DRE AV-S User manual

AV-S Ventilator
User Manual
This manual contains information for software upgrade v.1.87.01
DATASHEET
USER
AV-S Ventilator User Manual
Software Version 1.87.01
Addendum to the User Manual
Keep this Datasheet with the User Instruction Manual for AV-S at all times
AVS software V. 1.87.01 introduces a new user interface. However, if you are familiar with the AV-S, please note that the basic operation and calibration procedures remain as per v.1.86.
Modifications to the User Interface
Setting up the Ventilator for use
1. Introduction Screen
1.1 Start-up
At start-up, the introduction screen allows the user to select one of three default settings:
SITE DEFAULTS
ADULT DEFAULTS
PEDIATRIC DEFAULTS
NOTE a) The user must select one of the above default groups before the ventilator will switch to standby in that default mode
b) SITE DEFAULT is editable in standby mode (see section 1.2, below)
c) Settings can be saved via the service menu to create a new site default
1.2 Default Settings
1.2.1 Selection The user can select ADULT, or PEDIATRIC, or SITE, and view the default parameter settings. The options will remain, even after the ventilator is turned off.
1.2.2 Site Default Settings Adjust the parameter values from within the Service menu (SITE DEFAULTS) Press to confirm the new settings for site defaults.
1.3 Calibrate Touchscreen
The introduction screen allows the user to calibrate the screen
2. Parameter Display Identification
2.1 Active Parameters
Active parameters that can be set for use in the current mode are displayed as:
White Text on Blue
2.2 Inactive Parameters
Inactive parameters that can be set for any non-current mode are displayed as:
White Text on Blue Label White values on Black
2.3 Measured Parameters
Yellow values on Black
2.4 T+PS INIT (target and pressure support initial value)
The initial pressure value can be changed so that when entering either PRESSURE or PSV modes the TARGET value or PSUPP value are pre-selected.
NOTE Changing either of these limits in their active modes will maintain the value when changing between PSV, PRESSURE, and STANDBY modes.
3. Gas Mixture
The Gas Mixture window is an active touch-selectable area (in any mode), with a drop down menu. Gas Mixture is also available through the menu structure. Selection of the required mixture is in the normal way with the scroll wheel.
Using the Ventilator - description of modes and functions
4. Modes
4.1 Access to Support Modes
Access is available in Standby mode (depending on the support mode options on the ventilator).
Support Mode
a) PSV
b) SIMV
c) SMMV
d) SIGH ENABLE
SIGH TO BREATH RATIO
e) INSP PAUSE
INSP PAUSE %
WARNING Modes a, b, and c are only available when Spirometry is enabled.
2
4.2 Standby Mode
a) Standby mode at ventilator start-up: The last used Volume mode settings will be displayed
b) Standby mode selected while the ventilator is in use: The screen will display the previous ventilation mode, highlighted in yellow, within the relevant box. The last used parameters will also be displayed.
4.3 Spontaneous Mode
a) Spontaneous mode at ventilator start-
up: Default values will be displayed in white on a black background if the ventilator has just been powered ON.
b) Spontaneous mode selected while the
ventilator is in use: The last used ventilation mode (underlined) will be displayed, with the last used set values in white on a black background
4.4 Sigh
Sigh is settable from 1:n, where n has a range of 10 to 100. The Sigh menu can also be accessed by touching the icon area of the screen.
NOTE 1:10 is one sigh to ten normal breaths.
4.5 Inspiratory Pause
Inspiratory pause can be varied in the menu from 0 - 60%. The inspiratory pause menu can also be accessed by touching the icon area of the screen.
WARNING This can affect the maximum Tidal Volume.
3
5. Apnoea Alarm Mute - Spontaneous mode only
NOTE The occurrence of another alarm event will override this feature
In spontaneous mode the mute button acts both to silence an existing apnoea alarm and inhibit new apnoea alarms for a given period (provided that no other alarm events are present) .
This time period is selectable (choose from 15, 30, 60, 120, or 180 seconds) through the alarm settings menu, or accessed by touching the alarm area of the screen.
To adjust the default setting, use the SITE DEFAULT menu option.
6. Touchscreen Access to Mode Configuration Options
Touch the screen in the area containing the green icons to access mode configuration options (including INSP PAUSE, SIGH, and APNOEA ALARM mute/inhibit).
7. Waveform Pause and Print
Waveform pause and print icons are located to the left hand side of the waveform displays.
Ensure that a compatible printer is connected, and switched On (see section
5.1.8).
To print the waveform information, press the pause icon. The print icon will be displayed. Press the icon to print. Press the pause icon to unfreeze the waveform.
8. Waveform Freeze Loop
The FREEZE LOOP icon is located at the left hand side of the top waveform.
9. Leak Test
a) Select LEAK TEST through the Menu in Standby Mode.
b) With the bag/vent switch in VENT position, this checks for a leak using an occluded breathing system
The leak test procedure is given in section 5.1.12 in the user manual.
