Dräger Oxylog 3000 plus Instructions For Use Manual

Instructions for Use
Oxylog 3000 plus
WARNING To properly use this medical device, read and comply with these Instructions for Use.
Emergency and Transport Ventilator Software 1.n

Typographical Conventions

1 Consecutive numbers indicate steps of action,
with the numbering restarting with "1" for each new sequence of actions.
z Bullet points indicate individual actions or differ-
ent options for action.
– Dashes indicate the listing of data, options, or
objects.
(A) Letters in parentheses refer to elements in the
related illustration.
A Letters in illustrations denote elements referred
to in the text.
Any text shown on the screen and any labeling on the device are printed in bold and italics, for exam­ple, PEEP, Air, or Alarm settings.

Trademarks

– Oxylog
– AutoFlow
– DrägerService
are trademarks owned by Dräger.
– Sekusept
is a trademark of ECOLAB.
– BIPAP*
®
®
®
®
The "greater than" symbol > indicates the naviga­tion path in a dialog window, for example, System setup > Monitoring > Basic settings. In this example, System setup represents the dialog win­dow title, Monitoring represents a horizontally aligned tab, and Basic settings a vertically aligned tab.

Screen images

Schematic renderings of screen images are used, which may differ in appearance or in configuration from the actual screen images.
* Trademark used under license

Safety Information Definitions

WARNING A WARNING statement provides important
information about a potentially hazardous situation which, if not avoided, could result in death or serious injury.
CAUTION
A CAUTION statement provides important information about a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or in damage to the equipment or other property.
NOTE
A NOTE provides additional information intended to avoid inconvenience during operation.

Definition of target groups

For this medical device, users, service personnel, and experts are defined as target groups.
These traget groups must have been instructed in the use of the medical device and must have the necessary expertise, training, and knowledge to use, install, reprocess, maintain or repair the medi­cal device.
Dräger emphasizes that the medical device must be used, installed, reprocessed, maintained or repaired exclusively by the defined target groups.

Users

Users are intended operators as defined on page 14 hereof for the use of the medical device in accordance with its intended use.
Instructions for Use Oxylog 3000 plus SW 1.n 3

Service Personnel

Service Personnel are persons who are responsi­ble for the maintenance of the medical device towards the operating organization.
Service Personnel are persons who are authorized to install, reprocess, or maintain the medical device.

Experts

Experts are persons who are authorized to carry out repair or complex maintenance work on the medical device.

Abbreviations and Symbols

Please refer to "Abbreviations" on page 23 and "Symbols" on page 24 for explanations.

Contents

Contents
For Your Safety and that of Your Patients .7
General Safety Information. . . . . . . . . . . . . . . 8
Product-specific Safety Information . . . . . . . . 11
Application . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Intended use. . . . . . . . . . . . . . . . . . . . . . . . . . 14
Indications / Contraindications . . . . . . . . . . . . 14
Environment of use. . . . . . . . . . . . . . . . . . . . . 15
System Overview . . . . . . . . . . . . . . . . . . . . . 17
Front panel with all options. . . . . . . . . . . . . . . 18
Range of functions . . . . . . . . . . . . . . . . . . . . . 22
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . 23
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Operating Concept . . . . . . . . . . . . . . . . . . . . 27
Switch on or off. . . . . . . . . . . . . . . . . . . . . . . . 28
Ventilation controls . . . . . . . . . . . . . . . . . . . . . 29
Display operating controls . . . . . . . . . . . . . . . 30
Additional function keys . . . . . . . . . . . . . . . . . 31
Screen window structure . . . . . . . . . . . . . . . . 32
Assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Internal rechargeable battery . . . . . . . . . . . . . 37
Connecting the power supply . . . . . . . . . . . . . 38
External power supply . . . . . . . . . . . . . . . . . . 39
Connecting the gas supply . . . . . . . . . . . . . . . 41
Assembling the adult reusable hose system . 43 Connecting the adult disposable hose system 45 Connecting the paediatric disposable hose
system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Connecting the bacterial filter or HME . . . . . . 47
Connecting the CO
2 sensor and the cuvette . 48
Hanging the Oxylog 3000 plus on standard
rail systems. . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Getting Started . . . . . . . . . . . . . . . . . . . . . . . 51
Charging the battery. . . . . . . . . . . . . . . . . . . . 52
Determining the approximate pneumatic
operating time. . . . . . . . . . . . . . . . . . . . . . . . . 53
Checking readiness for operation. . . . . . . . . . 54
2 zero calibration and filter check before
CO
ventilation (optional) . . . . . . . . . . . . . . . . . . . . 58
Preparation for use after system check, CO
2
zero calibration and filter check. . . . . . . . . . . 60
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Starting operation . . . . . . . . . . . . . . . . . . . . . 62
Preparing ventilation mode . . . . . . . . . . . . . . 64
VC-CMV, VC-AC . . . . . . . . . . . . . . . . . . . . . . 65
VC-SIMV, VC-SIMV/PS. . . . . . . . . . . . . . . . . 68
PC-BIPAP, PC-BIPAP/PS . . . . . . . . . . . . . . . 71
SpnCPAP, SpnCPAP/PS . . . . . . . . . . . . . . . . 73
NIV – Non-invasive ventilation (Mask
ventilation) . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Special functions . . . . . . . . . . . . . . . . . . . . . . 77
O2 concentration with "O2 blending". . . . . . . 79
Setting HME correction . . . . . . . . . . . . . . . . . 80
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Screen brightness . . . . . . . . . . . . . . . . . . . . . 81
Alarm volume . . . . . . . . . . . . . . . . . . . . . . . . 81
Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Types of alarms. . . . . . . . . . . . . . . . . . . . . . . 84
In the event of an alarm. . . . . . . . . . . . . . . . . 85
Setting alarm limits . . . . . . . . . . . . . . . . . . . . 87
Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Displaying curves . . . . . . . . . . . . . . . . . . . . . 90
Displaying measured values . . . . . . . . . . . . . 90
2 measurement (optional) . . . . . . . . . . . . 91
CO
Configuration. . . . . . . . . . . . . . . . . . . . . . . . 95
Setting configuration parameters / display
information. . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Displaying configuration and information . . . 97
Customer Service Mode . . . . . . . . . . . . . . . . 98
Customer service manual . . . . . . . . . . . . . . . 110
Problem Solving . . . . . . . . . . . . . . . . . . . . . 111
Alarm - Cause - Remedy. . . . . . . . . . . . . . . . 112
Messages in the alarm window . . . . . . . . . . . 112
Messages in the information window . . . . . . 120
Error messages during the device check . . . 122
Instructions for Use Oxylog 3000 plus SW 1.n 5
Contents
Cleaning, Disinfection and Sterilization . . . 123
Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . 124
Safety Information on reprocessing . . . . . . . . 127
Reprocessing procedure . . . . . . . . . . . . . . . . 127
Reprocessing list . . . . . . . . . . . . . . . . . . . . . . 131
Assembling parts . . . . . . . . . . . . . . . . . . . . . . 131
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . 133
Maintenance intervals of Oxylog 3000 plus . . 134
Safety inspections . . . . . . . . . . . . . . . . . . . . . 135
Exchanging the internal battery . . . . . . . . . . . 136
In case of ventilator failure . . . . . . . . . . . . . . . 136
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Disposing of the medical device. . . . . . . . . . . 138
Disposal instructions . . . . . . . . . . . . . . . . . . . 138
Technical Data . . . . . . . . . . . . . . . . . . . . . . . 139
Ambient conditions . . . . . . . . . . . . . . . . . . . . . 140
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Performance data. . . . . . . . . . . . . . . . . . . . . . 143
Measured values and curves display . . . . . . . 144
Monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Operating data . . . . . . . . . . . . . . . . . . . . . . . . 146
Device specifications . . . . . . . . . . . . . . . . . . . 149
Materials used . . . . . . . . . . . . . . . . . . . . . . . . 150
Technical Documentation for the Oxylog 3000 plus according to EMC
standard IEC 60601-1-2 . . . . . . . . . . . . . . . . . 152
Principles of Operation . . . . . . . . . . . . . . . . 157
Ventilation modes. . . . . . . . . . . . . . . . . . . . . . 158
AutoFlow . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162
Dead space . . . . . . . . . . . . . . . . . . . . . . . . . . 164
Determining the cycle time, inspiratory time
and expiratory time. . . . . . . . . . . . . . . . . . . . . 164
Functional description . . . . . . . . . . . . . . . . . . 165
List of Accessories . . . . . . . . . . . . . . . . . . . 167
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169
6
Instructions for Use Oxylog 3000 plus SW 1.n

For Your Safety and that of Your Patients

For Your Safety and that of Your Patients
General Safety Information . . . . . . . . . . . . . 8
Strictly follow these Instructions for Use. . . . . 8
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . 8
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Connected Devices . . . . . . . . . . . . . . . . . . . . 9
Safe connection with other electrical
equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Modifications. . . . . . . . . . . . . . . . . . . . . . . . . . 9
Patient safety . . . . . . . . . . . . . . . . . . . . . . . . . 9
Patient monitoring. . . . . . . . . . . . . . . . . . . . . . 9
Information on Electromagnetic Compatibility 9
Functional safety . . . . . . . . . . . . . . . . . . . . . . 10
Appropriate monitoring . . . . . . . . . . . . . . . . . . 10
Product-specific Safety Information. . . . . . 11
Installing accessories . . . . . . . . . . . . . . . . . . . 11
Instructions for Use only available once . . . . . 11
Instructions for Use Oxylog 3000 plus SW 1.n 7
For Your Safety and that of Your Patients

General Safety Information

The following WARNINGS and CAUTIONS apply to general operation of the device. WARNINGS and CAUTIONS specific to subsystems or particular features appear with those topics in later sections of the manual.

Strictly follow these Instructions for Use

WARNING Risk of incorrect operation and of incorrect
use
Any use of the medical device requires full understanding and strict observation of all sections of these Instructions for Use. The medical device must only be used for the purpose specified under "Intended use" on page 14 and in conjunction with appropriate patient monitoring.
Strictly observe all WARNING and CAUTION statements throughout these Instructions for Use and all statements on medical device labels. Failure to observe these safety information statements constitutes a use of the medical device that is inconsistent with its intended use.

Maintenance

WARNI NG Risk of medical device failure and of patient
injury
The medical device must be inspected and serviced regularly by service personnel. Repair and complex maintenance work car­ried out on the medical device must be per­formed by experts.
Dräger recommends that a service contract is obtained with DrägerService and that all repairs are performed by DrägerService. For maintenance Dräger recommends the use of authentic Dräger repair parts.
If the above are not complied with, medical device failure and patient injury cannot be excluded.
Observe chapter "Maintenance".

Accessories

WARNI NG Risk due to unreleased accessories
If unreleased accessories are used, there is a risk of patient injury due to medical device failure.
Only use the medical device together with released accessories listed in the current list of accessories.
8
Instructions for Use Oxylog 3000 plus SW 1.n
For Your Safety and that of Your Patients

Connected Devices

WARNING Risk of electric shock and of device malfunc-
tion
Any connected devices or device combina­tion, not complying with the requirements mentioned in these Instructions for Use can compromise the correct functioning of the medical device and lead to an electric shock. Before operating the medical device, strictly comply with the Instructions for Use of all connected devices or device combinations.

Safe connection with other electrical equipment

CAUTION
Risk of patient injury
Electrical connections to equipment not listed in these Instructions for Use or these Assembly Instructions must only be made when approved by each respective manufacturer.
This publication excludes references to various hazards which are obvious to a medical professional and operator of this medical device, to the consequences of medical device misuse, and to potentially adverse effects in patients with abnormal conditions.
Medical device modification or misuse can be dangerous.

Patient monitoring

The operators of the medical device are responsible for choosing appropriate safety monitoring that provides adequate information on medical device performance and patient condition.
Patient safety may be achieved through a wide variety of means, ranging from electronic surveil­lance of medical device performance and patient condition, to simple, direct observation of clinical signs.
The responsibility for the selection of the best level of patient monitoring lies solely with the medical device operator.
Information on Electromagnetic

Modifications

WARNING Modification of the device is not allowed.

Patient safety

The design of the medical device, the accompany­ing literature, and the labeling on the medical device are based on the assumption that the use of the equipment is restricted to trained professionals, and that certain inherent characteristics of the med­ical device are known to the trained operator. Instructions, warnings, and caution statements are limited, therefore, largely to the specifics of the Dräger design.
Instructions for Use Oxylog 3000 plus SW 1.n 9
Compatibility
General information on electromagnetic compatibility (EMC) pursuant to international EMC standard IEC 60601-1-2:
Electromedical devices are subject to special precautionary measures concerning electromagnetic compatibility (EMC) and must be installed and put into operation in accordance with the EMC information. Refer to section "Technical Documentation for the Oxylog 3000 plus according to EMC standard IEC 60601-1-2" on page 152.
WARNING Do not use portable and mobile HF communi-
cations equipment, e.g., mobile phones, in the vicinity of the medical device.
For Your Safety and that of Your Patients

Functional safety

The essential performance of the Oxylog 3000 plus is defined as:
Appropriate delivery of ventilation to the patient­connection port or generation of an alarm condition.

