Drucker Diagnostics 853VES User Manual

MODEL 853VES
Model 853VES Operator’s Manual
Laboratory Centrifuge
Providing Quality Centrifuges since 1932
P/N 7711012
Rev. B
Table of Contents
Providing Quality Centrifuges since 1932
Model Description .................................................................. 3
Supplied Equipment ............................................................... 3
Warranty Information .............................................................3
Specifications ......................................................................... 4
Control Panel / Parts of the Centrifuge ................................ 5
Setup Location ........................................................................ 6
Initial Setup Procedure ..........................................................6
Operation ................................................................................ 7
Tube Holder Configurations
Control Panel Functions
Advanced User Settings
Memory Locations / Storing and Recalling ....................... 12
Digital Cycle Counter ........................................................... 12
Care and Preventative Maintenance ...................................13
Troubleshooting .................................................................... 14
Safety ..................................................................................... 15
Emergency Rotor Chamber Entry ....................................... 15
Calibration and Earth Ground Testing ................................ 15
Replacement Parts ............................................................... 16
02
WARNING
For the safety of both the operator and service personnel, care should be taken when handling substances that are known to be toxic, radioactive or contaminated with pathogenic microorganisms when using this centrifuge. When Risk Group II materials are used (as identified in the World Health Organization “Laboratory Bio-Safety Manual”), a Bio-Seal should be employed. More than one level of protection must be provided in the case of materials of a higher risk group. The use of flammable or explosive materials as well as those materials which chemically react vigorously is prohibited.
Model Description
The Horizon Model 853VES, Plasmafuge-12, is a medium capacity centrifuge with unmatchedperformance. The 853VES offers control over almost every aspect of the unit’s operation from variable acceleration and deceleration to easy time and speed input. Save settings in one of 10 memory locations and recall them at the touch of a button. High speed and force capability produces platelet-poor/free plasma and greatly reduces processing times. The exclusive Drucker horizontal rotor allows for quick and easy sample loading and complete horizontal separation.
Intended Use
General purpose laboratory centrifuge for sample separation.
Supplied Accessories (Standard)
See next page for optional rotor/accessories
*The rotor and rotor accessories are rated for a rotational speed of 5000 RPM.
One (1) 12-Place
Horizontal Rotor
7786024*
Also included (not shown):
• One (1) 10´ Line Cord
• One (1) Operator’s Manual
Twelve (12) 100 mm Carriers
7713031*
Twelve (12) 75 mm Carriers
7713033*
See the next page for a description of optional accessories.
WARRANTY:
The Drucker Company warranties that this centrifuge is free from
defects in workmanship and parts for 2 years.
Made in the USA by
Providing Quality Centrifuges since 1932
200 Shadylane Drive • Philipsburg, PA 16866
Phone: 814-342-6205 or 814-692-7661 • Fax: 814-692-7662 • www.druckercompany.com
03
Specifications
General Specications for the Horizon Model 853VES Centrifuge
Overall Dimensions (H x W x D): ............... 8.5˝ x 12.5˝ x 15.5˝
Centrifuge Motor: ...................................... 1/2 H.P. Brushless DC
Protection Breaker: ................................... 4 Amp. re–settable
Timer: .......................................................... Electronic, with hold or 0 to 99 minutes, +/- 1%
Weight: ........................................................ 30 lbs.
Permitted Environmental Conditions
Ambient Temperature During Operation: 2 oC - 40 oC
Maximum Relative Air Humidity: .............. 90%
Electrical Requirements
Power (Watts): ............................................ 200
Voltage (Volts): standard: ........................... 115 (+/- 10)*
optional: ............................ 230 (+/- 20)*
853VES: ............................ 115 Volts
853VES 230V: ................... 230 Volts
Frequency (Hz): .......................................... 50/60
*Check Serial Number Badge for Your Model Specific Voltage Requirement.
