Duomo Vent Check Pro User Manual

Keypad (Fig.1)
3
56
4
21
A
A
B
B
B C
Vent Check Pro (CO + CO2)
Operation
Press button 1 to turn the meter on and off.
After a 30 second warm-up the meter is ready to use and will display the normal operating screen (Fig. 2). To sk ip the warm up, press and hold button 6 for 2 seconds. It is advisable, however, to allow the unit to perform its full setup.
During operation the display will automatically cycle through the current level of CO
2
and CO. Press button 5 to see the current air temperature.
Pressing button 6 will cycle through minimum, maximum, average and current readings. In this mode the time at the bottom will change to show you how long youhavebeeninthismode.
Pressing any button will turn the backlight on for 10 seconds.
The unit will turn off automatically after 20 minutes of inactivity.To override this function, turn the unit off then restart it by holding down button 2 then turning it on with button 1.
The meter will take a new reading every second. When the operating conditions change, the unit will respond in 30 seconds for CO
2
.
When battery supply is low, 'Lob' will be displayed. Change the batteries or connect the DC adaptor.
NOTE: Do not hold the meter close to the face as exhalation can affect CO
2
readings.
Normal Operation (Fig.2)
A =Current
measurement type (CO
2
,CO
or TA)
B = Current level C = Current date /
time
Triple reading and display of CO2, CO and temperature NDIR sensor for CO
2
detection Datalogging capability (downloadable via USB) Statistics of weighted averages BS8494 compliant (from 10 - 60 ºC) Backlight for working in dark areas Programmable and audible warning alarms Easy manual fresh air correction on CO
2
PC connection via USB interface With hard case and power adaptor Can be used with rechargeable batteries
w: www.duomo.co.uk - e: sales@duomo.co.uk - t: 01905 797989 - f: 01905 774296 | © Duomo (UK) Ltd. 2007
A
cycle through the minimum, maximum and average readings of each parameter automatically. To return to normal mode,press button 3.
Pressing button 2 will hold all current readings and HOLD flashes in the top left of the screen. Press again to exit hold mode.
Recording (Manual)
The Vent Check Pro can manually record up to 99 points of data.
In normal mode, press button 4 to record a data point. The REC icon flashes on screen for about 3 seconds and the screen displays the data point number (1 - 99). Each point recorded contains information on the current CO
2
,
CO and air temperature.
You can view all manually recorded data points at any time by using button 5. Press and hold it for 2 seconds
V
e
nt
Che
c
kPro
or more until 'Recall' appears on screen.
Press buttons 2 and 4 to cycle up and down through all recorded data points stored on the unit. The data point number will appear on screen first, followed by the stored reading. Pressing button 5 will then cycle through each parameter. The time at which the data was recorded will also appear at the bottom of the screen.
To exit the recall mode, press button 3.
Recording (Automatic)
As well as recording data manually,the Vent Check Pro also allows you to datalog automatically.The unit can store up to 30,000 data points (10,000 data points each for CO
2
,CO and temperature). The sampling rate can be altered in the advanced setup menu and can be set from 1 second to 4 hours, 59 minutes and 59 seconds.
To begin datalogging press button 3 for 2 seconds. The REC icon will flash at the bottom of the screen. Pressing button 3 again for 2 seconds will stop the datalogger. NOTE: Each time you star t the datalogger you will overwrite the previously recorded data.
During datalogging you can still cycle through all three parameters.
Setup
To enter setup, make sure the unit is turned off then press and hold button 1 for more than a second. To exit this at any time,simply press button 3.
There are six setup options within this menu;
Use buttons 2 and 4 to cycle through each of the six options and their parameters and then button 6 to confirm a selection.
CLr
To delete all manually stored data points scroll to 'yes' and confirm your selection. Press button 3 at any time to escape back to the top menu.
ALAr
To set the audible alarm threshold for CO & CO
2
detection, scroll through to the level required and confirm your selection. Available levels range from 15 to 200ppm in 5ppm increments for CO and 400 to 2000ppm in 50ppm increments for CO
2
.
unit
Select whether to display the temperature on screen using degrees centigrade or fahrenheit.
rATE
Press button 6 to start setting the sampling rate. The hours will flash first, use buttons 2 and 4 to set. Pressing button 6 to move on to the minutes and repeat with the seconds. The final push of button 6 will confirm your selection and return to the main menu. The range is from 1 second to 4 hours,59 minutes and 59 seconds.
PrES
If operating at high altitudes the barometric pressure value of this unit may need adjusting from the standard of 1013hPa.(Range is 700 - 1990hPa).
rtC
The first option that appears is whether to use the 24 hour clock or not. Select 24 or 12, confirm your selection then set the date (day, month, year) using buttons 2 and 4, followed by 6 to set it. The time can then be set (hours, minutes,seconds) by the same method.
