Dwyer Instruments AT2200 Specifications

Series
6-19/64
[160.00]
1/8˝ N
PT
1-31/32
[50.00]
19/32
[
15.00]
Ø
5/16
[Ø8.00]
15/16
[
24.00]
2-9/16
[65.00]
1
9/32
[15.00]
1/8˝
NPT
5-9/32
[134.00]
8-5/64
[205.00]
7
-11/64
[182.00]
1-37/64
[40.00]
7-7/8
[
200.00]
6
-19/64
[160.00]
5
-23/64
[136.00]
INCHES OF WATER
MAX. PRESSURE 15 PSIG
0
.20
.40
.
60
.
80
1.0
AT2
2000
®
ATEX Approved Series 2000 Magnehelic
PRESSURE
Differential Pressure Gage
Magnehelic®Gage in Flame-Proof ATEX Enclosure
Differential Pressure Gages
The popular Magnehelic®Differential Pressure Gage is now available in a flame­proof ATEX enclosure with the new ATEX approved AT22000 Series. This gage can indicate positive, neg ative or dif ferential pressures and is accurate wit hin 2%. The M resistance, air velocity, furnace draft, pressure drop across orifice plates, liquid levels with bubbler systems and pressures in fluid amplifier or fluidic systems. Optional gage construction is available for use to either 35 psig (2.42 bar) or 80 psig (5.51 bar). Flame­proof ATEX enclosures are available in aluminum and include a glass window for viewing process pressures status on gage face.
Important notes for installation:
• Cables must be fitted through 1/2˝ NPT cable gland or ATEX conduit (not supplied
• Make sure after cabling to close tight cover and cable gland, in order to keep IP66
• Open cover only after de-energizing instrument.
Attention: Check local safety rules and warnings on unit and manual for a correct
RANGE CHART
Model 2000-00N 2000-00 2000-0 2001 2002 2003 2004 2005 2006 2008 2010 2012 2015 2020 2025 2030 2040 2050 2060 2080 2100 2120 2150 2160
8
®
agnehelic
with instrument).
rating (IP65 with option OPV, overpressure relief valve).
use of the instrument in hazardous area.
age is the industry standard to measure fan and blower pressures, filter
g
Range in w.c.
.05 to 0 to .2 0 to .25 0 to .50 0 to 1.0 0 to 2.0 0 to 3.0 0 to 4.0 0 to 5.0 0 to 6.0 0 to 8.0 0 to 10 0 to 12 0 to 15 0 to 20 0 to 25 0 to 30 0 to 40 0 to 50 0 to 60 0 to 80 0 to 100 0 to 120 0 to 150 0 to 160
DWYER INSTRUMENTS, INC. | www.dwyer-inst.com
Example Housing Range Pressure Rating
Temperature Rating Housing Material Cover Process Connection Overpressure Plug
Tag
AT2 AT2
2001
2XXX
X
-X
-MP
-HP
X
X LT
A
-A
O
O
1
1 2
X
X OPV
T2
T2
SPECIFICATIONS Service: Air and non-combustible, compatible gases. Wetted Materials: Consult factory. Magnehelic
Exterior finish is coated gray to withstand 168 hour salt spray corrosion test.
Housing Material: Aluminum. Finishing: Texture epoxy coat RAL7038. Accuracy: ±2% of FS (±3% on -0, -100 Pa, -125 Pa, -10 mm and ±4% on -00, -60
Pa, -6 mm ranges), throughout range at 70°F (21.1°C). Pressure Limits: -20 in Hg to 15 psig (-0.677 bar to 1.034 bar); MP option; 35 psig (2.41 bar), HP option; 80 psig (5.52 bar). Overpressure: Relief plug opens at approximately 25 psig (1.72 bar), standard gages only. Temperature Limits: 20 to 140°F (-6.67 to 60°C); Low temperature option: -20°F (-28.8°C); Case: -76 to 140°F (-60 to 60°C) (Note: Product temperature limits are less than case limits).
Mounting Orientation: Diaphragm in vertical position. Enclosure Rating: IP66. IP65 with option OPV, overpressure relief valve. Process Connections: 1/8˝ NPT female brass (SS optional). In presence of
acetylene it is necessary to use SS.
Weight: 8.6 lb (3.9 kg). ATEX Approved Product from Comhas with ECN: BVI 14ATEX0072. Agency Approvals: CE 1370 II2 GD Ex d IIC Gb T6; -60°C ≤ Ta ≤ +60°C Ex tb
IIIC Db T 85°C.
AT22001-XX-A01XT2 ATEX Approved Series 2000 Magnehelic Specify range by using Magnehelic Standard from -20 in Hg to 15 psig static pressure Medium pressure-max. static 35 psig High pressure-max. static 80 psig Standard temperature limits -6.67 to 60°C Low temperature limit to -28.8°C Aluminum
Glass cover 1/8˝ NPT female brass ports 1/8˝ NPT female SS ports Standard without overpressure relief valve Overpressure relief valve Material same as ports SS information label
®
Housing: Die cast aluminum case & bezel with acrylic cover;
u
®
Differential Pressure Gage
®
model number. See Ranger Chart
uOver Protection Note: See page 5 (Series 2000)
Loading...