Durr AZ50 Amalgam Separator Installation and user manual

INSTALLATION AND OPERATING INSTRUCTIONS
1999/12
DÜRR AMALGAM SEPARATOR AZ 50
25
26
CONTENTCONTENT
CONTENT
CONTENTCONTENT
Important information
1.1 CE-marking ...................................... 26
1.2 General Information .........................26
1.3 Electromagnetic Compatibility ........ 26
1.4 General Safety Information .............. 26
1.5 Electric Safety Instructions ..............26
1.6 Warning Information and Symbols .. 27

2. Product information .............................. 27

2.1 Use for the Intended Purpose ......... 27
2.2 Use other than that for the Intended
Purpose ........................................... 27
2.3 Product Description.........................27
3. Overview of models .............................. 28
4. Scope of delivery................................... 28
4.1 Special Accessories ........................28
5. Technical Data ....................................... 29
6. Overview of functions ........................... 30
7. Functional description .......................... 31
7.1 Method of Functioning..................... 31
7.2 Amalgam separation Separation..... 31
7.3 Filling Level Measurement .............. 31
7.4 Equipment Failure ........................... 32
7.5 Service Button ................................. 32
Assembly
8. Installation hints ....................................33
8.1 Installation Room ............................. 33
8.2 Installation Site ................................ 33
8.3 Combination Facilities ..................... 34
8.4 Planning a Pressure Equalising Pipe or
Air Demixer ......................................34
8.5 Hose Material .................................. 34
8.6 Hose Laying .................................... 34
8.7 Odour Trap ......................................35
9. Installation of AZ 50 .............................. 35
9.1 Fixing the AZ 50 in Place ................ 35
9.2 Fixing the Pressure Equalising Pipe in
10. Electrical connection............................. 36
11. Putting into operation ........................... 37
12. Tips for tracing faults............................ 38
Place ................................................35
10.1 Mains Connection via Earthing-
Contact Type Socket ....................... 36
10.2 Fixed Mains Connection ..................36
10.3 Electrical Connection in AZ 50 ........36
10.4 Display Module................................37
11.1 Functional Check ............................. 37
11.2 Visual Check ................................... 37
11.3 Service Button .................................37
11.4 Emergency Program ....................... 37
13. Service-Programm ................................ 40
Functional sequence ........................... 41
13.1 Start .................................................41
13.2 Indicator Test .................................. 41
13.3 Safety Switch ..................................41
13.4 Motor Running and Braking Action 41
13.5 Sediment Scanning ........................41
13.6 Output Signals ................................41
13.7 Stop .................................................41
14. Transport of the AZ 50 ........................ 42
Use
15. Operation................................................43
15.1 Ready for Operation State ...............43
15.2 Amalgam Collecting Vessel
95% full ............................................43
15.3 Amalgam Collecting Vessel
100% full ..........................................43
15.4 Fault ................................................. 44
16. Cleaning and disinfection..................... 44
17. Replacing the amalgam
collecting vessel.................................... 44
17.1 BEFORE replacing the amalgam
collecting vessel, please observe the
following:.......................................... 44
17.2 Replacement of Amalgam Collecting
Vessel .............................................. 44
17.3 Disposal of Amalgam Collecting
Vessel .............................................. 44
18. Maintenance........................................... 45
18.1 Annual Testing of Indicators on
Display Module ................................ 45
18.2 Testing for Proper Working Order -
Every 5 Years...................................46
18.3 Inspection of the Centrifugal Pump -
Every 2 Years...................................46
19. Disposal of the machine .......................46
Spare parts
AZ 50 ....................................................... 48-49
27
IMPORTANT INFORMATION
1. NOTE
1.1 CE-marking
•CE-marking in accordance with the Machinery Directives.
•The product bears the CE-marking in accordance with the directive governing Machinery Directives 89/392/EWG and conforms to the basic requirements of that Directive.
1.2 General Information
• The Installation and Operating Instructions form an integral part of the machine. They must be kept close to the machine and in readiness whenever required. Precise observance of these instructions is a prior condition for use of the machine for the intended purpose and for its correct operation.
• Safety for the operator as well as trouble­free operation of the machine are only ensured if use is made of original equipment parts. Moreover, use may only be made of those accessories that are specified in the technical documentation or that have been expressly approved and released by Dürr Dental for the intended purpose. If and where use is made of accessories or consumer supplies from outside sources, Dürr Dental are unable to assume any guarantee for safe operation or safe functioning.
