our GFM3/4 Mass Flow Meter was carefully packed in a sturdy cardboard carton,
Y
with anti-static cushioning materials to withstand shipping shock. Upon receipt,
inspect the package for possible external damage. In case of external damage to
the package contact the shipping company immediately.
.2 - Unpack the Mass Flow Meter
1
pen the carton carefully from the top and inspect for any sign of concealed
O
hipping damage. In addition to contacting the shipping carrier, please forward a
s
copy of any damage report to your distributor or Dwyer Instruments, Inc
.
directly.
When unpacking the instrument please make sure that you have all the items
indicated on the Packing List. Please report any shortages promptly.
.3 - Maintenance/Repair
1
pon final installation of the Series GFM3/4, no routine maintenance is required.
U
The Series GFM3/4 is not field serviceable and should be returned if repair is
needed. Field repair should not be attempted and may void warranty.
Warranty/Return
efer to “Terms and Conditions of Sales” in our catalog and on our website. Contact
R
ustomer service to receive a Return Goods Authorization number before shipping
c
he product back for repair. Be sure to include a brief description of the problem
t
plus any additional application notes
2. INSTALLATION
2.1 - Primary Gas Connections
lease note that the GFM3/4 Mass Flow Meter will not operate with liquids. Only
P
lean gases are allowed to be introduced into the instrument. If gases are
c
ontaminated they must be filtered to prevent the introduction of impediments into
c
the sensor.
AUTION
C
application. For more information, contact your distributor or Dwyer
GFM3/4 transducers should not be used for monitoring
OXYGEN gas unless specifically cleaned and prepared for such
®
Instruments,
Inc.
Attitude limit of Mass Flow Meter is ±15deg from calibration position (standard
calibration is in horizontal position). This means that the gas flow path of the Flow
Meter must be within this limit in order to maintain the original calibration accuracy.
Should there be need for a different orientation of the meter, re-calibration may be
necessary. It is also preferable to install the GFM3/4 transducer in a stable
environment, free of frequent and sudden temperature changes, high moisture, and
drafts.
Prior to connecting gas lines, inspect all parts of the piping system including
ferrules and fittings for dust or other contaminants. Be sure to observe the direction
of gas flow as indicated by the arrow on the front of the meter when connecting the
gas system to be monitored. Insert tubing into the compression fittings until the
ends of the properly sized tubing home flush against the shoulders of the fittings.
Compression fittings are to be tightened according to the manufacturer's
instructions to one and one quarter turns. Avoid over tightening which will seriously
damage the Restrictor Flow Elements (RFE's)!
WARNING
For models GFM4 (Multi Parameter versions) the maximum
pressure in the gas line should not exceed 100 PSIA
(6.8 bars). Applying pressure above 100 PSIA (6.8 bars) for extended periods of
time will seriously damage the pressure sensor and may cause serious injury or
death. Burst pressure is 200 PSIA (13.6 bar)!
GFM3 transducers are supplied with standard 1/4˝ or 3/8˝, or optional 1/8˝ inlet and
outlet compression fittings which should not be removed unless the meter is being
cleaned or calibrated for a new flow range.
Using a Helium Leak Detector, or other equivalent method, perform a thorough
leak test of the entire system. (All GFM3/4’s are checked prior to shipment for
leakage within stated limits. See specifications in this manual.)
2.2 - Electrical Connections
GFM3/4 is supplied with a 25-pin "D" connector. Pin diagram is presented in figure b-
1.
2.2.1 - Power Supply Connections
GFM3/4 transducers are supplied for three different power supply options:
±15Vdc (bipolar power supply)
DC Power (+) --------------- pin 1 of the 25-pin "D" connector
DC Power Common --------------- pin 18 of the 25-pin "D" connector
DC Power (-) --------------- pin 14 of the 25-pin "D" connector
12Vdc or +24Vdc (unipolar power supply)
+
CAUTION
dc! (See power requirements label at the rear of the GFM3/4 meter)
V
.2.2 - Output Signals Connections
2
C
C Power (+) --------------- pin 1 of the 25-pin "D" connector
D
C Power (-) --------------- pin 18 of the 25-pin "D" connector
D
O NOT CONNECT 24 Vdc POWER SUPPLY UNLESS YOUR
D
FM3/4 METER WAS ORDERED AND CONFIGURED FOR 24
G
AUTION
When connecting the load to the output terminals, do not
exceed the rated values shown in the specifications. Failure to
do so might cause damage to this device. Be sure to check that the wiring and
he polarity of the power supply is correct before turning the power ON. Wiring
t
rror may cause damage or faulty operation.
