Dwyer VF Manual

Page 1
Series VF Visi-Float® Flowmeters
F
L
FULL OPEN
®
Specications - Installation and Operating Instructions
M
Bulletin F-33
Model VFB Model VFA-SSV
The Series VF Visi-Float® Flowmeters are furnished in two models (see drawing
air, gas or water. Installation, operation and maintenance are very simple and only a few common sense precautions must be observed to assure long, trouble-free service.
CALIBRATION
Each owmeter is calibrated at the factory. If at any time during the meter’s life, you wish to recheck its calibration, do so only with devices of certied accuracy. DO NOT
attempt to check the Visi-Float® Flowmeter with a similar owmeter as seemingly unimportant variations in piping and back pressure may cause noticeable differences
in the indicated reading. If in doubt, return your owmeter to the factory. Before
proceeding with the installation of your Visi-Float® Flowmeter, check to be sure you
have the model and ow range you require.
LOCATION Temperature, Pressure, Atmosphere, and Vibration:
Visi-Float® Acrylic Flowmeters are exceptionally tough and strong. They are designed for use at pressures up to 100 psi (7 bar) and temperatures up to 150°F (66°C).
DO NOT EXCEED THESE LIMITS! The installation should not be exposed to strong
chlorine atmospheres or solvents such as benzene, acetone, carbon tetrachloride, etc. The mounting panel should be free of excessive vibration since it may prevent the unit from operating properly.
Inlet Piping Run: It is good practice to approach the owmeter inlet with as few
elbows and restrictions as possible. In every case the inlet piping should be at least as
large as the connection to the owmeter i.e. 1/8˝ Iron Pipe Size. Length of inlet piping makes little difference for normal pressure fed owmeters.
For ow meters on vacuum air service the inlet piping should be as short and open as
possible. This will allow operation near atmospheric pressure and thereby insure the
accuracy of the device. (Note that for vacuum air service the ow control valve if any, should be on the discharge side of the owmeter. Either the TMV unit or a separate in
line valve may be applied.)
Discharge Piping: As on the inlet, discharge piping should be at least as large as the
owmeter connection. In addition, for pressure fed owmeters on air or gas service the
discharge piping should be as short and open as possible. This will allow operation
of the ow tube at near atmospheric pressure and insure the accuracy of the device. This is of less importance on water or liquid owmeters since the owing medium is
generally incompressible and moderate back pressure will not affect the accuracy of the instrument as calibrated.
A
SPECIFICATIONS Service: Compatible gases and liquids. Wetted Materials: Body: Acrylic plastic; O-ring: Buna-N (uoroelastomer available);
Metal parts: brass standard, SS optional; Float: SS, black glass, aluminum, K monel
depending on range.
Temperature and Pressure Limits: Without valve: 100 psig (6.9 bar) @ 150°F (65°C); 150 psig (10 bar) @ 100°F (38°C); With valve: 100 psig (6.9 bar) @ 120°F (48°C).
Accuracy: VFA = 5% FS; VFB= 3% FS. Process Connection: 1/8˝ female NPT. VFB ranges 85 and 86 have 1/4˝ NPT back
connections or 3/8˝ NPT end connections. These ranges not available with brass
valves.
Scale Length: VFA 2˝ typical length; VFB 4˝ typical length. Mounting Orientation: Mount in vertical position. Weight: VFA: 4.0-4.8 oz (.11-.14 kg). VFB: 7.2-8.8 oz (.20-.25 kg).
DIMENSIONS - FLOWMETER
Model VFA Model VFB
A
4 [101.6]
B
3 [76.20]; 1/8 NPT conn.
C
1-5/8 [41.28]; 10-32 thd
D
1/2 [12.70]
E
1-3/16 [30.16]
F
1-1/4 [31.75]
I
2-1/16 [52.39]; Open
K
4-3/32 [104.0]
L
1 [25.40]
M
7/8 [22.23] ; 1/8 NPT
N
3/32 [2.381]
POSITION AND MOUNTING
All Visi-Float® Flowmeters must be mounted in a vertical position with the inlet connection at the bottom and outlet at the top.
Surface Mounting: Drill appropriate holes in panel using the dimensions shown in the drawing above. Hold the owmeter in position in front of the panel and install the mounting screws through the panel from the rear. Mounting screws must not be longer than the panel thickness plus 1/4˝, or the screw will hit the plastic and may damage the meter. The screws will require additional force during the initial installation, since
the insert boots are of a collapsed thread type and must be expanded into the plastic for the knurled surface to take hold. Insert boots will not have the proper 10-32 threads
until the rst screw has been inserted to expand the boot. Pipe up inlet and discharge
using pipe thread sealant tape or pipe thread sealant to insure against leakage.
C
B
E
D
M
I
6-1/2 [165.1] 5-1/2 [139.7]; 1/8 NPT conn. 3-1/2 [88.90]; 10-32 thd 1/2 [12.70] 1-1/2 [38.10] 1-1/4 [31.75] 2-1/16 [52.39]; Open 6-11/16 [169.9] 1-3/8 [34.93] 7/8 [22.23]; 1/8 NPT 3/32 [2.381]
K
N
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
www.dwyer-inst.com
e-mail: info@dwyermail.com
Page 2
RANGE CHART - VFA 2˝ SCALE - POPULAR RANGES Range No. SCFH Air Range No. LPM Air
1 2 3
4
5 6 7
8 9
Range No. CC/Min. Water Range No. GPH Water
32 33
34
RANGE CHART - VFB 4˝ SCALE - POPULAR RANGES Range No. SCFH Air Range No. LPM Air
50
91
51 52 53
54
55
Range No. SCFM Air Range No. GPH Water
90 .3 to 3 80
Range No. CC/Min. Air
60 100 to 1000
Surface Mounting on Piping Only: An alternate method of surface mounting is by omitting the mounting screws and supporting the Visi-Float® Flowmeter on the connecting piping only. For this method extra long or straight pipe threads should be used so that nuts may be run onto the pipe and later tightened against the back of the
panel to retain the unit in proper position. Use the appropriate hole layout information
from the drawing on the previous page, but omit the small holes.
