Dwyer Instruments 477A Data sheet

Series
SERIES 477A DIGITAL MANOMETER
HIGH
PRESSURE
LOW
PRESSURE
1-1/16
[27.13]
5/8
[16.21]
5-7/8
[149.69]
29/32
[22.99]
2-13/16
[71.58]
477A
Handheld Digital Manometer
Precise Air Pressure Measurement, ±0.1% F.S. Accuracy
TEST EQUIPMENT
Manometers,
Portable
The Popular Model 477 is now available with 0.1% full-scale accuracy in the new Series 477A. The 477A contains a highly accurate differential pressure sensor that offers a 0.1% full-
scale accuracy on air ranges from 20 in w.c. t o 100 psid. Series 477A Handheld Digital Manometers are packed faster, easier and more accurate than ever. Instantly select from up to nine of the most widely used pressure units without having to waste time and risk mistakes with tedious conversions. A non-volatile memory function enables storage of up to 40 readings — perfect for HVAC technicians making Pitot tube traverses of airflow readings across a duct. 477A is also ideal for maintenance personnel or technicians that require a highly accurate standard to check their instrumentation or equipment to ensure proper performance. When working in poorly lighted areas, just switch on the handy backlight feature. The manometer automatically shuts itself off after 20 minutes to minimize battery drain. Electronic zeroing means you simply touch a single key to perfectly null out any minor pressure differences. A display HOLD key freezes the current pressure for those all-too­common situations where readings fluctuate. Included is an audible alarm to warn of overpressure plus a visual alarm warning in case ambient noise levels are too high to hear the alarm. Audible alarm also confirms a value has been stored, eliminating the need to observe display during a duct traverse. Clear, concise operating instructions for all functions are printed on the rear of the rugged extruded aluminum case for quick reference. One-piece front membrane fully protects all keys from dust and moisture; wipes clean in seconds. Detailed written instructions, a wrist strap and 9 V alkaline battery are included.
FEATURES
• Measure positive, negative, or differential pressures
• Instant selection from up to nine english/metric units
• Stores 40 readings in memory for later reference
• Both audible and visual overpressure alarms
• Operates up to 100 hours on a single 9 volt battery
• New adjustable damping feature for averaging fluctuating reading
See page 567 for process tubing options.
with features needed to make pressure measurement and recording
The
from 32 to 60 and 78 to 104°F (0-15.6 and 25.6-40°C).
Pressure Hysteresis: ±0.1% of full-scale. Pressure Limits: See chart. Temperature Limits: 32 to 104°F (0 to 40°C). Storage Temperature Limits: -4 to 176°F (-20 to 80°C). Display: 0.42˝ (10.6 mm) 4 digit LCD. Resolution: See chart. Power Requirements: 9 V alkaline battery, installed non-functional, user
replaceable.
Weight: 10.2 oz. (289 g). Connections: Two barbed connections for use with 1/8˝ (3.18 mm) or 3/16˝ (4.76
mm) ID tubing for 477A-1, 477A-2, 477A-3, 477A-4 and 477A-5 only. Two compression fittings for use with 1/8˝ (3.18 mm) ID x 1/4˝ (6.35 mm) OD tubing for 477A-6 and 477A-7 only.
ACCESSORIES A-402A CARRYING CASE — Tough gray nylon pouch protects any
Series 477A Manometer. Double zippered for quick and easy access. With belt loop that snaps closed.
7-1/2˝H x 3˝W x 2-1/4˝D (191 x 76 x 57 mm)
UHH-C1, Soft carrying case
OPTION
For NIST traceable calibration certificate, add suffix -NIST to model numbers. Example: 477A-1-NIST.
Model 477A-1 477A-2 477A-3 477A-4 477A-5 477A-6 477A-7
Range
0-20.00 in w.c. 0-40.00 in w.c. 0-200.0 in w.c. 0-10.00 psi 0-30.00 psi 0-50.00 psi 0-100.0 psi
bar
.0498 .0996 .4982 .6895
2.069
3.447
6.895
psi
.7225
1.445
7.225
10.00
30.00
50.00
100.0
in Hg
1.471
2.942
14.71
20.36
61.08
101.8
203.6
Available Pressure Units
kPa
in w.c.
20.00
40.00
200.0
276.8
830.4 1384 2768
mm Hg
37.36
74.73
373.6
517.1 1551 2585 5171
4.982
9.96
49.82
68.95
206.9
344.7
689.5
mbar
49.82
99.6
498.2
689.5 2069 3447 6895
FTWC
1.667
3.333
16.67
23.07
69.20
115.3
230.7
mm w.c.
508.0 1016 5080 7031
Pa
4982 9964
Maximum Pressure
3 psig 3 psig 15 psig 30 psig 60 psig 100 psig 200 psig
CONTACT US | U.S. 219/879-8000 | U.K. (+44) (0)1494-461707 | A.U. (+61) (0) 2 4272 2055 | China +852-23181007
423