DÜRR Compressor Duo, Compressor Tandem, Compressor Quattro User and service manual

Duo, Trio, Quattro, Duo Tandem, Quattro Tandem
Installation and operating instructions
EN
0297
9000-610-45/30

1811V007
Contents
Important information
1 About this document � � � � � � � � � � � � � � � 3
1�1 Warnings and symbols � � � � � � � � � � 3 1�2 Copyright information � � � � � � � � � � � 3
2 Safety � � � � � � � � � � � � � � � � � � � � � � � � � � � 4
2�1 Intended purpose � � � � � � � � � � � � � � 4 2�2 Intended use � � � � � � � � � � � � � � � � � � 4 2�3 Improper use � � � � � � � � � � � � � � � � � 4 2�4 General safety information � � � � � � � � 4 2�5 Qualified personnel � � � � � � � � � � � � � 4 2�6 Electrical safety � � � � � � � � � � � � � � � � 4 2�7 Only use original parts � � � � � � � � � � � 4 2�8 Transport � � � � � � � � � � � � � � � � � � � � 5 2�9 Disposal � � � � � � � � � � � � � � � � � � � � � 5
Product description
3 Overview� � � � � � � � � � � � � � � � � � � � � � � � � 6
3�1 Scope of delivery � � � � � � � � � � � � � � 6 3�2 Optional accessories � � � � � � � � � � � � 6 3�3 Wear parts and replacement
parts � � � � � � � � � � � � � � � � � � � � � � � � 6
4 Technical data � � � � � � � � � � � � � � � � � � � � 7
4�1 Duo � � � � � � � � � � � � � � � � � � � � � � � � 7 4�2 Duo � � � � � � � � � � � � � � � � � � � � � � � � 9 4�3 Trio � � � � � � � � � � � � � � � � � � � � � � � � 11 4�4 Quattro � � � � � � � � � � � � � � � � � � � � � 13 4�5 Duo Tandem � � � � � � � � � � � � � � � � � 15 4�6 Quattro Tandem � � � � � � � � � � � � � � 17 4�7 Distance between rubber feet � � � � 19 4�8 Type plate � � � � � � � � � � � � � � � � � � � 19
5 Operation � � � � � � � � � � � � � � � � � � � � � � � 20
5�1 Duo with membrane drying unit � � � 20
Assembly
6 Requirements� � � � � � � � � � � � � � � � � � � � 21
6�1 Installation/setup room � � � � � � � � � 21 6�2 Setup � � � � � � � � � � � � � � � � � � � � � � 21 6�3 Information about electrical
connections � � � � � � � � � � � � � � � � � 22
7 Transport � � � � � � � � � � � � � � � � � � � � � � � 22 8 Quattro Tandem assembly and
installation � � � � � � � � � � � � � � � � � � � � � � 23
9 Installation � � � � � � � � � � � � � � � � � � � � � � 23
9�1 Remove the transport locks � � � � � 23 9�2 Establishing the compressed air
connection � � � � � � � � � � � � � � � � � � 23 9�3 Pressure reducer � � � � � � � � � � � � � 24 9�4 Place a collector tray underneath � 24 9�5 Electrical connections � � � � � � � � � � 24
10 Commissioning� � � � � � � � � � � � � � � � � � � 25
10�1 Check the motor protection
switch � � � � � � � � � � � � � � � � � � � � � � 25 10�2 Checking the switch-on/cut-off
pressure � � � � � � � � � � � � � � � � � � � � 25 10�3 Checking the safety valve � � � � � � � 26 10�4 Draining the condensation water � � 26 10�5 Adjusting the rate of flow at the
pressure reducer � � � � � � � � � � � � � � 27
11 Adjustment options � � � � � � � � � � � � � � � 27
11�1 Adjusting the pressure switch � � � � 27 11�2 Adjusting the motor protection
switch � � � � � � � � � � � � � � � � � � � � � � 28
12 Circuit diagrams� � � � � � � � � � � � � � � � � � 29
12�1 1/N/PE AC 110-127 V, 230 V
layout � � � � � � � � � � � � � � � � � � � � � � 29 12�2 3/N/PE AC 400 V layout � � � � � � � � 29 12�3 1/N/PE AC 230 V layout, Duo
Tandem � � � � � � � � � � � � � � � � � � � � 30 12�4 3/N/PE AC 230 V layout, Quattro
Tandem � � � � � � � � � � � � � � � � � � � � 31
EN
Contents
Usage
13 Operation � � � � � � � � � � � � � � � � � � � � � � � 32
13�1 Switching the unit on/off � � � � � � � � 32
14 Maintenance� � � � � � � � � � � � � � � � � � � � � 33
14�1 Maintenance schedule � � � � � � � � � 33 14�2 Replacing the air intake filter � � � � � 34 14�3 Replacing the filter in the
condensate separator � � � � � � � � � � 34
EN
14�4 Replacing the filter of the
membrane drying unit � � � � � � � � � � 35
15 Taking out of use � � � � � � � � � � � � � � � � � 36
15�1 Taking the unit out of use � � � � � � � 36 15�2 Storage of the unit � � � � � � � � � � � � 37
Troubleshooting
16 Tips for operators and service
technicians � � � � � � � � � � � � � � � � � � � � � � 38
2 9000-610-45/30 1811V007

Important information

1 About this document

These installation and operating instructions represent part of the unit�
If the instructions and information in these installation and operating instruc­tions are not followed, Dürr Dental will not be able to offer any warranty or as­sume any liability for the safe operation and the safe functioning of the unit�

1.1 Warnings and symbols

Warnings
The warnings in this document are intended to draw your attention to possible risks of personal injury or material damage�
The following warning symbols are used:
General warning symbol
Warning – dangerous high voltage
Warning – hot surfaces
Warning - automatic start-up of the unit
The warnings are structured as follows:
SIGNAL WORD Description of the type and source of
danger
Here you will find a description of the possible consequences of ignoring the warning�
i Follow these measures to avoid the
danger�
The signal word differentiates between four levels of danger:
DANGER
Immediate danger of severe injury or death
WARNING
Possible danger of severe injury or death
CAUTION
Risk of minor injuries
ATTENTION
Risk of extensive material/property damage
Other symbols
These symbols are used in the document and on or in the unit:
Note, e�g� specific instructions regarding efficient and cost-effective use of the unit�
Observe the operating instructions�
Switch off and de-energise the unit (e�g�unplug from mains)�
Air
Dispose of correctly in accordance with EU Directive 2012/19/EU (WEEE)�
CE labelling with the number of the noti­fied body
Order number
Serial number
Manufacturer

