DRTECH EVS3643 User Manual

DRTECH EVS 3643 System
Safety and Regulatory Information
with User’s Manual
EVS 3643 User’s Manual To Customers
Thank you for purchasing the DRTECH Radiography EVS 3643 (hereinafter this Product). This User’s Manual
be sure to read this manual thoroughly in order to utilize it more effectively. Also, read the Operation Manual for
EVS Calibration and configuration Software (hereinafter ECali1).
Important information on usage and management of equipment
1. Only a physician or legally certified operator should use this product.
2. The equipment should be maintained in a safe and operable condition by maintenace personal.
3. Use only computers ans image display monitors complying with IEC 60601-1 or IEC 60950-1 and under asystem configuration complying with IEC 60601-1-1. For details, consult your sales representative or local DRTECH dealer.
4. Use only the dedicated cables. Do not use any cables other than those supplied with this product.
Disclaimer
1. In no event shall DRTECH be liable for any damage or loss arising from fire, earthquake, any action or accident by a third party, any intentional negligent action by users, any trial usage, or other usage under abnormal conditions.
2. Roentgenography, image processing, image reading, and image data storage must be performe in accordance with the laws of the country or region in which the product is being used. The user is resposible for manufacturing the privacy of image data.
3. In no event shall DRTECH be liable for personal physical harm or property damage that is sustained when the instructions are not followed or the product is misused.
4. It is the resposibility of the attending physicians to provide medical care services. DRTECH will not be liable for faulty diagnoses.
5. In no event shall DRTECH be liable for direct or indirect cosequential damages arising from the use or unavailbility of this product. DRTECH shall not be liable for loss of image data for any reason.
6. In no event shall DRTECH be liable for any damage arising from moving, alteration, inspection or repair by a person other than authorized service engineers.
7. Specifications, composition, and appearance of this product may change without prior notice.
10-EVS1MADB002 2
EVS 3643 User’s Manual Contents
To Cust omers…………………………………………………………….……………………………… 2 Contents……………………………………………………………………………………………..…… 3 Conventions………………………………………………………………….……………………..…… 5
1. Safety nformation……………….…………………………………………….…….6
1-1. Safety precautions……………………………………………………………………………..6 1-2. Notes for using the equipment……………………………………………………..……..….10
2. Introduction…………………………………………………………………...……..12
2-1. Features……………………………………………………………….……………………….12 2-2. Intended use…………………………………………………………. ………..…………..….12 2-3. System configuration………………………………………………………………..………..13
3. Product description………………………………………………………………..14
3-1. Product components………………………………………….…………..….…….…..……..15 3-2. X-ray imaging condition…….………………….……………………………..…..…………16
4. Parts name and functions………………………………………………………..17
4-1. Detector specification………………….……………………………………….…..17 4-2. Detector component………………….……………………………………….…..18 4-3. Detector dimension………………….……………………………………….…..19 4-4. SSU (System syncronization unit)……………………………………..………….………20 4-5. Wireless acess point………………………………………………………………………..20
5. Operating pocedure…….………………………………………………………..22
5-1. Preparing to use the detector……………………….…………………………………….…..23 5-2. Operating the detector…….………………………………………………..………….………23 5-3. Ending use of the detector…………………………………………………………………….27 5-4. Detector status list………………………………………………………………………………27
10-EVS1MADB001 5
EVS 3643User’s Manual 5. Operating procedure
6. Extension facility….…….………………………………………………………..28
6-1. Using the wirless detector with other EXPRIMER series detectors in a same system …28 6-2. Sharing the wireless detector among different systems………………..………….………29 6-3. AED mode……………………………………….………………………………………………30 6-4. Wiring connections……………………………………….………………………………….31
7. How to Install………………………..…….….…….………………………………32
8. Troubleshooting….……….….…….………………………………………………89
9. Maintenace ………….….…….……………………………………………………..90
10. Specification…………………………………………………………………………91
10-1.Main specifications……………………………………………………………………………91 EVS 3643 Detector…………………………………………………………………………….91 SSU(system syncronization unit)…………………………………………………………….92 10-2. Chrateristicss………………………………………………………………………………….93
11. Regulatory information………….….….…………………………………………94
11-1. Medical equipment safety standards……………………………………………………….94
Medical equipment classification………………………………………………………….94
Product safety standards…………………………………………………………………..94
11-2. Radio frequency (RF) compliance information………………………………………………95 Declaration comforty………………………………………………………………………….95 11-3. Labels and markings on the equipment………………………………………………………99
Detector………………………………………………………………………………………99
Other components of the EVS 3643 system……………………………………..101
11-4. Guidance and manufacturer’s declaration for EMC………………………………………103
Electromagnetic emissions……………………………………………………………………103 Electromagnetic immunity…………………………………………………………………103 Electromagnetic separation distances………………………………………………………103
12. Warranty ………….….…….………………………………………………………..106
10-EVS1MADB001 4
EVS 3643 User’s Manual
Contents
Safety notices
The following safety notices are used to emphasize certain safety instructions. Follow the safety instructions in this user’s maual along with warning and cautions sysbols. Ignoring such warnings or cautions while handling the product may results in serious injury or accient. It is important for you to read and understand the contents of this users manual before attemting to use the product
WARNING
CAUTION
CAUTION
Prohibited
.
This notice used to idendifiy conditions under which improper use of the product may cause death or serious personal injury.
This notice used to idendifi\y conditions under which improper use of the product may cause minor personal injury.
This notice used to idendifiy conditions under which improper use of the product may cause property damage.
This is used to indicate a prohibited operation.
This is used to indicate an action that must be performed.
This is used to indicate important operations and restrictions.
IMPORTANT
Be sure to read this notice to prevent property damage or malfunction.
This is used to indicate operations for reference and complementary information. User are recommended to read this notice
10-EVS1MADB001 5
EVS 3643 User’s Manual
1. Safety information
1-1. Safety precautions
Follow these safeguards and properly use the equipment to prevent injury and damage to any equipment/data.
WARNING
Installation and environment of use
- Do not use or stroe the equipment near flammable chemical such as alcohol, thinner,
benzine, etc.
Prohibited
If chemicals are spilled or evaporate, it may result in fire or electric shock through contact with electric parts inside the equipment. Also, somedisinfectants are flammable. Be sure to take care when using them.
- Do not connect the equipment with anything other than specified. Doing so may result in fire or electric shock.
Power supply
- Do not use operate the equipment using any type of power supply other than the one
indicated on the rating label.
Otherwise, it may result in fire or electric shock.
Prohibited
10-EVS1MADB001 6
- Do not handle the equipment with wet hands. You may experience an electric shock that could result in death or serious injury.
- Do not place heavy object such as medical equipment on cables and cords, or do not
pull, bend, bundle, or step on them to prevent their sheath from being damaged, and do not alter them neither.
Doing so may damage the cords which could result in fire or electric shock.
- Do not supply power to more than one of equipment using the same AC outlet. Doing so may result in fire or electric shock.
- Do not turn on the system power when condensation has formed on the equipment. Doing so may result in fire or electric shock.
- Do not connect a multiple portable socket-outlet or extension cord to the system. Doing so may result in fire or electric shock.
- Securely plug the power cord into the AC outlet.
If contact failure occurs, or if dust or metal objects come into contact with the exposed metal prong of the plug, fire or electric shock may result.
- Be sure to turn OFF the power to each of equipment before connecting or
disconnecting the cords.
Otherwise, you may get an electric sock that could result in death or serious injury.
- Be sure to hold the plug or connector to disconnect the cord.
If you pull the cord, the core wire may be damaged, resulting in fire or electric shock.
EVS 3643 User’s Manual
1. Safety information
WARNING
Handling
The system, in whole or in part, cannot be modified in any way without written approval from DRTECH.
- No modification of this equipment is allowed
- Never disassemble or modify the equipment. Doing so may result in fire or electric shock. Also, since the equipment incorporates parts
Prohibited
that may cause electric shcok as well as other hazardous parts, touching them may cause death or serious injury.
- Do not place anything on top of the equipment. The object may fall and cause an injury. Also if metal objects such as needles or clips fall into he equipment, or if liquid isspilled, if may result in fire or electric shock.
- Do not hit or drop the equipment. The may be damaged if it recieves a strong jolt, which may result in fire or electric shock if the equipment is used without geing repaired.
- Have the patient take a sixed posture and do not let patient touch parts unnecessarily. If the patient touched connectors or switches, it may result in electric shock or malfunction of the equipment.
When a problem occurs
- Should any of the following occur, immediately turn OFF the power to each piece of
equipment, unplug the power cord from the AC outlet, and contact your sales representative or local DRTECH dealer:
When there is smoke, an odd smell or abnormal sound When liquid has been spilled into equipment or metal object has entered through an opening. When the equipment has been droppen and is damaged.
Maintenance and inspection
- When the equipment is going to be cleaned, be sure to turn OFF the power of each
equipment, and unpulg the power cord from the AC outlet. Never use alcohol,
Prohibited
benzine, thinner or any other flammable cleaning agent.
