Dräger Oxylog 3000 plus Instructions For Use Manual

Instructions for Use
Oxylog 3000 plus
WARNING To properly use this medical device, read and comply with these Instructions for Use.
Emergency and Transport Ventilator Software 1.n

Typographical Conventions

1 Consecutive numbers indicate steps of action,
with the numbering restarting with "1" for each new sequence of actions.
z Bullet points indicate individual actions or differ-
ent options for action.
– Dashes indicate the listing of data, options, or
objects.
(A) Letters in parentheses refer to elements in the
related illustration.
A Letters in illustrations denote elements referred
to in the text.
Any text shown on the screen and any labeling on the device are printed in bold and italics, for exam­ple, PEEP, Air, or Alarm settings.

Trademarks

– Oxylog
– AutoFlow
– DrägerService
are trademarks owned by Dräger.
– Sekusept
is a trademark of ECOLAB.
– BIPAP*
®
®
®
®
The "greater than" symbol > indicates the naviga­tion path in a dialog window, for example, System setup > Monitoring > Basic settings. In this example, System setup represents the dialog win­dow title, Monitoring represents a horizontally aligned tab, and Basic settings a vertically aligned tab.

Screen images

Schematic renderings of screen images are used, which may differ in appearance or in configuration from the actual screen images.
* Trademark used under license

Safety Information Definitions

WARNING A WARNING statement provides important
information about a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
A CAUTION statement provides important information about a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or in damage to the equipment or other property.
NOTE
A NOTE provides additional information intended to avoid inconvenience during operation.

Definition of target groups

For this medical device, users, service personnel, and experts are defined as target groups.
These traget groups must have been instructed in the use of the medical device and must have the necessary expertise, training, and knowledge to use, install, reprocess, maintain or repair the medi­cal device.
Dräger emphasizes that the medical device must be used, installed, reprocessed, maintained or repaired exclusively by the defined target groups.

Users

Users are intended operators as defined on page 14 hereof for the use of the medical device in accordance with its intended use.
Instructions for Use Oxylog 3000 plus SW 1.n 3

Service Personnel

Service Personnel are persons who are responsi­ble for the maintenance of the medical device towards the operating organization.
Service Personnel are persons who are authorized to install, reprocess, or maintain the medical device.

Experts

Experts are persons who are authorized to carry out repair or complex maintenance work on the medical device.

Abbreviations and Symbols

Please refer to "Abbreviations" on page 23 and "Symbols" on page 24 for explanations.