1.2 Copyright information

All circuits, processes, names, software pro­grams and units mentioned in this document are protected by copyright�
The Installation and Operating Instructions must not be copied or reprinted, neither in full nor in part, without written authorisation from Dürr Dental�
EN
Important information

2 Safety

Dürr Dental has designed and constructed this unit so that when used properly and for the in­tended purpose it does not pose any danger to people or property� Nevertheless, residual risks can remain� You should therefore observe the following notes�

2.1 Intended purpose

The compressor is designed to supply com­pressed air for dental applications�
EN

2.2 Intended use

The air supplied by the compressor is suitable for driving dental tools�
The compressed air generated by the compres­sor is delivered to the pipeline system of the surgery� The entire compressed air system must be designed in such a way that the quality of the compressed air generated by the compressor is not impaired�
With this prerequisite, the air provided by the compressor is also suitable for blow-drying tooth preparations�

2.3 Improper use

Any other usage or usage beyond this scope is deemed to be improper� The manufacturer ac­cepts no liability for damages resulting from im­proper use� In these cases the user/operator will bear the sole risk�
WARNING Risk of explosion due to ignition of combustible materials
i Do not operate the unit in any rooms
in which inflammable mixtures may be present, e�g� in operating theatres�
i The unit is not suitable for providing an air
supply to respirators�
i This unit is not suitable for drawing up fluids
or for compressing aggressive gases or po­tentially explosive gases�

2.4 General safety information

i Always comply with the specifications of all
guidelines, laws, and other rules and regula­tions applicable at the site of operation for the operation of this unit�
i Check the function and condition of the unit
prior to every use�
i Do not convert or modify the unit� i Comply with the specifications of the Installa-
tion and Operating Instructions�
i The Installation and Operating Instructions
must be accessible to all operators of the unit
at all times�

2.5 Qualified personnel

Operation
Unit operating personnel must ensure safe and correct handling based on their training and knowledge�
i Instruct or have every user instructed in han-
dling the unit�
Installation and repairs
i Installation, readjustments, alterations, up-
grades and repairs must be carried out by
Dürr Dental or by qualified personnel specifi-
cally approved and authorized by Dürr Dental�

2.6 Electrical safety

i Observe and comply with all the relevant elec-
trical safety regulations when working on the
unit�
i Replace any damaged cables or plugs imme-
diately�

2.7 Only use original parts

i Only use Dürr Dental parts or accessories and
special accessories specifically approved by
Dürr Dental�
i Only use only original wear parts and replace-
ment parts�
Dürr Dental accepts no liability for dam­ages or injury resulting from the use of non-approved accessories or optional accessories, or from the use of non-origi­nal wear parts or replacement parts�
The use of non-approved accessories, optional accessories or non-genuine wear parts / replacement parts (e�g� mains cable) can have a negative effect in terms of electrical safety and EMC�
4 9000-610-45/30 1811V007

2.8 Transport

The original packaging provides optimum pro­tection for the unit during transport�
If required, original packaging for the unit can be ordered from Dürr Dental�
Dürr Dental will not accept any responsi­bility or liability for damage occurring dur­ing transport due to the use of incorrect packaging, even where the unit is still un­der guarantee�
i Only transport the unit in its original packag-
ing�
i Keep the packing materials out of the reach of
children�

2.9 Disposal

Unit
The unit must be disposed of properly� Within the European Union, the unit must be disposed of in accordance with EU Directive 2012/19/EU (WEEE)�
i If you have any questions about the correct
disposal of parts, please contact your dental trade supplier�
Important information
EN

Product description

3 Overview

3.1 Scope of delivery

The following items are included in the scope of delivery (possible variations due to country-spe­cific requirements and/or import regulations):
Duo 110V, 1~, with condensate sep­arator � � � � � � � � � � � � � � � � � � � � � � 5250100004
Duo 110V, 1~, with membrane dry­ing unit � � � � � � � � � � � � � � � � � � � � � 5252100001
EN
Duo 230V, 1~, with membrane dry­ing unit � � � � � � � � � � � � � � � � � � � � � � � � 5252-01
Duo 230V, 1~, with membrane dry­ing unit � � � � � � � � � � � � � � � � � � 5252-01/Service
Duo 400V, with membrane drying unit � � � � � � � � � � � � � � � � � � � � � � � � � � � 5252-51
Trio 230V, 1~, with membrane drying unit � � � � � � � � � � � � � � � � � � � � � � � � � � � 5352-01
Trio 230V, 1~, with membrane drying unit � � � � � � � � � � � � � � � � � � � � � � � � 5352100001
Quattro 400V, with membrane drying unit * � � � � � � � � � � � � � � � � � � � � � � � � � 5452-51
Quattro 400V, with membrane drying unit * � � � � � � � � � � � � � � � � � � � � � � 5452100001
Duo Tandem 230V, 1~, with mem­brane drying unit � � � � � � � � � � � � � � � � 4252-01
Duo Tandem CAD/CAM 230V, 1~, with membrane drying unit * � � � � 4252100004
Quattro Tandem 230V, 3~, with membrane drying unit * � � � � � � � � � � � 4682-52
Quattro Tandem 400V, 3~, with membrane drying unit * � � � � � � � � � � � 4682-53
– Fabric reinforced hose – Hose nozzle – Hose clamp – Installation and operating instructions – Appliance log book – Collector tray
* with sterile filter

3.2 Optional accessories

The following items can optionally be used with the unit; these items do not bear the CE mark:
Pressure reducer � � � � � � � � � � � � � � 6040-992-00 Sterile filter � � � � � � � � � � � � � � � � � � 1640-981-00 Wooden cabinet for for sound insula-
tion of Duo Compressors � � � � � � � 5150-500-00 Wooden cabinet for for sound insula-
tion of DuoTandem, Trio and Quattro Compressors � � � � � � � � � � � � � � � � 4251-500-00