4
Modifications to Operational Envelope
(sections 4.5 and 4.6 in user manual)
Flow Range: 2 – 70 Litres per min.
Volume Range: 20 ml – 1.6 Litres (tidal)
2 – 50 Litres per min (minute vol.)
Rate 4 – 100 bpm
I:E Ratio 1:0.2 – 1:8.0 (normal)
1:2.0 – 1:8.0 (effective in support modes)
Inspiratory Time 0.3 – 10 seconds (normal)
0.3 – 5 seconds (effective in support modes)
ExpiratoryTime 0.3 – 10 seconds
(effective dependent on Inspiratory time)
Additional information on Sigh and Inspiratory Pause
(section 3.7.2.4)
Inspiratory Pause
Inspiratory Pause has a range of 0 to 60% of the inspiratory time.
Sigh
This function must be enabled in the mode menu but is only operational in volume ventilation. The sigh ratio is 1 to n (1:n) with n giving a range of 10 – 100 breaths
between sighs.
5
EXIT MENUS
O2 MONITOR & SPIROMETRY
LEAK TEST
FRESH GAS COMPENSATION:ON
MODES
WAVEFORM
ALARM SETTINGS
GAS MIXTURE: O2+AIR
SERVICE MENU
O2 Monitor & Spirometry
ESCAPE FROM MENU O2 MONITOR: on CALIBRATION: 100% HIGH ALARM SET: 105 LOW ALARM SET: 18 SPIROMETER: on SPIRO CALIBRATION: 0 L/min
off / on (Toggle option 21 / 100% (Toggle option) 19 -105 (Integer) 18 - 99 (Integer) off / on (Toggle option) 0 L/min / 10 L/min (Toggle option)
off/on (Toggle option)
Fresh Gas Compensation
ON / OFF
Special Modes
See next page
Waveform
ESCAPE FROM MENU SECOND WAVEFORM: off
Second waveform pick list
off vol. vs time vol. vs press.
Alarm settings
ALARM MENU ESCAPE FROM MENU ALARM MODE : default HIGH TIDAL VOLUME: off VM MIN: 3 L VM MAX: 9 L VT MIN: 300 mL VT MAX: 900 mL APNOEA ALARM LIMIT: 15 secs ALARM VOLUME: 50%
default / user (Toggle option) off / on (Toggle option)
0.0 - 7.4 (Integer)
0.1 - 7.5 (Integer) 10 - 1600 (Integer) 20 - 2400 (Integer)
0.3 - 3.5 (Integer) 50 - 100% (Integer)
Gas mixture: O2+Air
O2+AIR O2+N2O
Leak Test
ESCAPE FROM MENU <START/STOP LEAK TEST> LEAK STATUS: unknown LEAK LEVEL: 0 mL/min BSYS COMP 7.0 mL/cmH2O
Service
See page 62
Menu Structure
Main Menu
6
The SPECIAL MODES menu is context sensitive, with the contents dependent on current mode.
In STANDBY the SPECIAL MODES menu is:
ESCAPE FROM MENU
SUPPORT MODE: SIMV, SMMV, PSV
(1)
VOLUME TYPE: Tidal
SIGH ENABLE:
(2)
SIGH TO BREATH RATIO:
INSP. PAUSE% : 0%
APPLY: SITE DEFAULT
In SPONT mode and VOLUME mode, and SIMV/ SMMV, the SPECIAL MODES menu is:
ESCAPE FROM MENU
VOLUME TYPE: Tidal
SIGH ENABLE:
(2)
SIGH TO BREATH RATIO:
INSP. PAUSE% : 0%
In PRESSURE mode and PSV modes the SPECIAL MODES menu is:
ESCAPE FROM MENU
SIGH ENABLE:
(2)
SIGH TO BREATH RATIO:
INSP. PAUSE% : 0%
Notes (1) Support mode depends on configuration options.
The SUPPORT MODE option will be missing from the SPECIAL MODE menu if: a) Options are not enabled b) ‘’SPIROMETRY: off’’ is displayed.
The support mode sub menu can include:
none / PSV / SIMV / SMMV
(2) The options here are: 0 - 60%
(3) The options here are:
on - off 1:10 to 1:100
Note 1:10 indicates 1 breath with sigh, then 10 breaths without sigh
(2) The TRIGGER values are L/min with SPIROMETRY enabled, or
cmH
2O when SPIROMETRY disabled.
