Dwyer GFM2 User Manual

Series GFM2 Digital Mass Flow Meter
Installation and Operating Instructions
Bulletin F-GFM2
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
Phone: 219/879-8000 www.dwyer-inst.com
ABLE OF CONTENTS
T
. Unpacking the GFM2 Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
.1 Inspect Package for External Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.2 Unpack the Mass Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.3 Returning Merchandise for Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
. Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
.1 Primary Gas Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
.2 Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
.2.1 Power Supply Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.2.2 Output Signal Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.2.3 Output Communication Parameters and Connections . . . . . . . . . . .3
. Principle of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
5. Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
5.1 Preparation and Warm Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
5.2 Swamping Condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
.3 GFM2 Parameters Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
.3.1 Engineering Units Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
.3.2 Gas Table Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
5.3.3 Totalizer Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
5.3.4 Flow Alarm Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
5.3.5 Relay Assignment Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
5.3.6 K-Factors Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
.3.7 Zero Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
.3.8 Self Diagnostic Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
.4 Analog Output Signals Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
. Calibration Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
.1 Flow Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
.2 Gas Flow Calibration of GFM2 Mass Flow Meters . . . . . . . . . . . . . . . . . .8
.2.1 Connections and Initial Warm Up . . . . . . . . . . . . . . . . . . . . . . . . . . .8
7.2.2 ZERO Check/Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
7.2.3 Gas Linearization Table Adjustment . . . . . . . . . . . . . . . . . . . . . . . . .8
7.3 Analog Output Calibration of GFM2 Mass Flow Meters . . . . . . . . . . . . . . .9
.3.1 Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
.3.2 Gas Flow 0 to 5 VDC Analog Output Calibration . . . . . . . . . . . . . . . .9
.3.3 Gas Flow 4 to 20 mA Analog Output Calibration . . . . . . . . . . . . . . . .9
8. RS-485/RS-232 Software Interface Commands . . . . . . . . . . . . . . . . . . . . . . .9
8.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
8.2 Commands Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
.3 ASCII Commands Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
. Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
9.1 Common Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
9.2 Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
9.3 Technical Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
0. Calibration Conversions from Reference Gases . . . . . . . . . . . . . . . . . . .13
ppendix I GFM2 EEPROM Variables . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
A
Appendix II Internal User Selectable Gas Factor Table
ppendix III Gas Factor Table (“K” Factors) . . . . . . . . . . . . . . . . . . . . . .16
A
ppendix IV Component Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
A
(Internal “K” Factors) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
6.2 Flow Path Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
6.2.1 Restrictor Flow Element (RFE) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
6.2.2 GFM2 Models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Appendix V Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Appendix VI Maintenance/Repair, Warranty/Return . . . . . . . . . . . . . . . . .19
Page 2
. UNPACKING THE GFM2 MASS FLOW METER
1.1 - Inspect Package for External Damage
Your GFM2 Mass Flow Meter was carefully packed in a sturdy cardboard carton,
ith anti-static cushioning materials to withstand shipping shock. Upon receipt,
w
nspect the package for possible external damage. In case of external damage to
i
he package contact the shipping company immediately.
t
1.2 - Unpack the Mass Flow Meter
Open the carton carefully from the top and inspect for any sign of concealed shipping damage. In addition to contacting the shipping carrier please forward a
opy of any damage report to your distributor or Dwyer Instruments, Inc directly.
c
hen unpacking the instrument please make sure that you have all the items
W
ndicated on the Packing List. Please report any shortages promptly.
i
1.3 - Maintenance/Repair
Upon final installation of the Series GFM2, no routine maintenance is required. The Series GFM2 is not field serviceable and should be returned if repair is needed.
ield repair should not be attempted and may void warranty.
F
arranty/Return
W
Refer to “Terms and Conditions of Sales” in our catalog and on our website. Contact customer service to receive a Return Goods Authorization number before shipping the product back for repair. Be sure to include a brief description of the problem plus any additional application notes.
. INSTALLATION
.1 - Primary Gas Connections
Please note that the GFM2 Mass Flow Meter will not operate with liquids. Only clean gases are allowed to be introduced into the instrument. If gases are contaminated they must be filtered to prevent the introduction of impediments into the sensor.