3.3 Wear parts and replacement parts

The following working parts must be replaced at regular intervals (refer also to "Maintenance"); these articles do not bear the CE mark:
Air intake filter � � � � � � � � � � � � � � � � 0832-982-00 Fine filter � � � � � � � � � � � � � � � � � � � � 1610-121-00 Sterile filter � � � � � � � � � � � � � � � � � � 1640-981-00 Sintered filter � � � � � � � � � � � � � � � � � 1650-101-00 Fleece filter � � � � � � � � � � � � � � � � � �4280-982-00
Any repairs exceeding routine mainte­nance may only be carried out by quali­fied personnel or our service�
Information about replacement parts is available from the portal for authorised specialist dealers at: www�duerrdental�net�
If the mains cable of this unit is damaged it must only be replaced by an original mains cable from the manufacturer�
6 9000-610-45/30 1811V007
Product description

4 Technical data

4.1 Duo

Electrical data 5252-01
5252-01/Service
Rated voltage V 230 400 Mains frequency Hz 50 60 50 60 Nominal current at 8 bar (0�8 MPa) A 6�3 7�0 3�1 2�5 Motor protection switch, recommended
settings A 6�5 7�6 3�1 2�5 Speed rpm 1360 1600 1410 1690 Type of protection IP 24 IP 24 Mains fuses * A 10 10 Max� permissible mains impedance in ac-
cordance with EN61000-3-11 ** W 0�142 ­* Circuit breaker fuse characteristics B, C or D in acc� with EN60898-1 ** Mains impedance at 6 switching cycles per hour� If the number of switching cycles per hour is
higher a lower mains impedance is required�
General technical data
Pressure tank volume l 20 20 Suction power, approx� l/min 210 255 210 255 Delivery at 5 bar (0�5 MPa) l/min 115 130 115 130 Pressure build-up phase 0 - 7�5 bar (0 -
0�75 MPa) approx� s 85 70 85 70 Duty cycle % 100 100 Start-up pressure bar (MPa) 6 (0�6) 6 (0�6) Cut-off pressure bar (MPa) 7�8 (0�78) 7�8 (0�78) Cut-off pressure, max� adjustable bar (MPa) 9�5 (0�95) 9�5 (0�95) Safety valve, maximum permissible oper-
ating pressure bar (MPa) 10 (1) 10 (1) Pressure dew point at 7bar (0�7MPa) * °C +5 +5 Dimensions (H x W x D) ** cm 69 x 49 x 46 69 x 49 x 46 Weight kg 47 47 Noise level *** without sound insulation with sound insulation * Value determined at an ambient temperature of +40°C ** Values without accessories and add-on parts *** Noise level in accordance with ENISO1680 "Airborne noise emissions"; measured in a room
with sound damping� The levels are average values with a tolerance of ± 1�5 dB(A)� Higher val­ues may be obtained in rooms with reverberating sound characteristics�
dB(A) dB(A)
66 55
69 61
5252-51
66 55
69 61
EN
Filter mesh size
Compressor air intake filter µm 3 Fine filter for membrane drying unit µm 3 Sterile filter for membrane drying unit µm 0�01 Sintered filter for membrane drying unit µm 35
Product description
Ambient conditions during storage and transport
Temperature °C -10 to +55 Relative humidity % max� 95
Ambient conditions during operation
Temperature °C +10 to +40 Ideal temperature °C +10 to +25 Relative humidity % max� 95
Classification
Medical Devices Directive (93/42/EEC) Class IIa
EN
8 9000-610-45/30 1811V007
Product description

4.2 Duo

Electrical data 5250100004 5252100001
Nominal voltage V 110-115 110-127 110-115 110-127 Electrical frequency Hz 50 60 50 60 Nominal current at 8 bar (0�8 MPa)
Motor protection switch, recommended settings A 17 16 17 16
Speed
Type of protection IP 24 IP 24 Mains fusing * A 20 20 * Circuit breaker fuse characteristics B, C or D in acc� with EN60898-1
General technical data
Pressure tank volume l 20 20 Suction power, approx� l/min 210 255 210 255 Delivery at 5 bar (0�5 MPa) l/min 130 145 115 130 Pressure build-up phase 0 - 7�5 bar (0 -
0�75 MPa) approx� s 75 65 85 70 Duty cycle % 100 100 Start-up pressure bar (MPa) 6 (0�6) 6 (0�6) Cut-off pressure bar (MPa) 7�8 (0�78) 7�8 (0�78) Cut-off pressure, max� adjustable bar (MPa) 9�5 (0�95) 9�5 (0�95) Safety valve, maximum permissible oper-
ating pressure bar (MPa) 10 (1) 10 (1) Pressure dew point at 7bar (0�7MPa) * °C - +5 Dimensions (H x W x D) ** cm 69 x 49 x 44 69 x 49 x 46 Weight kg 46 50 Noise level *** without sound insulation with sound insulation * Value determined at an ambient temperature of +40°C ** Values without accessories and add-on parts *** Noise level in accordance with ENISO1680 "Airborne noise emissions"; measured in a room
with sound damping� The levels are average values with a tolerance of ± 1�5 dB(A)� Higher val­ues may be obtained in rooms with reverberating sound characteristics�
A
rpm
dB(A) dB(A)
14�1-
14�6
1350-
1370
66 55
14�2-
13�3
1560-
1640
69 61
14�1-
14�6
1350-
1370
66 55
14�2-
13�3
1560-
1640
69 61
EN
Filter mesh size
Compressor air intake filter µm 3 Fine filter for membrane drying unit µm 3 Sterile filter for membrane drying unit µm 0�01 Sintered filter for membrane drying unit µm 35 Fleece filter for automatic condensate
drain µm 5
Ambient conditions during storage and transport
Temperature °C -10 to +55 Relative humidity % max� 95
Product description
Ambient conditions during operation
Temperature °C +10 to +40 Ideal temperature °C +10 to +25 Relative humidity % max� 95
Classification
Medical Devices Directive (93/42/EEC) Class IIa
EN
10 9000-610-45/30 1811V007
Product description

