Drucker Diagnostics Horizon 24, Horizon 12, Horizon 24 AV-AH, Horizon 12 Flex Service Manual

Page 1
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
1
Page 2
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
2
Page 3
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
3
Page 4
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
4
Maximum Speed
3400 RPM
Maximum RCF
2000
Dimensions (in)
9.0 (H) x 15 (W) x 17 (L)
Environmental Operating Range
16-32 deg C
Typical Noise Level (At Maximum Speed)
< 59 dB A
Electrical Rating
95-253 VAC
PROBLEM
POSSIBLE CAUSE
SOLUTION
The lid does not open.
No Power
Check removable line cord or wall outlet.
Lid knob is ajar
Rotate the lid knob fully clockwise before pressing the ‘UNLOCK’ button
Lid lock is active (Unlock timed out)
Press the ‘UNLOCK’ button to de-activate the lid
Lid tray is unplugged from PCB or defective
Requires service PCB is damaged
Requires service
To gain access to the rotor - Remove the label directly in front of the lid knob and slide the lid latch lever toward the front of the centrifuge. This will unlock the lid.
PROBLEM
POSSIBLE CAUSE
SOLUTION
Excessive vibration
Rotor improperly loaded
Load equally filled tubes symmetrically in the rotor. All carriers and/or tube holders must be present in the rotor, whether loaded, or empty.
Debris lodged within the rotor or tube carriers
Carefully inspect all rotor pockets, tube holders and crevasses for debris.
Centrifuge housing is loose
Requires service
Missing/damaged feet
Requires service
Motor failure
Requires service
Rotor damaged
Replacement required
PROBLEM
POSSIBLE CAUSE
SOLUTION
Rotor does not spin
No Power
Check removable line cord
No Power
Check circuit breaker on underside of centrifuge.
No Power
Check wall outlet
Lid not properly latched
Press down firmly on lid and rotate lid knob fully clockwise.
Internal connection failure
Requires service
PCB failure
Requires service
Motor Failure
Requires service
PROBLEM
POSSIBLE CAUSE
SOLUTION
Clicking noise during braking
Rotor is loose
Tighten rotor screw per section 7-2
PROBLEM
POSSIBLE CAUSE
SOLUTION
Whistling noise while running Debris in air intake / exhaust ports
Remove power before clearing debris.
Gasket failure
Requires service
Page 5
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
5
PROBLEM
POSSIBLE CAUSE
SOLUTION
Cycle ends prematurely. The centrifuge reports the error in a repeating BUZZ / BEEP sequence
BALANCE ERROR 1 ERROR BUZZ + 2 REPEATING BEEPS Rotor improperly loaded
Load equally filled tubes symmetrically in the rotor.
Debris lodged within the rotor or tube carriers
Carefully inspect all rotor pockets, tube holders and crevasses for debris.
Rotor damaged
Replacement required
LOW SPEED ERROR 2 ERROR BUZZES + 2 REPEATING BEEPS
Rotor Speed is under set speed for too long.
Check the lid and guard bowl gaskets. Voids in the rotor chamber gaskets change the airflow, resulting in increased stress on the motor drive
OVERCURRENT ERROR 3 ERROR BUZZES + 3 REPEATING BEEPS
Motor or PCB failure
Service Required
NO RPM ERROR 6 ERROR BUZZES + 4 REPEATING BEEPS
Sensor not connected to PCB
Check connections
Sensor failure
Replace motor
Locked rotor
Service Required
EEPROM ERROR 2 ERROR BUZZES + 6 REPEATING BEEPS
PCB failure
Replace PCB
ACCELEROMETER ERROR 2 ERROR BUZZES + 7 REPEATING BEEPS
PCB failure
Replace PCB
MOTOR DRIVE ERROR 2 ERROR BUZZES + 9 REPEATING BEEPS
PCB failure
Replace PCB
Page 6
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
6
Page 7
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
7
Page 8
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
8
Page 9
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
9
Tiedown
SPLICE
SPLICE
THIS SIDE
Page 10
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
10
Page 11
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
11
Page 12
DRUCKER DIAGNOSTICS HORIZON 24 SERVICE MANUAL
SM016
REV A
12
Loading...