Spirometry enabled Spirometry disabled
0.7 L/min 0.5 cmH
2O
0.8 L/min 0.6 cmH
2O
0.9 L/min 0.7 cmH
2O
1.0 L/min 0.8 cmH
2O
1.5 L/min 0.9 cmH
2O
2.0 L/min 1.0 cmH
2O
2.5 L/min 1.2 cmH
2O
3.0 L/min 1.5 cmH
2O
3.5 L/min 1.7 cmH
2O
4.0 L/min 2.0 cmH
2O
SPECIAL MODES MENU
ESCAPE FROM MENU
SUPPORT MODE: SIMV, SMMV, PSV
VOLUME TYPE: Tidal
SIGH ENABLE:
SIGH TO BREATH RATIO:
INSP. PAUSE% : 0%
APPLY: SITE DEFAULT
7
UPGRADE MENU
ESCAPE FROM MENU
I/O HARDWARE: 2
I/O FIRMWARE: vx.xx [Build xx]
MAIN FIRMWARE: vx.xx [Build xx]
REGISTRATION KEY: unknown
UPGRADE FIRMWARE: unavailable
ADD NEW FEATURE: unavailable
DISPLAY HISTORY
ESCAPE FROM MENU
MANUFACTURER DATE : 03/03/05
TOTAL HOURS RUN: 100
LAST SERVICE DATE: 13/08/04
HOURS SINCE SERVICE: 100
DRIVE VALVE CYCLES: 1253
PATIENT VALVE CYCLES: 822
CUTOFF VALVE CYCLES: 72
Service
ESCAPE FROM MENU
LANGUAGE: ENGLISH
PATIENT LOG MENU
SITE DEFAULTS
SERIAL MODE: none
ABSORBER SWITCH; ON
CLOCK MENU
UPGRADE MENU
AMBIENT PRESSURE: 988 mBar
DISPLAY HISTORY
*SERVICE PIN: 0
*ENGINEER MENU
CLOCK MENU
ESCAPE FROM MENU
YEAR: 2005
MONTH: 3
DATE: 16
HOUR: 9
MINUTE: 57
UPDATE CLOCK
DAYLIGHT SAVING: off
PATIENT LOG MENU
ESCAPE FROM MENU
PRINT PATIENT DATA
LOGGING: off
LOG STATUS: disabled
CLEAR LOG DATA
LOGGING WINDOW: 10 min
Clock pick list (integer)
2005 - 2099 (integer) 1 - 12 (integer)
1 -31 (integer)
0 - 23 (integer)
0 - 59 (integer)
off / on (toggle
option)
*NOTE Sub-menus for Service PIN and Engineer Menu are not accessible by users.
SERVICE MENU
SITE DEFAULTS
ESCAPE FROM MENU
SAVE TO SITE
VIEW: SITE DEFAULTS
VOLUME TYPE : tidal
Vt SET: 550 ml
Vm SET: 5.5 Litres
T+PS INIT: 10 cmH2O
SET BPM : 10
I : E : 1:1.0
PEEP : OFF
LIMIT : 38 cmH2O
TRIGGER : 10 L/min
APNOEA ALARM LIMIT : 15 Sec
BACK LIGHT LEVEL : 50 %
Doc No. AVS 0408DS (U) September 2008
DRE, Inc. 1800 Williamson Court Louisville, KY 40223 USA
Tel: (502) 244-4444
Fax: (502) 244-0369
Web: www.dremed.com
IMPORTANT
(i)
Servicing and Repairs
In order to ensure the full operational life of this ventilator, servicing by an engineer trained by the manufacturer should be undertaken periodically.
The ventilator must be serviced to the following schedule:
(a) Six monthly service - inspection and
function testing.
(b) Annual / two year / four year service -
inspection and function testing, and component replacement.
Details of these operations are given in the Service Manual for the AV-S, available only for engineers trained by the manufacturer.
For any enquiry regarding the servicing or repair of this product, contact DRE, Inc.
Technical Support
DRE, Inc. 1800 Williamson Court Louisville, KY 40223
USA
Tel: (502) 244-4444 Fax: (502) 244-0369 Web: www.dremed.com
Always give as much of the following information as possible:
1. Type of equipment
2. Product name
3. Serial number
4. Approximate date of purchase
5. Apparent fault
FOREWORD
(ii)
Parameter / Device Relevant Standard
Pressure Measuring ISO 8835-2 Pressure Limitation Device EN 60601-2-13:2006 - 51.101.1 Exhaled Volume Monitor EN 60601-2-13:2006 - 51.101.4 Breathing System Integrity Alarm System EN 60601-2-13:2006 - 51.101.5 Continuing Pressure Alarm EN 60601-2-13:2006 - 51.101.6 Oxygen Monitor ISO 7767 Carbon Dioxide Monitor ISO 9918 Breathing Circuit ISO 8835-2 Agent Monitor ISO 11196 Gas Scavenging ISO 8835-3
For information on installing and connection of any of these systems or devices, please refer to the relevant manufacturer’s instructions.
This manual has been produced to provide authorized personnel with information on the
function, routine performance and maintenance checks applicable to the AV-S Anesthesia Ventilator.
Information contained in this manual is correct at the date of publication. The policy of the manufacturer is one of continued improvement to its products. Because of this policy, the manufacturer reserves the right to make any changes which may affect instructions in this manual, without giving prior notice.
Personnel must make themselves familiar with the contents of this manual and the machine’s function before using the apparatus.