PART NO. # MAX TUBE SIZE SPEED RANGE FORCE RANGE
7786024 Comes with Twelve 12 of each of the tube holders listed
12-Place Horizontal Rotor (Standard)
75 mm tube holder 7713043 12 17 mm x 75 mm 500 to 4,000 RPM 40 to 2,000 xg 100mm tube holder 7713040 12 17 mm x 100 mm 500 to 4,000 RPM 40 to 2,400 xg
7786010 Comes with Twelve 12 of each of the tube holders listed
12-Place Performance Plus Horizontal Rotor
75 mm tube holder 7713043 12 17 mm x 75 mm 500 to 5,000 RPM 40 to 1,100 xg 100 mm tube holder 7713040 12 17 mm x 100 mm 500 to 5,000 RPM 40 to 3,800 xg
20-Place Fixed-angle Rotor
7786042 Comes with Twenty 20 of each of the tube holders listed
75 mm tube holder 7713043 20 17 mm x 75 mm 500 to 4,000 RPM 40 to 1,700 xg 100 mm tube holder 7713040 20 17 mm x 100 mm 500 to 4,000 RPM 40 to 1,900 xg 125 mm tube holder 7713044 20 17 mm x 125 mm 500 to 4,000 RPM 40 to 2,000 xg
6-Place 50 mL
7786044 Comes with 6 of the holders listed
Fixed-angle Rotor
50 mL Fixed-angle 7787051 6 34 mm x 125 mm 500 to 4,000 RPM 40 to 2,000 xg
Contact your authorized dealer or The Drucker Co. to order accessories.
75 mm Tube Holder
p/n 7713043
100 mm
Tube Holder
p/n 7713040
125 mm
Tube Holder
p/n 7713044
Tube Holder Cap
p/n 7713011
50 mL Tube Holder
Fixed-angle p/n
7787051
1˝ Cushion Spacer
p/n 1525
04
Any use other than those specified by the Manufacturer is explicitly prohibited. Maximum sample density is 1.15 grams / mL, (water density = 1.0 grams / mL)
Control Panel / Parts of the Centrifuge
Lid
Lid Knob
Lid Safety
Interlock System
Control Panel
(see below)
Cabinet
RPM Display
RCF Conversion
Button
Time Display
Up/Down RPM
Change Buttons
Up/Down Time Change Buttons
Running Indicator Light
Brake Display
Program Button
Latched Indicator Light
Memory Preset Display
Unlocked Indicator Light
Start Button
Open / Stop Button
Memory Button
05
Setup Location
1. Unpack the centrifuge and verify that all of the supplied equipment is present.
2. Choose a setup location which meets the following criteria:
a) A bench top clearance height of 20” is required in order to open the lid.
b) The clearance envelope is the space around the centrifuge which is required for safety. Choose a
setup location which will allow for a clearance envelope of at least 30” x 30”, (with the centrifuge at the center). No person or hazardous material shall be permitted in the clearance envelope during operation. The operator time within the envelope shall be limited to the time necessary for loading, unloading and centrifuge operation only.
c) Proper ventilation is necessary to prevent the overheating of samples as well as premature
failure of the centrifuge. Choose an area which will allow unencumbered air ow.
d) The centrifuge is designed to rest on its four rubber feet. No adjustment is necessary for leveling
the centrifuge, however, the surface should be flat and level.
e) Be sure the outlet is always within reach as the line cord is the means of emergency disconnection.
Initial Setup Procedure
If any problems are found during the initial setup procedure, refer to the troubleshooting section. For further assistance, contact The Drucker Company at 814-342-6205 or +1-814-692-7661.
1. Plug the female end of the line cord into the rear of the centrifuge. Plug the male end into an approved electrical outlet. For electrical safety, the unit must always be properly grounded.
2. For operator safety, the locking system is always active; requiring power and direction from the user to disengage it (the lid also automatically unlocks at the end of a run when it is safe to do so). To unlock the lid (in order to access the rotor chamber) press the ‘OPEN / STOP’ button
on the control panel. The ‘UNLOCKED’ indicator light should illuminate. If it does not, refer to
the section on troubleshooting. The lid will remain unlocked for 15 seconds after pressing the ‘OPEN / STOP’ button.
