Dwyer TM2 User Manual

Series TM2 Digital Rate Meter & Totalizer
1-17/32 [38.89]
2-31/32 [75.41]
1-37/64 [40.08]
1-1/8
[28.58]
Specifications - Installation and Operating Instructions
Bulletin F-41-TM2
Monitor flow rate or total flow with Model TM2 Flow Totalizer
Section I. UNPACKING YOUR DISPLAY
Checking for external damage to shipment
Your Model TM2 Display was packed by the manufacturer in such a way that you should receive it with no damage. If external damage is noted upon receipt of the package, please contact the shipping company (not Dwyer Instrument, Inc.) immediately. Dwyer Instruments, Inc. will not be liable for damage to the meter once it has left the manufacturing premises.
Unpacking the display
.
SPECIFICATIONS Input: 7.5 VDC max. pulse. Accuracy: ±0.2%. Input Impedance: 27 k Ω at 3 VDC. Totalizer Type: UP count. Rate Indication Type: Frequency. Count Input Speed: 10 kHz @ 50% duty cycle. Totalizing Range: 0.0001 to 100.0000 with five position
decimal point. Rate Indication Range: 0.001 to 9999 (five position decimal point).
Update Time: 0.7 seconds. Display: 8-digit LCD, 7/16" H. Temperature Limits: 32 to 131°F (0 to 55°C). Housing Material: Cycolac X-17. Power: One 3V lithium battery (included). Average Battery Life: Approximately 5 years. Weight: 1 Ib (0.5 kg). Agency Approvals: CE.
Cautions against damaging the unit
Any damage inflicted on the display by the customer will not be repaired under warranty by Dwyer Instruments, Inc. See Section 6 for more information.
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
Phone: 219/879-8000 www.dwyer-inst.com
Section II. INSTALLING YOUR MODEL TM2
Mounting the display
When choosing a place to mount the display, keep these three things in mind:
a. The display should not be mounted near a solenoid or
other inductive device.
b. Enough ventilation should be supplied to keep the
totalizer operating within the temperature specifications (see Selection 5).
c. This display should not be mounted in a heavy
vibration area.
The recommended mount for the Model TM2 display is panel­mount. The recommended panel cutout for the display is 2.677˝ wide (68 mm) by 1.299˝ tall (33 mm).
To install the unit in a panel, first remove the two fasteners located on the side of the unit. These slide tightly in grooves and should slide out without too much trouble. DO NOT MISPLACE THESE FASTENERS!!!
With the fasteners removed, the display should slide into the panel cutout. The included gasket should be between the panel and the face of the display. Making sure the gasket is properly aligned and not twisted, slide the display in place. Reinstall the fasteners by sliding them in their grooves until they stop. The screws on the fasteners should be facing the panel and the head of the screw facing away from the panel.
Tighten the set screws of the fasteners using a flathead screwdriver in a clockwise motion. They should be equally tightened until the display is snug and the gasket is reasonably compressed.
Pin #4: This pin is used for remote reset. When
momentarily connected to ground, this pin will reset the totalizer just like the front panel reset button.
Pin #5: This pin is used to enter program mode. When
connected to the ground, the Model TM2 enters program mode, described in Section 3.
Pin #6-Pin #8: These pins are not functional.
See the last page for wiring diagrams of connections to various Flo-Sensors and other products.
Wiring Recommendations
Following these suggestions will increase noise immunity and lengthen unit life.
Cable: The connection between the count source and the totalizer/ratemeter should be made with a two-conductor shielded cable. The shield should be connected to earth ground at one end only. The connecting cable should not be run in conduits with cables switching high inductive loads.
Relay Coil Suppression: If a relay contact is used as a count source, the relay coil should be suppressed. This can be accomplished with an RC network for AC coils or a diode for DC coils.
Mounting: This totalizer/ratemeter should not be mounted near a solenoid or other inductive devices. Enough ventilation should be supplied to keep the totalizer operating within the temperature specifications. Do not mount this unit in a heavy vibration area.
When mounted with the provided fasteners, gasket, and specific rectangular panel cutout, the Model TM2 display achieves a NEMA-4X rating.
Electrical connections to the display
The Model TM2 is self-powered using an internal lithium battery, thus there is no need for external power. The typical life of this battery is 5 years.
After connections are made, power the display by removing the temporary insulator on the back of the unit. Simply pull the white tab out -- there is no need to open the battery cover.
Pin #1
Signal & Reset Ground
Pin #1: This pin should be connected to the ground of the
Pin #2: This pin is not used with any Dwyer Instruments,
Pin #3: This is the pulse signal input. The Model TM2
Pin #2
Not Used
pulse signal. In most cases, this will be the same as the ground of the Flo-Sensor that the Model TM2 Display is connected to.
