2.3.3 Attaching the transducer holder to the pipe 11
2.3.4 Fitting the transducers 11
2.3.5 Transducer attachment (diagonal mode) 13
3: Operating Procedures 15
3.1 Setting-up the Instrument 16
3.1.1 Using the instrument for the first time 16
3.1.2 Changing the user language 16
3.2 Using the Quick Start Menu 17
3.3 Instrument Calibration 21
3.3.1 Adjusting the zero cut-off 21
3.3.2 Adjusting the set zero flow offset 21
3.3.3 Adjusting the calibration factor 22
3.3.4 Adjusting the roughness factor 23
3.3.5 Adjusting the damping factor 23
3.4 Outputs 25
3.4.1 Current output 25
3.4.2 Pulse output 27
3.4.3 Alarm outputs 28
i
3.5 How to Measure Totalized Flows (manually) 31
3.6 Display of totalizers 32
3.7 Operation with an Energy Meter 32
3.7.1 Pulse output 32
3.7.2 Configuring the UFB 32
4: Maintenance & Repair 34
4.1 Introduction 34
4.2 General care 34
4.3 Warranty / Return 34
5: Troubleshooting 35
5.1 Overview 35
5.2 General Troubleshooting Procedure 36
5.3 Warning and Status Messages 37
5.4 Diagnostics Display 39
6: Options 40
6.1 Large Pipe Diameter Transducers 40
6.2 Transducer Holder Options 40
6.3 Extended Signal Cable Options 40
7: Specification 41
ii
1: General Description
1.1Introduction
This manual describes the operation of the UFB flowmeter. The flowmeter is designed to work with clamp-on
transducers to enable the flow of a liquid within a closed pipe to be measured accurately without needing to insert
any mechanical parts through the pipe wall or protrude into the flow system.
Using ultrasonic transit time techniques, the UFB is controlled by a micro-processor system which contains a
wide range of data that enables it to be used with pipes having an outside diameter ranging from 0.5 to 79 inches
(13 to 2000 mm) and constructed of almost any material. This can be extended to pipes of up to 197 inches (5000
mm) using the optional type D sensors. The instrument will also operate over a wide range of fluid temperatures.
UFB standard features:
• Large, easy to read graphic display with backlight.
• Simple to follow, dual function keypad.
• Simple ‘Quick Start’ set up procedure.
• Continuous signal monitoring.
• Isolated pulse output (volumetric or frequency).
• Isolated current output (4 to 20 mA, 0 to 20 mA, or 0 to 16 mA).
• 2x Isolated programmable alarm outputs.
• Password-protected menu operation for secure use.
• Signal diagnostics.
• Multi-function alarm outputs.
• Operates from Mains, 24 Vac, or 24 Vdc.
Volumetric flow rates are displayed in L/h, L/min, L/sec, gal/min, gal/h, USgals/min, USgals/h, Barrel/h, Barrel/
day, m³/s, m³/min, m³/h. Linear velocity is displayed in meters or feet per second. When operating in the ‘Flow
Reading’ mode the total volumes, both positive and negative, are displayed up to a maximum 12-digit number.
The flowmeter can be used to measure clean liquids or oils that have less than 3% by volume of particulate
content. Cloudy liquids such as river water and effluent can be measured along with cleaner liquids such as
demineralized water.
Typical UFB applications include:
• Sea or River water.
• Potable water.
• Demineralized water.
• Treated water.
The UFB is available in two model options. Model UFB-A is supplied with type ‘A’ transducers which are designed
to work with pipe diameters between 0.5 to 4.5 inches (13 to 115 mm). Model UFB-B is supplied with type ‘B’
transducers which are designed to work with pipe diameters between 2 to 79 inches (50 to 2000 mm). Both sets
of transducers use a common mounting system for pipe attachment, and throughout this manual any reference to
‘UFB’ applies to both ‘A’ and ‘B’ model variants unless otherwise stated.
Note: In addition to the 'A' and 'B' type sensors, type 'D' sensors (option) are available for use on pipes up to 197
inches (5000 mm). These sensors have a different mounting method. See Paragraph 1.3 for further details.
1
1: General Description
Fluid flow
Fluid flow
Fluid flow
Fluid flow
Fluid flow
U
U
U
U
D
D
D
D
Separation
Distance
Separation
Distance
Separation
Distance
Separation
Distance
Reflex mode
Reflex mode (double bounce)
Reflex mode (triple bounce)
Diagonal mode
This is the mode most commonly used.