• No warranty claims are accepted in respect of damage arising from the use of accessories or consumer supplies from outside sources.
• Dürr Dental only regard themselves as being responsible for the machines regarding safety, reliability and proper functioning if assembly, resettings, changes or modifications, extensions and repairs have been carried out by Dürr Dental or an agency authorised by Dürr Dental and if the machine is used in conformity with the Installation and Operating Instructions.
• The Installation and Operating Instructions conform to the relevant version of the machine and the underlying safety standards valid at the time of going to press. All proprietary rights are reserved in respect
of the specified circuitry, methods, names, software programs and equipment.
• Any reprinting of the technical documentation, in whole or in part, is subject to written prior approval of Dürr Dental.
1.3 Electromagnetic Compatibility
EN 60601-1-2 and IEC 1000-4-3 EN Stan­dards governing Electromagnetic Compatibility - Medico-electrical Equipment are fully complied with.
1.4 General Safety Information
• Dispose of the packaging material in the proper manner, and make sure that it is kept out of the reach of children.
• The product is classified as medico­technical equipment and may only be used by such persons as can guarantee that they will handle it properly on account of their training and/or technical knowledge.
• Each time prior to using the machine the user must check to ensure that it is functionally safe and in proper condition.
• The user must be familiar with how to operate the machine.
• The product is not intended for use in areas of rooms employed for medical purposes that are exposed to risk of explosion. Explosion hazards may result from the use of inflammable anaesthetics, skin cleansing agents, oxygen and skin disinfectants. Moreover the machine is only suitable to a limited extent for operation in a combustion­stimulating atmosphere.
1.5 Electric Safety Instructions
• The AZ 50 must be connected up to a properly installed earthing-contact type socket. For fixed connections please observe points 9 and 10 of these instructions.
• Prior to connecting up the machine a check must be carried out to see whether the mains voltage and mains frequency specified on the machine conform to the data applicable to the supply mains.
• No extension cables or leads must be used for power supply purposes.
28
• Prior to being put into operation, the machine and the lines for damage. Damaged lines, socket outlets and plugs must be replaced immediately.
• Never touch the patient and non-protected socket outlets or plugs of the machine at the same time!
need to be inspected
for damage resulting therefrom. All risk is borne solely by the user. Use for the intended purpose includes observance of the Installation and Operating Instructions as well as the installation, operating and maintenance conditions.
The machine must not be used in operating rooms.
1.6 Warning Information and Symbols
In the Installation and Operating Instructions use is made of the following terms or symbols to denote data or information of special importance:
Angaben bzw. Ge- und Verbote zur Information and/or mandatory regulations or prohibitions for the prevention of personal injury or substantial property damage.
Warning of dangerous electrical voltage.
Special information regarding the economical use of the machine and other information
Floating Application Part Type BF
CE Marking without Notified Body Number
Observe Instructions for Use
To protect operating personnel working on the AZ 50, protective gloves must be worn.
2. PRODUCT INFORMATION
2.1 Use for the Intended Purpose
The function of the Dürr Amalgam Separator is to separate and intercept heavy metals and amalgam dust drawn off by dental suction equipment in the form of fillings that have been drilled open.
The amalgam separator is intended for installation of a maximum number of two treatment units in dental surgeries or dental clinics.
Any other use or use beyond what is specified is deemed to be not for the intended purpose. The manufacturer accepts no liability
2.2 Use other than that for the Intended Purpose
Any other use or use beyond what is specified is deemed to be not for the intended purpose. The manufacturer accepts no liability for damage resulting therefrom. All risk is borne solely by the user.
2.3 Product Description
Among other substances, heavy metals and amalgam dust accumulate, drawn off by dental suction equipment in the form of fillings that have been drilled out. Biodegradation of amalgam that has found its way into the waste-water system is a difficult undertaking. To reduce contamination in waste water caused by heavy metal stemming from dental treatment equipment, it is essential - in compliance with the General Sewage Admini­stration Regulations (Annex 50, Dental Treatment) in force in Germany - to install amalgam separators ahead of the dental waste-water drain furnished with a seal of approval issued by the German Institute of Structural Engineering in Berlin. The AZ 50 Amalgam Separator has been designed by Dürr Dental to produce the required separation efficiency factor of over 95% proceeding based on a rate of flow of 8 ltr./min. Amalgam slurry is intercepted in an amalgam collecting vessel underneath the centrifuge drum. Depending on the volume accumulating, the amalgam collecting vessel needs replacing once or twice a year. The AZ 50 used in conjunction with a Dürr Suction Unit VS 300. The AZ 50 employed in connection with water ring pumps.