Inc. products.
accepts all TF10xx Series Flo-Sensor pulse outputs. These are 7.5VDC peak squarewave pulses directly proportional to the flow rate.
Pin #3
Signal In (Pulse Input)
Pin #4
Remote Reset (optional)
Pin #5
Enable Program Mode
Section III. PROGRAMMING THE TM2
The Dwyer Instruments, Inc. Model TM2 Display is a versatile meter that will indicate flow rates and totalize flow in virtually any unit of measurement. To do this, it uses 2 numerical factors (programmed by the user) to customize and match each display with each particular Flo-Sensor.
To successfully program this display, you must first understand how it works and what you are programming into it. Program mode uses six different screens to program rate and total flow. They are used as follows:
TOTALIZER FUNCTION:
Screen #1: The Significant digits of the Totalizer Multiplier
are entered here. The Model TM2 uses these to calculate total flow.
Screen #2: This screen is simply a decimal placer. Since
only the significant digits of the Totalizer Multiplier are entered in screen #1, this allows you to make up for the insignificant zeros not entered there.
RATE FUNCTION:
Screen #3: The significant digits of the Rate Multiplier are
entered here. The Model TM2 uses these to calculate current rate of flow.
Screen #4: This screen is another decimal placer. Since
only the significant digits of the Rate Multiplier are entered in screen #3, this allows you to make up for the insignificant zeros not entered there.
MISCELLANEOUS:
Screen #5: This screen should always read “1”. It is not
used with any Model TF10xx Flo-Sensors.
Screen #6: This screen enables/disables the front panel
reset button.
To understand the insignificant/significant digit importance, you must understand that many of your Rate and Totalizer Multipliers will be very small, with several zeros in front of them, such as
0.0001234. However, screens 1 and 3, where those numbers are entered, will only allow 6 to 4 digits, respectively. There is, therefore, no room for those preceding zeros. For each zero removed, the decimal place in the following screen shifts one left. This is further described in Section 3. d.
Since the Model TM2 operates on pulse inputs, the pulse output (frequency) of the Flo-Sensor at 100% flow must be determined. This number is located in one of three places:
1. On a tag attached to a fitting of the Flo-Sensor
2. On a calibration data sheet, included with the Flo-Sensor
3. On a label near the connector of the Flo-Sensor
The label should provide pulse output (PPS) information of the particular Flo-Sensor at the rated 100% flow of the Flo-Sensor. For example, for a Model TF1032 Flo-Sensor, rated from 0.1 to
2.0 L/min, you would need the pulse output data of that Flo­Sensor at 2.0 L/min. (The pulse output is usually given when the voltage output is 5.00 VDC, or 100% flow.)
Note: If the tag listing the pulse rate is missing, please contact Dwyer Instruments, Inc. for pulse output information on that particular Flo-Sensor. Please provide the model number, serial number, and flow range of the Flo-Sensor you are requesting information on. It is also possible to determine the pulse output frequency by using a volt meter and the Model TM2 together. Using methods of programming described in Section 3.d, set screen #3 to read “1.000” and screen #4 to read “0000”. Hook the volt meter up to the analog voltage output of the Flo-Sensor. Using the voltage output as a guide, provide enough flow through the sensor to wear the sensor outputs 5.00 VDC. Place the TM2 in rate mode and the number displayed will be the pulse output of the sensor at 100% rated flow.
Figuring the Rate Multipler
The Rate Multiplier is the number that the Model TM2 uses to interpret pulses into rate. To determine this factor, first convert the 100% flow rate of the Flo-Sensor (listed in the manual of the Flo-Sensor) into the units in which you wish the Model TM2 to display rate. Use the following formula to figure the Rate Multiplier:
100% Rated Flow
Time Unit New Time Unit Factor Pulse Output (PPS)
where 100% Rated Flow = 100% flow rate of Flo-Sensor, where pulse output is given (Ex: 2.0L)
Time Units = time units of 100% Rated Flow (Ex: 1 min.) Pulse Output = marked pulse output of sensor at 100% Flow (PPS) New Flow Unit Factor = see table on the next page New Time Unit Factor = see table on the next page
New Flow Unit Factor 1
X
X
Entering Program Mode
In order to program the factors into the display, it must first be put into program mode (of course, the unit must be powered, see Section 2.b). To do this, simply connect the ground input (pin #1) to the enable program input (pin #5) with a short wire, jumper, or alligator chip. Make sure the connection will hold well enough for you to program a series of numbers into the unit. When you are finished and wish to exit program mode, simply remove the connection.
