Series TH2-10 Thermo-Hygrometer Pen
1-3/8
[
35.05]
5
-9/16
[141.38]
4
5/64
[18.01]
2
-5/8
[
65.34]
2-5/8
[
65.34]
2
[
52.32]
2
[
52.32]
BATTERY COMPARTMENT
TH2-10
33% SALT SOLUTION
SALT SOLUTON KIT
T
H2-CAL
7
5% SALT SOLUTION
15ºC
20ºC
25ºC
30ºC
35ºC
15ºC
20ºC
25ºC
3
0ºC
35ºC
xxxx±xxx
xxxx±xxx
xxxx±xxx
x
xxx±xxx
xxxx±xxx
Made in xxxx
3
3%RH Mgcl
Sat. salt solution
Temp. Mgcl
75%RH Mgcl
Sat. salt solution
Temp. Mgcl
xxxx±xxx
xxxx±xxx
xxxx±xxx
xxxx±xxx
xxxx±xxx
Made in xxxx
S/N: xxxxxxxx
Battery:
Made in xxxxxx
AAA (UM-4)
CR2032
gVoltage
Thermo-Hygrometer
P
ower Hold Mode
(
HI. WBGT TH2-10)
Specifications - Installation and Operating Instructions
TE-TH2-10
The Model TH2-10 Thermo-Hygrometer Pen is a compact handheld device that
measures temperature, humidity, dew point, wet bulb, and heat index. For
ncreased accuracy, calibration salt baths can be used to field calibrate the
i
umidity sensor before taking measurements. Minimum, maximum, and average
h
eadings can be displayed on the large 4 digit LCD display. High and Low alarms
r
provide audible and visual indication of changes to the ambient humidity.
FEATURES
1) Temperature and relative humidity measurement.
) Calculate dew point temperature and wet bulb temperature.
2
) Calculate heat index and wet bulb global temperature.
3
) Hi Temperature alarm setting (Heat Index/WBGT).
4
5) Data hold & Maximum/Minimum/Average function.
6) Auto power off function.
7) Self-calibration by salt bottle (Optional) at 33% & 75% humidity points.
8) Quick response and high accuracy.
OPERATING INSTRUCTIONS
Please read the manual completely before using this device.
KEY FUNCTION
(1) PWR: Turn on / off the meter.
(2) HOLD: Data hold mode.
(3) MODE: Select measurement modes: Temp. (temperature), RH (Relative
Humidity), HI (heat index), DP (dew point), WB (wet bulb), HI WB (wet bulb global
temperature / WGBT).
OPERATION
Power On/Off
Press the POWER button to turn on the meter. The meter will display TEMP
reading and icon with a beep sound. To turn off the meter, press the POWER
button for more than 2 seconds.
Maximum and Minimum Mode
With the meter already on, press the POWER button less than 2 seconds. The
MAX icon with the highest reading appears on the LCD. Press the POWER button
SPECIFICATIONS
Range: RH: 0 to 99.99%; Temperature: -4 to 122°F (-20 to 50°C).
ccuracy: ±3% @ 25°C/10 to 90% RH (±5% @ 0 to 10%, 90 to 100%) ±1°F
A
0.6°C).
(
isplay: 4-digit LCD.
D
Resolution: 0.1°F/°C.
ower Requirements: One CR-2032 lithium battery.
P
eight: 3.2 oz (91 g).
W
gency Approvals: CE.
A
DATA HOLD
Press the HOLD button to freeze readings. Press the HOLD button again to return
to normal measurement.
Auto Power Off
The meter will shut off automatically after approximately 15 minutes of no action.
To disable auto power off, press the HOLD and POWER button at the same time
until “S-no” appears. Then release the button and you will return to normal
measurement.
Once the meter is powered off, it will return to the default auto power off settings.
SELECT TEMP UNIT °C/°F
When Power off,
press MODE+POWER to enter setting function
Press MODE button to select °C/°F
Press POWER button again to save and move to
next setting.
NOTE: (a) During set operation, if you want to
escape any mode setting, please go through all
setting modes by keeping press POWER button to
normal measurement.
(b) If you want to set alarm threshold, please follow
steps below.