3. Turn the latch counterclockwise and open the lid.
4. Spin the rotor by hand; check for free and level rotation. If the rotor does not spin freely, refer to the section on troubleshooting.
5. Place the test tube holders (or carriers depending on your rotor type) inside the rotor and verify that they are seated properly.
6. Close the lid. Rotate the lid knob clockwise to its complete stop position. The ‘LATCHED’ indicator
light should be illuminated. If it is not, make sure that the lid is latched properly. The centrifuge
will not run unless the lid is latched properly and the ‘LATCH’ indicator light is illuminated.
7. Use the up arrow button next to the RPM display to set the speed to full.
8. Initiate a test run by pressing the ‘START’ button.
(Continued on next page)
06
Initial Setup Procedure (continued)
9. The ‘RUNNING’ indicator light will illuminate.
10. The test tube carriers will slide up into the horizontal position (horizontal rotor only)
and the unit will accelerate to the current set speed.
11. Listen to the sound of the centrifuge. A smooth whirring sound should be heard. If there are any loud or unusual sounds, stop the centrifuge by pressing the ’OPEN / STOP’ button immediately and refer to the section on troubleshooting.
12. Press the ‘OPEN / STOP’ button to terminate the test run. The rotor will decelerate to a complete stop and the lid will then unlock automatically for sixty (60) seconds.
13. Take time now to familiarize yourself with the various additional user settings. Turn to page 10/11.
14. The centrifuge is now ready for operation.
Operation
NOTE: Follow the initial setup procedure before initial operation.
1. Press the ‘OPEN / STOP’ button to unlock the lid and then open the lid.
2. Insert cushions (if needed) into the tube carriers (holders) for the tube size you are using. Refer
to ‘Tube Holder Congurations’ for assistance.
3. Place the test tube samples into the tube carriers (holders). Be sure to follow the rules for balanced loads.
4. Close the lid and turn the lid knob clockwise to its complete stop position. The ‘LATCHED’
indicator light should illuminate to indicate that the latch is closed properly. If the lid knob is not
completely latched, the ‘LATCHED’ indicator light will not illuminate and the centrifuge cannot
be operated.
5. Set the desired speed and run time using the appropriate up and down arrow buttons.
6. If desired, adjust the braking level, acceleration rate, countdown delay, imbalance sensitivity and audible indication. Turn to page 11 for a description of these features.
7 Begin the run by pressing the ‘START’ button on the control panel.
8. The centrifuge should begin to spin. The ‘RUNNING’ indicator light should illuminate.
IF A PROBLEM IS FOUND DURING A SPIN THAT REQUIRES THE CENTRIFUGE TO SHUT DOWN, PRESS THE ‘OPEN / STOP’ BUTTON IMMEDIATELY !
9. After time has elapsed, the ‘RUNNING’ indicator light will extinguish and the rotor will decelerate to a complete stop.
10. The ‘UNLOCKED’ indicator light will illuminate and the locking mechanism will disengage,
allowing entry into the rotor chamber. If the automatic unlock times out (after 60 seconds) simply press the ‘OPEN / STOP’ button.
11. Turn the lid knob counterclockwise and open the lid.
12. Remove the samples.
13. The centrifuge may be used again immediately.
07
Tube Holder Configurations
The next two pages will describe proper balancing and tube loading. Use the appropriate section for the rotor installed in your centrifuge. Improper loading can result in damage to the centrifuge or test tube samples.
Twelve (12) Place Horizontal Standard and Performance Plus Rotors
This rotor is designed to hold up to twelve (12) tube holders. It comes with two tube holder types. The purple holders are designed to hold tubes up to 10 mL (up to 17 mm x 100 mm). The blue holders are designed to hold tubes up to 5 mL (up to 17 mm x 75 mm).