TIME UNIT FACTOR
Current Time Unit
minutes minutes minutes minutes hours hours hours hours
FLOW UNIT FACTOR
Current Time Unit
mL (milliliters) mL mL mL L(liters) L L L US Gallons US Gallons US Gallons US Gallons
Desired Time Unit
seconds minutes hours days seconds minutes hours days
Desired Time Unit
mL L US Gallons Fluid Ounces mL L US Gallons Fluid Ounces mL L US Gallons Fluid Ounces
Factor
60.00
1.00
0.016667
0.0006944
3600.00
60.00
1.00
0.0416667
Factor
1.000
0.0010
0.00026418
0.033814
1000.0
1.0000
0.26418
33.814
3785.4
3.7854
1.0000
128.0000
Figuring the Totalizer Multiplier
The Totalizer Multiplier is the number that the Model TM2 uses to interpret pulses into total flow accumulated. To determine this factor, first convert the 100% flow rate of the Flo-Sensor (listed in the manual of the Flo-Sensor) into the units in which you wish the Model TM2 to display total flow. Use the following formula to figure the Totalizer Multiplier:
100% Rated Flow Pulse Output (PPS) Totalizer Time Factor
where 100% Rated Flow = 100% flow rate of Flo-Sensor,
Pulse Output = Marked pulse output of sensor at
100% flow (PPS) New Flow Unit Factor = see table on the left. Totalizer Time Factor = see table below
TOTALIZER TIME FACTOR
Time Unit Sensor was Calibrated For
minutes hours
If this seems complicated, take a look at these examples:
Totalizer Time Factor
60 3600
New Flow Unit Factor
X
where pulse output is given (Ex: 2.0 L)
If this all seems confusing, please note the following examples:
EXAMPLE ONE
A certain Model TF1032 Flo-Sensor is rated from 0.1-2.0 L/min. The pulse output of the particular sensor is 274 PPS @ 2.0 L/min (each sensor will vary in pulse output). You want the Model TM2 to display rate in mL/min. The formula would look something like this:
2
1 1 274
A certain Model TF1053 Flo-Sensor is rated from 1.0-10.0 L/min. The pulse output of the particular sensor is 473 PPS @ 10.0 L/min (each sensor will vary in pulse output). However, you want the Model TM2 to display rate in Gallons/min. The formula should be used this way:
10 0.26418 1
1 1 473
A certain Model TF1062 Flo-Sensor is rated from 1-10 GPH. The pulse output of that particular sensor is 347 PPS @ 10 GPH (each sensor will vary in pulse output). However, you wish the Model TM2 to display rate in mL/second. Use the formula like this:
10 3785.4 1
1 3600 473
1000 1
X
EXAMPLE TWO
X
EXAMPLE THREE
X
X
X
X
= 7.299270
=
0.005585
0.030303
=
EXAMPLE ONE
A certain Model TF1032 Flo-Sensor is rated from 0.1-2.0 L/min. The pulse output of that particular sensor is 274 PPS @ 2.0 L/min (each sensor will vary in pulse output). You want the Model TM2 to display total flow in mL. The formula would look something like this:
2 1000 274 60
A certain Model TF1053 Flo-Sensor is rated from 1.0-10.0 L/min. The pulse output of the particular sensor is 347 PPS @ 10.0 L/min (each sensor will vary in pulse output). However, you want the Model TM2 to display total flow in Gallons. The formula should be used this way:
10 0.26418 473 60
A certain Model TF1062 Flo-Sensor is rated from 1-10 GPH. The pulse output of that particular sensor is 347 PPS @ 10 GPH (each sensor will vary in pulse output). However, you wish the Model TM2 to display total flow in mL. Use the formula like this:
10 3785.4 347 3600
X
EXAMPLE TWO
X
EXAMPLE THREE
X
= 0.1216545
= 0.1216545
= 0.030303
When you derive your particular Rate Multiplier using the formula and/or examples above, write it down -- it will be entered into screen #3 of the Model TM2 later.
Programming the Rate and Totalizer Multipliers to the Model TM2
Once in program mode as described in Selection 3. a, the Model TM2’s display should read a “1” on the far left. On the right, you should see “01.0000” with the far right digit flashing (it may read a different value if it has been previously programmed). You are now on screen #1, where the significant digits of the Totalizer Multiplier are entered.
First, you must determine which digits are significant. As you can see, only 6 digits can be programmed into screen #1. You must choose the 6 most important, or the 6 farthest left that are not zeros. The following chart shows examples of totalizer multipliers and which numbers you would actually enter:
Totalizer Multiplier
.0002499 .003456789 .000098765
1.66666667
To program your significant digits in, use the RST button to set the flashing digit and the T/R button to select the flashing digit.