Othewise, it may result in fire or electric shock
- Clean the plug of the power cord periodically by unplugging it from the AC outlet and
removing dust or dirt from the plug, its periphery and AC oulet with a dry cloth.
If the cord is kept plugged in for a long time in a dusty, humid or sootyplace, dust around the
plug will attract moisture, and this could cause insulation failure that could result in a fire.
- For safety reasons, be sure to turn OFF the power to each piece of equipment when
the insections indicated in this manual are going to be performed..
Otherwise, electric shocks may occur.
10-EVS1MADB001 7
EVS 3643 User’s Manual
1. Safety information
Installation and environment of use
- Do not install the equipment in any of the locations listed below. Doing so may result
in failure or malfuction, equipment faillng, or fire or injury.
Close to facilities where water is used where it may be exposed to direct sunlight Close to the air outlet of an air-conditioner or ventilation equipment Close to a eat source such as a heater where the power supply is unstable In a saline or sulfurous environment where temperature or humidity is high where there is freezing or condensation In area prone to vibration On an incline or in an unstable area
- Because the equipment cable is long, take care that cables do not become tangled during use. Also, be careful not to get your feet caught in the cable.
Otherwise, it may cause a malfunction of the equipment or the injury of the user due to tripping over the cable.
- Non-medical equipment such as the battery charger, and access point unit cannot be
used in patient’s vicinity.
Patient Vicinity
WARNING
Power supply
- Always connectthe three-core power cord plug to a grounded AC power outlet.
- To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth”.
- To make it easy to disconnect the plug at any time, avoid putting any obstacles near
the outlet. Otherwise, it may not be possible to disconnect the plug in an emergency.
- Be sure to ground the equipment to an indoor grounded connector. Also, be sure to connect all the grounded for the system to a common ground.
- Do not use any power source other than the one provided with this equipment. Otherwise, fire or electric shock may be caused due to leakage
Handling
- Do not spill liquid or chemicals onto the equipment or, in cases where the patient is
injured, allow it to come in contact with blood or body fluids.
Doing so may result in fire or electric shock.
10-EVS1MADB001 8
In such a situation, protect the equipment with a disposable covering as necessary.
- Turn OFF the power to each piece of equipment for safety when not belong used.
EVS 3643 User’s Manual
1. Safety information
CAUTION
Handling the equipment
The Equipment must be handled with care to avoid personal injury or damage to the internal image sensor.
- Handle the equipment carefully.
- Do not submerge the equipment in water
- The internal image sensor may be damaged if something hits against it, or if it is
dropped, or recieves a strong jolt.
- Do not place excessive weight on the detector. Otherwise, the internal image sensor may be damaged.
<Load Limit>
Uniform load: 150 kg over the whole area of the detector
Local load: 100 kg on an area 40 mm in dimeter
- Be sure to use the detector on a flat surface do it will not bend. Otherwise, the
internal image sensor may be damaged. Be sure to securely hold the detector while using it in upright positions. Otherwise, the detector may fall over, resulting in injury to the user or patient, or may flip over, resulting in damage to the inner device.
10-EVS1MADB001 9
EVS 3643 User’s Manual
1. Safety information
1-2. Notes for using the equipment
When using the equipment, take the following precautions. Otherwise, problems may occur and the equipment may not function correctly
System Diagnostic
The Ecali1 software runs a system diagnostic. Run Ecali1 software after installing the system and at least once a year. If an error occurs, report the detailed error information to DRTECH local dealer or distributor
The owner is responsible for ensuring that the system diagnostic is
CAUTION
Calibration
To ensure optimal performance of the system, it is important to verify that the system is calibrated.
performed every year. Do not try to use the system if the system diagnostic is failed.
The owner is responsible for ensuring that the system calibration is
CAUTION
performed after the system installation is completed or the system is repaired. Do not try to use the system if system calibration is not performed.
Before exposure
- Be sure to check the equipment daily and confirm that it works properly.
- Sudden heating the room in clod area will cause condensation to form on the equipment. In this case, wait until the condensation evaproates before performing an exposure. If the equipment is used while the condensation formed on it, problems may occur in the quality of captureed image. When an air-conditioner is used, be sure to raise/lower the temperature gradually so that a difference in temperature in the room and in the equipment does not occur, to prevent condensation.
During exposure
- Do not use the selected frequency chanel for other wireless device. Mutual interference may affect the image data transmission rate.
- Do not use the detector near devices generating a strong magnetic field. Doing so may produce image noise or artifacts.
10-EVS1MADB001 10
EVS 3643 User’s Manual 1. Safety information
Electric Shock Hazards
- To reduce the electric shock hazard, the system must be connected to an electrical ground.
- A three onductor AC power cable is supplied with this system to provide the proper electrical grounding. The power cable must be plugged into an UL-approved three-contact electrical outlet.
- Do not disassemble or modify the product as it may result in fire or electric shock. There are no operator serviceable parts or adjustments inside the systems. Only trained and qualified personnel should be permitted access to the internal parts of the system.
- If an APPLIANCE COUPLER or Mains Plug or other separable plug is used as the isolation means to safety
Disinfection and cleaning
- Wipe it with a dry cloth slightly damed with a neutral detergent.
- Do not use solvents such as alcohol, thinner or benzene. Doing so may damge the surface of the equipment
- Do not clean the system with turning the power on.
Operating/storage environment
- Be sure to use and store this equipment under the conditions described below:.
Temperature Humidity Atmospheric pressure
Operating environment
Transport & Storage
environment
- Do not expose this equipment to high temperatures and/or high humidity. Malfunction is occur.
- When not in use, keep the detector, handle unit, and grid in a designated location or in a location where they are safe and cannot fall down.
10 to 35
-15 to 55
30 to 85 % RH 700 to 1060 hPa
10 to 90 % RH 500 to 1060 hPa
Notes on disposal
- Disposal of this product in an unlawful manner may have a negative impact on health and on the environment. When disposing of this product, therefore, be absolutely sure to follow the procedure which is in conformity with the laws and regulations applicable in your area.
- The expected life span of EVS 3643 system is about 3 years
Handling the equipment
-The equipment must be handle with care to avoid personel injury or damage to the internal image sensor.
11
EVS 3643 User’s Manual 2. Introduction
The EVS 3643 Wireless is advanced wireless digital radiographic equipment in the DRTECH Exprimer series. This equipment is designed to provide the highest resolution and sensitivity in the series. In addition, the wireless LAN (IEEE 802.11n*) communication feature improves the operability, and high-speed processing.
2-1. Features
- Wireless LAN communication (IEEE 802.11n*) includes a lightweight and thin detector that is easy to handle.
- The shape of the detector, which is identical to that of a conventional film cassette complying with ISO4090, enables digital radiography in the existing analog radiography configuration
- The new sensor with 140 μm of pixel pitch and CsI (Cesium Iodide) used for the scintillator produces high resolution (approx. 7.86 Mega pixels) digital images within the effective imaging area (358 x 430 mm) with low doses of X-rays
- Depending on the operating conditions at each site, the wiring unit (optional) enables the equipment to be used through expansion to a wired connection
- At the time of installation, set a specific channel in the frequency band of 5.0 GHz before using the LAN. Note that the available frequency band for this standard varies, depending on the local radio laws, regulations and system requirements.
2-2. Intended use
- The EVS 3643 Digital X-ray detector is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy. This device is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. This device is not intended for mammography applications
12
EVS 3643 User’s Manual 2. Introduction
2-3. System Configuration
Generally, the EVS 3643 detector is used in system configuration as illustrated below:
Figure 2.1 EVS 3643 system configuration
Wireless communication is established between the EVS 3643 Wireless detector and System Control Unit. The EVS-3643 system is compliant with IEEE 802.11a/b/g/n (2.4 / 5 ). The available frequency band may vary depending on local radio laws and system requirements. Consult your local dealer for the frequency available in
your area.
- Use of multiple WLAN devices within the same frequency band may interference with
each wireless communication and cause a decline in transmission speed
- Do not cover or block the wireless module of the detector. Otherwise, the
transmission speed or operable distance may be reduced.
- Recommended maximum operating distance of wireless communication between the
detector and system synchronization unit is 8 meters.