The Importance of Patient Monitoring
WARNING Anesthetic systems have the capability to deliver mixtures of gases and vapours to the patient which could cause injury or death unless controlled by a qualified anesthetist.
There can be considerable variation in the effect of anesthetic drugs on individual patients so that the setting and observation of control levels on the anesthesia systems does not in itself ensure total patient safety. Anesthesia system monitors and patient monitors are very desirable aids for the anesthetist but are not true clinical monitors as the condition of the patient is also dependent on his respiration and the functioning of his cardio-vascular system.
IT IS ESSENTIAL THAT THESE ELEMENTS ARE MONITORED FREQUENTLY AND REGULARLY AND THAT ANY OBSERVATIONS ARE GIVEN PRECEDENCE OVER MACHINE CONTROL PARAMETERS IN JUDGING THE STATE OF A CLINICAL PROCEDURE.
Before using any monitoring system or device, the user must check that it conforms to the relevant standard, as listed in the table below.
CONTENTS
(iii)
Page No.
USER RESPONSIBILITY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1. WARNINGS AND CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2. PURPOSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3. DESCRIPTION
3.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Ventilation Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Pneumatic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.3.1 System Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4 Electrical System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 14
3.5 Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.5.1 Touchscreen Operation and Navigator wheel / push-button . . . . . . . . . . . . 15
3.5.2 User Adjustable Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.5.3 Operational capability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.5.4 Output Compensation Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.6 Interface with Integra AV-S and A200SP . . . . . . . . . . . . . . . . . . . . . . 20
3.7 Ventilation Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.7.1 Standby Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.7.2 Volume Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.7.3 Pressure Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.7.4 Spontaneous Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.7.5 Advanced Spontaneous Breathing Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.7.5.1 SIMV (Synchronised Intermittent Mandatory Ventilation) . . . . . . . . . . . . . 26
3.7.5.2 SMMV (Synchronised Mandatory Minute Ventilation) . . . . . . . . . . . . 27
3.7.5.3 PSV (Pressure Supported Ventilation) . . . . . . . . . . . . . . . . . . . . . . 28
3.7.5.4 PEEP ( Positive End Expiratory Pressure) . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.8 On-screen Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.9 Spirometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.10 Display Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.11 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.12 Oxygen Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.12.1 System Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.12.2 The Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.12.3 Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.12.4 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.12.5 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.12.6 Alarm Mute. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4. SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Oxygen Monitor . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
CONTENTS
(iv)
5. PRE-OPERATION PROCEDURES
5.1 Ventilator Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.1.1 Mounting the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.1.2 Electrical Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.1.3 Ventilator Gas Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.1.4 Breathing System Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.1.5 Bellows Drive Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.1.6 Anesthetic Gas Scavenging System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.1.7 Remote Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.1.8 Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.1.9 Breathing System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.1.10 Spirometer Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
5.1.11 Pressure Monitor Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
5.1.12 Leak Test / Compliance value Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . 52
5.1.13 Bellows Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
5.2 Pre-use Checklists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
5.2.1 Daily Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
5.2.1.1 Alarm System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
5.2.1.2 Ventilator Internal Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
5.2.1.3 Function Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5.2.2 Weekly Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
5.3 Oxygen Monitor Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
5.3.1 Installation . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
5.3.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
5.3.3 Sensor Low Indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
5.3.4 Setting the High and Low O
6. MAINTENANCE
6.1 Service Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
6.2.1 Outside Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
6.2.2 Bellows Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
6.2.3 Spirometer Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
6.2.4 Oxygen Monitor Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
6.2.5 Patient Connector Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
6.3 Sterilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.4 Oxygen Monitor Sensor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
7. APPENDIX
1. Back-up Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
2. Menu System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
3. Ventilator Spirometry System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
4. Disposal after use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
5. Approved Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
2 Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Statements in this manual preceded by the following words are of special significance:
WARNING means there is a
possibility of injury to the user or others.
CAUTION means there is a possibility
of damage to the apparatus
or other property.
NOTE indicates points of
particular interest for more efficient and convenient
operation.
Always take particular notice of the warnings, cautions and notes provided throughout this manual.
USER RESPONSIBILITY
1
This anesthesia ventilator has been built to conform with the specification and operating procedures stated in this manual and/or accompanying labels and notices when checked, assembled, operated, maintained and serviced in accordance with these instructions.
To ensure the safety of this device it must be checked and serviced to at least the minimum standards laid out in this manual. A defective, or suspected defective, product must not under any circumstances be used.
The user must accept responsibility for any malfunction which results from non­compliance with the servicing requirements detailed in this manual.
Additionally, the user must accept responsibility for any malfunction which may result from misuse of any kind or non­compliance with other requirements detailed in this manual.
Worn, broken, distorted, contaminated or missing components must be replaced immediately. Should such a repair become
necessary it is recommended that a request for service advice be made to DRE, Inc.
This device and any of its constituent parts must be repaired only in accordance with written instructions issued by the manufacturer and must not be altered or
modified in any way without the written approval of the manufacturer. The user of this equipment shall have the sole responsibility for any malfunction which results from improper use, maintenance, repair, damage or alteration by anyone other than the manufacturer.
