Durr Dental XR 24 Pro 1700-08 Installation And Operating Instructions Manual

EN
9000-600-74/30

2018/02SE
Dürr Dental X-ray Film Developer XR 24 Pro
Installation and Operating Instructions
EN
2 2018/02SE2
Content
Important Information
1. General . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 Note on Conformity . . . . . . . . . . . . . 4
1.2 General Notes . . . . . . . . . . . . . . . . . 4
1.3 Notes on medical products . . . . . . . . 4
1.4 Intended purpose . . . . . . . . . . . . . . . 4
1.5 Notes on EMC and medical prod-
ucts . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.6 Correct Usage . . . . . . . . . . . . . . . . . 5
1.7 Incorrect usage . . . . . . . . . . . . . . . . 5
1.8 Appliance disposal . . . . . . . . . . . . . . 5
1.9 Disposal of Chemicals . . . . . . . . . . . 5
2. Safety . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1 General Safety Notes . . . . . . . . . . . . 5
2.2 Electrical safety instructions . . . . . . . 6
3. Warnings and Symbols . . . . . . . . . . . . . 6
3.1 Model identification plate . . . . . . . . . 6
4. Delivery Contents . . . . . . . . . . . . . . . . . 7
4.1 Special accessories . . . . . . . . . . . . . 7
4.2 Disposable materials . . . . . . . . . . . . 7
5. Technical Data . . . . . . . . . . . . . . . . . . . 7
5.1 XR 24 Pro, 1700-08 . . . . . . . . . . . . . 7
5.2 Environmental conditions . . . . . . . . . 7
6. XR 24 Pro functional layout . . . . . . . . . 8
7. XR 24 Pro functional description . . . . . 9
7.1 PCB (main board) layout . . . . . . . . . . 9
Mounting
8. Set-up . . . . . . . . . . . . . . . . . . . . . . . . . 10
8.1 Room for set-up . . . . . . . . . . . . . . . 10
8.2 Set-up alternatives . . . . . . . . . . . . . 11
8.3 Set-up . . . . . . . . . . . . . . . . . . . . . . 11
9. Installation . . . . . . . . . . . . . . . . . . . . . 12
9.1 Water connections . . . . . . . . . . . . . 12
9.2 Waste water connections . . . . . . . . 12
9.3 Waste water hose connection
from developer and fixer . . . . . . . . . 13
10. Start-up and first use . . . . . . . . . . . . . 14
10.1 Clean unit . . . . . . . . . . . . . . . . . . . 14
11. Electrical connection . . . . . . . . . . . . . 15
12. Circuit diagram . . . . . . . . . . . . . . . . . . 16
12.1 PCB (main board) operating func-
tions . . . . . . . . . . . . . . . . . . . . . . . 16
12.2 PCB (main board) control func-
tions . . . . . . . . . . . . . . . . . . . . . . . 16
12.3 PCB (main board)
power functions 230V . . . . . . . . . 17
13. Start-up and first use . . . . . . . . . . . . . 18
13.1 Settings / service operation for
the Service Technician . . . . . . . . . . 19
13.2 Check temperature of the devel-
oper and carry out commission-
ing test . . . . . . . . . . . . . . . . . . . . . 20
14. Transport . . . . . . . . . . . . . . . . . . . . . . . 21
14.1 Remove drainage hoses . . . . . . . . 21
14.2 Protection from damage . . . . . . . . 21
EN
2018/02SE 3
Use
14.3 Program overview . . . . . . . . . . . . . 22
14.4 Service / Fault message . . . . . . . . 22
15. Operation . . . . . . . . . . . . . . . . . . . . . . . 22
15.1 In the morning or before surgery
start . . . . . . . . . . . . . . . . . . . . . . . 23
15.2 To change program: . . . . . . . . . . . 24
15.3 In the evening or after surgery
closing . . . . . . . . . . . . . . . . . . . . . 24
15.4 Surgery or clinic closures . . . . . . . . 24
15.5 Film recommendation list for Dürr Dental Developer
XR 24 ... . . . . . . . . . . . . . . . . . . . . 25
16. Maintenance . . . . . . . . . . . . . . . . . . . . 26
16.1 Change the chemicals . . . . . . . . . . 27
16.2 Every 2 months cleaning tanks
and roller sets . . . . . . . . . . . . . . . . 30
Trouble-shooting
17. Tips for Service Technicians . . . . . . . . 33
17.1 Fault message on display panel
with audible signal . . . . . . . . . . . . . 34
18. Tips for Service Technicians . . . . . . . . 35
19. Tips for Trouble-shooting . . . . . . . . . . 36
19.1 Before starting . . . . . . . . . . . . . . . 36
19.2 Remove front panel . . . . . . . . . . . . 36
19.3 Remove rear panel . . . . . . . . . . . . 36
19.4 Remove side panels . . . . . . . . . . . 36
19.5 After completing work . . . . . . . . . . 36
19.6 Service mode plan . . . . . . . . . . . . 37
19.7 Changing display . . . . . . . . . . . . . . 39
19.8 Check unit fusing . . . . . . . . . . . . . 39
19.9 Changing the process heater and
PTC-sensor . . . . . . . . . . . . . . . . . . 40
EN
4 2018/02SE
1. General
1.1 Note on Conformity
This product has been tested for conformity ac­cording to the appropriate directives as laid down by the European Union and has been found to conform to all the required specifications.
1.2 General Notes
• These Installation and Operating Instructions form an integral part of the unit. They must be kept close to the unit at all times. Precise ob­servance of these instructions is a precondi­tion for use of the unit for the intended pur­pose and for its correct operation. New per­sonnel must be made aware of the contents, and they should be passed on to future oper­ating staff.
• Safety for the operator as well as trouble-free operation of the unit are only ensured if use is made of original equipment parts. In addition, only those accessories expressly recommend­ed in the Installation and Operating Instructions or explicitly recommended by Dürr Dental may be used with this appliance. If other accessories are used with this appli­ance, Dürr Dental cannot guarantee safe op­eration or proper functioning. No liability on the part of the manufacture will be accepted in the case that damage arises through the use of non-approved accessories.
• Dürr Dental are only responsible for the equip­ment with regard to safety, reliability and proper functioning where assembly, reset­tings, changes or modifications, extensions and repairs have been carried out by Dürr Dental or an agency authorized by Dürr Dental and if the equipment is used in conformity with the Installation and Operating Instructions.
• These Installation and Operating Instructions conform to the relevant version of the equip­ment and the underlying safety standards val­id at the time of going to press. All circuits, processes, names, software and appliances quoted are protected under industrial property rights.
• The translation of these Installation and Operating Instructions has been carried out in good faith. Dürr Dental will not accept any claims arising from incorrect translation. The enclosed German version of these Installation and Operating Instructions is the final refer­ence - in case of doubt contact the manufac­turer or their agent.
• Any reprinting of the technical documentation, in whole or in part, is subject to prior approval of Dürr Dental being given in writing.