Contents

Contents
For Your Safety and that of Your Patients .7
General Safety Information. . . . . . . . . . . . . . . 8
Product-specific Safety Information . . . . . . . . 11
Application . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . 14
Indications / Contraindications . . . . . . . . . . . . 14
Environment of use. . . . . . . . . . . . . . . . . . . . . 15
System Overview . . . . . . . . . . . . . . . . . . . . . 17
Front panel with all options. . . . . . . . . . . . . . . 18
Range of functions . . . . . . . . . . . . . . . . . . . . . 22
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . 23
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Operating Concept . . . . . . . . . . . . . . . . . . . . 27
Switch on or off. . . . . . . . . . . . . . . . . . . . . . . . 28
Ventilation controls . . . . . . . . . . . . . . . . . . . . . 29
Display operating controls . . . . . . . . . . . . . . . 30
Additional function keys . . . . . . . . . . . . . . . . . 31
Screen window structure . . . . . . . . . . . . . . . . 32
Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Internal rechargeable battery . . . . . . . . . . . . . 37
Connecting the power supply . . . . . . . . . . . . . 38
External power supply . . . . . . . . . . . . . . . . . . 39
Connecting the gas supply . . . . . . . . . . . . . . . 41
Assembling the adult reusable hose system . 43 Connecting the adult disposable hose system 45 Connecting the paediatric disposable hose
system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Connecting the bacterial filter or HME . . . . . . 47
Connecting the CO
2 sensor and the cuvette . 48
Hanging the Oxylog 3000 plus on standard
rail systems. . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Getting Started . . . . . . . . . . . . . . . . . . . . . . . 51
Charging the battery. . . . . . . . . . . . . . . . . . . . 52
Determining the approximate pneumatic
operating time. . . . . . . . . . . . . . . . . . . . . . . . . 53
Checking readiness for operation. . . . . . . . . . 54
2 zero calibration and filter check before
CO
ventilation (optional) . . . . . . . . . . . . . . . . . . . . 58
Preparation for use after system check, CO
2
zero calibration and filter check. . . . . . . . . . . 60
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Starting operation . . . . . . . . . . . . . . . . . . . . . 62
Preparing ventilation mode . . . . . . . . . . . . . . 64
VC-CMV, VC-AC . . . . . . . . . . . . . . . . . . . . . . 65
VC-SIMV, VC-SIMV/PS. . . . . . . . . . . . . . . . . 68
PC-BIPAP, PC-BIPAP/PS . . . . . . . . . . . . . . . 71
SpnCPAP, SpnCPAP/PS . . . . . . . . . . . . . . . . 73
NIV – Non-invasive ventilation (Mask
ventilation) . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Special functions . . . . . . . . . . . . . . . . . . . . . . 77
O2 concentration with "O2 blending". . . . . . . 79
Setting HME correction . . . . . . . . . . . . . . . . . 80
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Screen brightness . . . . . . . . . . . . . . . . . . . . . 81
Alarm volume . . . . . . . . . . . . . . . . . . . . . . . . 81
Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Types of alarms. . . . . . . . . . . . . . . . . . . . . . . 84
In the event of an alarm. . . . . . . . . . . . . . . . . 85
Setting alarm limits . . . . . . . . . . . . . . . . . . . . 87
Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Displaying curves . . . . . . . . . . . . . . . . . . . . . 90
Displaying measured values . . . . . . . . . . . . . 90
2 measurement (optional) . . . . . . . . . . . . 91
CO
Configuration. . . . . . . . . . . . . . . . . . . . . . . . 95
Setting configuration parameters / display
information. . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Displaying configuration and information . . . 97
Customer Service Mode . . . . . . . . . . . . . . . . 98
Customer service manual . . . . . . . . . . . . . . . 110
Problem Solving . . . . . . . . . . . . . . . . . . . . . 111
Alarm - Cause - Remedy. . . . . . . . . . . . . . . . 112
Messages in the alarm window . . . . . . . . . . . 112
Messages in the information window . . . . . . 120
Error messages during the device check . . . 122
Instructions for Use Oxylog 3000 plus SW 1.n 5
Contents
Cleaning, Disinfection and Sterilization . . . 123
Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Safety Information on reprocessing . . . . . . . . 127
Reprocessing procedure . . . . . . . . . . . . . . . . 127
Reprocessing list . . . . . . . . . . . . . . . . . . . . . . 131
Assembling parts . . . . . . . . . . . . . . . . . . . . . . 131
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . 133
Maintenance intervals of Oxylog 3000 plus . . 134
Safety inspections . . . . . . . . . . . . . . . . . . . . . 135
Exchanging the internal battery . . . . . . . . . . . 136
In case of ventilator failure . . . . . . . . . . . . . . . 136
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Disposing of the medical device. . . . . . . . . . . 138
Disposal instructions . . . . . . . . . . . . . . . . . . . 138
Technical Data . . . . . . . . . . . . . . . . . . . . . . . 139
Ambient conditions . . . . . . . . . . . . . . . . . . . . . 140
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Performance data. . . . . . . . . . . . . . . . . . . . . . 143
Measured values and curves display . . . . . . . 144
Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Operating data . . . . . . . . . . . . . . . . . . . . . . . . 146
Device specifications . . . . . . . . . . . . . . . . . . . 149
Materials used . . . . . . . . . . . . . . . . . . . . . . . . 150
Technical Documentation for the Oxylog 3000 plus according to EMC
standard IEC 60601-1-2 . . . . . . . . . . . . . . . . . 152
Principles of Operation . . . . . . . . . . . . . . . . 157
Ventilation modes. . . . . . . . . . . . . . . . . . . . . . 158
AutoFlow . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162
Dead space . . . . . . . . . . . . . . . . . . . . . . . . . . 164
Determining the cycle time, inspiratory time
and expiratory time. . . . . . . . . . . . . . . . . . . . . 164
Functional description . . . . . . . . . . . . . . . . . . 165
List of Accessories . . . . . . . . . . . . . . . . . . . 167
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
6
Instructions for Use Oxylog 3000 plus SW 1.n