e
FM3/4 Mass Flow Meters are equipped with either calibrated 0 to 5 VDC (0 to 10
G
DC optional) or calibrated 4 to 20 mA output signals (jumper selectable). This
V
inear output signal represents 0 to 100% of the flow meter's full-scale range. Multi
l
Parameter versions (GFM3/4) are in addition equipped with either a calibrated 0 to
5 VDC (0 to 10 VDC optional) or a calibrated 4 to 20 mA output signal (jumper
selectable) for pressure and temperature. Pressure linear output signal represents
0 to 100 PSIA (46.9 kPa). Temperature linear output signal represents 0 to 50
CAUTION
utput signals!
o
All 4 to 20 mA current loop outputs are self-powered (non-
solated). Do not connect an external voltage source to the
i
°
C.
Flow 0 to 5 VDC or 4 to 20 mA output signal connection:
Plus (+)------------------- pin 2 of the 25-pin "D" connector
Minus (-) ------------------- pin 15 of the 25-pin "D" connector
emperature 0 to 5 VDC or 4 to 20 mA output signal connection
T
lus (+) -------------------- pin 3 of the 25-pin "D" connector
P
Minus (-) -------------------- pin 16 of the 25-pin "D" connector
Pressure 0 to 5 VDC or 4 to 20 mA output signal connection
Plus (+)-------------------- pin 4 of the 25-pin "D" connector
Minus (-)-------------------- pin 17 of the 25-pin "D" connector
To eliminate the possibility of noise interference, use a separate cable entry for the
DC power and signal lines.
2.2.3 - Communication Parameters and Connections
The digital interface operates via RS485 (optional RS-232 is available), and
provides access to applicable internal data including: flow, temperature, pressure
reading, auto zero, totalizer and alarm settings, gas table, conversion factors and
engineering units selection, dynamic response compensation and linearization
table adjustment.
The RS485 converter/adapter has to be configured for: multidrop, 2-wire, half
duplex mode. The transmitter circuit has to be enabled by TD or RTS (depending
on which is available on the converter/adapter). Settings for the receiver circuit
usually should follow the selection made for the transmitter circuit in order to
eliminate echo.
RS-485 T(-) or R(-) -------- pin 11 of the 25-pin "D" connector (-)
RS-485 T(+) or R(+) --------pin 24 of the 25-pin "D" connector (+)
RS-485 GND (if available) --------pin 20 of the 25-pin "D" connector (GND)
RS-232 Communication Interface Connection:
Crossover connection has to be established:
RS-232 RX
(pin 2 on the DB9 connector) -------- pin 11 of the 25-pin "D" connector (TX)
RS-232 TX
(pin 3 on the DB9 connector) -------- pin 24 of the 25-pin "D" connector (RX)
RS-232 GND
(pin 5 on the DB9 connector) -------- pin 20 of the 25-pin "D" connector (GND)
standardized. There are, however, some connectors with
nonconforming patterns and the numbering sequence on your mating
connector may or may not coincide with the numbering sequence shown in our
pin configuration table above. It is imperative that you match the appropriate wires
in accordance with the correct sequence regardless of then particular numbers
displayed on the mating connector.
NOTICE
Make sure power is OFF when connecting or disconnecting
any cables in the system.
The (+) and (-) power inputs are each protected by a 400 mA M (medium time-lag)
resettable fuse. If a shorting condition or polarity reversal occurs, the fuse will cut
power to the flow transducer circuit. Disconnect the power to the unit, remove the
faulty condition, and reconnect the power. The fuse will reset once the faulty
condition has been removed. DC Power cable length may not exceed 9.5´ (3
meters).
he flow rate can be displayed in volumetric flow or mass flow engineering units for
T
tandard or actual (temperature, pressure) conditions. Flow meters can be
s
rogrammed locally via the four button keypad and LCD, or remotely, via the RS-
p
32/RS-485 interface. GFM3/4 flow meters support various functions including:
2
flow totalizer, flow, temperature, pressure alarms, automatic zero adjustment, 2
SPDT relays output, 0 to 5 Vdc / 0 to 10 Vdc / 4 to 20 mA analog outputs for flow,
pressure and temperature.