Appropriate monitoring

CAUTION Always use a separate SpO2 monitor for
patients who are dependent on an exact O concentration.
The monitoring functionality of the Oxylog 3000 plus ensures appropriate monitoring of ventilation therapy. To ensure appropriate monitoring during ventilation, always set the following alarm limits:
– Airway pressure, Paw
– Expiratory minute volume, MVe
– Respiratory rate (if applicable), RR
–etCO
2 (if applicable)
Not setting appropriate alarm limits could suppress alarms related to:
– Acute changes in the patient’s condition
– Incorrect settings and faulty handling
– Hose system leakage
2
10
Instructions for Use Oxylog 3000 plus SW 1.n

Product-specific Safety Information

For Your Safety and that of Your Patients
WARNING Ventilation monitoring is mandatory at all
times! Whenever a patient is connected to the ventilator, constant attention by qualified medical staff is required in order to provide immediate corrective action in case of a malfunction.
The operator shall not solely rely on the built­in monitoring of the ventilator and must always assume full responsibility for proper ventilation and patient safety in all situations.
WARNING Keep a manual breathing bag available.
If a failure is detected in the ventilator and its life-support functions can no longer be guaranteed (e.g. in case of a power supply failure or interruption in the O2 supply), ventilation must be started without delay with an independent ventilator (breathing bag) – using PEEP and/or increased inspired O concentration as necessary.
WARNING To ensure proper ventilation, consider the
total dead space of the breathing circuit, espe­cially when using small tidal volumes. Observe for signs of rebreathing.
Risk of CO2 rebreathing.
2

Installing accessories

CAUTION
Installations on the Oxylog 3000 plus must be done in accordance with these Instructions for Use. Make sure that the connections are securely fitted onto the basic system.
Strictly follow the Assembly Instructions and Instructions for Use.

Instructions for Use only available once

NOTE
Only one copy of the Instructions for Use is included in the clinical package and should there­fore be kept in an accessible location for users.
NOTE
An etCO2 value by itself is insufficient as a basis for medical decisions.
Instructions for Use Oxylog 3000 plus SW 1.n 11
This page intentionally left blank

Application

Intended use . . . . . . . . . . . . . . . . . . . . . . . . . 14
Indications / Contraindications. . . . . . . . . . 14
Environment of use . . . . . . . . . . . . . . . . . . . 15
Application
Instructions for Use Oxylog 3000 plus SW 1.n 13
Application

Intended use

The Oxylog 3000 plus is a time-cycled, volume­controlled and pressure-controlled emergency and transport ventilator for patients requiring mandatory or assisted ventilation with a tidal volume from 50 mL upwards.
Intended operator: the device is intended for use by and under the supervision of trained healthcare professionals, e.g. doctors, nurses, emergency medical technicians, respiratory therapists, and paramedics.

Indications / Contraindications

The device is intended for use with patients with a tidal volume of 50 mL upwards.
WARNING The Oxylog 3000 plus ventilator may only be
used under the supervision of trained healthcare professionals in case immediate corrective action is required in the event of a device failure.
14
Instructions for Use Oxylog 3000 plus SW 1.n

Environment of use

Intended environment of use:
– Mobile use for emergency patients, in both
outdoor and indoor environments.
– During transport in ambulances or aircraft,
including helicopters.
– In accident and emergency departments.
– When moving ventilated patients around the
hospital.
– In the recovery room.
WARNING Do not use the medical device in hyperbaric
chambers.
The medical device may malfunction, causing danger to the patient.
WARNING Do not use the medical device in conjunction
with magnetic resonance imaging (MRI, NMR, NMI).
The medical device may malfunction, causing danger to the patient.
Application
WARNING This medical device is neither approved nor
certified for use in areas where combustible or explosive gas mixtures are likely to occur.
Instructions for Use Oxylog 3000 plus SW 1.n 15
This page intentionally left blank

System Overview

Front panel with all options. . . . . . . . . . . . . 18
Side view, right . . . . . . . . . . . . . . . . . . . . . . . . 19
Rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Adult hose system, reusable . . . . . . . . . . . . . 20
Adult hose system, disposable. . . . . . . . . . . . 20
Paediatric hose system, disposable . . . . . . . . 21
Range of functions . . . . . . . . . . . . . . . . . . . . 22
Ventilation functions of the Oxylog 3000 plus. 22
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . 23
Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
System Overview
Instructions for Use Oxylog 3000 plus SW 1.n 17
System Overview
ABCDEFG
H
I J K
L
M
NOPQRS
T
U
V

Front panel with all options

A Screen with screen pages for the specific
application
B Key Alarms to display the alarm settings
in the "Settings and Alarms" window and to change screen pages
C Key Settings to display ventilation
parameters (ventilation screen) in the "Settings and Alarms" window and to change screen pages
D Key for setting the ventilation mode SpnCPAP
E Key for setting the ventilation modes VC-CMV /
VC-AC
F Key for setting the ventilation mode VC-SIMV
G Key for setting the ventilation mode PC-BIPAP
H Red and yellow alarm indicators
I Key for suppressing the acoustic alarm sig-
nal for 2 minutes
J Key Alarm Reset for acknowledging alarm
messages
K Key O
100 % O2 for 100 % O2 application, depending
on the option installed at manufacture
2 inhalation for O2 inhalation or key
L Key Insp. Hold for initiating a manual inspira-
tion or for extending the current inspiratory time.
M Key Start/Standby
N Display symbols for the power supply
Charge status of the internal battery Mains power supply connected
O Rotary knob for making selections, changing
and confirming settings
P Control knob for setting the O
2 concentration
FiO2
Q Control knob for setting the maximum
inspiratory pressure Pmax
R Control knob for setting the respiratory rate RR
S Control knob for setting the tidal volume VT
T Explanation of color codes for quick pre-setting
of RR and VT
U Key Curves to change between the
pressure, flow or CO and large presentation
V Key Values to change screen pages in
the "Measured Values" window
Curves
2 (optional) curve in small
Front
18
Instructions for Use Oxylog 3000 plus SW 1.n
System Overview
B
C
D
E F
A
G
H
A
B
C

Side view, right

A Emergency air intake
CAUTION
Do not block the emergency air intake. This may result in ventilator malfunction.
B Knob to secure the battery compartment cover
C Connectors for flow measuring lines
D Gas outlet for breathing hose
E Connector for O
F Connector for power supply
G Connector for CO
H Connector for data communication cable
2 supply
2 sensor

Rear view

Side
A Emergency air intake
CAUTION
Do not block the air intake. This may result in ventilator malfunction.
B Fresh-gas intake with a filter cartridge
C Protection bracket
CAUTION
Do not use the protection bracket as a handle. Tilting the device to a vertical position may lead to airway pressure oscillation.
Rear
Instructions for Use Oxylog 3000 plus SW 1.n 19
System Overview
A
E
B
C
D
E
D
C
A
B

Adult hose system, reusable

A Angled connector
B Flow sensor
C Breathing valve
D Breathing hose
E Flow and pressure measuring hoses

Adult hose system, disposable

001
A Angled connector
B Flow sensor
C Breathing valve
D Breathing hose
E Flow and pressure measuring hoses
002
20
Instructions for Use Oxylog 3000 plus SW 1.n

Paediatric hose system, disposable

D
A
E
C
B
A Angled connector
B Flow sensor
C Breathing valve
D Breathing hose
E Flow and pressure measuring hoses
System Overview
08137170
Instructions for Use Oxylog 3000 plus SW 1.n 21
System Overview

Range of functions

Ventilation functions of the Oxylog 3000 plus

Ventilation modes:
– Volume-controlled ventilation:
– VC-CMV / VC-AC,
–VC-SIMV.
– Pressure-controlled ventilation:
–PC-BIPAP.
– Support of spontaneous breathing:
– SpnCPAP.
Additional settings for ventilation:
– Pressure Support: in the ventilation modes
VC-SIMV, PC-BIPAP, SpnCPAP,
– Apnoea ventilation: in the ventilation mode
SpnCPAP,
– AutoFlow (optional): in the ventilation modes
VC-CMV, VC-AC and VC-SIMV.
– NIV: in the ventilation modes: SpnCPAP (/PS),
PC-BIPAP (/PS), VC-CMV / AF, VC-AC / AF and VC-SIMV / AF.
For a detailed description of the ventilation modes and the additional settings, refer to "Principles of Operation" on page 157. For abbreviations, see "Abbreviations" on page 23.
NOTE
In these Instructions for Use the unit of measure­ment for airway pressure is expressed in [mbar]. However, in some languages the display of the Oxylog 3000 plus shows [cmH
One [mbar] equals approximately one [cmH
2O].
2O].
Special procedures:
– Inspiration hold,
–O
2 inhalation (optional), with an inhalation
mask.
22
Instructions for Use Oxylog 3000 plus SW 1.n

Abbreviations

System Overview
Abbreviation Explanation
100 % O
2 100 % O2 flow
AF AutoFlow
BF Body Floating
bpm Breaths per minute
BTPS Body Temperature, Pressure
Saturated
C Lung compliance
2 Carbon dioxide
CO
CSM Customer Service Mode
ΔPsupp Positive pressure above PEEP
EMC Electromagnetic Compatibility
ESD Electrostatic Discharge
2 Endtidal CO2 concentration
etCO
2 Fractional inspired oxygen
FiO
concentration
FRC Functional Residual Capacity
HME Heat and Moisture Exchange
I:E Ratio inspiratory time to expiratory
time
IPX2 Ingress Protection level 2
IPX4 Ingress Protection level 4
MVe Expiratory minute volume
MVi Inspiratory minute volume
MVespon Spontaneous expiratory minute
volume
NIV Non-invasive ventilation –
mask ventilation
2 Oxygen
O
2-Inhalat. O2 inhalation
O
Paw Airway pressure
PC-BIPAP Pressure Controlled - Biphasic
Positive Airway Pressure
PEEP Positive End Expiratory Pressure
Abbreviation Explanation
PIF Peak Inspiratory Flow
Pinsp Inspiratory pressure
PIP Peak Inspiratory Pressure
Pmax Maximum allowed inspiratory
pressure
Pmean Mean airway pressure
Pplat Plateau pressure
PS Pressure Support
R Resistance
RF Radio Frequency
RR Respiratory Rate (frequency)
RRapn Respiratory Rate during apnoea
ventilation
RRsp Spontaneous Respiratory Rate
SpnCPAP Spontaneous Continuous Positive
Airway Pressure
2 Saturation of peripheral oxygen
SpO
Tapn Time before apnoea is recognized
Te Expiratory time
Ti Inspiratory time
Tplat % Plateau time in % of inspiratory time
UN United Nations
VC-AC Volume Controlled - Assist Control
VC-CMV Volume Controlled - Controlled
Mandatory Ventilation
VC-SIMV Volume Controlled - Synchronized
Intermittent Mandatory Ventilation
VT Tidal volume
VTapn Tidal volume during apnoea
ventilation
VTe Expiratory tidal volume
VTi Inspiratory tidal volume
Instructions for Use Oxylog 3000 plus SW 1.n 23
System Overview
Insp. hold
Alarms

Symbols

Symbol Explanation
Key for initiating a manual
inspiration or for extending the current inspiratory time.
Key to display ventilation
Settings
parameters (ventilation screen) in the "Settings and Alarms" window and to change screen pages.
Key to display the alarm settings in
the "Settings and Alarms" window and to change screen pages.
Key to change screen pages in the
Values
"Measured Values" window.
Key to change between the
Curves
pressure, flow or CO
2 (optional)
curve in small and large presentation.
Key for suppressing the acoustic
alarm signal for 2 minutes.
Key for acknowledging alarm
Alarm
Reset
messages.
Rotary knob
Start / Standby key
Upper alarm limit
Lower alarm limit
! Advisory message
!! Caution message
Symbol Explanation
* Trigger indicator
Warning
Defibrillation-proof type BF applied
part
Charge status of the internal battery
External power supply connected
Battery charge (example: three quarters full)
Class II equipment, device pro­tected against electric shock with additional safety precautions such as double or reinforced insulations, without protective earthing.
Do not dispose of the device as municipal waste.
Manufacturing date
Manufacturer
DC input
Operating instructions
Follow Instructions for Use
!!! Warning message
24
Instructions for Use Oxylog 3000 plus SW 1.n
System Overview
Symbol Explanation
For dry locations only
Caution, hot surface!
Warning, dangerous voltage!
Temperature limitations
Latex free
Do not reuse
Do not use oil and grease
Non-sterile
Symbol Explanation
Slope (steep, medium, flat)
Quantity
Keep away from sunlight
For indoor use only
Do not open
Prohibition: Do not obstruct emer­gency air intake or fresh gas intake
Pediatric
Adult
Paediatric
Adult
Instructions for Use Oxylog 3000 plus SW 1.n 25
This page intentionally left blank

Operating Concept

Switch on or off . . . . . . . . . . . . . . . . . . . . . . 28
Switch on . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Switch off . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Ventilation controls . . . . . . . . . . . . . . . . . . . 29
Display operating controls . . . . . . . . . . . . . 30
Additional function keys . . . . . . . . . . . . . . . 31
Screen window structure. . . . . . . . . . . . . . . 32
Operating Concept
Instructions for Use Oxylog 3000 plus SW 1.n 27
Operating Concept
Oxylog 3000 plus
AB

Switch on or off

Switch on

z To switch the device ON, briefly press the
key (A).
00337170

Switch off

Refer to "Shutdown" on page 82.
28
Instructions for Use Oxylog 3000 plus SW 1.n

Ventilation controls

BC
A
Oxylog 3000 plus
Operating Concept

General ventilation controls

A Keys for selecting the ventilation modes:
– VC-CMV / VC-AC
–VC-SIMV
– SpnCPAP
– PC-BIPAP.
B Rotary knob.
C Ventilation parameter controls:
– Inspiratory tidal volume VT [mL],
– Ventilation respiratory rate RR [/min],
– Maximum inspiratory pressure Pmax
[mbar],
–O
2 concentration FiO2 [%].

