Dräger Oxylog Instructions For Use Manual

!
Oxylog
Emergency Ventilator
Instructions for Use
Deutscher Text: Bitte umdrehen
For Your Safety and that of
Your
Patients
For correct and effective use of the apparatus and to avoid hazards it is essential to read the following recommendations and to act accor­dingly’):
Strictly follow the instructions for use
~ Any use of the apparatus requires full
understanding and strict
observatton
of these instructions. The apparatus is on/y to be used for purposes specified here.
MaintenancE
The apparatus must be inspected” and
serviced” by experts at regular 2 years intervals (and a record kept).
We recommend obtaining a service contract with
DMgerService.
Repairs21
and general overhaul on the
apparatus may on/y be carried out by
experts. General overhaul by
DragerService
of pres­sure reducers should occur every 6 years. On/y original
Drtiger
spare parts may be used
for maintenance.
e
owner or operator to the extent that the apparatus has been
serviced
or repaired by
personnel not employed or
authonzed
by
DrtigerService
or when the apparatus was used /n a manner not conforming to its
untended
use.
Dragerwerk
Akt/engese//schaft
cannot be he/d responsible for damage caused by non-compliance with the recommendations given above. The warranty and
//ab/hty
provisions of the terms of sale and delivery of Dragerwerk
Aktiengesellschaft
are likewise
not
mod/fled
by the recommendations
gfven
above. Dragerwerk
Akt/engese//schaft
standards, these are based on
the
legal system of
the federal
Republic
of Germany
2,
In accordance
wth
DIN
31051’
Inspection
= exammarIon
of actual
cond/t/on
Serwce
= measures to
mafntaln
desired
I
cond,t,on
Liability for proper function or damage
RepaIr
= measures to restore
desired
condmon
The liability for the proper function of the
Maintenance
tnspectfon, serwce
and.
of
apparatus is irrevocably transferred to the
applfcable, repa,r
Contents
Page
lntented Use
..............................................
3
What’s What?
............................................
3
Mounting and Usage of Holder
.................
6
Initial Preparation ......................................
6
Determination of Compressed-Gas
1
Supply and Usage Period .........................
11
Functional Check
......................................
11
Operational Use
........................................
11
Shut-Down Actions
...................................
13
Page Oxylog with demand valve to facilitate spontaneous breathing
...............
14
Care
..........................................................
19
Inspection
..................................................
20
Storage ......................................................
20
Trouble Shooting .......................................
21
Technical Data
..........................................
22
Order List
..................................................
24
Parts List
...................................................
27
2
Intended Use
The Oxylog is a ventilator for providing controlled ventilation of infants as of 5
kg body weight and adults on a
time-
cycled, volume-constant basis. The device is designed for mobile use by rescue services, for transportation to hospitals in ambulances or helicopters,
What’s What?
(Figs. 1-3)
Oxylog (Figs. 1 and 2)
1 Airway pressure gauge (scale
range -10 mbar to + 80 mbar)
2
Zero adjustment of airway pressure
gauge
3
Rotary knob for setting ventilation ratio
3a Heart symbol - adjustment aid for
ventilation during cardiopulmon­ary resuscitation: ratio 12
min’
4 Rotary knob for setting minute
volume (MV)
5
Pneumatic main switch I-O
6
Switch
asAir Mix<< - >>No
Air
Mix<<
for transferring patients by road or air, for ventilation in the emergency admis­sions department and for transferring patients receiving ventilation from one department to another, e. g. from the operating theatre to the intensive-care ward.
7 Stud for attaching carrying strap
(also serves to secure device in holder)
Patient system (Items 8-10)
8 Ventilation tubing 9 Non-rebreathing valve (with exter-
nal taper
ISO,
dia. 22) for breathing mask or catheter connector (with internal taper
ISO,
dia. 15) for tube
15
Operating instructions in brief
Fig. 1 Oxylog with patient system connected
I
n
Fig. 2 Oxylog -gas inlets and outlets
11
Socket for ventilation tubing
13
Compressed-gas connection (male
12 Filter for purifying ambient air
thread M 15 x 1)
drawn in
14 Slot for alternative fastening of
carrying strap, e. g. when using
holder
4
Waste-gas socket or
PEEP valve connection
Patient connection
Connection to Oxylog
Fig. 3 Non-rebreathing valve
Non-rebreathing valve
(Fig. 3, Items 9.1-9.3)
9.1 Cover
with
connection for ventila-
tion tubing (external taper
ISO,
dia. 22)
9.2
Valve housing with patient connec­tion and waste-gas socket. The patient connection features an external taper
ISO,
dia. 22 and an
internal taper
ISO,
dia. 15.
The
waste-gas socket has an internal
taper
ISO,
dia. 22 for connection of
a PEEP valve.
