Drucker 642VES, 642VFD, 642VFD-Plus Operator's Manual

Operator’s Manual
Model 642VES, 642VFD & 642VFD-Plus • Laboratory Centrifuge
P/N 03-0-0002-0096 Rev. B
(642VES Shown)
Page 2
Model Description:
The centrifuge is controlled in the following manner depending on the model:
642VES: Controlled by five electronic push buttons and a backlit display, for easily controlling
centrifugation parameters. Up to nine commonly used settings can be saved and recalled at the touch of a button.
642VFD-Plus:Controlled by two electronic push-button timers that have been preset for ten (10)
minutes at 3,800 RPM and five (5) minutes at 1,800 RPM, for precise spin times and ease of use.
642VFD: Controlled by one electronic push-button timer that has been preset for ten (10)
minutes at 3,800 RPM, for precise spin times and ease of use.
WARNING: Use universal precautions when handling laboratory specimens. All human specimens of blood, blood compounds, or
bodily uids are potentially biohazardous and
may contain Hepatitis B virus, Hepatitis C
virus, Human immunodeciency virus or other
disease causing agents. For the safety of both the operator and service personnel, care should be taken when handling substances that are known to be toxic, radioactive, or contaminated with pathogenic microorganisms when using this centrifuge. When Risk Group II materials
are used (as identied in the World Health Organization “Laboratory Bio-Safety Manual”) the use of a Bio-Seal should be employed. More
than one level of protection must be provided in the case of materials of a higher group. The use
of ammable or explosive materials as well as
those materials which chemically react vigorously is prohibited. Any use of the equipment in a
manner not specied in these instructions may
impair the level of protection provided by the equipent.
Table of Contents
Model Description ...........................pg. 2
Supplied Equipment ........................pg. 2
Features ........................................... pg. 3
Specifications ..................................pg. 3
Setup Location and Procedure ........pg. 4
Front Panel Controls ........................pg. 5
Operation ......................................... pg. 7
Rotor Removal and Installation .......pg. 7
Care and Maintenance ....................pg. 8
Cleaning and Disinfecting ................ pg. 8
Troubleshooting ...............................pg. 9
Safety ............................................... pg. 10
Calibration and Ground Testing .......pg. 10
Replacement Parts ..........................pg. 11
Available Accessories ...................... pg. 11
Supplied Equipment*:
The following items come standard with each Horizon 642VES, 642VFD, and 642VFD-Plus
centrifuge:
1. One (1) Six-Carrier Horizontal Rotor
2. Six (6) 75mm-100mm Carriers
3. One (1) Operator’s Manual
4. One (1) Line Cord
Intended Use
General purpose laboratory centrifuge for sample separation.
The use of any line cord other that what is supplied by the manufacturer may not carry an adequate rating and is therefore prohibited.
*The rotor and rotor accessories are rated for rotational frequency of 5,500 RPM.
p/n 7786067* p/n 7713079*
p/n 03-0-0002-0096
p/n 7760006 (North America) p/n 7760005 (Europe)
Page 3
Features:
Swing–out horizontal rotor design, incorporating a unique test tube holder that produces horizontally
separated samples while requiring no additional parts
• Cool–Flow air ow design that prevents overheating of samples
• Lid safety switch that prevents the centrifuge from operating unless the lid is closed and latched
• Removable rotor for easy cleaning
• Locking lid that allows entry into the centrifuge only after the rotor has completely stopped
• Clear lid for safe observation of samples and optical calibration of speed
• Electronically controlled timed operation
• Push-button operation
• Indicator Lights
‘RUNNING’ Green – lights when power is applied to motor ‘RUN 1’ VFD-Plus only Green – lights when power is applied to motor ‘RUN 2’ VFD-Plus only Green – lights when power is applied to motor ‘LATCHED’ Yellow – lights when lid is closed and latched ‘UNLOCKED’ Red – lights when the locking system is deactivated
Specications 642VFD, 642VFD-Plus, and 642VES:
Speed Range:
(Horizontal) 1,000 - 3,800 (+/- 100) RPM (Fixed Angle) 1,000 – 3,900(+/- 100) RPM
RCF Range:
(Horizontal) 114 – 2016 (+/- 90) xg (Fixed Angle) 90 – 1860 (+/- 90) xg Maximum Capacity: 60 mL (6 × 10 mL)* Overall Dimensions (H x W x D): 8.75 in. × 11.75 in. × 14 in.