Selecting the ventilation mode

Oxylog 3000 plus
00137170
BC
z Press the appropriate ventilation mode key (A)
for approximately 3 seconds.
Or
1 Press the appropriate ventilation mode key (A).
2 Press the rotary knob (B) to confirm.
The selected ventilation mode will be activated.
The active ventilation mode is displayed in the upper left corner of the display (C).
A
08537170
Instructions for Use Oxylog 3000 plus SW 1.n 29
Operating Concept
Oxylog 3000 plus
B C
D
E
A
Curves

Display operating controls

A Rotary knob for making selections, changing
and confirming settings.
NOTE
Different parameters can be set in the display window via the rotary knob (e.g. Ti, PEEP, ΔPsupp, Pinsp). – To select the parameter: turn rotary knob. – To activate the parameter: press rotary knob. – To set the value: turn rotary knob. – To confirm the value: press rotary knob.
00237170
B Key Values to change screen pages in
the "Measured Values" window.
C Key Curves to change between the
pressure, flow or CO2 (optional) curve in small and large presentation.
D Key Settings to display ventilation
parameters (ventilation screen) in the "Settings and Alarms" window and to change screen pages.
E Key Alarms to display the alarm settings
in the "Settings and Alarms" window and to change screen pages.
30
Instructions for Use Oxylog 3000 plus SW 1.n

Additional function keys

Oxylog 3000 plus
A B C D
Additional keys are positioned on the right side of the front panel:
A key for suppressing the acoustic alarm sig-
nal for 2 minutes.
B Key Alarm Reset for acknowledging alarm
messages.
C Key Insp. Hold for initiating a manual inspira-
tion or for extending the current inspiratory time.
D Key O
2 inhalation for O2 inhalation or key
100 % O2 for 100 % O2 application, depending
on the option installed at manufacture.
Operating Concept
03237170
Instructions for Use Oxylog 3000 plus SW 1.n 31
Operating Concept
A
B
D
C
E
F
BA D
E
C

Screen window structure

General window structure

A Ventilation mode.
B Alarm messages field.
C Measured values window.
D Curve window.
D and E are combined for a large curve screen.
E Setting and alarm window.
F Information window. For information on the
content, refer to "Messages in the information window" on page 120.

Measured values window

05437171
A Parameter measured.
B Measured value.
C Unit of measure.
D Measured values 1/7: 1st page of 7 available
pages. If CO
2 option is not installed: 1/6 available
pages.
E Trigger indicator.
The last page shows an overview of all measured values.
0583717115637171
32
Instructions for Use Oxylog 3000 plus SW 1.n
Operating Concept
BA
A B

Settings window

A Menu for setting supplementary ventilation
parameters in accordance with the desired ventilation mode.
– AutoFlow (optional)
– Brightness
–CO
2 filter check (optional)
–CO
2 zero calibration (optional)
– Cuvette type (optional)
– HME correction
– Hose type
– I:E / Ti
–NIV
–PEEP
–Pinsp
– RRapn and VTapn
– Slope
– Tapn
– Tplat
– Trigger
Δ
Psupp
B Page number: e.g. 1st page of 3 available
pages.
z Press Settings key.
The pages are displayed consecutively.

Alarms window

05737171
A Menu for alarm limits and alarm parameters.
For detailed operating instructions, see "Setting alarm limits" on page 87.
B Page number: e.g. 1st page of 2 available
pages.
To advance to the next page:
z Press the Alarms key.
The pages are displayed consecutively.
10937171
Instructions for Use Oxylog 3000 plus SW 1.n 33
Operating Concept

Pressure curve large view

Curve window showing the airway pressure curve Paw.
Curves
z Press Curves key multiple times.

Flow curve large view

055371710563717115437171
Curve window showing the flow curve.
CO
2 curve large view
Curve window showing the CO2 curve.
34
Instructions for Use Oxylog 3000 plus SW 1.n

Assembly

Internal rechargeable battery . . . . . . . . . . . 37
Removing the battery . . . . . . . . . . . . . . . . . . . 37
Checking the charge status of the battery . . . 37
Installing the battery . . . . . . . . . . . . . . . . . . . . 37
Connecting the power supply . . . . . . . . . . . 38
External power supply . . . . . . . . . . . . . . . . . . 38
External power supply . . . . . . . . . . . . . . . . . 39
External power supply with DC/DC converter 39 External power supply from mains voltage
(AC/DC power pack). . . . . . . . . . . . . . . . . . . . 40
Connecting the gas supply . . . . . . . . . . . . . 41
Supply from an O2 cylinder . . . . . . . . . . . . . . 41
Supply from a piped O2 system . . . . . . . . . . . 42
Assembling the adult reusable hose
system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Breathing valve assembly. . . . . . . . . . . . . . . . 43
Hose connections. . . . . . . . . . . . . . . . . . . . . . 44
Assembly
Connecting the adult disposable hose
system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Connecting the paediatric disposable
hose system . . . . . . . . . . . . . . . . . . . . . . . . . 46
Connecting the bacterial filter or HME . . . . 47
Connecting the CO
cuvette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Hanging the Oxylog 3000 plus on
standard rail systems. . . . . . . . . . . . . . . . . . 49
Instructions for Use Oxylog 3000 plus SW 1.n 35
2 sensor and the
Assembly
WARNING Avoid tripping on, or ensnaring, the breathing
hose, CO
2 sensor cable, AC/DC supply cables,
DC/DC supply cables or compressed gas hose.
There is a risk of injury and a risk of accidental extubation of the patient.
WARNING Do not kink the patient breathing hoses while
ventilating.
Risk of asphyxiation or hypoventilation.
WARNING Do not use any damaged parts or accessories.
Damaged or deformed parts must be replaced.
WARNING Data communication between the
Oxylog 3000 plus and other equipment is only supported when using the MEDIBUS protocol.
WARNING Electrical connections to equipment, which
are not listed in these Instructions for Use, should only be made following consultation with the respective manufacturers.
Equipment malfunction may result as well as risk of patient injury.
WARNI NG Always use the angled connector of the hose
system.
If an angled connector is not used, the minute volume may be measured incorrectly.
WARNI NG Do not use an adult breathing hose for tidal
volumes below 100 mL.
Risk of CO
2 rebreathing.
CAUTION
Do not use the Oxylog 3000 plus without a dust filter.
Risk of patient inhaling dust or device damage.
CAUTION
Do not use electrically conductive hoses.
This can endanger the operator and damage the device during defibrillation.
Risk of electric shock.
WARNING All equipment connected to the
Oxylog 3000 plus must comply with IEC 60601-1-2.
WARNING Do not combine parts of different hose sys-
tems, especially for paediatric applications.
Risk of CO2 rebreathing.
36
Instructions for Use Oxylog 3000 plus SW 1.n

Internal rechargeable battery

A
B
C
Assembly
Internal power is provided by means of a removable rechargeable battery. For technical information, refer to "Technical Data" on page 139.

Removing the battery

1 Turn the knob (C) on the battery compartment
cover (B) counterclockwise to release the cover.
2 Open the battery cover.
3 Remove the battery (A) by pulling the tab.

Installing the battery

1 Insert the battery into the battery compartment.
2 Close the battery cover.
3 Tighten the knob by turning it.
CAUTION
The Oxylog 3000 plus will interrupt ventilation when the battery is replaced while the device is switched on and the external power supply is not connected. Ventilation will always resume with the last values settings approximately 3 seconds after inserting a recharged battery.
015

Checking the charge status of the battery

z Press the button on the rechargeable battery.
The charge status is indicated as a percentage by an indicator.
Instructions for Use Oxylog 3000 plus SW 1.n 37
Assembly

Connecting the power supply

External power supply

To recharge the battery and to extend the electrical operation time, use either:
– DC/DC converter, or
– AC/DC power pack.
For more information refer to page 146.
WARNING A fully charged battery must always be
installed for safety reasons, even when operating from an external power supply.
CAUTION
Without a charged battery installed, ventilation will be interrupted in case of an external power failure.
NOTE
It is recommended to have a fully charged spare battery available when using the Oxylog 3000 plus.
Always position the device so that the external power connector can be easily disconnected from the ventilator.
38
Instructions for Use Oxylog 3000 plus SW 1.n

External power supply

Oxylo
g 3000 plus
12 VDC 24 VDC 28 VDC
D
B
A
C

External power supply with DC/DC converter

WARNING The DC/DC converter should be used in dry
locations only.
Risk of electric shock or equipment damage.
The DC/DC converter must be used to connect the Oxylog 3000 plus to onboard DC power supply systems, e.g. in ambulances. It can be used with the following voltages: 12 VDC, 24 VDC or 28 VDC. The onboard power supply shall have a fuse of 10 to 16 A, suitable for DC current. Outside this range the Oxylog 3000 plus cannot use the DC input power.
Mount the DC/DC converter on a flat wall and make sure the wall is solid enough to support the bracket. Use all four mounting holes (screw size M4).
Assembly
016
1 Plug the large connector (A) of the DC/DC
converter into the on-board supply.
2 Plug the small connector (B) into the DC
connector (C) of the Oxylog 3000 plus.
3 When the Oxylog 3000 plus is correctly
connected to an external power supply, the indicator (D) lights up.
Instructions for Use Oxylog 3000 plus SW 1.n 39
Assembly
Oxylog 3000 plus
A
B
D
C

External power supply from mains voltage (AC/DC power pack)

WARNING The AC/DC power pack must not be used
outdoors.
Risk of electric shock or equipment damage.
1 Connect the mains plug (A) to the mains outlet.
2 Connect the DC connector (B) to the DC
connector (C) of the Oxylog 3000 plus.
3 When the Oxylog 3000 plus is correctly
connected to an external supply, the indicator
(D) lights up.
To isolate the ventilator system from mains, discon­nect the power cable from the wall connector.
017
40
Instructions for Use Oxylog 3000 plus SW 1.n

Connecting the gas supply

Assembly
Take care when handling O2:
WARNING Secure O2 cylinders so they cannot fall over.
Keep away from excessive heat.
Risk of explosion.
WARNING Do not grease or lubricate O2 fittings, such as
cylinder valves and pressure reducers and do not handle with greasy hands.
Risk of fire.
WARNING Operate cylinder valves by hand and rotate
slowly to prevent the risk of fire or explosion.
Do not use tools.
WARNING Only use medical grade oxygen.
WARNING Always provide adequate ventilation in the
area where the ventilator is being operated, in order to maintain ambient O2 concentration below 25 %, to prevent risk of fire.