Fresh Air Correction & Calibration
CO2Fresh Air Correction
It is strongly suggested that the unit be corrected in a sunny, outdoor environment that is well ventilated. Do not correct the unit in places crowded with people or close to areas with high CO
2
concentration such as
ventilating outlets or fireplaces.
To correct the unit for CO
2
make sure that it is turned off and place it in the calibration site. Turn it on by pressing and holding buttons 1, 3 and 4 for 3 seconds.The unit will enter calibration mode. There are three different calibration options available to perform. Select the CO
2
option and pres button 6 to start calibrating. After about 5 minutes the unit will stop flashing and calibration will be completed.
To abort correction at any time,simply turn the meter off.
CAUTION: The meter is corrected at ambient air of around 400ppm. Do not correct the meter in an environment with an unknown CO
2
concentration. Doing
this could lead to inaccurate measurements.
CLr - Clears all manually stored data points ALAr - Determines the CO & CO
2
alarm threshold
unit - Changes the display between
o
C andoF
rATE - Sets the sampling rate for the datalogger PrES - Pressure compensation for CO
2
measurement
rtC - Sets the date and time for the unit
w: www.duomo.co.uk - e: sales@duomo.co.uk - t: 01905 797989 - f: 01905 774296 | © Duomo (UK) Ltd. 2007
CO2Levels & Guidelines
ASHRAE Standard 62-1989: 1000ppm CO
2
concentration in occupied building should not
exceed 1000ppm.
Building Bulletin 101 (BB101): 1500ppm UK standards for schools say that CO
2
averaged over the whole day (i.e. 09:00 - 15:30) should not exceed 1500ppm.
Troubleshooting
Specification
Can't power on
Press button 1 for more than 0.3 seconds and try again. Check whether batteries are in good contact and correct polarity is observed.
Fixed readings
Check whether data hold function was activated.
Slow response
Check whether the air flow channels on the rear are blocked.
Error messages
E01: CO
2
sensor damaged E02:The value is under range E03:The value is over range E04:The original data error results in this error E07: Voltage to measure CO
2
. Replace batteries or use an adaptor E17: Retry CO
2
calibration
E31:Temperature sensor damaged
CO
2
Range of 0 - 9999ppm Accuracy ± 30ppm ±5% of reading Response time <30 seconds
CO
Range of 0 - 1000ppm Accuracy: ± 10ppm at <100ppm
± 10% of reading for 101 - 500ppm ± 20% of reading for 501ppm and above
Response time <60 seconds
Temperature
Range of -20 - 60 ºC (-5 - 140 ºF) Accuracy ± 0.6ºC (0.9 ºF) Response time <2 minutes
Warm Up
30 seconds
Operating Ranges
From0~50ºC , 0 ~ 95% RH (avoid condensation)
Storage
From -20 ~ 60 ºC ,0 ~ 99% RH (avoid condensation)
Power
4 pcs AA batteries (rechargeable option),DC Adaptor
Battery Life
> 10 hours (Alkaline)
w: www.duomo.co.uk - e: sales@duomo.co.uk - t: 01905 797989 - f: 01905 774296 | © Duomo (UK) Ltd. 2007
OSHA: 5000ppm
Time weighted average over five 8-hour work days should not exceed 5000ppm.
NIOSH recommendations:
250 - 350ppm Normal outdoor ambient concentrations.
600ppm Minimal air quality complaints.
1000ppm Indicates adequate ventilation; complaints such as headaches, fatigue and eye/throat irritation will be more widespread. 1000ppm should be used as an upper limit for indoor levels.
CO Levels & Guidelines
ppm Symptoms and applicable standard
0 - 1 Normal background levels
1600 Nausea within 20 min., death within2-3hrs
9
Maximum indoor air quality level:
Maximum allowable concentraion per
ASHRAE Residential standards 62-1989 for
living area
25
Maximum limit 8 hrs of continuous
exposure per California OSHA workplace
standards
35
Maximum 8 hrs average exposure level per
US OSHA workplace standards
50
Maximum concentration for continuous
exposure in any 8 hrs average level per
OSHA standards
100
Remove employees from enclosed space if the CO concentration exceeds 100ppm per
OSHA exposure limit
200
Mild headache, fatigue, nausea and
dizziness within2-3hrs
800
Dizziness, nausea, convulsions, death within
2-3hrs
400
Frontal headache,life threatening after 3
hrs. Maximum concentration in flue gas per
the US EPA and AGA standards
Loading...
+ 5 hidden pages