Thank you for choosing our electronic balance.
Outline
Thank you for purchasing YP series electronic balance. The YP models are high performance electronic balances that we confidently recommend based on 20 years of precision balance manufacture. While there models are of course capable of fast and accurate weighing, the YP models all work on high precision strain gauge load cell which implements high speed stabilization and high reliability. The YP series balances also feature operation keys, improving operating convenience and making the balances easier to use. These balances also feature a variety of other functions that make it more convenient for customers to use for their own applications, including the Window Direct communication function, which enables measuring results to be transferred to a PC without installing any software. To ensure that you can make full use of the performance and function of your YP series balances, read this instruction manual carefully and use the balance correctly in accordance with the directions in the manual. When you have finished reading the manual, keep it in a safe place together with the balance so that you can refer to it at any time.
2
YP series precision balance work on high precision strain gauge load
Spec
ifica
tions,
cell which implements high speed stabilization and high reliability.
Model No. (YP) 1000 2000 3000 4000 5000 Max. Capacity (g) 1000 2000 3000 4000 5000 Tare Range (g) 0-1000 0-2000 0-3000 0-4000 0-5000 Readability PanSize(mm) Dimensions (LxWxH) Power supply AC 110-120V/220-240V
Model No. (YP) 10001 20001 30001 40001 50001 Max. Capacity (g) 1000 2000 3000 4000 5000 Tare Range (g) 0-1000 0-2000 0-3000 0-4000 0-5000 Readability PanSize(mm) Dimensions (LxWxH) Power supply AC 110-120V/220-240V
1g
143*192mm
280x180x80mm
0.1
143*192mm
280x180x80mm
Model No. (YP) 2002 3002 5002 6002 10002 20002 Max. Capacity (g) 200 300 500 600 1000 2000 Tare Range (g) 0-200 0-300 0-500 0-600 0-1000 0-2000 Readability PanSize(mm) Dimensions (LxWxH) Power supply AC 1 10-120V/220-240V
0.01
133*182mm
280x180x80mm
Operating Temperature Range: 5-35 Degree Centigrade Moisture: 50-85%
3
1 BEFOREWEIGHING
Main body
Pan
Placethe objecttobe weighed here
Displaypanel
Shows the weighing results, information for making function settings, the current functionsetting
Name and Function of the Components
Numericvaluearea
Level screws
Adjust to level the balance
Operation keys
Used to open and close the balance, tare the balance, perform calibration and print.
Unit display area
Operation keys
4
Operation Keys
During Weighing
Key
ON/OFF Turn on or turn off the balance
COU Select the counting function
Press once and Release
Press and hold for a while
TAR/CAL
UNIT Select the unit
Choosing the Installation Site
The measuring performance of the balance is greatly influenced by the environment where it is installed. Observe the following points to ensure safe and accurate weighing.
Tare the balance(setting it to zero)
Performs calibration
Caution
Do not use the balance anywhere exposed to explosive, combustible or corrosive gases.
This could cause fire or trouble.
5
Use the correct power supply and voltage with the balance.
Using an incorrect power supply or voltage with the balance will lead to fire or trouble with the balance. Note also that if the power supply or voltage is unstable or if the power supply capacity is insufficient, it will not be possible to obtain satisfactory performance from the balance.
Precautions on Use
Avoidlocations where thebalance will be exposed to any of the following.
¤Air flow from an air conditioner, ventilator, door or window
¤Extreme temperature changes
¤Vibration from surroundings or nearby equipment
6
¤Direct sunlight
¤Dust, electromagnetic waves or a magnetic field
Install the balance on a strong and stable flat table or floor.
Placing the balance in an unstable site could lead to injury or trouble with the balance. When selecting the installation site, take into account the combined weight of the balance and the item to be weighed.
Unpacking and Delivery Inspection
Check that all of the items indicated below are included in the package, and that nothing has been damaged. [1] Main Body [2] Pan [3] Introduction Manual [4]Adapter
7
Turning the Power On
1 Insert the plug of the Power Cord into the DC IN connector on the back of the balance. 2 Connect the Power Cord to the power outlet. 3 Press “ON”. The display will automatically go through the changes indicated below.
Warming Up
Before performing span calibration on the balance or measuring its accuracy, you must ensure that is it in a stable state. When stabilizing the balance, it is important that its temperature is stable. Putting the balance in weighing mode and leave it with the power on for at least half an hour in advance of calibration. This is called “warming up”.
Performing Span Calibration
Always perform span calibration for a balance after moving it. Weights are required for external calibration balances. Before performing span calibration, warm up the balance in advance.
8
Also, carry out the adjustment at a location where there are few people moving around and there i s no air flow or vibration.
1 Press “CAL”
The weight value will flash. The calibration record is being output. When the output has finished, span calibration will start automatically
2 Place the calibration weight on the pan.
Wait until the flashing weight value display changes to a steady 500.00.
3 T ake the weight off the pan.
“0.00” will be displayed and the balance will return to the weighing mode
9
USING THE BALANCE 1 Enter the weighing mode
2 Place a container on the pan. 3 Once the display has stabilized, press “zero/tar”.
The indication changes to zero.
4 Insert the sample into the container. 5 When the display has stabilized, (the stability mark) lights up, read the
display.
Selecting the Display Switching Units
You can display different units from among those set to be available. 1 press “UNIT” in the weighing mode. Repeatedly pressing this key will cycle you through the registered units.
10