withoutwithout
without Pump (Model 7112-01) is
withoutwithout
withwith
with Pump (Model 7112-02) is
withwith
29
3. OVERVIEW OF MODELS
4. SCOPE OF DELIVERY
Amalgam Separator
AZ 50 without Pump.......................... 7112-01
Amalgam Separator
AZ 50 with Pump ............................... 7112-02
Amalgam Separator AZ 50
without Pump..................................... 7112-01
Amalgam Separator .................... 7112-100-50
Pressure Equalising Pipe ............ 7112-101-00
Accessories, complete ............... 7112-001-00
Installation and Operating
Instructions AZ 50 .................. 9000-605-95/01
Instructions for Use - Disinfection and Cleaning of Suction Systems 9000-605-10
Amalgam Separator AZ 50
with Pump ..........................................7112-02
Amalgam Separator .................... 7112-100-51
Accessories, complete ............... 7112-001-00
Installation and Operating
Instructions AZ 50 .................. 9000-605-95/01
Instructions for Use - Disinfection and Cleaning of Suction Systems 9000-605-10
4.1 Special Accessories
The parts listed below are scope of delivery. Please order as required.
Base for PAL 14 .......................... 7112-992-00
Base for PAL 50 .......................... 7112-993-00
Hood ............................................ 7112-200-00
Remote Indicator ......................... 7110-994-00
Test Set, complete ...................... 7110-064-00
notnot
not included in the
notnot
30
7112-01
7112-02
5. TECHNICAL DATA
Model 7112-01 without Pump Model 7112-02 with Pump
Electrical Rating
Voltage (V) 230 Frequency (Hz) 50 Max. Current Consumption (A) 1,7
Centrifuge Motor
Voltage (V) 230 Frequency (Hz) 50 Max. Current Consumption (A) 0,95 Rated Speed (1/min) 2780 Rated output (W) 90
Elektronic System
Protective Low Voltage (V) AC/DC24 Current Delivery (A) max.0,7
Rate of Flow
Volume of Secretion (l/min) max. 8
Amalgam Collecting Vessel
Usable Volume (ccm) ca.150 Replacement Interval (Monate) ca. 6-9
A Vakuum-Anschluß, Schlauch D=25mm B Abwasser-Anschluß,Schlauch D=25mm
Dimensions
7112-01: D = 14cm, L = 32,8cm, H = 36cm 7112-02: D = 14cm, L = 38cm, H = 36cm
Temperature Range
Machine in operation ....... +10 to +40 °C
Storage and transport .......... -25 bis +70 °C
Atmospheric Humidity
Machine in operation ...................... max. 80%
Storage and transport ..................... max. 95%
Weight
7112-01 ................................ approx. 7.5 kg
Inspection label
• CE-label
• Institut für Bautechnik, Berlin:
General Structural Supervisory Approval with Inspection label: Z-64.1-13
• Electromagnetic Compatibility Test in
accordance with IEC 1000-4-3 EN V 50 140 regarding high-frequency irradiation 3 V/M corresponds to EN 60 601-1-2.
31
6. OVERVIEW OF FUNCTIONS
32
16 15
14 13
17
7. FUNCTIONAL DESCRIPTION
1 Centrifuge Drum 2 Light Barrier 3 Solenoid Coil 4 Probe 5 Level Pump 6 Amalgam Particles 7 Collecting Vessel 8 Motor
9 Speed Monitoring Unit 10 Motor Fan 11 Safety Switch 12 Centrifugal Pump 13 Yellow LED Indicator 14 Pushbutton 15 Orange LED Indicator 16 Green LED Indicator 17 Display Module
7.1 Method of Functioning
The function of the AZ 50 Amalgam Separator is to provide continuous amalgam separation in all waste water stemming from the treatment units.
During the process of suction removal, the secretion that is drawn off is separated from the suction air in the upstream suction unit.
