Dräger Pac 3000 Instruction Manual

D
Pac
®
3000 CO, H
2
S, O
2
Instructions for Use
ϝΎϤόΘγϻ΍ ΔϘϳήσ
1 For Your Safety
The use of the Dräger Pac 3000 instruments assumes a complete
knowledge and adherence to the users manual.
— The Dräger Pac 3000 may only be used in areas subject to
explosion hazards which are explicitly covered under the Ex Approvals which have been given to the Dräger Pac 3000.
— Dräger Pac 3000 is not for use in oxygen-enriched
atmospheres. — Please check calibration before safety relevant use. — The performing of calibration and bump testing shall be
conducted according to local regulations. — The gas opening is equipped with a dust and water filter.
This filter protects the sensor against dust and water.
Do not destroy the filter. Replace destroyed or clogged filter
immediately. — Dräger Safety cannot be held responsible for damages incurred
when the above guidelines are not explicitly adhered to. The
warrenty and liability provisions of the terms of sale and delivery
of Dräger Safety are not modified by the recommendations given
above.
2 Intended Use
— Personal gas alarm in the workplace.
3 What is what
4 Standard configuration
1)
1) Please be aware of special settings by customer requirements.
1 Alarm LED 6 Gas Opening 2 Horn 7 Screw 3 Remaining Display 8 Clip 4 OK Key On/Off/Alarm Acknowledge 9 Label 5 + Key Off/Bump Test 10 IR interface
CO H2S O
2
Measuring Range 0 to 500
ppm
0 to 100
ppm
0 to 25
vol.-%
Vibrating alarm Yes Yes Yes
1
6
5
2
3
9
4
7
7
1
7
8
10
7
9
St-5056/5053-2004.eps
2) For O2 A1 is the lower alarm threshold, used to indicate oxygen deficiency
5 Operation
5.1 Remaining life of the instrument
— Once activated check the remaining life by pressing
[+]
while instrument is turned off. The remainig time in days will be shown. After another press "d" will be shown. After another press the gas to be measured will be shown, e. g. "750", "d", "CO".
5.2 Activating a new instrument
— Press and hold
[+]
for approx. 3 seconds while "3, 2, 1" appears in the display. The instrument’s usable life is now started. The gas to be measured will be shown. After 10 seconds the display will turn off or if
[+]
is pressed again, "d" will be shown. After 10 seconds the display will turn off or after another press the remaining time in days will be shown. After 10 seconds the display will turn off.
5.3 Turning the instrument on
— Press and hold
[OK]
. The display counts down until startup:
"3, 2, 1".
— All display segments are lit. Next, the LED, Alarm and Vibrating
alarm are activated in sequence. Please check these before each use.
— The instrument will perform a self test. — The software version and the gas name are displayed. — The number of days of remaining operation are shown,
e. g. "750", "d".
— The A1 and A2 alarm limits are displayed. —For the O
2
sensor: after the first activation of the instrument, a sensor warm up time of up to approx. 15 minutes is needed. The gas name flashes until the warm up time has passed.
5.4 Before entering a working place
— After turning the instrument on, the gas name to be measured
will normally be shown in the display.
— Check for the notice icon
[!]
. When lit, it is recommended that
you perform a bump test as described in section 5.5.
CO H2S O
2
Alarm Threshold A1 30 ppm 10 ppm 19 vol.-%
2)
acknowledgeable Yes Yes No latching No No Yes Alarm Threshold A2 60 ppm 20 ppm 23 vol.-% acknowledgeable No No No latching Yes Yes Yes Display Gas Name Gas Name Gas Name Bump Test Mode off off off Life Signal off off off Turning the instru-
ment off
always
allowed
always
allowed
always
allowed
— Clip the instrument to clothing before working in or near
potential gas hazards.
Insure that the gas opening is not covered and that the instrument is also near to your breathing area.
5.5 Performing a “bump test” with gas
— Prepare a Dräger test gas cylinder with 0.5 l/min and a gas
concentration higher than the alarm threshold to be tested.
— Connect Dräger Pac 3000 and the test gas cylinder to the
calibration adapter or to the Dräger Bump Test Station.
