Notice of Rights:
All rights reserved. No part of this manual including the interior design, cover design and icons
may be reproduced, transmitted or utilized in any form or by any means, electronic, mechanical,
photocopying, recording, or by any information storage and retrieval system, without written
permission from Dukane Corporation.
Notice of Liability:
The information contained in this manual is distributed on an “As is” basis, without warranty.
While every precaution has been taken in the preparation of this manual, Dukane Corporation
shall not have any liability to any person or entity with respect to any liability, loss, or damage
caused or alleged to be caused directly or indirectly by the instructions contained in this manual,
or by the hardware products described herein.
Specications subject to change without notice.
This user’s manual documents product features, hardware, and controls software available
at the time this user's manual was published.
Printed in the United States of America.
Dukane Part Number: 403-575-00
Dukane ultrasonic equipment is manufactured under one or more of the following U.S. Patents:
(* = Inactive)
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
This page intentionally left blank
Page 2
Dukane Manual Part No. 403-575-00
General User Information
Read This Manual First
Before operating your ultrasonic system, read this User’s
Manual to become familiar with the equipment. This will
ensure correct and safe operation. The manual is organized
to allow you to learn how to safely operate this equipment.
The examples given are chosen for their simplicity to illustrate basic operation concepts.
This manual provides information to set up, operate, and
interface this generator/power supply as an integral part of
Dukane’s iQ Series ES press system.
Particular models are listed in Section 10 - Specications
Notes, Cautions and Warnings
Throughout this manual we use NOTES to provide information that is important for the successful application and
understanding of the product. A NOTE block is shown to
the right.
In addition, we use special notices to make you aware
of safety considerations. These are the CAUTION and
WARNING blocks as shown here. They represent increasing levels of important information. These statements help
you to identify and avoid hazards and recognize the consequences. One of three different symbols also accompany
the CAUTION and WARNING blocks to indicate whether
the notice pertains to a condition or practice, an electrical
safety issue or a operator protection issue.
Drawings and Tables
The gures and tables are identied by the section number followed by a sequence number. The sequence num-
ber begins with one in each section. The gures and
tables are numbered separately. The gures use Arabic
sequence numbers (e.g. –1, –2, –3) while the tables use
roman sequence numerals (e.g. –I, –II, –III). As an example, Figure 3–2 would be the second illustration in section three while Table 3–II would be the second table in
section three.
Section 1 - Introduction
NOTE
Note statements provide additional
.
information or highlight procedures.
CAUTION
Caution statements
identify conditions or
practices that could result
in damage to the equipment or other
property.
WARNING
Warning statements
point out conditions or
practices that could result in personal injury or
loss of life.
Condition
or Practice
Electrical
Safety Issue
Operator
Protection
(hearing)
Dukane Manual Part No. 403-575-00
Page 3
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
Press System Overview
An iQ Series Ultrasonic Press System ES consists of
these components: the iQ generator, press (with thruster,
switches, controls, cables, transducer, booster, horn, and
xture) and iQ Explorer software. The software allows
the operator to control the system either remotely or at the
press location.
The iQ Series ES generator has rugged internal ultrasonic
circuitry and ensures a continuous resonant frequency lock
at the start of each weld.
The generator’s compact size allows multiple units to be
placed into an industrial equipment cabinet. This generator will operate at the same international line voltage input
specications as the other generators of this product family (unless the 120V option is installed). It also includes an
RFI line lter that passes FCC and strict CE test specications for global applications.
Key Generator Features
• Digital Control of all power supply functions, includ-
ing digital frequency synthesis, and parameters allows
for unique congurations and allows for future upgrades.
• Process Parameters such as frequency, amplitude
and power are all updated at a 0.5 ms rate.
• Pulse Width Modulation incorporates patented
circuitry giving the power supply the ability to
efciently change the output amplitude. This makes
it possible to start large horns with reduced power. It
also provides more power-efcient switch-mode generator operation and increased reliability.
• Linear Ramp Soft-Start circuitry allows the acoustic
stack to be brought to operating amplitude smoothly,
minimizing start-up surges and abnormal stress to the
stack and generator.
• Digi-Trac Tuning tracks the resonant frequency of
the acoustic stack (horn, booster, transducer) and ad-
justs the generator output frequency to match it. This
is done for every weld cycle and eliminates the need
to manually tune the generator.
