This document contains information in which Dräger claimed proprietary
rights. The information may not be reproduced in whole or in part except as
authorized in writing by Dräger. This information is the property of Dräger;
it is provided solely for the use intended.
Repairs/Modifications
Repairs on this device shall be performed only by DrägerService, a division
of North American Dräger (NAD), or its Authorized Service Centers.
Information about repairs can be obtained from Dräger or Authorized
Dealers. NAD or Dräger will not be responsible for injury to persons or
damage to property arising directly or indirectly out of unauthorized repairs
or modifications to this device. Furthermore, any unauthorized repairs or
modifications void any warranty extended by Dräger.
This document is provided for your information only. It will not be
exchanged or updated without request.
Trademarks
The Dräger name and logo and the name Babytherm® are registered
trademarks of Dräger Medizintechnik GmbH.
For correct and effective use of the product and in order to minimize
hazards it is mandatory to carefully read and to observe all portions of this
manual.
The design of the equipment, the accompanying literature, and the labeling
on the equipment take into consideration that the purchase and use of the
equipment are restricted to trained professionals, and that certain inherent
characteristics of the equipment are known to the trained operator.
Instructions, warnings, and caution statements are limited, therefore,
largely to the specifics of the Dräger design. This publication excludes
references to various hazards which are obvious to a medical professional
and operator of this equipment, to the consequences of product misuse, and
to potentially adverse effects in infants with abnormal conditions. Product
modification or misuse can be dangerous. Dräger disclaims all liability for
the consequences of product alterations or modifications, as well as for the
consequences which might result from the combination of this product with
other products whether supplied by Dräger or by other manufacturers if
such a combination is not endorsed by Dräger.
Introduction
Introduction
The operators of the heated bassinet system must recognize their
responsibility for choosing appropriate safety monitoring that supplies
adequate information on equipment performance and infant condition.
Infant safety may be achieved through a wide variety of different means
ranging from electronic surveillance of equipment performance and infant
condition to simple, direct observation of clinical signs. The responsibility
for the selection of the best level of infant monitoring lies solely with the
equipment user.
Limitation of Liability
Dräger's liability, whether arising out of or related to manufacture and sale
of the goods, their installation, demonstration, sales representation, use,
performance, or otherwise, including any liability based upon Dräger's
Product Warranty, is subject to and limited to the exclusive terms and
conditions as set forth, whether based upon breach of warranty or any other
cause of action whatsoever, regardless of any fault attributable to Dräger
and regardless of the form of action (including, without limitation, breach of
warranty, negligence, strict liability, or otherwise).
1
Introduction
Dräger shall not be liable for, nor shall buyer be entitled to recover any
special incidental, or consequential damages or for any liability incurred by
buyer to any third party in any way arising out of or relating to the goods.
Limited Warranty
All Dräger products are guaranteed to be free of defects in material and
workmanship for a period of one year from date of delivery.
The following are exceptions to this warranty:
1. Defects caused by misuse, mishandling, tampering, or by modifications
THE STATED EXPRESSED WARRANTlES ARE IN
LlEU OF ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING, WITHOUT LIMITATION,
WARRANTlES OF MERCHANTABILITY, FITNESS
FOR ANY PARTICULAR PURPOSE, OR
NONINFRINGEMENT.
not authorized by Dräger or its representatives are not covered.
2. All consumable and disposable products are guaranteed to be free from
defects upon shipment only.
Any product which proves to be defective in workmanship or material will
be replaced, credited, or repaired with Dräger holding the option. Dräger is
not responsible for deterioration, wear, or abuse. In any case, Dräger will
not be liable beyond the original selling price.
This warranty is rendered void and Dräger shall not be held liable for
condition as resultant from:
1. Damage to the unit is incurred as a result of mishandling.
2. The customer fails to maintain the unit in a proper manner.
3. The customer uses any parts, accessories or fittings not specified or sold
by Dräger.
4. Sale or service is not performed by an authorized service/dealer.
Application of this warranty is subject to the following conditions:
1. Dräger or its authorized representative must be promptly notified, in
writing, upon detection of the defective material or equipment.