4.3 Trio

Electrical data 5352-01
5352100001
Rated voltage V 230 Mains frequency Hz 50 Nominal current at 8 bar (0�8 MPa) A 8�6 Motor protection switch, recommended
settings A 8�6 Speed rpm 1350 Type of protection IP 24 Mains fuses * A 10 Max� permissible mains impedance in ac-
cordance with EN61000-3-11 W 0�078 * Circuit breaker fuse characteristics B, C or D in acc� with EN60898-1
General technical data
Pressure tank volume l 50 Suction power, approx� l 315 Delivery at 5 bar (0�5 MPa) l/min 160 Pressure build-up phase 0 - 7�5 bar (0 -
0�75 MPa) c� s 140 Duty cycle % 100 Start-up pressure bar (MPa) 6 (0�6) Cut-off pressure bar (MPa) 7�8 (0�78) Cut-off pressure, max� adjustable bar (MPa) 9�5 (0�95) Safety valve, maximum permissible oper-
ating pressure bar (MPa) 10 (1) Pressure dew point at 7bar (0�7MPa) * °C +5 Dimensions (H x W x D) ** cm 76 x 74 x 52 Weight kg 70 Noise level *** without sound insulation with sound insulation * Value determined at an ambient temperature of +40°C ** Values without accessories and add-on parts *** Noise level in accordance with ENISO1680 "Airborne noise emissions"; measured in a room
with sound damping� The levels are average values with a tolerance of ± 1�5 dB(A)� Higher val­ues may be obtained in rooms with reverberating sound characteristics�
dB(A) dB(A)
69 54
EN
Filter mesh size
Compressor air intake filter µm 3 Fine filter for membrane drying unit µm 3 Sterile filter for membrane drying unit µm 0�01 Sintered filter for membrane drying unit µm 35
Ambient conditions during storage and transport
Temperature °C -10 to +55 Relative humidity % max� 95
Product description
Ambient conditions during operation
Temperature °C +10 to +40 Ideal temperature °C +10 to +25 Relative humidity % max� 95
Classification
Medical Devices Directive (93/42/EEC) Class IIa
EN
12 9000-610-45/30 1811V007
Product description

4.4 Quattro

Electrical data 5452-51
5452100001
Rated voltage V 400 Mains frequency Hz 50 60 Nominal current at 8 bar (0�8 MPa) A 4�4 4�8 Motor protection switch, recommended
settings A 5�0 5�0 Speed rpm 1440 1700 Type of protection IP 24 Mains fuses * A 10 Max� permissible mains impedance in ac-
cordance with EN61000-3-11 ** W 0�24 * Circuit breaker fuse characteristics B, C or D in acc� with EN60898-1 ** Mains impedance at 6 switching cycles per hour� If the number of switching cycles per hour is
higher a lower mains impedance is required�
General technical data
Pressure tank volume l 50 Suction power, approx� l/min 420 505 Delivery at 5 bar (0�5 MPa) l/min 215 240 Pressure build-up phase 0 - 7�5 bar (0 -
0�75 MPa) c� s 105 95 Duty cycle % 100 Start-up pressure bar (MPa) 6 (0�6) Cut-off pressure bar (MPa) 7�8 (0�78) Cut-off pressure, max� adjustable bar (MPa) 9�5 (0�95) Safety valve, maximum permissible oper-
ating pressure bar (MPa) 10 (1) Pressure dew point at 7bar (0�7MPa) * °C +5 Dimensions (H x W x D) ** cm 76 x 74 x 52 Weight kg 82 Noise level *** without sound insulation with sound insulation * Value determined at an ambient temperature of +40°C ** Values without accessories and add-on parts *** Noise level in accordance with ENISO1680 "Airborne noise emissions"; measured in a room
with sound damping� The levels are average values with a tolerance of ± 1�5 dB(A)� Higher val­ues may be obtained in rooms with reverberating sound characteristics�
dB(A) dB(A)
70 54
75 55
EN
Filter mesh size
Compressor air intake filter µm 3 Fine filter for membrane drying unit µm 3 Sterile filter for membrane drying unit µm 0�01 Sintered filter for membrane drying unit µm 35
Product description
Ambient conditions during storage and transport
Temperature °C -10 to +55 Relative humidity % max� 95
Ambient conditions during operation
Temperature °C +10 to +40 Ideal temperature °C +10 to +25 Relative humidity % max� 95
Classification
Medical Devices Directive (93/42/EEC) Class IIa
EN
14 9000-610-45/30 1811V007
Product description

4.5 Duo Tandem

Electrical data 4252-01 4252100004
Rated voltage V 230 230 Mains frequency Hz 50 60 50 60 Nominal current at 8 bar (0�8 MPa) A 12�2 14�3 12�2 14�3 Motor protection switch, recommended
settings A 6�5 7�6 6�5 7�6 Speed rpm 1410 1690 1410 1690 Type of protection IP 24 IP 24 Mains fuses * A 20 20 * Circuit breaker fuse characteristics B, C or D in acc� with EN60898-1
General technical data
Pressure tank volume l 50 50 Suction power, approx� l/min 420 505 420 505 Delivery at 5 bar (0�5 MPa) l/min 230 260 230 260 Pressure build-up phase 0 - 7�5 bar (0 -
0�75 MPa) c� s 115 100 115 100 Duty cycle % 100 100 Start-up pressure bar (MPa) 6 (0�6) > 7 (0�7) Cut-off pressure bar (MPa) 7�8 (0�78) 9 (0�9) Cut-off pressure, max� adjustable bar (MPa) 9�5 (0�95) 9�5 (0�95) Safety valve, maximum permissible oper-
ating pressure bar (MPa) 10 (1) 10 (1) Pressure dew point at 7bar (0�7MPa) * °C +5 +5 ** Dimensions (H x W x D) *** cm 76 x 79 x 52 76 x 79 x 52 Weight kg 90 90 Noise level **** without sound insulation with sound insulation * Value determined at an ambient temperature of +40°C ** With ORANGE rinsing nozzle *** Values without accessories and add-on parts **** Noise level in accordance with ENISO1680 "Airborne noise emissions"; measured in a room
with sound damping� The levels are average values with a tolerance of ± 1�5 dB(A)� Higher val­ues may be obtained in rooms with reverberating sound characteristics�
dB(A) dB(A)
69 55
74 56
69 55
EN
74 56
Filter mesh size
Compressor air intake filter µm 3 Sterile filter for membrane drying unit µm 0�01 Sintered filter for membrane drying unit µm 35
Ambient conditions during storage and transport
Temperature °C -10 to +55 Relative humidity % max� 95
Product description
Ambient conditions during operation
Temperature °C +10 to +40 Ideal temperature °C +10 to +25 Relative humidity % max� 95
Classification
Medical Devices Directive (93/42/EEC) Class IIa
EN
16 9000-610-45/30 1811V007
Product description