The secretion that accumulates continuously enters the centrifuge drum (1) in which the particles of amalgam are retained by centrifugal forces. Mounted below the centrifuge drum (1) is a replaceable collecting vessel (7) into which the separated particles of amalgam (6) are flushed after the motor (8) has been switched off. A probe (4) serves to check the filling level inside the collecting vessel (7) and indicates on the display module (17) when the vessel needs changing. A secure turn-lock fastener serves to facilitate replacement and closure of the collecting vessel (7). The compact dimensions of the AZ 50 makes it possible to install the separator close to the treatment units, thus resulting in short secretion-flow channels. Sturdy features of design on the amalgam separator serve to warrant reliable functioning. After the centrifuge drum has been switched off, a braking cycle provides a self-cleaning effect. This caters for extremely quiet running combined with a reliable separation efficiency factor exceeding 95% even when subjected to maximum loading.
7.2 Amalgam separation Separation
Microswitches located in the hose manifold or in the spittoon serve to activate the motor (8) (see wiring diagram). The amalgam­contaminated secretion arriving from the suction unit is retained by centrifugal forces in the centrifuge drum (1) of the AZ 50. Hooking the suction hoses into the hose manifold serves to switch off the motor (8). The motor (8) receives a switch-off signal and immediate braking takes place of the centrifuge drum (1) which is brought to an immediate standstill so that inertia produced by the rotating water ring flushes the separated particles out of the centrifuge drum (1) downwards into the collecting vessel (7). A level pump (5) connected to the centrifuge drum (1) maintains the liquid inside the collecting vessel (7) at a constant level. This prevents any escape of liquid when changing the collecting vessel (7).
7.3 Filling Level Measurement
The filling level in the collecting vessel (7) is checked by a probe (4) each time the main switch is switched on. The filling level is also scanned every 12 hours when the AZ 50 is switched on. The solenoid coil (3) triggers the scanning process, whereupon the probe (4) attracted by gravity sinks under its own weight. The filling level is measured via a light barrier (2) and indicated on the display module (17) when reaching 95%. As soon as the collecting vessel (7) has filled up with amalgam slurry to a level of 95 %, the green LED (16) will be accompanied by a permanent yellow LED (13) appearing on the display module (17). In addition a hooter will sound which can be silenced by briefly pressing pushbutton (14). Afterwards the amalgam separator will be “READY FOR OPERATION” once more. The yellow LED (13) lights up as a reminder to change the vessel. Each time the main switch is switched on these warnings will continue to be displayed until the vessel has been replaced. As soon as the filling level has attained 100 % the green LED (16) will go out and the yellow LED (13) will light up permanently, the orange LED (15) will start flashing and the hooter will sound. In this case, the hooter silenced by pushbutton control. The amalgam separator will then be placed out of service
cannotcannot
cannot be
cannotcannot
33
new collecting vesselnew collecting vessel
until a
new collecting vessel (7) has been
new collecting vesselnew collecting vessel fitted. After the vessel has been replaced the AZ 50 can be put into normal operation once again.
16
7.4 Equipment Failure
15
If failure occurs on the amalgam separator due to a technical defect, a fault alarm will be
14
indicated on the orange LED (15). In addition,
13
a hooter will sound which can be silenced by pressing pushbutton (14). The “FAULT” indicator is triggered for example by a motor (8) defect via a speed monitoring unit (9), defect on the motor fan (10) or by absence of collecting vessel (7) unleashed by a safety switch (11).
7.5 Service Button
If pushbutton (14) is pressed for longer than 2 seconds, the motor (8) will start up. If the suction unit is activated during this process by lifting up the suction hoses, the hoses can be cleaned and disinfected in this way along with the amalgam separator. To do this, switch on the rinsing basin or place the hoses onto the Orocup filled with disinfectant. If the orange LED (15) lights up, the amalgam separator can be started up manually by pressing pushbutton (14) for at least 2 sec. If, in response to pressing pushbutton (14) several times, a fault alarm is registered each time or if the amalgam separator fails to start up after pressing pushbutton (14), this is a sign of a technical defect having occurred.
34
ASSEMBLY
8. INSTALLATION HINTS
8.1 Installation Room
• The room temperature must not be allowed to fall below + 10 °C in winter or rise above + 40 °C in summer.
• Installation in purpose-built rooms, e.g. heating plant room must be clarified beforehand as regards building legislation.
• Installation in wet rooms is not permissible.
8.2 Installation Site
• The Amalgam Separator AZ 50 must be placed directly next to or at a distance not exceeding 30 cm away from the suction unit.
