Durr XR25 Operating Manual

Operating manual DÜRR X-RAY film processor XR25
1995/04
9000- 600- 04
51
52
CONTENTS
1. SAFETY ADVICE
1. Safety advice .......................................... 51
1.1 Electrical safety advice ................... 51
1.2 General information ......................... 51
2. Scope of XR 25 ...................................... 52
3. Model review .......................................... 52
4. Operation ............................................... 53
5. Changing of chemicals ......................... 54
5.1 Cleaning the roller sets....................54
5.2 Cleaning the bath ............................ 55
5.3 Before draining the chemicals ........ 56
5.4 Draining the chemicals .................... 56
5.5 Cleaning the tanks ........................... 57
5.6 Filling up with chemicals ................. 58
5.7 Inserting roller sets .......................... 58
5.8 Putting on lid .................................... 58
5.9 Switching on XR25 for operation ..... 58
6. Daylight attachment .............................. 59
7. Film storage ........................................... 59
8. Surgery breaks ...................................... 60
9. Fault finding info . ................................. 61
1.1 Electrical safety advice
Work on electrical apparatus must only be carried out by qualified electricians (VBG 4) or through our trained Service Technicians. With the combination of equipment, as well as with electrical repairs, the relevant safety and check regulations must be observed. (VDE 0100, or else, IEC 364, 0105 pt.1, 0701, EN603351, VBG4, in other countries the relevant regulations such as IEC, CEE must be complied with). For repairs only use original parts otherwise the guarantee and test certification will be invalidated.
1.2 General advice
In the BRD, as well as some other countries, used x-ray chemicals are regarded as special waste and must be collected and disposed of separately.
53
2. XR 25 OUTLINE
3. MODEL SURVEY
• Automatic film developing of various film sizes from 2x3 to max. 24x30 cm.
• A fresh mix of chemicals lasts approx. 3 weeks based on a usage rate of 10 intra-oral and 3 panorama films per day. This statement only refers to Durr Automat XR
Chemicals.
• Films can be fed in continuously.
• The run through time is approx. 6 mins. With model 1702-03 (variable) it is between 3-6 mins.
• The transport of films in processing is via roller sets.
• In the Processor the developer and fixer solutions are warmed and kept to a constant temperature within an accepted tolerance range. The warm-up phase is approx 20 mins. depending upon ambient temperature.
• If the XR 25 is installed in a darkroom, no daylight attachment is required.
XR25 X-ray film processor Model: 1702-01 (230V)
1702-02 (110V) 1702-03 (110V, variable processing time)
3.1 The whole programme at a glance
Please open up left fold-in page! Pos./Description Order n. ((1) XR25 X-ray film processor
230V, 50Hz .....................................1702-01
or
XR25 X-ray film processor 110V, 50-
60Hz ............................................... 1702-02
or
XR25 X-ray film processor 110V, 50-60Hz
(variable)........................................1702-03
(2) Darkroom light ...................... 1482-000-00
(3) *Collecting container
for Developer ........................ 1700-062-00
Collecting container
for Fixer ................................. 1700-061-00
(4)* Storage container
for Developer ....................... 1700-060-00
Storage container
for Fixer ................................. 1700-070-00
(5) Daylight attachment DL24
with filter pane ...................... 1700-910-00
Darkroom attachment
DA 24 ..................................... 1700-930-00
suitable for commonly used dental films e.g. 2x3 up to 18x24 or 15x30 cm films
(6) Daylight attachment DL26
with filter pane ...................... 1700-920-00
suitable for film sizes up to 24x30
* It is possible to order a Collection system
consisting of: 1 Collecting container and 3 Storage containers
Collection system for Developer .. 1700-060-00
Collection system for Fixer .......... 1700-070-00
54
XR25
DÜRR DENTAL
Film­Transport
A
4. OPERATION
MORNING: Open water tap EVENING: Close water tap
Never let machine run when dry and never without D/F roller in situ. Pull out power plug on emptied machine.
Morning
• Open water tap
• Mains switch (G) ON
Warming up phase begins (approx, 20mins.)
Yellow indicator light (D) shines
Ready for use
The correct processing temperature (28°C) is reached as soon as the green indicator (C) lights up.
Film transport
• Film transport switch (A) ON Red light indicator (B) lights up
Advice: Insert DURR cleaningfilm, order no.