Your centrifuge must contain a balanced load in order to work properly. To ensure that the load is balanced, keep these rules in mind when inserting test tube samples into the 12 place horizontal rotor:
1. Opposing tube holders must be identical.
2. Opposing tube holders must be empty or loaded with equally weighted samples.
3. If an odd number of samples is to be spun, use a water-filled tube to mate with the unpaired one.
Directions:
A.
B.
1. Compare the tube to be spun with the boxes shown here.
2. Find the box that most closely matches the tube’s length.
NOTE: The tube length with its stopper or cap must be no longer
than the chosen box.
3. Match the letter from the chosen box with one of the configurations
below. For Example: A tube is found to be as long as box B.
Accordingly, we can use a 100 mm tube holder with a 1525 cushion or a 75 mm tube holder with no cushion, (configurations B1 or B2).
Standard
Performance Plus (optional)
A.
PURPLE
C.
100mm Tube Holder
B1.
PURPLE BLUE
B2.
C.
BLUE
Six (6) Place 50 mL Fixed-angle Rotor (optional)
This rotor is designed to hold up to six (6) 50 mL tube holders.
Your centrifuge must contain a balanced load in order to work properly. To ensure that the load is balanced, keep these rules in mind when inserting test tube samples into the 6 place, 50 mL fixed-angle rotor:
1. Opposing tube holders must be identical.
2. Opposing tube holders must be empty or loaded equally weighted samples.
3. If an odd number of samples is to be spun, use a water-filled tube to mate with the unpaired one.
75mm Tube Holder with 1525 cushion
08
Twenty (20) Place Fixed-angle Rotor (optional):
This rotor is designed to hold up to twenty (20) tube holders. It comes with three tube holder types. The orange holders are designed to hold tubes up to 15 mL (up to 17 mm x 125 mm). The purple holders are designed to hold tubes up to 10 mL (up to 17 mm x 100 mm). The blue holders are designed to hold tubes up to 5 mL (up to 17 mm x 75 mm).
Your centrifuge must contain a balanced load in order to work properly. To ensure that the load is balanced, keep these rules in mind when inserting test tube samples into the 20 place fixed-angle rotor:
1. Opposing tube holders must be identical.
2. Opposing tube holders must be empty or loaded equally weighted samples.
3. If an odd number of samples is to be spun, use a water-filled tube to mate with the unpaired one.
Directions:
1. Compare the tube to be spun with the four boxes shown here.
2. Find the box that most closely matches the tube’s length.
NOTE: The tube length with its stopper or cap must be shorter than the chosen box.
3. Match the letter from the box with one of the configurations shown.
For Example: A tube is found to be as long as box B. Accordingly, we can use a
125 mm tube holder with a 1525 cushion or a 100 mm tube holder with no cushion, (configurations B1 or B2).
A.
B.
C.
D.
ORANGE ORANGE
125 mm Tube Holder
C1. C2. D.
A.
PURPLE
B1. B2.
125 mm Tube Holder with 1525 cushion
BLUE
or
PURPLE
BLUE
100 mm Tube Holder
75 mm Tube Holder with 1525 cushion
09
Control Panel Functions
Time Adjustment and Timer Operation: The run time may be set from 30 seconds to 99 minutes
and 30 seconds. Press the up and down arrow buttons next to the time display to adjust the run time. Adjustments may be made prior to a run or after a run begins. A quick tap will adjust the time by 30 seconds. Hold down the button for 1 minute adjustments - continue to hold for 5 minute adjustments.
Press the down button once while 00:30 is displayed to access the “hold” feature. When a “hold” run
is initiated, the timer will count up to keep track of total run time. The unit will run for a maximum of 99 minutes and 30 seconds and will then shut off automatically. The “hold” run may also be terminated by pressing the STOP/OPEN button. The total run time during a “hold” run is displayed on the time display until the operator unlocks the lid by pressing the OPEN button or opens the lid (during the automatic unlock after the run ends).