. SPECIFICATIONS
4
FM3
G
ervice: Clean gases compatible with wetted parts.
S
Wetted Materials: 316 SS, 416 SS; Fluoroelastomer, Buna-N, EPR or PTFE O-
rings.
Accuracy: ±1% FS.
epeatability: ±0.25% FS.
R
esponse Time: 0.6 to 1.0 s to within ±2% of setpoint over 20 to 100% FS.
R
utput Signal: Linear 0 to 5 VDC (3000 Ω min. load impedance); 0 to 10 VDC
O
(6000 Ω min. load impedance); 4 to 20 mA (500 Ω max. loop resistance).
Max. Particulate Size: 5 microns.
Temperature Limits: Ambient: 32 to 122°F (0 to 50°C); Dry Gases: 14 to 122°F (-
10 to 50°C).
ower Supply: ±12 VDC; ±15 VDC; ±24 VDC.
P
rocess Connections: 1/8˝ compression fitting for flow rates ≤ 10 L/min; 1/4˝ for ≤
Accuracy: ±1% FS.
Repeatability: ±0.25% FS.
Response Time: 0.6 to 1.0 s to within ±2% of setpoint over 20 to 100% FS.
Output Signal: Linear 0 to 5 VDC (3000 Ω min. load impedance); 0 to 10 VDC
(6000 Ω min. load impedance); 4 to 20 mA (500 Ω max. loop resistance).
Max. Particulate Size: 5 microns.
Temperature Limits: Ambient: 32 to 122°F (0 to 50°C); Dry Gases: 14 to 122°F (-
10 to 50°C).
Power Supply: ±12 VDC; ±15 VDC; ±24 VDC.
Process Connections: 1/8˝ compression fitting for flow rates ≤ 10 L/min; 1/4˝ for ≤
Use of the GFM3/4 flow transducer in a manner other than that specified in this
manual or in writing from Dwyer
®
Instruments, Inc, may impair the protection
provided by the equipment.
3. PRINCIPLE OF OPERATION
The stream of gas entering the Mass Flow transducer is split by shunting a small
portion of the flow through a capillary stainless steel sensor tube. The remainder of
the gas flows through the primary flow conduit. The geometry of the primary conduit
and the sensor tube are designed to ensure laminar flow in each branch. According
to principles of fluid dynamics, the flow rates of a gas in the two laminar flow
conduits are proportional to one another. Therefore, the flow rates measured in the
sensor tube are directly proportional to the total flow through the transducer. In
order to sense the flow in the sensor tube, heat flux is introduced at two sections of
the sensor tube by means of precision wound heater sensor coils. Heat is
transferred through the thin wall of the sensor tube to the gas flowing inside. As gas
flow takes place heat is carried by the gas stream from the upstream coil to the
downstream coil windings. The resultant temperature dependent resistance
differential is detected by the electronic control circuit. The measured gradient at
the sensor windings is linearly proportional to the instantaneous rate of flow taking
place. An output signal is generated that is a function of the amount of heat carried
by the gases to indicate mass molecular based flow rates. Additionally, the GFM3/4
model Mass Flow Meter incorporates a Digital Signal Processor (DSP) and nonvolatile memory that stores all hardware specific variables and up to 10 different
calibration tables. Multi parameter flow meters provide accurate data on three
different fluid parameters:
• flow
• pressure
• temperature
Page 4
Flow rates are stated for Nitrogen at STP conditions [i.e. 70°F (21.1°C) at 1
*
tm]. For other gases use the K factor as a multiplier from APPENDIX 2.