9.3 Diaphragm (complete) comprising control diaphragm and checkvalve.
The control diaphragm and check valve are marked yellow and red to enable their presence and correct
installation position in the valve housing to be checked.
e
’ .;,,,:,,
‘;
/’
_ .-.-.-.-:i:&.. .A.!.i
Mounting and Usage
of Holder
The Oxylog is secured in position in the
rescue vehicle by way of the holder
8404560 (Fig. 4). The screws for attaching the basic
element are provided.
Mounting of holder
The basic element 16 is provided with sufficient holes for the fastening screws.
At least 3 screws (with maximum
possible spacing between them) are to be used in each case. The installation location is arbitrary.
Insertion of Oxylog into holder
Push device into holder such that stems of two studs 7 on housing slide into slots
Fig. 4 Holder 6404560 (without bracket)
6
16a of holder. Studs must engage in
hole in brackets 17. Press on brackets 17 to ensure that the
device is firmly secured in position in the holder.
The guide stud 19 secures the device on the back. Fig. 5 illustrates the Oxylog in position in the holder.
Removing Oxylog
Pressing open the two brackets 17 releases the studs 7 and enables the
device to be removed from the holder. Holder with rail bracket (Fig. 6)
The holder is used for attachment to the
Drager
wall rail system (10 x 25 mm
section).
Handling is the same as for the holder
8404560.
Fig. 5 Oxylog in holder
Fig. 6
Holder with
rail
bracket
,,
--- _...___.. _.,.L._
:
Attachment of carrying strap on back
event of constant mobile use outside the
of device
rescue vehicle - the carrying strap can
If the carrying strap is constantly in
use-
be secured in position in the slots in the
e. g. including cases where Oxylog is
Oxylog housing. See attachment
dia-
accommodated in holder and in the
gram in Fig. 7.
Fig. 7 Attaching carrying strap on back of device
Initial Preparation
An inlet pressure at the device
(com-
The functioning of the Oxylog can be
pressed-gas connection 13 in Fig. 2) of
checked with the compressed-gas
at least 2 bar at a flow rate of 60
L/min is
source (see
B>Functional Check<<
on
required for operation of the Oxylog.
page 11).
The drive sources used (central gas
1
supply or gas cylinder with pressure reducer) must always comply with the
(
aboveprerequisite.
Operation from a central gas supply
Screw
OJair
connecting hose (3 m or 5
Any upstream closing or metering
m, see Order List) to device and insert
elements must be fully open!
connector into wall outlet valve.
8
,. .
Operation from a gas cylinder
Screw pressure reducer to cylinder. Check 2-6 bar delivery-pressure set­ting. Screw connecting hose (1.5 m or 3 m, see Order List) to device and to pressure reducer. Fully open cylinder valves.
Use with
DrSger OxatoP
The connection piece 2M 19051 is exclusively designed for use of the Oxylog on the Oxator head (Fig. 8). This connection piece is screwed to the Oxator head and consists of a
self-
closing, standard
O2
coupling for connection of the Oxylog using the standard, gas-specific connector
(02/
air). A self-closing inlet screw connec­tion for O2 makes it possible (in addition to the use of cylinders) to supply the Oxylog and the other components connected to the Oxator head (e. g.
humidifier/nebulizer or aspiration ejec­tor) from a central gas supply unit (see Oxator Operating Manual). The Oxylog may not be fed via the flow control valves at the Oxator head!
Fig. 8 Oxator head with connection piece
for Oxylog
Use of gas blenders
.
In the case of lengthy ventilation, e. g. during repatriation flights, low, defined
O2
concentrations may be required.
For this purpose a compressed-gas blender can be connected upstream of the Oxylog (see Order List).
Caution! The inlet pressure at the Oxylog must however be at least 2 bar at a flow rate of approx. 60
L/min
(see
BjFunctional
Check<<
on page 11).
The switch is to be set to
a>No
Air
Mix<<.
Equipping
(Fig. 9)
0
Assemble non-rebreathing valve as
per Fig. 10. Ensure that entire diaphragm is correctly positioned and take particular care to ensure that red check valve is present and not out of shape. Screw cover to valve housing (turn 45” in clockwise direction).
0
Attach ventilation tubing to socket at
Oxylog as well as to
~~lnspiration~~
socket on non-rebreathing valve.
0
If use is made of an (optional)
externally adjustable pressure limit­ing valve (see Order List), attach this valve first to the socket at the Oxylog and then connect the ventilation tubing to the socket of the pressure limiting valve.
0
If PEEP is being applied, insert PEEP
valve (see Order List) into waste-gas socket of non-rebreathing valve.
0
Set airway pressure gauge to zero.
0
Preselect O2 concentration
a>Air Mix<<
setting reduces drive-gas consumption by roughly 50% as ambient air is sucked in.
Loading...
+ 20 hidden pages