4.6 Quattro Tandem

Electrical data 4682-52 4682-53
Rated voltage V 230 / 3~ 400 / 3~ Mains frequency Hz 50 60 50 60 Nominal current at 8 bar (0�8 MPa) A 15�2 16�6 8�8 9�6 Motor protection switch, recommended
settings A 9 9 5 5 Speed rpm 1440 1700 1440 1700 Type of protection IP 24 IP 24 Mains fuses * A 20 20 Max� permissible mains impedance in ac-
cordance with EN61000-3-11 ** W 0�445 0�18 * Circuit breaker fuse characteristics B, C or D in acc� with EN60898-1 ** Mains impedance at 6 switching cycles per hour� If the number of switching cycles per hour is
higher a lower mains impedance is required�
General technical data
Pressure tank volume l 90 90 Suction power, approx� l/min 845 1010 845 1010 Delivery at 5 bar (0�5 MPa) l/min 430 480 430 480 Pressure build-up phase 0 - 7�5 bar (0 -
0�75 MPa) c� s 90 80 90 80 Duty cycle % 100 100 Start-up pressure bar (MPa) 6�5 (0�65) 6�5 (0�65) Cut-off pressure bar (MPa) 8�5 (0�85) 8�5 (0�85) Cut-off pressure, max� adjustable bar (MPa) 9�5 (0�95) 9�5 (0�95) Safety valve, maximum permissible oper-
ating pressure bar (MPa) 10 (1) 10 (1) Pressure dew point at 7bar (0�7MPa) * °C +5 +5 Dimensions (H x W x D) ** cm 76 x 102 x 62 76 x 102 x 62 Weight kg 143 143 Noise level *** dB(A) 69 74 69 74 * Value determined at an ambient temperature of +40°C ** Values without accessories and add-on parts *** Noise level in accordance with ENISO1680 "Airborne noise emissions"; measured in a room
with sound damping� The levels are average values with a tolerance of ± 1�5 dB(A)� Higher val­ues may be obtained in rooms with reverberating sound characteristics�
EN
Filter mesh size
Compressor air intake filter µm 3 Fine filter for membrane drying unit µm 3 Sterile filter for membrane drying unit µm 0�01 Sintered filter for membrane drying unit µm 35
Ambient conditions during storage and transport
Temperature °C -10 to +55 Relative humidity % max� 95
Product description
Ambient conditions during operation
Temperature °C +10 to +40 Ideal temperature °C +10 to +25 Relative humidity % max� 95
Classification
Medical Devices Directive (93/42/EEC) Class IIa
EN
18 9000-610-45/30 1811V007
Product description
1
1
b
a
1

4.7 Distance between rubber feet

Distances between the rubber feet for different pressure vessel volumes:
a (cm) b (cm) 20 l 23 27 50 l 32�5 45 90 l 32�5 59

4.8 Type plate

Complete system
The type plate of the overall system is located on the mounting plate for the compressor unit�
Compressor unit
The type plate of the compressor unit is located on the crankcase below the cylinder�
EN
1 Compressor unit type plate
Membrane drying unit
The type plate of the membrane drying unit is located on the side of the membrane drying unit�
1 Membrane drying unit type plate
1 Type plate for the complete system
Product description
1
2
34
14
15
16
17
18 19

5 Operation

5.1 Duo with membrane drying unit

13
6
EN
7
5
1 Compressor unit 2 Air intake filter 3 Pressure tank 4 Condensate drain valve 5 Non-return valve 6 Pressure switch 7 Safety valve 8 Pressure gauge/display 9 Pressure relief valve 10 Cooler with ventilating fan 11 Membrane drying unit 12 Pressure limiting valve 13 Humidity display 14 Fine or sterile filter 15 Rinsing nozzle 16 Membrane fibre 17 Sinter filter 18 Water collection chamber 19 Water outlet valve
The compressor unit draws in atmospheric air and compresses it without oil� It then transports the oil­free compressed air to the membrane drying unit� The cooler and the membrane dryer extract mois­ture from the compressed air� The oil-free, hygienic and dry air is stored in the pressure tank ready for use in connected devices�
20 9000-610-45/30 1811V007
9 8
10 11
12

Assembly

6 Requirements

The unit must not be set up or operated within the vicinity of the patients (within a radius of 1�5m)�
The unit can be installed either at the same level as the surgery room or on a floor below (e�g�cel­lar)�
Due of the amount of noise generated, we rec­ommend that the unit is installed in an adjoining room�
The pipes provided on-site must at least meet the country-specific requirements for drinking water�
Further information can be found in our separate planning information leaflet for compressed air�

6.1 Installation/setup room

The room chosen for set up must fulfil the fol­lowing requirements:
– Closed, dry, well-ventilated room – Should not be a room made for another pur-
pose (e�g� boiler room or wet cell)
– If the unit is installed in a machine room, e�g�in
an adjoining room or cellar, the requirements set out in ISO-TS22595 must be complied with�
NOTICE Risk of overheating due to insuffi­cient ventilation
The units generates heat� Possibility of heat damage and/or reduced service life of the unit�
i Do not cover the unit� i Install a fan for auxiliary ventilation in
rooms where ambient temperatures exceed 40 °C while the unit is in op­eration�
≥ 40 °C
EN

6.2 Setup

The following conditions must be taken into ac­count for installation:
The air is filtered when it is sucked in� This does not alter the composition of the air� For this reason it is important to keep the sucked-in air free of harmful substances (e�g�do not suck in exhaust gases or contaminated exhaust air)�
– Clean, level and sufficiently stable subsurface
(note the weight of the unit)� – Type plate easy to read� – Unit easy to access for operation and main-
tenance� – Easy-to-access power outlet to which the unit
is connected� – Maintain sufficient distance to the wall (at least
20cm)� – The compressed air pipe should be routed as
closely as possible to the place of installation
(note the length of the hose supplied)�
Assembly

6.3 Information about electrical connections

i Ensure that electrical connections to the
mains power supply are carried out in accord­ance with current valid national and local reg­ulations and standards governing the installa­tion of low voltage units in medical facilities�
i Install an all-pole disconnect switch with a
contact opening width of at least 3mm in the electrical connection to the mains power sup­ply�
i Observe the current consumption of the de-
vices that are to be connected�
EN