For Your Safety and that of Your Patients

For Your Safety and that of Your Patients
General Safety Information . . . . . . . . . . . . . 8
Strictly follow these Instructions for Use. . . . . 8
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . 8
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Connected Devices . . . . . . . . . . . . . . . . . . . . 9
Safe connection with other electrical
equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Modifications. . . . . . . . . . . . . . . . . . . . . . . . . . 9
Patient safety . . . . . . . . . . . . . . . . . . . . . . . . . 9
Patient monitoring. . . . . . . . . . . . . . . . . . . . . . 9
Information on Electromagnetic Compatibility 9
Functional safety . . . . . . . . . . . . . . . . . . . . . . 10
Appropriate monitoring . . . . . . . . . . . . . . . . . . 10
Product-specific Safety Information. . . . . . 11
Installing accessories . . . . . . . . . . . . . . . . . . . 11
Instructions for Use only available once . . . . . 11
Instructions for Use Oxylog 3000 plus SW 1.n 7
For Your Safety and that of Your Patients

General Safety Information

The following WARNINGS and CAUTIONS apply to general operation of the device. WARNINGS and CAUTIONS specific to subsystems or particular features appear with those topics in later sections of the manual.

Strictly follow these Instructions for Use

WARNING Risk of incorrect operation and of incorrect
use
Any use of the medical device requires full understanding and strict observation of all sections of these Instructions for Use. The medical device must only be used for the purpose specified under "Intended use" on page 14 and in conjunction with appropriate patient monitoring.
Strictly observe all WARNING and CAUTION statements throughout these Instructions for Use and all statements on medical device labels. Failure to observe these safety information statements constitutes a use of the medical device that is inconsistent with its intended use.

Maintenance

WARNI NG Risk of medical device failure and of patient
injury
The medical device must be inspected and serviced regularly by service personnel. Repair and complex maintenance work car­ried out on the medical device must be per­formed by experts.
Dräger recommends that a service contract is obtained with DrägerService and that all repairs are performed by DrägerService. For maintenance Dräger recommends the use of authentic Dräger repair parts.
If the above are not complied with, medical device failure and patient injury cannot be excluded.
Observe chapter "Maintenance".

Accessories

WARNI NG Risk due to unreleased accessories
If unreleased accessories are used, there is a risk of patient injury due to medical device failure.
Only use the medical device together with released accessories listed in the current list of accessories.
8
Instructions for Use Oxylog 3000 plus SW 1.n
For Your Safety and that of Your Patients

Connected Devices

WARNING Risk of electric shock and of device malfunc-
tion
Any connected devices or device combina­tion, not complying with the requirements mentioned in these Instructions for Use can compromise the correct functioning of the medical device and lead to an electric shock. Before operating the medical device, strictly comply with the Instructions for Use of all connected devices or device combinations.

Safe connection with other electrical equipment

CAUTION
Risk of patient injury
Electrical connections to equipment not listed in these Instructions for Use or these Assembly Instructions must only be made when approved by each respective manufacturer.
This publication excludes references to various hazards which are obvious to a medical professional and operator of this medical device, to the consequences of medical device misuse, and to potentially adverse effects in patients with abnormal conditions.
Medical device modification or misuse can be dangerous.

Patient monitoring

The operators of the medical device are responsible for choosing appropriate safety monitoring that provides adequate information on medical device performance and patient condition.
Patient safety may be achieved through a wide variety of means, ranging from electronic surveil­lance of medical device performance and patient condition, to simple, direct observation of clinical signs.
The responsibility for the selection of the best level of patient monitoring lies solely with the medical device operator.
Information on Electromagnetic

Modifications

WARNING Modification of the device is not allowed.