If the spacing distance between the AZ 50 and the suction unit is too great combined with failure to ensure correct hose laying, there is a risk of the secretion not being conveyed as it should.
• If installing the amalgam separator in conjunction with a Dürr Suction Unit VS 300 space needs to be provided for the pressure equalising pipe.
• The amalgam separator can be mounted as follows:
Directly on the wall. Use Mounting Set, Part no. 7801-003-00 (included in accessories).
35
8.3 Combination Facilities
The Amalgam Separator AZ 50 can be combined with the following suction machines:
VS 300
• AZ 50 combined with the Dürr Suction Unit. VS 300
• AZ 50 combined with water ring pumps.
8.4 Planning a Pressure Equalising
Pipe or Air Demixer
To cater for proper pressure equalisation in the suction system between the amalgam separator and the suction unit, a pressure
6
PAL
equalising pipe or air demixer needs to be mounted in conjunction with the following combinations:
7112-017112-01
7112-01 - Pressure equalising pipe from
7112-017112-01 Dürr Dental - to be mounted in conjunction with the Dürr Suction Unit VS 300!
7112-02 7112-02
7112-02 - Air demixers,
7112-02 7112-02 Dental must be mounted with a water ring pump, e.g. Dental EZ!
notnot
not supplied by Dürr
notnot
8.5 Hose Material
In conjunction with the amalgam separator use may only be made of flexible PVC spiral hoses with integral spiral or equivalent hoses.
The following must not be used: Hoses not resistant to dental disinfectants and chemicals as well as rubber hoses or solid PVC hoses
7
that are insufficiently flexible.
36
8
Dental EZ
10cm
100cm
8.6 Hose Laying
Hose laying i.e. between the suction unit or water ring pump and the AZ 50 must proceed observing a slope. The hose length must not be allowed to exceed 30 mm.
Hoses positioned 5050
50 must observe a slope.
5050
upstream or ahead of the AZ 50upstream or ahead of the AZ 50
upstream or ahead of the AZ 50,
upstream or ahead of the AZ 50upstream or ahead of the AZ 50
On the VS 300 the entrance to the amalgam separator must not be higher than the exit of the suction machine, otherwise there will be a risk of liquid and sediment flowing back into the suction machine.
downstream or after the AZdownstream or after the AZ
downstream or after the AZ
downstream or after the AZdownstream or after the AZ
8.7 Odour Trap
The drain on the AZ 50 must be connected via an odour trap by the user.
9. INSTALLATION OF AZ 50
Integration of the Amalgam AZ 50 into existing or new suction drawoff systems:
9.1 Fixing the AZ 50 in Place
• If mains connection proceeds via an earthing-contact type socketearthing-contact type socket
earthing-contact type socket (wired to the
earthing-contact type socketearthing-contact type socket equipment or surgery main switch) the 50 need not be fixed in position50 need not be fixed in position
50 need not be fixed in position.
50 need not be fixed in position50 need not be fixed in position
• Where
• Fix the amalgam separator in a suitable
fixed mains connectionfixed mains connection
fixed mains connection of the AZ 50
fixed mains connectionfixed mains connection takes place on terminal strips via the equipment or surgery main switch the must be fixed in positionmust be fixed in position
must be fixed in position.
must be fixed in positionmust be fixed in position Refer to point 10 “Electrical Connection” for this.
directly next to or at a distance notdirectly next to or at a distance not
place
directly next to or at a distance not
directly next to or at a distance notdirectly next to or at a distance not
exceeding 30 cmexceeding 30 cm
exceeding 30 cm from the suction unit or
exceeding 30 cmexceeding 30 cm water ring pump with a dowel and screw connection made to a shelf, wall or angle frame.
If the spacing distance between the AZ 50 and the suction unit is too great combined with failure to ensure correct hose laying, there is a risk of the secretion not being conveyed as it should.
AZAZ
AZ
AZAZ
AZ 50AZ 50
AZ 50
AZ 50AZ 50
9.2 Fixing the Pressure Equalising Pipe in Place
The pressure equalising pipe as supplied needs to be mounted in the connecting hose between the Dürr Suction Unit VS 300 and the Amalgam Separator AZ 50.
37
10. ELECTRICAL CONNECTION
Electrical connection must proceed via the main switch of the treatment unit or the surgery main switch.