CAUTION
application. For more information, contact your distributor or Dwyer.
Attitude limit of Mass Flow Meter is ±15° from calibration position (standard calibration is in horizontal position). This means that the gas flow path of the Flow Meter must be within this limit in order to maintain the original calibration accuracy. Should there be need for a different orientation of the meter, re-calibration may be necessary. It is also preferable to install the GFM2 transducer in a stable environment, free of frequent and sudden temperature changes, high moisture, and drafts.
Prior to connecting gas lines inspect all parts of the piping system including ferrules and fittings for dust or other contaminants. Be sure to observe the direction of gas flow as indicated by the arrow on the front of the meter when connecting the gas system to be monitored. Insert tubing into the compression fittings until the ends of the properly sized tubing home flush against the shoulders of the fittings. Compression fittings are to be tightened according to the manufacturer's instructions to one and one quarter turns. Avoid over tightening which will seriously damage the Restrictor Flow Elements (RFE's)!
WARNING
pressure above 500 PSIA (34.47 bars) will seriously damage the flow sensor and may cause serious injury or death.
GFM2 transducers are supplied with standard 1/4˝ or 3/8˝ inlet and outlet compression fittings which should not be removed unless the meter is being cleaned or calibrated for a new flow range.
FM2 transducers should not be used for monitoring OXYGEN
G
as unless specifically cleaned and prepared for such
For models GFM2 models the maximum pressure in the gas line should not exceed 500 PSIA (34.47 bars). Applying
CAUTION
2.2.2 - Output Signals Connections
AUTION
C
do so might cause damage to this device. Be sure to check that the wiring and
he polarity of the power supply is correct before turning the power ON. Wiring
t
rror may cause damage or faulty operation.
GFM2 Mass Flow Meters are equipped with either calibrated 0 to 5 VDC (0 to 10
DC optional) or calibrated 4 to 20 mA output signals (jumper selectable). This
V
inear output signal represents 0 to 100% of the flow meter's full scale range.
l
AUTION
C
output signals!
Flow 0 to 5 VDC or 4 to 20 mA output signal connection:
P M
To eliminate the possibility of noise interference, use a separate cable entry for the DC power and signal lines.
2.2.3 - Communication Parameters and Connections
he digital interface operates via RS-485 (optional RS-232 or Profibus DP is
T
vailable) and provides access to applicable internal data including: flow, CPU
emperature, pressure reading, auto zero, totalizer and alarm settings, gas table,
t conversion factors and engineering units sel ection, dynamic response compensation and linearization table adjustment.
Communication Settings:
RS-485 Communication Interface Connection:
The RS-485 converter/adapter has to be configured for: multidrop, 2 wire, half duplex mode. Settings for the receiver circuit usually should follow the selection made for the transmitter circuit in order to eliminate echo.
RS-485 T(-) or R(-) -------- pin 8 of the 15 pin "D" connector (TX-) RS-485 T(+) or R(+) -------- pin 15 of the 15 pin "D" connector (RX+) RS-485 GND (if available) -------- pin 9 of the 15 pin "D" connector (GND)
RS-232 Communication Interface Connection:
Crossover connection has to be established:
RS-232 RX (pin 2 on the DB9 connector) -------- pin 8 of the 15 pin "D" connector (TX) RS-232 TX (pin 3 on the DB9 connector) -------- pin 15 of the 15 pin "D" connector (RX) RS-232 GND (pin 5 on the DB9 connector) -------- pin 9 of the 15 pin "D" connector (GND)
Do not apply power voltage above 26 VDC. Doing so will cause
FM2 damage or faulty operation.
G
When connecting the load to the output terminals, do not exceed the rated values shown in the specifications. Failure to
All 4 to 20 mA current loop outputs are self-powered (non­isolated). Do not connect an external voltage source to the
lus (+) ------------------- pin 2 of the 15 pin "D" connector
inus (-) ------------------- pin 1 of the 15 pin "D" connector
aud rate: -------- 9600 baud
B
top bit: -------- 1
S Data bits: -------- 8 Parity: -------- None Flow Control: -------- None
Using a Helium Leak Detector or other equivalent method perform a thorough leak test of the entire system. (All GFM2’s are checked prior to shipment for leakage within stated limits. See specifications in this manual.)
