7. Calibrating Units for different Liquids ......................................................... 19
1. General .................................................................................................... 19
2. Cleaning and Flushing .............................................................................. 20
3. Returning Units For Repair Or Recalibration ............................................... 20
F-20-TF-L, pg. 2 of 30
A.Introduction
1.Unpacking
All sensors are suitably packaged to prevent damage during shipping. If
external damage is noted upon receipt of the package, please contact
shipping company
to the device once it has left the manufacturing premises.
Open the package from the top, taking care not to cut too deeply into the
package. Remove all the documentation and contents. Take care to remove
all the items and check them against the packing slip. The products should
also be checked for any concealed shipping damage. If any shortages or
damage is noted, please contact the shipping company and/or Dwyer
Instruments to resolve the problem.
immediately. Dwyer Instruments is not liable for damage
Typical Contents of Box
(Model TFP-LP shown, other models similar).
2.Product Overview and Principle of Operation
CAUTION: Take care not to
drop
your sensor or subject it to
sudden impact. Read the installation section of this manual before
providing power or tubing connections to the unit. Any damage
caused by improper installation or careless handling will not be
repaired under warranty (see limited warranty on page 30 for
more details).
TF Series Flow Meters and Flow Sensors for liquids are capable of
measuring flows as low as 13-100 ml/min or as high as 1.0-10.0 l/min.
Highly repeatable results are achieved using a patented Pelton Turbine
F-20-TF-L, pg. 3 of 30
Wheel flow sensor design. This proven design minimizes zero drift while
maintaining fast response and linear outputs with virtually no maintenance.
Patented technology measures flow using a miniature turbine wheel similar
in size to a U.S. dime (16 mm diameter, 0.75 mm thick). The micro-turbine
wheel is supported on a very small sapphire shaft that is held in position by
two sapphire bearings. The micro-turbine assembly is so light that it virtually
floats in the liquid. This relieves force on the bearings and almost eliminates
wear.
As flow passes through the flow sensor, a precision machined nozzle directs
the fluid onto the very small teeth of the micro-turbine wheel. This causes the
wheel to spin at a speed proportional to the flow rate.
The micro-turbine wheel has alternating white and black sections evenly
spaced on one side of the wheel. An infrared light beam is directed onto the
wheel. As the wheel rotates the infrared beam is reflected off each white
section. The reflected beam is detected by a phototransistor that converts the
reflections into electrical pulses. As the wheel spins faster the pulse rate
increases. Processing circuitry provides analog and/or pulse output that are
linearly proportional to the flow rate
When the wheel stops (under zero flow conditions), no pulses are
generated. Consequently, zero drift is not possible and zero adjustments are
never required.
Every unit is supplied with a calibration certificate detailing the results
obtained during calibration. Units are calibrated using deionized water as
the reference media. Flowing liquids with different specific gravities or
viscosities may affect the calibration.
F-20-TF-L, pg. 4 of 30
B.Installation
CAUTION: Do not flow any gas through a liquid flow sensor.
This may damage the micro-turbine assembly and void the
warranty. Do not exceed the pressure, temperature or power
operating ranges detailed in the Specifications section of this
manual. Dwyer Instruments shall not be liable for any damage or
injury caused by incorrect operation of their products.
1.General Considerations
It is recommended that a safety shut-off valve be installed upstream of
(before) the sensor.
All wetted parts should be checked for compatibility with the liquid to be
used. If there are any incompatibilities e.g. highly corrosive liquid, then the
unit may be damaged or fail prematurely. Such damage will not be
repaired under warranty.
Units should be installed in a clean, dry environment with an ambient
temperature that is as stable as possible. Avoid areas with strong magnetic
fields, strong air flows or excessive vibration.
If the liquid to be used may contain particles larger than 25 microns then a
filter (25 microns or less) should be installed upstream of (before) the unit.
The required differential press (or pressure drop across the unit) decreases
exponentially with decreases in flow rate.
Pressure Drop vs. Percentage of the Full Scale
Rated Flow (may vary ±10% of indicated psid)
12
0.8
10
8
6
Delta P psid
4
2
0
102030405060708090100
% o f rate d flow
100-1000 ml/min Units
All Other Ranges
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
Delta P Bar
F-20-TF-L, pg. 5 of 30
To calculate the pressure drop at a certain flow rate use the formula
PD = (YourFlow / MaxFlow)2 x PDmax
PD = Pressure drop at YourFlow.
YourFLow = flow rate (ml/min or l/min)
MaxFlow = 100% rated flow for the sensor (in same units as YourFlow)
max = Pressure drop at 100% rated flow (see chart above)
PD
If the pressure available for the flow sensor is BELOW the pressure drop at
the required flow then flow through the unit will not be possible.