7 Transport

WARNING Risk of explosion of the pressure tank and pressure hoses
i The pressure tank and the pressure
hoses must be vented before they are stored or transported�
i Protect the unit against moisture, dirt and ex-
treme temperatures during transport ("4 Tech­nical data")�
i Always make sure that the condensate collec-
tor chamber is empty before transporting the unit ("15 Taking out of use")�
i Always transport the unit in an upright posi-
tion�
i Only transport the unit using the transport
handles provided�
i Check the unit for transport damage�
22 9000-610-45/30 1811V007
Assembly
1 21
8 Quattro Tandem assembly
and installation
For weight reasons the unit is not supplied fully assembled� Instead, the compressor units are installed at the site of use�
i Set up the tank at the planned installation site� i Screw the vibration reducers into the motor
mounting�
i Place the compressor units on the vibration
reducers�
i Attach the compressor units with the lock
washers and nuts�

9 Installation

9.1 Remove the transport locks

For safe transport, the unit is secured with foam padding blocks and a transport strap�
i Cut and remove the transport strap� i Remove the foam padding blocks�

9.2 Establishing the compressed air connection

The supplied flexible pressure hose be­tween the pipe system and the compres­sor prevents vibrations from being trans­mitted and thus reduces noise� This en­sures safe and reliable operation�
i Connect the pre-assembled connecting
sleeve of the pressure hose to the quick re­lease coupling�
i Measure the required length of the pressure
hose and shorten if necessary�
i Press a fitting hose connector (not included in
the scope of delivery) onto the pressure hose (internal diameter 10mm) and secure it with a hose clip�
i Connect the connecting sleeve of the pres-
sure hose to the compressed air tube�
EN
i Insert the compressed air connections from
the compressor unit into the cooler�
i Plug in the electrical connections of the com-
pressor units at the control box� Connect the left-hand unit to the left-hand plug connector and the right-hand unit to the right-hand plug connector�
1 Compressed air connections 2 Electrical connections
Assembly
i Place a collector tray under the condensate
separator or the membrane drying unit (de­pending on type)�
EN
Figure 1: Duo with condensate separator

9.3 Pressure reducer

i Insert the pressure reducer into the quick re-
lease coupling�
i Insert the pressure hose into the quick release
coupling on the pressure reducer�
9.4 Place a collector tray under­neath
During operation, condensation water on the unit is continuously separated and automatically drained� In order to prevent water damage due to drained condensation, it is collected in the collector tray�
As an option, the condensation can also be drained through a hose into the waste water system� Always comply with appli­cable national regulations for waste water systems�

9.5 Electrical connections

Safety when making electrical connections
The unit has no main power switch� For this reason it is important that the unit is be set up in such a way that the plug can be easily accessed and unplugged if re­quired�
i The unit must only be connected to a correct-
ly installed power outlet�
i Make sure that none of the electrical cables
leading to the unit are under any mechanical tension�
i Before initial start-up check that the mains
supply voltage and the voltage stated on the type plate match (see also "4� Technical data")�
Establishing the electrical connections
DANGER Risk of electric shock due to defec­tive mains cable
i Mains cables must not be allowed to
come into contact with any hot sur­faces on the unit�
i Connect the mains plug to an earthed power
outlet�
24 9000-610-45/30 1811V007
Assembly
1
2

10 Commissioning

In many countries technical medical products and electrical devices are sub­ject to regular checks at set intervals� The owner must be instructed accordingly�
i Turn on the unit power switch or the main sur-
gery switch�
i Carry out an electrical safety check in accord-
ance with applicable local regulations (e�g� the German Ordinance on the Installation, Opera­tion and Use of Medical Devices / Medizinpro­dukte-Betreiberverordnung) and record the re­sults as appropriate, e�g� in the technical log book�

10.1 Check the motor protection switch

After installation of the compressor the motor protection switch needs to be checked and ad­justed if it is not set up correctly� It was set to the recommended setting in the factory (see "4 Technical data")�
i Switch on the unit at the pressure switch by
rotating the switch to the "I" position�

10.2 Checking the switch-on/cut-off pressure

The switch-on/cut-off pressure is preset at the factory� Check the setting during first start-up�
i Switch on the unit at the pressure switch by
rotating it to the position "I AUTO"�
i Read off the cut-off pressure from the pres-
sure gauge�
i Drain the air from the pressure tank (e�g� via
the condensate drain valve) until the unit starts and then close it again�
i Read off the pressure when the unit switches
on� If the readings deviate from the values preset
at the factory, adjust the pressure switch to the factory settings�
EN
1 On/off switch 2 Pressure gauge
i Measure the maximum current consumption
(this is the value just before the cut-off pres­sure is reached)�
If the reading deviates from the recommend­ed setting then the motor protection switch needs to be adjusted (see "11�2 Adjusting the motor protection switch")�
Assembly

10.3 Checking the safety valve

Correct operation of the safety valve must be checked when the unit is started up for the first time and again subsequently at regular intervals�
At the factory, the safety valve is set to 10 bar (1 hPa), checked and stamped�
DANGER Risk of explosion of the pressure tank and pressure hoses
i Do not change the safety valve set-
EN
tings�
i Switch on the unit at the pressure switch and
fill the pressure tank to the cut-off pressure�
WARNING Risk of damage to the safety valve
Risk of explosion of the pressure tank and pressure hoses due to a defective safety valve
i Do not use the safety valve to vent the
pressure tank�
i To open, rotate the screw of the safety valve
anti-clockwise until the valve begins to blow off� Only allow the safety valve to blow for a short period�
i Then turn the screw clockwise as far as it will
go to close the valve� The valve must now be closed again�
10.4 Draining the condensation wa­ter
Temperature changes during transport may cause condensation water to accumulate in the pressure tank� The condensation water can only be drained from the pressurised pressure tank�
i Switch on the unit at the pressure switch and
wait until the cut-off pressure is reached�
Pressure tank
i At maximum tank pressure, open the conden-
sate drain valve�
i Close the condensate drain valve as soon as
all of the accumulated condensation water has been blown out�
Condensate separator
i At maximum tank pressure, unscrew the bot-
tom screw fitting on the condensate separa­tor�
26 9000-610-45/30 1811V007
Assembly
+
1.
i Close the screw connection as soon as all of
the condensate has been blown out�
Figure 2: Duo with condensate separator

10.5 Adjusting the rate of flow at the pressure reducer

The pressure reducer regulates the rate of flow in the system and adjusts it to the required op­erating pressure� In order to adjust the rate of flow air needs to be extracted via a consumer�
i Activate the air consumer unit� i Lift the rotary knob at the pressure reducer� i Adjust the rate of flow via the rotary knob�
Turn the knob in the "+" direction to increase the rate of flow�
Turn the knob in the "-" direction to reduce the rate of flow�
i Press in the rotary knob until it engages and
cannot be adjusted�
2.