2.2 - Electrical Connections
GFM2 is supplied with a 25 pin "D" connector. Pin diagram is presented in figure b-1.
2.2.1 - Power Supply Connections
The power supply requirements for GFM2 transduers are: 11 to 26 VDC, (unipolar power supply)
DC Power (+) --------------- pin 7 of the 15 pin "D" connector DC Power (-) --------------- pin 5 of the 15 pin "D" connector
Page 3
IN GFM2 FUNCTION
P
1 Common, Signal Ground
For Pin 2 (4 to 20 mA return)
2 0 to 5 VDC or 4 to 20 mA
low Signal Output
F
Relay No. 2 - Normally Open
ontact
C
4 Relay No. 2 - Common
Contact
5 Common, Power Supply
- DC power for 11 to 26 VDC)
(
Relay No. 1 - Common
ontact
C
Plus Power Supply
(+ DC power for 11 to 26 Vdc) 8 RS485 (-) (Optional RS232 TX) 9 RS232 Signal GND (RS-485
ND Optional)
G
0 Do not connect
Test/Maintenance terminal)
(
11 Relay No. 2 - Normally Closed
Contact 12 Relay No. 1 - Normally Open
Contact
3 Relay No. 1 - Normally Closed
ontact
C
4 Do not connect
(Test/Maintenance terminal)
5 RS485 (+) (Optional RS232
X) Shield Chassis Ground
R
OTICE
N
onconforming patterns and the numbering sequence on your mating
NOTICE
In general, "D" Connector numbering patterns are
tandardized. There are, however, some connectors with
s
Make sure power is OFF when connecting or disconnecting any cables in the system.
dditionally, the GFM2 Mass Flow Meter incorporates a precision analog
A
icrocontroller and non-volatile memory that stores all hardware specific variables
m
nd up to 10 different calibration tables. The flow rate can be displayed in 23
ifferent volumetric or mass flow engineering units. Flow meter parameters and
ero adjustment (activated via local button or communication interface), 2
z
rogrammable SPDT relays output, 0 to 5 VDC / 4 to 20 mA analog outputs (jumper
electable), self diagnostic alarm, 36 internal and user defined K-factor. Optional
s
ocal 2x16 LCD readout with adjustable back light provides flow rate and total
l volume reading in currently selected engineering units and diagnostic events indication.
PECIFICATIONS
S
ervice: Clean gases compatible with wetted parts.
S
etted Materials:
W
GFM2-X-X-A: Anodized aluminum, brass, 316 SS fluoroelastomer O-rings; GFM2-X-X-S: 316 SS, and fluoroelastomer O-rings; Buna-N, EPR and PTFE
Accuracy: ±1% FS.
epeatability: ±0.25% FS.
R
esponse Time: 2 seconds to within ±2% of actual flow.
R
utput Signal: Linear 0 to 5 VDC (3000 Ω min. load impedance) and 4 to 20 mA
O
(500 Ω max. loop resistance).
Max. Particulate Size: 5 microns. Temperature Limits: 32 to 122°F (0 to 50°C). Power Supply: 11 to 26 VDC.
rocess Connections: 1/8˝ compression fitting for flow rates ≤ 10 L/min; 1/4˝ for ≤
P
0 L/min; 3/8˝ for ≤ 100 L/min.
isplay: 2 x 16 character LCD.
D Pressure Limits: 500 psig (34.5 bar). Leak Integrity: 1 x 10-9 smL/sec of helium. Weight: 1.05 lb (0.48 kg).
O-rings optional.
The (+) and (-) power inputs are each protected by a 300 mA M (medium time-lag) resettable fuse. If a shorting condition or polarity reversal occurs, the fuse will cut power to the flow transducer circuit. Disconnect the power to the unit, remove the faulty condition, and reconnect the power. The fuse will reset once the faulty condition has been removed. DC Power cable length may not exceed 9.5 feet (3 meters).
Use of the GFM2 flow transducer in a manner other than that specified in this manual or in writing from Dwyer, may impair the protection provided by the equipment.
