Drucker Diagnostics 642VES User Manual

MODEL 642VES
Model 642VES Operator’s Manual
Laboratory Centrifuge
Providing Quality Centrifuges since 1932
P/N 7711072
Rev. C
Table of Contents
Providing Quality Centrifuges since 1932
Model Description .................................................................. 3
Intended Use ........................................................................... 3
Device Classification .............................................................. 3
Supplied Equipment ............................................................... 3
Warranty Information .............................................................3
Specifications ......................................................................... 4
Care and Preventative Maintenance .....................................5
Setup Location ........................................................................ 6
Initial Setup Procedure ..........................................................6
Transportation ........................................................................ 6
Operation ................................................................................ 7
Programming Example ..........................................................7
Executing a Program ..............................................................7
Tube Holder Configurations
Troubleshooting .................................................................... 10
Safety ..................................................................................... 11
Emergency Rotor Chamber Entry ....................................... 11
Calibration and Earth Ground Testing ................................ 11
Speed Verification ................................................................11
Replacement Parts ............................................................... 12
02
WARNING
Use universal precautions when handling laboratory specimens. All human specimens of blood, blood compounds, or bodily fluids are potentially biohazardous and may contain Hepatitis B virus, Hepatitis C virus, Human immunodeficiency virus or other disease causing agents. For the safety of both the operator and service personnel, care should be taken when handling substances that are known to be toxic, radioactive, or contaminated with pathogenic microorganisms when using this centrifuge. When Risk Group II materials are used (as identified in the World Health Organization “Laboratory Bio-Safety Manual”), the use of a Bio-Seal should be employed. More than one level of protection must be provided in the case of materials of a higher group. The use of flammable or explosive materials as well as those materials which chemically react vigorously is prohibited.
Model Description
The Horizon Model 642VES is a continuous-duty, electronically­controlled, variable-speed, laboratory centrifuge with a lid safety interlock system. The unit also incorporates a backlit display, making accessing and controlling centrifugation parameters easier than ever. The 642VES offers the user control over many parameters, including variable runtime, speed control, and deceleration. The exclusive Drucker horizontal rotor allows for quick and easy sample loading and complete horizontal separation. Up to ten commonly used settings can be saved and recalled at the touch of a button.
Intended Use
General purpose laboratory centrifuge for sample separation.
Device Classification
FDA Class One
Supplied Accessories (Standard)
See next page for optional rotor/accessories
*The rotor and rotor accessories are rated for a rotational speed of 5000 RPM.
One (1) 6-Carrier Horizontal Rotor
7786021*
Six (6) 100 mm Carriers
7713031*
Six (6) 75 mm Carriers
7713033*
Also included (not shown):
• One (1) 10´ Line Cord 7760006 (United States) 7760005 (Europe)
• One (1) Operator’s Manual 7711072
See the next page for a description of optional accessories.
WARRANTY:
The Drucker Company warranties that this centrifuge is free from
defects in workmanship and parts for 2 years.