a
low Rate
odel
M
std liters/min)
(
GFM3/4-X-010
GFM3/4-X-050
GFM3/4-X-100
T
. OPERATING INSTRUCTIONS
5
F
p to 10
u
0
5
100
able IV - Pressure Drops
aximum Pressure Drop
M
(mm H
0)
(psid)
2
25
0.04
200
2
500
5
.23
3
.08
8
(kPa)
0.276
2.3
2
5.7
5
5.1 - Preparation and Warm Up
It is assumed that the Digital Mass Flow Meter has been correctly installed and
horoughly leak tested as described in section 2. Make sure the flow source is OFF.
t
hen applying power to a flow meter, within the first two seconds you will see on
W
he LCD display: the product name, the software version, and revision of the
t
EEPROM table. After two seconds the LSD display switches to the main screen
with the following information:
• Temperature and Pressure reading (for models GFM4 only).
Mass Flow reading in current engineering units.
•
Current Gas Table and Gas Name.
•
Totalizer Volume reading in current volume based engineering units.
•
• Totalizer , Alarm, and Relays status.
¡
F
GFM3/4 Main Screen
NOTICE
Allow the Digital Mass Flow Meter to warm-up for a minimum of
15 minutes.
During initial powering of the GFM3/4 transducer, the flow output signal will be
indicating a higher than usual output. This is an indication that the GFM3/4
transducer has not yet attained its minimum operating temperature. This condition
will automatically cancel within a few minutes and the transducer should eventually
zero.
5.2 - Swamping Condition
If a flow of more than 10% above the maximum flow rate of the Mass Flow Meter
is taking place, a condition known as "swamping" may occur. Readings of a
"swamped" meter cannot be assumed to be either accurate or linear. Flow must
be restored to below 110% of maximum meter range. Once flow rates are lowered
to within calibrated range, the swamping condition will end. Operation of the meter
above 110% of maximum calibrated flow may increase recovery time. -
5.3 - Programming GFM3/4 using LCD and Keypad
All features of the flow meter can be accessed via the local four button keypad and
LCD. The LCD incorporates an energy-saving auto shut-off backlit feature. If
enabled, after 15 minutes of operation without user intervention the LCD backlight
turns off. In order to turn on the LCD backlight press any key on the keypad. The
LCD backlight energy-saving auto shut-off feature can be disabled or enabled by
user (see p. 5.3.12 "LCD backlight Energy-saving Setting").
.3.1 - Changing Units of Measurement for Temperature and Pressure
5
eading
R
y default after power up, the temperature reading is displayed in °F and pressure
B
n PSI. Pressing ( ) [Enter] button from main screen will alter the units of measure
i
to °C for temperature and kPa for pressure reading respectively. In order to change
units of measure back to °F for temperature and PSI for pressure, press ( )
[Enter] button while in the main screen one more time.
.3.2 - Monitoring GFM3/4 Peripheries Settings
5
he last row at the bottom of the main LCD screen reflects settings and status for
T
Totalizer, Flow Alarm, and Relays (see Figure b-2).
¡
F
igure B-2
F
Totalizer Status:
TOT: R - totalizer is running (Enabled).
TOT: S - totalizer is stopped (Disabled).
low Alarm Status:
F
: S- flow alarm is disabled.
A
A: R,N - flow alarm is enabled and currently there are no alarm conditions.
A: R,L - flow alarm is enabled and currently there is Low alarm condition.
A: R,H - flow alarm is enabled and currently there is High alarm condition.
Relay Settings:
N - No assignment (relay is not assigned to any events).
H - High Flow Alarm condition.
L - Low Flow Alarm condition.
R - Range between High and Low Flow Alarm condition.
T - Totalizer reached set limit.
A - High Temperature Alarm condition.
B - Low Temperature Alarm condition.
C - High Pressure Alarm condition.
D - Low Pressure Alarm condition.
Continued pressing of the ( ) [Up] button from the main screen will switch the
status line to display the following information:
- Calibrated full-scale range in standard L/min for current Gas Table.
- Device Digital Communication interface type (RS-485 or RS-232).
Pressing the ( ) [Dn] button from any of the status line will
switch the status display to one step back.