13
EVS 3643 User’s Manual
EVS 3643 Wireless system consists of detector, system synchronization unit (SSU), CDs and relevant accessories. (Refer to chapter 3-1 “Product Components” for CD information)
X-ray Detector (EVS 3643)
System Syncronize Unit
(EVS-SSU01) Option
3. Product description
Access Point
(D-Link DIR-850L)
USB Switch Box Hand Switch
LAN Cable
(Gigabit LAN 10 m)
AC Power Cable (2m) Option Tether Interface (3 m)
Generator Interface Cable
(15m) Option
Interface Cable(7 m), Option Power adaptor (12V, 7.08A) Adaptor cable (3 m) Option
14
EVS 3643 User’s Manual
3. Product description
Functional cable (0.5 m)
CD(Software / Calibration) Maual (Hard Copy)
Battery charger
(EVS-BCS)
Battery pack
(EVS-MBP)
Table 3.1 EVS 3643 packaging
10-EVS1MADB001 15
EVS 3643 User’s Manual
3-1. Product component
Item Product name
3. Product description
Flat panel detector
System syncronization unit (SSU)
Battery chaarger and Battery pack
CD (Software / Calibration)
User’s Maual
EVS 3643(Scintillator: CsI : Tl) 2.98 kg
EVS-SSU01 : 2.2 kg Optional
EVS-BCS (charger): 0.5 kg EVS-MBP (battery pack): 0.24 kg
Document: User ‘s Manual (PDF) Calibration Data: Defect Map, Gain, Post Offset, Pre Offset Calibration & Configuration S/W: ECali1
Hard Copy
AC Power Cable (2m) Generator Interface Cable (15m, Optional) USB Switch Box Hand switch LAN Cable (10m, Direct, 1000BASE-T)
Accessories
Tether Cable (3m, Optional) Adaptor cable (3m, Optional) Functional cable (0.5 m) Power adapter Interface Cable (7m, Optional) License Dongle Key (USB)
Table 3.2 Product componets
WARNING
The use of accessories and cables other than those specified, with the exception of EVS 3643 Wireless accessories and cables sold by DRTECH Co., LTD. as replacement parts for internal components, may result in increased emissions or decreased immunity of the equipment. Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC standards. All combinations of equipment must be in compliance with IEC 60601-1-1 system requirements. Any person who connects additional equipment to the signal input or signal output ports configures a medical system, and is therefore responsible for ensuring that the system complies with the requirements of the system standard IEC 60601-1. If in doubt, consult DRTECH technical support representative
.
16
EVS 3643 User’s Manual 3. Product description
Workstation (Recommended and minimum but NOT included)
Item Specification
Operating system
CPU
Memory
Hard disk
LAN card
Monitor 1024 x 768 or higher
Optional disc drive CD or DVD R/W
Windows 7 64 bit SP1 (Professional Edition or higher)
Intel Core i5 2600 or higher (or compatible CPU)
4GB or higher
1TB or higher
Gigabit (Detector only) Intel® PRO 1000 Series (Gigabit LAN Card for network interface) Min. Requirements: 1Gbps, Jumbo Frames: 9K Receive Descriptors: 2K (higher than 1024) This is not dedicated to DICOM
Table 3.3 Workstation
Grid (Recommended but Not included)
Item Description
SID
Ratio
Frequency
Inter spacer
100 / 130 / 150 / 180 cm
8 : 1
215 Line/inch
Al
Table 3.4 Grid specifications
3-2. X-ray Imaging condition
X-ray Energy Range
40kVp ~ 150kVp
Reliability (Lifetime Dose)
More than 74Gy (35uGy x 365days x 24hours x 60minutes x 60seconds/15sec)
17
EVS 3643 User’s Manual 4. Part name and function
4-1. Detector specification
Item Description
Model EVS 3643 Purpose General Radiography Pixel Pitch 140 um Scintillator CsI (Cesium Iodide) Image Matrix Size 2560 × 3072 pixels Effective Imaging Area (H x V) 358 x 430 mm Image Acquisition and Transfer Time < 3 sec.
Spatial Resolution
Rated Power Supply Wireless Wired
Power Consumption Max. 24 W Network Interface 14 bit Digital Output Ethernet (1000BASE-T)
Dimensions (㎜)
Weight 2.98 kg
Min. 3.5 line pair/
DC +12V 2 A Powered by the battery pack (4,000 mA h) Powered by SSU using tether interface Powered by Power adopter using tether interface
386 (H) × 460 (V) × 14 (D)
Environmental Requirements
Temperature: +10 ~ +35
Operation
Humidity: 30 ∼ 85% RH(Without Condensing)
Atmospheric pressure: 700 ∼ 1060 hPa
Temperature: -15 ~ +55
Storage and Transportation(unpacked)
Table 4.1 Detector specifications
Tether Interface: Allows the detector to communicate with SSU via Ethernet cabling when wireless
communications is not available or higher speed data transfer is necessary
Humidity: 10 ∼ 90% (Without Condensing)
Atmospheric pressure: 500 ∼ 1060 hPa
18
EVS 3643 User’s Manual
4. Part name and function
4-2. Detector component
The detector is designed to capture radiographic images. Captured images are transmitted to the EVS 3643 image-capture computer using the wireless/wired data transfer
Figure 4.1 Detector components
A. Wireless antena :Transmits image data with wireless comunication (IEEE802.11n). B. Battery Pack : Supplies electricla power to the detector while communication wirelessly C. Sataus indicators
AP: Alter button for Wireled/Wireless(2.4 GHz / 5 Hz) Link: Shows detector registraion and connection status. Ready:shows data comunication status Power: Shows power on/off status of the detector.
Simultaneous blinking of two or more LED lamps indicates a system error. D. AP Button: registration detector. E. Power Button: Detector power on / off F. Connecter: Data communication and power supplying thought tether cable
19
EVS 3643 User’s Manual
4-3. Detector dimension
4. Part name and function
Figure 4.2 Detector dimension
20
EVS 3643 User’s Manual 4. Part name and function
4-4. Power supply and SSU (System syncronization unit)
4-4-1 SSU Specifications
Description
Item
(System synchronization unit)
Model EVS-SSU01 Power Supply Dimensions (W x H x D) 260 mm x 248 mm x 49 mm Weight 2.2 kg
Input: AC100 to 240V, 50/60Output: DC +12V 8.3A, 75W
SSU
Environmental Requirements
Temperature: +10 ~ +35
Humidity: 30 ∼ 85% RH(Without Condensing)
Operation
Atmospheric pressure: 700 ∼ 1060 hPa
Altitude: Max. 2 km
Temperature: -15 ~ +55
Storage and
Humidity: 10 ∼ 90% (Without Condensing)
Transportation(unpacked)
Atmospheric pressure: 500 ∼ 1060 hPa
Altitude: Max. 2 km
Table 4.2 System syncronization unit specifications
21
EVS3643 User’s Manual 4. Part name and function
4-5. Battery Charger and Battery Pack
4-5-1 Battery Charger
Item Description
Model
Simultaneous Charging Battery Pack 2 EA
Charging Time 3 hours
Rated Power Supply DC +12 V, 6 A Max.
Dimensions (W x H x D) 180 mm x 255 mm x 35 mm
Weight 0.5 kg
Table 4.3 Battery charger specifications
EVS-BCS
4-5-1-1 Battery charger components
Figure 4.4 Battery charger
A. Power indicator : indicates the power on/off status.. B. Charging indicator : C. Battery compartment: D. DC Input :
Connect the DC adapter to supply electrical power to the battery charger
Indicates the charging status.
Insert the battery pack to charge.
EVS 3643User’s Manual
4-5-1-2 Battery charger dimension
Unit : mm
5. Operating procedure
Figure 4.4 Battery charger system
EVS 3643 User’s Manual 4. Part name and function
4-5-2 Battery Pack specification
Item Description
Model
Cell Type Lithium Polymer
Number of Cell 2S1P (2series 1 Parallel)
Rated Power Supply Output : DC +7.4 V
Life Approx. 500 times (charge / discharge 1 cycle)
Dimensions (W x H x D) 163 mm x 148 mm x 7 mm
Weight 0.24 kg
Table 4.5 Battery charger specifications
EVS-MBP
4-5-2-1 Battery charger components
A. Charging indicator : Indicates the charging status B. Latch knob :
rotating on /off for battery swap
RA-EVS-UM-100
Figure 4.5 Battery pack
EVS3643 User’s Manual
4-5-1-2 Battery pack dimension
Unit : mm
4. Part name and function
Figure 4.6 Battery pack dimension
EVS 3643User’s Manual
4-5-3 Charging Battery Pack
5. Operating procedure
The battery pack supplies power to the detector during wireless connection. Be sure to use only the dedicated battery pack and fully charge it before use.
Connect the power adapter to the DC Input port of the battery charger. the power LED lights in blue indicating
the presence of direct current (DC) power.
Insert the battery pack into the battery charger. Charging starts automatically. The charge LED lights green
when the battery pack is being charged. After the battery pack is charged completely, all level of the charge LED will going to luminance
Gently pull the charged battery pack to remove from the battery charger
Securely plug the power cord into the power source. If contact failure
WARNING
CAUTION
CAUTION
occurs, or if dust or metal objects come into contact with the exposed metal prongs of the plug, fire or electrical shock may occur
Be sure to stop charging the battery pack when the charge LED lights in green beyond the specified charging time. Not doing so may result in battery pack overheating or smoking or in explosion or fire.
You must use the power adaptor that is certified with IEC 60950 or IEC 60601-1.
Two batteries can be charged at the same time.
It takes approximately two hours to fully charge a battery pack. The required charging time may vary depending on the temperature and remaining battery level.
EVS 3643 User’s Manual
4. Part name and function
4-6 Wireless access point(Optional)
This antenna equipment relays captured images from the detector to the control system. No operation is
required for this equipment while using the EVS Digital Radiography system.
