Pa x - ­kPa x x x hPa x x ­MPa - x x mmH
Ox x -
2
mH
O- -x
2
mmHg x x ­psi x x x inH
Ox xx
2
inHg x x -
Linearity, hysteresis and repeatability (10 °C to 35 °C) ± 0,2 % F.S. (50 °F to 95 °F) (standard) ± 1 digit
± 0,1 % F.S. (option) ± 1 digit ± 0,05 % F.S. (option) ± 1 digit (according to measuring range)
Operating temperature -5 °C to 50 °C
(23 °F to 122 °F)
Storage temperature -20 °C to 60 °C
(-4 °F to 140 °F) Humidity 30 to 95 % rH Case protection class IP 54 Power supply 9 V-battery (IEC
6LR61) or
accumulator
regulated plug-in
mains supply unit
(7 to 14 VDC) Current consumption <9 mA Operating time (battery) appr. 70 h Baud rate RS232 9600/4800/2400/
1200 baud
Measuring rate in RS232- mode
-Class 0,2 20 measurements/s
-Class 0,1 and 0,05 10 measurements/s Measuring rate, 2 ½ or normal mode 5 measurements/s Memory capacity max. 964
measurements
Memory interval manual,
1,5,10,20,30,60 s 2,3,5,10,30,60 min
Display LCD matrix,
2 lines of 16 characters
Pneum. ports hose 4/6 mm or
NPT 1/8" Case dimensions 152x83x34/29 mm Weight incl. battery 270 g
11
www.GlobalTestSupply.com
4.2 Measuring ranges
Metric Range English Range Max. load capacity Max. static pressure
(rounded)
0 ... 2.5 kPa 0 ... 10 inH2O 12.5 kPa 50 inH2O 700 kPa 100 psid 0 ... 7 kPa 0 ... 28 inH2O 35 kPa 140 inH2O 700 kPa 100 psid 0 ... 20 kPa 0 ... 80 inH2O 150 kPa 600 inH2O 700 kPa 100 psid 0 ... 30 kPa 0 ... 120 inH2O 150 kPa 600 inH2O 700 kPa 100 psid 0 ... 50 kPa 0 ... 200 inH2O 400 kPa 1600 inH2O 700 kPa 100 psid 0 ... 100 kPa 0 ... 14.5 psid/g 400 kPa 58 psid/g 700 kPa 100 psid 0 ... 110 kPa 0 ... 15.9 psia 400 kPa 58 psia -- -­0 ... 200 kPa 0 ... 29 psia/d/g 700 kPa 100 psia/d/g 700 kPa 100 psid 0 ... 700 kPa 0 ... 100 psia/d/g 1700 kPa 245 psia/d/g 1700 kPa 245 psid 0 ... 1000 kPa 0 ... 145 psid/g 2700 kPa 390 psid/g 2700 kPa 390 psid 0 ... 1700 kPa 0 ... 245 psid/g 2700 kPa 390 psid/g 2700 kPa 390 psid 0 ... 3000 kPa 0 ... 435 psig 7000 kPa 1000 psig -- -­0 ... 7000 kPa 0 ... 1000 psig 14000 kPa 2000 psig -- --
a = absolute pressure d = differential pressure g = relative pressure
Conversion factors 1 mbar = 0,1 kPa 1 mbar = 0,0010 bar
1 mbar = 10,20 mmH 1 mbar = 0,7501 mmHg 1 mbar = 0,0145 psi 1 mbar = 0,4015 inH 1 mbar = 0,02953 inHg 1 mbar = 1,0 hPa
The maximum load capacity applies for
))
)
))
relative overpressure and negative pressure. The instrument is calibrated
from 0 to100% of the measuring range. Exceeding or underrunning this
range by up to about 10% is still displayed.
Differential pressure sensors (d)
))
)
))
doesn’t measure the same value on the P and S side mandatorily because of her geometry.
O
2
O
2
www.GlobalTestSupply.com
4.3 Power supply unit connection
The HM28 can be operated by a regulated plug-in power supply unit.
Input 115V, 60 Hz Output 9 V DC ± 20 %, 100 mA
(7 to 14 V DC)
5 Maintenance
The HM28 requires no maintenance. It can be cleaned with a damp cloth. Do not use cleaning agents containing solvents! See the relevant chapters for battery change and recalibration.
12
6 Warning messages and faults
Fault/indication Possible cause Remedy
ERROR OUT OF RANGE
CHANGE BATTERY
No change in measurement
Does not switch on
Instrument inaccurate
10% exceeding or under­running of measuring range
Battery voltage too low
Over pressure applied to pressuresensor
No power supply
Inaccurate recalibration
Not zeroed
Natural aging of the pressure sensor
7 Accessories
Standard 1 9 V block battery
1 operating instructions
Option Plug-in power supply unit
115V, 60Hz Leather case with carrying strap Service set Hand pump with variobellows5bar Handpump 20 bar SCS test certificate Adapter RS232 9M-25F Adapter NPT1/8" Communication SW package
comprising:
- RS232-IF cable (9 pole fem.)
- Communication-software for MS Windows
- Measurement places management software for MS Windows
Apply permissible measuring pressure
Insert new battery
Dispatch instrument for repair
Fit new battery as required Battery possibly not correctly inserted Plug in power supply unit correctly
Repeat recalibration
Vent and actuate zero
Carry out recalibration
13
www.GlobalTestSupply.com
8 Calibration instructions
Recalibration to be carried out by specially skilled staff only!
Actuate the following key combinations:
• First press MODE and keep it pressed
• Next and additionally, press CLEAR/ ZERO and keep pressed until CALIBRATION is displayed
Exit from this mode is possible at any time via MODE or OFF.
The instrument must be in operation for at least 30 minutes (warm up stabilization time).
New calibration values are only stored when all five setpoints have been running through correctly and
Calibration at 22 °C (71,6 °F) environmental temperature
An incorrect internal instrument tempera­ture can lead to accuracy fluctuations!
The pressure set points have to be set in [mbar]!
Check the instrument accuracy after recalibration!
acceptable values acquired. In the event of a maloperation, the calibration process is interrupted and the previous calibration data are retained.
noitarbilaC
tnioP
.S.F%0
.S.F%52rabmxTES
.S.F%05rabmxTES
.S.F%57rabmxTES
.S.F%001rabmxTES
yalpsiDetucexEsetoN
NOITARBILAC
C°22BILAC rabmxTES
KO
KO
KO
KO
KONOITARBILAC
14
www.GlobalTestSupply.com
xtnioptesteS-
RAELC/OREZsserP-
rosnes
elbaliavaton.S.F%0
erusserpetulosbahtiw
xtnioptesteS-
RAELC/OREZsserP-
xtnioptesteS-
RAELC/OREZsserP-
xtnioptesteS-
RAELC/OREZsserP-
xtnioptesteS-
RAELC/OREZsserP-
www.GlobalTestSupply.com