• Line Voltage Regulation automatically maintains
constant amplitude regardless of line voltage deviation. The available output power is maintained with
any voltage input within the specied range. This
provides consistent system performance regardless of
line voltage uctuations. It also eliminates the need
for bulky, external constant–voltage transformers.
• Load Regulation provides constant ultrasound
amplitude automatically regardless of power draw.
The ultrasonic output amplitude level is held to
within ±1% to provide weld process consistency
and reduced weld cycle times.
• High Line Voltage Power Supply means that
standard systems will operate worldwide at the local high line voltage level, whether it is 200VAC
@60Hz in Japan, 240VAC @50Hz in Europe or
208VAC @60Hz in the United States. There are no
internal transformer taps to change for worldwide
operation.
• Low Line Voltage Power Supply - This optional
120V power supply is designed for North American applications.
• Flow Through Cooling Tunnel with a matched
high–performance heatsink and thermostatically
controlled fan reduces thermal gradients and increases component life.
• AC Power Inrush protection reduces electri-
cal stress on the internal components by protecting them from AC power startup transient current
surges.
prevent instantaneous component failure in the
event of extreme output overload conditions. The
overload power limit is based on the actual true RMS
power output level.
• Process Limits include: distance, time only, time
and energy, and peak power. These programmable
limits provide the means to adapt to a wide variety
of welding applications.
• Rear Panel Expansion Slot is available to allow
for custom congurations for OEM and cost effec-
tive solutions.
• RS232SerialCongurationPortis used for eld
software upgrades, troubleshooting and advanced
hardware setup with optional PC-based congurator.
• CECertication means that the system meets the
required European standards to be sold and used in
Europe (high line voltage models only).
• ISO9001Certication means that this system has
been manufactured to high quality standards and
assures you of manufacturing excellence.
International Power Grounding .................... 9
Lifting the Equipment .........................10
Dukane Manual Part No. 403-575-00
Page 5
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
This page intentionally left blank
Page 6
Dukane Manual Part No. 403-575-00
General Considerations
Please observe these health and safety recommendations for safe, efcient, and injury-free operation of your
equipment.
Proper Installation - Operate system components only
after they are properly installed.
NoUnauthorizedModications- Do not modify your
system in any way unless authorized to do so by Dukane
Corporation. Unauthorized modications could cause
equipment damage and/or injury to the operator. In addition, unauthorized modications will void equipment
warranty.
Keep the Cover On - Do not remove any equipment
cover unless directed to do so by Dukane Corporation. The
generator produces hazardous electrical voltages which
could cause injury.
Section 2 - Health and Safety
NOTE
These recommendations apply to the
welding system. System in this manual
refers to a complete group of components
associated with the welding of parts, also
known as an ultrasonic assembly system.
A typical iQ Series System consists of
the iQ generator, a press with thruster,
switches, controls, cables, transducer,
booster, horn, and xture, and iQExplorer
software.
WARNING
Grounded Electrical Power - Operate this equipment
only with a grounded electrical connection.
See Electrical Safety for grounding instructions, Page 9.
Comply with Regulations - You may be required to add
accessories to bring the system into compliance with
applicable regulations (OSHA in the USA) for machine
guarding and noise exposure.
Use Eye Protection - Wear ANSI approved safety impact
goggles.
Acoustic Stack Hazard - When an acoustic stack (transducer, booster, horn and tip) is energized by the ultrasound
signal, it presents a potential hazard. Stay clear of an
energized stack.
System Abort Switch - Install a system abort switch at
each operator station when ultrasonic plastic assembly
equipment is used with automatic material handling equipment in an automated system.
Foot Switch - Using a foot switch in place of the optical
Never operate the generator with the cover off. This
is an unsafe practice and
may cause injury.
CAUTION
At some time you may be asked to
remove equipment covers by the
Dukane Service Dept. personnel.
Before doing so, disconnect the
unit electrically from the incoming line AC
power. If the unit is a press/thruster, lock the
Air Lockout Valve, located on the rear panel,
in its closed position.
Dukane Manual Part No. 403-575-00
Continued
Page 7
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
General Considerations
Pre-trigger Switch Adjustment - The pre-trigger switch
option starts the horn vibrating before contacting the part
to be welded. To ensure safe operation, adjust the pre-
trigger so the ultrasound signal will not activate if the
horn is more than ¼ in (7 mm) from the part to be welded.