Patient safety

The design of the medical device, the accompany­ing literature, and the labeling on the medical device are based on the assumption that the use of the equipment is restricted to trained professionals, and that certain inherent characteristics of the med­ical device are known to the trained operator. Instructions, warnings, and caution statements are limited, therefore, largely to the specifics of the Dräger design.
Instructions for Use Oxylog 3000 plus SW 1.n 9
Compatibility
General information on electromagnetic compatibility (EMC) pursuant to international EMC standard IEC 60601-1-2:
Electromedical devices are subject to special precautionary measures concerning electromagnetic compatibility (EMC) and must be installed and put into operation in accordance with the EMC information. Refer to section "Technical Documentation for the Oxylog 3000 plus according to EMC standard IEC 60601-1-2" on page 152.
WARNING Do not use portable and mobile HF communi-
cations equipment, e.g., mobile phones, in the vicinity of the medical device.
For Your Safety and that of Your Patients

Functional safety

The essential performance of the Oxylog 3000 plus is defined as:
Appropriate delivery of ventilation to the patient­connection port or generation of an alarm condition.

Appropriate monitoring

CAUTION Always use a separate SpO2 monitor for
patients who are dependent on an exact O concentration.
The monitoring functionality of the Oxylog 3000 plus ensures appropriate monitoring of ventilation therapy. To ensure appropriate monitoring during ventilation, always set the following alarm limits:
– Airway pressure, Paw
– Expiratory minute volume, MVe
– Respiratory rate (if applicable), RR
–etCO
2 (if applicable)
Not setting appropriate alarm limits could suppress alarms related to:
– Acute changes in the patient’s condition
– Incorrect settings and faulty handling
– Hose system leakage
2
10
Instructions for Use Oxylog 3000 plus SW 1.n

Product-specific Safety Information

For Your Safety and that of Your Patients
WARNING Ventilation monitoring is mandatory at all
times! Whenever a patient is connected to the ventilator, constant attention by qualified medical staff is required in order to provide immediate corrective action in case of a malfunction.
The operator shall not solely rely on the built­in monitoring of the ventilator and must always assume full responsibility for proper ventilation and patient safety in all situations.
WARNING Keep a manual breathing bag available.
If a failure is detected in the ventilator and its life-support functions can no longer be guaranteed (e.g. in case of a power supply failure or interruption in the O2 supply), ventilation must be started without delay with an independent ventilator (breathing bag) – using PEEP and/or increased inspired O concentration as necessary.
WARNING To ensure proper ventilation, consider the
total dead space of the breathing circuit, espe­cially when using small tidal volumes. Observe for signs of rebreathing.
Risk of CO2 rebreathing.
2

Installing accessories

CAUTION
Installations on the Oxylog 3000 plus must be done in accordance with these Instructions for Use. Make sure that the connections are securely fitted onto the basic system.
Strictly follow the Assembly Instructions and Instructions for Use.

Instructions for Use only available once

NOTE
Only one copy of the Instructions for Use is included in the clinical package and should there­fore be kept in an accessible location for users.
NOTE
An etCO2 value by itself is insufficient as a basis for medical decisions.
Instructions for Use Oxylog 3000 plus SW 1.n 11
This page intentionally left blank

Application

Intended use . . . . . . . . . . . . . . . . . . . . . . . . . 14
Indications / Contraindications. . . . . . . . . . 14
Environment of use . . . . . . . . . . . . . . . . . . . 15
Application
Instructions for Use Oxylog 3000 plus SW 1.n 13
Application

Intended use

The Oxylog 3000 plus is a time-cycled, volume­controlled and pressure-controlled emergency and transport ventilator for patients requiring mandatory or assisted ventilation with a tidal volume from 50 mL upwards.
Intended operator: the device is intended for use by and under the supervision of trained healthcare professionals, e.g. doctors, nurses, emergency medical technicians, respiratory therapists, and paramedics.