USA and Canada: Federal Law restricts the sale and use of this device to, or on the order of, a licensed practitioner.
1. WARNINGS AND CAUTIONS
2
The following WARNINGS and CAUTIONS must be read and understood before using this ventilator.
WARNINGS
General Information
1. Personnel must make themselves familiar with the contents of this manual and the machine’s function before using the ventilator.
Before Using the Ventilator
2. Before the AV-S ventilator is used clinically for the first time a Calibration Check and Output Check must be successfully completed.
Calibration and output checks must be carried out by a DRE-trained technician, following the procedure in Appendix 6 in the AV-S Service Manual.
3. Before the ventilator is used clinically for the first time, verify that the hospital engineering department has carried out an earth continuity test. If the integrity of the protective conductor is in doubt, the ventilator must not be used.
Before the ventilator is used clinically for the first time, the commissioning engineer must confirm that the air/oxygen selection is set correctly for the drive gas that is to be used. The use of any other gas will cause inaccurate operation and may damage the ventilator, resulting in potential injury to the patient.
8. The driving gas is discharged through the opening in the back of the ventilator control unit. The discharged gas may contaminate the environment, and should therefore be extracted using a gas scavenging system.
9. The bellows can only support approximately 1 kPa (10 cmH differential positive pressure, above which it may be dislodged from the mounting ring, resulting in dangerous malfunction of the ventilator.
Do not connect a positive end expiratory pressure (PEEP) valve or other restrictive device to the exhaust port on the bellows base.
This would increase the pressure inside the bellows and the bellows could detach from the base, causing serious malfunction.
2O)
4. Excessive electronic noise caused by other poorly regulated devices, such as an electrocautery unit, may adversely interfere with the proper functioning of the ventilator.
To avoid this problem, do not connect the ventilator’s power cord into the same electrical wall outlet or adaptor strip into which an electrocautery unit is connected.
5. If used with a mains extension cord, the unit may be subject to electro-magnetic interference.
6. The driving gas supply must be clean and dry to prevent ventilator malfunction.
7. This ventilator is designed to be driven by oxygen or medical air only. The drive gas is set during manufacture and the ventilator is calibrated for that gas.
10. Breathing System
The breathing system which conveys gases from the anesthetic machine to the patient, and disposes of expired gases, must conform to the requirements of ISO 8835-2. Because breathing systems require frequent cleaning and disinfection they are not a permanent part of the anesthetic ventilator and therefore cannot be directly under the control of the anesthetic ventilator manufacturer. However, we strongly recommend that only breathing systems which have been approved and authorized by the manufacturer for use with AV-S should be employed.
Do not use conductive breathing system hoses.
When mechanical ventilation is employed the patient breathing system must be connected directly to a pressure relief valve to prevent the possibility of barotrauma.
WARNINGS AND CAUTIONS
3
11. The spirometer sensors are mounted within the A200SP absorber. Do not fit a spirometer sensor to any other location. The device will not measure exhaled volumes in any other position.
12. The operation of each alarm function should be verified daily.
Periodically check the alarms at clinically suitable intervals. If the audible alarm or the visual indicator of any alarm function fails to activate during any alarm condition or fails to reset after the alarm has been cleared, refer the unit to an authorized service technician.
13. Before using the ventilator check that all connections are correct, and verify that there are no leaks.
Patient circuit disconnects are a hazard to the patient. Extreme care should be taken to prevent such occurrences.
It is recommended that Safelock fittings are used throughout the breathing circuit.
Any problem arising from an improperly functioning scavenging system is solely the user’s responsibility. Do not use a scavenging system that restricts drive gas flow when negative pressure is exerted on it.
18. When the ventilator is connected to a patient, it is recommended that a qualified practitioner is in attendance at all times to react to an alarm or other indication of a problem.
19. In compliance with good anesthesia practice, an alternative means of ventilation must be available whenever the ventilator is in use.
20. It is recommended that the patient oxygen concentration should be monitored continuously.
21. If the drive gas supply pressure drops below a nominal 241 kPa (35 psi), the LOW DRIVE GAS SUPPLY alarm will activate both audibly and visually. Patient minute volume may be reduced due to lowered flow rates
14. Check that the cable between the control unit and remote display screen unit is connected before use. Always use a cable type recommended by the manufacturer.
Using the Ventilator
15. The AV-S ventilator is not intended for use in intensive care applications.
16. This apparatus must not be used with, or in close proximity to, flammable anesthetic agents. There is a possible fire or explosion hazard.
17. Anesthesia apparatus must be connected to an anesthetic gas scavenging system (AGSS) to dispose of waste gas and prevent possible health hazards to operating room staff. This requirement must be observed during test procedures as well as during use with a patient.
The scavenging transfer and receiver system must conform to ISO 8835-3.