The two transducers (U & D) are attached
to the pipe in line with each other and the
signals passing between them are reflected
by the opposite pipe wall.
The separation distance is calculated by
the instrument in response to entered data
concerning the pipe and fluid characteristics.
In this mode the separation distance is
calculated to give a double bounce. This is
most likely to occur if the pipe diameter is
so small that the calculated reflex mode
separation distance would be impractical
for the transducers in use.
This illustration goes one step further to show
a triple bounce situation. This would normally
apply when working with very small pipes
relative to the transducer range in use.
This mode might be selected by the
instrument where relatively large pipes are
concerned. In this mode the transducers are
located on opposite sides of the pipe but the
separation distance is still critical in order
for the signals to be received correctly.
This mode can be used with the standard
‘A’ & ‘B’ transducer sets but for really large
pipe installations the optional transducer set ‘D’
might be recommended.
Upstream
transducer
1.2Principles of Operation
When ultrasound is transmitted through a liquid, the speed at which the sound travels through the liquid is
accelerated slightly if it is transmitted in the same direction as the liquid flow, and decelerated slightly if
transmitted against it. The difference in time taken by the sound to travel the same distance but in opposite
directions is therefore directly proportional to the flow velocity of the liquid.
The UFB system employs two ultrasonic transducers attached to the pipe carrying the liquid and compares the
time taken to transmit an ultrasound signal in each direction. If the sound characteristics of the fluid are known,
the instrument’s microprocessor can use the results of the transit time calculations to compute the fluid flow
velocity. Once the flow velocity is known the volumetric flow can be easily calculate d for a given pipe diameter.
Figure 1.1 Operating modes
2
The system can be set up to operate in one of four modes, determined mainly by the pipe diameter and th e type
Acoustic
Steel
T ransducer
Transducers
(Sensors)
UFB
Instrument
Mounting
High
Transducer
Cables (x2)
User
Documentation
holder
Temperature
Interface
Cables (x2)
Couplant
Applicator
(Ax2, or Bx2)*
Clamps (x2)
Bands (x2)
of transducer set in use. The diagram in Figure 1.1 illustrates the importance of applying the correct separation
distance between the transducers to obtain the strongest, and therefore most reliable, signal.
1.3Supplied Hardware
The supplied UFB components are shown in Figure 1.2.
1: General Description
3
UFB Standard equipment
• Instrument with backlit graphic display.
• Transducer cables (x2) 16.5 feet (419 mm) in length.
• High temperature interface cables (x2).
• Transducers ‘A-ST’ x2 (UFB-A) for use with pipes rangi n g 0.5 to 4.5
• Transducers ‘B-ST’ x2 (UFB-B) for use with pipes ranging 2 to 79 Inches (50 to 2000 mm).
• Transducer holder for use with ‘A’ or ‘B’ transducers.
• Steel bands used to secure the transducer holder to the pipe.
• Acoustic couplant.
• User documentation.
UFB Optional equipment
• Transducer set 'D' can be used for monitoring pipes of 59 inches to 197 inches (1500 to 5000 mm) outside
diameter, over a temperature range -4°F to +176°F (-20°C to +80°C). This optional kit is supplied in a
separate case and includes the type 'D' transducers together with ratchet straps and holders for attaching
the transducers to the pipe.
Figure 1.2 Standard UFB equipment
inches (13 to 115 mm).
1: General Description
Keypad
LCD Display
Coaxial cables to transducers
Cable glands for
Small & Large Glands
for power connection
Alarms & I/O connections
Blue = Downstream
Red = Upstream
1.4UFB Instrument
The UFB is a microprocessor controlled instrument operated through a menu system using an inbuilt LCD display
and keypad. It can be used to display the instantaneous fluid flow rate or velocity, together with totalized volumes.
The instrument also provides an isolated current output, or variable pulse output, that is proportional to the
measured flow rate and can be scaled to suit a particular flow range. Two isolated alarm outputs are provided
which can be configured in a number of ways. For example, to operate when the flow rate exceeds a specified
maximum or minimum value.
1.4.1 Connections
Transducer connections
The transducers are connected to two coaxial sockets located on the bottom left-hand of the instrument. The silkscreen above these connectors show a red and blue triangle and a direction of flow symbol. For a positive flow
reading, it is important that the upstream transducer is connected to the RED socket and the downstream
transducer to the BLUE one. It is safe to connect or disconnect these cables while the instrument is switched on.