System Electrical Cabling - Electrical power must be
off when connecting or disconnecting electrical cables.
Do Not Wear Loose Clothing or Jewelry - They can
become caught in moving parts.
Stay Alert - Watch what you are doing at all times. Use
common sense. Do not operate the press when you are
tired or distracted from the job at hand.
Do not Operate the Equipment - Your judgement or
reexes could be impaired while taking prescription medi-
cations. If so, do not operate the equipment. Be familiar
with warning labels and recommended activity restrictions
that accompany your prescription medications. If you have
any doubt, do not operate the equipment.
Continued from previous page
Plastics Health Notice
Certain plastic materials, when being processed, may
emit fumes and/or gases that may be hazardous to the
operator’s health. Proper ventilation of the work station
should be provided where such materials are processed.
Inquiries should be made to the U.S. Department of Labor
concerning OSHA regulations for a particular plastic prior
to processing with Dukane ultrasonic equipment.
CAUTION
Parts being joined ultrasonically will at times vibrate at
audible frequencies. Wear ear
protectors to reduce annoying
or uncomfortable sounds. In
addition, ultrasound bafes, sound enclosures, or materials that absorb sound may
be located to surround the system.
WARNING
Keep head, hands, limbs
and body at least six
inches (152 mm) away
from an operating press/
thruster. A vibrating, descending horn
can cause burns and/or crushing
injuries.
Page 8
CAUTION
When making cable connections to system equipment
or disconnecting cables
from system equipment,
make sure electrical power to the system
is turned off, and AC power cords are
removed from their receptacles. After the
cables have been securely connected
and the connections and cable routing
checked a nal time, the power may be
restored.
Dukane Manual Part No. 403-575-00
Section 2 - Health and Safety
Electrical Safety
Domestic Power Grounding
For safety, the power cords used on this product have a
three-wire, grounding-type power cord. Figure 2-1 illustrates the appropriate electrical outlet to use with the
power cord that is included with systems shipped to North
America.
Approved 2 pole, 3 wire grounding
receptacle HUBBELL No. 5652 or equivalent
to NEMA 6–15R or 6–20R
Figure 2–1Example of 220/240 Volt,
Grounded, 3-Prong Receptacle
International Power Grounding
The power cable normally provided for international use is
compatible with the power outlet used in many Continental European countries (Refer to Figure 2–2.) However,
if your application requires another type of power cord,
check with your equipment supplier, and follow local regulations concerning proper wiring and grounding.
Grounding Contacts
CAUTION
If you have a two-prong electrical receptacle, we strongly
recommend that you replace
qualied electrician replace it following the
National Electric Code and any local codes
and ordinances that apply.
See Figures 2–1 and 2–2.
it with a properly grounded
three-prong type. Have a
CAUTION
If there is any question about
the grounding of your receptacle, have it checked by a
ing prong, or alter the plug in any way.
If an extension cord is needed, use a
three-wire cord that is in good condition. The cord should have an adequate
power rating to do the job safely. It must
be plugged into a grounded receptacle.
Do not use a two-wire extension cord with
this product.
qualied electrician. Do not
cut off the power cord ground-
Typical Outlet
Provided Cable
Figure 2–2 International 220/240V Grounding
Dukane Manual Part No. 403-575-00
100/120 Volt Systems (North America or
Japan)
The power cord (including strain relief) supplied
with the 100/120 AC systems is permanently
attached to the rear of the generator. Units with this
power cord are for use in North America or Japan.
Page 9
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
Lifting the Equipment
High ProleLow Prole
lbkglbkg
Generator
Only
Generator
+ Packing
Materials
Table 2–I iQ Generator Weights
How to Lift Safely
• Before lifting, take a moment to think about what
you’re about to do.
• Examine the object for sharp corners, slippery spots
or other potential hazards. Know your limit and don’t
try to exceed it.
• Ask for help if needed, or if possible, divide the load
to make it lighter.
2511.3209.1
3013.62511.3
CAUTION
Take care when lifting the
equipment. We recommend
using a mechanical lift device
to assist.
• Know where you are going to set the item down, and
make sure it and your path are free of obstructions.
Then follow these steps:
Step 1. Stand close to the load with your feet
spread apart about shoulder width, with
one foot slightly in front of the other for
balance.