Speed Adjustment / RCF Display: Press the up and down arrow buttons next to the speed
display to change to run speed. Adjustments may be made prior to a run or after a run begins. A quick
tap will adjust the speed by 50 RPMs. Hold down the button for 100 RPM adjustments. Continue to
hold for 500 RPM adjustments.
Press the RCF button at any time to convert the current speed setting to the corresponding force value (xg). The speed may be set by RCF by holding down the RCF button while using the up/down speed set buttons. The jumps made in the RCF value while holding down the RCF button and pressing the up and down arrow buttons correspond to 50 RPM adjustments in speed. Since the RCF is a
function of the square of the speed, this jump size will vary.
Starting and Stopping a Run: With the lid switch closed, press the START button to begin
a run. Press the STOP/OPEN button at any time during a run to terminate it.
Unlocking the Lid: The lid is unlocked whenever the red UNLOCKED indicator light is illuminated.
The lid unlocks automatically for 60 seconds at the end of a run. The lid can also be unlocked for an additional 15 seconds by pressing the OPEN/STOP button while the unit is idle. The lid cannot be unlocked while the rotor is spinning.
Imbalance Detection Reset: This centrifuge is equipped with imbalance detection. If an imbalance
is detected, the centrifuge will terminate the current run and will begin to brake to a stop. The words
“BALANC” and “ERROR” will ash on the speed display. Once the rotor has stopped, open the lid
the cancel the error reporting, balance the load and begin a new run. Alternately, the error reporting can be canceled by pressing the OPEN/STOP button
Advanced User Settings
This centrifuge allow for total control of the run by the user. You may adjust the acceleration rate, adjust the braking rate, adjust the imbalance sensitivity level, select from several audible settings,
and use the optional countdown delay.
Acceleration Rate: The user may adjust the acceleration rate from minimum (1) to maximum (10)
using the following procedure:
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “ACCEL” is displayed in the
speed display.
3. The current acceleration rate is displayed in the time display. Use the ‘UP/ DOWN arrow buttons next to the time display to adjust the acceleration rate from 1 (minimum) to 10 (maximum).
4. Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to the speed display to adjust additional settings.
10
Advanced User Settings (continued)
Braking Rate: The user may adjust the braking rate from minimum (slowest deceleration, 0) to
maximum (quickest deceleration, 9) using the following procedure:
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “BRAKE” is displayed in the
speed display.
3. The current braking rate is displayed in the brake display. Use the UP/DOWN arrow buttons next to the brake display to adjust the braking rate from 0 (minimum) to 9 (maximum).
4. Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to the speed display to adjust additional settings.
Countdown Delay: When the countdown delay is turned on, the timer will not begin decrementing
until the rotor has reached full speed as set by the user. This feature ensures that the samples see full force for the total run time.
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “CNTDLY” is displayed in the
speed display.
3. The current status of this feature (ON or OFF) is displayed in the time display. Press UP/DOWN arrows next to the time display to toggle the status between ON and OFF.
4. Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to the speed display to adjust additional settings.
Imbalance Detection: This centrifuge is equipped with imbalance detection. This ensures
that the centrifuge will safely shut down in the event that an imbalanced load is run or a malfunction occurs. This feature is intended to provide for operator safety and to extend the life of the centrifuge. The LOW setting will allow for a greater imbalance before shutting the centrifuge down.
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “BALANC” is
displayed in the speed display.
3. The current status of this feature (LOW or HI) is displayed in the time display.
Press UP/DOWN arrows next to the time display to toggle the status between LOW and HI.
4. Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to
the speed display to adjust additional settings.
Audible Settings: The user may select from six different audible indicator levels:
Setting Button Press Beep? End-of-run audible notification
1 .............NO .............................. NONE
2 .............YES ............................ NONE
3 .............NO .............................. YES, 5 beeps
4 .............YES ............................ YES, 5 beeps
5 .............NO .............................. YES, continuous, user must press STOP/OPEN button to cancel
6 .............YES ............................ YES, continuous, user must press STOP/OPEN button to cancel
1. While the centrifuge is idle, press the PROGRAM button.
2. Use the UP/DOWN arrow buttons next to the speed display until “BEEPER” is displayed in the speed display.
3. The current audible selection is displayed in the time display. Use the UP/ DOWN arrow buttons next to the time display to select a setting (1 to 6).