2. Contact Dräger Customer Service at 1-800-4DRAGER for a Return
Authorization Number.
3. Defective material or equipment must be returned, shipping prepaid, to
Dräger or its authorized representative.
4. Examination by Dräger or its authorized representative must confirm
that the defect is covered by the terms of this warranty.
2
Introduction
5. Notification in writing, of defective material or equipment must be
received by Dräger or its authorized representative no later than two (2)
weeks following expiration of this warranty.
The above is the sole warranty provided by Dräger. No other warranty
expressed or implied is intended. Representatives of Dräger are not
authorized to modify the terms of this warranty.
Dräger, Telford, PA
3
Introduction
g
Definitions and Symbols
Warnings
Cautions
Notes
Terminology
A WARNING statement refers to conditions with a possibility of personal
injury or death if disregarded.
A CAUTION statement designates the possibility of damage to equipment
or improper operation if disregarded.
A NOTE provides additional information intended to clarify apparently
contradictory or confusing situations.
Inspectionexamination of actual condition
Servicemeasures to maintain specified condition
Repairmeasures to restore specified condition
Maintenanceinspection, service, and repair, where
necessary
Preventive
Maintenance
Maintenance measures at regular
intervals
Typing conventions in this manual:
LED messages are printed in bold type:
Symbols
Example:
CAUTION
!
ATTENTION
SEt
:
:
Refer to accompanying documents before
operatin
Consulter l es documents ci -joints avant de faire
fonctionner l’appareil.
Denotes IEC601-1 type of applied part.
Protective earth symbol.
equipment.
4
General WARNINGS and CAUTIONS
Warning: Strictly follow this Operator's Instruction Manual. Any use of
the product requires full understanding and strict
observation of all portions of these Instructions. The
equipment is only to be used for the purpose specified under
"Intended Use." Observe all WARNINGS and CAUTIONS as
rendered throughout this manual and on labels on the
equipment.
Warning: This device is to be used only in rooms with line power
installations complying with national safety standards for
hospital infant rooms. To maintain grounding integrity,
connect only to a "hospital grade" receptacle. Always
disconnect power before servicing.
Warning: Do not use in the presence of flammable anesthetics. Turn off
and disconnect oxygen supply when performing cleaning and
service. Keep matches, lighted cigarettes, and all other
sources of ignition out of the room in which the Babytherm is
located. Small quantities of flammable agents, such as ethyls
and alcohol, can cause a fire in connection with oxygen left in
the Babytherm in the presence of ignition sources.
Introduction
Warning: Do not use cellular phones or similar wireless communication
equipment within 33 feet (10 m) of the Babytherm 8000 LDR.
Wireless phones may cause malfunction in electromedical
equipment.
Caution: Restriction of Distribution
: Federal Law restricts this device
to sale by or on the order of a physician.
Caution: Maintenance
your local DrägerService
: In case of malfunction of this device, contact
SM
technician or our Factory
Authorized Technical Service Center. The device must be
inspected by competent and factory authorized technical
service representatives at regular intervals. A record must be
kept on this preventive maintenance. Maintenance or repair
of the Babytherm 8000 LDR shall be performed only by
Dräger authorized technical service representatives.
Warning: Never operate the heated bassinet if it has suffered physical
damage or does not seem to operate properly. In this case,
always refer servicing to properly trained and factory
authorized service personnel.
5
Introduction
Precautions During Preparation
Warning: Use of gel mattress when mattress heater is switched off will
result in infant becoming hypothermic. Use foam rubber
mattress when mattress heater is not in use for extended
periods (> 15 min.).
Caution: Never route infant circuits or lines over top of panels – they
may get pinched when folding up panels or when installing
cover.
Precautions During Operation
Warning: To prevent infant fall, do not leave infant unattended when
any panels are down. All panels must be erect and secure
during transport to prevent infant fall.
Warning: When folding up surrounding panels, take care not to pinch
infant body parts, circuits, or other items, such as sheets.
Warning: Temperature Monitoring
skin condition at regular intervals and adjust temperature
setting to individual infant needs. Infant temperature must
be monitored particularly carefully during phototherapy or
while a radiant warmer is used in conjunction with the
Babytherm heated bassinet.