The leads or cables to the machine must be laid without any stress or strain (strain relief clamps).
10.1 Mains Connection via Earthing­Contact Type Socket
Where the mains connection, 230V, proceeds
earthing-contact type socketearthing-contact type socket
via an
earthing-contact type socket, this must
earthing-contact type socketearthing-contact type socket
be wired to the equipment or surgery main switch.
10.2 Fixed Mains Connection
In the case of fixed mains connection, 230V connection proceeds on suitable terminals behind the main switch of the treatment unit via the equipment or surgery main switch.
Line cross section: 0.75 mm
2
10.3 Electrical Connection in AZ 50
Connection of the control contact on the suction unit or water ring pump takes place on the marked terminals inside the terminal box of the AZ 50. Connection of the contact in the hose manifold, dental mouth-rinsing basin valve, position selection valve and flushing system (built into the treatment unit) proceeds via the control cable. 1, 2 and 3 in terminal box of AZ 50.
,,
,
,,
38
10.4 Display Module
The display module should be visible and audible at all times. For that reason, it needs to be mounted in a place where it can be easily observed. A display module is mounted to the AZ 50. If the machine is located in a clearly visible place in the surgery, no further display module will be required. If the separator is not located in the dental surgery, an mounted. Again, this should be located in a place where it can be easily observed, e.g. mounted to the treatment unit. Connection to the AZ 50 proceeds via a 6-core cable and distribution box.
additionaladditional
additional display module can be
additionaladditional
11. PUTTING INTO OPERATION
• Switch on equipment or surgery main switch. As soon as the green LED lights up, the AZ 50 will be ready for operation.
• Carry out functional check and also check the connections for tightness.
11.1 Functional Check
• To perform this check, switch on the equipment or surgery main switch. The sensor will be heard to fall as it proceeds to measure the filling level.
11.2 Visual Check
Check connections, hoses and AZ 50 for tightness, sealing if necessary. For this purpose, proceed to lift up the suction hose and switch on the basin flushing system. This starts up the AZ 50 and the liquid is transported to the centrifuge drum from where it flows into the drain.
11.3 Service Button
Press the service button (14, page 8) for at least 2 sec. This serves to start up the centrifuge motor which will be braked to a standstill after approx. 30 sec.
11.4 Emergency Program
In the event of failure occurring on the AZ 50 for technical reasons, short-time operation of the suction unit will be possible despite this failure. In that case, the secretion drawn off by suction from the patient’s mouth will continue to be pumped into the AZ 50. The AZ 50 will then act as a sediment separator but will need to be repaired as quickly as possible.
39
12. TIPS FOR TRACING FAULTS
Faults Possible Cause
1. No indication on display module, separator not “READY FOR OPERATION”
2. Yellow LED appears on display module, centrifuge motor fails to start up ­Fault
• Main switch on treatment unit or surgery main switch not switched on.
• Cable fault or display module defective.
• Orange LED lights up, safety switch (11, page 30) activated or winding burned out.
• Mains fuse defective.
• Motor jammed.
40
Solution
• Switch main switch of treatment unit or surgery main switch to “ON” position.
• Replace cable or display module.
• Collecting vessel incorrectly inserted or safety switch defective; if necessary replace switch or AZ 50.
• Centrifuge motor defective, remove any particles that may be present in the pump area or centrifuge drum. Replace fuse.
• Check to ensure that the centrifuge motor is running properly, remove any particles that may be present in the pump area or centrifuge drum.
41
13. SERVICE-PROGRAMM
42
FUNCTIONAL SEQUENCE
In order to be able to check whether the AZ 50 is working properly, a service program can be activated. The various program steps are as follows:
• To effect STARTUP press start button on display module while simultaneously switching the main switch to ON.
• Indicator Test: Green, orange and yellow LEDs light up.
• Safety limit switch Switching-on signal for AZ 50.
• Motor run.
• Sediment filling-level measurement.
• Output signal AZ 50 for suction unit: Yellow LED lights up.
• Stop
Pressing the service button continue the sequence of program steps. Pressing the service button repeat a program step. Startup of the service program is indicated by an audible signal tone in addition to the 3 LEDs.
13.1 Start
To start the service program, keep the service button pressed down on the display module and switch on the voltage supply to the AZ 50. As soon as the signal tone is heard, let go of the service button.
13.2 Indicator Test
During the indicator test all three LEDs light up in addition to a signal tone being heard.
