Drive Voyager Pro User Manual

18018
Voyager
Pro Nebulizer
Federal (US) Law restricts this device to sale by or on the order of a physician.CAUTION:
01
Getting acquainted 02 Safety information and warnings 02 Important information on the device 07
Accessories
Protecting against leaks
Automatic switch-off
Know Your Voyager™ Pro Nebulizer 08 Initial use 09
Before using the device for the first time
Assembly
Battery Replacement
Operating the device with the switching power supply
21 noitarepO
Preparing the nebulizer
Filling the nebulizer
Closing the nebulizer
Fitting accessories
Inhaling correctly
Cleaning and disinfection 15
Warning
Cleaning
Disinfection
Disassembly
Drying
Material resistance
Storage
How to replace the mesh nebulizer 20
12 gnitoohselbuorT
32 snoitacificepS Manufacture Date 24 Warranty Information 25
Getting acquainted
The Voyager™ Pro nebulizer is a vibrating mesh nebulizer for treating the upper and lower respiratory tract.
The nebulization and inhalation of medication prescribed or recommended by a doctor can prevent against diseases of the respiratory tract, treat the side effects of these disorders and accelerate healing.
For more information on possible applications, consult your doctor or pharmacist. The device is suitable for inhalation at home or on the move. Medicines should only be inhaled when instructed by a doctor.
Inhalation should be performed in a calm and relaxed atmosphere. Inhale slowly and deeply to enable the medicine to reach the small bronchi deep into the lungs. Exhale normally.
02
Area of application
Safety information and warnings
Warning
Before use, ensure that there is no visible damage to the device or accessories.
When in doubt, do not use the device and contact your equipment provider.
Using the device is not a substitute for consultation with or treatment from a
physician. Whenever you have any pain or illness, always contact your physician first.
If you have health concerns of any kind, contact your general practitioner!
When using the nebulizer, always follow the basic hygiene procedures.
Always follow your doctor’s instructions regarding the type of
medicine to
be used and the dose, frequency, and duration of inhalation.
Only use medicines that have been prescribed or recommended by your
doctor or pharmacist.
03
If the device does not function correctly or if you begin to feel unwell or feel
pain, stop using the device immediately.
During use, hold the device away from the eyes, as some nebulized
medicines may cause damage to the eyes.
Do not operate the device in the presence of flammable gases.
Do not use the device near electromagnetic transmitters.
This device is not intended to be used by children or persons with restricted
physical, sensory (e.g. insensitivity to pain) or mental abilities or persons lacking in the required experience or knowledge for safe operation of the device, unless supervised or instructed by a person responsible for their safety.
Keep children away from packaging materials (risk of suffocation).
Do not use attachments that are not recommended by the manufacturer.
Do not dip or submerge the device into water and do not use it in wet
rooms. Do not allow any liquid to penetrate the device.
Protect the device from strong impacts.
Do not use the device if it has fallen or been dropped on the floor, been
exposed to extreme humidity or is otherwise damaged. In case of doubt, contact our customer service department or your local dealer.
The Voyager
Pro Nebulizer must only be operated along with the appropriate Drive Medical accessories. The use of other accessories can impair therapeutic efficiency and may damage the device.
Remove the batteries from the device if you will not be using it for an
extended per
iod. Leaking batteries may damage the device.
Always keep mesh nebulizer in the device. A drop or fall in 4 in (10 cm)
height may cause it broken or damage.
Always dry up the device, including mesh nebulizer and medicine container.
Oxides may block the mesh hole and impair the nebulizer efficiency.
04
Important
Power loss, sudden faults or other unfavorable conditions may render the
device unusable. You are therefore recommended to keep a spare device or replacement medicine (as agreed with the doctor).
The device must be stored away from sources of heat.
Do not use the device in a room in which a spray has previously been used.
If this is the case, ventilate the room before commencing treatment.
For reasons of hygiene , each user should use their own set of accessories (mesh
nebulizer, mask and mouthpiece).
Store the device in a location protected against environmental influences.
The device must be stored in the specified ambient cond
itions.
Do not use any medicines in powder (even if dissolved).
Do not shake the nebulizer, as this may cause liquid to leak, thereby
damaging the device.
05
General notes
Important
Use the device only:
On humans
For its intended purpose (aerosol inhalation) and in the manner described
in these instructions for use.
Any improper use can be dangerous!
In acute emergencies, first aid takes priority.
In addition to medicines, use only a saline solution.
This device is not intended for commercial use, but only for individual
use.
Before using the device for the first time
Important
Before you use the device for the first time, remove all packaging materials.
Protect the device from dust, dirt and moisture.
Do not use the device in very dusty environments.
Before use, ensure the device and accessories are at room temperature.
Switch the device off immediately if it is defective or malfunctioning.
The manufacturer shall not be held liable for damage or injuries caused by
improper or incorrect use.
06
Notes on handling batteries
Keep batteries out of the reach of children. Children can put batteries in their
mouths and swallow them. This can cause severe harm to their health. In this case, consult a physician immediately!
Normal batteries must not be recharged, heated or thrown into open flame
(risk of explosion).
Leaking batteries may damage the device. If you do not intend to use the
device for prolonged periods, remove the batteries from the battery compartment.
Repairs
Note
Never open or attempt to repair the device by yourself, as otherwise proper
function is no longer guaranteed. Failure to observe this regulation shall void the warranty.
If you need to have the device repaired, contact your equipment provider.
Caution
Leaking or damaged batteries can cause burns upon contact with the skin. When removing such batteries, wear suitable protective gloves.
Important information on the device
The safe functioning of the device can only be guaranteed if the accessories recommended by the manufacturer are used.
Accessories
When filling the medicine container with medicine, ensure that you do not exceed the maximum mark (8 ml). The recommended filling amount is between 2 and 8 ml.
During use, you can tilt the device up to a maximum angle of 45˚ in any direction, without impairing its nebulizing function or success of the treatment. However, spraying only takes place if the substance to be sprayed is in contact with the mesh. If it is not, spraying stops automatically.
If this happens, please try to hold the device as upright as possible.
Protecting against leaks
The device has an automatic switch-off function. To prevent damage to the mesh, the devic
e switches off automatically when the liquid has been
almost used up. Do not use the device if the medicine container or water tank is empty. The device switches off automatically if the substance to be sprayed is no
longer in contact with the mesh.
Automatic switch-off
07
Caution