11 Adjustment options

11.1 Adjusting the pressure switch

WARNING Risk of explosion of the pressure vessel
The pressure vessels used in the com­pressors are designed to withstand continuous pressure changes of 2bar and can be used continuously under these pressure changes�
i For load changes > 2bar (max� per-
missible: 3bar), comply with the maxi­mum load change cycles specified in the operating instructions of the pres­sure vessel�
DANGER Exposed live parts
Risk of electric shock due to live parts
i Disconnect all power from the device� i Use insulated tools� i Do not touch live parts�
The cut-off pressure must be at least 0�5 bar (0�05 hPa) below the maximum pres­sure of 10 bar (1 hPa) of the safety valve� Otherwise the safety valve can open too early, which will prevent the compressor unit from attaining the cut-off pressure, as a result of which it will run continuous­ly� The maximum permitted pressure is marked by a red line on the attached pressure gauge�
If the read-off values differ from the factory set­tings or if other settings are required, the cut-off pressure of the compressor can be adjusted at the adjusting screw on the pressure switch� The start-up pressure can then be adjusted using the pressure difference Δp�
i Take off the pressure switch cover� i Adjust the cut-off pressure P at the adjust-
ment screw� The cut-off pressure increases in the "+"
arrow direction and decreases in the "-" arrow direction� The pressure difference ∆p is also influenced by this adjustment�
EN
-
3.
Assembly
i Adjust the start-up pressure via the pressure
difference ∆p at the adjustment screw� The pressure difference increases in the "+"
arrow direction and decreases in the "-" arrow direction�
The maximum permissible pressure difference must not be set to more than 3bar�
EN

11.2 Adjusting the motor protection switch

Pressure switch
i Take off the pressure switch cover� i Adjust the motor protection switch with the
adjustment screw to the measured value (ob­serve the range between the MIN permissible setting and the MAX permissible setting of the motor protection switch, see "4 Technical data")�
Controller
i Remove the cover from the controller� i Adjust the motor protection switch with the
adjustment screw to the measured value (ob­serve the range between the MIN permissible setting and the MAX permissible setting of the motor protection switch, see "4 Technical data")�
28 9000-610-45/30 1811V007

12 Circuit diagrams

X1 M1
1~
LP
X1 M1
1~

12.1 1/N/PE AC 110-127 V, 230 V layout

135
P>
I >
I >I >
Q1
246
EN
X1 Mains connection L/N/PE AC 230V Q1 Pressure switch M1 Compressor unit M2 Fan motor, membrane drying unit M3 Fan motor, noise insulation (if required)

12.2 3/N/PE AC 400 V layout

Z2/U1U2Z1
M
1~
Assembly
PE
EN
PE
1~
M2 M3
X1 Mains connection 3/N/PE AC 400V Q1 Pressure switch M1 Compressor unit M2 Fan motor, membrane drying unit M3 Fan motor, noise insulation (if required)
L1 L2 L3 PEN
135
P>
I >
I >I >
Q1
246
3~
PE
N
1~
M2 M3
Assembly
Q1
M1
M2 M5

12.3 1/N/PE AC 230 V layout, Duo Tandem

A1
246
135
P>
333 222 111 777
X1 X4
PE PE PE
PE
2211
EN
L1/PE/N
X5
X5 Mains connection L/N/PE AC 230V Q1 Pressure switch A1 Controller X1 Distributor rail X2 Plug connection of compressor unit X3 Plug connection of compressor unit X4 Distributor rail Q1�2 Motor protection switch Q1�3 Motor protection switch M1 Compressor unit M2 Compressor unit M5 Fan motor, membrane drying unit
13
21
Q1.2 Q1.3
246
14 22
I>
I>
I>
135
X2
64
31
PE
3~ 1~
X3
246
I>
135
13
21
14 22
I>
I>
31
64
PE
1~
30 9000-610-45/30 1811V007

12.4 3/N/PE AC 230 V layout, Quattro Tandem

Q1
M1X1 M2 M3 M4
A1
246
135
P>
444555
X5 X4
PE
Assembly
PE
2211
EN
Q2
I> I>
X2
L1/L2/L3/N/PE
X1 Mains connection 3/N/PE AC 230V Q1 Pressure switch A1 Control box X2 Plug connection of compressor unit X3 Plug connection of compressor unit X4 Distributor rail X5 Distributor rail Q2 Motor protection switch Q3 Motor protection switch K1 Time-lag relay M1 Compressor unit M2 Compressor unit M3 Fan motor, membrane drying unit M4 Fan motor, membrane drying unit
246
135
321
3~
14 22
13 21
I>
PE
64
135
A1
K1
A2
246
Q3
X3
I> I>
246
135
321
3~
14 22
13 21
I>
PE
64
1~
1~

Usage

13 Operation

Prior to working on the device or in case of danger, disconnect it from the mains (e�g� pull the mains plug)�

13.1 Switching the unit on/off

i Switch on the unit at the pressure switch by
rotating it to the position "I AUTO"� The compressor unit will start up automati-
EN
cally and fill the pressure tank� When the cut-off pressure is reached the compressor unit switches itself off automatically�
i The unit can be switched off when required by
turning the pressure switch to the "0 OFF" setting�
32 9000-610-45/30 1811V007

14 Maintenance

Prior to working on the device or in case of danger, disconnect it from the mains (e�g� pull the mains plug)�
CAUTION Risk of infection due to burst filters
Particles enter the compressed air network and can therefore enter the mouth of the patient�
i Replace filters in accordance with the maintenance schedule�

14.1 Maintenance schedule

Usage
NOTICE Risk of damage to the unit due to blocked filters
Continuous running due to reduced delivery� Damage to the unit due to burst filters�
i Replace filters in accordance with the maintenance schedule�
Unit with condensate separator
Maintenance interval
At regular intervals i Empty the collector tray under the condensate separator (the interval may
Annually i Replace the air intake filter in the compressor unit – every six months if the
Every 4 years i Replace the vibration dampers� In accordance
with national law
Unit with membrane-drying unit
Maintenance interval
At regular intervals i Empty the collector tray under the membrane drying unit (the interval may
Annually i Replace the air intake filter in the compressor unit – every six months if the
Every 4 years i Replace the vibration dampers� In accordance
with national law
Maintenance work
vary depending on the ambient conditions and method of working; empty it daily if the humidity is high)�
concentration of dust is high�
i Replace the fleece filter in the condensate separator�
i Check the safety valve� i Carry out recurring safety inspections (e�g� pressure tank inspections, electri-
cal safety inspections) in accordance with applicable national laws�
Maintenance work
vary depending on the ambient conditions and method of working; empty it daily if the humidity is high)�
concentration of dust is high�
i Replace the fine or sterile filter� i Replace the sintered filter�
i Check the safety valve� i Carry out recurring safety inspections (e�g� pressure tank inspections, electri-
cal safety inspections) in accordance with applicable national laws�
EN
Usage

14.2 Replacing the air intake filter

i Switch off the compressor at the pressure
switch�
i Pull off the noise reducer from the air intake fil-
ter�
i Remove the air intake filter� i Insert a new air intake filter� i Push on the noise reducer onto the air intake
filter�
1
EN
1 Noise reducer 2 Air intake filter
2
14.3 Replacing the filter in the con­densate separator
i Switch off the compressor at the pressure
switch�
i Unplug the mains plug� i Release the pressure from the pressure tank�
To do this, unscrew the bottom screw fitting
on the condensate separator�
i Push the slider of the housing locking mecha-
nism downwards�
i Open the housing by rotating and pulling
downwards�
i Rotate the black filter mount (approx� 45°)� i Remove the filter mount upwards from the
water accumulation chamber�
45°
i Unscrew the rejected part downwards� i Remove the filter element�
34 9000-610-45/30 1811V007
Usage
i Replace the filter element� i Re-assemble the condensate separator�
During assembly pay attention to de­tents and markings�
14.4 Replacing the filter of the mem­brane drying unit
Fine/sterile filter
i Switch off the unit� i Disconnect all power from the device� i Unscrew and remove the filter cover� i Remove the fine/sterile filter� i Insert the new fine/sterile filter� i Replace the filter cover and close�
Sintered filter
i Unscrew and remove the filter housing� i Remove the sintered filter� i Insert a new sintered filter� i Replace the filter housing and close�
EN
Usage

15 Taking out of use

15.1 Taking the unit out of use

If the unit is not to be used for a prolonged peri­od of time, we recommend that it is properly shut down and taken out of use�
To do this, the accumulated condensation water from the unit must be drained�
i Switch on the unit and wait until the cut-off
pressure is reached�
Condensate separator
i Open the condensate drain valve on the con-
EN
densate separator� Once the start-up pressure has been
reached the compressor will switch on�
i With the compressor switched on and the
condensate drain valve open, wait until no
more condensation water emerges�
i Switch off the unit� i Close the condensate drain valve when no
more air emerges�
i Disconnect the mains plug� i Disconnect the compressed air connection on
the quick release coupling�
Membrane drying unit
i While the compressor unit is running, open
the condensate drain valve on the membrane drying unit� When no more condensation wa­ter emerges, close the condensate drain valve�
i Switch off the device�
Pressure tank
i Open the condensate drain valve�
Once the start-up pressure has been reached the compressor will switch on�
i With the compressor switched on and the
condensate drain valve open, wait until no
more condensation water emerges�
i Switch off the unit� i Close the condensate drain valve when no
more air escapes�
i Disconnect all power from the device�
Figure 3: Duo with condensate separator
36 9000-610-45/30 1811V007
i Disconnect the compressed air connection on
the quick release coupling�

15.2 Storage of the unit

WARNING Risk of explosion of the pressure tank and pressure hoses
i The pressure tank and the pressure
hoses must be vented before they are stored or transported�
i Protect the unit against moisture, dirt and ex-
treme temperatures during transport (refer to the section on "Ambient conditions")�
i Only store the unit when it has been com-
pletely emptied�
Usage
EN

Troubleshooting

16 Tips for operators and service technicians

Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service�
Prior to working on the device or in case of danger, disconnect it from the mains (e�g� pull the mains plug)�
Fault Probable cause Solution Compressor will not start No mains voltage� On three-
EN
phase units: one phase is missing or not connected (generation of a humming sound)
i Check the mains fuse; if neces-
sary, switch the circuit breaker back on� If the fuse is defective, replace it� Check the mains supply voltage�
Humming noise from motor
Compressor does not switch off
Compressor switches on from time to time even though no air is being tak­en for a consumer unit
Undervoltage or overvoltage i Measure the supply voltage; call
Relief valve defective, unit starts against pressure
Mechanical sluggishness of a unit (piston is stuck); motor protection has tripped
Motor capacitor is defective i Replace the capacitor�
Wrong size of compressor, air in­take too high
Leak in the compressed air sys­tem
Defective membrane drying unit i Check whether there is an
Leak in the compressed air sys­tem
an electrician if necessary�
i Check that the relief valve
discharges after switching off the unit� Free up the movement of the relief valve operable or replace it�
i Switch the unit off and discon-
nect it from the power supply, remove the fan hood from the blocked compressor and turn the fan wheel� If this is not possible, replace the piston and cylinder or the complete unit�
i Calculate the air requirement
(this can be up to 50 l/min per treatment unit), if necessary install a larger compressor�
i Locate and seal the leak� i Inform a service technician�
increased flow of air at the filter housing of the membrane drying unit (bottom), if necessary re­place the membrane drying unit�
i Locate and seal the leak� i Inform a service technician�
38 9000-610-45/30 1811V007
Fault Probable cause Solution Knocking or loud noises
on the compressor
Reduced delivery. Com­pressor needs longer to charge the pressure tank, see charging times in "4 Technical data"
Water dripping from air consumers
Compressor unit defective i Disconnect all power from the
Air intake filter dirty i Replace the air intake filter at
Defective membrane drying unit i Replace the membrane drying
Maintenance work not carried out regularly (without membrane dry­ing unit)
Defective membrane drying unit i Inform a service technician�
device and inform a service technician�
least 1x per year� The air intake filter must never be cleaned�
unit�
i Inform a Service Technician� i Regularly drain the condensation
water from the pressure tank, see "10�4 Draining the conden­sation water"
Troubleshooting
EN
Hersteller/Manufacturer:
DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com
Loading...