Supply from an O2 cylinder

WARNING Always use gas cylinders and pressure regu-
lators that comply with all applicable regula­tions.
CAUTION
Always use a full O2 cylinder.
Risk of asphyxiation.
1 Connect the pressure reducer (270 to 600 kPa
delivery pressure, 500 kPa nominal pressure) to the O2 cylinder.
WARNING Only use a pressure reducer with a relief valve
at the outlet to limit the delivery pressure to a maximum of 1000 kPa in case of a malfunction, to prevent damage to the ventilator through excessive O pressure on the input.
2 supply
WARNING No smoking or open flames.
O
2 enhances combustion of other substances
and can intensify fires.
Instructions for Use Oxylog 3000 plus SW 1.n 41
Assembly
O
2
A
B
C
A
B

Supply from a piped O2 system

2 Connect the O2 hose (A) to the
Oxylog 3000 plus.
3 Connect the O
(B).
4 Rotate the cylinder valve (C) slowly and open
fully.
WARNING Do not connect flow control valves or flowme-
ters in the gas supply to the Oxylog 3000 plus.
The ventilator could malfunction.
WARNING Always check the O2 pressure of cylinder
before use, to prevent insufficient oxygen supply during use.
2 hose to the pressure reducer
42
018
1 Connect the O2 hose (A) to the
Oxylog 3000 plus.
2 Connect the gas probe (B) to the O
unit until it has properly engaged and the supply
2 is assured.
of O
Instructions for Use Oxylog 3000 plus SW 1.n
019
2 terminal

Assembling the adult reusable hose system

A
A
B
C
D
E
Reusable parts must always be sterilized before use!

Breathing valve assembly

008
WARNING The rubber disc (A) in the housing must not be
removed, damaged or bent, otherwise the valve will not work properly and will endanger the patient.
Risk of CO2 rebreathing.
1 Place the diaphragm (B) in the breathing valve
2 Fit the cover (A) and turn it approximately 60
3 Push the flow sensor (D) onto the breathing
4 Push the angled connector (E) onto the flow
Assembly
housing (C). Ensure that it is inserted correctly.
o
clockwise to secure into position (a click can be felt).
valve (C). Note the correct alignment of the parts by the groove in the flow sensor (D) and the notch on the breathing valve (C).
sensor (D).
009
Instructions for Use Oxylog 3000 plus SW 1.n 43
Assembly
A
B B
B
A

Hose connections

1 Connect the breathing hose (A) to the breathing
valve.
2 Connect the flow measuring lines (B) to the noz-
zles on the flow sensor. Note the different diam­eters of the hoses and the nozzles when connecting the flow measuring lines and con­nect to the correct side.
3 Connect the flow measuring lines (A) to the
Oxylog 3000 plus. Correct alignment is indicated by a notch on the connector, which must point away from the breathing hose. Otherwise, the set will not fit and the measured values will be incorrect.
4 Connect the breathing hose (B) to the gas outlet
on the Oxylog 3000 plus.
When connecting a hose, check that the hose
011012
setting in the Settings window corresponds to the connected hose.
44
Instructions for Use Oxylog 3000 plus SW 1.n

Connecting the adult disposable hose system

A
B
NOTE
Using a disposable hose may reduce the risk of cross-infection.
Assembly
013
1 Connect the flow measuring lines (A) to the
Oxylog 3000 plus. Correct alignment is indi­cated by a notch on the connector, which must point away from the breathing hose. Otherwise, the set will not fit and the measured values will be incorrect.
2 Connect the breathing hose (B) to the gas outlet
on the Oxylog 3000 plus.
When connecting a hose, check that the hose setting in the Settings window corresponds to the connected hose.
Instructions for Use Oxylog 3000 plus SW 1.n 45
Assembly
B
C
A

Connecting the paediatric disposable hose system

1 Connect the blue flow measuring line (B) to the
blue labeled connector.
2 Connect the transparent flow measuring line (A)
to the other connector.
3 Connect the breathing hose (C) to the gas outlet
on the Oxylog 3000 plus.
When connecting a hose, check that the hose setting in the Settings window corresponds to the connected hose.
46
013
Instructions for Use Oxylog 3000 plus SW 1.n

Connecting the bacterial filter or HME

Assembly
WARNING Bacterial filters, HME, and masks increase the
resistance and dead space volume of the ventilation equipment. Note the manufacturer’s directions.
Risk of CO2 rebreathing.
NOTE
When using an HME, the measured flow may deviate from the actual expiratory flow, as temperature and humidity of the gas are reduced.
The flow and volume measurements can be corrected for use with an HME. Refer to "Setting HME correction" on page 80.
It is recommended to use a bacterial filter between ventilator and patient, to reduce the risk of bacteria, viruses, fungi or spores being present in the inspiratory flow.
z Connect the bacterial filter or HME to the angled
connector as follows.
Adult reusable hose:
Adult disposable hose:
014050
Paediatric hose:
010
Instructions for Use Oxylog 3000 plus SW 1.n 47
Assembly
A
B
C
D

Connecting the CO2 sensor and the cuvette

51
1 Disconnect the angled connector (A) from the
flow sensor (D).
2 Attach the cuvette (C) to the flow sensor (D),
with the cuvette windows facing the side.
3 Attach the angled connector (A) to the
cuvette (C).
4 Push the CO
with the cable towards the device.
5 Plug the CO
Oxylog 3000 plus. For the connector location, refer to the section "Side view, right" on page 19.
6 Insert the CO
the hose.
Alternatively, connect the cuvette (C) directly to the patient side of the angled connector (A), without disconnecting the angled connector from the flow sensor (D).
The CO maximum of one extension cable. Refer to "List of Accessories" on page 167.
For CO2 zero calibration and filter check before ventilation, see page 58. For CO2 zero calibration and filter check during ventilation, see page 91. For CO page 91. For CO2 configuration in the Customer Service Mode see page 105.
2 sensor cable can be extended with a
2 measurement and cuvette type selection see
2 sensor (B) onto the cuvette (C),
2 sensor into the connector of the
2 sensor cable in the cable clips on
48
Instructions for Use Oxylog 3000 plus SW 1.n

Hanging the Oxylog 3000 plus on standard rail systems

The Oxylog 3000 plus can be hung on various rail systems measuring up to 35 mm diameter by means of the claw.
– Ensure that the rail is completely inserted in the
claw.
– To ensure optimal functioning of the claw, a
distance of at least 25 mm between rail and wall is required.
WARNING Be careful when placing the ventilator on the
rail or bed rim.
Risk of damage to property or personal injury.
CAUTION
The Oxylog 3000 plus is only held by its own weight when hung on a bar or rail. The Oxylog 3000 plus must be secured additionally when being transported, otherwise vibrations may cause accidental dislodgement.
Assembly
Instructions for Use Oxylog 3000 plus SW 1.n 49
This page intentionally left blank

Getting Started

Charging the battery . . . . . . . . . . . . . . . . . . 52
Indication of battery capacity / battery
operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Determining the approximate pneumatic
operating time. . . . . . . . . . . . . . . . . . . . . . . . 53
Checking readiness for operation . . . . . . . 54
Perform device check. . . . . . . . . . . . . . . . . . . 54
2 zero calibration and filter check
CO
before ventilation (optional) . . . . . . . . . . . . 58
Zero calibration before ventilation . . . . . . . . . 58
2 filter check before ventilation . . . . . . . . . 59
CO
Preparation for use after system check, CO
2 zero calibration and filter check . . . . . 60
Getting Started
Instructions for Use Oxylog 3000 plus SW 1.n 51
Getting Started
Oxylog 3000 plus
BA

Charging the battery

The actual screen display may differ in appearance or configuration.
NOTE
The ambient temperature must be between 0 and 35 °C when charging the batteries.
When an external supply is available:
1 The green indicator (A) lights up when an
external power source is connected.
2 A three colored indicator (B) lights up to
show the current charge status of the internal battery:
– Green: the battery is fully charged.
– Yellow: the battery is being charged.
– Red: a battery is not inserted or cannot be
charged.
– Indicators (A) and (B) remain off while the
ventilator is being operated from the internal battery.

Indication of battery capacity / battery operation

The remaining capacity of the battery is indicated by Oxylog 3000 plus in 25 % increments in the lower right section of the information window when power is ON.
As an example, in the above screen the battery is 75 % charged.
– The accuracy of the battery capacity indicator
02437170
can vary, depending on the age and condition of the battery. Refer to "Technical Data" on page 139 for additional information.
– The capacity indication is overwritten when
other messages need to be shown in the Information window.
– Additional alarms can draw attention to the
remaining operating time of the battery. Refer to the table "Alarm - Cause - Remedy" on page 112.
For screen brightness during battery operation, refer to "Screen brightness" on page 81.
10637171
An external battery charging station connected to the mains power supply can be used to charge an extra battery. Refer to the "List of Accessories" on page 167 for additional information.
52
Instructions for Use Oxylog 3000 plus SW 1.n

Determining the approximate pneumatic operating time

O2 supply [L] (MV +0.5*) [L/min]
420
5.8
A
Getting Started
Example for supply of O2:
– Cylinder pressure measured on the pressure
gauge of the pressure reducer: 20,000 kPa (200 bar)
– Liquid capacity of the O
Supply of O
2:
2 cylinder: 2.1 L
2.1 L x 20,000 kPa = approx. 420 L at environmen­tal pressure level.
Example for pneumatic operation time:
– VC-CMV mode, frequency 10 breaths/min,
VT = 0.53 L, O
2 = 100 %
– Minute volume = 10 breaths/min x 0.53 L =
5.3 L/min
Operation time =
* Calculated with average gas consumption of ventilator: 0.5 L/min
Example:
13637171
A O2 consumption = 5.3 L/min
Operation time = = approx. 72 minutes
The pneumatic operation time increases when Oxylog 3000 plus operates with O of less than 100 % O2, as ambient air is drawn into the device.
The amount of gas from the high-pressure supply, which is currently being consumed, is indicated by the Oxylog 3000 plus in the lower left section of the information window in L/min. This display is over­written when a higher priority message is activated.
Instructions for Use Oxylog 3000 plus SW 1.n 53
2 concentration
Getting Started
A
C
B

Checking readiness for operation

The device check should be performed:
– Before every use of the device if the breathing
hose was changed.
– At least every six months.
The Oxylog 3000 plus interrupts the device check if a fault is detected.
The relevant fault is indicated on the screen.
WARNING The patient may be endangered if the device
check is not completed successfully.

Perform device check

The device check consists of the following steps:
– Connect the test lung
– Switch the device ON
– Check connections
– System check
– Power supply failure check.
The duration of the device check is approximately 3 minutes.
Connect the test lung
021
1 Make sure that the angled connector (A) is
connected to the flow sensor.
2 Connect the catheter mount (B) of the test lung,
diameter 7 mm, to the angled connector. The catheter mount simulates the resistance of the airways.
3 Connect the balloon (C) of the test lung. Refer
to "List of Accessories" on page 167.
NOTE
BTPS values of a test lung are not the same as the BTPS values of a patient. The Oxylog 3000 plus measures and adapts according to BTPS values of a patient. Therefore, when a test lung is connected, the MVe and VTe indicated on display may differ from the MVe and VTe that is set by the operator.
54
Instructions for Use Oxylog 3000 plus SW 1.n
Getting Started
Oxylog 3000 plus
A
B
Switch the device ON
1 To switch the device ON briefly press
the key (A).
The device performs a selftest and the operator is prompted, on the display, to activate the configuration menu or device check:
Press rotary knob for device check and configuration
2 Press the rotary knob (B) to confirm, before the
black progess bar is complete. The start-up screen appears:
3 Select Device check in the start-up menu and
02211937171
confirm.
NOTE
The device check can be discontinued at any time by pressing the Alarm Reset key.
NOTE
During device check, the connections (gas supply, hose type) and the system (flow, pressure levels, alarm signals and knobs) are checked.
14737171
Instructions for Use Oxylog 3000 plus SW 1.n 55
Getting Started
Oxylog 3000 plus
A
Check connections
1 Ensure that the gas supply has been
connected.
2 Select and confirm the appropriate hose type.
3 Ensure that the test lung has been connected.
The Oxylog 3000 plus automatically checks if a test lung has been connected. The device check is interrupted if a test lung is not detected within one minute. The check is continued when the test lung is detected.
4 The Oxylog 3000 plus automatically checks if
the detected hose differs from the selected hose type.
If the wrong hose type is selected:
– Press the Alarm Reset key to cancel the
device check.
– Restart the device check.
– Select the correct hose type.
System check
023
5 Set the control knobs (A) below the display to
the required values.
The Oxylog 3000 plus successively activates the audible and optical alarm signals and prompts the operator to acknowledge each signal.
6 Confirm the audible and optical alarm signals.
The device check continues automatically.
During the automatic test sequence, the Oxylog 3000 plus checks the flow, pressure lev­els and alarm signals. Corresponding sounds are heard.
The bar graph shows the progress made by the check.
The result is displayed on the last page of the device check screens. If all tests are completed successfully, the device will go to the last page. If a test fails, the device will go directly after the failed test to the last page, without performing the other tests.
After confirmation, the system returns to the menu screen.
56
If the service inspection date has been passed with­out servicing, the text Service date overdue ! will appear in the window after finishing the device check. In this case the device must be serviced immediately.
Instructions for Use Oxylog 3000 plus SW 1.n
Getting Started
High airway pressure and disconnection alarm check
Check the high airway pressure alarm
1 Ventilate the test lung in CMV mode.
2 Press the test lung manually, until the airway
pressure exceeds the set Pmax.
3 Check if the P
AW high alarm occurs.
Check the alarm in case of hose system disconnec­tion
1 Ventilate the test lung in CMV mode.
2 Disconnect the breathing hose and/or flow mea-
suring lines from the ventilator.
3 Check if an applicable alarm occurs.
Power supply failure check
A monthly check of the power supply failure alarm is recommended.
1 Switch the device on.
2 Disconnect the external power supply.
3 Remove the battery to activate the acoustic
alarm signal.
4 Listen for the acoustic alarm signal.
If the device check is not completed successfully:
1 Refer to "Error messages during the device
check" on page 122 of the section "Problem Solving".
2 Contact your local DrägerService for support.
NOTE
If no alarm is heard, contact DrägerService.
5 When the power supply failure alarm test is
completed, reinstall the battery into the battery compartment of the Oxylog 3000 plus.
6 Connect the external power supply.
Troubleshooting
WARNING The ventilator is ready for operation only after
all functional tests have been successfully performed.
Instructions for Use Oxylog 3000 plus SW 1.n 57
Getting Started
Oxylog 3000 plus
A
B