3. Principle of Operation
The stream of gas entering the Mass Flow transducer is split by shunting a small portion of the flow through a capillary stainless steel sensor tube. The remainder of the gas flows through the primary flow conduit. The geometry of the primary conduit and the sensor tube are designed to ensure laminar flow in each branch. According to principles of fluid dynamics the flow rates of a gas in the two laminar flow conduits are proportional to one another. Therefore, the flow rates measured in the sensor tube are directly proportional to the total flow through the transducer.
In order to sense the flow in the sensor tube, heat flux is introduced at two sections of the sensor tube by means of precision wound heater-sensor coils. Heat is transferred through the thin wall of the sensor tube to the gas flowing inside. As gas flow takes place heat is carried by the gas stream from the upstream coil to the downstream coil windings. The resultant temperature dependent resistance differential is detected by the electronic control circuit. The measured temperature gradient at the sensor windings is linearly proportional to the instantaneous rate of flow taking place.
An output signal is generated that is a function of the amount of heat carried by the gases to indicate mass-molecular based flow rates.
Page 4
low Rates
F
low rates are stated for Nitrogen at STP conditions [i.e. 70°F (21.1°C) at 1 atm].
F
or other gases use the K factor as a multiplier from APPENDIX III.
F
low Rate
odel
M
FM2-X-101
G
odel
M
std liters/min)
(
up to 10
GFM2-X-101 Transmitter
F
Maximum Pressure Drop
)
mm H
(
2
130
eight
W
2.20 lb (1.00 kg)
psid)
(
hipping Weight
S
kPa)
(
.18
.275
3.70 lb (1.68 kg)
Table IV - Pressure Drops
5. OPERATING INSTRUCTIONS
NOTICE
uring the first 6 minutes of the initial powering of the GFM2
D transducer, the status LED will emit a constant amber light.
or the GFM2 transducer with LCD option: If the LCD diagnostic is activated, the
F second line of the LCD will display the time remaining until the end of the warm up period (Minutes:Seconds format) and will alternatively switch to Totalizer reading indication every 2 seconds.
F: 50.0 L/min
** WarmUp 2:39 **
5.1 - Preparation and Warm Up
t is assumed that the Digital Mass Flow Meter has been correctly installed and
I
horoughly leak tested as described in section 2. Make sure the flow source is OFF.
t
hen applying power to a flow meter within the first 2 seconds you will see on the
W LCD display: the product name, the software version, and revision of the EEPROM table (applicable for LCD option only).
S: Ver1.4 Rev.A0
igure b-2: GFM2 First Banner Screen
F
ithin the next 2 seconds, the RS-485 network address, the analog output settings,
W and currently selected gas calibration table will be displayed (applicable for LCD option only).
Ad: 11 Out: 0 to 5 VDC
Gas# 1 AIR
Figure b-3: GFM2 Second Banner Screen
NOTICE
Actual content of the LCD screen may vary depending on the model and device configuration.
After 2 seconds, the LSD display switches to the main screen with the following information:
- Mass Flow reading in current engineering units (upper line).
- Totalizer Volume reading in current volume or mass based engineering units (lower line).
F: 50.0 L/min
T: 75660.5 Ltr
Figure b-4: GFM2 Main Screen
NOTICE
During initial powering of the GFM2 transducer, the flow output signal will be indicating a higher than usual output. This is an indication that the GFM2 transducer has not yet attained its minimum operating temperature. This condition will automatically cancel within a few minutes and the transducer should eventually indicate 0.
Allow the Digital Mass Flow Meter to warm-up for a MINIMUM of 6 minutes.
igure b-5: GFM2 Main Screen During Sensor Warm Up Period.
F
OTICE
N
After 6 minutes of the initial powering of the GFM2 transducer, the status LED will emit a constant GREEN light (normal
operation, ready to measure). For GFM2 with LCD option, the screen will reflect
low and totalizer reading. (see Figure b-4).
f
.2 Swamping Condition
f a flow of more than 10% above the maximum flow rate of the Mass Flow Meter
I is taking place, a condition known as "swamping" may occur. Readings of a "swamped" meter cannot be assumed to be either accurate or linear. Flow must be restored to below 110% of maximum meter range. Once flow rates are lowered to within calibrated range, the swamping condition will end. Operation of the meter
bove 110% of maximum calibrated flow may increase recovery time.