Step 2. Squat down bending at the knees (not your
waist). Tuck your chin while keeping your
back as vertical as possible.
Step 3. Get a rm grasp of the object before be-
ginning the lift. Begin slowly lifting with
your LEGS by straightening them. Never
twist your body during this step.
Step 4. Once the lift is complete, keep the object
as close to the body as possible. As the
load’s center of gravity moves away from
the body, there is a dramatic increase in
stress to the lumbar region of the back.
Step 5. If you must turn while carrying the load,
turn using your feet-not your torso. To
place the object below the level of your
waist, follow the same procedures in reverse order. Remember, keep your back
as vertical as possible and bend at the
knees.
Page 10
Dukane Manual Part No. 403-575-00
Section 3 - Installation
SECTION 3
Installation
Before Installation...........................13
When to Use Lockout/Tagout Devices............... 13
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
This page intentionally left blank
Page 12
Dukane Manual Part No. 403-575-00
Before Installation
As you plan for the installation of your generator,
please consider these important subjects as listed
below:
• When to use lockout / tagout devices
• Lifting the generator safely - See Section 2 -
Health and Safety, Page 10.
When to Use Lockout /
Tagout Devices
Section 3 - Installation
Figure 3-1 Lockout Device In Open Position, Unlocked
The typical kind of LOTO device for this generator
is a clam shell type device (with lockout capability).
The LOTO device is placed over the plug end of the
generator electrical cord. This effectively prevents
access to the energy isolation point. See the example
of one such device in the gure above.
The gure to the right shows the lockout device in the
closed, locked position.
Continued
Figure 3-2 Bottom Lockout Device
In Closed Position, Locked
WARNING
Electrical safety hazards
exist inside the generator
chassis. Before making
any internal adjustments
to the generator, apply a
lockout/tagout (LOTO)
device to the generator
chassis.
Dukane Manual Part No. 403-575-00
Page 13
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
Lockout/Tagout
Continued from previous page
Procedure to use BEFORE making any internal
adjustments to the generator:
1. Push the generator's AC power switch/breaker
to the OFF position.
2. Unplug the generator's electrical cord from its
source.
3. Authorized personnel apply a lockout/
tagout (LOTO) device to the plug end of the
generator's electrical cord. Using a typical clam
shell type LOTO device:
1) Open the clam shell.
2) Place the electrical cord plug end inside the
shell.
3) Close the shell.
4) Secure the shell with its lock, and lock it.
4. Wait a minimum of ve minutes for the
generator to discharge its electrical energy.
5. After taking these steps, make the necessary
adjustments to the generator.
Assuming the generator is being put back into
service. . .
Procedure to use AFTER making any internal
adjustments to the generator:
1. Authorized personnel remove the lockout/
tagout device from the plug end of the
generator's electrical cord. Using a typical clam
shell type LOTO device:
1) Unlock the protective shell.
2) Open the shell, exposing the electrical cord
end.
3) Remove the LOTO device, and set it aside.
2. Plug the generator's electrical cord into its AC
power source.
3. Push the generator's AC power switch/breaker
to the ON position.
Page 14
Dukane Manual Part No. 403-575-00
Unpacking
Carefully open your shipping container, and make sure
it contains the items shown on the shipping documents.
Inspect all items, and report any damage immediately.
Section 3 - Installation
Placement
Generator placement and cable routing should permit
easy access and not interfere with normal system operation.
Allow at least 5 inches (13 cm) of space on both ends of
the generator chassis for air circulation. Allow space between 3-6 inches, [8-15 cm] - at the rear of the chassis
for cable clearance. (The amount of space depends on the chassis type and what rear panel connections there are.)
See Section 10 - Specications, for generator drawings.
[435.86]
3.45
[87.63]
EXHAUST AIR
OUTLET
17.16
CAUTION
Allow 5 inches for air ventilation at the cooling air inlet and
at the exhaust air outlet. The
fan draws in fresh air to cool
the internal components, reduce thermal
gradients and increase component life.
3.81
Q
i
[96.8]
COOLING AIR
INLET
IN
[mm]
Figure 3-3 Generator Front View (typical) Low Prole
Continued
Dukane Manual Part No. 403-575-00
OPTIONAL
POWER BUTTON
Page 15
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
Placement in a Seismic Region
If the iQ generator is to be used in an active seismic region,
secure the unit by rack-mounting it or by securing the unit
to a benchtop.