• Retain the packaging for possible return of the product to the manufacturers. Only the origi­nal packaging provides adequate protection during transport of the unit. Should return of the product to the manufacturers be neces­sary during the guarantee period, Dürr Dental accepts no responsibility for damage occur­ring during transport where the original pack­aging was not used! Ensure that the packag­ing is kept out of the reach of children.
1.3 Notes on medical products
• This product is a technical medical appliance and, as such, may only be operated by such persons who, as a result of training or experi­ence, can be confidently expected to operate it correctly and safely for patients and staff alike.
• Multi-socket power outlet extensions may not be placed on the floor. Where multi-socket power strips are used then the requirements under EN 60601-1-1 must be strictly observed.
• Other appliances should not be connected to the same multi-socket power outlet extensions.
1.4 Intended purpose
The unit is intended solely for the development of extraoral and intraoral X-ray film in dental ap­plications in accordance with the Dürr Dental Film Recommendation List. The machine con­tains a roller transport system that transports the film after exposure with X-ray beams through the developer, fixer and water baths and then subsequently through the drying sections.
Important Information
EN
2018/02SE 5
1.5 Notes on EMC and medical products
Medical products should be treated with respect when it comes to electro-magnetic compatibility and special safety measures must be taken. Notes on electro-magnetic compatibility of products for medical applications can be found in our brochure number 9000-606-67/30 or the information can be found on the internet at www.duerr.de in the Download section for Technical Documentation
1.6 Correct Usage
The unit is intended solely for the development of extraoral and intraoral X-ray film in dental ap­plications in accordance with the Dürr Dental Film Recommendation List. The machine con­tains a roller transport system that transports the film after exposure with X-ray beams through the developer, fixer and water baths and then subsequently through the drying sections. For operation of the unit, only use parts, acces­sories and optional accessories that have been specified or approved by Dürr Dental. For cleaning and disinfection of the device, only use disinfectants and cleaning agents that have been specified or approved by Dürr Dental. All legal requirements that apply at the site of use in relation to health and safety in the workplace and to the disposal of chemicals must be com­plied with.
1.7 Incorrect usage
Any use of this appliance/these appliances above and beyond that laid down in the Installation and Operating Instructions is deemed to be incorrect usage. The manufactur­er cannot be held liable for any damage result­ing from incorrect usage. The operator will be held liable and bears all risks. This unit is not suitable for monitoring patients over longer periods of time. The unit is not desi­gned for operation within the patient environment. Only photographic chemicals that are coordina­ted with the development process may be used. The waste water drainage line must have a downward gradient. The developer and fixer must be disposed of separately as special was­te in suitable containers in accordance with nati­onal regulations and guidelines.
Due to the bath temperature that needs to be maintained for the process, optimum developing quality can only be ensured up to an ambient temperature = bath temperature. The unit must not be used in operating theatres or similar rooms in which potential hazards could arise from the ignition of flammable mixtures.
1.8 Appliance disposal
• EU directive 2012/19/EU (WEEE / Waste Electric and Electronic Equipment) dated 4 July 2012 and its current application in national law clearly states that dental products are covered by the above-mentioned directive and, as such, are covered by special waste disposal require­ments within the European Union.
• If you have any questions regarding the cor­rect waste disposal procedures for this prod­uct please contact Dürr Dental or your dental supplier who will be pleased to assist.
1.9 Disposal of Chemicals
The empty concentrate bottles of developer and fixer must be disposed of according to national and local regulations. Quantities of used devel­oper and fixer solutions, as well as any cleaning fluids, should be disposed of according to local rules and regulations concerning waste water.
2. Safety
2.1 General Safety Notes
This appliance has been designed and con­structed by Dürr Dental that any danger arising from correct usage is virtually impossible. In spite of this, we feel it is our duty to mention the following safety measures in order to prevent any possible danger.
• When using this appliance all local and rele­vant regulations must be observed! Converting or modifying the appliance in any way is strictly prohibited. In such cases, any and all guarantees immediately become invalid. The operation of modified appliances can be punishable by law. In the interests of trouble-free operation the operator is respon­sible for observing these regulations.
• Installation must be carried out by a technical expert.
• Before every use the operator must check the functional safety and the condition of the appliance.
EN
6 2018/02SE
3. Warnings and Symbols
In the operating instructions the following warn­ings and symbols have been used:
Information and/or mandatory re-
gulations or prohibitions for the prevention of personal injury or sub­stantial property damage
Information and/or instructions or prohi-
bitions regarding personal safety or ex­tensive material damage.
Warning - dangerous electrical
voltage.
• The operator must be knowledgeable in the operation of the appliance.
• The product is not designed to be used in medical treatment areas where there exists the danger of explosion. Areas where explo­sions could occur are those where flammable anesthetic material, skin cleansers, oxygen and skin disinfectants are present. This appli­ance is not to be used in areas where the at­mosphere could cause fire.
2.2 Electrical safety instructions
• The appliance may only be operated using a correctly installed power socket.
• Before connecting to the electricity supply the appliance must be inspected and checked that the supply voltage and the supply fre­quency correspond to that of the local electri­cal supply.
• Before initial use and start-up the appliance and all supply lines must be checked for any signs of damage. Damaged supply lines and connections must be replaced immediately.
• Never come into contact with patients and open plug-in connections on the appliance at the same time.
• When using the appliance observe all the rele­vant electrical safety procedures.
• Only use the supplied mains cable to connect the device.
Power switch ON / OFF
/ Select parameters
Confirm selection
Switch on power
Switch off power
Each morning - open the water tap
Each evening - turn water tap off
28 x 24 h
replace chemicals every 4 weeks
Drain off the chemicals
Clean unit
3.1 Model identification plate
Order no. / model no.
Serial-No.
Observe the operating instructions
CE-Labeling without Notified Body
Number
Manufacturer
Fuses in accordance with IEC 60127-2
Health Inustry Bar Code (HIBC)
To be disposed of properly in accord-
ance with EU directive 2012/19/EU (WEEE)
EN
2018/02SE 7
4. Delivery Contents
X-ray film developer XR 24 Pro, model 1700-08 (230 V AC, 50/60 Hz)
accessories, complete . . . . . . . . . 1700-001-00
4.1 Special accessories
The parts listed as special accessories
are not part of the standard scope of delivery but can be ordered separately.
