The Series HATX transmitter has been carefully calibrated before
shipment to you, and it should be handled with the same care given any
precision instrument. Pressure range and dimensions are reported on the
specifications bulletin for the transmitter.
Ambient Conditions
Do not use in ambient conditions corrosive to anodized aluminum, or
submerge in liquids, subject to spray or vibration environment.
Electrical Connections:
Standard
˝Belden˝ Cable Lead
Function
Positive Excitation
Positive Output
Negative Output
Negative Excitation
Case
(#8723 Grey Cable)
Red
Green
White
Black
Shield
Electrical
The electrical circuit is equivalent to a 4-terminal network which can be
grounded at only one point, either at the negative excitation or the
negative signal output lead, but must not be commoned or grounded at
more than one point.
The unit must be operated with the case connected either to the
negative excitation terminal or to the negative output terminal. Failure to
do this may result in damage to, or unsatisfactory
operation of the unit. This connection may be made by connecting shield
and black (negative excitation) leads or alternatively by connecting the
shield and negative output signal leads together. Best shielding against
noise will be obtained by connecting the shield and negative excitation
leads. Circuit is reversed voltage protected for at least 5 minutes.
Internal transient suppression network is provided for short duration
transients to 150 volts.
SPECIFICATIONS
Service: Non-conducting air/gas.
Wetted Materials: Anodized aluminum, alumina ceramics, gold,
fluorocarbon elastomer sealant & Buna-N O-ring.
Accuracy: <±0.05% FS.
Stability: <±0.1% FS over 6 months @ 70°F.
Pressure Limits:
Barometric Ranges: 800 to 1100 hPa/mb;
600 to 1100 hPa/mb: 20 psia.
Absolute and Gage Ranges: 1.5x of std range.
Temperature Limits:
Operating: 0 to 175°F (-18 to 71°C);
Storage: -65 to 250°F (-53 to 121°C).
Compensated Temperature Range: 30 to 120°F (-1 to 49°C).
Thermal Effect:
Barometric Ranges: <±0.2% FS / 100°F;
PSIG and PSIA Ranges: <±0.1% FS / 100°F.
Power Requirements: 22 to 32 VDC.
Output Signal: 0 to 5 VDC.
Zero and Span Output:
Zero output: Factory set to within ±5 mV;
Span (FS) output: Factory set to within ±5 mV.
Zero and Span Adjustments: None.
Response Time: <10 msec.
Current Consumption: <10 mA.
Electrical Connections: 2 ft multi-conductor cable.
Process Connections: 1/8˝-27 NPT internal.
Mounting Orientation: Pressure port 90° parallel to ground.
Weight: 9 oz (255 g).
Agency Approval: CE.
In some instances, use of long cables (several hundred feet long), may
introduce enough cable capacitance into the output circuit to cause
output oscillation. If encountered, this oscillation may be eliminated by
connecting a 100 ohm resistor (1/8 watt or larger) in series in each of the
output leads at the end of the 2 ft transducer cable. These series
resistors of course add to the output resistance.
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
Adjustments (with cover removed)
CAUTION
1. Before making any adjustments, determine which circuit board
diagram on page 2 matches the circuit board in your transmitter.
2. Turn off excitation power during cover removal or replacement.
3. Touching any adjustments other than zero or span may
necessitate recalibration and voids the warranty.
ZERO OUTPUT
Output can be adjusted to zero by potentiometer as shown in diagram. Unit
factory adjusted to 0.00 VDC output (±5 mV).
SPAN
Can be adjusted by potentiometer as shown in diagram. Unit factory
adjusted to 5.00 VDC FS output (±5 mV).
LINEARITY - DO NOT TOUCH
Factory adjusted for best linearity.
Circuit Board 1
INSTALLATION INSTRUCTIONS FOR HOOK-UP
OF 4-WIRE PRESSURE TRANSMITTERS
Single Power Supply, Single Output
GREEN
OUT
WHITE
RECEIVER
PRESSURE
TRANSMITTER
NOTE: GROUND 1 & 2 MUST BE ELECTRONICALLY ISOLATED
RED
EX
BLACK
SHIELD
Power Supply (Nominal 24 VDC)
HATX - Pressure Transmitter (4 Wire Circuit)
Receiver - Digital Volt Meter or Data Acquisition System
Single Power Supply, Multiple Outputs
POWER
SOURCE
22 - 32 VDC
Circuit Board 2
NOTE: When circuit cover is removed, a zero shift can be noticed. The
zero will return by the same amount when the cover is replaced. This
displacement of the output should be taken into account when making
adjustments or during recalibration.
*Adjustments: Due to the high accuracy of this transmitter, adjustments to
zero and span should only be made with the use of primary pressure
standards of known accuracy higher than that of the Series HATX.
NOTE: The shield is internally commoned to the case and pressure port
of the transmitter. When the shield is connected to ground the case and
pressure port of the transmitter will also be commoned to that ground.
MAINTENANCE
Upon final installation of the Series HATX High Accuracy Pressure
Transmitter, no routine maintenance is required. A periodic check of
system calibration is recommended. The Series HATX is not field
serviceable and should be returned if repair is needed (field repair should
not be attempted and may void warranty). Be sure to include a brief
description of the problem plus any relevant application notes. Contact
customer service to receive a return goods authorization number before
shipping.
©Copyright 2010 Dwyer Instruments, Inc. Printed in U.S.A. 12/10 FR# R1-443859-00
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com