Indications / Contraindications

The device is intended for use with patients with a tidal volume of 50 mL upwards.
WARNING The Oxylog 3000 plus ventilator may only be
used under the supervision of trained healthcare professionals in case immediate corrective action is required in the event of a device failure.
14
Instructions for Use Oxylog 3000 plus SW 1.n

Environment of use

Intended environment of use:
– Mobile use for emergency patients, in both
outdoor and indoor environments.
– During transport in ambulances or aircraft,
including helicopters.
– In accident and emergency departments.
– When moving ventilated patients around the
hospital.
– In the recovery room.
WARNING Do not use the medical device in hyperbaric
chambers.
The medical device may malfunction, causing danger to the patient.
WARNING Do not use the medical device in conjunction
with magnetic resonance imaging (MRI, NMR, NMI).
The medical device may malfunction, causing danger to the patient.
Application
WARNING This medical device is neither approved nor
certified for use in areas where combustible or explosive gas mixtures are likely to occur.
Instructions for Use Oxylog 3000 plus SW 1.n 15
This page intentionally left blank

System Overview

Front panel with all options. . . . . . . . . . . . . 18
Side view, right . . . . . . . . . . . . . . . . . . . . . . . . 19
Rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Adult hose system, reusable . . . . . . . . . . . . . 20
Adult hose system, disposable. . . . . . . . . . . . 20
Paediatric hose system, disposable . . . . . . . . 21
Range of functions . . . . . . . . . . . . . . . . . . . . 22
Ventilation functions of the Oxylog 3000 plus. 22
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . 23
Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
System Overview
Instructions for Use Oxylog 3000 plus SW 1.n 17
System Overview
ABCDEFG
H
I J K
L
M
NOPQRS
T
U
V

Front panel with all options

A Screen with screen pages for the specific
application
B Key Alarms to display the alarm settings
in the "Settings and Alarms" window and to change screen pages
C Key Settings to display ventilation
parameters (ventilation screen) in the "Settings and Alarms" window and to change screen pages
D Key for setting the ventilation mode SpnCPAP
E Key for setting the ventilation modes VC-CMV /
VC-AC
F Key for setting the ventilation mode VC-SIMV
G Key for setting the ventilation mode PC-BIPAP
H Red and yellow alarm indicators
I Key for suppressing the acoustic alarm sig-
nal for 2 minutes
J Key Alarm Reset for acknowledging alarm
messages
K Key O
100 % O2 for 100 % O2 application, depending
on the option installed at manufacture
2 inhalation for O2 inhalation or key
L Key Insp. Hold for initiating a manual inspira-
tion or for extending the current inspiratory time.
M Key Start/Standby
N Display symbols for the power supply
Charge status of the internal battery Mains power supply connected
O Rotary knob for making selections, changing
and confirming settings
P Control knob for setting the O
2 concentration
FiO2
Q Control knob for setting the maximum
inspiratory pressure Pmax
R Control knob for setting the respiratory rate RR
S Control knob for setting the tidal volume VT
T Explanation of color codes for quick pre-setting
of RR and VT
U Key Curves to change between the
pressure, flow or CO and large presentation
V Key Values to change screen pages in
the "Measured Values" window
Curves
2 (optional) curve in small
Front
18
Instructions for Use Oxylog 3000 plus SW 1.n
System Overview
B
C
D
E F
A
G
H
A
B
C

Side view, right

A Emergency air intake
CAUTION
Do not block the emergency air intake. This may result in ventilator malfunction.
B Knob to secure the battery compartment cover
C Connectors for flow measuring lines
D Gas outlet for breathing hose
E Connector for O
F Connector for power supply
G Connector for CO
H Connector for data communication cable
2 supply
2 sensor

Rear view

Side
A Emergency air intake
CAUTION
Do not block the air intake. This may result in ventilator malfunction.
B Fresh-gas intake with a filter cartridge
C Protection bracket
CAUTION
Do not use the protection bracket as a handle. Tilting the device to a vertical position may lead to airway pressure oscillation.
Rear
Instructions for Use Oxylog 3000 plus SW 1.n 19
System Overview
A
E
B
C
D
E
D
C
A
B

Adult hose system, reusable

A Angled connector
B Flow sensor
C Breathing valve
D Breathing hose
E Flow and pressure measuring hoses

Adult hose system, disposable

001
A Angled connector
B Flow sensor
C Breathing valve
D Breathing hose
E Flow and pressure measuring hoses
002
20
Instructions for Use Oxylog 3000 plus SW 1.n