Warning: Mattress Heater
mattress must be placed directly on the heated tray surface.
Always take into account warm-up time of the mattress
(approximately 1 hour) before placing an infant in the
bassinet.
Warning: To minimize infant overheating, always consider effect of
additional heat sources, such as phototherapy lamps or
radiant warmers. They increase the heat directed to the
infant. Avoid direct sunlight.
Warning: Do not obstruct conductive heat transfer by placing
additional blankets or other insulating covers underneath the
infant.
Warning: Set temperatures in the lower extended temperature range
only on the order of a physician. Check the infant's
temperature and skin condition at regular intervals.
: When mattress heating is used, the gel
: Check the infant’s temperature and
Warning: Check the infant's temperature and skin condition at regular
intervals.
6
Introduction
Warning: Always use a proper fitting thin sheet (Part No. 2M21272).
Do not fold the sheet under the gel mattress more than 3". To
prevent infant fall in Trendelenberg or reverse Trendelenberg
position always use proper fitting mattress sheet.
Warning: When used with an infant radiant warmer, check its
operator’s manual for directions of the correct heat level and
distance to the infant.
Warning: Use of radiant warmers may increase insensible water loss.
Measures to maintain desired fluid levels should be
considered.
Warning: Use of Radiant Warmer
recommended. Dräger cannot warrant or endorse the safe
performance of third party radiant warmers for use with the
Babytherm 8000 LDR. Specifically, combinations with
radiant warmers that are not intended for use with heated
bassinets bear risks of excessive overall heat transfer to the
infant.
Caution: Remove from mattress all objects with sharp edges. Do not
fold gel mattress. The gel mattress may become damaged.
Precautions During Care
Warning: Always follow accepted hospital procedures for handling
equipment contaminated with body fluids.
Warning: Always disconnect AC line cord before cleaning and
disinfecting.
Caution: Do not launder mattress in an automatic washing machine.
Do not autoclave mattress.
Precautions During Maintenance
Warning: To minimize the risk of infection, clean and disinfect bassinet
and accessories before any maintenance according to
established hospital procedures – this applies also when
returning units or parts for repair.
: Use of the Dräger RH600 warmer is
Warning: Maintenance work on the Babytherm 8000 LDR may be
performed by trained and factory authorized staff only.
Warning: Never operate the heated bassinet if it has suffered physical
damage or does not seem to operate properly. In this case,
always refer servicing to properly trained and factory
authorized service personnel.
7
Introduction
Intended Use
Babytherm 8000 LDR is a heated bassinet intended as a therapeutic device
providing warmth, with or without the use of a canopy, for premature and
full-term neonates and infants up to 16 lbs.
8
Mattress Heating
The Babytherm 8000 LDR features a mattress heating system that uses
conductive heat to warm up infants and to keep them warm.
The system consists of a gel mattress, a heating plate and an electronic
control and monitoring unit. The gel is enclosed by a soft, skin compatible
film.
The mattress molds itself to the body contours of the infant lying on it,
resulting in a large contact surface, a property that helps promote thermal
heat transfer to the infant and minimizes pressure points. A cold gel
mattress, on the other hand, dissipates infant thermal energy. The gel
material has an excellent capacity for storing heat. When the power supply
is disconnected (e. g. during transport within the hospital), the infant is
kept warm for approximately 15 minutes if sufficiently insulated (blanket,
cover).
Description
Description
Features
The heating plate consists of an aluminum plate with a heater element
attached to its under side. It distributes heat evenly over the entire bed
area.
The control and monitoring unit is located under the heating plate and faces
the side most frequently accessed by caregivers.
The micro-processor controlled system continuously monitors the status of
the electronic temperature control and alarm system, and alerts the user if
any faults are detected.
With its ergonomic design including:
•foldable side panels for increased access to the infant
•tilting bed for Trendelenberg and reverse Trendelenberg position
•pop-up side table – integrated hand rail
•cable management system
•storage space in the cabinet
•5 inch diameter castors, two with locks
•optional canopy
it serves an important function for newborn babies from the first minute
after birth until discharge to home.
9