DUNIWAY Stockroom Terranova 908A Instruction Manual

DUNIWAY STOCKROOM CORP.
Instruction Manual
Terranova
Model 908A Dual Capacitance Diaphragm
Gauge Controller
Copyright © 2000 by Duniway Stockroom Corp.
rev100410sr
1 of 29
DUNIWAY STOCKROOM CORP.
Table of Contents
I Safety Information page 4
II Overview page 5
A. Front View, Back View and Dimensions B. General Description C. Specifications D. Controls and Indicators C. Capacitance Diaphragm Gauge Suppliers
III Installation page 9
A. Unpack the Controller B. Mount the Controller C. Select the CDG D. Connect the CDG E. Attach the CDG Cable F. Make Accessory Connections G. Check Supply Voltage H. Attach the Power Cord
IV Operation page 13
A. Turn Power On B. Front Panel Controls C. Set Pt 1 High D. Set Pt 1 low E. Set Pt 2 High F. Set Pt 2 Low G. Units mTorr/Torr/kTorr -- Bar/mBar/microBar -- Pascal/kPascal -- Arb H. Calibrate I. Full Scale J. Zero K. Reset of Stored (Default) Values L. Set Point Operation M. Reading Pressure N. Analog Output O. Serial Interface P. Pascal/kPascal Pressure Units Only Set-Up (-J Mode)
2 of 29
DUNIWAY STOCKROOM CORP.
V Maintenance page 25
A. Changing Fuses B. Schematic Diagrams
VI Trouble Shooting page 26
VII Return Procedure page 27
VIII Warranty page 28
List of Figures
Figure 1: Model 908A front view page 5 Figure 2: Model 908A rear view page 5 Figure 3: Model 908A dimensions page 5 Figure 4: Sensor Connector -- 9 Pin Female D-Sub page 10 Figure 5: MKS Baratron Pin Connections (15 pin) page 10 Figure 6: Figure 7. Figure 8: Figure 9: Figure 10: Protective Ci
MKS Baratron Pin Connections (5 pin) page 11 MKS Baratron Pin Connections (9 pin ‘D’) page 11 Accessory Connector -- 15 Pin Female D-Sub page 12 Accessory Connector Signals and Pins
page 12
rcuits/Set Point Relays page 17 Figure 11: Choosing Components/Set-Point Circuits page 18 Figure 12 908A Ranges: Sensor, High, Low, Set Points page 19 Figure 13: Analog Output and
Displayed Pressure
page 20 Figure 14: Serial Interface Cables frim PC to 908A page 21 Figure 15: Serial Output and Displayed Pressure page 22 Figure 16: Serial Output and Full Scale Range Setting Figure 17: Torr-Pascal Relationship - Terranova 908J Figure 18: Error Codes p
page 22
page 24
age 26
3 of 29
DUNIWAY STOCKROOM CORP.
I Safety Information
Explosive Gases
WARNING! Do not use the Model 908A Dual Capacitance Diaphragm Gauge Con­troller to measure the pressure of combustible gas mixtures. The gauge normally operates at low temperatures, but it is possible that momen­tary transients or controller malfunction may cause ignition of combus­tible mixtures, which then might explode and cause damage to equipment and injury to personnel.
Limitation on use of Compression Mounts
WARNING! Do not use a compression mount (quick-connect) for attaching the gauge tube to the vacuum system in applications that may develop pos­itive pressures. Positive pressures may cause the tube to be blown out of a compression fitting and damage equipment and injure personnel.
Chemicals
WARNING! Many organic cleaning solvents, such as acetone, produce fumes that are toxic or flammable. Use such solvents only in areas that are well ventilated to the outdoors and away from electronic equipment, open flames, or other potential ignition sources.
Please let us Know...
Terranova products are the most advance instruments of their type available from any manufacturer. We have made this Instruction Manual as complete and clear as possible. Let us know if you have any comments that can make this manual or our products more useful.
4 of 29
DUNIWAY STOCKROOM CORP.
A. Front View, Back View and Dimensions
Figure 1: Model 908A front view
II Overview
Figure 2: Model 908A rear view
Figure 3: Model 908A dimensions
5 of 29
DUNIWAY STOCKROOM CORP.
B. General Description
The Terranova Scientific Model 908A Dual Capacitance Diaphragm Gauge (CDG) Controller displays vacuum pressure as measured from capacitance diaphragm gauges. The 908A supplies ±15 volts at up to 0.75 amp; this is sufficient to operate most heated capacitance diaphragm gaug­es. The 908A precisely measures the 0 to 10 volt signal from the CDG to determine pressure. Model 908A controller covers full scale ranges from 20 mTorr to 10,000 Torr. The 908A is housed in a 1/8 DIN enclosure and is simple to operate.