22. An audible alarm indicates an anomalous condition and should never go unheeded.
23. The characteristics of the breathing circuit connected between the ventilator and the patient can modify or change patient ventilation. To assist the maintenance of the delivered patient tidal volume, the ventilator control system software includes: A) a compliance compensation algorithm, B) a fresh gas compensation algorithm.
However, patient ventilation must be monitored independently from the ventilator. It is the responsibility of the user to monitor patient ventilation.
24. Care must be taken to ensure that the flow sensors are connected correctly to the inspiratory and expiratory ports of the absorber.
WARNINGS AND CAUTIONS
4
25. The Vent Inop (ventilator inoperative) alarm indicates that one of the following conditions has occurred:
a) The drive gas solenoid has failed.
b) The flow control valve has failed.
c) Internal electronic fault.
d) Internal electrical fault.
e) Software error. Note that if a ventilator error is
detected, ‘Ventilator Inoperative’ will be displayed on the front control panel display.
26. The High and Low Airway Pressure Alarms are important for patient care. It is important that the sensor is properly located in the expiratory limb of the circuit - refer to section 5.1.10.
27. The patient must be continuously
attended and monitored when Advanced Breathing Modes are in use.
User Maintenance
28. User maintenance is restricted to cleaning the outside surfaces of the ventilator, see section 6. Other procedures detailed in this manual must be carried out by trained technicians.
Service and repair operations must only be carried out by an engineer trained by the manufacturer. The warranty for this product is void if the product is not maintained in accordance with the service schedule detailed in section 6.1, and the procedures published in the Service Manual for this product.
Control Unit
29. Opening the control unit by unauthorized personnel automatically voids all warranties and specifications.
Prevention of tampering with the control unit is exclusively the user’s responsibility. If the control unit seal is broken, the manufacturer assumes no liability for any malfunction or failure of the ventilator.
30. For continued protection against fire hazards, any replacement fuses must be the identical type and rating as the original components. Replacement must be carried out by trained technician. See section 4 for fuse rating.
31. If the internal battery is fully discharged, the ventilator will not function in the event of mains power failure. The battery must be recharged before the ventilator is used clinically, otherwise backup cannot be guaranteed. See Appendix for battery maintenance. See also CAUTION No. 7.
Used or defective batteries must be disposed of according to hospital, local, state, and federal regulations.
32. No oil, grease or other flammable lubricant or sealant must be used on any part of the ventilator in close proximity to medical gas distribution components. There is a risk of fire or explosion.
33. Exterior panels must not be removed by unauthorized personnel and the apparatus must not be operated with such panels missing. There is a possible electric shock hazard.
Bellows Assembly
34. The valve seat on the patient gas exhalation diaphragm valve in the base of the bellows assembly must be cleaned regularly. Note that the bellows assembly is built into the A200SP Absorber - please refer to User Manual for this product. Failure to keep the valve seat clean could result in the diaphragm sticking, thus preventing exhalation.
Great care must be taken not to damage the precision surface of the valve seat on the patient gas exhalation diaphragm valve in the base of the bellows assembly.
Never use any hard object or abrasive detergent to clean it; use only a soft cloth. If the valve seat is damaged, the valve will leak and may cause serious ventilator malfunction.
10. Circuit compliance is not activated until Fresh Gas Compensation is switched OFF.
NOTES
1. The term ‘cycle’ is used to designate the transition to the exhalation phase.
2. The term ‘trigger’ is used to indicate the transition to the inhalation phase.
WARNINGS AND CAUTIONS
5
CAUTIONS
1. Do not sterilize the ventilator control unit. The patient block assembly must be removed from the control unit before sterilization ( see section 6.2.5). All other internal components are not compatible with sterilization techniques and damage may result.
2. For ventilator components which require sterilization, peak sterilization temperatures should not exceed 134oC (275oF) to prevent possible damage. (See section 6).
3. Care must be taken not to let any liquid run into the control unit; serious damage may result.
4. The exhalation valve located in the bellows base assembly and the pediatric bellows adaptor must be cleaned and sterilized separately. Note that the bellows assembly is built into the A200SP Absorber - please refer to User Manual for this product.
5. Always check for correct fitment, and carry out a full function test before clinical use, if the bellows has been removed and refitted for any reason. Note that the bellows assembly is built into the A200SP Absorber - please refer to User Manual for this product.
6. Always check for correct fitment, and carry out a full function test before clinical use, if the bellows has been removed and refitted for any reason. See section 6.
7. Damage may occur to the battery if it is allowed to remain in a discharged state. Check the battery frequently if the ventilator is in storage (see Appendix 1).
8. Fresh gas compensation is disabled if :
a) The spirometry system is turned OFF through the menu system, or b) The spirometry system is not functioning correctly.
9. Fresh gas mixture compensation is disabled
if : a) The spirometry system is turned OFF through the menu system, or b) The spirometry system is not functioning correctly.
c) The O2 monitor is switched OFF.