4-5-1 Access point Specifications
Description
Item
Model D-Link DIR-850L Power Communications Dimensions (W x H x D) 85.73 x 111.13 x 145.28 mm Weight 0.28 kg
IEEE 802. 11n/g(2.4 GHz), IEEE 802.11ac/n/a (5GHz)
Access Point
(AP)
DC 12V/2A
Environmental Requirements
Temperature: 0 ~ +40
Operation
Humidity: 10 ∼ 90% RH(Without Condensing)
CAUTION
RA-EVS-UM-100
Do not use the wireless access point within the patient’s vicinity
EVS 3643User’s Manual 5. Operating procedure
The EVS 3643 Detector is properly adjusted at installation by the service engineer. If you encounter any
problems during normal operation or daily inspections, consult your sales representative or local DRTECH
dealer.
General workflow
The following workflow indicates the procedures after startup of the EConsole1 and other system equipment
5-1. Preparing to use the detector
Attach a fully-charged battery pack to the detector.
5-2. Operating the detector
1. Turn on the detector
2. Resister1)the detector and make connection2)to the EConsole1 power supply to the detector
3. Conduct Examination
Select or resister the patient information
Select the protocol (selection3)of the detector)
- Arrange the patient in the correct posture
- Position the X-ray generator to adjust the exposure field
- Check all the conditions
1)
: A procedure in order to register the detector
to a specific digital radiography system
2)
: Network connection between the EVS 3643
wireless detector and the EConsole1
3)
: Selection of EVS 3643 wireless
from the Exprimer series detector Selection of wireless/wired data transfer
Loop back procedure for each patient
Check the captured images
List the images
Transmit the images
Sterilize the portion of the detector that has been in contact with a patient
5-3. Ending use of the detector
1. Turn off the detector
2. Remove the battery pack
Loop back procedure for each body part
Conduct the next examination
For details, refer to the operation manual or setup guide of the EConsole1
EVS3643 User’s Manual
5-1. Preparing to use the detector
5. Operating procedure
CAUTION
Be sure to use only the dedicated power supply for the EVS-3643 detector
Standard Configuration
Configuring with SSU (power supply) to the detector
Make sure that the cable is placed to the connector (detector side)
Figure 5.1 EVS 3643 System Configuration
5-2. Operating detector
1
Turn on the detector
Before operating the detector, start up the EConsole1
Press and hold the POWER button (approx. 1 second)
Power lamp(Blue) light up
10-EVS1MADB001 2924
EVS 3643User’s Manual
2
Register the detector and make connection to the EVS control system
Registration
Press and realease the AP button Link lamp flashes.
5. Operating procedure
When the AP Lamp flashes 1 time in 2sec, wired mode status has showing.
When the AP Lamp flashes 2 times in 2sec, wireless (2.4 GHz) mode status has showing.
When the AP Lamp flashes 3 times in 2sec, wireless (5 GHz) mode status has showing.
Connection
Network connection between the internal
wireless module of the detector and the
wireless access point/EVS control system is secured automatically The link lamp lights up when the detector is resistered and the communication connection is established
The LINK lamp does not light up when the detector is not registered or the communication
connection is not established.
When the Ready and LINK Lamp flash, a communication error has occurred.
See Troubleshooting.
3
Conduct examination
For details about operation, refer to the Operation Manual for the ECali1.
i) Select the patient information or protocols on the screen and start the examination.
EVS 3643 User’s Manual
The READY lamp flashes during preparation for examinations.
The READY lamp lights up when the detector and EConsole1 change to exposure ready status.
• Arrange the patient in the correct posture and position the detector aligning it with the target body part.
• Position the X-ray generator to adjust the exposure field.
• Check all conditions before exposure. Make sure that four LED lamps (POWER, READY, LINK, AP) are lit. This means that the system is
ready to start an examination
A communication error has occurred when two or more lamps flash.
5. Operating procedure
When the READY lamp (green color) flashes slowly, the detector is in detector selection status
(Sleep).
The detector enters detector selection status automatically when it has not been used for a certain
period of time.
ii) Press the exposure switch of the X-ray generator.
Images captured with the detector are transmitted to the EConsole1 and appear on the monitor.
Check the images on the monitor.
If any uncompleted protocols remain, repeat the
procedure ii).
ii) Click the button for ending the examination
images are stored automatically
To conduct examination for another patient, repeat the step 3.
IMPORTANT
Sterilize the portion of the detector that has been in contact with a patient to prevent infection
10-EVS1MADB001 25
EVS 3643 User’s Manual
A signal strength indicator appears on the screen of the ECali1 computer. It shows the wireless
communication level between the detector and the ECali1.
Keep the wireless communication level stable on capturing or transmitting images.
5. Operating procedure
Display
Signal strengt
(comunication stability)
Very high (Stable) Normal
High (Stable) Normal
Unstable
Low
(Unstable)
No signal
(Comunication failed)
comunication. Comunication
comuncations
Status Required actions
-
-
Check whether there is any obtacle (e.q., your hands)between the wireless module and the wireless acess point. If there is any obtacle, remove it.
speed is
lowered
Cannot
If the problem cannot be resolved. Consult your sales representative or local DRTECH dealer Confirm that detector and the access point are turned on. If the problem cannot be resolved. Conult your sales representative or local DRTECH dealer
10-EVS1MADB001 26
EVS 3643 User’s Manual 5. Operating procedure
5-3. Ending use of the detector
Turn off the detector
Press the SSU POWER button. All the LED lamps are off.
Detector status list
Lamp type
Color
Power ON During detector registration Detector registration completed (1 Sec,) Communication established During exposure preparation
Ready status or performing an examination (Ready)
Detector selection status (Sleep) During image data transmission
Power OFF
: Light on : Flashes (On/Off status changes every second)
х : Lights off : Flashes slowly (On/Off status changes every 2 seconds)
Power
Lamp
Blue Green Orange Green Green
- -
○ / ★ - -
х х
х
х
х
х
Ready
Lamp
Link
Lamp
х
-
-
х
х
AP
Lamp
х
-
- : Unspecified status
10-EVS1MADB001 27
EVS 3643 User’s Manual 6. Extension facility
6-1. Using the wireless detector with other EVS series detectors in a
same system
The ECali1 with the EVS 3643 detector enables connection with EXPRIMER series detectors other than that EVS 3643 detector, according to preset settings. Select a detector when conducting an examination. For details, consult your sales representative or local DRTECH dealer.
Figure 6.1 Detector dimension
10-EVS1MADB001 28
EVS 3643 User’s Manual 6. Extension facility
6-2. Sharing the detector among different systems
The EVS 3643 detector can be used in two or more systems in which protocols for the detector have already been preset. For details, consult your sales representative or local DRTECH dealer
EVS 3643 Detector
X-ray
generator
X-ray
generator
Hospital Network
Figure 6.2 Sharing the detector among different systems
10-EVS1MADB001 29
EVS 3643 User’s Manual 6. Extension facility
6-3. AED mode
AED mode can make exposure as using internal storage in the detector with out a wireless connection
The images can be transmitted to a PC and used by connecting the detector to Viewer. The transmitting
images in the detector are removed automatically.
Image processing and correction are available after connecting the detector to Viewer
Figure 6.3 AED mode
10-EVS1MADB001 30
EVS 3643 User’s 6. Extension facility
6-4. Wiring connections
The EVS 3643 detector has a cable connector. With the wiring unit (optional), it is possible to expand from a wireless configuration to a wired configuration (see the figure below). Wired configuration is suitable for use where the detector is set in a Bucky stand and table in place of a film cassette. In this configuration, for the reason that data communication and power supply are made via a cable, users do not have to be concerned about power and wireless communication failure. This reduces the time and labor involved in charging and replacing.
Figure 6.4 Wiring connection with SSU
X-ray Generator
X-Ray Exposure
EVS_Tether_Cable
EVS_Tether_Cable
SSU
Hospital Network
USB SW BOX
Figure 6.5 Wiring connection with USB S/W BOX
For details about wired configuration, consult your sales representative or local DRTECH dealer.
10-EVS1MADB001 31
EVS 3643User’s Manual
7-1. Hardware Installation
This section describes how to connect the EVS 3643 system (Detector)
5. Operating procedure
CAUTION
7-1-1 EVS 3643
1
Connect the one end of the generator interface cable to the X-ray port of SSU, and the other to the port of
the x-ray generator.