DÜRR Cleaning film, clear x 100 . . 1410-007-00
Water hose with Aqua-Stop . . . . . 1330-001-51
Regeneration unit (230 V AC) . . . . 1700-820-00
Large collector, 20 liter
for developer. . . . . . . . . . . . . . . . . 1416-021-00
Large collector, 20 liter
for developer. . . . . . . . . . . . . . . . . 1416-011-00
Preparation container E with lid . . . 1330-015-00 Preparation container F with lid . . . 1330-016-00 Collecting system for developer . . . 1700-062-00
Collecting system for fixer . . . . . . . 1700-061-00
Daylight cover DL 24
with filter disc . . . . . . . . . . . . . . . . 1700-910-00
suitable for film sizes 2 x 3 cm to max. 15 x 30 cm
Daylight cover DL 26 . . . . . . . . . . . 1700-920-00
suitable for conventional film sizes to max. 24 x 30 cm
Darkroom adapter DA 24 . . . . . . . 1700-930-00
suitable for film sizes: 2 x 3 cm to max. 24 x 30 cm
Electrical fuse,
230 V AC, T 6.3 AH . . . . . . . . . . . 9000-115-25
4.2 Disposable materials
DÜRR-AUTOMAT XR/C (with vitamin C) developer and fixer concentrated liq­uids
for 2 applications . . . . . . . . . . . CXB160C9940
Spray cleaner WR 2000, for transport roller sets, 6 x Pack . . CCB800C5540
Cleaning film Dürr Clear 100,
100 pieces . . . . . . . . . . . . . . . . . . 1410-007-00
XR Clean Set . . . . . . . . . . . . . . . . 1700-007-00
5. Technical Data
5.1 XR 24 Pro, 1700-08
Model 1700 -08 Voltage V 230 Current A 4,0 Electrical
frequency
Hz 50-60
Power output W 950 Heat output
Bath heating Drying W
400 450
G-fuse link IEC60127-2/V
A T6.3 AH
Warm-up time min ~20
(0.5 °C/min.)
Film processing time
min 3,25 - 20
Bath circulation process
Developer Fixer (l/min)
~1,8
~2
Water circulation rate
(l/min) ~2
Water pressure
min. water pressure max. water pressure
bar
2 6
Tank volume
developer, fixer (l) 5 each
Fuse type IP 20 Protection class I Surge voltage
category
II
Medical device Class I Duty cycle ED % 100 Weight kg 25 Dimensions
(DxWxH)
cm 44 x 42 x 51
5.2 Environmental conditions
Storage and transport conditions
Temperature (°C) . . . . . . . . . . . . . . . . -10 to +60
Rel. humidity (%) . . . . . . . . . . . . . . . . . . max. 95
Operating conditions
Temperature (°C) . . . . . . . . . . . . . . . +10 to +25
Rel. humidity (%) . . . . . . . . . . . . . . . . . . max. 70
EN
8 2018/02SE
6. XR 24 Pro functional layout
1 Film feed 2 Start flap 3 Roller set - film feed 4 Drive shaft
5 Roller sets developer/fixer (D/F) 6 Roller set - water 7 Water inflow 8 Roller set - dryer
9 Safety switch “dryer” 10 Dryer heating 11 Fan 12 Drive motor
12a Speed sensor 13 Fluid level indicator sensor “Water”
14 Rotary pump (for D and F) 15 Water intake valve 16 Circulation heater
17 PTC-sensing device
18 Overflow pipe F with drainage plug 19 Lifter 19a Light barrier for lifting bath 20 Lift motor 21 Overflow pipe D with drainage plug 22 Lifting bath D Developer bath F Fixer bath W Water bath A Power switch
on main display
Stand-by Position
2 A 3 4 5 6 7
20 19 19a 18 17 16
8
9 10
11
12
13
14
15
22
21
1
F
W
E
12a
EN
2018/02SE 9
7. XR 24 Pro functional description
Press the power switch (min. 2 secs)
and the display lights up. The warming-up pha­se for the developer and fixer baths begins. The necessary waiting time required before the re­quired temperature has been reached is shown: "WAITING . . MIN" (display panel blinks). The unit will not start during this warm-up stage.
The bath circulation process starts parallel to the bath heating. During the circulation step
the rotatory pump (14) feeds the chemicals from the developer and fixer baths (tank volume of each is 5 liters) to the lifting bath (22).
Once the set temperature has been reached the display will show "MACHINE
EMPTY", i.e. the unit is now ready for use. As soon as the set temperature falls by more than 0.5 °C the process heater and rotary pump (14) will start up again.
When a film is inserted, the film feed flap (2) opens and the display will show a blinking mes­sage: "MACHINE LOADED". At the same time a signal is triggered to lower the lifting bath (22), the water intake valve (15) is opened, and the dryer heater (10), drive motor (12) and the rotatory pump (14) (for developer and fixer bath circulation) all start up. The processing time can be checked using the speed sensor (12a). The raising and lowe­ring of the lifting bath is controlled by the rotati­on of the disc (driven by the lift motor (20)) and is monitored by the infrared light sensor (19a).
When the lifting bath (22) is lowered the film is transported between the rollers and through the developer and fixer baths, and then through the rinsing and drying zones (8) in the preset pro­cessing time.
During the film processing, water is released via the water inflow (7) into the rinsing zone. The water is drained off via the waste water system. There is a continuous flow of water of ca. 2 l/ min, at 2 bar. If the waste water drainage gets blocked, then the level indicator sensor (13) sends an audible signal, the water intake valve (15) is closed and the water inflow is halted.
Once the film has been taken up the film feed flap (2) closes. As soon as the blinking mes­sage "MACHINE LOADED" on the display panel is extinguished, then a further film or 6 intraoral films side-by-side can be fed in.
If, after the film processing stage (message is audible signal), no further film is inserted, then the lifting bath (22) is raised (stand-by position "MACHINE EMPTY"). The drive motor (12), wa­ter intake valve (15) and dryer heater (10) will then be switched off. In the stand-by position the developer and fixer roller set (5) is complete­ly submerged in the chemicals. This serves to prevent the majority of any deposits attaching themselves to the rollers.
7.1 PCB (main board) layout
23 A Operating parts 23 B Control functions 23 C Power functions
23 A
23 B 23 C
EN
10 2018/02SE
8. Set-up
8.1 Room for set-up
• The XR 24 Pro developer may only be operat­ed in a clean, dry and well-ventilated room.
• Room temperature must not fall below +10 °C in winter and must not exceed +28 °C in summer.
• The room chosen for set-up must have ac­cess to a water connection (water tap) (27) R3/4" with pipe anti-vacuum device, waste water drainage (28) with aeration (anti-odour) via plastic siphon and an electrical safety socket (26). We strongly recommend that a wash basin be available in the vicinity of the developer.
The water tap (27) and electrical safety
socket (26) must be easily accessible to the operating personnel.
• We strongly recommend that set-up of the XR 24 Pro is effected in a darkroom. If the ap­pliance cannot be set up in a darkroom, then a daylight cover (24) (DL 24 or DL 26) must be fitted.
• When operating an appliance with daylight cover (24) please note:
Never expose the machine to direct
sunlight! Danger of the chemicals overheating or pre-exposure of film materials. Never store film material in the vicinity of the X-ray room!
Once film material is inserted into the
film depot (25) of the daylight cover the­re is the danger that pre-exposure can occur due to stray ambient radiation from the X-ray appliance.
• If necessary, it is possible to set up a separat­ing wall or barrier using lead protection.
27
26
28
24
25
Mounting
1
2
3
EN
2018/02SE 11
24
8.2 Set-up alternatives
• Place the developer unit onto a horizontal, stable, flat, and single-piece working surface at a comfortable working height.