13.3 Safety Switch
During this program step various input signals undergo evaluation and are visible on the display module. In addition to LED indication a signal tone will sound. In the normal state, the LEDs are switched off.
twicetwice
twice serves to
twicetwice
onceonce
once serves to
onceonce
13.4 Motor Running and Braking Action
Starting up this program step automatically activates the motor and causes the motor to slow down after approx. 12 sec. This test step can be repeated by pressing the service button once only. In that case the motor running action must be switched off by pressing the service button once more. During the program step Motor Running and Braking Action the various speed ranges are represented on the display module by different LEDs. The no-load speed is set to 2800 RPM.
13.5 Sediment Scanning
This program step serves to check the sediment probe for proper functioning. Each time the service button is pressed, the sediment probe will descend. If use is made
test vesseltest vessel
of a
test vessel, the various filling levels can
test vesseltest vessel be scanned and rendered visible on the display module.
13.6 Output Signals
In the course of this program step, various output signals are activated simultaneously which can then be measured as a voltage on different connectors. The specified voltages are optimum values.
• Yellow LED on display module of AZ 50.
• Control signal for suction unit X 6 contact 1
& 2 has connected through (relay responded).
13.7 Stop
Quit the service program (stop=display off) by switching off the voltage supply on the treatment unit. Pressing the service button twice at the end of the program will return you to the beginning of the program again.
• Open collecting vessel, safety switch is actuated => orange LED lights up.
• Remove suction hose for hose manifold => green LED lights up.
43
14. TRANSPORT OF THE AZ 50
Should it be necessary to move the Amalgam Separator AZ 50 to a different place, the following action needs to be taken prior to transporting the unit:
• Clean and disinfect the suction system.
• Remove the contaminated collecting vessel and replace by a new collecting vessel.
• Do not turn the AZ 50 upside down as otherwise dirt will enter the centrifuge drum and cause congestion.
44
USE
16
15. OPERATION
15
15.1 Ready for Operation State
14 13
Green LED (16) lights up.
15.2 Amalgam Collecting Vessel 95% full
Green LED (16) lights up. and yellow LED (13) lights up. and hooter sounds.
• As soon as the 95% filling level
the hooter can be switched off by pressing pushbutton (14). Afterwards the
green LED (16) - Amalgam Separator “Ready for Operation” will light up.
1
2
• The yellow LED
of the necessity to replace the amalgam collecting vessel.
The amalgam collecting vessel needs to be replaced as soon as a filling level of 95% is attained. Refer to point 17 for this.
(13) lights up as a reminder
is attained,
15.3 Amalgam Collecting Vessel 100% full
Yellow LED (13) lights up. and orange LED (15) starts flashing. and hooter sounds.
• Once the 100% filling level
will no longer be possible to switch off the hooter by pressing pushbutton (14).
• The AZ 50 will not be “Ready for Operati-
on” again until the amalgam collecting vessel has been replaced.
is attained, it
45
15.4 Fault
Orange LED (15) starts flashing and hooter sounds
• The hooter can be silenced by briefly pressing pushbutton (14).
• If pushbutton (14 ) is held down for
than 2 seconds
be started up once more.
• Green LED (16) lights up denoting “Ready for Operation”.
• If, in response to pressing pushbutton (14) several times, a fault alarm is registered each time, this is a sign of a technical defect having occurred.
Call technician. For hints to assist the dental technician in locating the fault, please refer to point 12.
the amalgam separator can
longer
16. CLEANING AND
DISINFECTION
A material-compatible, non-foaming disinfectant for suction systems is prescribed for cleaning and disinfecting the entire suction system including the amalgam separator and the water volume linked to the collecting vessel.
We recommend cleaning and disinfection of the suction system twice a day. Further information will be gathered from the Instructions for Disinfecting and Cleaning Suction Systems, Part no. 9000-605-10.
17. REPLACING THE AMALGAM
COLLECTING VESSEL
17.1 BEFORE replacing the amalgam
collecting vessel, please observe the following:
Replace the amalgam collecting vessel in the MORNING!
We recommend that the amalgam collecting vessel always be replaced in the morning prior to the commencement of work. This serves to prevent liquid from dripping out of the drum during replacement.
17.2 Replacement of Amalgam Collecting Vessel
• Switch off the main switch of the treatment
unit.