CO2 zero calibration and filter check before ventilation (optional)

The CO2 zero calibration and filter check only work if the CO sensor is present.
1 To switch the device ON briefly press the
The device performs a selftest and the operator is prompted, on the display, to activate the configuration menu or device check:
Press rotary knob for device check and configuration
2 option has been installed and if the CO2
key (A).
3 Select CO
2 zero calibration and filter check
in the start-up menu and confirm.
The function CO
2 Zero Calibration and Filter
check is displayed only if the option is avail-
able.
NOTE
02211937171
The CO2 zero calibration and filter check can be discontinued at any time by pressing the Alarm Reset key.

Zero calibration before ventilation

The zero calibration is performed with a clean CO2 sensor that has been removed from the cuvette!
14737171
2 Press the rotary knob (B) to confirm, before the
black progress bar is complete.
58
NOTE
Do not breathe on the CO2 sensor during zero calibration, otherwise the zero calibration can fail or the zero calibration can pass with an invalid zero value.
Instructions for Use Oxylog 3000 plus SW 1.n
Getting Started
A
B
A
B

CO2 filter check before ventilation

NOTE
Before the CO2 filter check, you need to have fin­ished a successful CO wise the CO
2 filter check may be outside of the
tolerance range.
17914837171
1 Remove the CO2 sensor (A) from the cuvette
1 Remove the CO2 sensor (A) from the cuvette
(B).
(B).
2 Attach the CO
3 Select Filter check.
2 Select and activate Zero calibration. The
screen displays the text Remove sensor from cuvette. Confirm with rotary knob.
3 Confirm. The zero calibration starts and the line
displays Zero calibration in progress. After a
4 Confirm. The filter check starts and the screen
displays Filter check in progress. After a successful filter check, the line briefly displays Filter check OK.
5 Press Alarm Reset to exit.
successful zero calibration, the line briefly displays Zero calibration OK.
6 Attach the CO
4 Press Alarm Reset to exit.
5 Attach the CO
2 sensor back to the cuvette.
If the check was not successful:
The Oxylog 3000 plus displays the alarm Filter check failed. The test value is outside the
If zero calibration was not successful:
The Oxylog 3000 plus displays the alarm Zero calibration failed.
z Repeat zero calibration.
If zero calibration is still not possible:
1 Check whether the sensor (A) is soiled and
clean it if necessary. If the sensor is defective, replace the sensor.
2 Repeat zero calibration.
Instructions for Use Oxylog 3000 plus SW 1.n 59
permissible tolerance.
z Check whether the sensor (A) or test filter (B) is
soiled and clean them if necessary. Repeat the
2 filter check.
CO
If the check was still not successful:
z Check the CO
For connecting the CO page 48. For CO2 zero calibration and filter check during ventilation, see page 91. For CO2 measure­ment see page 91. For CO Customer Service Mode see page 105.
2 zero calibration. Other-
2 sensor (A) to the test filter (B).
2 sensor back to the cuvette.
2 calibration with test gas.
2 sensor and cuvette see
2 configuration in the
177
Getting Started

Preparation for use after system check, CO2 zero calibration and filter check

1 Assemble the Oxylog 3000 plus for operation.
Refer to "Assembly" on page 35.
2 Connect to the power supply and gas supply.
Refer to "Internal rechargeable battery" on page 37 and "Connecting the gas supply" on page 41.
3 Start the ventilator:
z Select Start ventilation from the menu and
confirm.
Or
z Press the Alarm Reset key.
14737171
60
Instructions for Use Oxylog 3000 plus SW 1.n

Operation

Starting operation. . . . . . . . . . . . . . . . . . . . . 62
Switch the device ON . . . . . . . . . . . . . . . . . . . 62
Preparing ventilation mode . . . . . . . . . . . . . 64
To activate a ventilation mode . . . . . . . . . . . . 64
Setting ventilation parameters . . . . . . . . . . . . 64
VC-CMV, VC-AC . . . . . . . . . . . . . . . . . . . . . . 65
Trigger (VC-AC) . . . . . . . . . . . . . . . . . . . . . . . 66
Setting AutoFlow (optional) . . . . . . . . . . . . . . 67
Cardio-pulmonary resuscitation (CPR). . . . . . 67
VC-SIMV, VC-SIMV/PS . . . . . . . . . . . . . . . . . 68
Setting Pressure Support VC-SIMV/PS . . . . . 69
Setting AutoFlow (optional) . . . . . . . . . . . . . . 70
PC-BIPAP, PC-BIPAP/PS . . . . . . . . . . . . . . . 71
Setting Pressure Support PC-BIPAP/PS . . . . 72
Operation
Sp nCPA P, SpnC PAP/ PS . . . . . . . . . . . . . . . . 73
Apnoea ventilation . . . . . . . . . . . . . . . . . . . . . 74
Setting Pressure Support Spn-CPAP/PS . . . . 75
Cardio-pulmonary resuscitation (CPR). . . . . . 75
NIV – Non-invasive ventilation (Mask
ventilation) . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Special functions . . . . . . . . . . . . . . . . . . . . . 77
Manual inspiration / Inspiration hold . . . . . . . . 77
100 % O2 (optional) . . . . . . . . . . . . . . . . . . . . 77
O2 inhalation (optional) . . . . . . . . . . . . . . . . . 77
O2 concentration with "O2 blending". . . . . 79
Setting HME correction . . . . . . . . . . . . . . . . 80
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Screen brightness . . . . . . . . . . . . . . . . . . . . 81
Alarm volume . . . . . . . . . . . . . . . . . . . . . . . . 81
Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Instructions for Use Oxylog 3000 plus SW 1.n 61
Operation
Oxylog 3000 plus
A
B

Starting operation

The actual screen display may differ in appearance or configuration.
WARNING Only use a ventilator that has been cleaned,
disinfected and successfully tested to be ready for operation, to prevent a health risk for the patient and user.
Refer to the chapter "Cleaning, Disinfection and Sterilization" on page 123.
CAUTION
When using the ventilator in very low ambient temperatures, always consider that the cold gas will expand due to the warming by the patients body. Carefully monitor the MVe.
Risk of hyperventilation.
CAUTION
In very high ambient temperatures, avoid mixing with ambient air: always set FiO
2 to 100 % O2.
Exposing the patient to very warm inspiratory gas may cause lung damage.

Switch the device ON

0133717011937171
z To switch the device ON, briefly press the
key (A).
62
The Oxylog 3000 plus performs a selftest. The selftest will be completed in approximately six seconds.
During the selftest, the system briefly displays the starting page with a bar graph indicating the progress of the selftest, the software version, the activated software options and a prompt for the operator to activate the device check by pressing the rotary knob (B).
Instructions for Use Oxylog 3000 plus SW 1.n
If the rotary knob (B) is not pressed during the self­test, the hose selection page is displayed.
Select the connected hose type, as seen in the above graphic, by rotating the rotary knob (B) and confirm by pressing the rotary knob (B). The ventilator now automatically begins ventilation with the default settings.
NOTE
As long as the hose selection page is shown, the patient is not being ventilated.
Operation
14937171
NOTE
The request to select the hose type can be configured. Refer to "Customer Service Mode" on page 98.
Starting screen with default settings.
The default settings can be configured in Customer Service Mode. Refer to "Set start-up settings" on page 100.
10437171
Instructions for Use Oxylog 3000 plus SW 1.n 63
Operation

Preparing ventilation mode

To activate a ventilation mode

1 Press and hold a ventilation mode key for
approximately 3 seconds.
Or
2 Press a ventilation mode key and confirm by
pressing the rotary knob.
The new ventilation mode selected is now effective.
For an overview of all ventilation modes, refer to "Range of functions" on page 22. For a detailed explanation on all ventilation modes, refer to "Principles of Operation" on page 157.

Setting ventilation parameters

1 Set the required control knob below the display.
Or
2 Select, set and confirm a parameter on the
display with the rotary knob.
If the changed settings are not confirmed after 5 seconds, the alarm ! Confirm settings appears. If the settings are still not confirmed after 10 seconds, the alarm ! Settings not confirmed appears. After that the former settings are restored.
12137171
When the PEEP setting is increased above 10 mbar, a message Confirm PEEP above 10 mbar? will appear to request confirmation of the change. The PEEP setting can be increased to the desired setting after the message is confirmed with the rotary knob.
The device can be configured to show Ti or I:E as a primary parameter that can be set. If Ti is configured as the primary parameter, I:E will be shown in the information window when Ti is selected, and vice versa. This configuration will apply to all ventilation modes. Refer to the "Customer Service Mode" on page 98.
64
Instructions for Use Oxylog 3000 plus SW 1.n

VC-CMV, VC-AC

Paw
Flow
T
insp
T
e
P
plat
Plateau time Tplat
1
RR
Insp. Flow
t
t
Pmax
PEEP
Exp. Flow
Operation
VC-CMV -Volume Controlled - Controlled Mandatory Ventilation
Volume-controlled ventilation with fixed mandatory minute volume MV, which is set with tidal volume VT and respiratory rate RR.
WARNING Only use VC-CMV for patients who are not
spontaneously breathing. Otherwise, the patient may be put at risk by not receiving sufficient ventilation.
Use VC-AC for patients with partial spontaneous breathing.
Set the ventilation pattern with the controls below the display:
– Tidal volume VT.
– Ventilation respiratory rate RR.
(minimum possible frequency: 5 per min).
– Maximum airway pressure Pmax.
–O
2 concentration FiO2.
The following can be set on the display:
– Positive end expiratory pressure PEEP.
– Sensitivity Trigger.
– Ventilation time ratio I:E or inspiratory time Ti.
When setting the ventilation respiratory rate RR, tidal volume VT or I:E / Ti, the associated values for inspiration flow and Ti / I:E are automatically
025
displayed in the information window.
– Plateau time Tplat %, in % of the inspiratory
time.
AutoFlow (optional).
1043717110537171
Instructions for Use Oxylog 3000 plus SW 1.n 65
Operation
– Hose type
The selected hose type must match the hose type in use. Otherwise a correct volume measurement cannot be guaranteed.

Trigger (VC-AC)

NOTE
When in VC-CMV and the trigger is set to a value, the ventilation mode changes to VC-AC.
VC-AC Volume Controlled - Assist Control
For synchronization with the patient's spontaneous inspiratory efforts.
1373717114537171
The mandatory breaths are synchronized with the patient's spontaneous inspiratory efforts when the trigger is activated and the trigger sensitivity is set.
In this case, the actual respiratory rate may be higher than the set ventilation respiratory rate RR.
Successful patient triggering is briefly indicated by an asterisk (*) on the left side of the curve window.
– Cuvette type (optional)
NOTE
The cuvette windows of the reusable cuvette and disposable cuvette have different optical proper­ties. Therefore, the correct cuvette type must be selected in the Setting menu. Otherwise the zero point is shifted by up to ± 8 mmHg of CO
2.
66
Instructions for Use Oxylog 3000 plus SW 1.n
Operation
Oxylog 3000 plus
A
Activating/setting the trigger
1 Press the key Settings (A) until the trig-
ger parameter is displayed.
2 Select the line Trigger on the display and then
set and confirm the value with the rotary knob. Small value = high sensitivity.
The ventilation mode AC
is shown on the display.
Deactivating trigger
1 Set a value less than 1 L/min or greater than
15 L/min (off is displayed instead of a value).
2 Press the rotary knob to confirm.
The last effective trigger value is adopted by the ventilator when changing from VC-AC to PC-BIPAP or SpnCPAP.