4 to 20 mA, ‘Pulse’, and Alarm I/O connections
The 4 to 20 mA, ‘pulse’, and alarm I/O cables, enter the bottom of the instrument via two cable glands and
connected internally to a terminal block. Full details of the terminal connections are provided in Chapter 2
(Installation).
Power supply
Two cable glands located on the bottom right-hand side of the instrument are available for the power supply
cable. Two sizes of glands are provided to accept cables of different diameters.
1.4.2 Keypad
The instrument is configured and controlled via a 15-key tactile membrane keypad, as shown in Figure 1.4.
Figure 1.3 Instrument details
4
1: General Description
Scroll UP
ENTER (SELECT)
Scroll DOWN
Scroll LEFT
Scroll RIGHT
Numerical keypad
with dual function keys
Figure 1.4 UFB Keypad
Menus and the menu selection keys
Note: As a security measure, once the instrument has been set-up for the first time, a password is required to
gain subsequent access to the operating menus (see page 19).
The UFB menus are arranged hierarchally with the MAIN MENU being at the top level. Menu navigation is
achieved by three keys located on the right hand side of the keypad which are used to scroll UP and DOWN a menu
list and SELECT a menu item. When scrolling through a menu, an arrow-shaped cursor moves up and down the
left hand side of the screen to indicate the active menu choice which can then be selected by pressing the ENTER
(SELECT) key.
Some menus have more options than can be shown on the screen at the same time, in which case the
overflowed choices can be brought into view by continuing to scroll DOWN past the bottom visible item. Menus
generally ‘loop around’ if you scroll beyond the first or last items.
If you select Exit on any menu it usually takes you back one level in the menu hierarchy, but in some cases it
may go directly to the ‘Flow Reading’ screen.
Some screens require you to move the cursor left and right along the display as well as up and down. This is
achieved using keys 5 (scroll LEFT) and 6 (scroll RIGHT).
Dual function numerical keypad
The block of keys shown in the center of the keypad in Figure 1.4 are dual function keys. They can be used to
enter straight-forward numerical data, select the displayed flow units, or provide quick access to frequently
required control menus.
1.4.3 Power supply
Mains supply
As standard, the UFB instrument is designed to work with a mains supply of 86 to 236 Vac and 50/60 Hz. A mains
supply fuse is located adjacent to the mains power connection (see Figure 2.2).
24V Supply
An alternative 24 V (ac/dc) power supply module is available as a factory fitted option.
Power failure
The instrument will automatically power-up and become operational when the input power is applied. In the event
of a power failure, the instrument’s configuration parameters are stored in non-volatile memory which then allows
the instrument to return to normal operation immediately power is restored.
5
2: Installation
WARNING
WARNING
WARNING
Caution
2: Installation
2.1Safety Precautions and Warnings
You may be exposed to potentially lethal (mains) voltages
when the terminal cover of this instrument is removed.
Always isolate the supply to this instrument before removing
This instrument must be installed by an electrically qualified
technician aware of the potential shock hazards presented
when working with mains powered equipment.
LETHAL VOLTAGES
the terminal cover.
LETHAL VOLTAGES
If the equipment is powered from a 24 Vac supply then the
Blanking plugs are fitted to the cable glands on leaving the manufacturer.
In order to preserve the enclosure’s IP65 rating, ensure that the blanking plugs
2.2Installing the UFB Instrument
2.2.1 Positioning the instrument
The UFB instrument should be installed as close as conveniently possible to the pipe-mounted ultrasonic
sensors. Standard transducer cables are 16.5 ft (5 m) in length with 33 ft (10 m) cables being optionally available.
Where, for operational reasons, it is not possible to mount the instrument this close to the sensors, bespoke
cables of up to 328 ft (100 m) can be provided. Consult Dwyer for further information and availability.
A suitable mains supply must be available to power the instrument (an optional 24V (ac/dc) supply module is
available). The external supply must be suitably protected and connected via an identifiable isolator. A 500 mA
fuse is fitted internally in the instrument’s input supply line.
SUPPLY EARTHING
supply must be isolated from earth.
IP65 Enclosure Protection
remain fitted in any unused cable gland.