Installation of this equipment should be made by licensed and authorized personnel.
wiring
connection
10-EVS1MADB001 32
EVS 3643 User’s
2
Connect the LAN cable to PC port of SSU, and the other to the LAN Card connector of workstation
assigned for the data transfer
7. How to Install
3
To transmit image data and connect power with Tether Interface, connect the Tether interface cable to FPD
1 or FPD 2 of SSU
10-EVS1MADB001 33
Tether
EVS 3643 User’s
4
Connect the power cable to the AC pport of the SSU to supply power
7. How to Install
CAUTION
5
Turn on the power switch in the front of the SSU
This equipment must only be connected to a main power with protective earth
10-EVS1MADB001 34
EVS 3643 User’s
7-2. ECali1 Installation
7-2-1 System Requirements
Table 7.1. System requirement
Items Minimum Recommended
CPU Intel i3-2100 Intel i5-3470
RAM 2 GB 4 GB
HDD 100 GB 500 GB
NVIDIA GeForce
VGA Intel HD GRAPHIC 2000
GT630 1GB
ODD CD-ROM DVD Recorder
7. How to Install
OS Windows XP Pro SP2
Display Size 17 inch 23 inch
Display Resolution 1024 x 768 1920 x 1080
Efficient operation is not guaranteed for PCs that do not comply with the recommended specification.
Windows 7 Pro
32 / 64bit
7-2-2 Installation & Removal
Installation prerequisites Windows 7
Windows 7 users must set the User Account Control (herein after ‘UAC’) permission and disable the
firewall.
1) UAC Setting
Click the start icon
Click ‘Control Panel’
Select ‘User Accounts’
In the User Accounts window, click ‘User Accounts’
In the User Accounts task window, click ‘Change User Account Control Setting’
Adjust the slider to the ‘Never notify,’ and then click ‘OK’.
10-EVS1MADB001 35
EVS 3643 User’s Manual
7. How to Install
2) Disabling the Firewall
Click the Start icon
Click ‘Control Panel’
Click ‘Security’
Click ‘Windows Firewall’
Click ‘Turns Windows Firewall on or off’ on the left
Click ‘Off (not recommended),‘ and then click ‘OK’
Windows 8
Windows 8 users must set the User Account Control (herein after ‘UAC’) permission,
Disable the firewall and run the program as the system administrator.
3) UAC Setting
Click ‘Setting’ on the right hand menu
Click ‘Control Panel’
Select ‘User Accounts’
In the User Accounts window, click ‘User Accounts’
In the User Accounts task window, click ‘Change User Account Control Setting’
Adjust the slider to the ‘Never notify,’ and then click ‘OK’
4) Disabling the Firewall
Click ‘Setting’ on the right hand menu
Click ‘Control Panel’
Click ‘Security’
Click ‘Windows Firewall’
Click ‘Turns Windows Firewall on or off’ on the left
Click ‘Off (not recommended),‘ and then click ‘OK’
5) Run as Administrator
Windows 8 users must run the Installation File and Program as the Administrator Install the program by right clicking (press and hold, if touch screen) and selecting ‘Run as Administrator’ After completing the installation, set the icon, so that it always runs as administrator
10-EVS1MADB001 36
EVS 3643 User’s Manual 7. How to Install
Right click (press and hold, if touch screen) the shortcut icon
Click ‘Properties’
Click the ‘Advanced’ icon in the ‘Shortcut’ tab‘
Check the ‘Run as administrator’ box, click on ‘OK’
7-1-3. Installation
Run the provided setup file.
Figure7.1. ECali1 Setup File
At least 2GB of available HDD space is required for the proper installation of ECali1!
Windows 8 users must run the file as ‘administrator’.
Right-click the file, select ‘Run as administrator’ in the menu.
Figure 7.2. Setup Wizard Window
10-EVS1MADB001 37
EVS 3643 User’s Manual 7. How to Install
Figure7.3. Language & Number of Detector Setting
Figure7.4. Setting of the Model of the First Detector
10-EVS1MADB001 38
EVS 3643 User’s Manual 7. How to Install
Figure7.5. Installation Path Setting
Figure 7.6. Start Menu Folder Selection
10-EVS1MADB001 39
EVS 3643 User’s Manual 7. How to Install
Figure7.7. Additional Task Setting
Figure 7.8. Verification of Settings
10-EVS1MADB001 40
EVS 3643 User’s Manual 7. How to Install
Figure 7.9. Installation Window
Figure7.10. Installation Window
10-EVS1MADB001 41
EVS 3643 User’s Manual 7. How to Install
Figure 7.11. Installation Window
Figure7.12. Installation Window
10-EVS1MADB001 42
EVS 3643 User’s Manual 7. How to Install
7-1-4 Settings Post Installation
1) Windows 8 users must set ECali1 so that it runs as administrator.
Right click (press and hold) the ‘ECali1.exe’ in the installation destination (in C:\Cali1 by default).
‘Click ‘Properties’.
Select ‘Advanced’ in the ‘Shortcut’ tab.
Check the ‘Run as administrator’ box, then click ‘OK’
7-1-5 IP Address Configuration
It is required when using a EVS 3643 detector using an Ethernet Interface. The IP Address of the Network Adapter connected with the EVS 3643 Detector System installed on the PC should be configured. For Windows 7, Control Panel – Network Connections – Network and Sharing Center – Manage Network Connections on the left menu bar – Right click on the network connected to the Detector – Properties – Double click Internet Protocol Version 4 (TCP/IPv4)
Figure7.13. IP Address Configuration
IP Address: 192.168.250.XXX
For F600: 192.168.250.100 Others: 1 XXX 254 (Excluding 150 and 200)
Subnet mask: 255.255.255.0
10-EVS1MADB001 43
EVS 3643User’s Manual 5. Operating procedure
7-1-6 Uninstall ECali1
1) Click ‘Uninstall ECali1’ in Windows start menu – All Programs - ECali1
Figure 7.14. Uninstall ECali1
2) Click ‘Yes’ and proceed with uninstall when a window appears verifying the removal of the program.
Figure7.15. Uninstall ECali1
10-EVS1MADB001 44
EVS 3643 User’s Manual 7. How to Install
7-2 ECali1 Operation & Detector Integration
7-2-1 Program Start-up
In order to run the program, you can select ECali1 in Windows Menu – All Programs – ECali1 Folder, or by double clicking ECali1.exe in C:\ECali1 (or the customized installation path)
Figure7.16. ECali1 Start-up
Note
An error notification as in Figure 7.17. Pixel / Gain map File Notification) will be presented during the
initial start-up of the program.
It is a normal notification informing the absence of Pixel map file and Gain map file, so proceed by
clicking ‘OK’.
Figure 7.17. Pixel / Gain map File Notification
Note
Please install the additional software in accordance with the specification of your OS if the program does not operate after proper installation
For Windows XP SP3, Windows 7, and Windows 8:
VC2008SP1 Runtime (vcredist_x86.exe)
http://www.microsoft.com/en-us/download/details.aspx?id=5582
For Windows XP SP2:
Wireless LAN API package
http://www.microsoft.com/en-us/download/details.aspx?id=2098
Microsoft Core XML Service (MSXML) 6.0
http://www.microsoft.com/en-us/download/details.aspx?id=3988
VC2008SP1 Runtime (vcredist_x86.exe)
http://www.microsoft.com/en-us/download/details.aspx?id=5582
10-EVS1MADB001 45
EVS 3643 User’s Manual 7. How to Install
For Windows XP SP1 or earlier OS:
ECali1 will only operate in Windows XP SP2 and later environments. Please install Service
Pack 2 or 3 if you are using Windows XP.
We do not guarantee proper operation of ECali1 in OS earlier than Windows XP
7-2-2 Detector Connection
1) The Detector which will be utilized should be registered prior to operating the program.
2) A Registration and Configuration window for the Detector will appear during the initial start-up of
the program.
3)This configuration window is available through Configuration in the Option menu, and please refers
to 1) Detector Configuration for detailed information.
Figure7.18. Detector Registration
displays the registered Detector.
If a specific item requires editing, select the Model to be edited in and click Edit.
Select a Model in and click Add in order to add a Detector.
The most recently registered model will be deleted if Delete is selected.
10-EVS1MADB001 46
EVS 3643 User’s Manual 7. How to Install
Changes in configurations must be saved.
Save the changes in configuration by selecting the save icon below
Figure 7.19. Configuration Save icon
When the save button is selected, a program restart message will appear.
Restart the program by selecting ‘OK’.
10-EVS1MADB001 47
EVS 3643 User’s Manual 7. How to Install
Verify Detector Connection after restarting the program.
Open the ‘Log’ tab after connecting and powering on all necessary devices including the Detector.
Figure7.20. Detector Connection Log
The Detector has been successfully integrated if a ‘Ready to Capture!’ appears in the window as
seen in the figure.
10-EVS1MADB001 48
EVS 3643 User’s Manual 7. How to Install
7-3 Grid Configuration
1)Grid type must be configured for each registered Detector..
2)The Grid type is set as ‘None’ by default, so this step may be skipped if a Grid isn’t used.
3) Grid configuration may be completed in the ‘Detector #’ tab. # will be designated with alphabets of A, B, C, ….
Figure7.21. Grid Configuration
4 ) Please verify the Model and Grid of the registered Panel since a limited number of Detectors
may use the Grid with the use of multiple Detectors.
10-EVS1MADB001 49
EVS 3643 User’s Manual 7. How to Install
5) Filter file must be registered after the configuration of Type.
Figure7.22. Registering Grid Filter File
Select the Filter file by selecting ‘…’ as in the figure.
