Penguin Pediatric Compressor Nebulizer
Nebulizador Compresor Pediátrico Pinguino
Compresseur Nébuliseur Pediatrique Manchot
MQ6002, MQ6002R, MQ6006
2
Penguin Pediatric Compressor Nebulizer
INDEX
1. Introduction
Product Identication
Important Safeguards
Operating Your Compressor Nebulizer
Cleaning
Replacement Parts & Accessories
Specications
Warranty
2.
------------------------------------------------------
------------------------------------------------------
--------------------------------
---------------------------------------
------------------------------------------------------------------------
-----------------------------------------------------------------
------------------------------------------------------------------------
------------------------------------------------------------------
3.
4.
5.
6.
7.
8.
1. INTRODUCTION
Nebulizer. This product is a compact medical device designed to eciently deliver
Thank you for purchasing the Drive Medical AIRIAL™ Penguin Pediatric Compressor
physician prescribed medication to the bronchial lung passages for the treatment of asthma, allergies, COPD, and other respiratory ailments. With proper care and use, it will provide you with many years of reliable treatment.
The AIRIAL
TM
is intended for use with a single adult, pediatric, or infant patient. This product creates a stream of air that travels through a clear tube to the nebulizer. When air enters the nebulizer, it will convert the prescribed medication into aerosol mist for easy inhalation.
Your AIRIAL
TM
should be used on the order of physician. We encourage you to read this guidebook thoroughly to learn about the features of this product. Any use of this product other than its intended use should always be avoided.
Page 2
Page 3
Page 3
Page 4
Page 5
Page 6
Page 7
Page 8
INSTRUCTIONS
2. PRODUCT IDENTIFICATION
3. IMPORTANT SAFEGUARDS
Note: Read all instructions carefully before use.
The following basic precautions are needed when using an electrical product: CAUTION: Failure to read and observe all precautions could result in personal
injury or equipment damage.
1. To avoid electrical shock: • Keep the unit away from water • Do not immerse the power cord or the unit in liquid • Do not use while bathing • Do not reach
for a unit that has fallen into water – immediately unplug the unit.
2. Never operate the unit if it has any damaged parts (including power cord), or if it has been dropped or submersed in water. Promptly send it to a service center for examination and repair.
3. The unit should not be used where products are being used.
4. Disconnect the unit from the electrical outlet before clea
after each use.
5. Do not use other attachments unless recommended by manufacturer.
6. This appliance has a polarized plug (one blade is wider than the other).
ITEMS
MQ6002
MQ6002R
MQ6006
Penguin Pediatric Nebulizer with Disposable Neb Kit & Carry Bag Penguin Pediatric Nebulizer with Reusable & Disposable Kit & Carry Bag Penguin Pediatric Nebulizer with Disposable Neb Kit, No Carry Bag
DESCRIPTIONS
Product Cautions:
3
4
1. Connect this product to an appropriate voltage outlet for your model.
To reduce the risk of electric shock, this plug is intended to t into a polarized outlet only one way. If the plug does not t fully into the outlet, reverse the plug. If it still does not t, contact a qualied electrician. Do not attempt to modify the plug in any way.
2. Do not run this product unattended.
3. Never operate the unit if it has any damaged parts (including power cord),
or if it has been dropped or submersed in water. Promptly send it to a service center for examination and repair.
1. Do not store the unit in direct sunlight, high temperatures, or high humidity.
2. Keep the unit out of reach of small children.
3. Always keep the unit unplugged while not in use.
Cleaning Cautions:
Storage Cautions:
Operating Cautions:
1. Do not immerse the unit in water. It may damage the unit.
2. Disconnect the unit from the electrical outlet before cleaning.
3. Clean all necessary parts after each use as instructed in this guidebook.
4. OPERATING YOUR COMPRESSOR NEBULIZER
Important: Prior to initial operation, this product should be thoroughly cleaned per the “Cleaning Procedures” in this guidebook.
2. Gently twist the top part of the nebulizer cup counter-clockwise to
diassemble the nebulizer cup.
1. Place your compressor Nebulizer on a at and stable surface. Be sure that
you can easily reach the controls when you are seated.
3. Fill the bottom section of the nebulizer
cup with the medication prescribed by your Physician. Be sure that the cone is put inside the bottom section.
Place the cone inside and ll medication.
4. Gently twist the top part clockwise to
reassemble the nebulizer cup. Be sure the two sections t well.