Rack- Mounting
Install the four brackets from Dukane's rack-mount kit to
the generator. See Table 3-I, and Figure 3- 4 (showing a
low prole unit) below.
Mount the generator to a 19-inch equipment rack.
17.16
[435.9]
ALLOW 5"
BOTH
SIDES
FOR
COOLING
12.99
[330.0]
0.12
[3.0]
ALLOW 5"
BOTH
SIDES
FOR
COOLING
ALLOW 3" IN BACK FOR CABLES
Continued from previous page
System TypeDukane Part Number
High Prole147- 4721
Low Prole147- 4720
Table 3- I Rack Mount Bracket Part Numbers
11.37
[288.8]
12.57
[319.3]
18.76
[476.5]
18.16
[461.2]
3.82
[97.0]
Figure 3-4 Rack Mounting iQ Generator
Page 16
Continued
0.81
[20.6]
0.25 X 0.50 SLOT
(2 PLCS)
3.44
[87.4]
OPTIONAL RACK
MOUNT BRACKETS
147-4720
3.25
[82.6]
IN
[mm]
Dukane Manual Part No. 403-575-00
Section 3 - Installation
Benchtop Mounting
Continued from previous page
If you choose to mount the generator on a benchtop follow
these instructions:
1) Install the four (4) optional hold down
brackets. See Figure 3- 5 below.
2) Secure the slot side of each bracket to the
generator's sheet metal cover.
3) Secure the hole side of each bracket to the
bench itself.
18.84
13.20
[335.28]
[478.47]
18.04
[458.22]
ALLOW 3" IN BACK FOR CABLING.
0.280 HOLE
1.14
[29.0]
ALLOW 5" BOTH
SIDES
FOR COOLING.
ALLOW 5" BOTH
SIDES
FOR COOLING.
17.16
[435.86]
IN
[mm]
Figure 3-5 Securing iQ Generator to Benchtop
10.00
[254.0]
12.54
[318.52]
OPTIONAL HOLD DOWN BRACKETS
147-1398
5.19
[131.83]
5.50
[139.7]
1.40
[35.6]
OPTIONAL HOLD
DOWN BRACKETS
Dukane Manual Part No. 403-575-00
Page 17
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
Rear Panel Layout Overview
This section provides an overview of the generator rear
panel layout, which includes panel areas dedicated to various standard system functions and options that are available. Figure 3-6 illustrates a typical panel layout.
A IEC Power Inlet Connector – Attaches to an IEC style
power cord.
B Power Switch – Circuit Breaker – Used to switch
system power ON and OFF.
C Chassis Grounding Stud – Chassis connection for a
protective earth ground.
The System I/O Panel
K
D
J10 RS-232
J11 ENCODER
SYSTEM INPUTS
SYSTEM OUTPUTS
CONFIGURATION
J4
J2
J3
ULTRASOUND
J1
E
F
G
Options Module Panel
H Communications board
Shown in the example above are these ports:
J9 - Ethernet, and
J10 - Serial RS232
K Press interface board.
Shown in the example above are these ports:
J6 - Base/Abort
D System Input Connector (J2) – Connections for sys-
tem control input signals.
E System Output Connector (J3) – Connections for sys-
tem status output signals.
F Ultrasound Output Connector (J1) – Coaxial high
voltage connection to ultrasonic stack.
G Conguration Port Connector (J4) – Digital control
port to modify system parameters.
Page 18
J5 - Thruster
J11 - Encoder
NOTE
The press interface board is available in several different congurations. See Section 8, Options for
more information.
Dukane Manual Part No. 403-575-00
Section 3 - Installation
AC Power Inlet Panel
The standard AC power inlet panel is described in this section.
IEC AC Power Inlet Connector
The IEC AC power inlet connector mounted on the system
AC power inlet panel requires a properly congured IEC
compliant power cord, which enables worldwide system
operation by simply changing the power cord.
Low prole systems are equipped with a 10 amp rated IEC
inlet connector. The high prole systems include a 16/20
amp rated IEC inlet connector. 120 VAC and 3600/4800W
systems include a non-detachable power cord.