A file compatible with the configured Grid Type must be selected.
Filter file is located in the [ECali1 Installation Folder]\Filter Folder.
It is in the C:\ECali1\Filter folder by default.
Figure 7.23. Filter File
6) Input the Lines/Inch and Angle in accordance with the Grid Type. The cm to Inch
button may be utilized to alter the units when entering the Lines/Inch, if the cm unit is known.
10-EVS1MADB001 50
EVS 3643 User’s Manual 7. How to Install
Figure7.24. Additional Grid Configuration
7) Save and restart the program after completing the configuration.
10-EVS1MADB001 51
EVS 3643 User’s Manual 7. How to Install
7-4 Map File Registration
Open the configuration tab for each registered Detector though Option Menu – configuration and
opening the Configuration window.
Figure7.25. MAP File Destination Configuration
Use the ‘…’ on the right or manually enter in a new directory to change the destination
name. Please set it as a file used or provided by the X-ray acquisition software.
Save the changes in configuration using the save button below after designating a
destination for all registered Panels.
Figure7.26. Save Button
ECali1 will automatically restart when the changes in configuration has been saved.
10-EVS1MADB001 52
EVS 3643 User’s Manual 7. How to Install
7-5 ECali1 UI Overview
7-5-1 Main Screen of the Program
The main screen of ECali1 is as illustrated in [7.27].
Figure 7.27. Main Screen of ECali1
Category Title Reference
Menu
Toolbar
Pixel Map Toolbar
Thumbnail Thumbnail (p. 67)
Image Viewer Image Viewer (p. 68)
Pixel Viewer
Mini Map / Pixel Value 7-7 Mini Map and Pixel Value (p. 70)
Histogram Histogram (p. 71)
Status Bar
Menu (p. 49)
Toolbar (p. 53)
Pixel Map Toolbar and Pixel Viewer (p. 66)
Pixel Map Toolbar and Pixel Viewer (p. 66)
Status Bar (p. 72)
10-EVS1MADB001 53
EVS 3643 User’s Manual 7. How to Install
1) Menu
Corresponds to in [Figure
7.27] and comprised of 5 menus, which are File, Tools, Option,
Detector and Help.
A. File Menu
Figure 7.28. File Menu
Table 7.1. File Menu Descriptions
Menu Descriptions
Load pixel map Loads the Pixel map file.
Save pixel map Saves the Pixel map file.
Load gain map Loads the Gain map file.
Load panel map Loads the Panel map file.
Load raw image Loads the RAW image file in .IMG1 format.
Saves the currently selected image as a RAW image file in .IMG
Save raw image
format.
Exit Exits ECali1.
1
.IMG file: Image Data file that has 16 bit gray-level pixel value
10-EVS1MADB001 54
EVS 3643 User’s Manual 7. How to Install
B.Tools Menu
Menu Descriptions
Gain map calibration
Create FGain map
Filter calibration
Filter calibration using files
Figure 7.29. Tools Menu
Table 7.2. Tools Menu Descriptions
Initiates the Gain Map Calibration. Please refer to7-10 Gain Calibration
for more details.
Generates a Filtered Gain Map2using the currently loaded Gain Map. Will not operate if the loaded Gain Map is a Filtered Gain Map. The generated Filtered Gain Map is saved in the configured Gain Map destination of the Detector (Panel) set in Configuration. Run the Grid Filter Calibration task.
Please refer to
수 없습니다.
오류! 참조 원본을 찾을 수 없습니다.. 오류! 참조 원본을 찾을
for more details.
Run the Grid Filter Calibration using raw files.
Please refer to
수 없습니다.
오류! 참조 원본을 찾을 수 없습니다.. 오류! 참조 원본을 찾을
for more details.
Adjusts the magnifying ratio of the image displayed in the Image Viewer
Scale Fit
domain.
Adjusts the magnifying ratio of the image displayed in the Image Viewer domain to set the size of the image corresponding to its actual size.
Scale 1:1
The Pixel displayed on the screen will be identical to the actual Pixel of the Detector.
Saves the configured scale ratio in the current Image Viewer and applies it to
Fixed Scale
the subsequent images.
2
Filtered Gain Map: Gain Map maintaining the uniformity of the image even if the uniformity of the X-
ray source is different, by applying filtering to the Gain Map
10-EVS1MADB001 55
EVS 3643 User’s Manual 7. How to Install
C. Option Menu
Menu Descriptions
Configuration
Engineer Mode
Terminal Mode
Remove grid ON/OFF
AED – Acq mode
AED – Stay mode
Ready
Figure 7.30. Option Menu
Table 7.3. Option Menu Descriptions
Opens the Configuration window. Please refer to. 7-12 Configuration for more details regarding Configuration. Initiates Engineer Mode of ECali1 for Advanced Configuration. Requires the input of a password. Initiates the Terminal Emulator which allows the internal configuration of the Detector. Sets the Grid Pattern Algorithm application. If a check mark () is available as noted in the figure, the algorithm is applied (ON). This category must be ON if Grid is utilized.
Activates AED mode to allow image acquisition.
Inactivates AED mode and reverts to Standby.
Converts the status of the Detector to Ready.
The Detector performs Exposure.
Exposure
Cancels the Ready status of the Detector and reverts to Standby.
Ready cancel
10-EVS1MADB001 56
EVS 3643 User’s Manual 7. How to Install
D. Detector Menu
Figure 7.31. Detector Menu
Select the active Detector from the registered Detector (Panel). Number of Menus reflecting the number of registered Detectors through Option – Configuration
- General will be generated. Only one Detector may be selected
E. Help Menu
Figure7.32. Help Menu
Table 7.5. Help Menu Descriptions
Menu Descriptions
Displays the ECali1 manual.
Manual
The manual is in PDF format, thus Adobe Reader program will be required.
http://get.adobe.com/kr/reader/
About Will display the version information of ECali1. Check update
* Requires internet connection. Checks and installs updates for ECali1.
10-EVS1MADB001 57
EVS 3643 User’s Manual 7. How to Install
7-5-2 Toolbar
Corresponds to ⓑ in [7.27] and provides most used menus as Toolbar icons.
Figure 7.33. Toolbar
Table 7.6. Toolbar Icon Descriptions
Icon Descriptions
Performs Gain map calibration. Corresponds to the Gain map calibration in the Tools Menu.
Performs Grid Calibration. Corresponds to the Grid map calibration in the Tools Menu.
Loads and applies the Pixel map file. Corresponds to the Load pixel map in the File Menu.
Stores the currently applied Pixel map as a file. Corresponds to the Save pixel map in the File Menu.
Loads and applies the Gain map file. Corresponds to the Load Gain map in the File Menu.
Loads and displays the projected image from the file. Corresponds to the Load raw image in the File Menu.
10-EVS1MADB001 58
EVS 3643 User’s Manual 7. How to Install
Stores the currently projected image as a file.
Corresponds to the Save raw image in the File Menu.
This category must be ON when using the Grid.
Corresponds to the Remove grid ON/OFF in the Tools Menu.
Adjusts the magnifying ratio of the image in ⓔ to fit on screen.
Corresponds to the Scale Fit in the Tools Menu.
Adjusts the magnifying ratio of the image in ⓔ to actual size of the
image.
Corresponds to the Scale 1:1 in the Tools Menu.
* Only available when the selected Detector (Panel) supports AED mode.
* Only available when AED mode is active.
Changes the status of the Detector to acquire images. Corresponds to
the AED - Acq mode in the Options Menu.
* Only available when the selected Detector (Panel) supports AED mode.
* Only available when AED mode is active.
Converts the Detector to Standby mode.
Corresponds to the AED - Stay mode in the Options Menu.
* Only available when the selected Detector (Panel) supports AED mode.
This category must be activated when using AED mode.
10-EVS1MADB001 59
EVS 3643 User’s Manual 7. How to Install
7-5-3 Pixel Map Toolbar and Pixel Viewer
Corresponds to and in [Figure 7.27], and utilized when creating a Pixel Map.
Figure7.34. Pixel Map Toolbar
Figure 7.35. Pixel Viewer
The selected area from the Image Viewer will be magnified and displayed in the Pixel Viewer. Please refer to 7-11 Pixel
10-EVS1MADB001 60
Correction for more details and methods of generating Pixel Maps.
EVS 3643 User’s Manual 7. How to Install
7-5-4 Thumbnail
Corresponds to in [Figure 7.27].
Displays the Thumbnail of the projected image, and allows the user to select a specific image between multiple images.
If there are a significant number of projected images, a scroll button on the top and bottom allows the user to select a specific image.
Select the Thumbnail to be deleted and select ‘Delete’ if a re-projection is required during the calibration. The selected image will be deleted and a re-projection may be performed.
The following information will be displayed if an image is loaded from a file.
Figure 1.36. Thumbnail
The file name will be displayed on the top left. STD: Standard Deviation Value of the Pixel value of the image. Mean: Mean value of the pixel value of the image.