• In order to have access to the hoses the working surface should have at least one opening of ca. 7 x 10 cm.
• Take into account that additional space may be required for service and maintenance and that sufficient ventilation is available for the unit: to the rear leave at least 10 cm, left and right allow ca. 50 cm free space
• Leave enough space at the front in case a daylight cover (24) is to be used. If necessary, unscrew the adjustable feet from the front and apply to the rear.
8.3 Set-up
• First remove the plastic tapes to the packaging and carefully lift the box upwards.
• Remove all small parts.
• Take the unit from the sides and place as
required.
• Set up the unit in the vicinity of water supply, waste water system and electrical safety socket (max. distance 1.5 m)
• Set up the appliance so that it has a slight tilt, ca. 0.3 ° to the fixer bath (fig. 6). Adjust the feet of the appliance (29) using the spanner provided (30) as required.
29
30
4
5
EN
12 2018/02SE
9. Installation
9.1 Water connections
• Check that clean, clear water flows from the water tap (27).
When installing for the first time make sure the connection is allowed to flow and rinse first!
• Because of great regional differences in the water quality a fine filter (32) must be fitted be­tween the water tap (27) and water hose (34).
The fine filter (32) protects the flow regu-
lator in the water intake valve of the ap­pliance from becoming dirty.
• Turn off the water tap (27).
• Connect the 90° elbow piece (33) to the water
tap (27), mounting the fine filter (32) (arrow in direction of flow) and the double nipple piece (31) at the same time.
Check that the coarse filter (34a) and
both seals (34b) are present in the threaded fitting of the water hose (34).
• Connect the water hose (34) to the water in­flow (35) of the appliance (threaded pipe fitting with coarse filter) and connect to double nip­ple piece (31).
9.2 Waste water connections
• The waste water connection must include a siphon trap.
• Lay the waste water hose (36) with a steady incline to the siphon trap: Shorten the hose to an appropriate length
34
35
36
27
31
32 33
34
34a
34b
6
7
8
EN
2018/02SE 13
40
41
39
43
41
39
Fixer
max. 3 cm
37 38
25
When laying out the waste water
hoses the following must be careful­ly observed:
- avoid sagging hoses! Avoid squeezing or reducing the pressure on hose!
• The external ends of the hoses should be coated with UHU-Plast.
• Screw the screw socket (37) onto the hose and fit to siphon trap muff (25) using a hose clamp (38).
• Secure the waste water hose using hose clamp, screws and rawlplugs.
9.3 Waste water hose connection
from developer and fixer
• Place the 10l-collector tanks (39) for develop­er (black level indicator marker) and fixer (red level indicator marker) below the appliance in a clearly visible position and free from the possibility of tilting.
• Lay hoses (40) with constant incline
• Shorten the hoses to the correct length.
• Feed the hose for developer (black markings)
into the black cover (41).
• Screw the cover (41) of the developer collec­tor (39) on tightly.
• Fix the hose for fixer (red markings) in the same way to the fixer collector.
Empty the collector vessel when the
level indicator marker (43) shows full - danger of overspill! Disposal of the developer and fixer requires observance of all regional rules and regulations! In Germany and many other countries, x-ray chemicals must be disposed of separate­ly as special waste: check local regulations.
9
10
11
EN
14 2018/02SE
44
45
8
8
6 5
3
90
90
10. Start-up and first use
10.1 Clean unit
• Remove the cover.
• Remove the floating covers to developer and
fixer baths.
• Open the green catch (44) and locking devices (45).
• Swing the dryer roller set (8) towards the back.
• Extract the roller sets for water (6) and devel­oper / fixer (5).
• Swing the display (90) upwards.
• The display can also be lifted upwards and
then laid to the side on the left.
• Extract the roller set film insertion (3).
12
13
14
EN
2018/02SE 15
61
22
11. Electrical connection
The unit is set up for a supply volta-
ge of 230 V (see model identification plate). Before connection of the main power cable it is important to check whether the required voltage is available, otherwise the appli­ance can be damaged. Use the Dürr mains cable supplied. The unit must only be used in completely closed state - with all side panels and co­ver fitted and closed - before con­necting to the mains power supply.
• Connect the power plug (61) into the appli­ance and into the main power socket.
• Press the lever to the side and remove the lift­ing bath (22).
• Clean the roller sets, lifting bath and tanks us­ing a damp sponge.
"Preparation of chemicals " and "Insertion of roller sets" see sections 16.1.3 and 16.1.4
15
16
EN
16 2018/02SE
12. Circuit diagram
12.1 PCB (main board) operating functions
12.2 PCB (main board) control functions
(see also parts 23 A and 23 B under section 7)
U1 Light barrier to cover
U2 Light barrier to cover
U3 Light barrier lifting bath
U4 Light barrier to film insertion slot
R1 Temperature sensor "Bath temperature"
X5 Level indicator sensor "Water"
M1 Speed sensor
12.1 12.2
23 B
23 A
EN
2018/02SE 17
M
1 Roller drive motor
M2 Ventilation fan
M3 Rotary pump
M4 Lifting bath motor
R1 Bath heating
R2 Dryer heater
Y1 Water intake valve
X1/X2 Power supply
X3 Regeneration unit
12.3 PCB (main board) power functions 230V
(see also part 23 C under section 7)
12.3
23 C
EN
18 2018/02SE
13. Start-up and first use
The developer unit XR 24 Pro has 3 different settings for the processing time, and is factory set at the following values.
Intraoral films 6 min (program INTRA)
Extraoral films 4 min (program EXTRA)
Endodontal films 3:15 min (program ENDO)
Please note: X-ray films with such short proces­sing times are not suitable for archiving.
-Developer / fixer bath temperatures 28°C
-Dryer performance 30%
When using other film types requiring different processing times and dryer performance the the values will need to be changed as appropriate. Refer to the film list, section 15.7.
For a program overview refer to the section "Usage"
The unit must only be used in com-
pletely closed state - with all side panels and cover fitted and closed
- before connecting to the mains power supply.
Every time the unit is switched on at the
power switch then program "INTRA" is started, i.e. after the unit is switched off there is an automatic reset to "INTRA".
Guidance on using the unit and the
display for the operator can be found in section "Usage", section 15.
A complete description of the service
operation can be found on the section covering Maintenance 19.6.
The developer unit must never be
operated when the roller set is not present!
Where the roller sets are not inserted then there is pressure being exerted vertically upwards on the developer and fixer fluids and could splash out.
Chemicals irritate eyes and skin and there is the danger of serious eye damage.
Without developer and fixer roller sets there can
be no return of chemicals to the ap­propriate containers.
There is also the danger of the chemicals mixing.
The developer unit must never be operated
without fluids (chemicals, water) as this can lead to the heating breaking down.
EN
2018/02SE 19
13.1 Settings / service operation for the Service Technician
• Open the water tap.