For example: You have a sensor with a 50-500ml/min flow range and want
to know the pressure drop (or minimum differential pressure required) at
300ml/min. According to the chart above, the pressure drop at 100% of
flow is approx 10psi. Using the formula above:
PD = (300 / 500)2 x 10 = 3.6 psid
Therefore, at 300ml/min the minimum required differential pressure for this
unit is 3.6 psid. This means that if the available pressure is below 3.6psid
then flow through the unit may not be possible.
Pressure drop through a system is cumulative. If the total pressure drop
across all the components in a system exceeds the minimum pressure
available then flow will not be possible.
For example: A system has a pressure of 30-40psi. There are several
components and the sum of their pressure drops at the required flow rate is
32psid. If the system is operating at 30psi flow would not be possible as the
total of the pressure drops would be greater than the pressure available.
The system will only operate if the system pressure is above 32psi.
If there is any possibility that there may be bubbles or entrapped gas in the
system then the outlet tubing should be elevated above the inlet port. This
will enable any gas that may become entrapped in the unit to escape (see
section C2 for details of how to remove entrapped air or gas).
2.Mounting the Flow Sensor or Flow Meter.
The TF Series of Flow Sensors and Flow Meters have no particular
installation requirements so may be mounted in any convenient position.
It is recommended that units be fixed to a suitable substrate with #4 screws
using the two mounting holes provided.
F-20-TF-L, pg. 6 of 30
Mounting The Sensor
(Mounting hardware not included with sensor)
3.Tubing Connections
CAUTION: DO NOT FLOW ANY GAS THROUGH A LIQUID
FLOW SENSOR OR FLOW METER. THIS WILL DAMAGE THE
MICRO-TURBINE ASSEMBLY AND VOID THE WARRANTY.
All tubing must be clean and without crimps, burrs or scratches.
Only use the fittings factory installed on the unit. If the fittings are removed
the calibration of the unit may be effected and leaking may occur. If
different fittings are required please contact Dwyer Customer Service for
assistance.
When connecting the sensor to the tubing a wrench should be used to stop
the fitting rotating in the sensor body. Take care not to over tighten the
fittings or leaking may occur.
Connecting and tightening the Fittings
F-20-TF-L, pg. 7 of 30
CAUTION: DO NOT over tighten the fittings into the sensor
body. Excessive force may damage the sensor body and will
NOT be repaired under warranty.
The flow direction for the unit is clearly marked on the label. Do not reverse
the flow direction or the unit will not function.
Close Up of Label Showing Flow Directions
(All models similar)
For the best results, precautions should be taken to minimize the amount of
turbulence in the flow stream. Avoid valves, fittings, curves in the tube, and
changes in the internal diameter or any other restrictions close to the inlet of
the sensor. The tubing ID should be as large as possible and preferably not
smaller than the ID of the fittings installed on the sensor.
For 0.2-2.0 L/min units a 10 cm straight length of tube before the sensor is
recommended. For higher flow range units (0.5-5.0 L/min and 1.0-10.0
L/min), a 20 cm straight length of tubing before the sensor is recommended.
If this is not possible it is recommended that straight lengths between all
connections on the inlet side of the sensor are as long as possible and 90
degree fittings (with a large enough ID) are used instead of curves in the
tubing.
4.Electrical Connections
CAUTION: Incorrect wiring may cause severe damage to the
unit. Applying an AC voltage (115VAC or 230VAC) directly to
the unit will cause damage. Read the following instructions
carefully before making any connections.
F-20-TF-L, pg. 8 of 30
a)Overview
The TFP-LP and TFM-LP series provide a 0-5VDC and pulse output
proportional to the flow rate. This output may be connected to a display,
data acquisition system or voltmeter.
The TFP-LI and TFM-LI feature an integral display that provides a local flow
reading. These units also have a 0-5VDC analog output available. If
required, this may be connected to another display, data acquisition system
or voltmeter.
A stable D.C. power supply is required to operate the unit. The voltage and
current requirements depend on the configuration of the unit. Full details
may be found in the Specifications section of this manual.
Connecting wires should be as short as possible to avoid voltage drops.
Twisted 2 pair conductor cable should be used if the length of the power
wires is to be longer than 1 meter.
Units are supplied with an integral 4 pin connector. Connections to the unit
are made using a mating cable assembly or power adapter package as
detailed in the following sections (parts 4.b to 4.e of this manual). A
connector pin and wire color cross reference may also be found in
Appendix G (Page 27) of this manual.
b)Connecting the Cable Assembly
The connector on the end of the cable assembly should be pushed into the
mating socket on the sensor taking care to ensure that it is the correct way
up.
Connecting the Cable Assembly
(All models similar)
Electrical Connections are made to the cable assembly as detailed in the
following sections.
F-20-TF-L, pg. 9 of 30
Loading...
+ 21 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.