Figure 7.37. Thumbnail Information
10-EVS1MADB001 61
EVS 3643 User’s Manual 7. How to Install
7-6 Image Viewer
Corresponds to in [Figure 7.27].
Figure 7.38. Image Viewer
10-EVS1MADB001 62
EVS 3643 User’s Manual 7. How to Install
Displays the projected image and supports the following functions.
Table 7.7. Image Viewer Function & Control Method
Function Control Method
Magnify Roll the mouse wheel forward in the Image Viewer.
Reduce Roll the mouse wheel backwards in the Image Viewer.
Move Image Left click, drag and drop the image in the Image Viewer.
W/L3 Width Increase Drag the right mouse button down.
W/L Width Reduction Drag the right mouse button up.
W/L Left Shift Drag the right mouse button left.
W/L Right Shift Drag the right mouse button right.
Magnify Pixel Double click the left mouse button in the area within the Image Viewer
in order to magnify the selected are in the Pixel Viewer screen.
3
W/L: Window Leveling
10-EVS1MADB001 63
EVS 3643 User’s Manual 7. How to Install
7-7 Mini Map and Pixel Value
Corresponds to in [Figure 7.27]
7-7-1 Mini Map
Figure 7.39. Mini Map
Allows the users to verify the area currently being viewed in the Image Viewer. The area within the green dotted boarder is the area currently displayed in the Image Viewer.
7-7-2 Pixel Value
Figure 7.40. Pixel Value
Displays the pixel value of the location the mouse cursor is hovering in the Image
Viewer or the Pixel Viewer
10-EVS1MADB001 64
.
EVS 3643 User’s Manual 7. How to Install
7-8 Histogram
Corresponds to in [Figure 7.24]
Displays the Histogram of the selected image
Figure7.41. Histogram
Supports Dynamic or Static mode subsequent to the Windows Leveling conditions
7-8-1 Dynamic
The minimum and maximum values will be fixed while the Histogram values may be adjusted if Window Leveling is attempted in the Image Viewer while in Dynamic mode.
7-8-2 Static
The Histogram values will be fixed while the minimum and maximum values may be adjusted if Window Leveling is attempted in the Image Viewer while in Static mode.
10-EVS1MADB001 65
EVS 3643 User’s Manual 7. How to Install
7-9 Status Bar
Corresponds to in [Figure 7.27]
7-9-1 Image Information Display
Figure7.42. Status Bar
Displays various information of the image
Current axis and Pixel value of the mouse cursor
Displayed as (X : [X Axis] Y : [Y Axis]) = [Pixel Value]
Mean: Mean Pixel Value of the image
STD: Standard Deviation of the Pixel value of the image
Min: Minimum value of the image Histogram
Max: Maximum value of the image Histogram
Peak: The Peak value and its Position of the image Histogram
Displayed as [Peak Value] at [Position]
7-9-2 Detector Status Display
* Supports only Ethernet Type and F600.
A green light will be displayed if the Detector is able to take a projection and a red light will be displayed if the Detector is unable to take a projection.
10-EVS1MADB001 66
EVS 3643 User’s Manual 7. How to Install
7-9-3 Wireless Information Display
* Support only Wireless Detector.
Information displayed when using a Wireless data communication.
7-9-4. Map File Loading Display
Green light will be displayed if a Map file required for Calibration has been loaded and the red light will be displayed if it has not been loaded.
Gain: Gain Map Loading Pixel: Pixel Map Loading Panel: Panel Map Loading
If a Filtered Gain Map has been loaded in place of a Gain Map, the following FGain will be displayed.
10-EVS1MADB001 67
EVS 3643 User’s Manual 7. How to Install
7-10 Gain Calibration
Gain Calibration Procedure will update or modify the x-ray characteristics to be combined
at the field to enhance the acquired image quality.
7-10-1 Gain Calibration Preparation
1)Detector and Grid Configuration
The Detector and Grid configuration must be completed prior to performing Gain
Calibration. Please refer to for details pertaining to the configuration.
2)MAP Data Location Configuration
The provided Pixel Map (*.MAP) and Gain Map (*.GMP) should be Loaded prior to
performing Gain Calibration. The destination location of the Map file provided by the X-ray image acquisition software (EConsole1 and etc.) should be registered in configuration.
Please refer to 7-4 Map File Registration for more details
10-EVS1MADB001 68
EVS 3643 User’s Manual 7. How to Install
7-10-2 Gain Calibration
Actual X-ray must be shot to project an image in order to perform Gain Calibration.
The Gain Calibration projection conditions are as follows. [Ta b l e 7- 8]
Table 7.8. Gain Calibration Condition
Conditions Descriptions
Projection must be taken without the application of Grid.
Grid application
Subject Projection must be done without a subject
X-ray exposure X-ray exposure is required
X-ray radiation dose
SID Stand: 150 cm, Table: 100 cm
Number of
projections
(Cautions) The Grid Type must be set ‘None’ in Configuration -
Detector - Grid selection - Type before Gain Calibration, and must
be restored after Gain Calibration.
Adjust the mean pixel value of the projected image to be between
3,000 ~ 4,000
X-ray Condition example) 70kVp, 200mA, 2mAs
Default Value: 7 counts
Configuration possible in Configuration - General - Calibration -
Count.
Default Value: 60 Seconds
Projection interval
Other things
Configuration possible in Configuration - General - Calibration -
Interval.
Collimator should be open maximally
X-ray should be set to be exposed in the whole Detector
Detector should be aligned in the center
10-EVS1MADB001 69
EVS 3643 User’s Manual 7. How to Install
1) Detector Selection
Select the Detector to perform the Gain Calibration from the Detector Menu.
Figure 7.43. Detector Selection
2) Starting Gain Map Calibration
Next, select Gain map calibration from Tools Menu or select the shortcut icon.
Figure7.44. Starting Gain Calibration
A notification message will appear. Select ‘OK’.
Figure 7.45.. Gain Calibration Start Notification
10-EVS1MADB001 70
EVS 3643 User’s Manual 7. How to Install
Next, ECali1 alerts whether the radiation dose is appropriate or whether it should be automatically calculated. Proceed by selecting ‘OK’.
Figure 2. X-ray Radiation Dose Check Activation
Next, a main screen will appear. Proceed with the X-ray projection at this point in order to acquire an image.
Figure 7.473. Projection Standby Main Screen
The program automatically determines the appropriateness of the X-ray radiation dose when the image is acquired.
Figure7.48. X-ray Radiation Dose Verification Notification
10-EVS1MADB001 71
EVS 3643 User’s Manual 7. How to Install
Figure7.49. Notification for Low X-ray Radiation Dose
If provided with a notification of low radiation dose, please select ‘OK’ then increase the X-ray
radiation dose and re-project.
A notification will also appear if the X-ray radiation dose is high. Please select ‘OK’ then reduce the X
ray radiation dose and re-project.
Figure 7.50. Notification for Appropriate X-ray Radiation Dose
If the X-ray radiation dose is appropriate, please select ‘OK’ and proceed with the projections.
Figure 7.51. Notification Informing 7 Images Requirement
A notification informing the requirement of 7 images will appear. The required number of images may
be set in Configuration.
10-EVS1MADB001 72
EVS 3643 User’s Manual 7. How to Install
Select ‘OK’ and proceed with the projections of 7 images.
Figure 7.52. Starting the Gain Calibration Projection
Figure7.53. Standby post Projection of 1 Image
Stand by until the next designated projection after the initial projection.
Figure7.54. Projection Standby Timer
The next projection will take place after the projection standby timer disappears. The projection
standby time may be set in configuration
.
73
EVS 3643 User’s Manual 7. How to Install
The Gain Calibration results are automatically saved as Gain Map (*.GMP) when all required images are
projected and the existing files are backed-up.
Figure7.55. Notification for Gain Calibration Completion
The existing GMP files are backed up in the [ECali1 Installation Folder]\MAP folder. The file name is as follows. [7-56]
Figure7.56. Example of a Back-up File
10-EVS1MADB001 74
EVS 3643 User’s Manual 7. How to Install
7-11 Pixel Correction
If there are pixel defects or line defects needed to be calibrated, users can correct them manually.
7-11-1 Pixel Correction Preparation
1) Detector and Grid Configuration The Detector and Grid configuration must be completed prior to performing Pixel correction
2)Map Data Location Configuration
The provided Pixel Map (*.MAP) should be Loaded prior to performing Pixel Correction. The destination location of the Map file provided by the X-ray image acquisition
Software (EConsole1 and etc.) should be registered in Configuration
.
10-EVS1MADB001 75
EVS 3643 User’s Manual 7. How to Install
7-11-2 Pixel Correction Performance
At least 1 count of RAW image acquired through X-ray Exposure is required to perform Pixel Correction.
Table 7.4. Pixel Correction Conditions
Condition Descriptions
Grid application Projection must be taken without the application of Grid
Subject Projection must be done without a subject
X-ray exposure X-ray exposure is required
Number of
projection
Other things Collimator should be open
More than 1 count.
