DUNIWAY Stockroom Terranova 906A Instruction Manual

DUNIWAY STOCKROOM CORP.
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: info@duniway.com
www.duniway.com
WWW.DUNIWAY.COM
Instruction Manual
Terranova
Model 906A Convection Gauge Controller
for use with a 275 CONVECTRON® or
MKS/HPS 317 CEP tube
rev082813sr
Copyright (c) 2006 by Duniway Stockroom Corp.
1 of 28
DUNIWAY STOCKROOM CORP.
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: info@duniway.com
www.duniway.com
WWW.DUNIWAY.COM
Table of Contents
I Safety Information page 4
II Front Panel, Rear Panel & Dimensions page 5
III Overview page 6
A. General Description
B. Specifications
IV Installation page 9
A. Unpack the Controller
B. Mount the Controller
C. Select the Gauge Tube
D. Connect the Gauge Tube
E. Attach the Gauge Cable
F. Make Relay Connections
G. Accessory Connector
H. Check Voltage Setting
I. Attach the Power Cord
V Operation page 13
A. Turn Power On
B. Choice of ‘275’ and ‘CEP’ Response Curves
C. Calibration
D. Setting to Zero
E. Setting Near Zero
F. Setting Atmosphere
G. Reset of Stored Values
H. Changing Units of Measurement
I. Set Point Operation
J. Reading the Set Points
2 of 28
DUNIWAY STOCKROOM CORP.
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: info@duniway.com
www.duniway.com
WWW.DUNIWAY.COM
K. Setting the Set Points
L. Analog Output
M. Serial Interface
N. Other Gases
VI Maintenance page 20
A. Changing Fuses
B. Schematic Diagrams
VII Trouble Shooting page 21
VIII Application Note - Set Point Relays page 25
IX Return Procedure page 26
X
Warranty page 27
XI Declaration of Conformity page 28
List of Figures
Figure 1: Model 906A: front view page 5 Figure 2: Model 906A: dimensions page 5 Figure 3: Model 906A, rear view page 5 Figure 4 Model 906A, Sensor Connect Figure 5 Access
ory Connector Pin Assignments page 11 Figure 6 Analog Voltage Output vs. Pressure page 17 Figure 7 S Figure 8 Figure 9
erial RS-232 Output vs. Pressure - Examples page 17 Serial RS-232 Cable page 18 Indicated Pressure for Various Gases page 19
Figure 10 Error Codes p Figure 11 Resistance Values & End View - 275 Sensor page 22 Figure 12 Resistance Valu
es & End View - MKS/HPS CEP page 23
or Pins page 11
age 20
3 of 28
DUNIWAY STOCKROOM CORP.
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: info@duniway.com
www.duniway.com
WWW.DUNIWAY.COM
I Safety Information
Explosive Gases
WARNING! Do not use the Model 906A Convection Gauge Controller to measure the pressure of combustible gas mixtures. The gauge normally operates at low temperatures, but it is possible that momentary transients or controller malfunction can raise the gauge above the ignition tempera­ture of combustible mixtures, which then might explode and cause damage to equipment and injury to personnel.
Limitation on use of Compression Mounts
WARNING! Do not use a compression mount (quick-connect) for attaching the gauge tube to the vacuum system in applications that may develop pos­itive pressures. Positive pressures may cause the tube to be blown out of a compression fitting and damage equipment and injure personnel.
Chemicals
WARNING! Many organic cleaning solvents, such as acetone, produce fumes that
are toxic or flammable. Use such solvents only in areas that are well ventilated to the outdoors and away from electronic equipment, open flames, or other potential ignition sources.
4 of 28
DUNIWAY STOCKROOM CORP.
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: info@duniway.com
www.duniway.com
WWW.DUNIWAY.COM
II Front Panel, Rear Panel & Dim e n s i o n s
Figure 1: Model 906A: front view
Figure 2: Model 906A: dimensions
Figure 3: Model 906A: rear view
CONVECTRON® is a registered trademark of Granville-Phillips Company
5 of 28
DUNIWAY STOCKROOM CORP.
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: info@duniway.com
www.duniway.com
WWW.DUNIWAY.COM
III Overview
A. General Description
The Terranova Scientific Model 906A Convection Gauge Controller displays vacuum pressure as measured from an HPS Model 317 Convection Enhanced Pirani (CEP) or Granville-Phillips
®
Model 275 CONVECTRON conductivity of air/nitrogen or argon. The Model 906A controller covers the range from 0.1 mTorr to 1000 Torr or 0.1 µbar to 1000 mbar, and controls two relays with independent set points.
B. Specifications
1. Useful Measuring Range
0.1 mTorr (µbar) to 995 Torr (mbar), for air or nitrogen; (10-4 torr (mbar) to 103 torr (mbar)); range selection is automatic
gauge tube. It displays vacuum measurements based on thermal
2. Display Range
-19 mTorr (µbar) to 995 torr (mbar); pressures lower than -19 mTorr (µbar) display LO; pressures higher than 995 torr (mbar) display HI; if cable is not plugged in display shows OFF
3. Accuracy of Interface to Gauge
Pressure calculation algorithm is accurate to ±1% of published data for the HPS Convection En-
®
hanced Pirani and the Granville-Phillips CONVECTRON
gauge (this is for the 906A controller
only, and does not include uncertainty of the gauge).