.3 GFM2 Parameters Settings
5.3.1 Engineering Units Settings
The GFM2 Mass Flow Meter is capable of displaying flow rate with 23 different Engineering Units. Digital interface commands (8.3 ASCII Command Set “GFM2 SOFTWARE INTERFACE COMMANDS”) are provided to:
- get currently active Engineering Units
- set desired Engineering Units.
The following Engineering Units are available:
Units of Measure
Number
Index
Flow Rate
Engineering
Units
% mL/sec mL/min mL/hr L/sec L/min L/hr
3
m
/sec
3
m
/min
3
m
/hr
3
ft
/sec
3
ft
/min
3
ft
/hr g/sec g/min g/hr kg/sec kg/min kg/hr Lb/sec Lb/min Lb/hr User
Totalizer
Engineering
Units
%s mL mL mL Ltr Ltr Ltr
3
m
3
m
3
m
3
ft
3
ft
3
ft g g g kg kg kg Lb Lb Lb UD
Description
Percent of full scale Mililiter per second Mililiter per minute Mililiter per hour Liter per second Liter per minute Liter per hour Cubic meter per second Cubic meter per minute Cubic meter per hour Cubic feet per second Cubic feet per minute Cubic feet per hour Grams per second Grams per minute Grams per hour Kilograms per second Kilograms per minute Kilograms per hour Pounds per second Pounds per minute Pounds per hour User Defined
Page 5
NOTICE
automatically.
Once Flow Unit of Measure is changed, the Totalizer’s Volume/Mass based Unit of Measure will be changed
5.3.2 Gas Table Settings
The GFM2 Mass Flow Meter is capable of storing calibration data for up to 10
ifferent gases. Digital interface commands are provided to:
get currently active Gas Table number and Gas name
­ set desired Gas Table.
-
OTICE
N
rdered). If instead of the valid Gas name (for example NITROGEN), the LCD
creen or digital interface displays Gas designator as “Uncalibrated”, then the
s user has chosen the Gas Table which was not calibrated. Using an “Uncalibrated” Gas Table will result in erroneous reading.
5.3.3 Totalizer Settings
The total volume of the gas is calculated by integrating the actual gas flow rate with respect to the time. Digital interface commands are provided to:
reset the totalizer to ZERO
­ start the totalizer at a preset flow
­ assign action at a preset total volume
-
- start/stop (enable/disable) totalizing the flow
- read totalizer via digital interface
The Totalizer has several attributes which may be configured by the user. These
ttributes control the conditions which cause the Totalizer to start integrating the gas
low and also to specify actions to be taken when the Total Volume is outside the
f
pecified limit.
s
OTICE
N
ntered in % FS engineering unit. The Totalizer will not totalize until the flow rate
ecomes equal to or more than the Totalizer Start value. Totalizer Stop values
ust be entered in currently active volume / mass based engineering units. If the
m Totalizer Stop at preset total volume feature is not required, then set Totalizer Stop value to 0.
Totalizer action conditions become true when the totalizer reading and preset "Stop at Total" volumes are equal.
y default the GFM2 is shipped with at least one valid
B
alibration table (unless optional additional calibrations were
c
Before enabling the Totalizer, ensure that all totalizer settings are configured properly. Totalizer Start values have to be
Latch Mode - Controls Latch feature when Relays are assigned to Alarm event. Following settings are available:
- Latch feature is disabled for both relays
- Latch feature is enabled for Relay#1 and disabled for Relay#2
- Latch feature is enabled for Relay#2 and disabled for Relay#1
- Latch feature is enabled for both relays.
NOTICE
NOTICE
onditions. If Relay is assigned to the Alarm event, in some cases, the Alarm
c
atch feature may be desirable.
The current Flow Alarm settings and status are available via digital interface (8.3 ASCII Command Set “GFM2 SOFTWARE INTERFACE COMMANDS”).