Page 5
5.3.3 - GFM3/4 Main Menu
Pressing of the ( ) [Esc] button from the main screen will switch the display to the
ain Menu. The following screen will appear:
M
i
If the User defined Unit of Measure option is selected the following screen will
appear:
igure b-5, User Defined Unit of Measure Screen (K factor)
F
Figure b-3, GFM3/4 Main Menu Screen
ressing of the ( ) [Up] or ( ) [Dn] buttons allows the user to scroll up or down
P
he menu options. Press ( ) [Enter] button to select the highlighted option of the
t
enu.
m
The following menu options are available:
1. Units of Measure - View or Change the Units of Measure for Flow process
ariable.
v
. Gas Table Select - View or Change the Gas Table.
2
. Totalizer - View or Change settings for Totalizer.
3
4. Alarm Settings - View or Change settings for Flow, Pressure and Temperature
larm.
A
. Relay Action - View or Change settings for each of two available Relays.
5
6. K Factors - View or Change settings for User defined or Internal K Factors.
7. Zero Calibration - Initiate Automatic Sensor Zero Calibration.
8. Flow Conditions - Allows the user to set the Actual or Standard Flow conditions.
9. BackLight Timer - Allows the user to turn On/Off the Energy-saving for LCD
acklight.
b
10.Exit - Returns to the Main Screen with process variables reading.
NOTICE
5.3.4 Gas Flow Engineering Units Settings
While in the Main Menu scroll with ( ) [Up] or ( ) [Dn] button to highlight the
Units of Measure option and press the ( ) [Enter] button. The following screen will
appear.
Pressing the ( ) [Esc] button from any level of the Menu will
switch the menu to one level higher (up to Main Screen).
Figure b-4, GFM3/4 Units of Measure Screen
In order to specify the User Defined Unit of Measure user has to set three key
parameters:
factor - Conversion factor relative to L/min unit of measure.
K
ime base - Hours, Minutes, or Seconds.
T
Density - Use density (YES / NO).
Press the ( ) [Enter] button to move the flashing cursor to the digit that has to be
adjusted. Pressing ( ) or ( ) will increment or decrement a particular digit
espectively. The numbers will change from 0 to 9 and next to the decimal point (.).
r
ressing the ( ) button one more time will change the digit on the highlighted
P
osition of the cursor back to 0. The same is true in reverse when pressing the
p
( ) button. Only one decimal point is allowed. If changing position of the decimal
oint is required, change decimal point to any desired digit then move the cursor to
p
he required position and adjust it to the decimal point with ( ) or ( ) button. When
t
complete with K-factor value settings, press the ( ) [Esc] button to move in to the
Time base settings screen. The following screen will appear:
Figure b-6, User Defined Unit of Measure Screen (Time base)
Use ( ) or ( ) buttons to highlight desired time base option. Press the ( ) [Enter]
button to set the Time base and move in to the density settings screen. The
following screen will appear:
i
The following Engineering Units menu options are available:
1. % FS - percent of full-scale.
2. L/min - Liters per minute.
3. L/h - Liters per hour.
4. mL/min - milliliters per minute.
5. mL/h - milliliters per hour.
6. SCFH - cubic feet per hour.
7. SCFM - cubic feet per minute.
8. LbPH - pounds per hour.
9. LbPM - pounds per minute.
10. User - User defined Unit of Measure.
11. Exit - Exit to Main Menu
Selecting option 1 to 9 sets the corresponding Unit of Measure and switches the
LCD back to Main Menu.
NOTICE
Once Flow Unit of Measure is changed the Totalizer's Volume
based Unit of Measure will be changed automatically.
Figure b-7, User Defined Unit of Measure Screen (Density)
Use ( ) or ( ) button to highlight desired density option. Press the ( ) [Enter]
button when done. The LCD will display the Units of Measure Screen and new
settings will be reflected at the bottom status line.
5.3.5 - Gas Table Settings
The GFM3/4 Mass Flow Meter is capable to store calibration data for up to 10
different gases.
NOTICE
ordered). If instead of the valid Gas Name (for example NITROGEN) the main
creen displays Gas designator as "Uncalibrated", then the user has chosen the
gas table which was not calibrated. Using an Uncalibrated Gas Table will result in
erroneous reading.
