Dräger EM200-E
Gaspreconditioner
Gas inlet
LED and
Infrared transmitter
USB interface
Graphic display, illuminated
Control keys „F, ▲, ▼, H“
Shock and impact resistant housing
Diesel Exhaust Measurement
The Dräger EM200-E is an electronic multigas instrument with internal sampling pump. It precisely determines CO and NOx content in undiluted exhaust gases of Diesel engines. The internal ambient pressure sensor allows the immediate use without preceding calibration in underground mines.
02 | Dräger EM200-E
Benefits
Shock and impact proof housing
The Dräger EM200-E has an impact and shockproof body. Easy handling and an excellent technology allow precise and fast results of measurement. The high power battery affords a high operating time of typically ten hours.
Precise measurement of CO and NOx content in exhaust gases from diesel engines
One multigas analysis instrument to be used for precise testing of diesel engines in underground mining applications: regulation compliance (in some countries maximum emissions from engines are regulated to be able to set ventilation rates), health risk (a build up of these gases can be dangerous to individuals; CO, NO and NO2 have regulated exposure concentrations), operating costs (increases in CO/NOx output can indicate
a problem with the engine or the scrubbers resulting in reduced fuel efficiency), reduce downtime (using these values as pre-indicator for engine breakdowns can enable preventive maintenance instead of reactive measures and loss of production).
Documentation
Saved data can be locally documented or transferred to the PC.
High precision measurement in all different depths
An internal pressure sensor automatically compensates ambient pressure influences in different depths. Therefore engines do not have to go to the workshop as a “fixed” testing area. The Dräger EM200-E comes to the engine. No matter where it is located: 4,000 m below or 4,000 m above sea level.
Accessories
D-544-2016
Dräger EM200-E | 03
Vehicle exhaust probe
For measuring exhaust gases.
Fabric tube, per meter
D-545-2016
D-541-2016
D-542-2016
Equipment case IP67 EM200 line
Plug-in charger
100 – 240 V
04 | Dräger EM200-E
Accessories
D-543-2016
Infrared printer MSI IR3
Related Products
D-17812-2016
Dräger EM200plus-f / Dräger EM200plus-i
The Dräger EM200plus-f and EM200plus-i are your partners for measuring and testing heating and large industrial boilers. The series performs measuring tasks such as flue gas temperature, NOx measurement, SO2 measurement, the determination of the combustion
efficiency or CO measurement up to 30,000 ppm just as reliably as solid fuel measurement or the calculation of flow velocities (version­dependent).
Dräger VARIOx-2
In addition to adjustment tasks, the single-channel device can optimise combustion for energy conservation or short-term emission measurement. It reviews limits and is suitable for process analysis.
D-50237-2015
Dräger EM200-E | 05
Technical Data
Display Liquid crystal graphic display, backlit, manual adjustable Interface USB for PC interface, infrared for printer, no multifunction jack for
additional instruments Operating temperature +5 °C … +40 °C Storage temperature -20 °C … +50 °C Power supply Internal: high power battery, 4.8 V 2,000 mAh, indication of state
of charge
External: plug-in charger Operating capacity Typically 10 hours Gas extraction Membrane pump for gas sampling Gas processing Integrated gas conditioning cartridge with condensate trap and
particle filter Weighting 1,100 g Dimensions 195 mm x 165 mm x 75 mm (H x W x D)
Display Principle of
measurement
CO, carbon monoxide
NO, nitric oxide El.-chem. sensor 0 … 2,000 ppm 1 ppm 0 … 600 ppm: ± 10 ppm or
NO2, nitrogen dioxide NOx, oxides of nitrogen Barometer Piezo-resistive 800 ... 1,700 hPa 1 hPa MV3 = Measuring Value
El.-chem. sensor 0 ... 8,000 ppm 1 ppm
El.-chem. sensor 0 … 200 ppm 1 ppm 0 … 100 ppm: ± 5 ppm or
El.-chem. sensor 0 up to 2,000 ppm 1 ppm
Measuring range Resolution Accuracy
0 … 200 ppm: ± 10 ppm or
10 % of MV³
200 … 2,000 ppm: ± 20 ppm or
5 % of MV³
2,000 … 8,000 ppm: ± 100 ppm or
10 % of MV³
5 % of MV³
5 % of MV³
06 | Dräger EM200-E
Notes
Not all products, features, or services are for sale in all countries. Mentioned Trademarks are only registered in certain countries and not necessarily in the country in which this material is released. Go to www.draeger.com/trademarks to find the current status.
CORPORATE HEADQUARTERS
Drägerwerk AG & Co. KGaA Moislinger Allee 53–55 23558 Lübeck, Germany www.draeger.com
Manufacturer:
Dräger MSI GmbH Rohrstr. 32 58093 Hagen, Germany
REGION EUROPE
Dräger Safety AG & Co. KGaA Revalstraße 1 23560 Lübeck, Germany Tel +49 451 882 0 Fax +49 451 882 2080 info@draeger.com
.
REGION MIDDLE EAST, AFRICA
Dräger Safety AG & Co. KGaA Branch Office P.O. Box 505108 Dubai, United Arab Emirates Tel +971 4 4294 600 Fax +971 4 4294 699 contactuae@draeger.com
.
REGION ASIA PACIFIC
Draeger Singapore Pte. Ltd. 61 Science Park Road The Galen #04-01 Singapore 117525 Tel: +65 6872 9288 Fax: +65 6259 0398 asia.pacific@draeger.com
.
REGION CENTRAL
AND SOUTH AMERICA
Draeger Panamá Comercial S. de R.L. 59 East Street, Nuevo Paitilla, House 31, San Francisco Town Panama City, Panama Tel +507 377-9100 / Fax -9130 servicioalcliente@draeger.com
.
Locate your Regional Sales Representative at: www.draeger.com/contact
91 04 112 | 21.02-3 | HQ | PP | Subject to modifications | © 2021 Drägerwerk AG & Co. KGaA
REGION NORTH AMERICA
Draeger, Inc. 3135 Quarry Road Telford, PA 18969 Tel +1 800 4DRAGER (+1 800 437 2437) Fax +1 215 723 5935 info.usa@draeger.com
.
Loading...