4. Press the PROGRAM button to exit or use the UP/DOWN arrow buttons next to the speed display to adjust additional settings.
11
Memory Locations / Storing and Recalling
The Performance Series centrifuges are capable of storing up to 10 user-defined presets. These memory presets contain all of the settings needed to define a specific run, (run time, speed, acceleration rate, braking rate, etc.). The user can use these memory locations to quickly configure the centrifuge for a
specific test type and ensure that the test is run in the same way each time by recalling this same setting.
Store a configuration in a memory location
1. Adjust the various settings (speed, time, acceleration rate, etc.) for the configuration to be saved.
2. Press and hold the PROGRAM button until the preset display begins to flash.
3. Press the MEMORY button to increment the memory location by one. Continue
to press the memory button until the desired location is displayed.
4. Press the program button to store the current configuration into the memory location that is displayed.
Recall a configuration from a memory location
1. Press the MEMORY button until the desired memory location is displayed in the memory display.
2. As you review the presets, the settings on the control panel will change to reflect the settings stored in the memory preset currently displayed.
3. When the desired preset is found, pause and the memory display will momentarily flash. The stored configuration is now loaded.
Notes: Once a configuration has been loaded, it will remain active until any change is made on the
control panel. If for instance the user adjusts the time, the current configuration will no longer match the previously loaded configuration and the memory display will change to a “-” dash. The previous configuration may be recalled using the procedure described above.
Memory locations may be overwritten by selecting them during the store procedure.
Digital Cycle Counter
This centrifuge is equipped with a digital cycle counter to monitor machine usage and for routine maintenance.
To display the current cycle count, perform the following:
1. Unlock the lid by pressing the OPEN/STOP button.
2. Open the lid latch so that the yellow LATCHED indicator light is not illuminated.
3. Press and hold the START button for approximately ve seconds. The word “CYCLES” will
be displayed in the speed display and the current cycle count will be displayed in the time display. If the current cycle count exceeds 9,999 cycles, the cycle count will be displayed
by showing the rst part of the number and then the end of the number. For example: the
1st shown is 0002 and the 2nd shown is 4537 represents 24,537 cycles.
12
Care and Preventative Maintenance
With proper care and maintenance your Horizon centrifuge will provide years of laboratory service. For
proper care, the following steps should be taken:
1. Provide Adequate Ventilation: For cooling purposes, the Horizon draws in ambient air through the
air intake cover on the top of the lid and exhausts this air in the rear of the base. The centrifuge should be placed on a hard smooth surface for good air circulation.
2. Always Spin Balanced Loads: This centrifuge is equipped with imbalance detection. However, it is
still possible to run loads that are minimally imbalanced, particularly at an imbalance setting of LOW SENSITIVITY. For maximum centrifuge life, set the imbalance detection to HI SENSITIVITY and always run balanced loads. Refer to pages 8 and 9 on for additional information on balancing the load.
3. Keep the Tube Carriers / Holders Clean: NOTE: Always follow the safety guidelines of your laboratory
to properly clean up and/or dispose of materials in the event that a substance known to be potentially toxic, radioactive or contaminated with a pathogenic microorganism is spilt in or on the centrifuge. Small glass fragments left in the tube carrier / holder after a tube breakage may adhere to the next test tube inserted in that carrier / holder. When this tube is handled, these fragments may puncture protective gloves and lacerate the operator’s fingers or hand. Remaining fragments may provide stress points on subsequent tubes and result in additional breakage. If a tube breakage occurs, carefully remove the tube holder. Properly dispose of the sample and tube fragments and thoroughly clean both the inside and outside of the tube carrier / holder. Insert a new tube cushion (if necessary) and replace the tube carrier / holder in the rotor.