Mounting on Piping Only Without Panel: For the temporary or laboratory type
installation, the panel may be omitted altogether and the owmeter installed directly in rigid piping. Its light weight permits this without difculty.
OPERATION
To start system, open the valve slowly to avoid possible damage. Control valves on BV and SSV models are turned clockwise to reduce ow, counter clockwise to increase ow (valve is designed for ow adjustment only, not intended to be used as an open/
shut-off valve). A nylon insert is provided in the threaded section of the valve stem to
give a rm touch to the valve and to prevent change of setting due to vibration.
The performance of low range units used in air or gas applications may be affected by
static electricity. Excessive static charge may cause the ball oat to behave erratically
or provide a false reading. To ensure the proper function of the unit, the application should be designed to minimize or dispel static electricity.
The standard technique for reading a Variable Area Flowmeter is to locate the highest point of greatest diameter on the oat, and then align that with the theoretical center of the scale graduation. In the event that the oat is not aligned with a grad, an extrapolation of the oat location must be made by the operator as to its location between the two closest grads. The following are some sample oats shown with reference to the proper location to read the oat.
.1 to 1 .2 to 2 .6 to 5 1 to 10 2 to 20
4 to 30
5 to 50 10 to 100 20 to 200
6 to 50 10 to 100 20 to 200
.3 to 3 1 to 10 2 to 20
4 to 40
10 to 100 15 to 150 20 to 200
21 22 23
24
25 26 27
41 42 43 44
65 66 67
68 69
Range No. CC/Min. Water
82 2 to 30
83 81
Range No. GPM Water
85 86
.06 to 0.5 .15 to 1 .6 to 5 1 to 10 3 to 25 6 to 50 10 to 100
.6 to 5 2 to 10 3 to 20
8 to 40
.2 to 4
1 to 10 1 to 20 3 to 30
4 to 40
.5 to 12 1 to 20 6 to 60
.2 to 2 .5 to 5
CAUTION
will allow gas or liquid to ow out front of valve body and could result in serious personal injury. For applications involving high pressure and/or toxic gases or uids, special
non-removable valves are available on special order. Contact factory for details.
Variable Area Flowmeters used for gases are typically labeled with the prex “S” or “N”, which represents “Standard” for English units or “Normal” for metric units. Use of this prex designates that the owmeter is calibrated to operate at a specic set of conditions, and deviation from those standard conditions will require correction for the calibration to be valid. In practice, the reading taken from the owmeter scale must
be corrected back to standard conditions to be used with the scale units. The correct
location to measure the actual pressure and temperature is at the exit of the owmeter, except under vacuum applications where they should be measured at the owmeter inlet. The equation to correct for nonstandard operating conditions is as follows:
Q2 = Q1 x
Where: Q1 = Actual or observed owmeter reading
Q2 = Standard ow corrected for pressure and temperature
P1 = Actual pressure (14.7 psia + gage pressure) P2 = Standard pressure (14.7 psia, which is 0 psig)
T1 = Actual temperature (460 R + temp °F) T2 = Standard temperature (530 R, which is 70°F)
Example: A owmeter with a scale of 10-100 SCFH Air. The oat is sitting at the 60 grad on the owmeter scale. Actual Pressure is measured at the exit of the meter as 5 psig. Actual Temperature is measured at the exit of the meter as 85°F.
Q2 = 60.0 x
Q2 = 68.5 SCFH Air
MAINTENANCE
The only maintenance normally required is occasional cleaning to assure reliable operation and good oat visibility.
Disassembly: The owmeter can be disassembled for cleaning by simply
disconnecting the piping, dismounting the unit from the panel and removing the top-
plug-ball stop. Take out the ball or oat by inverting the body and allowing the oat
to fall into your hand. (Note: It is best to cover the discharge port to avoid losing the
oat through that opening.) When removing the oat guide assembly on VFB models 85 and 86, be careful not to lose the short pieces of plastic tubing on both ends of the guide which serve as oat stops.
Cleaning: The ow tube and owmeter body can best be cleaned with a little pure soap and water. Use of a bottle brush or other soft brush will aid the cleaning. Avoid benzene, acetone, carbon tetrachloride, alkaline detergents, caustic soda, liquid
soaps (which may contain chlorinated solvents), etc. and avoid prolonged immersion.
Reassembly: Reinstall the oat, remount, connect and place the unit back in service. A little stop cock grease or petroleum jelly on the O-rings will help maintain a good seal as well as facilitate assembly. No other special care is required.
For VFB models 85 and 86, rst install the lower tting, next the oat guide and oat and nally the upper tting and plug. Be certain that both ends of the oat guide are properly engaged and the oat is correctly oriented.
ADDITIONAL INFORMATION
For additional owmeter application information, conversion curves, factors and other data covering the entire line of Dwyer Instruments, Inc. full-line catalog.
Do not completely unscrew valve stem unless owmeter is unpressurized and drained of any liquid. Removal while in service
P1 x T2
P2 x T1
(14.7 + 5) x 530
14.7 x (460 + 85)
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
Printed in U.S.A. 7/20 FR# 440241-00 Rev. 8©Copyright 2020 Dwyer Instruments, Inc.