(One for check and correct defects, and others to verify
correction)
10-EVS1MADB001 76
EVS 3643 User’s Manual 7. How to Install
1) Detector Selection
Select the Detector to perform the Pixel Correction from the Detector Menu.
Figure7.56. Panel Selection
2) Pixel Correction
Acquire the image by performing an X-ray projection from the main screen.
Figure 7.57. Projection Standby Screen
Figure 7.58. Image Acquisition Screen post Projection
10-EVS1MADB001 77
EVS 3643 User’s Manual 7. How to Install
Figure 7.59. Display the Horizontal Line Defect Area in Pixel Viewer
A defect of a point or line may exist for each Detector. [7-59] illustrates a Horizontal Line Defect.
There are 3 types for Pixel Correction.
Figure 7.60. Pixel Defects
10-EVS1MADB001 78
EVS 3643 User’s Manual 7. How to Install
The following are utilized to perform Pixel Correction. [Figure7-61]
Figure 7.61. Pixel Correction Icon
Pixel Correction is performed in the following process.
Display the Pixel Defect in the Pixel Viewer
Select the appropriate shortcut icon
Ctrl+Left Click the area of Defect in the Pixel Viewer
Figure 7.62. Pixel Correction (Horizontal Line Defect)
10-EVS1MADB001 79
EVS 3643 User’s Manual 7. How to Install
The selected Defect will be displayed in green. The selected Defect will be displayed in blue if the Defect is already saved.
Figure7.63. Example Post Performance of Pixel Correction
Perform Pixel Corrections for all existing Defect and save to complete the process.
Figure 7.64. Save Pixel Map
10-EVS1MADB001 80
EVS 3643 User’s Manual 7. How to Install
7-12 Configuration
The Configuration will appear through Option Menu - Configuration.
Figure7.65. Configuration
: Configuration detailed tab
: Save the modified contents and close the window.
Automatically restart the program when necessary.
: Close the window without saving the contents.
10-EVS1MADB001 81
EVS 3643 User’s Manual 7. How to Install
7-12-1 General
Configure the number and type of Detector being utilized. In addition, Configuration matters required in the process of calibration may be modified.
Figure 7.66. General
10-EVS1MADB001 82
EVS 3643 User’s Manual 7. How to Install
1) Detector Configuration
Figure 7.67. Detector Configuration
Table 7-10. Model Configuration Descriptions
Category Descriptions
Type of Panel Model which will be added or modified
Type of Panel added (registered)
Adds the Panel Model selected in ①
Modifies the Model of the Panel selected in ②
Here, the Model is replaced by the Model selected in ①
Deletes the most recently registered Panel.
(With disregards to selection, the item at the bottom of the list will be deleted.)
10-EVS1MADB001 83
EVS 3643 User’s Manual 7. How to Install
2) Calibration Configuration
Figure7.68. Calibration Configuration
Table 7-5. Calibration Configuration Descriptions
Category Descriptions
Select whether to set the Grain Map as Filtered Gain Map or Gain Map.
Configures the number of images to be project when performing Gain/Grid
Calibration. The Default value is 7 counts.
Checks the X-ray radiation dose when performing Gain Calibration and provides
an option for re-projection if the radiation dose is unfit.
It is recommended for accurate Calibration.
Configures the interval between each image projections when performing
Gain/Grid/Panel Calibration. The default value is 60 sec.
10-EVS1MADB001 84
EVS 3643 User’s Manual 7. How to Install
7-12-2 Options
Figure 7.69. Options Configuration
Table 7-6. Options Descriptions
Category Descriptions
Configures the use of Auto Update function.
Configures the use of the virtual keyboard for Touch environments.
Selects the editing method utilized during the Pixel Map Calibration process.
Selects the port connected with USB Switch Box.
Checks ‘Fix Min / Max’ and inputs Min and Max pixel value to use fixed window
level.
85
EVS 3643 User’s Manual 7. How to Install
7-13 Detector
Allows the modification of necessary configuration for each Detector. Number of tabs corresponding to the number of registered Panels through Configuration – General will be
generated
The following may be configured through Panel Configurations
.
Figure 4. Configuration – Changes in the Number of Panel Tabs
Map files save path
Save path of the Map file required for Calibration
Edge cut setting
The boarders of the image
Grid selection
Configurations relevant to the type of Grid and Filter Type being used
F600 Model Configuration
Configurations only for F600
10-EVS1MADB001 86
EVS 3643 User’s Manual 7. How to Install
7-13-1 Map File Path
Figure7.71. Map File Path
Table 7.7. Map File Path Descriptions
Category Descriptions
Configures the save path of the Pixel Map file.
The File Dialogue may be opened by selecting ‘…’ on the right.
Configures the save path of the Gain Map file.
The File Dialogue may be opened by selecting ‘…’ on the right.
Configures the save path of the Panel Map file.
The File Dialogue may be opened by selecting ‘…’ on the right.
Previous Map files are backed-up when Map files are saved multiple times.
Utilized when restoring the Map file with the previous file.
However, restore cannot be processed if a backed-up Map file does not exist.
10-EVS1MADB001 87
EVS 3643 User’s Manual 7. How to Install
7-13-2 Edge Cut Setting
Figure 7.72. Edge Cut Setting
Fixed placement of a boarder on the image. The Pixel value and width may be designated in this configuration.
Table 7.8. Edge Cut Setting Descriptions
Category Descriptions
Configures the Threshold Value of the image.
The default is determined by each model.
If the Pixel Value of the image imported from the Detector exceeds 11000, the
values will be automatically adjusted to 11000 if this value is set to 11000.
Configures the Pixel Value of the additional boarder which will be placed on
the image.
The default value is 11000.
Configures the width of the top, bottom, left and right boarder. Unit is in Pixel.
The Default value is all 10.
10-EVS1MADB001 88
EVS 3643 User’s Manual 8. Troubleshooting
When you encounter problems or error messages in the ECali1 while using this equipment, search the table below for the problem or error message and try the solutions. If the problem persists, turn off the detector and consult your sales representative or local DRTECH dealer. Please refer to the details of the problem or error messages.
Symptom
Failed to turn on. Power calbe is not attached Attach a power cable with SSU
LINK lamp does not light up
Cause/Error messages in
the EVS Control Software
The communication circuit is not secured
Remedy
Register the detector and make connection to the ECali1. In wireless configuration, confirm the following conditions:
• There is no obstacle between the wireless module of the detector and the wireless access points.
• The wireless access point is turned on. In wired configuration, confirm the following conditions:
• The detector cable is firmly connected to the SSU(power box).
• The SSU(power supply) and the AP are firmly connected via cable.
• The AP is turned on and network communication is functioning properly.
In wireless configuration, confirm the following conditions:
The access point is turned on.
Wireless communication status is stable
While the POWER lamp
Data transmission error has occurred.
lights, READY and LINK lamps flash.
The access point are firmly connected via cable.
In wired configuration, confirm the following conditions:
The detector cable is firmly connected to the detector.
The SSU is turned on.
Exposure preparation error has occurred. / Error Code: 21
Confirm that the cables with the X-ray interface unit are connected securely and the X-ray interface box is turned on.
All the LED lamps flash. Two LED lamps flash and Turn off the detector and
Hardware error has occurred /Error Code: other than 21
Turn off the detector and turn it on again.
turn it on again. the other flashes slowly.
10-EVS1MADB001 89
EVS 3643 User’s Manual
9. Maintenance
Maintenance and inspection
In order to ensure that the equipment is used safely and normally, be sure to inspect the equipment before use. If any problem is found during the inspection and cannot be corrected, please contact your sales representative or local DRTECH dealer.
 Daily inspection
WARNING
Cable
1) Ensure that cables are not damaged and cable jackets are not torn.
2) Ensure that the power cord plugs are securely connected to both the equipment AC inlet and the AC
outlet
.
Detector
For safety reasons, be sure to turn OFF the power to each piece of equipment before the following. Otherwise, an electric shock may result.
1) Ensure that there are no loose screws or breaks.
2) Ensure that there is no dust or foreign matter on the external connector.
3) Ensure that there are no breaks or short-circuits in the power supply conector.
After turning on the power
Be sure to start the ECali1 before performing the following inspection.
1) Perform test exposure
.
 Monthly inspection
1) Conduct a Performance Test.
2) Regularly conduct a Self-diagnosis. For details, refer to the Setup Guide for the EVS Control Software.
 Yearly inspection
1) Perform a Performance Test or Self-diagnosis using a phantom or resolution chart, etc.
 Irregular inspection
Calibration
1) Perform Calibration when exposure conditions have changed significantly. For details, refer to the Setup Guide for the ECali1.
10-EVS1MADB001 90
EVS3643 User’s Manual
10-1. Main specifications
EVS 3643 X-ray detector
10. Specifications
[Dimensional diagram]
(unit mm)
10-EVS1MADB001 91
EVS 3643 User’s Manual
SSU (System Syncronization Unit)
10 Specifications
[Dimensional diagram]
(unit mm)
Rated power supply Input: 100-240 VAC, 50/60Hz Output: DC +12 V 8.3A, 75W
10-EVS1MADB001 92