4. Motor and Electrical Maintenance: This centrifuge uses a brushless-DC motor. There are no
brushes to replace and it should not need routine servicing for the life of the centrifuge. The electrical components are selected for high reliability and should not need service.
5. Keep the Centrifuge Clean: The cabinet, rotor top and accessories shall be thoroughly cleaned using
either isopropyl alcohol, soap and water, or a mild bleach solution. The use of Fully/Partially Halogenated
Hydrocarbons, Ketones, Esters and all other chemicals not prescribed by the manufacturer may cause
damage to the rotor and tube carriers / holders and shall not be used.
If it is necessary to remove the rotor for additional cleaning it is required that a qualified technician remove
the outside housing and rotor assembly. Contact The Drucker Company for additional information.
Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge in water or other cleaning solutions as this will cause damage and void your warranty!
6. Tube Carrier / Holder Replacement: It is recommended that the tube holders be replaced after 24
months of use. Inspect tube holders regularly for cracks. If cracks are discovered, replace immediately.
13
Troubleshooting:
1. Problem: The rotor does not spin freely.
Solutions:
Make sure that nothing has fallen into the rotor chamber.
If there is nothing obstructing the rotor, the motor may be damaged.
Contact your authorized dealer or The Drucker Company for further assistance.
2. Problem: There is excessive noise when the machine is running.
Solutions:
Check to see that the load is balanced.
Make sure that nothing has fallen into the rotor chamber.
Make sure that the screw in the center of the rotor is snug.
The motor may be damaged.
Contact your authorized dealer or The Drucker Company for further assistance.
3. Problem: The centrifuge does not run or an error message is displayed.
Solutions:
Make sure that the centrifuge is getting power. Does the control panel come on? Check the electrical outlet that the unit is plugged into. Check the circuit breaker on the base near the rear of
the centrifuge.
If the unit is getting power, make sure that the lid latch is closed properly. The latch is closed properly when the yellow ‘LATCHED’ indicator lightis illuminated.
If the centrifuge stops soon after start-up and ‘ERROR’ is displayed on the speed display, refer to the following for fault information:
14
ERROR / SPEED
The centrifuge cannot reach full speed due to a problem with the rotor, an inadequate power supply, or other electrical problems. Press the OPEN/ STOP button to cancel the error and then check the rotor
and the line voltage.Contact your authorized dealer or The Drucker Company for further assistance.
ERROR / BALANCE
The centrifuge has detected an imbalance. Press the OPEN/STOP button to cancel the error and balance the load. If the load is balanced, make sure that the centrifuge is installed on a secure, level location.
4. Problem: The lid knob cannot be turned / the lid cannot be unlocked.
Solutions:
Make sure the centrifuge has power and that the rotor is stopped. Press the ‘OPEN/STOP’ button. The red ‘UNLOCKED’ indicator light should illuminate and the unlocked mechanism should
disengage, allowing entry into the lid.
If the lid is still locked, make sure that the red ‘UNLOCKED’ indicator is illuminated and turn the lid
knob first completely clockwise and then counterclockwise.
If the red ‘UNLOCKED’ indicator will not illuminate or the locking mechanism will not disengage,
the electronics or locking mechanism may be damaged. Have a technician service the centrifuge
or contact The Drucker Company for further assistance.
