Dwyer HM28 User Manual

Operating instructions Digital handheld pressure gauge
HM28
Contents
1 Description.......................................4
2 Safety details ...................................4
3.1 Switching on and off ........................4
3.2 Pneumatic connection .....................4
3.3 Operating modes .............................4
3.4 Configuration ...................................6
3.5 RS 232 mode ................................... 7
3.6 Battery change ..............................10
3.7 Recalibration..................................10
4 Specifications ................................ 10
4.1 Technical data ............................... 10
4.2 Measuring ranges ..........................12
4.3 Mains supply unit connection ........12
5 Maintenance ..................................12
6 Warning messages and faults .......13
7 Accessories ...................................13
8 Calibration instructions ..................14
2
Operating instructions
ON = On OFF = Off MODE = Select operating mode
- Pressure measurement
- Min./Max.
- Leak rate
- Datalogging CLEAR/ZERO = Zero setting PRINT/HOLD = Print and freeze display START/STOP = Data recording / printing SET = Select config. modes SELECT = Select config. data
c Matrix display d Connection for pneumatic hoses e Connection for plug-in power supply unit f Connection for RS 232 interface
d
c
e
f
Important!
Please note warning symbol in the operating instructions.
Copyright (c) by LOVE CONTROLS. The contents of this publication may not be copied without the express permission of LOVE CONTROLS. Subject to dimension and design changes.
3
1 Description
The „LOVE CONTROLS“ handheld pressure gauge HM28 is a digital pressure measuring instrument with integrated pressure sensor for the measurement of differential, relative or absolute pressures and vacuum. Its versatile range of functions and high precision makes it suitable for a wide range of applications.
Configuration possibilities:
Autom. switching off, selectable time
Selectable resolution and damping
Selectable units of time
Selectable measurement units
General reset to standard setting
Selectable data transfer rate
Configuration printout
Selectable measuring intervals
2 Safety details
Switching off Press the OFF key, or auto-
matic switching off 1,10 or 60
min after the last keystroke. In case of a temperature change the instrument requires at least 30 minutes to adjust to the new ambient temperature.
3.2 Pneumatic connection
For measuring ranges up to 7000 mbar, 4/6 mm or NPT 1/8" connectors can be connected depending on model. Higher ranges are equipped with NPT 1/8".
Ensure correct connection of the pneumatic hoses!
+P high pressure S- low pressure (not available
with the absolute and relative pressure version)
The pressure values stated on the rating plate and mentioned in these operating instructions must not be exceeded otherwise the pressure sensor can be destroyed.
Do not use the instrument in danger zones (explosives zones)! Wear eye protection for pres­sures > 1bar!
3 Operating
3.1 Switching on and off
Switching on Press the ON key (the maxi-
mum permissible measuring range and the class appears on the display). For precision measurements the instrument must be switch­ed on for at least 1 minute (warm-up phase).
Measuring negative differential or relative pressures:
connect “S-” or
change ports “S-“ and “+P”.
The HM28 will display a positive value.
3.3 Operating modes
The modes in the matrix below can be selected in succession by pressing the MODE key.
After switching ON, the instrument is always in the mode „normal“.
4
Î negative pressure,
When screwing on a coupling it is important to hold with a span­ner in the opposite direction. Never hold at case only!
Switching on
Normal (differential) and analogue/hold
Normal (absolute) and differential
Zoom
Min/Max
Leak rate
Tendency (absolute)
Record 1)
Print Record
print hold
print
print
print
print
print
start/stop
print
quick keystroke: max. output speed
Keystroke approx.1s: single or fast output start/stop printing
zero
display
zero
diff
zero
display
zero
(reset)
zero
display
clear
memory
Display example
Max. Range:
200 mbar
Full scale error
0.05 %
123.45 mbar ====== or HLD 123.4 mbar
1013.2 mbar
DIF 1013.2 mbar
123.45
MAX 150.0 mbar MIN 100.0 mbar
1234.5 mbar
LEK 2.1 /min
1013.2 mbar
TND - 15.7 /min REC 432.1 mbar
STOP 30 s MEM 1 REC 432.1 mbar
RUN 30 s MEM 321
432.1 mbar STOP 30 s MEM 901
PRINT RECORD PUSH PRINT
PRT 432.1 mbar 321 00:13:30
PRT Record stopped
Notes
Measuring range displayed in desired unit
Accuracy class as a % of full scale reading
Only with differential or relative pressure models
Hold, store indicated value
Only with absolute pressure mode Zero
sets Diff to 0 Measurement in large
figures With absolute pressure,
reset to actual value
Only with Diff./Rel.
(unit/time unit)
Only with absolute pressure (unit/time unit)
If measurement memory cleared
Recording runs up to 964 values
Recording stopped
Appears only if values are stored
Printing / display with time
Recording stopped
1) at 70 bar the measurements are stored in [bar]
5