(22 cm x 30 cm x 35 cm) Weight: 12 lbs. (5.4 kg)
Centrifuge Motor: 1/30 HP, PSC motor Nominal Acceleration Time: 20 seconds Protection: 4 Amp, resettable circuit breaker (x2) Timer (642 VES): Electronic, 1 to 99 minutes, preset
to 10 minutes +/- 1% Timer (642 VFD, 642VFD-Plus): Electronic, 1 to 30 minutes, preset to 10 minutes +/- 1%
Power Rating: 200 Watts Voltage Requirements: 115/ 230 VAC (+/- 10%)
Frequency: 50/60 Hz
Permitted Environmental Conditions
Ambient Temperature During Operation: 60°F - 90°F (16°C - 32°C)
Maximum Relative Air Humidity: 90%
* Maximum sample density is 1.15 grams / mL, (water density = 1.0 grams / mL)
Any use other than those specied by the Manufacturer is explicitly prohibited.
Page 4
Setup Location:
1. Unpack the centrifuge and verify that all of the supplied equipment is present.
2. Choose a setup location which meets the following criteria:
a) A bench top clearance height of 20” is required in order to open the lid.
b) The clearance envelope is the space around the centrifuge which is required for safety. Choose a setup location which will allow for a clearance envelope of at least 24” x 24”, (with the centrifuge at the center). Do not position the centrifuge in such a way that it would be difcult to disconnect the power in the case of emergency. No person or hazardous material shall be permitted in the clearance envelope during operation. The
operator time within the envelope shall be limited to the time necessary for loading, unloading and centrifuge operation only.
c) Proper ventilation is necessary to prevent the overheating of samples as well as
premature failure of the centrifuge. Choose an area which will allow unencumbered air
ow.
d) The centrifuge is designed to secure to the operating surface by four suction feet. No adjustment is necessary for leveling the centrifuge, however, the surface should be at
and level.
e) Be sure the outlet is always within reach as the line cord is the means of emergency disconnection!
Initial Setup Procedure:
If any problems are found during the initial setup procedure, refer to the troubleshooting section on page 9.
1. Plug the female end of the supplied line cord into the power input module located on the rear of the centrifuge. Plug the male end into an approved electrical outlet. For electrical safety, the unit must always be properly grounded.
2. Flip the switch on the power input module located on the rear ofthe
centrifuge to the ON ( | ) position.
3. For safety purposes, the locking system is always activated. To
deactivate the system, (in order to insert or retrieve samples), press the OPEN/STOP button on the
control panel. The UNLOCKED indicator light should illuminate. If it does not, refer to page 9 on troubleshooting. The lid will be unlocked for 15 seconds after pushing the OPEN/STOP button.
4. Turn the latch counterclockwise and open the lid.
5. Spin the rotor by hand; check for free and level rotation. If the rotor does not spin freely,
refer to page 9 on troubleshooting.
6. Place the six test tube holders inside the rotor (as shown to the right), and verify that
they are seated properly.
7. Close the lid. Rotate the lid knob clockwise to its complete stop position. The ’LATCHED’
indicator light should be illuminated. If it is not, make sure that the lid is latched properly.
The centrifuge will not run unless the lid is latched and the ‘LATCHED’ light is
illuminated.
8. Start a centrifugation cycle by pushing the START (BLOOD) button.
9. The ‘RUNNING’ indicator light will illuminate.
10.The test tube holders will slide up into the horizontal position and the unit will accelerate
to full speed.
11.Listen to the sound of the centrifuge. A smooth whirring sound should be heard. If
there are any loud or unusual sounds, stop the centrifuge by pushing the OPEN/STOP
button immediately and refer to page 9 on troubleshooting.
12.Push the OPEN/STOP button. The ‘RUNNING’ indicator light should go out and the
motor should slow to a stop.
12.The lid should remain locked until the rotor has nearly stopped. If the machine unlocks
prematurely, contact Drucker Diagnostics for assistance. Once the rotor has stopped, the interlock system will become disengaged for sixty (60) seconds. The ‘UNLOCKED’
indicator light will illuminate during this time.
13.To gain entry into the centrifuge after this period has ended, simply press the OPEN/
STOP button. The lid will unlock for fteen (15) additional seconds.
After the centrifuge has passed this procedure it is ready for operation.
Horizontal Rotor Shown
Loading...
+ 8 hidden pages