Value to Enter in Scrn #1
24.9900
34.5679
98.7650
16.6667
Enter the digits much in the same way as you did for the Totalizer Multiplier. The RST button sets the flashing digit and the T/R button selects the flashing digit. The difference is that screen #3 will also allow you to set the decimal place of the Rate Multiplier. Usually, it is best to move the decimal point on this screen as far to the left as possible (it will only go as far left as X.XXX). For instance, if your Rate Multiplier is 1.234, then that exact value could be entered into this screen. If your Rate Multiplier is something like .0004567, though, then you would want to enter
4.567 in this screen. However, if your Rate multiplier is more than 9, such as 57.6789, then you would want to enter 57.68 in as your value, not 5.768.
To program this value into the Model TM2, use the RST button to set the flashing digit and the T/R button to select the flashing digit. Once you have cycled through the digits once, a “d” appears on the far right of the display. This signifies that the Model TM2 is ready to accept the decimal position of this screen. Move the decimal point as far left as possible for your value using the RST button. When this is completed, move on to screen #4 by pressing both the T/R button and the RST button simultaneously. The far left digit on the screen should now read “4”.
When you have ompleted screen #1, move on to screen #2 by pressing both the T/R button and the RST button simultaneously. The far left digit on the screen should now read “2.”
Screen #2 allows you to make up for the decimal places you lost by only entering the significant digits in screen #1. To determine the decimal place for screen #2, divide the exact number you entered in screen #1 by the actual Totalizer Multiplier. For example, if your Totalizer Multiplier is .009567, you should have entered “95.6700” in screen #1. To determine screen #2, divide
95.67 by .009567, which gives you 10000. The result, in the case 10000, determines how many zeros after the decimal point you should have in screen #2. Since 10000 contains 4 zeros, screen #2 should have 4 zeros after the decimal point; screen #2 would read “00.0000”. If the result of dividing screen #1 by the actual Totalizer Multiplier equals one, then screen #2 should not have a decimal point.
To select the decimal point, use the RST button. Each press of the button will move the decimal point one to the right. When you have selected the correct decimal position, press both the RST and the T/R button to go on to screen #3. Once you have done this, the far left digit of the screen should read “3”.
Screen #3 allows you to enter the significant digits of the Rate Multiplier. It, however, is limited to only 4 digits. Select the significant digits of the Rate Multiplier using the same method as you used for the Totalizer Multiplier, except this time selecting 4 digits instead of 6.
Screen #4 makes up for any lost decimal places in screen #3. For this screen, divide the exact number you entered in screen #3 by the actual Rate Multiplier. For example, if your Rate Multipliers .004321, you should have entered “4.321” in screen #3. To determine screen #4, divide 4.321 by .004321, which give you 1000. The result, in this case 1000, determines how many zeros after the decimal point you should have in screen #4. Since 1000 contains 3 zeros, screen #4 should have 3 zeros after the decimal point; screen #4 would read “0.000”. If the result of dividing screen #3 by the actual Rate Multiplier equals one, then screen #4 should not have a decimal point.
When this is completed, move on to screen #5 by pressing both the T/R button and the RST button simultaneously. The far left digit on the screen should now read “5”.
The far right value of screen #5 should be “1”. If it reads “10”, press the RST button to change it to “1”. When this is completed, move on to screen #6 by pressing both the T/R button and the RST button simultaneously. The far left digit on the screen should now read “6”.
Screen #6 allows you to disable/enable the front panel reset button. This button resets the totalizer value to zero. If you wish to disable this button, use the RST button to make screen #6 read “no R”. If you wish the button to operate, screen #6 should read “R”.
Note: If the front panel reset button is disabled, the totalizer may not still be reset by connecting the rear terminal pin #4 (RST) to pin #1 (GND).
When you are finished programming the Model TM2, disconnect the programming jumper (running from pin #1 to pin #5).
Making minor changes in calibration (of the display)
If you notice a small error in the display of either rate or total flow, a small change in the Rate or Totalizer Multiplier may correct the problem. To establish the error, you must provide a constant, steady signal to the Model TM2 Display. This usually means putting a steady, constant flow through the Flo-Sensor that the display is connected to. Higher flows (within the flow range of the Flo-Sensor) will allow you to more accurately correct the error.
To recognize the error, the display should be in normal operating mode, not programming mode. With a constant signal, the Model TM2 will display a constant rate and will totalize flow steadily. If you are experiencing a minor error in the Rate display, set the display to show rate (see Operating Instructions). If you are experiencing a minor error in the Total Flow display, then set the display to show total flow.