Setting AutoFlow (optional)

The following can also be set on the display for VC-CMV and VC-AC:
– The AutoFlow function AutoFlow-AF.
00537170
When AutoFlow is switched on, the setting Tplat% is no longer valid, and Slope must be set.
For more information on AutoFlow, refer to "AutoFlow" on page 162.

Cardio-pulmonary resuscitation (CPR)

During CPR, the airway pressure Paw is increased because of chest compressions.
The Oxylog 3000 plus will try to limit the airway pressure Paw to the set Pmax, without ending the inspiration prematurely.
However, if due to compressions the airway pressure Paw exceeds the set Pmax by 5 mbar, the Oxylog 3000 plus cycles to the expiratory phase.
Therefore in general, if Pmax is set to a higher value, a higher minute volume is possible. However, this increases the intra-thoracic pressure and may reduce coronary perfusion.
10737171-1
Instructions for Use Oxylog 3000 plus SW 1.n 67
Operation

VC-SIMV, VC-SIMV/PS

Volume Controlled - Synchronized Intermittent Mandatory Ventilation
Paw
PEEP
P
max
Pressure
support PS
short
rise time
long
rise time
t
i
T
Flow
1
RR
Insp. Flow
Trigger window
t
For patients with inadequate spontaneous breath­ing, or for patients who are to be weaned gradually.
Fixed mandatory minute volume MV is set with tidal volume VT and ventilation respiratory rate RR. The patient can breathe spontaneously between the mandatory breaths and thus contribute to the total minute volume. Spontaneous breathing can be assisted with PS.
Set the ventilation pattern with the controls below the display:
– Tidal volume VT.
– Respiratory rate RR.
(minimum possible respiratory rate: 2 per min).
– Maximum airway pressure Pmax.
–O
2 concentration FiO2.
The following can be set on the display:
– Positive end expiratory pressure PEEP.
– Pressure Support
Δ
Psupp above PEEP.
– Sensitivity Trigger.
Successful patient triggering is indicated by an asterisk (*) on the left side of the curve window.
– Ventilation time ratio I:E or inspiratory time Ti.
027
When setting the ventilation respiratory rate RR, tidal volume VT or I:E / Ti, the associated values for inspiration flow and Ti / I:E are automatically dis­played in the information window.
– Plateau time Tplat %, in % of the inspiratory
time.
AutoFlow (optional).
10737171-115837171-1
68
Instructions for Use Oxylog 3000 plus SW 1.n
Insp.term.%PIF
Inspiration termination criteria of pressure sup­ported strokes, as percentage of the peak inspiratory flow (PIF).
Hose type
The selected hose type must match the hose type in use. Otherwise a correct volume mea­surement cannot be guaranteed.
Operation

Setting Pressure Support VC-SIMV/PS

The following can also be set on the display for VC-SIMV:
1443717114337171
– Setting on page 1: Pressure Support
above PEEP.
– Setting on page 2: When
Δ
Psupp is set above 0 mbar, the pressure rise time Slope can be set.
Flat slope = long pressure rise time Medium slope = medium pressure rise time Steep slope = short pressure rise time.
Δ
Psupp
14237171-1
Cuvette type (optional)
NOTE
The cuvette windows of the reusable cuvette and disposable cuvette have different optical proper­ties. Therefore, the correct cuvette type must be selected in the Setting menu. Otherwise the zero point is shifted by up to ± 8 mmHg of CO
Instructions for Use Oxylog 3000 plus SW 1.n 69
2.
Operation

Setting AutoFlow (optional)

The following can also be set on the display for VC-SIMV and VC-SIMV/PS:
– The AutoFlow function AutoFlow-AF.
When AutoFlow is switched on, the setting Tplat% is no longer valid, and Slope must be set.
For more information on AutoFlow, refer to "AutoFlow" on page 162.
10737171-1
70
Instructions for Use Oxylog 3000 plus SW 1.n

PC-BIPAP, PC-BIPAP/PS

Operation
Pressure Controlled - Biphasic Positive Airway Pressure
Without spontaneous breathing
Short
Paw
rise time
PEEP
T
i
Flow
1
RR
With spontaneous breathing
Long rise time
insp
P
Trigger window
Pressure support PS
Short rise time
Long rise time
t
t
Pressure-controlled ventilation combined with spontaneous breathing throughout the respiratory cycle and variable Pressure Support at CPAP level.
For patients without spontaneous breathing, to spontaneously breathing patients shortly before extubation. The patient is weaned by gradually reducing the mandatory portion of the total minute volume MV and by reducing the Pressure Support ΔPsupp.
The following can be set on the display:
– Positive end expiratory pressure PEEP.
– Inspiratory pressure Pinsp.
– Pressure Support
Δ
Psupp above PEEP.
– Sensitivity Trigger.
Successful patient triggering is indicated by an asterisk (*) on the left side of the curve window.
028
1083717114137171
– Ventilation time ratio I:E or inspiratory time Ti.
The mandatory portion of the total minute volume MV is set via the inspiratory pressure Pinsp, PEEP and ventilation respiratory rate RR.
Refer to the description on page 161 for details.
Set the ventilation pattern with the controls below
– Pressure rise time Slope (effective for the PC-
BIPAP stroke and Pressure Support ΔPsupp).
NIV - Non-invasive ventilation.
Refer to "NIV – Non-invasive ventilation (Mask ventilation)" on page 76.
the display:
– Respiratory rate RR.
– Maximum airway pressure Pmax.
–O
2 concentration FiO2.
Instructions for Use Oxylog 3000 plus SW 1.n 71
Operation
Insp.term.%PIF
Inspiration termination criteria of pressure sup­ported strokes, as percentage of the peak inspiratory flow (PIF).
Hose type
The selected hose type must match the hose type in use. Otherwise a correct volume mea­surement cannot be guaranteed.

Setting Pressure Support PC-BIPAP/PS

The following can also be set on the display for PC-BIPAP:
1403717115037171
Δ
– Setting on page 1: Pressure Support
above PEEP.
– Setting on page 2: Pressure rise time Slope.
Flat slope = long pressure rise time Medium slope = medium pressure rise time Steep slope = short pressure rise time.
Psupp
12437171
Cuvette type (optional)
NOTE
The cuvette windows of the reusable cuvette and disposable cuvette have different optical proper­ties. Therefore, the correct cuvette type must be selected in the Setting menu. Otherwise the zero point is shifted by up to ± 8 mmHg of CO
2.
72
Instructions for Use Oxylog 3000 plus SW 1.n

SpnCPAP, Sp n C PA P / P S

Spontaneous Continuous Positive Airway Pressure
WARNING Only use SpnCPAP (/PS) for patients with
sufficient spontaneous breathing.
Otherwise there is a risk of the patient receiving insufficient ventilation.
Operation
Set the ventilation pattern with the controls below the display:
– Maximum airway pressure Pmax.
–O2 concentration FiO2.
The following can be set on the display:
– Positive end expiratory pressure PEEP.
– Pressure Support
Δ
Psupp above PEEP.
NIV - Non-invasive ventilation.
Refer to "NIV – Non-invasive ventilation (Mask ventilation)" on page 76.
Insp.term.%PIF
Inspiration termination criteria of pressure sup­ported strokes, as percentage of the peak inspiratory flow (PIF).
Hose type
The selected hose type must match the hose type in use. Otherwise a correct volume mea­surement cannot be guaranteed.
15737171
Cuvette type (optional)
NOTE
The cuvette windows of the reusable cuvette and disposable cuvette have different optical properties. Therefore, the correct cuvette type must be selected in the Setting menu. Otherwise the zero point is shifted by up to ± 8 mmHg of CO
1513717115237171
2.
Instructions for Use Oxylog 3000 plus SW 1.n 73
Operation
Paw
Flow
Apnea alarm time
Tapn
Start
apnea ventilation with apnea alarm
t
t
1
RRapn
Pmax
Oxylog 3000 plus
A B

Apnoea ventilation

Apnoea back-up ventilation is only applicable when using the SpnCPAP mode. In the event of an apnoea, the ventilator will automatically activate volume-controlled mandatory ventilation - (VC­CMV).
When an apnoea occurs, the device simultaneously issues an alarm signal and switches to volume controlled ventilation with the parameters respiratory rate RRapn, tidal volume VTapn, and the maximum airway pressure Pmax when the apnoea time Tapn has been reached. The ventilation time ratio I:E = 1:1.5 and the plateau time Tplat % = 0 are preset during apnoea ventilation.
Setting apnoea ventilation
On the display:
z Set Tapn with the rotary knob (B) to a value
between 15 and 60 seconds.
029US
The parameters RRapn and VTapn, which are required for setting apnoea ventilation, are now displayed:
3 Set RRapn and VTapn.
4 Set Pmax. This determines the maximum
airway pressure allowed during apnoea ventilation.
06837170screen 13
74
Instructions for Use Oxylog 3000 plus SW 1.n
Operation
To end apnoea ventilation
z Press the Alarm Reset key.
The ventilator resumes ventilating with the original mode and parameter settings.
To deactivate apnoea ventilation
z Set Tapn to OFF.
NOTE
Apnoea ventilation can only be activated in the ventilation mode SpnCPAP without NIV.
The minimum ventilation required by the patient must be monitored via the lower alarm limit MVe .

Setting Pressure Support Spn-CPAP/PS

If ΔPsupp is set above 0 mbar, the following can also be set on the display for SpnCPAP:

Cardio-pulmonary resuscitation (CPR)

Refer to "Cardio-pulmonary resuscitation (CPR)" on page 67.
12537171
– Sensitivity Trigger.
Successful patient triggering is indicated by an asterisk (*) on the left side of the curve window.
– Pressure rise time Slope (effective for Pressure
Δ
Support
Instructions for Use Oxylog 3000 plus SW 1.n 75
Psupp).
Operation
NIV – Non-invasive ventilation (Mask ventilation)
Use of NIV
NIV can only be activated as a supplementary function in the ventilation modes SpnCPAP (/PS), PC-BIPAP (/PS), VC-CMV / AF, VC-AC / AF and VC-SIMV / AF. The Oxylog 3000 plus automatically adjusts to the requirements of mask ventilation. Mask leakages are detected by the device and compensated for. Therefore, the displayed measured values VTe and MVe do not include the leakage. The leakage alarm is inactive.
WARNING If NIV is not activated, measured values for VTe
and MVe will be inconsistent if there are leakages during ventilation.
WARNING Ensure that NIV is not activated for intubated
patients.
Risk of undetected leakages and inadequate ventilation.
WARNING Bacterial filters, HME and masks increase the
resistance and dead space volume of the ventilation equipment. Note the manufacturer’s directions.
Risk of CO2 rebreathing.

To switch on NIV

1 Activate NIV off.
2 Select NIV on and confirm.
– The supplement NIV appears in the ventilation
mode window.
12837171
WARNI NG Set the lower alarm limit MVe according to
the minimum ventilation required for the patient.
Otherwise, there is a risk of the patient receiving insufficient ventilation.
NOTE
Apnoea ventilation is not possible when NIV is active.
WARNING Check MVe alarm limits after deactivating NIV
mode.
WARNING Avoid high airway pressure.
Risk of aspiration.
76
Instructions for Use Oxylog 3000 plus SW 1.n

Special functions

Oxylog 3000 plus
A
B
Operation

Manual inspiration / Inspiration hold

The function Manual inspiration / Inspiration hold will either initiate a new (manual) breath or hold the inspiratory phase of the current breath for a maxi­mum of 15 seconds.
The pattern of the manually started breath corre­sponds with the set ventilation mode.
This function is not available for:
– SpnCPAP without PS,
–O
2 inhalation (optional).
To activate Manual inspiration or Inspiration hold

100 % O2 (optional)

To apply 100 % O2 for 3 minutes regardless of the momentarily set value.
z Briefly press key 100 % O
2 (B).
Its indicator lights up for 3 minutes.
The set value is resumed by the ventilator upon expiry of these 3 minutes, or when the 100 % O
2 is
pressed again. The indicator dims.