— To enter the bump test mode press the
[+]
3 times within 3 seconds. The instrument beeps twice, quickly. The Display begins to flash slowly.
— Open the regulator valve to let test gas flow over the sensor. — If gas concentration exceeds the alarm thresholds A 1 or A 2 the
corresponding alarm will occur.
— To finish the bump test press
[OK]
, the
[!]
icon is removed from
the display and the instrument returns to the measuring mode.
— If during the bump test no alarm occurs within 1 minute, the
instrument alarm mode is entered to indicate failure. The error icon
[X]
and the notice icon
[!]
are flashing; error code 240 is
shown upon acknowledgement.
"– – –"
is shown instead of the
gas name, and the
[X]
and
[!]
icon are lit. In this case the bump
test can be repeated or the instrument can be calibrated.
— The bump test can also be made by an automatic function. This
function can be activated using the PC software Pac Vision or CC Vision (see section 5.8).
— If the bump test mode was entered by mistake, while the notice
icon
[!]
was not lit and no gas flow over the sensor, press
[+]
to cancel the bump test mode and switch over to measurement screen.
5.6 During operation
— If the allowable measurement range is exceeded or a negative
drift occurs, the following will appear in the display:
“”
(too high concentration) or
“”
(negative drift). — Alarms are indicated as described in section 6. — Continuous function of the instrument is indicated by the life
signal, which is a beep every 60 seconds, if configured (see section 4).
5.7 Turning the instrument off
— Simultaneously hold both keys for approximately 2 seconds until
“3” appears in the display. Continue to hold both keys until the countdown is finished. The alarm and LED will be activated momentarily.
5.8 Calibration and configuration
— For calibration or individual configuration connect Dräger
Pac 3000 to a PC using the connecting cradle or the E-Cal System. Calibration and configuration can be done with installed software Pac Vision or CC Vision. A calibration “due date” can be set using the operation timer (in days). Strictly follow the instructions for use of the modules and software in use.
5.9 Adjustable operation timer (in days)
— Dräger Pac 3000 is equipped with an adjustable operation
timer. The operation timer can be used to set an individually operation period e. g. to adjust a "calibration due date", an "inspection due date", an "out of order date" etc.
— To adjust the operation timer connect Dräger Pac 3000 to a PC
using the connecting cradle or the E-Cal System. The adjustment can be done with installed software Pac Vision or CC Vision.
6 Alarms
6.1 Concentration Pre/Main Alarms
An optical and acoustical alarm are activated whenever alarm threshold A1 or A2 is exceeded.
The instrument is equipped
with a vibrating alarm. It vibrates in parallel to these alarms. — During an A1, the LED will blink and the alarm will sound. — During an A2, the LED and alarm tone will repeat in a double
repeating pattern. — The display will alternate between the measurement value and
"A1" or "A2". — The alarms may, according to the selected configuration, be
acknowledged or turned off. (See section 4.) "Acknowledgeable":
alarms and LED can be acknowledged by pressing
[OK]
.
"Latching": The alarm will only deactivate when the concentration
falls under the alarm threshold and then
[OK]
is pressed.
— If the alarm is not latching, the alarm will deactivate as soon as
the concentration falls under the alarm threshold.
6.2 Battery pre/main alarms
— When the battery pre-alarm is activated, the audible alarm
sounds and the LED blinks, and the “low battery” icon » «
flashes. — To acknowledge the pre-alarm, push
[OK]
.
— After the first battery pre-alarm, the battery will last for
approx. 1 further week and the "low battery" icon stays lit. — When the battery main alarm is activated, the audible alarm
sounds in a repeating pattern of 2 repeating tones and the LED
blinks in the same pattern. — The battery main alarm is not acknowledgeable; the instrument
will automatically turn off after approx. 1 minute. — In case of a very low battery, the internal voltage monitor could
activate the LED’s.
6.3 Changing the battery
— Do not change the battery in explosion-hazard areas!