6
2.2.2 Mounting the instrument
GND
TxD
RxD
mA+
mA-
PULSE+
PULSE-
ALARM1+
ALARM1-
ALARM2+
ALARM2-
EXPIO#1
EXPIO#2
EXPIO#3
EXPIO#4
EXPIO#5
EXPIO#6
EXPIO#7
EXPIO#8
24V+
24V-
230V -L
230V-N
230V-E
GND
TxD
RxD
mA+
mA-
PULSE+
PULSE-
ALARM1+
ALARM1-
ALARM2+
ALARM2-
EXPIO#1
EXPIO#2
EXPIO#3
EXPIO#4
EXPIO#5
EXPIO#6
EXPIO#7
EXPIO#8
24V+
24V-
230V -L
230V-N
230V-E
FUSE
4.5in
7.8in
Screw Slot
Keyhole
Mounting Details
The instrument should be
securely wall-mounted using the
three fixing points shown.
Cable connections
All power and control cables enter
through cable glands located on the
bottom of the instrument and connect
to terminal blocks as shown.
Ideally, the UFB enclosure should be fixed to a wall using three screws – see Figure 2.2.
1. Remove the UFB terminal cover.
2. Fix a screw into the wall at the required point to align with the mounting keyhole on the back of the enclosure.
3. Attach the enclosure to the wall using the keyhole screw mounting.
4. Align the enclosure then mark out the positions for the two remaining screw fixings through the slots near the
bottom corners of the enclosure. Then remove the enclosure, and drill (and plug) the two fixing points.
5. Clear the site of any dust/debris, then mount the enclosure on the wall.
2: Installation
Figure 2.1 UFB Mounting and connection details
7
2: Installation
WARNING
WARNING
2.2.3 Connecting the instrument
All cables enter the instrument through the (4) cable glands and then connected to terminal blocks which are
located behind a safety cover. The terminal blocks use a spring-loaded securing mechanism which is opened by
lifting the orange tab situated on the top of the terminal connection.
Control & monitoring cables
Depending on the fitted options, any of the following control and monitoring cables may be required:
• Current output – a 4 to 20 mA, 0 to 16 mA, or 0 to 20 mA monitoring signal is output at terminal mA+ and
mA-. (mA+ is the current output terminal and mA- is the return terminal).
• Pulse output – an opto-isolated pulse output is available at terminals PULSE+ and PULSE- (PULSE+ is the
pulse output terminal and PULSE- is the return terminal).
• Alarm Outputs – two programmable, multifunction alarm outputs are available using MOSFET, SPNO
relays. The relays, which are rated at 48 V/500 mA continuous load, are connected to terminals
ALARM1+, ALARM1-, ALARM2+ and ALARM2- respectively.
1. Remove the terminal block cover.
2. Route the control and monitoring cables through the two smaller cable glands.
3. Cut the wires to length, strip back the insulation by approximately 0.4in (10 mm) and connect them to the
required terminals identified in Figure 2.1.
4. On completion, tighten the cable glands to ensure the cables are held securely.
Power connections
LETHAL VOLTAGES
Ensure the power cable is isolated from the mains supply.
Do not apply mains voltage with the terminal cover removed.
SUPPLY EARTHING
If the equipment is powered from a 24V AC supply then the
supply must be isolated from earth.
The UFB instrument can be powered from a mains supply (86 to 264 Vac, 47 to 63 Hz) or from a 24 V(ac/dc)
supply if it is fitted with a 24 V supply module.
1. Route the power cable through one of the two cable glands located below the power connection terminals,
using the gland most suitable for the power cable diameter.
2. Cut the wires to length, strip back the insulation by approximately 0.4 inches (10mm), and connected to them
to the correct power supply terminals identified in Figure 2.1.
3. On completion, tighten the cable glands to ensure the cables are held securely.
4. Refit the terminal block cover.
8
2.3Installing the Ultrasonic Transducers
The UFB equipment expects a uniform flow profile as a distorted flow will produce unpredictable measurement errors. Flow
profile distortions can result from upstream disturbances such as bends, tees, valves, pumps and other similar obstructions.
To ensure a uniform profile, the transducers must be mounted a sufficient
distance away from any cause of distortion.