O2 inhalation (optional)

WARNING The O2 inhalation function is not a ventilation
mode.
It may only be used for patients with spontane­ous breathing who receive a constant O2 flow of between 0 and 15 L/min via a mask.
If a tracheal stenosis or other obstruction occurs, the flow is interrupted by the ventilator for 5 seconds at an airway pressure of 30 mbar and the airway pressure is reduced to 0 mbar. The !!! Paw high alarm is active.
NOTE
The options 100 % O2 and O2 inhalation are mutually exclusive.
01237170
z Press key Inspiration hold (A) for as long as
inspiration is required.
Instructions for Use Oxylog 3000 plus SW 1.n 77
Operation
To activate O2 inhalation
A B
Oxylog 3000 plus
C
1 Connect the inhalation mask to the gas outlet
for breathing hose (B).
2 Press and hold key O
approx. 3 seconds.
O
2 inhalation is performed with the previously
effective setting.
3 Set and confirm the required O2 flow via the
rotary knob (C).
Display (example):
2 inhalation (A) for
0833717007137170
78
Instructions for Use Oxylog 3000 plus SW 1.n

O2 concentration with "O2 blending"

Flow (L/min.)
O
2
= f(Flow, Paw)
40
20
0
15
30
85 (%)
100
O
2
(%)
Flow (L/min.)
O2 = f(Flow, Paw)
40
20
0
20
93 (%)
100
O
2
(%)
The FiO2 can be set between 40 % and 100 % O2, regardless of the ventilation mode. An inspiratory
2 concentration below 100 % is produced by
O drawing in ambient air, with the injector principle realized in the Oxylog 3000 plus.
However, the O2 concentration which can be realized depends on the mean airway pressure and the inspiratory flow. The O never be lower than 40 %.
This is shown in the following graphics:
(%)
O
2
40
20
0
3
O
2 concentration which can be realized at a Pmean
40
of 5 mbar.
(%)
2
O
40
20
0
10
2 concentration which can be realized at a Pmean
O
35
of 15 mbar.
2 concentration can
O2 = f(Flow, Paw)
O
2
= f(Flow, Paw)
80 (%)
100
82 (%)
100
Flow (L/min.)
Flow (L/min.)
Operation
O
2 concentration which can be realized at a Pmean
of 30 mbar.
031032
2 concentration which can be realized at a Pmean
O of 60 mbar.
The O
2 concentration is a calculated value. It is not
measured by an internal O2 sensor.
If the Oxylog 3000 plus cannot achieve the set O concentrations, the signal "Check settings FiO prompts the user to correct the setting.
z Correct setting by control FiO
2
When the O2 concentration has been set, the value will be displayed after approximately 30 seconds.
033034
2
2"
Instructions for Use Oxylog 3000 plus SW 1.n 79
Operation
When patients are breathing spontaneously, the achievable O2 concentration will depend on the profile of the inspiratory flow.
WARNING In toxic surroundings:
– The patient must be ventilated with 100 %
medical grade oxygen so that toxic constit­uents do not enter into the breathing gas.
– The patient must be immediately trans-
ferred to a breathable atmosphere in order to prevent inhalation of toxic air when spontaneous breathing resumes.

Setting HME correction

The temperature and moisture influence of HME have an effect on the flow measurement. The Oxylog 3000 plus can compensate for the presence of an HME.
When using HME, select, set and confirm HME - On in the Settings window with the rotary knob.
WARNI NG In infectious environments:
– The patient must be ventilated with 100 %
medical grade oxygen so that bacteria, viruses, fungi or spores do not enter the breathing gas.
– The patient must be immediately trans-
ferred to a breathable atmosphere in order to prevent inhalation of infectious air when spontaneous breathing resumes.
When HME - ON is selected, the flow sensor expects an expiration gas temperature of 35 °C and a relative humidity of 0 %.
When HME - Off is selected, the flow sensor expects an expiration gas temperature of 37 °C and a relative humidity of 100 %.
80
13837171
Instructions for Use Oxylog 3000 plus SW 1.n

Calibration

Operation
The pressure and flow sensors are automatically calibrated by the device at regular intervals without interrupting ventilation.
The saved calibration values are retained even when the device is switched OFF.

Screen brightness

The screen brightness level can be set on the last page of the Settings menu, from level 1/4 to 4/4.
During battery operation, when no controls have been set for a period longer than one minute, the screen will automatically be dimmed (power save mode).

Alarm volume

The alarm volume level can be set on the last page of the Alarms menu, from level 1/4 to 4/4.
The CO2 sensor must be calibrated if the test values are not adhered to during the test gas check.
For calibrating CO Service Mode" on page 98.
The screen brightness level in power save mode can be adjusted in Customer Service Mode. Refer to "Customer Service Mode" on page 98.
2 sensor, refer to "Customer
Instructions for Use Oxylog 3000 plus SW 1.n 81
Operation
Oxylog 3000 plus
AB
A
B

Shutdown

– After disconnecting the patient
Switch the device OFF:
1 To switch the device OFF, hold down the
key (A) for approximately 3 seconds.
Ventilation is now stopped and a high-priority alarm is issued. This alarm can be silenced with the key.
2 Either:
– Press the rotary knob (B) to confirm switch
off.
Or
– Press the key (A) to resume ventilation
with the previous settings.
NOTE
When the device is switched OFF, the battery is still being charged if the device is connected to an external power source.
When O
2 is supplied from a cylinder:
3 Close the cylinder valve.
WARNI NG The cylinder valve must be closed completely
to avoid gas flow leakage by the device.
When O
2 is supplied from the pipeline system:
4 Disconnect the high-pressure connection from
the source.
5 Disconnect the breathing hose.
00337170
CAUTION
When disconnecting the breathing hose, always grip the sleeve (A) and not the corrugations (B).
If this is not done, the corrugations or hose may be torn from the sleeve.
048
82
Instructions for Use Oxylog 3000 plus SW 1.n

Alarms

Types of alarms . . . . . . . . . . . . . . . . . . . . . . 84
In the event of an alarm . . . . . . . . . . . . . . . . 85
Setting alarm limits . . . . . . . . . . . . . . . . . . . 87
Alarms
Instructions for Use Oxylog 3000 plus SW 1.n 83
Alarms
Oxylog 3000 plus
C
B
A
!!! Apnoea ventilation

Types of alarms

The actual screen display may differ in appearance or configuration from these illustrations.
The Oxylog 3000 plus assigns a priority to the alarm message. This message highlights the text with the appropriate number of exclamation marks and generates different tone sequences for the respective alarms. !!! = Warning !! = Caution ! = Advisory
Refer to the list "Alarm - Cause - Remedy" on page 112 for information on how to remedy the alarms.

Caution

An alarm of medium priority. z The alarm indicator (A) flashes yellow.
Caution messages are preceded by two exclamation marks.
Example: !! No int. battery ?
z The Oxylog 3000 plus generates a three-tone
sequence, which is repeated approximately every 20 seconds.

Advisory

An alarm of low priority. z The yellow alarm indicator (A) lights up.
Advisory messages are preceded by one exclamation mark.
Example:
! Settings not confirmed
z The Oxylog 3000 plus generates a 2-tone
alarm sequence, which sounds only once.

Warning

An alarm with high priority z The alarm indicator (B) flashes red.
Warnings are preceded by three exclamation marks and displayed in inverted form (C).
Example:
z The Oxylog 3000 plus generates a sequence of
five tones, which sound twice and are repeated approximately every 7 seconds.
84
036
Instructions for Use Oxylog 3000 plus SW 1.n

In the event of an alarm

Oxylog 3000 plus
C B
A
Oxylog 3000 plus
A
z The indicator (A) flashes red or yellow, or lights
up yellow. And
z The alarm message appears on the upper right
corner of the screen (C). In addition, alarm tones are issued.
When the fault has been remedied the alarm tone is canceled. Alarms, which have been remedied and remain on the display, can be acknowledged (reset):
z Press the Alarm Reset key (B).
The alarm message is now removed from the
display. Every alarm which has been remedied, but not acknowledged, will be overwritten when a new alarm is issued.
Alarms
Suppress alarm tones
WARNING Check the display regularly for alarm
messages when the alarm tones are silenced.
Otherwise, alarms can be missed.
037
038
1 Press the key (A).
The alarm indicator remains active and all cur­rent alarm tones are suppressed for approxi­mately 2 minutes. Alarms with a higher priority override the alarm silence and sound with a tone sequence. All alarm tones are resumed by the device after these 2 minutes.
WARNING Set the alarm volume to a suitable level for the
environment.
Pay additional attention in environments where the surrounding noise interferes with hearing the maximum alarm tone of the device
WARNING
(e.g. in a helicopter).
NOTE
To be notified of new acoustic alarm signals, the 2 minutes alarm silence must be reset.
Instructions for Use Oxylog 3000 plus SW 1.n 85
Alarms
NOTE
The volume of alarm tones can be adjusted. Refer to section "Alarm volume" on page 81.
If alarm tones are to be heard again before the 2 minutes have expired:
2 Press the key (A) again.

In the event of a gas failure

WARNING In the event of a gas failure, proper ventilation
cannot be guaranteed. The Oxylog 3000 plus issues the alarm !!! Supply pressure low.
Disconnect the patient from the device and continue ventilation without delay using another ventilator.

In the event of an internal power supply failure

WARNING In the event of an internal power supply fail-
ure, automatic ventilation, volume measure­ment and alarms do not operate.
An acoustic alarm signal is given to indicate the internal power supply failure.
Spontaneous breathing can continue through the emergency air intake.
Disconnect the patient from the device and continue ventilation without delay using another ventilator.
86
Instructions for Use Oxylog 3000 plus SW 1.n

Setting alarm limits

Oxylog 3000 plus
A B
Alarms
CAUTION
Set alarm values carefully.
Extreme alarm values can render the alarm system useless.

Setting upper alarm limit for Paw

Regardless of the set ventilation mode, the airway pressure Paw is controlled by the ventilator and lim­ited to the set maximum inspiratory pressure Pmax. The airway pressure is limited when Pmax is reached; inspiration will not be terminated prema­turely. For more details refer to "Cardio-pulmonary resuscitation (CPR)" on page 67.
Pmax appears in the pressure curve as a dashed line. When this dashed line is reached, the Oxylog 3000 plus issues a !!! Paw high alarm.
inspiratory and expiratory pressure; for more than 20 seconds.
Setting alarm limits for MVe, RRsp and optional
2
etCO
WARNING Set the lower alarm limit MVe according to
the minimum ventilation required for the patient.
Otherwise, there is a risk of the patient receiv­ing insufficient ventilation.
1 Press the key Alarms (B).
Display (example):
2 Select and activate the low alarm limit or
high alarm limit for MVe, RRsp or etCO
2 on
the display.
3 Set and confirm the value.
If the CO2 sensor cable is disconnected, etCO2
039
1 Set the maximum airway pressure Pmax via the
Pmax control (A).
alarm limits are not visible. If the CO
2 sensor cable is disconnected and then
reconnected, the previously set alarm limits will still be valid.

Lower alarm limit for Paw

The Oxylog 3000 plus
automatically generates an
For the alarm limit ranges refer to "Alarm limit ran­ges" on page 141.
alarm when it no longer detects a pressure difference of more than 5 mbar between the
Instructions for Use Oxylog 3000 plus SW 1.n 87
11037171-2
Alarms

Setting alarm limits automatically

WARNING After using the function Alarms: Autoset,
check if the new alarm limits are appropriate for the patient.
Risk of hypoventilation.
The function Alarms: Autoset sets the alarm limits on the basis of the actual measured values at the time of activation. This automatic setting of alarm limits is performed only once, when confirmed with the rotary knob.
1 Press the key Alarms (B).
2 Select and activate the line Alarms: Autoset
on the display.
3 Press the rotary knob to confirm
Alarms: Autoset, or press the Alarm Reset key to leave the settings unchanged.
The auto alarm limits are based on the actual mea­sured values as follows:
Alarm Setting
MVe Current value –20 %, with a
minimum of 0.5 L/min. Below
0.5 L/min. the limit remains unchanged.
MVe Current value +30 % or
+2 L\min, whichever is smaller
RRsp Current value +5/min, with a
minimum of 10/min.
2 Based on the current value
etCO
The etCO
2 / auto alarm limits are based
on the actual etCO2 value as follows:
Lower alarm limit [mmHg]
Current measured value
Upper alarm limit [mmHg]
[mmHg]
Unchanged <15 Unchanged
Current –5 15 to 35 Current +15
Current –7 35 to 45 Current +10
Current –10 >45 Current +5
88
Lower alarm limit [kPA] or [Vol.%]
Current measured value [kPA] or
Upper alarm limit [kPA] or [Vol.%]
[Vol.%]
Unchanged <2.0 Unchanged
Current –0.7 2.0 to 4.7 Current +2.0
Current –0.9 4.7 to 6.0 Current +1.3
Current –1.3 >6.0 Current +0.7
Instructions for Use Oxylog 3000 plus SW 1.n

Monitoring

Displaying curves. . . . . . . . . . . . . . . . . . . . . 90
Displaying measured values . . . . . . . . . . . . 90
CO
2 measurement (optional). . . . . . . . . . . . 91
Cuvette type setting . . . . . . . . . . . . . . . . . . . . 91
Checking the CO2 sensor during ventilation . 91
Zero calibration during ventilation. . . . . . . . . . 92
2 filter check during ventilation . . . . . . . . . 93
CO
Monitoring
Instructions for Use Oxylog 3000 plus SW 1.n 89
Monitoring