6
WARNINGS AND CAUTIONS - Oxygen Monitor
Oxygen Monitor
Note that the sensor for the oxygen monitor is built into the A200SP Absorber - for additional information, please refer to the A200SP User Manual.
WARNINGS
1. We recommend calibration of the oxygen monitor every time the system is turned on, as a safety precaution.
2. Do not attempt to open the fuel cell. The sensor contains small quantities of : a) electrolyte, classified as a harmful irritant which is potentially hazardous, and b) lead.
Used or defective cells must be disposed of according to hospital, local, state, and federal regulations.
CAUTIONS
1. Do not sterilize any oxygen monitor component.
2. Do not autoclave or expose the sensor to high temperatures.
3. If the sensor shows signs of being affected by condensation, dry the sensor with soft tissue. Do not use heat to dry the sensor.
NOTES
1. The O2 SENSOR FAULT alarm indicates that one of the following conditions has occurred. a) Internal electrical fault b) Software/electronics fault c) Oxygen sensor fault.
2. The concentration read-out may, in certain conditions of excess pressure, show a value above 100%. To accommodate these conditions it is possible to set the high alarm value up to 105% (see section 5).
3. ALWAYS check the integrity of the sensor assembly before use.
4. Once exhausted, the sensor must be disposed of according to hospital, local, state and federal regulations.
5. The sensor measures oxygen partial pressure, and its output will rise and fall due to pressure change. An increase in pressure of 10% at the sensor inlet will produce a 10% increase in sensor output.
6. The oxygen sensor is not suitable for sterilization. If contamination is suspected, fit a new sensor (see section 6.4) and dispose of the contaminated unit according to hospital, local, state and federal regulations.
3. To maintain maximum sensor life: i) always switch off the anesthetic machine after use, to ensure that the basal flow ceases. ii) disconnect the breathing circuit after use.
4. The accuracy of flow and volume measurements may be reduced if the oxygen monitor is not in use.
5. Fresh gas mixture compensation is disabled
if the oxygen monitor is switched OFF.
2. PURPOSE
7
The AV-S Ventilator is a software controlled, multi-mode ventilator, designed for mechanical ventilation of adult and pediatric patients under general anesthesia. In addition, in spontaneous mode, it can be used to monitor spontaneously breathing patients It is designed for use in closed-circuit anesthesia.
Indications for use of the device:
The AV-S Ventilator is intended to provide continuous mechanical ventilatory support during anesthesia. The ventilator is a restricted medical device intended for use by qualified trained personnel under the direction of a physician. Specifically the ventilator is applicable for adult and pediatric patients. The ventilator is intended for use by health care providers, i.e. Physicians, Nurses and Technicians with patients during general anesthesia. The AV-S ventilator is not intended for use in intensive care applications.
Oxygen Monitor
The Oxygen Monitor is intended to continuously measure and display the concentration of oxygen in breathing gas mixtures used in anesthesia, and is intended for adult and pediatric patients. The oxygen monitor is an integral part of the ventilator. The oxygen monitor is intended for use by health care providers, i.e. Physicians, Nurses and Technicians for use with patients during general anesthesia.
3. DESCRIPTION
8
1
2
Spontaneous Mode Patient Support
SIMV - Synchronised Intermittent Mandatory Ventilation
SMMV - Synchronised Mandatory Minute Ventilation
PSV - Pressure Supported Ventilation
PEEP - Positive End Expiratory Pressure
3.1 General Description
The AV-S Ventilator is a pneumatically driven, software controlled, multi-mode ventilator.
The ventilator is a time-cycled, volume/pressure controlled, and pressure limited.
The ventilator has compliance compensation and a user selectable option of an inspiratory pause fixed at 25% of the inspiratory time. In addition, fresh gas compensation and user selectable gas mixture compensation is a standard feature.
Ventilation Modes
Volume Mode - continuous mandatory ventilation Pressure Mode - pressure controlled ventilation Spontaneous, with advanced patient support -
SIMV, SMMV, PSV, PEEP
Patient Monitoring
Airway pressure, measured from the expiratory limb of the breathing circuit.
Tidal Volume and Minute Volume measurement is provided by a dual spirometry system
An integral oxygen monitor system measures oxygen concentration in the breathing circuit inspiratory limb.
The print function provides a permanent record of function activity for up to eight hours during a procedure, or can be used to record waveforms.
Screen
210 mm (8.4 inch) high definition, colour TFT screen, with single/dual waveform display. Mounting: Remote, arm-mounted as illustrated (1) or optional combined control unit / screen (see section 5.1.1).
Bellows unit
The bellows unit (2) is built into the A200SP absorber. A pediatric bellows assembly is available as an option
Drive gas supply
The drive gas supply can be oxygen or air. The supply must be at 310 to 689 kPa (45 to 100 psi). Note that the drive gas is specified by the customer, and set during manufacture. Conversion from one drive gas to another must only be carried out by an
authorized service engineer trained by the manufacturer.