Page 6
By default the GFM3/4 is shipped with at least one valid
calibration table (unless optional additional calibrations were
rom the Main Menu, the user would traverse the menu tree until reaching the "Gas
F
able Select" menu. The following screen will appear:
T
se ( ) or ( ) button to highlight "Start at Flow" option and press the ( ) [Enter]
U
utton. The following screen will appear:
b
’
Figure b-8, Current Gas Table Settings
se ( ) or ( ) button to select desired Gas Table, and press the ( ) [Enter]
U
utton. The LCD will display the Main Menu screen. If desired, press the ( ) [Esc]
b
button to go back to the Main GFM3/4 screen. The new gas table number and the
name of the gas will be reflected on the Main GFM3/4 screen.
5.3.6 - Totalizer Settings
he total volume of the gas is calculated by integrating the actual gas flow rate with
T
espect to time. Both keypad menu and digital interface commands are provided to:
r
• Set the totalizer to ZERO.
• Start the totalizer at a preset flow.
• Assign action at a preset total volume.
Start/stop (enable/disable) totalizing the flow.
•
Read totalizer.
•
The totalizer has several attributes which may be configured by the user. These
attributes control the conditions which cause the totalizer to start integrating the gas
flow, and also specify actions to be taken when the Total Volume is outside the
specified limit.
NOTICE
entered in %FS engineering unit. Totalizer will not totalize until the flow rate
becomes equal or more than the Totalizer Start value. Totalizer Stop values have to
be entered in volume / mass based engineering units.
Totalizer action conditions become true when the totalizer reading and preset "Stop
at Total" volumes are equal.
From the Main Menu, the user would traverse the menu tree until reaching the
"Totalizer" menu. The following screen will appear:
Before enabling the Totalizer, ensure that all totalizer settings
are configured properly. Totalizer Start values have to be
Figure b-11, Totalizer Settings (Start)
ressing ( ) or ( ) will increment or decrement Start Flow value per 0.1% FS
P
espectively. When done with adjustment, press the ( ) [Enter] button.
r
Use ( ) or ( ) button to highlight "Stop at Total" option and press the ( ) [Enter]
button. The following screen will appear.
:
Figure b-12, Totalizer Settings (Stop)
Press the ( ) [Enter] button to move the flashing cursor to the digit that has to be
adjusted. Pressing ( ) or ( ) will increment or decrement a particular digit
respectively. The numbers will change from 0 to 9. Pressing the ( ) button one
more time will change the digit on the highlighted position of the cursor back to 0.
The same is true in reverse,when pressing the ( ) button. Only one decimal point
is allowed. If changing position of the decimal point is required, change decimal
point to any desired digit then move the cursor to the required position and adjust
it to the decimal point with ( ) or ( ) button. When done with adjustment, press
the ( ) [Esc] button.
5.3.7 - Alarm Settings
GFM3/4 provides the user a flexible alarm/warning system that monitors the
Process Variables (Gas Flow, Pressure and Temperature) for conditions that
fall outside configurable limits, and then provides feedback to the user visually
via the LCD (only for Flow) or via a Relay contact closure.
Figure b-9, Totalizer Settings
Mode Run/Stop - Allows the user to Enable/Disable Totalizer.
Start at Flow - Allows the user to enter Gas flow rate in %FS at which Totalizer
starts integrating of the gas flow.
Stop at Total - Allows the user to enter Totalizer Limit Volume when user defined
action will occur.
Reset to Zero - Allows the user to reset Totalizer reading to zero.
Use ( ) or ( ) button to highlight "Mode Run/Stop" option and press the ( )
[Enter] button. The following screen will appear:
’
Figure b-10, Totalizer Settings (Stop/Run)
There are three different alarms:
• Gas Flow - (GFM 3/4)
• Gas Temperature - (GFM4)
• Gas Pressure - (GFM4)
Each alarm has several attributes which may be configured by the user. These
attributes control the conditions which cause the alarm to occur, and also specify
actions to be taken when the Process Variable is outside the specified conditions.
NOTICE
conditions.
From the Main Menu, the user would traverse the menu tree until reaching the
"Alarm Settings" menu. The following screen will appear:
Page 7
All three alarms are non-latching. That means the alarm is
indicated only while the monitored value exceeds the specified
Figure b-13, Alarm Settings
Loading...
+ 15 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.