1410-007-00- before starting the day’s processing by feeding in film at the film entrance.
• Feed in film at film entrance.
Advice: Always feed in films straight and do
not have them overlapping, to avoid bending or blocking the transportation.
• When the last film in a sequence has been processed, switch OFF transport switch (A). Red light (B) goes out.
B C D G
(A) Film transport switch ON/OFF (B) Film transport ON
RED indicator lights up
(C) Ready for use
Green indicator lights up
(D) Warm up phase
Yellow indicator lights up
(G) Mains switch ON/OFF
55
5. CHANGING CHEMICALS
Chemicals should be changed every 3 weeks. Advice: The 3 weekly chemical change
G
1
5.1 Cleaning the roller sets
2
Advice: Any hard particle deposits can
Left and right fold-in pages:
• Mains switch (G) off, see fig. 1
• Pull out mains plug.
• Lift off lid (7) and put to one side.
• Open all latches (25) and catches (24),
• Remove guide plate (8) from dryer set.
• Tilt dryer set (23) backwards and then lift out.
• Remove water roller set (26) and film
specifically refers to: a) The use of „DURR Automat XR
chemicals“.
b) A daily max. film throughput of 10
intra-oral and 3 panoral films.
In the BRD, as well as some other countries, used x-ray chemicals are regarded as special waste and must be collected and disposed of separately. We recommend the use of a DURR Collection & Storage system, see left fold-in page (3) and (4).
safely and easily be removed with the special DURR Spray Cleaner WR 200 (P/N CCB800C55..). For Cleaning the roller racks periodically (3-6 mos) please use DURR Cleaning Bath set (P/N 1700-007-
00).
see right fold-in page.
entrance set (28).
56
3
• Remove both bath covers (9) (only on model 1702-03 variable) from the developer and fixer bath, let drip off and then lay aside.
4
• Take out Developer/Fixer roller set (27). Firstly, lift the roller set vertically with both hands, then carefully pour the remainder of the chemicals into the baths over the front side of the roller set.
• Clean all roller sets under warm water (approx.40°C).
When cleaning the roller sets use the correct coloured sponge. Fixer - red Developer - black To disregard this advice could cause contamination of fresh chemicals.
5
5.2 Cleaning the bath
• Lift bath (11) with both hands. Pour remaining chemicals back into the tanks of the XR25.
• Rinse empty bath with warm water using the appropriate sponges provided.
6
57
5.3 Prior to draining the chemicals
Please observe the following:
If collection vessels (3) are present, check whether they are empty. Otherwise empty the chemicals from them (4). To avoid overflow it is important that the vessels (3) are previously emptied. Sobald die Chemiekalien vom Entwickl­ungsgerät abgelassen werden, fließen sie in die Auffangbehälter.
7
5.4 Draining the chemicals
• Unscrew the red and black plugs (29).
58
8
With collection vessels (3) in position, the
used chemicals flow into them.
• Decant the collected and used chemicals from the collection vessel (3) into the storage containers (4).
9
5.5 Cleaning the tanks
• With the appropriate wet sponge, wipe both Tanks (13)(Developer-black, Fixer-red).
• Screw in red and black plugs (29), see fig.8.
9
10
• Fill both tanks up to the mark (14) with warm water (approx.40°C).
• Insert Developer/Fixer roller set, close latches and catches (24), see right fold-in page.
Never let Processor run when dry.
If there are no chemicals or water in the tanks and the mains switch is activated, this can lead to a technical defect.
If the Developer/Fixer roller set is not inserted, the water will spurt upwards
as the conduit for the solution is not in place.
• Replace lid (7) again, see fig.2.
• Insert mains plug.
• Mains switch (G) ON, see p. 5.
• Film transport switch (A) ON, let XR25 run for approx. 10 mins. The remaining chemicals are removed from the hoses and circulation pump.
• Mains switch (G) OFF
• Pull out mains plug.
• Remove lid (7).
• Take out developer/fixer roller set (27), see fig.5.
• Unscrew red and black plugs (29). The water will flow into the empty collection vessels (3).
• Empty vessels (3) again.
11
59
12
13
EE
E
EE
EE
E
EE
FF
F
FF
FF
F
FF
5.6 Filling with chemicals
• Screw in red and black plugs (29), see right fold-in page.