SETTING MODE- ALARM SETTING (AFTER SELECTING TEMP UNIT °C/°F)
again and the MIN icon will appear with the lowest reading on the LCD. Press the
POWER button again and the AVG icon appears with the average reading on the
LCD. Press POWER again and the meter will return to normal measurement.
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
Note:
1. The alarm sound can’t be disabled unless you enter setting mode to change the
threshold setting.
2. The default settings of Heat Index and WBGT alarm settings are off.
3. Please go through and select Temp. unit (C/F) before alarm setting.
Phone: 219/879-8000 www.dwyer-inst.com
tep 1. Heat Index Hi alarm select
S
Press MODE to select
HI (Heat Index) alarm ON or OFF.
(a) When alarm function is on, press POWER button to
nter Heat Index setting (move to Step 2).
e
b) When alarm function is off, press POWER button to
(
nter WBGT (HI WB) alarm setting (step 3).
e
tep 2. Heat Index Hi Temp. setting.
S
When Hi alarm is on, HI (Heat Index) icon with the
efault value 27.8°C (82°F) appears on the LCD. Keep
d
ressing MODE button to increase digit sequencely
p
rom hundred digit and release the button when the
f
igit up desired appears.
d
(Please refer to Example 1).
Press POWER button to save HI setting and go to
WBGT alarm selecting.
Step 3. WBGT alarm select
ress MODE to select HI WB (WBGT) alarm ON/OFF.
P
(a) When the alarm function is on, press POWER
button to enter WBGT setting (move to Step 4).
(b) When the alarm function is off, press POWER
button to go back to normal measurement.
tep 4. WBGT Hi Temp. setting
S
ATTERY REPLACEMENT
B
hen low battery indication “BAT” icon appears on the LCD, the battery needs to
W
e replaced.
b
se a coin to remove the battery compartment cover on the rear of the meter.
U
eplace the CR-2032 lithium battery with a new battery face up (+) in the
R
compartment.
# Example 1 - Set Hi Temp. Alarm at 128.3°C
tep 1:
S
eep pressing the MODE button, the
emp.
T
H
emp.
T
Hi
i
•
•
K
umbers will cycle from 0 to 9. It will jump
n
to the next digit to the right when the
desired number (0 to 9) is reached. So for
this example, we need a “1” in the
thousand unit column. When the number
esired is reached, it will jump to the next
d
igit to the right.
d
Step 2:
Keep pressing the MODE button, the
ame action as before until the hundreds
s
igit shows ”2". Then release the MODE
d
button.
When WGBT alarm is on, HI WB (WBGT) icon with the
default value 28.0°C (82.4°F) appears on the LCD.
Keep pressing MODE button to increase digit
sequencely from hundred digit and release the button
when the digit you desired appears. (Please refer to
Example 1). Press POWER button to save WBGT
setting and the meter will go to normal measurement.
Note:
HEAT INDEX: The Heat Index is determined using the dry bulb temperature and
relative humidity. It is based upon charts available from the U.S. National Weather
Service. The Heat Index represents how an average person feels relative to climate
conditions. For a given temperature, the higher the humidity, the higher the heat
index will be.
TROUBLE SHOOTING
Error messages:
Humidity sensor is failed.
E-2
Temperature sensor is failed.
E-3
Operation temperature is too high.
E-4
Operation temperature is too low.
E-5
Some hardware has failed.
E-6
Note: If above error messages appear on the display, please contact with local
distributor for technical service.
Temp.
Hi
•
Temp.
Hi
•
MAINTENANCE/REPAIR
Upon final installation of the Series TH2-10, no routine maintenance is required.
The Series TH2-10 is not field serviceable and should be returned if repair is
needed. Field repair should not be attempted and may void warranty.
WARRANTY/RETURN
Refer to “Terms and Conditions of Sales” in our catalog and on our website. Contact
customer service to receive a Return Goods Authorization number before shipping
the product back for repair. Be sure to include a brief description of the problem
plus any additional application notes.
Step 3:
eep pressing the MODE button, the
K
ame action as before until the tens digit
s
shows “8”. Then release the MODE button.
Step 4:
Keep pressing the MODE button, the
same action as before until the last unit
digit shows “3”. Then release the MODE
button.
©Copyright 2012 Dwyer Instruments, Inc. Printed in U.S.A. 8/12 FR# R6-443998-00
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com