Paediatric hose system, disposable

D
A
E
C
B
A Angled connector
B Flow sensor
C Breathing valve
D Breathing hose
E Flow and pressure measuring hoses
System Overview
08137170
Instructions for Use Oxylog 3000 plus SW 1.n 21
System Overview

Range of functions

Ventilation functions of the Oxylog 3000 plus

Ventilation modes:
– Volume-controlled ventilation:
– VC-CMV / VC-AC,
–VC-SIMV.
– Pressure-controlled ventilation:
–PC-BIPAP.
– Support of spontaneous breathing:
– SpnCPAP.
Additional settings for ventilation:
– Pressure Support: in the ventilation modes
VC-SIMV, PC-BIPAP, SpnCPAP,
– Apnoea ventilation: in the ventilation mode
SpnCPAP,
– AutoFlow (optional): in the ventilation modes
VC-CMV, VC-AC and VC-SIMV.
– NIV: in the ventilation modes: SpnCPAP (/PS),
PC-BIPAP (/PS), VC-CMV / AF, VC-AC / AF and VC-SIMV / AF.
For a detailed description of the ventilation modes and the additional settings, refer to "Principles of Operation" on page 157. For abbreviations, see "Abbreviations" on page 23.
NOTE
In these Instructions for Use the unit of measure­ment for airway pressure is expressed in [mbar]. However, in some languages the display of the Oxylog 3000 plus shows [cmH
One [mbar] equals approximately one [cmH
2O].
2O].
Special procedures:
– Inspiration hold,
–O
2 inhalation (optional), with an inhalation
mask.
22
Instructions for Use Oxylog 3000 plus SW 1.n

Abbreviations

System Overview
Abbreviation Explanation
100 % O
2 100 % O2 flow
AF AutoFlow
BF Body Floating
bpm Breaths per minute
BTPS Body Temperature, Pressure
Saturated
C Lung compliance
2 Carbon dioxide
CO
CSM Customer Service Mode
ΔPsupp Positive pressure above PEEP
EMC Electromagnetic Compatibility
ESD Electrostatic Discharge
2 Endtidal CO2 concentration
etCO
2 Fractional inspired oxygen
FiO
concentration
FRC Functional Residual Capacity
HME Heat and Moisture Exchange
I:E Ratio inspiratory time to expiratory
time
IPX2 Ingress Protection level 2
IPX4 Ingress Protection level 4
MVe Expiratory minute volume
MVi Inspiratory minute volume
MVespon Spontaneous expiratory minute
volume
NIV Non-invasive ventilation –
mask ventilation
2 Oxygen
O
2-Inhalat. O2 inhalation
O
Paw Airway pressure
PC-BIPAP Pressure Controlled - Biphasic
Positive Airway Pressure
PEEP Positive End Expiratory Pressure
Abbreviation Explanation
PIF Peak Inspiratory Flow
Pinsp Inspiratory pressure
PIP Peak Inspiratory Pressure
Pmax Maximum allowed inspiratory
pressure
Pmean Mean airway pressure
Pplat Plateau pressure
PS Pressure Support
R Resistance
RF Radio Frequency
RR Respiratory Rate (frequency)
RRapn Respiratory Rate during apnoea
ventilation
RRsp Spontaneous Respiratory Rate
SpnCPAP Spontaneous Continuous Positive
Airway Pressure
2 Saturation of peripheral oxygen
SpO
Tapn Time before apnoea is recognized
Te Expiratory time
Ti Inspiratory time
Tplat % Plateau time in % of inspiratory time
UN United Nations
VC-AC Volume Controlled - Assist Control
VC-CMV Volume Controlled - Controlled
Mandatory Ventilation
VC-SIMV Volume Controlled - Synchronized
Intermittent Mandatory Ventilation
VT Tidal volume
VTapn Tidal volume during apnoea
ventilation
VTe Expiratory tidal volume
VTi Inspiratory tidal volume
Instructions for Use Oxylog 3000 plus SW 1.n 23
System Overview
Insp. hold
Alarms