9
DESCRIPTION
2
7
3
13 14
15
12
1110
9
8
6
5
4
1
Control Unit Rear Panel
Gas Connections
1. Ventilator drive gas inlet
- connect to anesthetic machine
auxiliary gas outlet
2. Bellows Drive Gas Output
- connect to bellows via A200SP
absorber - see section 5.1.5
3. Outlet - Exhaust Valve
- connect to scavenge system - see section 5.1.6
Electrical Connection
4. Electrical mains input and fuse unit
Interface and Parameter inputs
5. A200SP Absorber Bag/Vent switch interface, and Spirometer connector
6. Integra AV-S Interface connector - (primary on/off switch)
7. Pressure Monitor Port
8. Input socket - Oxygen monitor sensor
Data and Printer Ports
9. Data Output
10. Output to remote display
11. Ethernet
12. USB
13. VGA
14. Printer port
15. RS232 (manufacturer’s use only)
NOTE USB port is for access only by engineers trained by the manufacturer. All other data ports are read only. For further information, please contact your distributor’s service department, or the manufacturer.
3.2 Ventilation Cycle
This section provides a simplified description of the ventilation cycle.
1. Inspiratory Phase
The drive gas proportional valve (1) in the control unit opens. Drive gas is delivered to the bellows housing (2). The patient proportional valve (3) opens, and gas flows through the bleed valve. The back pressure ensures that the exhaust valve (4) is kept closed. Drive gas pressure builds up above the bellows (5), which starts to move down. The diaphragm (6) in the bellows assembly base is held closed, and patient gas is forced out of the bellows base (7) into the breathing system.
2. Beginning of Expiratory Phase
The drive gas proportional valve (1) closes. The patient proportional valve (3) closes. The exhaust valve (4) opens. Patient gas returns to the bellows (5). As the bellows rises, redundant drive gas is pushed out through the exhaust valve.
DESCRIPTION
10
1
4
4
6
7
5
5
2
3
3
1
3
1
4
DESCRIPTION
3. End of Expiratory Phase
With the bellows at the top of its housing fresh gas continues to flow. To prevent a high pressure build up the exhalation diaphragm (6) lifts and allows gas to exit through the exhaust valve (4).
4. PEEP Positive End Expiratory Pressure (user selectable)
The patient proportional valve (3) applies PEEP pressure plus 20 cmH2O to the exhaust valve, which remains closed at this stage. As fresh gas flows in the patient circuit, any pressure increase above PEEP pressure in the bellows (5) will cause gas to bleed past the exhaust valve (4). If there is a fall in pressure in the breathing circuit, the continuous flow from the drive gas proportional valve (1) helps maintain the set PEEP pressure.
11
6
4
5
5
12
DESCRIPTION
A
Pneumatic Flow
Diagram
C
1817
5
8
14
9
12
6
16
13
15
11
7
1
2
10
4
3
B
0 - 80 cmH2O
100 cmH
2O
0 - 90 cmH2O
241 kPa (35 psi)
3 to 7 bar
3.3 Pneumatic System
3.3.1 System Operation
Refer to the pneumatic system diagram on the previous page.
A) Gas inlet manifold block
The AV-S Ventilator is designed to operate on a 310 - 689 kPa (45 -100 psi) drive gas supply (oxygen or air - to customer’s requirement).
1. DISS Connector
The gas source is connected to the DRIVE GAS SUPPLY fitting on the rear of the ventilator control unit. The gas supply should be capable of a flow rate of 80 L/min while maintaining a minimum pressure in excess of 310 kPa (45 psi).
2. Filter The drive gas is filtered with a 40-micron Input Gas Filter which protects the pneumatic components from incoming particulate matter.
3. The Low Supply Pressure Detector The pressure switch is set at a predetermined level to detect a loss or reduction of the input gas source pressure. When the pressure falls below 235 kPa (35 psi ± 1 psi), the LOW SUPPLY PRESSURE indicator will be displayed and the high priority audible alarm will activate.
4. Input Pressure Regulator Regulates the input drive gas to 260 kPa ± 21 kPa (38 psi ± 3 psi).
5. Cut-off Valve The valve isolates the the gas supply : a) when the ventilator is switched off b) when a fault condition occurs.
6. Airway Pressure Sensor Connected to expiratory limb of breathing circuit.
B) Pneumatic Control Manifold Block
7. Drive Gas Proportional Valve
8. Drive Gas Flow Sensor
9. Drive Gas Pressure Sensor
10. Low Pressure Regulator
11. Patient Proportional Valve
12. PEEP pressure sensor
13. Restrictor The restrictor allows a flow of up to 2 L/min (<2 L/min bleeding)
C) Exhaust Manifold Block
14. Check Valve
15. Diaphragm Valve
16. Pressure Relief valve - Set to 100 cmH
2O
17. Exhaust Port ( to AGSS)
18. Bellows drive gas outlet (to bellows assembly)
DESCRIPTION
13
Loading...
+ 61 hidden pages