• Check unit and all connections for signs of
leakages
• Start service operation
press + simultaneously
(min. 4 secs) unit is now in Service mode
• Select service function
Use key
and Service Function to
select
• Change settings
with
or change parameter values
using
1 s, confirm new values
• *1 The selected Language, Frequency or
Value is immediately displayed on restart
*2 Factory settings
Change values using and in 5 and 5% steps, confirm with
• End Service operation Switch off unit at power switch (press
min. 2 secs)
• Set parameters e. g. Language, Electrical
frequency, etc. at the display using pro­cess steps:
EN
20 2018/02SE
13.2 Check temperature of the de­veloper and carry out commis­sioning test
The following procedure is shown as an
example for use in Germany. When al­tering for other countries take local situ­ations (e.g. regulations) into account!
• Switch on the unit and wait for the warm-
ing-up phase of the developer bath to finish, see also "15.1 In the morning or before sur­gery start"
• To check the temperature of the developer,
switch off the unit.
• Unplug at the mains and remove all power.
• Close the water tap.
• Remove the cover.
• Measure the temperature of the developer
(front left to a depth of ca. 20 cm). The temperature of the developer must agree with temperatures given in the film list, e.g. +28 °C +/- 0.5 °C.
• Record temperature in Acceptance Log and
Constancy Test Records.
• Replace cover, plug in at the mains, open wa-
ter tap, use
to switch on unit.
• Insert test film and carry out Acceptance Test
according to x-ray regulations. Record test re­sults in test report. - Archive the test film. Inform operating staff that regular constancy checks must be carried out.
17
EN
2018/02SE 21
31
39
63
35
27
39
14. Transport
Before transport of the developer please note the following points:
The developer unit may only be
transported with completely empty tanks.
• Switch off unit at main switch.
• Turn off the water tap (27).
• Remove power connection between mains
cable and from the electrical safety socket.
• Unscrew the water hose (31) from the unit and from the water tap (27).
• Empty the chemicals from the collector vessel (39) into the appropriate canister (63).
• Let the chemicals flow from the tanks into col­lector vessels (39), see section 16.1
• Clean the developer unit, see section 10.1
14.1 Remove drainage hoses
• Remove the drainage hose (35).
• Loosen the hose from the collector vessel (39)
for developer and fixer, being careful to avoid any drops of chemicals leaking out.
14.2 Protection from damage
• Protect the developer from any mechanical pressure and damage.
• Ensure that the appliance during transport cannot move within the packaging.
• Inform transport personnel.
18
19
20
EN
22 2018/02SE
14.3 Program overview
The programs Intra, Extra and Endo are facto­ry set (see below); however, they can be indivi­dually altered by the Service Technician accor­ding to the film list.
INTRA - Process time: t=6min
- bath-temperature: T=28 °C
- dryer performance: 30%
EXTRA - Process time: t=4min
- bath-temperature: T=28 °C
- dryer performance: 30%
ENDO - Process time: t=3min 15 sec
- bath-temperature: T=28 °C
- dryer performance: 30%
Every time the unit is switched on at the
power switch then program "INTRA" is started, i.e. after the unit is switched off there is an automatic reset to "INTRA".
14.4 Service / Fault message
See section 17.1 Fault messages on display with audible signal
15. Operation
Before operating the appliance please note the following points:
The developer unit must never be
operated when the roller set is not present!
Where the roller sets are not inserted then there is pressure being exerted vertically upwards on the developer and fixer fluids and could splash out.
Chemicals irritate eyes and skin and there is the danger of serious eye damage.
Without developer and fixer roller sets there can
be no return of chemicals to the ap­propriate containers.
There is also the danger of the chemicals mixing.
The developer unit must never be ope-
rated without fluids (chemicals, water) as this can lead to the hea­ting breaking down. Set heating controls.
In order to ensure constant image
quality the unit should be switched on daily.
Also note that the ambient outside temperature should not be higher than the temperature of the developer bath, see section 5.2 Environmental conditions.
Stand-by Mode
Once the film has left the unit it switches to Stand-by Mode. In Stand-by Mode the unit is ready for use and is waiting for insertion of the next film material. Before inserting the film, check the program selection!
Only when activated by the Service Technician!
SERVICE INTERVAL
REACHED!
ERROR WATER TOO
LOW!
ERROR
ROLLER DRIVE
ERROR WATER TOO
HIGH!
BEFORE OPENING
SWITCH OFF AT MAINS
Use
Power - switch
Confirmation - key
, "Browse menu"
EN
2018/02SE 23
15.1 In the morning or before sur­gery start
• Open the water tap.
• Switch on unit:
•
Press power switch ca. 2 s until
audible signal is heard
>> DURR DENTAL<<
* XR 24 PRO V.. *
WARM-UP PHASE
WAITING .. min
<INTRA> 6:00
MACHINE EMPTY
<INTRA> 6:00
MACHINE LOADED
<INTRA> 6:00 REMAIN ..MIN
<INTRA> 6:00
MACHINE LOADED
<INTRA> 6:00
MACHINE EMPTY
To change program:
see next page
1 Display LEDs light up. Lifting bath moves au-
tomatically into the correct position, ca. 30 seconds.
2 The unit is in the warming-up phase until the
correct temperature in the developer bath is reached. Warm-up time: 0.5 °C/min. During the warm-up stage the display blinks with message "WAITING .. MIN"
3 As soon as the set temperature has been re-
ached the display is lit.
The unit is ready for use. Film material can be inserted.
INTRA is selected as default program; to
change, press key
and select EXTRA or
ENDO.
4 Once a film has been inserted, the display
blinks: "MACHINE LOADED", as soon as
the message stops and stops blinking, the next film can be inserted.
Up to six intraoral films side-by-side
within the markings can be inserted si­multaneously, see next page.
5 The remaining time before the film is ejected
is displayed (automatic reset). An audible signal is heard when film is ejected.
6 If no further film is inserted the lifting bath re-
turns to "Stand-by position"
7 When the unit is in "Stand-by" mode the dis-
played message is: "MACHINE EMPTY"
Non-exposed or developed film
must not be left on or near the unit, as the LCD-display panel could lead to additional exposure.
EN
24 2018/02SE
15.2 To change program:
• 6 Up to six intraoral films can be laid next to each other between the markings and insert­ed. When inserting 2x3 films take care that they lie correctly and the marking is complete­ly fed into the developer unit all the way.
• For rapid test pictures (e.g. when carrying out root canal-treatment) change to program "ENDO" (shorter process time: 3:15 min).
• Please also refer to the list of recommended films in section 15.4!
Whe creating test pictures (e.g. for root canal-treatment) under program "ENDO", please note the following:
X-ray films developed at this shorter
processing time (3:15 min) cannot be archived. If the film is to be stored in the surgery archive, immediately following diagnosis ensure that the film is fixed again for a minimum of 30 seconds and then washed.