.3.5 Relay Assignment Settings
wo sets of dry contact relay outputs are provided to actuate user supplied
T equipment. These are programmable via digital interface such that the relays can be made to switch when a specified event occurs (e.g. when a low or high flow alarm limit is exceeded or when the totalizer reaches a specified value).
he user can configure each Relay action from 6 different options:
T
o Action : (N) No assignment (relay is not assigned to any events and not
N
nergized).
igh Flow Alarm : (H) High Flow Alarm condition.
H
ow Flow Alarm : (L) Low Flow Alarm condition.
5.3.6 K-Factors Settings
onversion factors relative to nitrogen for up to 36 gases are stored in the GFM2
C (see APPENDIX II). In addition, provision is made for a user-defined conversion factor. Conversion factors may be applied to any of the ten gas calibrations via digital interface commands.
f the alarm condition is detected, and the Relay is assigned to
I
larm event, the corresponding Relay will be energized.
A
y default, flow alarm is non-latching. That means the alarm is
B
ndicated only while the monitored value exceeds the specified
i
Local maintenance push button is available for manual Totalizer reset on the field. The maintenance push button is located on the right side of the flow meter inside the maintenance window above the 15 pin D-connector (see Figure C-1 “GFM2 configuration jumpers”).
NOTICE
sequence is recommended:
1. Disconnect GFM2 from the power.
2. Press maintenance push button (do not release).
3. Apply power to the GFM2 while holding down the maintenance push button.
4. Release maintenance push button after 6 seconds. For GFM2 with optional
LCD, when GFM2 Main Screen appears (see Figure b-4).
5.3.4 Flow Alarm Settings
GFM2 provides the user with a flexible alarm/warning system that monitors the Gas Flow for conditions that fall outside configurable limits as well as visual feedback for the user via the status LED and LCD (only for devices with LCD option) or via a Relay contact closure.
The flow alarm has several attributes which may be configured by the user via a digital interface. These attributes control the conditions which cause the alarm to occur and to specify actions to be taken when the flow rate is outside the specified conditions.
Mode Enable /Disable - Allows the user to Enable/Disable Flow Alarm. Low Alarm - The value of the monitored Flow in % FS below which is considered an alarm condition.
NOTICE
High Alarm - The value of the monitored Flow in % FS above which is considered an alarm condition.
NOTICE
Action Delay - The time in seconds that the Flow rate value must remain above the high limit or below the low limit before an alarm condition is indicated. Valid settings are in the range of 0 to 3600 seconds.
In order to locally Reset Totalizer, the reset push button must be pressed during power up sequence. The following
The value of the Low alarm must be less than the value of the High Alarm.
The value of the High alarm must be more than the value of the Low Alarm.
The available K Factor settings are:
• Disabled (K = 1).
• Internal Index The index [0-35] from internal K factor table (see APPENDIX II).
• User Defined User defined conversion factor.
NOTICE
5.3.7 Zero Calibration
The GFM2 includes an auto zero function that, when activated, automatically adjusts the mass flow sensor to read zero. The initial zero adjustment for your GFM2 was performed at the factory. It is not required to perform zero calibration unless the device has zero reading offset with no flow conditions.
NOTICE
condition is established.
Shut off the flow of gas into the Digital Mass Flow Meter. To ensure that no seepage or leak occurs into the meter, it is good practice to temporarily disconnect the gas source. The Auto Zero may be initiated via digital communication interface or locally by pressing the maintenance push button, which is located on the right side of the flow meter inside the maintenance window above the 15 pin D-connector (see Figure C-1 “GFM2 configuration jumpers”).
NOTICE
will prevent initiating Auto Zero function via the maintenance push button before the 6 minutes sensor warm up period has elapsed.
To start Auto Zero locally, press the maintenance push button. The status LED will flash not periodically with the RED light. On the GFM2 with optional LCD, the following screen will appear:
The conversion factors will not be applied for % FS engineering unit.
Before performing Zero Calibration, make sure the device is powered up for at least 15 minutes and absolutely no flow
The same maintenance push button is used for Auto Zero initiation and Totalizer reset. The internal diagnostic algorithm
AUTOZERO IS ON!
Figure b-6: GFM2 Screen in the Beginning of Auto Zero Procedure
Page 6
Loading...
+ 14 hidden pages