C. Specifications
1. Useful Measuring Range
4 decades; for CDG tubes with full scale of 20 mTorr to 10.00 ktorr; full scale range selection is entered on the front panel by the user
2. Display Range
-9.9 torr to 10.00 ktorr; pressures higher than 130% of Sensor Full Scale displays HI;
The
3. Display Resolution
varies according to full scale range, from 0.01 mTorr to 1 Torr
4. Gauge Interface
The 908A incorporates a high-resolution input circuit which allows the use of the capacitance di­aphragm gauge over its entire 4 decades.
5. Input to the 908A Controller
0 to 10 volts for full scale of the gauge
6. Units of Display
torr (mtorr), mBar (Bar), pascal (kpascal), or arb (no units); user selectable
7. Full Scale
user selectable range to match CDG FULL SCALE : 20, 50, 100 mTorr; 1, 2, 10, 100, 1000, 5,000 and 10,000 torr
8. Calibration Adjust
for calibration of display; allows user to multiply CDG response by 0.50 to 2.00
9. Vacuum Gauge
one or two capacitance diaphragm gauges which have 1-10 VDC signal output and require up to
0.75 amp total from ± 15 volt supplies; this is sufficient to operate most heated gauges
6 of 29
DUNIWAY STOCKROOM CORP.
10. Operating Temperature Range
+2 to +50 deg. Celsius
11. Process Control Set Points
two, with independent High and Low set points for each relay, for flexible control of hysteresis
12. Process Control Relays
two relays; contacts rated for resistive loads: AC: 120VA: 2.0 Amps @ 60VAC, 1.0 Amps @120 VAC, 0.5 Amps@/240 VAC. DC: 60 Watts: 2.0 Amps @ 30 VDC, 1.0 Amps @ 60 VDC, 0.4 Amps @ 150 VDC.
13. Nonvolatile Memory
for all user specified parameters
14. Analog Output
logarithmic, 0.5 volts/decade; 0.10 mTorr=0.5 volts, displays pressure for selected gauge
15. Output Power
+15 at 0.75 amp and -15 volts at 0.75 amp; sufficient to operate temperature-controlled gauges
16. Mounting
The 908A may be used as a bench-top instrument or it may be mounted in an instrument panel. Clips are provided for panel mounting.
17. RS-232 Input/Output
allows user to read pressure and set points; 9600 baud, 8-N-1; available through the accessory connector
18. Operating Voltage
The Model 908A has a universal power supply, which operates on input voltages from 90 VAC to
VAC 47 to 65 Hz; input is through a standard IEC 320 instrument power input receptacle on
240 the rear panel; input power is protected by fuses in both lines of the input power.
19. Weight, Model 908A only; does not include cable or CDG
0.9 lb. /0.4 kg
D. Controls and Indicators
1. GAUGE SELECT button
allows the user to select which of the two gauges are shown on the digital display
2. SELECT button
Allows selection of parameters to be adjusted, e.g. Set Points
7 of 29
DUNIWAY STOCKROOM CORP.
3. RAISE and LOWER buttons
used for adjustment of gauge and instrument parameters
4. Digital Display
4-digit 7-segment bright red LED, 10 mm high
5. Display Indicators
bright red individual LED for miscellaneous indicators
E. Capacitance Diaphragm Gauge Suppliers
CDGs may be ordered from several sources, including the following:
Duniway Stockroom Corp. 1305 Space Park Way Mountain View, California 94043 USA telephone (650) 969-8811 toll-free (800) 446-8811 facsimile (650) 965-0764 email info@duniway.com internet www.duniway.com
MKS Instruments, Inc. Six Shattuck Rd. Andover, Massachusetts 01810 USA telephone (978) 975-2350 toll-free (800) 227-8766) facsimile (978) 975-0093 internet www.mksinst.com
8 of 29
DUNIWAY STOCKROOM CORP.
III Installation
A. Unpack the Controller
Carefully unpack the Model 908A Dual Capacitance Diaphragm Gauge Controller. The shipment includes these components:
• controller unit
• power cord
• mounting clips, fuses
• D-sub 15 ac
• this instruction manual
cessory connector
If your controller does not have all of these items, call Du
niway Stoc
kroom. If anything appears
to have been damaged in shipment, contact the shipper.
If you need a gauge cable for connection to the Model 908A, please contact Duniway Stockroom.
Do not plug the power cord in yet.
B. Mount the Controller
You can rest the controller unit on a bench, table top, or shelf, or you can mount it in a rack or cabinet. The controller unit is housed in a standard 1/8 DIN box. If you are mounting the unit in a panel, the cutout dimensions are 1.78 inch by 3.60 inch (45.2 mm by 91.4 mm), see fig. 3, page
9. One mounting clip attaches to each of the sides of the controller unit. To attach the clip, slide the beveled surfaces of the clip under the cutout on the side of the box and push the clip toward the back of the unit.
Be sure to leave enough clearance at the back of the controller unit for easy ac
cess to ca
nections.
C. Select the CDG
The Model 908A controller is designed to work with standard capacitance diaphragm gauges which operate on ±15 volts. If you have difficulty obtaining a CDG, please contact us at Duniway Stockroom.
ble con-
CAUTION Use of a CDG other than those which operate from ±15 volts may cause damage to the CDG.
9 of 29