An appropriately rated power cord must be securely attached to the welding system’s IEC inlet connector. If the
correct power cord conguration is not included with the
system for the local AC power outlet at your location, an
appropriate IEC power cord should be available from a
local electrical parts supplier. Note that the system undervoltage lockout will inhibit system operation if a North
American power cord congured for 120V is connected to
the system. A minimum of 200V is required for the system
to operate.
Chassis Grounding Stud
The chassis grounding stud is used to attach a protective earth ground to the generator. This will aid in the
suppression of electrical interference or radio frequency
interference (RFI) that is common in a industrial envi-
ronment. The chassis ground stud is C in Figure 3-7.
Proper system grounding is discussed on Page 9.
I
LINE VOLTAGE:
200-240 Vac
0
50/60Hz, 10A
PE
C
Figure 3-7 IEC AC Power Inlet Connector
System I/O Panel
The standard system I/O panel is described in this section.
Power Switch/Circuit Breaker
The power switch/circuit breaker has a rocker type actuator switch that will activate or deactivate the AC power
to the system. The power ON position is marked with the
internationally recognized I symbol, the power OFF position is marked with the 0 symbol. This power switch also
integrates an appropriately sized over-current protection
circuit breaker function in the generator.
If an over-current condition trips the circuit breaker, it will
automatically switch to the OFF position. If the overload
current that caused the circuit breaker to trip is due to a
transient condition, the circuit breaker can be reset by
switching the actuator back to the ON position. If when
resetting the circuit breaker after it has tripped, it immediately trips again, there is likely an internal system malfunction, and the generator will require service.
Do not repeatedly try to reset the circuit breaker. If it trips,
this will only cause more damage to the generator.
System Inputs Connector
The SYSTEM INPUTS connector mounted on the system I/O panel includes connections for all of the basic
system control input signals, that will typically come
from an automated control system. The cable attached to
this connector includes all of the available system control signals, which will be controlled by an output card
or output port on the automation controller.
The user can determine which signals to use for each
particular welding application, but there must be at least
one connection to this connector in order to activate the
ultrasound output. All of the input signals on this connector are electrically isolated (signals are NOT referenced
to chassis ground) and activated when a 24VDC voltage
source is connected to the signal input pin, referenced
to the isolated common pin. The electrically isolated input signals can be driven from an automation controller
output that is either sinking (NPN) or sourcing (PNP),
depending upon how the isolated common connection
is terminated. All inputs sink or source 10mA of current
from a 24VDC power supply.
Continued
Dukane Manual Part No. 403-575-00
Page 19
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
Continued from previous page
Note that a simple switch closure (relay contact) connected to a control input can not activate the input without adding an external power supply to power the input.
Adding jumper connections to pins available on the System Inputs connector, can congure switch closure inputs
to operate referenced to generator chassis ground (nonisolated), without adding a separate power supply, if desired.
Refer to Application Note AN507 (Automation Interface
Guidelines) at http://www.dukane.com/us/DL_ApplDa-ta.asp for wiring diagrams of example applications.
System Inputs Connector Pinout
The SYSTEM INPUTS connector is a HD-15F (high
density D-subminiature 15 circuit female) connector.
Connector pin numbers for this connector are shown in
Figure 3-9. The male connector on the cable is a mirror
image of the panel-mounted connector and is shown in
Figure 3-10. Table 3-II lists the signal names and descriptions, with more detailed descriptions that follow. The
wire color coding for the system input cable is listed in
Table 3-II, to assist with custom automation system wiring and assembly.
Figure 3-8 System I/O Panel (standard panel shown)
NOTE
Refer to Section 8, Options for information
on optional features.
Page 20
Dukane Manual Part No. 403-575-00
1 2 3 4 5
6
10
11 12 13 14 15
Section 3 - Installation
Pin Signal NameCable Color CodeSignal Option Requirements
1+22VBLK
2Power GroundWHT
3Remote Setup Selection Bit 0 InputRED
4Remote Setup Selection Bit 1 InputGRN
5Remote Setup Selection Bit 2 InputORN
6Remote Setup Selection Bit 3 InputBLU
7Remote Setup Selection Bit 4 InputWHT/BLK
8Ultrasound Activation/Cycle Start InputRED/BLK
9Not UsedGRN/BLKNot Used
10Front Panel Control Lock InputORN/BLKNot Used
11Press Inhibit for Hand ProbesBLU/BLKHand Probe
12System Latch Reset InputBLK/WHT
13Isolated CommonRED/WHT
14Not UsedGRN/WHT
15Automation Cycle Stop InputBLU/WHT
Table 3-II Generator Input Signals (J2)
System Inputs Signal Descriptions
Pin 1 (+22V)
This pin can supply +22VDC at up to 250mA to power
the user’s automation controls.
