Drucker Diagnostics 842VES User Manual

p/n 7711011
Rev. A
Operator/Service Manual
Model 842VES • Laboratory Centrifuge
Table of Contents
Model Description page 3
Supplied Equipment page 3
Warranty Information page 3
Specications page 4
Control Panel / Parts of the Centrifuge page 5
Setup Location page 6
Initial Setup Procedure pages 6,7
Operation page 7
Tube Holder Congurations Performance Plus Horizontal Rotor page 7 6-Place Horizontal Rotor page 8 6-Place Fixed Angle Rotor page 9
Advanced User Settings Acceleration Rate page 10 Braking Rate page 11 Countdown Delay page 11 Imbalance Detection page 11 Audible Setting page 11
Memory Locations / Storing and Recalling page 12
Digital Cycle Counter page 12
Care and Preventative Maintenance page 13
Rotor Removal and Installation page 14
Troubleshooting page 15
Safety page 16
Emergency Rotor Chamber Entry page 16
Calibration and Earth Ground Testing page 16
Replacement Parts page 17
page 2
WARNING: For the safety of both the operator and service personnel, care should be taken when handling substances that are known to be toxic, radioactive or contaminated with pathogenic microorganisms when using this centrifuge. When Risk Group II materials are used (as identied in the World Health Organization “Laboratory Bio-Safety Manual”), a Bio-Seal should be employed. The rotor and rotor accessories should not be considered as Bio-Seals. More than one level of protection must be provided in the case of materials of a higher group. The use of ammable or explosive materials as well as those materials which chemically react vigorously is prohibited.
Model Description
The Horizon Model 842VES is a continuous-duty, digitally-controlled centrifuge designed for the separation of uid samples. The 842VES oers control over almost every aspect of the unit’s operation from variable acceleration and deceleration to easy time and speed input. High speed and force capability produces platelet-poor/ free plasma and greatly reduce processing times. The exclusive Drucker horizontal rotor allows for quick and easy sample loading and complete horizontal separation. Save settings in one of 10 memory locations and recall them at the touch of a button.
Intended Use
General purpose laboratory centrifuge for sample seperation.
Supplied Accessories (Standard)
See the next
page for a
description of
optional
One (1) 6-Place Horizontal Rotor
p/n 7786021
Six (6)
100mm Tube Holders
p/n 7713040
Six (6)
75mm Tube Holders
p/n 7713043
accessories.
Also included (not shown):
• One (1) 10’ Line Cord • One (1) Operator’s Manual • 5/32” Allen Wrench
WARRANTY:
The Drucker Company warranties that this centrifuge is free from
defects in workmanship and parts for 2 years.
Made in the USA by
200 Shadylane Drive • Philipsburg, PA 16866
Phone: 814-342-6205 or 814-692-7661 • Fax: 814-692-7662 • www.druckercompany.com
page 3
Specications:
General Specications for the Horizon Model 842VES Centrifuge
Overall Dimensions (H x W x D): 8.0 in. x 11.0 in. x 13.5 in. Centrifuge Motor: 1/2 H.P. Brushless DC Protection Breaker: 4 Amp. re–settable Timer: electronic, with hold
or 0 to 99 minutes, +/- 1%
Weight: 23 lbs. Permitted Environmental Conditions Ambient Temperature During Operation: 2 oC - 40 oC Maximum Relative Air Humidity: 90% Electrical Requirements: Power: 200 Watts Voltage: North America: 115 Volts (+/- 10%)
Japan: 100 Volts (+/- 10%) Europe: 230 Volts (+/- 10%) Frequency: 50/60 Hz
Any use other than those specied by the Manufacturer is explicitly prohibited.
Maximum sample density is 1.15 grams / mL, (water density = 1.0 grams / mL)
PART NO. # MAX TUBE SIZE SPEED RANGE FORCE RANGE 6-Place Horizontal (Standard) 77860211 75mm tube holders 77130431 6 17mm x 75mm 1000 to 4,500 RPM 140 to 2,260 xg 100mm tube holders 77130401 6 17mm x 100mm 1000 to 4,500 RPM 140 to 2,800 xg
Refer to serial number label
on rear of centrifuge to
conrm voltage requirement.
6-Place Fixed-Angle Rotor 7786047
2
75mm tube holder 77130431 6 17mm x 75mm 1000 to 4,000 RPM 140 to 1,500 xg 100mm tube holder 77130401 6 17mm x 100mm 1000 to 4,000 RPM 140 to 1,730 xg 125mm tube holder 77130442 6 17mm x 125mm 1000 to 4,000 RPM 140 to 1,950 xg
6-Place Performance Plus Horizontal 7786017
3
75mm tube holder 77130563 6 13mm x 75mm 1000 to 6,500 RPM 140 to 5,000 xg
12-Place combo Fixed-Angle STAT 7786003
75mm tube locations n/a3 6 13mm x 75mm 1000 to 7,000 RPM 140 to 4,880 xg
3
100mm tube locations n/a3 6 13mm x 100mm 1000 to 7,000 RPM 140 to 5,700 xg 12-Place Fixed-Angle STAT 7786001
NOTE: Contact the Drucker Company for assistance when changing the installed rotor. The installed rotor may only be changed by qualied personnel.