— The instrument contains a replaceable lithium battery. — The battery is part of the Ex approval. — Only the following battery types shall be used: — Duracell 123 Photo, Lithium, 3 V — Duracell 123 Ultras, Lithium, 3 V — Panasonic CR 123A, Lithium, 3 V — Energizer EL 123A, Lithium, 3 V — Powerone CR 123A, Lithium, 3 V — Turn the instrument off. — Unscrew the 4 screws from the back case. — Open the front case and remove the depleted battery. — Insert the new battery according to specified polarity (+/–).
— Place front case back and fasten it by tightening the 4 screws of
the back case.
— After changing the battery a sensor warm up time of approx.
15 minutes is needed. The gas name flashes until the warm up time has passed.
6.4 Handling of exhausted batteries
—Caution:
— Never throw them into a fire! — Never attempt to charge them! — Never attempt to open them, danger of explosion! — Dispose of exhausted batteries only as special waste in
accordance with local regulations.
— Spent batteries may be returned to Dräger Safety for disposal.
6.5 Usable life alarm
— Before the end of the instrument’s usable life, a warning period
begins. During this period the remaining life time flashes just after turning the instrument on, e. g. "30" / "d".
— To acknowledge this message
[OK]
must be pressed. After that,
the instrument can be used further.
— After the usable life has expired the text "0" / "d" will alternate in
the display and cannot be acknowledged. Dräger Pac 3000 will not longer measure and may be returned to Dräger Safety for recycling.
6.6 Instrument alarm
— The alarm and LED will be activated three times, periodically. —The
[X]
icon is flashing, a 3 digit error code is shown in the
display.
— If an error appears in the display see section 6.7 and if
necessary please contact DrägerService.
6.7 Trouble shooting errors
Code Cause Remedies
100 Flash / EEprom write
fail
Contact DrägerService
102 AD system defect Contact DrägerService 104 Flash check sum wrong Contact DrägerService 107 Self test failed Contact DrägerService 109
Configuration incomplete
Configure again
210
Fresh air calibration failed
Repeat operation 220 Span calibration failed Repeat operation 240 Bump test failed Repeat bump test or
calibrate instrument
7 Technical Specifications
8 Sensor Specifications
_____________ Date of production: see label at the back case.
Environmental Conditions During opera-
tion
–30 to 50
o
C / –22 to 122 oF 700 to 1300 hPa 10 to 90% relative humidity
Conditions for storage
0 to 40
o
C / 32 to 104 oF
30 to 80% relative humidity
Ingress protec­tion
IP 65
Operating times
2 years (typical at 25
o
C)
Battery life (typical at 25
o
C)
24 hours of use per day, 1 minutes alarm per day: CO, H2S: >10.400 hours O
2
: >3.600 hours
Intensity of alarm
typical 90 dB (A) at a distance of 30 cm (11.8 in.)
Dimensions 64 x 84 x 20 (battery compartment 25) mm
2.5 x 3.3 x 0.8 (battery compartment 1) in.
Weight 106 g (3.8 oz.) Approvals CE
0158
(89/336/EEC, 94/9/EC)
DEMKO 04 ATEX I/II M 1/1 G EEx ia I/IIC T4, –30oC≤ Ta ≤ +55oC (–22oF≤ Ta ≤ +131oF) UL Class I, II Div I, Group A, B, C, D, E, F, G, Temp Code T4 –30oC≤ Ta ≤ +55oC (–22oF≤ Ta ≤ +131oF) cUL Class I, II Div I, Group A, B, C, D, E, F, G, Temp Code T4 –30oC≤ Ta ≤ +55oC (–22oF≤ Ta ≤ +131oF)
CO H2S O
2
Temperature Influence Zero point:
±5 ppm no effect≤±0.2 vol.-%
Sensitivity:
±0.3 %
of measured
value / K
±5 % of mea­sured value
±2 % of
measured value
Drift (20
o
C)
Zero point:
±2 ppm/a≤±1 ppm/a≤±0.5 vol.-%/a
Sensitivity:
±1 % of
measured value
/ month
±1 % of
measured value
/ month
±1 % of
measured value
/ month
Please be aware of possible sensor cross sensitivities (see Data Sheet CO: 9023816, H
2
S: 9023819, O2: 9023820).
Loading...