With the display in the correct mode, measure the amount of flow passing though the Flo-Sensor in exactly one minute. Compare this to what the Model TM2 shows. If there is a small error, correct it by dividing the actual flow by the indicated flow. Take that result and multiply it by the factor located in screen #1 or screen #3, the Totalizer Multiplier and Rate Multiplier screens, respectively. If you are correcting a minor error in the Totalizer function, then multiply it by the factor in screen #1. If you are correcting a minor error in the Rate function, then multiply the number by the factor in screen #3. The final result should be programmed into the appropriate screen as outlined in Section 3. d. No decimal points should need to be moved for corrections of minor errors.
As an example, a Model TM2 display indicates a flow rate of 5.0 L/min when the flow rate is actually 5.2 L/min. If you divide 5.2 by 5.0, you get approximately 0.962. The Rate Multiplier in screen #3 is currently 1.345. If you multiply 0.962 by 1.345, you arrived at 1.294. The number should then be programmed into screen #3, and the minor error will be corrected.
Section IV. OPERATING INSTRUCTIONS
The entire operation of the Model TM2 is dependent on the two buttons located on the front panel. The unit will not operate in programming mode. If it is currently in program mode (with a jumper from pin #1 to pin #5), then remove the jumper and the Model TM2 will automatically exit program mode and return to normal operation.
Two screens are available on the Model TM2. The Totalizer screen displays the accumulated pulse inputs, or total flow. This unit will totalize up to 8 digits, automatically blanking any leading zeros. The Rate screen shows the current process, or flow rate, passing through the Flo-Sensor the display is connecting to. This screen updates ever 0.7 seconds. To determine which mode you are currently in, look at the far left of the display. If the letter “R” is displayed there, then the Model TM2 is displaying rate. If there is no “R”, then the display is showing total flow. To toggle between the two modes, press the T/R button.
If you are currently in Rate mode, the RST button and/or reset terminal have no effect. If you are operating in Totalizer mode, then the RST button will reset the totalizer back to zero. If this button has no effect, it may be disabled (see Programming Instructions). The totalizer may also be reset by momentarily connecting pin #4 (on the rear of the unit, RST) to pin #1 (also on the rear, GND). This function cannot be disabled.
MAINTENANCE
Upon final installation of the Series TM2 Digital Rate Meter and Totalizer, no routine maintenance is required. A periodic check of the system calibration is recommended. The Series TM2 is not field servicable and should be returned if repair is needed (field repair should not be attempted and may avoid warranty). Be sure to include a brief description of the problem puls any relevant application notes. Contact customer service to receive a return goods authorization number before shipping.
If the totalizer is not reset before all 8 digits read “9”, then it will roll over back to zero and continue monitoring total flow.
TO WIRE A TF10xx SERIES FLO-SENSOR AND A TM2 DISPLAY USING THE OPTIONAL TF01 POWER ADAPTER
MODEL TM2 DISPLAY
PUSH-BUTTON
REMOTE RESET
(NOT INCLUDED)
4
5
2
3
1
TF10xx FLO-SENSOR
JUMPER/ CONN.
WHITE
GREEN
YELLOW
TF01 POWER ADAPTER
TERMINAL STRIP (NOT INCLUDED)
+ -
0-5VDC OUT
WIRING DIAGRAMS
TF10xx FLO-SENSOR
MODEL TM2 DISPLAY
5
JUMPER/ CONN.
BLACK
GREEN
WHITE
RED
4
3
2
1
PUSH-BUTTON
REMOTE RESET
(NOT INCLUDED)
-
0-5VDC
+
OUT
TF05 CABLE
TO WIRE A TF10xx SERIES FLO-SENSOR AND A TM2 DISPLAY USING THE OPTIONAL TF05 CABLE
TERMINAL STRIP (NOT INCLUDED)
+ -
TO 12 VDC
POWER SOURCE
MODEL TM2 DISPLAY
TF10xx FLO-SENSOR
JUMPER/
TO WIRE A TF10xx SERIES FLO-SENSOR AND A TM2 DISPLAY USING THE OPTIONAL TF02 POWER ADAPTER
©Copyright 2010 Dwyer Instruments, Inc. Printed in U.S.A. 8/10 FR# R2-443561-00 Rev.1
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
TF02 POWER ADAPTER
CONN.
BLACK
GREEN
BLACK/ WHITE
TERMINAL STRIP (NOT INCLUDED)
Phone: 219/879-8000 www.dwyer-inst.com
PUSH-BUTTON
REMOTE RESET
(NOT INCLUDED)
Loading...