For additional assistance, contact technical support Monday-Friday, 8:30 AM - 5:00 PM, EST
Providing Quality Centrifuges since 1932
200 Shadylane • Philipsburg, PA 16866 U.S.A. Phone: 814-342-6205 or 814-692-7661 • Fax: 814-692-7662 • www.druckercompany.com
Made in the USA by
The Horizon model 853 complies with all requirements of UL’s Electrical Equipment for Measurement, Control, and Laboratory Use; Part 1: General Requirements, UL 61010-1, Issued: 7/12/2004, Ed: 2nd, Rev: 10/28/08; Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use Part 1: General Requirements, CSA/C22.2 #61010-1, Issued: 7/12/2004, Ed: 2nd
Safety:
Horizon Lid Safety Switch: The Horizon lid is secured to the top of the cabinet by a latching knob
and pawl system. When the knob is rotated clockwise, the pawl grips the underside of the cabinet opening and prevents the lid from opening. A mechanical stop positions the pawl and prevents it from rotating completely. When rotated to the stop position, the pawl makes contact with a micro–switch mounted underneath the cabinet top.The lid safety switch prevents the centrifuge from operating
while the lid is open. The yellow ‘LATCHED’ indicator light on the front of the machine will illuminate
when the lid has been latched properly.
Horizon Lid Safety Interlock System: In addition to the Lid Safety Switch, the Horizon has a true
“0 RPM” lid locking system. The lid safety interlock system keeps the lid locked at all times, (even during power failure), and requires that the rotor be at rest in order to unlock the lid. The centrifuge will not allow entry into the rotor chamber unless the centrifuge has power and the rotor is stopped. To open the lid, make sure that the centrifuge is plugged in and, with the rotor stopped, press the
‘OPEN / EMERGENCY STOP’ button.
NOTE: After the centrifuge has started spinning, it may be possible to rotate the lid knob enough to
cause the pawl to lose contact with the lid safety switch. If this happens, the centrifuge motor may lose power, but the lid will still remain locked. If the knob is accidentally moved and this situation should occur, rotate the knob fully clockwise to its stop position and the centrifuge will resume operation.
Circuit Breaker: The Horizon is protected with a 4 Amp circuit breaker located at the rear of
the machine mounted to the base. Any electrical short circuit will cause the breaker to cut power to the machine.
Emergency Rotor Chamber Entry:
In the event of power failure, it may be impossible to unlock the lid by conventional means. In this case, entry into the rotor chamber may be made by removing the latch label and using a pen to manually disengage the locking mechanism (see photo). Pull the mechanism towards the control panel and then unlatch and open the lid. If the unit is damaged, contact your authorized
dealer or The Drucker Company.
Calibration and Earth Ground Testing:
It is recommended that the top speed, ground continuity and line leakage be tested every two years for
continued safe operation. Contact The Drucker Company for further information or testing availability.
15
Replacement Parts:
Part No. Description
7724177 ................. Foot, rubber
7751068 ................. Lid Tray Micro-Switch
7723002 ................. Lid Tray Solenoid
7735016 ................. Motor, 1/2 H.P. Brushless DC
7717041 ................. PC Control Board
3056002 ................. 853VES Transformer, 115V-16V
3056003 ................. 853VES Transformer, 220V-16V
7751043 ................. Circuit Breaker
7724044 ................. Front Panel Label
7760003 ................. Line cord
7714101 ................. Pawl, latch, lid
7714103 ................. Knob, latch, lid
7712263 ................. Lid
7724071 ................. Hinge, friction
7732018 ................. Gasket, cabinet opening
7732019 ................. Gasket, rotor chamber
12-Place Horizontal Rotor p/n 7786024
853VES Performance Plus Horizontal Rotor p/n 7786010
6-Place 50 mL Fixed-Angle Rotor p/n 7786044
20-Place Fixed-Angle Rotor p/n 7786042
75 mm Tube Holder p/n 7713043
Tube Holder Cap
p/n 7713011
Contact your authorized dealer or The Drucker Company
Providing Quality Centrifuges since 1932
100 mm Tube Holder p/n 7713040
50 mL Tube Holder Fixed-angle p/n 7787051
to order replacement parts or accessories.
Made in the USA by
125 mm Tube Holder p/n 7713044
1˝ Cushion Spacer
p/n 1525
16
200 Shadylane Drive • Philipsburg, PA 16866
Phone: 814-342-6205 or 814-692-7661 • Fax: 814-692-7662 • www.druckercompany.com
Loading...