If the collecting vessel is unscrewed without the treatment unit having been switched off, the orange LED will start flashing and the hooter will sound.
• Remove the empty collecting vessel from
the carton and screw off the lid.
To reduce the risk of infection, we recommend that liquid-proof gloves be worn while replacing the collecting vessel.
• Hold the collecting vessel attached to the
AZ 50 from underneath, turn in direction of arrow and remove downwards.
• Cut the disinfectant bag open at one of the
corners and pour the contents into the full collecting vessel.
The disinfectant causes acid burns. If coming into contact with the eyes, rinse thoroughly with water and seek medical advice. If coming into contact with the skin, wash off with soap and water.
• Seal the lid on the full
paying careful attention to the markings on the lid. If the lid is properly sealed, these markings should face one another.
• Place the sealed collecting vessel in the
transport carton, seal and secure with adhesive tape to prevent accidental opening.
• Insert the empty collecting vessel in the AZ
50 from underneath and turn in the direction of the arrow as far as it will go. The AZ 50 is now “Ready for Operation” once more.
collecting vessel,
17.3 Disposal of Amalgam Collecting Vessel
Write the name and address of the recycling enterprise on the transport carton and send to authorised disposal company by post or by other means. Relevant national regulations must be observed.
46
18. MAINTENANCE
• Depending on the volume accumulating, the
amalgam collecting vessel will need
16
replacing once or twice a year.
15
14 13
To reduce the risk of infection, we recommend that liquid-proof gloves be worn while replacing the collecting vessel.
18.1 Annual Testing of Indicators on
Display Module
This test must be carried out by suitably trained personnel.
Accessories needed for testing
1 Test vessel
••
Testing the green LED (16)
••
After the main switch is switched on, the green LED should light up denoting that the machine is “Ready for Operation”.
••
Testing the orange LED (15) and
••
pushbutton (14)
- Remove collecting vessel. The orange LED (15) should now light up and the hooter should sound as well. Pressing pushbutton (14) should silence the hooter, while the orange LED will continue to flash. To reduce the risk of infection, we recommend that liquid-proof gloves be worn while replacing the collecting vessel.
• Testing the yellow LED (13)
- Position main switch to OFF
- Remove collecting vessel, insert test vessel, close and set to 100% filling level.
- Position main switch to ON After sediment querying has been carried out, the yellow LED should now light up, the orange LED will respond by flashing and the hooter will sound.
- Position main switch to OFF
- Remove test vessel.
- Insert collecting vessel (7) and close. Remember to record the results of the test in operating log book!
47
18.2 Testing for Proper Working Order - Every 5 Years
This test should be carried out by an inspector every 5 years observing national statutory regulations (in accordance with the General Sewage Handling Regulations, Annex 50, Dental Treatment).
Required tools and equipment:
1 Phillips screwdriver 1 Test vessel 1 Measuring cup
• Remove collecting vessel. During this
process the orange LED on the display module should respond by flashing and the hooter should sound as well. The hooter can be silenced by pressing pushbutton (14, page 23) on the display module.
• Insert test vessel.
• Press pushbutton (14) on the display
module and the green LED denoting “READY FOR OPERATION” should light up once more.
• Turn on dental mouth-rinsing basin and
draw off at least 200 ml water via suction hose. Hook suction hose back in place.
• After the AZ 50 has switched off, remove the
test vessel and pour the water contained in the vessel into a measuring cup.
If contents exceed 70 ml this denotes that the AZ 50 is in proper working order.
An operating log book must be kept by the user covering all maintenance work. See “OPERATING LOG BOOK FOR DÜRR AMALGAM SEPARATOR”, Part no. 9000-605-72
18.3 Inspection of the Centrifugal Pump - Every 2 Years
Part no. 7112-991-00. Test and clean, replacing if necessary.
19. DISPOSAL OF THE MACHINE
Substances including heavy metals and amalgam dust in the form of fillings that have been drilled open are drawn off via the suction hoses. To reduce heavy metal contamination in sewage stemming from dental treatment equipment, the waste water accumulating must be disposed of observing the valid general sewage handling regulations in force in the country of use.
Components contaminated with amalgam, such as strainers, filters and hoses, etc. must likewise be disposed of in accordance with the respective national regulations.
The AZ 50 can be disposed of in the normal way.
The built-in control unit, electronics board and electronic components must be disposed of as electronic scrap.
48
Loading...