• Insert separating plate (13) between Fixer tank „F“ and Developer tank „E“, black side towards developer, red side towards fixer.
1.Filling up tank F-fixer (red):
• Pour in large red bottle A and top up with cold water to the lower mark (14).
• Now pour in small red bottle B and top up with cold water to the upper mark (15).
2.Filling up tank E-developer (black):
• Pour in large black bottle A and top up with cold water to the lower mark (16).
• Now pour in small black bottle B and top up with cold water to the upper mark (17).
• Take out separating plate (13), clean it and store it.
• Replace bath (11), see fig.6.
5.7 Insert roller sets
see right fold-in page and fig.14.
• Firstly, lift all latches (25) ready to receive roller sets.
• Insert roller sets. Secure all latches (25) and catches (24).
5.8 Replace lid
see fig.2 (7).
The lid should fit easily. If not, check that all catches (24) and latches (25) are properly closed.
60
5.9 Making XR25 ready for use
• Insert mains plug
• Power switch (G) ON, see p.5. Yellow indicator (D) lights up, warm up phase
begins (approx.20mins.) As soon as the green indicator (C) lights up, the correct temperature has been reached and the XR25 is ready for use.
14
15
16
6. DAYLIGHT ATTACHMENT
If no darkroom is present, two daylight attachments, DL24 und DL26, are available to choose from, see left fold-in page, fig.(5) and (6). The order no. as well as the recommended max. film size for each is on p.4 under pt. 3.1.
„The whole programme at a glance“ can be seen. For easier handling of wrapped intra-oral films, the lid (18) can be lifted whilst the filter pane (19) remains closed.
When processing thin coated (Extra­oral) RP films, the lid must be definitely closed. Otherwise the film will be pre-exposed.
When necessary, individual elements of the daylight attachment can be altered. Through this, the arm sleeves (20) can be adjusted to suit individual requirements.
7. FILM STORAGE
In the film compartment (21) which is situated in the daylight attachment, a max. of 10 extraoral films can be stored.
When storing films in the compartment it is important to observe the following:
When using the film storage compartment always use up the films before refilling. This will prevent films being stored too long which, if ignored, will cause ‘veiling`of the films.
• Films 12x30, 13x18 and 15x30 place in film holder (22).
- Hang film holder in storage compartment
- Lay films in film holder.
• Films 24x30 and 24x18 do not require the film holder
- Lay films directly into film compartment
- The 24x18 films are stood upright in the film compartment.
Close film compartment (21) each time after taking out a film. If the film storage area is not closed there is always the danger of exposing the films whenever the lid is lifted.
17
61
8. SURGERY BREAKS
The max. lifespan of the chemicals in the tanks is 4 weeks. If this period is exceeded during a Surgery break,e.g. holidays, the chemicals should be drained beforehand. The tanks and roller sets cleaned with warm water (approx. 40°C) and then left standing empty. The procedure as under pt.5 „CHANGING THE CHEMICALS“.
62
9. FAULT FINDING INFO.
Condition
1. Film too light
2. Films too dark
3. Yellow/Green veiling on the film
4. Dark veiling on films
Possible causes
• Developer temperature too low
• X-ray exposure time too short
• Developer solution exhausted
• Developer temperature too high
• X-ray exposure time too long
• Film not properly fixed
• Light getting into darkroom, e.g. through keyhole
• Darkroom light wrongly positioned
• Film has been affected by Processor control lights
• Lid of daylight attachment not properly closed
Remedy
Set temperature to 28°C
If possible, longer exposure time
Drain chemicals and mix fresh solutions.
Set temperature to 28°C
Shorten x-ray exposure time
Test whether Fixer solution is OK: Swish film to and fro for approx. 40 secs. If the film is transparent, the fixer is fine. Should veiling persist: Drain chemicals and mix fresh ones.
Check darkroom for light leakage and seal off source.
Only indirect light allowed
Check all control lights and if necessary dim them
When inserting extraoral and panoral films the cover of the filter pane must be closed.As films are light sensitive, the lid must be properly closed.
5. Increased background veiling on the film
6. Brownish veiling on the film
• Films out of date
• Film falsch gelagert.
• Film not properly washed
Check sell-by date Store films in a cool place. Films should
not be kept near the x-ray machine. (Scattered radiation)
Open water tap. Safety switch in water bath has switched off water inlet valve - Check water drain..
63
64
Loading...