15.3 In the evening or after surgery
closing
• Switch off unit
•
Close the water tap
15.4 Surgery or clinic closures
of more than two weeks:
• Drain of chemicals, see section 16.1.1
• Clean tanks and roller sets thoroughly and
leave to dry, see section 16.3,
For error messages, see section
17.1 "Fault messages on display
21
EN
2018/02SE 25
15.5 Film recommendation list for Dürr Dental Developer XR 24 ...
Intraoral Films
Bath temperature 28 °C Process time 6 min. Dryer performance 30% Film types
Kodak Ultra Speed 2.2 x 3.5 cm
3.1 x 4.1 cm
2.7 x 5.4 cm
5.7 x 7.6 cm
Kodak Insight 2.2 x 3.5 cm
3.1 x 4.1 cm
2.4 x 4.0 cm
2.7 x 5.4 cm
5.7 x 7.6 cm
Agfa Dentus M2
Comfort
2 x 3 cm
3 x 4 cm 5 x 7 cm
Extraoral Films
Bath temperature 28 °C Film formats to max. 24 x 30 cm Film types Process time Dryer
capacity Kodak
T-Mat L/RA 4min 20 % T-Mat G/RA 4min 20 % Ektavision L 6min 30 % Ektavision G 6min 30 %
Agfa
Dentus Ortholux 4 min 20 %
For other film types check the "Dental
Sensitivity Index" and "Dental Contrast Index" for given process times and alter parameters as necessary.
EN
26 2018/02SE
16. Maintenance
Every week
Use a cleaning film
• Once a week before beginning work use a cleaning film, e.g. "Dürr Clear 100".
• Carry out a Constancy Check according to X-ray regulations:
Every month
Change the chemicals, where no regeneration
unit (91) is fitted, see section 16.1
This interval is recommended for a daily
film usage of 10 intraoral and 3 extraoral films
Every 2 months
Clean tanka and roller sets, see section 16.3 See also the leaflet enclosed with the x-ray film chemicals, Order no. 9000-600-91/01
Every 3 months
Change the chemicals, where regeneration unit (91) is fitted, see section 16.1
Annually
Check the darkroom for its light sealing quality.
Maintenance intervals
Every week
Every
month
Every
2 months
Every
3 months
Annually
Use a cleaning film
e.g. Dürr Clear 100
X
Constancy Check
according to x-ray regulations
X
Change chemicals
• without regeneration
unit
• with regeneration unit
X
X
Cleaning
• Developer tank and roller sets
X
• Check darkroom
• light conditions
X
EN
2018/02SE 27
5
8
6
3
39
63
60
61
91
16.1 Change the chemicals
16.1.1 Drain off the chemicals
• Switch off unit at main switch.
• Unplug (61a) from mains.
• Remove the cover (60).
• Check whether collector vessel (39) is empty,
if necessary empty into the large collector canister (63).
It is extremely important that the collec-
tor vessels are empty, otherwise there is the danger of fluids overflowing when the chemicals are now drained from the tanks.
• Remove the two floating covers from the de­veloper and fixer baths, place to one side and allow to dry.
• First unscrew the smaller plug from the lifting bath, then the larger plugs from the tanks.
• Empty collector vessel.
• Undo the cover to collector vessels, let the
ends of the hoses hang in a neutral container.
This serves to prevent water from get-
ting into the collector vessel or canister during the cleaning stage.
• Open all green flaps and catches.
• Swing the dryer roller set (8) towards the
back.
• Swing the display panel upwards. When using a daylight cover lift the display panel upwards and then lay to one side.
• Extract the roller sets for water (6) and film in­sertion (3).
• Lift the roller sets developer/fixer (5), tilt to the side and allow the chemicals to drain off.
• Thoroughly rinse the roller sets with the corre­sponding sponges under running hot water.
22
23
24
EN
28 2018/02SE
22
25
• Press the side levers and remove the lifting bath (22).
For cleaning, use the black sponge for
developer areas and the red sponge for fixer areas.
• Clean the lifting bath (22) thoroughly with the
corresponding sponges under running hot water (at least 40°C).
• Rinse the tanks with hot water and clean thor­oughly with the corresponding sponges.
• Thoroughly rinse the roller sets with the corre­sponding sponges under running hot water.
EN
2018/02SE 29
52
16.1.2 " Preparation of chemicals "
The preparation of chemicals decribed is
based on the application of "DÜRR-AUTOMAT XR/C" chemicals, as supplied.
• Screw the large plugs for developer (21) and fixer (18) into the respective tanks (see page
28).
• Connect the developer and fixer hoses again to collector vessel.
• Place the separating panel (52) between developer and fixer tanks in position. This se­parating barrier prevents mixing of chemicals when refilling.
When replenishing chemicals wear
protective gloves and protective glasses.
If any fluid comes into contact with
the eyes, rinse thoroughly with wa­ter and visit a doctor as soon as possible.
After completion rinse hands
thoroughly under running water.
Please also refer to our safety notes which are available on-line: http://www.duerr.de or order directly from Dürr Dental.
1. Chemicals for fixer bath (F)
• Cover the developer tank with a cloth or paper towel in order that no splashes lead to mixing of chemicals.
• Pour the fixer concentrate carefully into the fi­xer tank.
•
Info: 1 bottle of concentrated liquid provides 5 l usable solution
• Fill the concentrate bottle with water and pour this too into the fixer tank.
• Fill the tank with water (max. 20 °C) up to the upper marking (54).
2. Chemicals for developer bath (D)
• Be sure to remove any splashes that might have occurred from fixer refilling from the em­pty developer bath!
• Pour the developer concentrate carefully into the developer tank.
• Fill the concentrate bottle with water and pour this too into the fixer tank.
• Fill the tank with water (max. 20 °C) up to the upper marking (54).
• Replace lifting bath and roller set and screw in the plug.
54
26
27
52
EN
30 2018/02SE
54
22
16.2 Every 2 months cleaning tanks and roller sets
The cleaning of tanks and roller sets de-
cribed is based on the application of "DÜRR-AUTOMAT XR/C" chemicals, as supplied.
• First drain off the chemicals, see section
16.1.1
• Lay the roller set in a wash basin and spray,
e.g. with "spray cleaner WR 2000", and leave to work for ca. 10 min.
18
21
• Press the lever to the side and remove the lift­ing bath (22).
Use the black sponge for developer, the
red sponge for fixer areas when cleaning.
• Clean the lifting bath (22) thoroughly with the
correct sponge under running warm water (min. 40 °C).
• Rinse the tanks with warm water (max. °C) and clean thoroughly with the appropriate sponges.
• Rinse the roller set thoroughly with the cor­rect sponge under running warm water .
• Screw the large plugs (
18), (21) back into the
tanks.
Only the developer tank is cleaned.
Chemical cleaning of the fixer tank is not necessary.
• 1.Fill the developer tank (D) with water (max. 20 °C) up to the upper marking (54).
• 2.Pour cleaner into the developer tank (D)
• 3.Fill the fixer tank (F) with water (max. 20 °C)
up to the upper marking (54).
Never use Dürr Universal Cleaner
when working with "DÜRR­AUTOMAT XR/C" chemicals
If even the smallest amount of
Universal Cleaner is left in the deve­loper bath then the chemicals must be prepared again.
28
29
30
EN
2018/02SE 31
60
59
6
5
3
22
44
8
45
58
60
1
• Place lifting bath in position.