Symbols

Symbol Explanation
Key for initiating a manual
inspiration or for extending the current inspiratory time.
Key to display ventilation
Settings
parameters (ventilation screen) in the "Settings and Alarms" window and to change screen pages.
Key to display the alarm settings in
the "Settings and Alarms" window and to change screen pages.
Key to change screen pages in the
Values
"Measured Values" window.
Key to change between the
Curves
pressure, flow or CO
2 (optional)
curve in small and large presentation.
Key for suppressing the acoustic
alarm signal for 2 minutes.
Key for acknowledging alarm
Alarm
Reset
messages.
Rotary knob
Start / Standby key
Upper alarm limit
Lower alarm limit
! Advisory message
!! Caution message
Symbol Explanation
* Trigger indicator
Warning
Defibrillation-proof type BF applied
part
Charge status of the internal battery
External power supply connected
Battery charge (example: three quarters full)
Class II equipment, device pro­tected against electric shock with additional safety precautions such as double or reinforced insulations, without protective earthing.
Do not dispose of the device as municipal waste.
Manufacturing date
Manufacturer
DC input
Operating instructions
Follow Instructions for Use
!!! Warning message
24
Instructions for Use Oxylog 3000 plus SW 1.n
System Overview
Symbol Explanation
For dry locations only
Caution, hot surface!
Warning, dangerous voltage!
Temperature limitations
Latex free
Do not reuse
Do not use oil and grease
Non-sterile
Symbol Explanation
Slope (steep, medium, flat)
Quantity
Keep away from sunlight
For indoor use only
Do not open
Prohibition: Do not obstruct emer­gency air intake or fresh gas intake
Pediatric
Adult
Paediatric
Adult
Instructions for Use Oxylog 3000 plus SW 1.n 25
This page intentionally left blank

Operating Concept

Switch on or off . . . . . . . . . . . . . . . . . . . . . . 28
Switch on . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Switch off . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Ventilation controls . . . . . . . . . . . . . . . . . . . 29
Display operating controls . . . . . . . . . . . . . 30
Additional function keys . . . . . . . . . . . . . . . 31
Screen window structure. . . . . . . . . . . . . . . 32
Operating Concept
Instructions for Use Oxylog 3000 plus SW 1.n 27
Operating Concept
Oxylog 3000 plus
AB

Switch on or off

Switch on

z To switch the device ON, briefly press the
key (A).
00337170

Switch off

Refer to "Shutdown" on page 82.
28
Instructions for Use Oxylog 3000 plus SW 1.n

Ventilation controls

BC
A
Oxylog 3000 plus
Operating Concept

General ventilation controls

A Keys for selecting the ventilation modes:
– VC-CMV / VC-AC
–VC-SIMV
– SpnCPAP
– PC-BIPAP.
B Rotary knob.
C Ventilation parameter controls:
– Inspiratory tidal volume VT [mL],
– Ventilation respiratory rate RR [/min],
– Maximum inspiratory pressure Pmax
[mbar],
–O
2 concentration FiO2 [%].

Selecting the ventilation mode

Oxylog 3000 plus
00137170
BC
z Press the appropriate ventilation mode key (A)
for approximately 3 seconds.
Or
1 Press the appropriate ventilation mode key (A).
2 Press the rotary knob (B) to confirm.
The selected ventilation mode will be activated.
The active ventilation mode is displayed in the upper left corner of the display (C).
A
08537170
Instructions for Use Oxylog 3000 plus SW 1.n 29
Operating Concept
Oxylog 3000 plus
B C
D
E
A
Curves

Display operating controls

A Rotary knob for making selections, changing
and confirming settings.
NOTE
Different parameters can be set in the display window via the rotary knob (e.g. Ti, PEEP, ΔPsupp, Pinsp). – To select the parameter: turn rotary knob. – To activate the parameter: press rotary knob. – To set the value: turn rotary knob. – To confirm the value: press rotary knob.
00237170
B Key Values to change screen pages in
the "Measured Values" window.
C Key Curves to change between the
pressure, flow or CO2 (optional) curve in small and large presentation.
D Key Settings to display ventilation
parameters (ventilation screen) in the "Settings and Alarms" window and to change screen pages.
E Key Alarms to display the alarm settings
in the "Settings and Alarms" window and to change screen pages.
30
Instructions for Use Oxylog 3000 plus SW 1.n
Loading...
+ 142 hidden pages