Pin 2 (Power Gnd)
Pin 2 is the 22VDC return and is tied to the system chassis ground.
Pin 3 (Remote Setup Selection Bit 0 Input)
Pin 3 is the Remote Setup Selection Bit 0, which is the
least signicant bit used to select different welding setups
with an automation control system.
Pin 4 (Remote Setup Selection Bit 1 Input)
Pin 4 is the Remote Setup Selection Bit 1, which is the
second least signicant bit used to select different welding setups with an automation control system.
5 4 3 2 1
5 4 3 2 1
10
10
15 14 13 12 11
15 14 13 12 11
Figure 3-9HD-15F, Generator Input Connector
Figure 3-10 HD-15M, Generator Input Cable
Connector
6
6
Dukane Manual Part No. 403-575-00
Continued
Page 21
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
.
Continued from previous page
Pin 5 (Remote Setup Selection Bit 2 Input)
Pin 5 is the Remote Setup Selection Bit 2, which is the
third least signicant bit used to select different welding
setups with an automation control system.
Pin 6 (Remote Setup Selection Bit 3 Input)
Pin 6 is the Remote Setup Selection Bit 3, which is the
second most signicant bit used to select different welding setups with an automation control system.
Pin 7 (Remote Setup Selection Bit 4 Input)
Pin 7 is the Remote Setup Selection Bit 4, which is the
most signicant bit used to select different welding setups
with an automation control system.
Pin 8(Ultrasound Activation/ Cycle Start
Input)
Pin 8 is used to activate the generator ultrasound output.
Activation of this control input will switch the ultrasound
output ON, and deactivating this signal will switch ultrasound OFF. This input signal will also function as a cycle
start input, where the ultrasound activation and timing
are completely under the control of the process controller. Depending on the welding process controller setup,
this input signal could be activated momentarily to start
a welding cycle.
Pin 9(Not Used)
Pin 10(Not Used)
Pin 11 (Press Inhibit for Hand Probes)
Pin 11 is used to disconnect power applied to a press or
thruster, if a hand probe is connected to the system input
connector, for safety considerations. The hand probe activation switch could unexpectedly start a welding cycle that
activates a press or thruster to the down position. The hand
probe cable connector is wired to apply chassis ground to
this pin, when it is attached to the system, which activates
a press inhibit relay that disconnects power from the pneumatic press valves. This pin must be left open whenever
a press control board is installed. Connecting this pin to
chassis ground will inhibit press operation.
Pin 12(System Latch Reset Input)
Pin 12 is used to reset the Any Fault or System Overload
status outputs (See Status Output descriptions.). If a fault
occurs during a weld cycle, these outputs will normally
remain active until the next weld cycle is initiated. Activating this input will reset the status output faults and may
simplify automation programming.
Pin 13(Isolated Common)
Pin 13 is electrically isolated from chassis ground. Using
isolated sourcing (PNP) output drivers, this common line
would be connected to isolated ground potential. Using
isolated sinking (NPN) output drivers, this common line
would be connected to the isolated positive supply voltage
output.
Pin 14 (Not Used)
Pin 14 is an open connection.
Pin 15 (Automation Cycle Stop Input)
Pin 15 is an input control signal that when enabled, can
be used by the automation control system as a redundant
signal to shut the ultrasound output off. This signal could
also be recongured through menu selections to function
as an automation end-of-weld control signal input.