Flow
Valid transducer location
10 x Diameter20 x Diameter
45°
Uniform Flow ProfileDistorted Flow Profile
Possible
sludge
Air
Flow
Transducer
Holder
45°
In many applications an even flow velocity profile over a full 360° is unattainable due to, for example, air turbulence at the
top of the flow and possibly sludge in the bottom of the pipe. Experience has shown that the most consistently accurate
results are achieved when the transducer holders are mounted at 45° with respect to the top of the pipe.
Possible
sludge
Air
Reflex Mode
Diagonal Mode*
Transducer
Transducer
Transducer Holder 2
Transducer Holder
Transducer Holder 1
Transducer Holder 2
45°
*Note: when using the UFB in the ‘diagonal’ mode an additional transducer holder and fixing kit is required.
Holder
Holder 1
2.3.1 Transducer positioning
2: Installation
Figure 2.2 Locating the transducers
9
Figure 2.3 Transduce r ho ld er attachment (reflex vs. diagonal mod e )
2: Installation
Rectangular
Upstream
Downstream
Stainless steel bands
transducer
opening
transducer
Nylon Flag
Transducer
Cable Connector
Thumb screw
Transducer Holder
To obtain the most accurate results, the condition of both the liquid and the pipe wall must be suitable to allow the
ultrasound transmission along its predetermined path. It is important also that the liquid flows uniformly within the
length of pipe being monitored and that the flow profile is not distorted by any upstream or downstream
obstructions. This is best achieved by ensuring there is a straight length of pipe upstream of the transducers of at
least 20 times the pipe diameter and 10 times the pipe diameter on the downstream side, as shown in Figure 2.2.
Flow measurements can be made on shorter lengths of straight pipe, down to 10 diameters upstream and 5
diameters downstream, but when the transducers are positioned this close to any obstruction the resulting errors
can be unpredictable.
Preparation
Before you attach the transducers you should first ensure that the proposed location satisfies the distance
requirements shown in Figure 2.2, otherwise the resulting accuracy of the flow readings may be affected.
Prepare the pipe by degreasing it and removing any loose material or flaking paint in order to obtain the best
possible surface. A smooth contact between the pipe surface and the face of the transducers is an important
factor in achieving a good ultrasound signal strength, and therefore maximum accuracy.
Key Point: Do not expect to obtain accurate results if the transducers are positioned
close to any obstructions that distort the uniformity of the flow profile.
Type ‘A’ or ‘B’ transducers are attached to the pipe using the adjustable transducer holder assembly shown in
Figure 2.5
Figure 2.6
Figure 2.4. The transducer holder itself is secured to the pipe using two wrap-around steel bands. For user
convenience, an imperial ruler (inches) is etched on the top side plate and a metric (millimeters) ruler is etched on
the bottom side plate – as shown in Figure 2.4. Once the transducer holder assembly is fully assembled, the
transducers are locked into position by tightening two thumb screws which are themselves then secured by
locking nuts.
Note: When using the UFB in the diagonal or reflex mode on pipes over 16 inches diameter, two transducer
holders are required with one transducer mounted in each – see Paragraph 2.3.5 for diagonal mode details.
2.3.3 Attaching the transducer holder to the pipe
1. Remove the two nylon flags from the transducer
holder.
2. Position the transducer holder horizontally on
the pipe at 45° to the horizontal and secure it in
position using the stainless steel banding, as
shown in Figure 2.5.
2: Installation
Note: In the following procedure the transducer
holder is installed with the rectangular opening
facing towards the upstream end of the pipe.
2.3.4 Fitting the transducers
1. Attach the nylon flags to the two
transducers, as shown in Figure
2.6.
2. Using a syringe applicator, apply a
narrow bead (0.1in) of acoustic
couplant to the base of both
transducers.
3. Insert the downstream transducer
cable (blue) through the holder’s
end plate and up through the
rectangular opening at the lefthand side of the transducer
holder, as shown in Figure 2.7.
Note: When carrying out the following steps, handle the transducer assembly with care to avoid smearing the
acoustic couplant on the pipe while attaching the transducer to the transducer holder.
4. Connect the downstream cable (blue) to one of the transducers, then lower the transducer and cable through
the rectangular opening as shown in Figure 2.8.
5. Align the slots in the sides of the nylon flag with the edges on the top of the transducer holder, then carefully
slide the transducer assembly to the middle of the holder.
6. Tighten the thum b scre w (clockwi se) unt il the tra nsdu cer is locked to the top of the transducer holder. This is
done to prevent the acoustic couplant from touching the pipe.
11
Loading...
+ 33 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.