• Insert developer and fixer roller sets.
The developer must never be
operated without roller sets!
If roller sets are missing then the developer and fixer fluids can come under pressure and be splashed verti­cally upwards and out of the developer.
Chemicals irritate eyes and skin and there is the danger of serious eye damage.
Without developer and fixer roller sets
there can be no return of chemicals to the appropriate containers.
Danger of chemicals getting into the eyes or of mixing of chemicals.
• Screw in small plugs into the lifting bath.
• Close all green flaps and catches.
• Return the display to its position.
• Replace cover (60).
• Plug in at the mains, press main switch and
switch on unit.
• Start unit, e.g. by placing 3x4 film material into the film insertion slot (1) until the film feed flap is activated: leave the film for ca. 4 min in this position.
• Allow the unit to run for ca. 6 minuten (this causes the hoses and the rotatory pump to be cleaned automatically).
• After ca. 6 min switch off by pressing the main switch; unplug at the mains and remove all power.
• Remove the cover (60).
• Swing the display panel upwards.
• Open all green flaps and catches.
• Unscrew the small plugs from the lifting bath.
• Lift, shake loose fluid off and then extract the
developer and fixer roller sets.
• Lift out the lifting bath.
• Rinse off the roller sets and lifting bath with
water.
• Unscrew the large plugs for developer (21) and fixer (18) tanks and drain off the water
.
• Rinse out the developer bath with water.
31
32
EN
32 2018/02SE
16.2.1 Replace the roller sets
• Place the lifting bath (22) in position.
• Insert the roller set for the dryer (8) and swing
to the rear.
• Place the roller sets for water (6), developer/ fixer (5) and then film insertion (3) in place in that order. Note carefully that the guide wheels mesh together correctly. It might be necessary to turn the cogged drive wheel (58) until it falls corrcetly into the worm screw of the drive shaft.
• Swing the dryer roller set (8) towards the front.
• Close the green catch (44) and locking
devices (45).
• Place the floating covers (59) for developer and fixer baths in psition.
• Display can now be swung forwards and placed in position.
• Replace cover (60). Note carefully that the lower edge of the cover must be completely engaged into the guide on the housing panels.
• Plug in unit.
60
59
6
5
3
22
44
8
45
58
33
EN
2018/02SE 33
17. Tips for Service Technicians
Any repairs above and beyond routine maintenance must be carried out by suitably qualified personnel or one of our service technicians.
Before any trouble shooting, switch off all power.
Problem Probable cause Solution
1. Appliance not functioning
• Power cable not plugged in correctly.
• Cover not (correctly) in place.
• Power switch released too
soon.
• Plug the power cable into the rear of the machine and into a suitable earthed socket.
• Correctly place cover in position.
• Press main switch for minimum 2 seconds.
2. Film is too light • Developer is empty.
• Film exposure time too short
on image device.
• Temperature of the developer too low.
• Change chemicals (section 16.1).
• Set correct exposure times on the image
device (where required).
• Call Service Technician to check and adjust if necessary.
3. Film is too dark • Film exposure time too long
on image device.
• Temperature of the developer too high.
• Set correct exposure times on the image device (where required).
• Call Service Technician to check and adjust if necessary.
4. Yellow/green fog on the film
• Film not fixed sufficiently.
• Film developed under pro-
gram ENDO (with process time 3:15).
• Check whether fixer is still fresh: agitate an intraoral film for ca. 40 s in fixer. If the film becomes transparent then the fixer is in or­der. If the fog effect remains, change the chemicals (see section 16.1 ff).
• Select program with longer process time. Observe the film processing times (see list of recommended films section 15.7).
• Coarse filter blocked. • Coarse filter at intake nozzle needs cleaning.
• Suction connection not
waterproof.
• Check the suction lines and connections for leaks and correct as necessary.
5. Dark foggy streaks on film
• Cover of daylight operating hood open.
• Light falling into the darkroom, e.g. through keyhole.
• Darkroom lighting incorrectly installed or angled.
• When working with orthochromatic (green sensitive) films, ensure that the cover is closed, otherwise the film will be pre-exposed.
• Check that the dark room is completely light proof, blackout if necessary.
• Lamps must be directed away from work area (indirect light only).
6. Brownish fog on film
• Chemicals exhausted • Mix new chemicals
Trouble-shooting
EN
34 2018/02SE
Display reading further message
follows ...
Possible cause Solution
At beginning and/or during the developer process
Safety relay on PCB (main board) is defect
Call service technician for maintenance. Machine can still be operated.
Before opening unplug at the mains and remove all power!
After switching on the unit
Service intervall has been reached.
Call service technician for maintenance. Machine can still be operated.
After insertion of film and with a delay of approx.35s
The water inflow to water bath has been interrupted
Check that water tap is open and/or inflow hose is connected.
At beginning and/or during the developer process
- Waste water drai­nage blocked, badly laid down or not vented
- Electrical short cir­cuit caused by lime scale on level indi­cator sensors.
Remove blockage or check whether waste water draina­ge hose is correctly laid (water pockets build up). Work can only continue when fault has been rectified.
At beginning and/or during the developer process
RPM-signal from motor interrupted, or cable or connector is defect
Call service technician for maintenance. Problem during film development: film is transported for only the shortest of process times (3:15min)
.
! Not enough fixer !
17.1 Fault message on display panel with audible signal
SERVICE INTERVAL
REACHED!
Problem Probable cause Solution
7. Severe fogging of film
• Films overlap
• Films incorrectly stored
• Temperature of the developer
too high caused be excessi­ve room temperature
• Check the use-by date
• Observe all instructions concerning
storage conditions, use-by-dates, or other manufacturer's instructions on the film packaging.
Do not store film material in the
vicinity of the imaging device. Stray light or radiation can cause pre-exposure of the film.
If necessary, it is possible to set up a separating wall or barrier using lead protection.
• Reduce room temperature
BEFORE OPENING
SWITCH OFF AT
MAINS
ERROR WATER TOO
LOW!
ERROR WATER TOO
HIGH!
ERROR
ROLLER DRIVE
EN
2018/02SE 35
Fault message on display panel with audible signal, see 17.1
18. Tips for Service Technicians
see also 18. Tips for Trouble-shooting
Problem Probable cause Solution
1. Appliance not functioning
• Mains fuse is defect
• Appliance fuse is defect
• Check fusing, change if necessary
• Check fusing, change if necessary (see
section 19.8)
2. No supply of fresh water
•Water tap is blocked or defect
• Water intake blocked
• Lime scale on fill level sensors
(electr. bridging)
• Check water valves, change if necessary (see section 19.10)
• Clean the fine filter or coarse filter in wa­ter hose
• Clean the fill level sensors (see section
19.12)
3. Temperature of the developer not correct
• Process heater is defect
• Air in the system, rotary pump
not vented. (Where unit filled with water or chemicals at ca. 20 °C the unit starts up wit­hout pre-heating.)
