The Series 634ES Pressure Transmitter senses a single
air, compatible gas or liquid pressure and converts it into
a standard 4-20 mA output signal. Ranges are available
from 0-10 through 0-6000 psi. All models are field
adjustable so any range within these limits can be
achieved by recalibration using the easily accessible
span and zero potentiometers.
Positive pressure can be measured within an accuracy of
±0.5% of span. The Series 634ES uses an isolated
piezoresistive pressure sensor to produce a resistance
change across a wheatstone bridge. The signal is conditioned and converted into a 4-20 mA output signal.
For applications requiring direct pressure or percent of
full span readings, the optional A-701 digital readout
makes an ideal companion device, providing a bright .6”
high, 3-1/2˝ digit LED and supplying power to the Series
634ES Transmitter.
PHYSICAL DATA
GENERAL
Maximum Pressure: See chart on this page.
Wetted Parts: 316, 316L Stainless Steel.
Housing: Designed to meet NEMA-4X.
ELECTRICAL
Power Supply: 12.3-35 VDC-2 wire.
Output Signal: 4-20 mA DC (limited at 38 mA DC).
Loop Resistance: 0 - 1100 ohms from 12.3 to 35 VDC.
R
L max
=
Current Consumption: DC: 38 mA max.
MATERIALS
Housing: Cast aluminum; textured gray polyurethane
finish.
Pressure Connection: Stainless Steel.
MECHANICAL
Weight: 1 lb., 12 oz. (.8 kg).
Span and Zero Adjustments: Protected potentiome-
ters, located in auxiliary housing.
Pressure Connection: 1/4˝ female NPT x 1/2˝ male
NPT.
PERFORMANCE AT 70°F (21.1°C)
Zero Output: 4 mA DC.
Full Span: 16 mA DC.
Accuracy: ±0.5% of span.
Warm-up Time: 10 minutes.
STABILITY/ENVIRONMENTAL
Operating Temperature: 20 to 120˚F (- 6.7 to 48.9˚C).
Thermal Errors: ± 0.02%/˚F typical.
Stability: 1% F.S./yr.
STANDARD ACCESSORIES
(3) “Z” mounting brackets.
(3) 10-32 x 4” RH machine screws.
Series 634ES Pressure Transmitter
Specifications - Installation and Operating Instructions
Bulletin E-69ES
Vps-12.3V
20 mA
Series 634ES Models and Ranges in PSI (bar)
634ES-0
634ES-1
634ES-2
634ES-3
634ES-4
634ES-5
634ES-6
634ES-7
634ES-8
634ES-9
10 (.69)
30 (2.07)
50 (3.45)
100 (6.9)
200 (13.8)
300 (20.7)
500 (34.5)
1000 (69)
2000 (138)
4000 (276)
10 (.69)
20 (1.38)
40 (2.76)
60 (4.14)
100 (6.9)
250 (17.2)
350 (24.1)
600 (41.4)
1250 (86)
2500 (172)
20 (1.38)
40 (2.76)
60 (4.14)
120 (8.3)
250 (17.2)
350 (24.1)
600 (41.4)
1250 (86)
2500 (172)
6000 (414)
30 (2.07)
60 (4.14)
100 (6.9)
200 (13.8)
400 (27.6)
500 (34.5)
1000 (69)
2000 (138)
4000 (276)
7500 (517)
Model Number As Stocked Min. Range Max. Range Max Pressure
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
Lit-By Fax: 888/891-4963
(3) #10-32 x 1/4 [6.35] DP HOLES
EQUALLY SPACED ON A 2-1/2 [63.50] B.C.
TYP
1/2NPT
1-9/16
[39.69]
3/4 NPT
03-3/8 [85.73]
1-11/16
5-5/16
[134.94]
2-1/2 [63.50]
[42.86]
1/4NPT
2-7/8 [73.03]
1/16 [1.59]
25/32 [19.84]

INSTALLATION
LOCATION: Select a location where temperature of the
unit will be between 20°F and 120°F. Distance from the
receiver is limited only by total loop resistance. See
“Electrical Connections.” The tube feeding the pressure
to the instrument can be run practically any length
required, but long lengths will slightly increase response
time. Avoid surfaces with excessive vibration.
POSITION: A vertical position is recommended, as all
stocked models are spanned and zeroed at the factory in
this position. They can be used at other angles, but final
spanning and zeroing must be done while transmitter is in
the alternative position.
PRESSURE CONNECTIONS: A single pressure connection is provided at the bottom of the transmitter housing.
It has1/4” female NPT and 1/2” male NPT threads. Attach
positive pressure to this port.
MOUNTING: The Series 634ES Transmitter can be
mounted three ways:
(A) Supported directly by pipe providing pressure.
(B) Attached to a mounting surface with 10-32 x 1/4”
machine screws (included). The machine screws are
installed through the mounting surface into tapped holes
on back of unit.
(C) Mounted with “Z” brackets (included). Attach “Z”
brackets to tapped holes on back of unit and fasten to
front of mounting surface.
ELECTRICAL CONNECTIONS
CAUTION: DO NOT EXCEED SPECIFIED SUPPLY VOLT-
AGE RATINGS. PERMANENT DAMAGE NOT COVERED
BY WARRANTY WILL RESULT. THIS UNIT IS NOT
DESIGNED FOR AC VOLTAGE OPERATION.
Electrical connections to the Series 634ES Transmitter
are made inside the enclosure. Remove the cover, feed
stripped and tinned leads through the conduit opening
and connect to terminal block screws marked + and -.
See Figure A for locations of terminal block, span and
zero adjust potentiometers. See Figure B (Pg. 3) for
schematic diagram.
FIG. A
An external power supply delivering 12.3 to 35 VDC with
minimum current capability of 40 mA must be used to
power the control loop in which the Series 634ES
Transmitter is connected. See Figure B for connection of
the power supply, transmitter and receiver.
The range of appropriate receiver load resistance (RL) for
the power supply voltage available is given by the formula and graph in Figure C on Pg. 3.
Shielded 2-wire cable is recommended for control loop
wiring, and the cable shielding may be grounded if
desired. Note also that the receiver may be connected in
either the negative or positive side of the loop, whichever
is most convenient.
Should polarity of the transmitter or receiver inadvertently become reversed, the loop will not function properly,
but no damage will be done to the transmitter.
Series 634ES Pressure Transmitter
Specifications - Installation and Operating Instructions
Bulletin E-69ES
Page 2