Made in the USA by
Providing Quality Centrifuges since 1932
200 Shadylane Drive • Philipsburg, PA 16866
Phone: 814-342-6205 or 814-692-7661 • Fax: 814-692-7662 • www.druckercompany.com
03
Specifications
DESCRIPTION PART NO. MAX TUBE SIZE SPEED RANGE FORCE RANGE
6-Place Horizontal Rotor 7786021
This rotor or rotor accessory is rated for a rotational speed of 5000 rpm and a force of 3,475 × g
7713033 (Green) 7
713043 (Blue)
75 mm Tube Holders 17 mm x 75 mm 1,000 to 3,200 rpm 111 to 1,134 × g
7713031 (Red) 7713040 (Purple)
100 mm Tube Holders 17 mm x 100 mm 1,000 to 3,200 rpm 139 t o 1,425 × g
,
DESCRIPTION PART NO. MAX TUBE SIZE SPEED RANGE FORCE RANGE
6-Place Fixed-Angle Rotor 7786047
This rotor or rotor accessory is rated for a rotational speed of 4,500 rpm and a force of 2,475 × g
7713033 (Green) 7713043 (Blue)
75 mm Tube Holders 17 mm x 75 mm
1,000 to 3,900 rpm
88 to 1,339 × g
7713031 (Red) 7713040 (Purple)
100 mm Tube Holders 17 mm x 100 mm
1,000 to 3,900 rpm
105 to 1,598 × g
7713032 (Black)
125 mm Tube Holders 17 mm x 125 mm
1,000 to 3,900 rpm
122 to 1,857 × g
DESCRIPTION PART NO. MAX TUBE SIZE SPEED RANGE FORCE RANGE
6-Place Horizontal Rotor 7786021
This rotor or rotor accessory is rated for a rotational speed of 5000 rpm and a force of 3,475 × g
7713033 (Green) 7
713043 (Blue)
75 mm Tube Holders 17 mm x 75 mm 1,000 to 3,200 rpm 111 to 1,134 × g
7713031 (Red) 7713040 (Purple)
100 mm Tube Holders 17 mm x 100 mm 1,000 to 3,200 rpm 139 to 1,425 × g
,
General Specifications for the Horizon Model 642VES Centrifuge
Speed Range (Horizontal Rotor): ............. 1,000 - 3,200 rpm (+/- 100 rpm)
Speed Range (Fixed-angle Rotor): ........... 1,000 - 3,900 rpm (+/- 100 rpm)
Maximum Capacity (Horizontal Rotor): ... 60 mL (6 × 10 mL) Maximum Capacity (Fixed-angle Rotor): .. 90 mL (6 × 15 mL)
Operating Voltage Ranges: ....................... 103 - 127 VAC or 207 - 253 VAC
Overall Dimensions (H x W x D): ............... 9.0˝ × 11˝ × 12.5˝ (22 cm × 28 cm × 32 cm)
Centrifuge Motor: ...................................... 1/30 HP, PSC motor
Nominal Acceleration Time: ..................... 10 seconds
Protection Breaker: ................................... 4 Amp. re–settable
Timer: .......................................................... Electronic, with hold or 0 to 99 minutes, +/- 1%
Weight: ........................................................ 18.5 lbs. (8.4 kg)
Permitted Environmental Conditions
Ambient Temperature During Operation: 2 oC - 40 oC
Maximum Relative Air Humidity: .............. 90%
Electrical Requirements
Power (Watts): ............................................ 200
Voltage (Volts): ............................................ 115 VAC (+/- 10), 230 VAC (+/- 20)*
*Check Serial Number Badge for Your Model Specific Voltage Requirement.
Frequency (Hz): ........................................... 50/60
PART NO. # MAX TUBE SIZE SPEED RANGE FORCE RANGE
6-Place Horizontal Rotor
7786021 75 mm tube holder 7713033 6 17 mm x 75 mm 1,000 to 3,200 RPM 111 to 1,134 xg 100 mm tube holder 7713031 6 17 mm x 100 mm 1,000 to 3,200 RPM 139 to 1,425 xg
This rotor or rotor accessory is rated for a rotational speed of 5000 rpm and a force of 3,475 xg
6-Place Fixed-angle Rotor
7786047 75 mm tube holder 7713033 6 17 mm x 75 mm 1,000 to 3,900 RPM 88 to 1,339 xg 100 mm tube holder 7713031 6 17 mm x 100 mm 1,000 to 3,900 RPM 105 to 1,598 xg 125 mm tube holder 7713032 6 17 mm x 125 mm 1,000 to 3,900 RPM 122 to 1,857 xg
This rotor or rotor accessory is rated for a rotational speed of 4,500 rpm and a force of 2,475 xg
Contact your authorized dealer or The Drucker Co. to order accessories.
75 mm Tube Holder
p/n 7713033
100 mm
Tube Holder
p/n 7713031
125 mm
Tube Holder
p/n 7713032
Tube Holder Cap
p/n 7713011
1˝ Cushion Spacer
p/n 1525
04
Any use other than those specified by the Manufacturer is explicitly prohibited. Maximum sample density is 1.15 grams / mL, (water density = 1.0 grams / mL)
Loading...
+ 8 hidden pages