• Replace process heater and PTC-sensor (see section 19.9)
• Vent the rotary pump by repeatedly star­ting the unit. Press main switch ON / OFF...
4. Film too light • Developer temperature s
et
too low
• Check the process heater and PTC­sensor, if necessary replace (see section
19.9)
• Measure the temperature enter in Service Menu "CALIBRATION?"
5. Film too dark • Developer temperature too
high or process heater is defect
• Check the process heater and PTC­sensor, if necessary replace (see section
19.9)
• Measure the temperature enter in Service Menu "CALIBRATION?"
6. Yellow/green fog on the film
• No chemicals in developer/fixer roller set (level of bath is too low)
• No agitation of chemicals
• Check the seal on lip and roller set and
change if necessary
• Check the circulation pump, there may be air in the hose
7. Green-brown fog on film
• Film not wet • Level-sensor in water bath has switched off water valve – check the water drainage and clean and lay correctly if necessary (see 9.2 )
• Further alternatives, see Solutions under section18, 2. "No fresh water supply"
8 . Film not com-
pletely dry
• Dryer setting incorrect
• Ventilation fan defective
• Heating defective
• Set dryer performance according to list
of recommended films (see section 15.7)
• Change ventilation fan
• Change temperature control fuse
EN
36 2018/02SE
19. Tips for Trouble-shooting
19.1 Before starting
• Switch off unit at mains switch
• Unplug (61a).
• Remove the cover (60).
• Remove the housing-panels as required for
working.
19.2 Remove front panel
• Swing the display panel upwards.
• Open the green catch (44) and locking devic-
es (3).
• Lift the film insert roller set (3) out.
• Undo the two grey rapid release catches (68)
from the front panel (56).
• Undo the 3 cross-headed screws (57) and re-
move the front panel (56).
19.3 Remove rear panel
• Open the green catches (44) and locking devi-
ces (45) of the dryer roller (8) set.
• Extract the roller set for the dryer (8) and
swing to the rear.
• Undo the 3 cross-headed screws (57) and re-
move the rear panel (67).
19.4 Remove side panels
• Remove front (56) and rear panels (67), lift the
side panels (55) upwards to remove.
19.5 After completing work
• Replace the housing panels correctly.
Never operate the unit without cor-
rectly inserted and secured housing including panels! The unit is con­nected to voltage supply - danger of electrical shock.
• Replace the cover (60), plug in unit.
56
55
57
57 67
68
3
55
60
61
44
45
8
34
35
36
EN
2018/02SE 37
19.6 Service mode plan
• Start Service mode
Switch off unit at power switch (press
min. 2 secs)
•
+ simultaneously press (min. 4 s)
Unit now in Service mode
• Select service function
Use key
and Service Function to
select
• Change settings
with
or changes parameter values
with 1s, confirm new values
*1 Set language, frequency or value
first appears on restart
*2 Factory settings
• End Service mode
Switch off unit at power switch (press
min. 2 secs)
EN
38 2018/02SE
A
1
SERV.INTERVAL?
SERVICE OFF SERVICE ON
BACK
COMP. TEST?
ROLLER DRIVE?
START MANUEL
START MANUEL OFF
START MANUEL ON
DRIVE OFF
DRIVE ON
BATH HEATER?
BATH HEATER OFF
BATH HEATER ON
CIRCULAT.PUMP?
CIRCULAT.PUMP OFF
CIRCULAT.PUMP ON
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=OK
[ENTER]=OK
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
[ENTER]=CONTINUE
DRYER?
DRYER OFF
DRYER ON
VENTILATOR?
VENTILATOR OFF
VENTILATOR ON
LIFTING PAN?
LIFTING PAN DOWN
LIGHT BARRIER
LIFTING PAN ON/OFF
LIGHT BARRIER
FILM FLAP ON/OFF
LIFTING PAN TOP
REGEN.PUMP?
REG.PUMP OFF
REG.PUMP ON
WATER VALVE?
WATER VALVE OFF
WATERVALVE ON
INPUTS
WATER TO HIGH
ON/OFF
WATER TO LOW
ON/OFF
CALIBRATION?
VALUE 28.0 °C
BACK
[ENTER]=CONTINUE
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
[ENTER]=OK
SERVICE RESET
SERVICE (TARGET)
0490H
DUE
0500H
EN
2018/02SE 39
62
64 65
90
91
92
86 93
19.7 Changing display
• Pull the connector of the display cable (92) from the PCB (main board) (86) and draw through opening (91).
• Swing the display (90) upwards and remove.
• Install the new display in reverse order.
19.8 Check unit fusing
• Open the flap (64) to the rear of the appliance by applying light pressure with a screwdriver.
• Press the spring retainer together and extract the fuse cartridge (65).
• Check fuse, change if necessary. Order no. see section 4.1 Special accessories.
37
38
39
EN
40 2018/02SE
16
17
34
81
15
80
79
19.9 Changing the process heater and PTC-sensor
• Remove all covers and panels.
• Drain off chemicals, see section 16.1.1
• Pull the connector (93, fig. 43) from PCB
(main board) (86).
• Unscrew the screw on rotatory pump mount-
ing and lift pump.
• Remove both connectors from the rotatory
pump and the connector to process heater (16).
• Remove process heater from its mounting and
uncouple hoses.
• Install the new process heater correctly.
• Attach all connectors to rotatory pump and
process heater.
• Attach the connector (93, fig. 43) onto PCB
(main board) (86).
• Replace the housing panels correctly.
19.10 Check water intake valve
• When the water intake valve is not switched
on:
• Start the Service mode (see section19.6), se-
lect menu point "Individual Test" and then "Water Intake Valve".
• If the valve does not switch on despite electri-
cal voltage being present then the complete valve (15) should be replaced, see section
19.11"Replacing water intake valve".
19.11 Replacing water intake valve
First check the water intake valve, see section
19.10 above, replace if defective.
• Close the water tap.
• Carefully remove the water hose from the wa-
ter inflow connection (34) of the unit.
• Pull the connecting cable (80) from the valve.
• Pull the inner water hose (79) carefully from
the valve from behind.
• Undo the screws (81) next to the water inflow
and replace the complete valve unit.
40
EN
2018/02SE 41
83
19.12 Cleaning fluid level indicator sensor "Water"
• Close the water tap.
• Remove the cover.
• Open the green flaps and catches.
• Swing the dryer roller set towards the back.
• Remove roller set for water.
• Clean the fluid level indicator sensors
(82)+(83), to prevent lime scale build up form­ing a bridge.
• To check function, the unit must be started up
without the water roller set in place.
• Simulate the insertion of a film, the display
should have message "WATER TOO LOW".
• Now build a bridge between both level indica-
tor sensors (82)+(83); the message "WATER TOO HIGH" appears, the sensors are now free. Cleaning is completed. In the event of any deviation, see Trouble­shooting, section 17.1
19.13 Changing light barrier sensors
Use the light barrier spare parts kit Order no. 1700-993-00.
A description is provided with the spare parts kit.
82
41
EN
42 2018/02SE
EN
2018/02SE 43
DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com
Loading...