5. Always unplug the product immediately after use.
6. Never block the air openings of the main unit or place it where the air
openings may be obstructed.
4. If any abnormality occurs, discontinue use immediately until the unit has
been examined and repaired.
5
5. Attach one end of air tube to the base of the nebulizer cup.
6. Attach the other end of airtube to the air tube connector located on the back of the compressor.
7.
Attach the interface accessory of your choice to the top section of the nebulizer cup.
8. Plug the power cord into an appropriate electrical outlet. Make sure at this stage, the power switch is at “OFF” status.
9. Press the power switch to “ON” status to begin your prescribed treatment.
10. When treatment is nished, shut o the unit and unplug it from the electrical outlet.
Important:
The compressor motor has a thermal protector, which will shut o the unit before the unit is overheated. When the thermal protector shuts the unit o, please:
b. Unplug the unit from the electrical outlet.
a. Switch o the unit.
c. Wait 30 minutes for the motor to cool down before another treatment. d. Make sure the air openings are not obstructed.
5. CLEANING
It is recommended that the nebulizer and the mask (or any other patient accessory interface that is used) be thoroughly cleaned with hot water after each use and cleaned with a mild detergent after the last treatment of the day. If your physician or respiratory therapist species a dierent cleaning procedure, follow their instructions.
CAUTION:
Before cleaning, make sure the power cord is disconnected from the main outlet.
Rinsing (after each treatment)
1. Disconnect the air tube from the nebulizer cup. Disconnect the mask or other patient accessory interface from the nebuilzer cup.
2. Gently twist the top part of the nebulizer cup counter-clockwise to disassemble the nebulizer.
3. Rinse the nebulizer cup and mask (or other patient interface) with water.
4. Let these components air dry, away from direct sunlight.
5. Reassemble the nebulizer cup when completely dry and put these parts in a dry, sealed container.
6
Disinfecting
Cleaning the Compressor
Please follow the following steps to disinfect your nebulizer unless otherwise sp
d by your physician. It is suggested that the unit is disinfected after the
last treatment of the day.
Use one part white vinegar with three parts distilled water. Make sure mixed solution is enough to submerge the nebulizer, mouthpiece and mask. Complete Rinsing Steps 1-3. Wash nebulizer, mouthpiece and mask in warm water and a mild detergent. Then wash them in hot tap water.
Submerge these parts in the vinegar and water solution for thirty minutes. Complete Rinsing Steps 3-5.
2.
1.
3.
4.
5.
2.1.Do not use any powdered cleaners or soap pads, which may damage the nish.
Wipe daily with a damp cloth.
Filter change
CAUTION:
Do not operate the compressor without a lter. Only use lters supplied by your distributor. Do not wash or clean the lter.
1. Changing proc
edure:
A
. B. C. Replace the lter cover.
Replace the used lter with a new lter.
Remove the lter cover.
6. REPLACEMENT PARTS AND ACCESSORIES
Contact your healthcare distributor for replacement parts or accessories.
7
7. SPECIFICATIONS
Medication Capacity
Power Current
Particle Size MMAD
0.5 to 10μm
4 μm Sound Level Maximum Medication capacity Average Nebulization Rate Compress or Pressure Range 25 to 42 Psi (0.17 to 0.29 Mpa) Operating Pressure Range 8 to 13 Psi (0.056 to 0.090 Mpa) Liter Flow Range 5 ~ 8 lpm
10˚C to 40˚C (50˚F to 104˚F)
10 to 95% RH
-20˚C to 70˚C (-4˚F to 158˚F)
Operating Temperature Range Operating Humidity Range
Storage Temperature Range
Storage Humidity Range Dimension (L x W x H) Weight 3.52 lbs (1.60 kg)
188. 0 × 169. 5 × 136.5mm
10 to 95 % R H
Standard Accessories
Nebulizer, Air Tube, Filters (5pcs),
Mouthpiece, Child Mask
Protection against electric shock
- Class II Equipment
- Type BF applied parts:
- Mouthpiece, mask
Protection against harmful ingress of water
- IPX0 (Ordinary)
Mode of operation—20 minutes on & 40 minutes o
- No AP/APG (Equipment not suitable for use in the presence of ammable anesthetic mixture with air. Not intended for use with oxygen or in oxygen enriched environments.)
Degree of safety in the presence of ammable anesthetics or oxygen
5ml
AC 120V, 60Hz / 230V, 50Hz
1.5 A / 0.8 A
Around 60 dBA
13ml (cc)
0.2ml/min