Displaying curves

The curve window can display the airway pressure curve Paw, the flow curve or the CO
2 curve
(optional). Refer to "Screen window structure" on page 32.
To display a different curve:
z Press the Curves key.
Curves

Displaying measured values

Measured values are displayed in the measured values window.
To switch between the values:
z Press the Values key: the next value pair
is displayed on the screen.
The following values can be displayed:
–MVe
,
–FiO2,
–RR,
–VTe,
– PEEP,
–Pmean,
–PIP
,
–Pplat,
–MVespon
,
–RRsp,
–etCO
2.
When the CO2 sensor is connected to the ventilator, the etCO2 value will be shown automatically in the measured values window.
The values are shown in pairs; the pairs of values can be configured as required. Refer to "Customer Service Mode" on page 98.
90
Instructions for Use Oxylog 3000 plus SW 1.n

CO2 measurement (optional)

Monitoring
The CO2 measurement only works if the CO2 option has been installed and if the CO2 sensor is present.
1 Connect the CO
to "Connecting the CO
2 sensor and the cuvette (refer
2 sensor and the cuvette"
on page 48).
2 Set the cuvette type in the Settings menu (refer
to Cuvette type setting).
The following will be activated:
– Curve window: CO
2 Curve
– Measured values window: the parameter etCO
will automatically be displayed
– Alarm window: etCO2 high and etCO2 low alarm
For more information on curves and measured values: refer to "CO
2 curve large view" on page 34.
For more information on configuring measured value pairs: refer to "Customer Service Mode" on page 98.
For connecting the CO
2 sensor and cuvette see
page 48. For CO2 zero calibration and filter check before ventilation, see page 58. For CO2 configura­tion in the Customer Service Mode see page 105.

Cuvette type setting

NOTE
The cuvette windows of the reusable cuvette and disposable cuvette have different optical proper­ties. Therefore, the correct cuvette type must be selected in the Setting menu. Otherwise the zero point is shifted by up to ± 8 mmHg of CO
To set the cuvette type (reusable or disposable):
1 Press the Settings key.
2 Select and activate the line Cuvette type.
3 Set the cuvette type and confirm.
2
NOTE
If a wrong cuvette type is selected, the Oxylog 3000 plus displays the alarm !!! Check cuvette type.

Checking the CO2 sensor during ventilation

The following checks are necessary for the CO2 sensor during ventilation:
Check Interval
2 zero cali-
CO bration
2 filter
CO check
Recommended before measure­ment and when changing the
2 sensor to another unit.
CO
Required in intervals of one month.
2.
Instructions for Use Oxylog 3000 plus SW 1.n 91
Monitoring
A
B

Zero calibration during ventilation

The zero calibration is performed with a clean CO2 sensor that has been removed from the cuvette!
NOTE
Do not breathe on the CO2 sensor during zero calibration, otherwise the zero calibration can fail or the zero calibration can pass with an invalid zero value.
To perform zero calibration:
1 Connect the CO2 sensor and wait at least
3 minutes for the CO2 sensor to complete its warm-up phase.
2 Remove the CO
cuvette (B).
3 Press the Settings key.
4 Select and activate the line CO
Run. The screen displays the text Remove the sensor from cuvette then press rotary knob.
5 Confirm. The zero calibration starts and the line
displays Busy. Note the possible warm-up time. During zero calibration, ventilation settings can be changed. After a successful zero calibration, the line briefly displays Pass.
6 Attach the CO
cuvette (B).
2 sensor (A) from the
2 Zero Calib -
2 sensor (A) back to the
If zero calibration was not successful:
The Oxylog 3000 plus displays the alarm !!! CO Zero calib. failed.
z Repeat zero calibration.
If zero calibration is still not possible:
1 Check whether the sensor (A) is soiled and
clean it if necessary. If the sensor is defective, replace the sensor.
2 Repeat zero calibration.
179
2
92
Instructions for Use Oxylog 3000 plus SW 1.n
Monitoring
A
B

CO2 filter check during ventilation

NOTE
Before the CO2 filter check, you need to have fin­ished a successful CO wise the CO
2 filter check may be outside of the
tolerance range.
To perform CO2 filter check:
1 Remove the CO
2 Attach the CO
3 Press the Settings key.
4 Select and activate the line CO
Run.
5 Confirm. The filter check starts and the line
displays Busy.
During the filter check, ventilation settings can
be changed.
After a successful filter check, the line briefly
displays Pass.
6 Attach the CO
2 zero calibration. Other-
2 sensor from the cuvette.
2 sensor (A) to the test filter (B).
2 Filter check -
2 sensor (A) back to the cuvette.
For connecting the CO
2 sensor and cuvette see
page 48. For CO2 zero calibration and filter check before ventilation, see page 58. For CO
2 configura-
tion in the Customer Service Mode see page 105.
177
If the check was not successful:
The Oxylog 3000 plus displays the alarm !!! CO2 Filter check failed. The test value is outside the permissible tolerance.
z Check whether the sensor (A) or test filter (B) is
soiled and clean them if necessary. Repeat the
CO2 filter check.
If the check was still not successful:
z Check the CO
2 calibration with test gas.
Instructions for Use Oxylog 3000 plus SW 1.n 93
This page intentionally left blank

Configuration

Setting configuration parameters / display
information . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Displaying configuration and information .97
Setting language. . . . . . . . . . . . . . . . . . . . . . . 97
Displaying the battery type . . . . . . . . . . . . . . . 97
Customer Service Mode. . . . . . . . . . . . . . . . 98
To enter Customer Service Mode . . . . . . . . . . 99
Settings in Customer Service Mode . . . . . . . . 99
To exit the parameter settings menu . . . . . . . 99
Set start-up settings . . . . . . . . . . . . . . . . . . . . 100
Hose/cuvette start-up settings . . . . . . . . . . . . 101
Set date and time . . . . . . . . . . . . . . . . . . . . . . 102
Set the measured values display window. . . . 102
Enter activation code . . . . . . . . . . . . . . . . . . . 102
Test buttons and potentiometer . . . . . . . . . . . 102
Test loudspeaker, buzzer, LEDs and display . 103
Display battery and supply data . . . . . . . . . . . 103
Check safety valve . . . . . . . . . . . . . . . . . . . . . 104
Display info logbook entries . . . . . . . . . . . . . . 104
Display user logbook entries . . . . . . . . . . . . . 104
Display maintenance and service contact
information . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
CO2 sensor . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Exit Customer Service Mode . . . . . . . . . . . . . 109
Configuration
Customer service manual . . . . . . . . . . . . . . 110
Instructions for Use Oxylog 3000 plus SW 1.n 95
Configuration

Setting configuration parameters / display information

The actual screen display may differ in appearance or configuration from these illustrations.
1 To switch the device ON briefly press the
key. The device performs a selftest and the operator is prompted, on the display, to activate the configuration menu or device check:
Press rotary knob for device check and configuration.
2 Press the rotary knob to confirm, before the
black progress bar is complete. The start-up menu is then displayed:
3 Select Configuration and information in the
start-up menu and confirm.
96
15337171
Instructions for Use Oxylog 3000 plus SW 1.n

Displaying configuration and information

Configuration
– The settings made via the "Configuration" are
retained after the ventilator is switched OFF.
– Configuration can be canceled by pressing the
Alarm reset key.
The following data can be displayed via Configuration and information:
– Language
– Identification No. (Device ID)
– Total hours of operation (Working hours)
– Hours of operation since the last inspection and
maintenance (Hours since service time)
– Battery type and battery capacity.

Setting language

1 Press the key Settings to select the
menu Configuration.
2 Select and activate the line Language.
3 Select the language and confirm.
The new language selected is effective immediately.
06637171screen 17

Displaying the battery type

z Press the key Settings to select the
menu Information. The performance data of the inserted battery are displayed on the device.
Instructions for Use Oxylog 3000 plus SW 1.n 97
Configuration

Customer Service Mode

NOTE
Ventilation is not possible when the device is in Customer Service Mode.
In Customer Service Mode, the ventilator performs function tests, outputs status information and permits configuration of parameter settings. Displays in Customer Service Mode appear in English and cannot be changed to any other language.
001 Set start-up settings. Configure start-up settings, restore manufacturer's
default settings.
002 Hose/cuvette start-up settings. Determine which hose system / cuvette type is used.
003 Set date and time. Set date and time.
004 Set measured values display window. Configure the layout of measured values in the measured
values window or restore manufacturer's default settings.
005 Enter activation code. Enter the activation code for options.
006 Test buttons and potentiometer. Check for correct functioning of keys and controls.
007 Test loudspeaker, buzzer, LEDs and
display.
008 Display battery and supply data. Display battery data and condition of the supply voltage.
009 Check safety valve. Checking of the safety valve.
010 Display info logbook. Calibration logbook and technical errors in chronological
011 Display user logbook. Logbook of operating phases, ventilator settings and
012 Display maintenance and service
contact information.
013 CO
2 sensor. Check or calibrate the CO2 sensor.
Check for correct functioning of loudspeaker, buzzer, LEDs and display.
sequence.
alarms.
Display the maintenance schedule and contact information for service personnel.
98
Instructions for Use Oxylog 3000 plus SW 1.n
Configuration
Oxylog 3000 plus
A B C
DE

To enter Customer Service Mode

1 Ensure that ventilation is OFF.
2 Turn controls VT (A) and RR (B) all the way to
the right. 3 Switch on the device by briefly pressing the
Start/Standby key (C) and simultaneously
press and hold the Curves key (D) and
the Values key (E) until the main
Customer Service Mode menu appears. 4 In the main menu, set the number of the
required test with the rotary knob.
Curves

Settings in Customer Service Mode

6 Select the required function with the cursor
(asterisk).
– To select a parameter: turn rotary knob.
– To activate the parameter: press rotary
knob.
– To set a value: turn rotary knob.
– To confirm the value: press rotary knob.

To exit the parameter settings menu

07737170screen 18
11837170
1 Select the line EXIT.
2 Press the rotary knob to confirm.
The set values are saved as default settings and remain effective.
5 Press the rotary knob to activate the test.
Instructions for Use Oxylog 3000 plus SW 1.n 99
Configuration

Set start-up settings

The ranges of the settings are:
Parameter Range Ventilation mode Trigger 0 (OFF) 1 to 15 L/min PEEP 0 to 20 mbar I:E or Ti Configurable I:E or Ti I:E 1:100 to 50:1 Ti 0.2 to 10.0 s Tplat % 0 to 50 % ΔPsupp 0 to 35 mbar Slope FLAT, MEDIUM, STEEP Pinsp 3 to 55 mbar O
2-Flow 0 to 15 L/min
NIV ON, OFF Tapn 0 (OFF) to 60 s VTapn 50 to 2000 mL RRapn 12 to 60 bpm MVe-high alarm 2.0 to 41 L/min MVe-low alarm 0.5 to 40 L/min RR-high 10 to 100 bpm HME correction ON, OFF AutoFlow ON, OFF Brightness-min* 1/4 to 4/4 Brightness 1/4 to 4/4 Alarm volume 1/4 to 4/4 etCO
2 high alarm 1 to 100 mmHg
etCO
2 low alarm 0 to 100 mmHg
CO
2 unit mmHg, kPa or Vol.%
Hose type Adult disposable
Adult reusable Paediatric disposable
Cuvette type Disposable
Reusable
* Brightness-min: screen brightness level in power save mode. Refer to "Screen brightness" on page 81.
The manufacturer's default settings are:
Parameter Manufacturer's default setting Ventilation mode VC-CMV Trigger 0 L/min at VC-CMV, VC-AC
and 3 L/min at VC-SIMV,
SpnCPAP, PC-BIPAP PEEP 5 mbar I:E or Ti I:E I:E 1.0:1.5 Ti 2.0 s Tplat % 0 % ΔPsupp 0 mbar Slope (pressure
Standard rise time)
Pinsp 20 mbar O
2-Flow 10 L/min
NIV OFF Tapn 0 s (when switched on
minimum of 15 s) VTapn 500 mL RRapn 12 /min MVe-high alarm 40.0 L/min MVe-low alarm 0.5 L/min RRsp-high alarm 100 /min HME correction OFF AutoFlow OFF Brightness-min* 1/4 Brightness 3/4 Alarm volume 3/4 etCO
2 high alarm 50 mmHg
etCO
2 low alarm 33 mmHg
CO
2 unit mmHg
Hose type Adult disposable Cuvette type Disposable
100
Instructions for Use Oxylog 3000 plus SW 1.n
Loading...