4. Units of Display
torr/mtorr or Bar/mBar; must be specified at time of ordering (see Section V-H for changing Units of Display in the field)
5. Vacuum Gauge
HPS Convection Enhanced Pirani gauge, series 317 or Granville-Phillips CONVECTRON gauge, series 275
6. Operating Temperature Range
®
+2 to +50 deg. Celsius
6 of 28
DUNIWAY STOCKROOM CORP.
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: info@duniway.com
www.duniway.com
WWW.DUNIWAY.COM
7. Pressure Display
3-digit bright red LED, 10 mm high
8. Display Indicators
bright red LED for VAC, ATM, SET PT 1, SET PT 2
9. Display Resolution
varies; from 0.1 mTorr (µbar) below 10 mTorr (µbar), to 5 Torr (mbar) above 100 Torr (mbar)
10. VAC and ATM Adjust
adjusted by front panel push-button
11. Process Control Set Points
two, adjusted by front panel push-button
12. Process Control Relays
two relays, 2 amp, 240 VAC contacts; independent Normally Open and Normally Closed; +5 volts is provided for TTL applications; available through the DB15 accessory connector
13. Nonvolatile Memory
for VAC, ATM and SET POINTS
14. Analog Output
calibrated, 12-bit resolution, logarithmic, 0.50 volts/decade; 0.0 mTorr (µbar) = 0 volts; 10 mTorr (µbar) = 0.50 volts; 100 mTorr (µbar)= 1.00 volts; etc.; available through the DB15 accessory connector
15. RS-232 Input/Output
allows user to read pressure and set points; 9600 baud, 8-N-1; available through the DB15 acces­sory connector
7 of 28
DUNIWAY STOCKROOM CORP.
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: info@duniway.com
www.duniway.com
WWW.DUNIWAY.COM
16. Operating Voltage
The Model 906A has a universal power supply, which operates on input voltages from 100 to 240 V 50/60 Hz or DC; standard IEC 320 power input receptacle on rear panel; replacement fuse type: 5 mm X 20 mm, regular, 1 amp
manufacturer fuse type Bussman GDB-1A or GDC-1A Littlefuse 217 001 or 218 001
17. Weight
1 lb. / 0.5 kg
18. Mounting
Side clips are provided for panel mounting in standard 1/8 DIN cutout.
19. Environmental Considerations
not for use with explosive or corrosive gases
20. Vacuum Gauge Tubes
are available from the following sources:
Duniway Stockroom Corp.
48501 Milmont Drive Fremont, CA 94538
Telephone (800) 446-8811 or (650) 969-8811 FAX (650) 965-0764
HPS Division, MKS Instruments, Inc. 5530 Sterling Drive Bo
ulder, Colorado 80301 Telephone (800) 367-1967 or (303) 449-9861 Facsimile (303) 442-6880
Granville-Phillips Company 5675 Arapahoe Avenue Boulder, Colorado 80303 Telephone (800) 776-6543 or (303) 443-7660 Facsimile (303) 443-2546
8 of 28
DUNIWAY STOCKROOM CORP.
TELEPHONE: 650-969-8811 TOLL-FREE (U.S. only): 800-446-8811 FAX: 650-965-0764 EMAIL: info@duniway.com
www.duniway.com
WWW.DUNIWAY.COM
A. Unpack the Controller
Carefully unpack the Model 906A Convection Gauge Controller. The shipment includes these components:
• controller unit
• power cord
• mounting clips DB15 accessory connector
• this instruction manual
IV Installation
If your contro
ller does not have all of these items, call Duniwa
y Stockroom. If it appears to have
been damaged in shipment, contact the shipper.
B. Mount the Controller
You can mount the controller unit freestanding on a bench, table top, or shelf, or you can mount it in a rack or cabinet. The controller unit is housed in a standard 1/8 DIN box.
If you are mounting the unit in a panel, the cuto ing clip attaches to each of the sides of the controller unit. To attach th
ut dimensions are 44 mm by 92 mm. One mount-
e clip, slide the beveled surfaces of the clip under the cutout on the side of the box and push the clip toward the back of the unit until the central tongue locks the clip in place. Then slide the unit into the panel; the clips will hold the unit in place.
Be sure to leave enough clearance at the back of the controller unit for easy ac
cess to c
nections.
C. Select the Gauge Tube
The Model 906A controller is designed to work with the Granville-Phillips CONVECTRON
275 or HPS/MKS Enhanced Convection Pirani, series 317 tubes. If you have difficulty obtaining a gauge, please contact us at Duniway Stockroom.
able con-
®
, series
CAUTION
Use of a gauge tube other than those listed above may cause improper readings, and may cause damage to the gauge tube.
9 of 28