MODEL 670
FUME HOOD MONITOR
Ensures Proper Fume Hood Performance
AIR QUALITY
The Model 670 Fume Hood Monitor continuously senses air ow through the face
of the fume hood, ensuring safe levels of fresh air are exhausting. The 670 provides
a highly accurate hot wire sensor to detect very low ows common on fume hoods.
The Model 670 comes with everything required to quickly install the unit including a
mounting bracket, 24˝ of tubing for connecting to the inside of the hood wall and a 120
Volt AC power adapter.
FEATURES/BENEFITS
• Flexible surface or ush mounting
• LED safe and alarm status indicators
• Audible alarm
• Sash alarm input
• Night time set-back
APPLICATIONS
• Fume hood ventilation monitoring
MODEL CHART
Model Description
670 Fume hood monitor
13/64
3-29/64 [87.88]
SPECIFICATIONS
Service: Fume hood face velocity air ow.
Alarm Range: 30-400 FPM (0.15-2.0 m/s).
Alarm Indication: Red LED & audible alarm.
Low Air Velocity Alarm Delay: Fixed 5 secs.
Visual LED Display: Red: Alarm; Green: Normal.
Horn Silence: Yes-temporary and permanent.
Accuracy: Face velocity ±10%.
Temperature Limits: Operating temperature: 55 to 86°F (13 to 30°C); Storage
temperature: -40 to 150°F (-40 to 65°C).
Power Requirement: 15 VDC 500 mA; 120 VAC, 60 Hz power transformer
included.
Relay Output Low Air Flow Alarm: 5 A @ 250 VAC.
Relay Input For Night Setback: 2 wire rated for 24 VDC usage.
Sash High Indication: Using a two wire micro switch or 3 wire proximity switch
input, rated for 24 VDC usage.
Mounting: Semi ush, ush or surface mounted when using included bracket.
Weight: 5.0 oz (141 g).
[5.08]
1-1/16
[27.10]
5-45/64
[145.03]
4-61/64
[125.98]
4-9/16
[116.08]
3-17/64 [83.06]