13 x 75mm Performance Tube Holder p/n 7713056
3
75mm tube locations n/a3 12 13mm x 75mm 1000 to 7,000 RPM 140 to 4,880 xg
100 mm Tube Holder p/n 7713040
1
125 mm
Tube Holder 75 mm Tube Holder p/n 7713043
1
This rotor or rotor accessory is rated for a rotational speed of 5000 RPM and a force of 3000 xg.
2
This rotor or rotor accessory is rated for a rotational speed of 4500 RPM and a force of 2500 xg.
3
This rotor or rotor accessory is rated for a rotational speed of 7000 RPM and a force of 5500 xg.
1
p/n 7713044
3
Tube Holder Cap p/n 7713011
2
1” Cushion Spacer p/n 1525
1.5 to 2mL tube Adapter p/n 7713065
13 x 75mm insert, p/n 7713064
13 x 100 mm insert, p/n 7713066
.5 to 1 ml tube Adapter p/n 7713068
page 4
Control Panel / Parts of the Centrifuge:
Lid
Lid Knob
Lid Safety Interlock System
Control Panel (see below)
Cabinet
RCF Conversion Button
Start Button
Open / Stop Button
Time Display
Up/Down RPM Change Buttons
RPM Display
Up/Down Time Change Buttons
Running Indicator Light
Brake Display
Latched Indicator Light
Program Button
Unlocked Indicator Light
Memory Preset Display
Memory Button
page 5
Setup Location:
1. Unpack the centrifuge and verify that all of the supplied equipment is present.
2. Choose a setup location which meets the following criteria:
a) A bench top clearance height of 16” is required in order to open the lid.
b) The clearance envelope is the space around the centrifuge which is required for safety. Choose a setup location which will allow for a clearance envelope of at least 24” x 24”, (with the centrifuge at the center). No person or hazardous material shall be permitted in the clearance envelope during operation. The operator time within the envelope shall be limited to the time necessary for loading, unloading and centrifuge operation only.
c) Proper ventilation is necessary to prevent the overheating of samples as well as premature failure of the centrifuge. Choose an area which will allow unencumbered air ow.
d) The centrifuge is designed to rest on its four rubber feet. No adjustment is necessary for leveling the centrifuge, however, the surface should be at and level.
e) Be sure the outlet is always within reach as the line cord is the means of
emergency disconnection!
Initial Setup Procedure:
If any problems are found during the initial setup procedure, refer to the troubleshooting
section. For further assistance, contact The Drucker Company at 814-342-6205 or 814-342-6205.
1. Plug the female end of the line cord into the rear of the centrifuge. Plug the male end into an approved electrical outlet. For electrical safety, the unit must always be properly grounded.
2. For operator safety, the locking system is always active; requiring power and direction from the user to disengage it (the lid also automatically unlocks at the end of a run when it is safe to do so). To unlock the lid (in order to access the rotor chamber) press the ‘OPEN / STOP’ button on the control panel. The ‘UNLOCKED’ indicator light should illuminate. If it does not, refer to the section on troubleshooting. The lid will remain unlocked for 15 seconds after pressing the ‘OPEN / STOP’ button.
3. Turn the latch counterclockwise and open the lid.
4. Spin the rotor by hand; check for free and level rotation. If the rotor does not spin freely, refer to the section on troubleshooting.
5. Place the test tube holders inside the rotor and verify that they are seated properly.
6. Close the lid. Rotate the lid knob clockwise to its complete stop position. The ’LATCHED’ indicator light should be illuminated. If it is not, make sure that the lid is latched properly. The centrifuge will not run unless the lid is latched properly and the ’LATCH’ indicator light is illuminated.
7. Use the up arrow button next to the RPM display to set the speed to 4000 RPM.
8. Initiate a test run by pressing the ‘START’ button.
9. The ‘RUNNING’ indicator light will illuminate.
10. The test tube carriers will slide up into the horizontal position (horizontal rotor only) and the unit will accelerate to the current set speed.
11. Listen to the sound of the centrifuge. A smooth whirring sound should be heard. If there are any loud or unusual sounds, stop the centrifuge by pressing the ’OPEN / STOP’ button immediately and refer to the section on troubleshooting.
12. Press the ‘OPEN / STOP’ button to terminate the test run. The rotor will decelerate to a complete stop and the lid will then unlock automatically for sixty (60) seconds.
13. Take time now to familiarize yourself with the various additional user settings. Turn to page 11.
14. The centrifuge is now ready for operation.
page 6
Loading...
+ 11 hidden pages