Page 22
Dukane Manual Part No. 403-575-00
Section 3 - Installation
Pin Signal NameCable Color CodeSignal Option Requirements
1+22VBLK
2Spare OutputWHT
3+22V Power GroundRED
4Programmable Status Output 1GRN
5Programmable Status Output 2ORN
6Ultrasound Active Status OutputBLU
7Any Fault Status OutputWHT/BLK
8Press Trigger Status OutputRED/BLK
9System Overload Status OutputGRN/BLK
10System Online Status OutputORN/BLK
11Press Top of Stroke Status OutputBLU/BLK
12Current Loop OK Status OutputBLK/WHTNot Used
13Analog Monitor Signal CommonRED/WHT
14Not UsedGRN/WHT
15Power Signal Monitor OutputBLU/WHT
16Amplitude Monitor OutputBLK/RED
17Amplitude/Power Regulation Status OutputWHT/RED
18MPC Ready Status OutputORN/REDNot Used
19System Power OK Status OutputBLU/RED
20Bad Part Status OutputRED/GRN
21Good Part Status OutputORN/GRN
22System Ready Status OutputBLK/WHT/RED
23Suspect Part Status OutputWHT/BLK/RED
24Isolated CommonRED/BLK/WHT
25Not UsedGRN/BLK/WHT
Table 3-III System Output Connector Signals (J3)
System Outputs Connector
The SYSTEM OUTPUTS connector mounted on the generator I/O panel includes connections for all of the basic
system status and monitor output signals, which will typically connect to an automated control system. The cable
attached to this connector includes all of the available
system output signals, which will be read or monitored by
a digital input card or analog inputs on the user-supplied
automation controller.
The user can determine which signals are appropriate for
each welding application.
The system monitor output signals are analog signals used
to monitor ultrasonic amplitude setting and ultrasonic output
power levels, referenced to the Monitor Common (Pin 13).
iQ Series Ultrasonic Generator/Power Supply ES User’s Manual
Continued from previous page
This is at system chassis ground potential (non-isolated).
All of the digital output status signals on this connector,
are isolated (signals are not referenced to generator chassis
ground). When a status output signal is activated, it will
sink current (500mA Max. sourced by a 24VDC supply)
to isolated common. In automation terms, the outputs are
NPN (sinking) and would drive a PNP (sourcing) input
that is referenced to the Isolated Common pin.
The digital status output signals can only sink current to
isolated common. They cannot be changed to a sourcing
(PNP) type output. If sourcing status outputs are required,
there is an optional isolated NPN to PNP conversion
module that is DIN rail mountable. The module converts
the standard isolated sinking (NPN) status output signals
to isolated sourcing (PNP) status output signals. Consult
your local sales representative for information about the
conversion module.
System Outputs Connector Pinout
The SYSTEM OUTPUTS connector is a DB-25F (standard
D-subminiature 25 circuit female) connector. Connector
pin numbers for this connector are shown in Figure 3-11.
The male connector on the cable is a mirror image of the
panel mounted connector and is shown in Figure 3-12.
Table 3-III lists the signal names. Detailed descriptions
are listed in System Outputs Signal Descriptions that follow. To assist with custom automation system wiring and
assembly, the wire color coding for the system outputs
cable is listed in Table 3-III.
System Outputs Signal
Descriptions
Pin 1(+22V Power Supply)
This pin can supply +22VDC at up to 250mA to power
the user’s automation controls.
Pin 2 (Not Used)
Pin 2 is an open connection.
Pin 3(+22V Power Ground)
Pin 3 is the 22VDC return and is tied to the system
chassis ground.
Pin 4(Programmable Status Output 1)
Pin 4 is a digital active low status output that can be
reprogrammed and assigned to other system status signals (from the available selections) from the System
tab in iQExplorer.
Pin 5 (Programmable Status Output 2)
Pin 5 is a digital active low status output that can
be reprogrammed and assigned to another system
status signal (from the available selections) from the
System tab in iQExplorer.
Pin 6(Ultrasound Active Status Output)
Pin 6 is a digital active low status output that activates
when the system is delivering ultrasonic power to the
load attached to the ultrasound output connector. This
output will be an open circuit when the ultrasound
output is off.
Pin 7(Any Fault Status Output)
Pin 7 is a digital active low status output that activates
whenever any fault condition is detected that inhibits
ultrasound output and normal system operation. This
output will be an open circuit when no system fault
conditions are detected.
Generator faults that will activate the Any Fault output:
• Overload (Average, Peak, or Frequency)
• Overtemperature Fault
• System Power Fault
Pin 8(Press Trigger Status Output)
Page 24
Pin 8 is a digital active low status output that activates
when the specied trigger type has occurred. It will
remain active until the thruster head raises to the top
of the stroke. This output will be an open circuit when
a trigger condition hasn’t been detected.
Dukane Manual Part No. 403-575-00
Continued
Loading...
+ 106 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.